Meter Testing

Embed Size (px)

Citation preview

  • 8/10/2019 Meter Testing

    1/1

    Technical manual 6-10 Testing

    Risk of personal injury or death!Use only authorized personnel and proper test procedures to test meteringequipment. Dangerous voltages are present. Personal injury, death, orequipment damage can result if safety precautions are not followed.

    To set up the A1800 ALPHA meter for testing:

    1. Install the meter in the stable mounting fixture.

    2. Place the precision Wh or varh reference standard and precision voltage and currenttransformers (as required) in series with the meter being tested.

    If voltage transformers are not required, then the meter source voltage should beplaced in parallel with the Wh or VARh reference standards. See Appendix D, Wiringdiagrams, for appropriate wiring diagrams for the A1800 ALPHA meter.

    3. Connect the control equipment for switching the source voltage to the precisionreference standard.

    4. Connect the measuring equipment for counting the standards output pulses.5. Apply the rated test current and voltage to the terminals of the meter.

    Meter testing

    Since no adjustments are required for the A1800 ALPHA meter in the field, meter testing isdone primarily to verify that the meter is operating within its specifications. Typically, meterspecifications are verified by checking the meter calibration. The accuracy of theA1800 ALPHA meter remains consistent over a wide range of ambient temperatures.Nevertheless, for precise test results, meters should be tested in an environment where themeter and test equipment are at the same ambient temperature, ideally 22 C (72 F).

    The test voltage should be applied to the meter for at least ten seconds prior to making testmeasurements. This allows the power supply circuitry to stabilize. When testing meters for

    Class 0.2 accuracy, a test cycle time of at least one minute is recommended at Inand withPF = 1.0. When using current values lower than Infor testing, test errors may occur becausenot enough time is allowed for the test. When using a lower test In, increase the test timeproportionally.

    The preferred test method is to apply full 3-phase voltage and current to both the meterand the precision reference standard. Nevertheless, if required, polyphase meters can betested with single phase loading. Single phase loading is done by connecting the voltageinputs in parallel and the current sensors in series to combine element operation. Theaccuracy test results for single phase and polyphase loading will be virtually identical andwell within A1800 ALPHA meter specifications.

    Using relay outputs for testing. The relay outputs can be used instead of the LED to testmeter calibration. To do so, the relay outputs need to be configured for pulse output. When

    using the relay outputs for testing, testing t ime should exceed 20 seconds for accurateresults at normal test current values of Ibor In. If more accurate testing is required, uselonger testing times. If precision testing is required, testing times should be as long as ittakes to attain a stable accuracy level when comparing the meter under test to theprecision standard. The testing time may vary because of the characteristics of theprecision reference standard and the amount of power flowing through the test circuits.Some experimentation may be required to determine the testing time needed to reach astable accuracy level.

    Using LCD pulse count for testing. For test shops that do not have photoelectric pulsesensors and related counters and do not want to use relay outputs for testing, the LCD canprovide a pulse count that reflects the energy measured during a test. Use the LCD pulsecount to determine the energy measured during the test cycle and compare it with theenergy delivered by precision reference standard.