Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data. Chapter 6 Howell, Murray, Pierre. Introduction. Maintenance is an important element of the engineering equipment life cycle. Failure data are the backbone of medical-related reliability studies. - PowerPoint PPT Presentation

Citation preview

Page 1: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

MEDICAL DEVICE MAINTENANCE AND SOURCES OF OBTAINING MEDICAL DEVICE- RELATED FAILURE DATA

Chapter 6Howell, Murray, Pierre

Page 2: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Introduction

Maintenance is an important element of the engineering equipment life cycle.

Failure data are the backbone of medical-related reliability studies.

Provides invaluable information to professional involved with the design, manufacture and maintenance of future medical devices/equipment.

This chapter presents various important aspects of medical device maintenance and sources for obtaining medical device-related failure data.

Page 3: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Medical Equipment Classifications

Imaging and radiation therapy Life Support and therapeutic Patient diagnostics Patient environmental, transport Laboratory apparatus Miscellaneous

Page 4: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Medical Equipment Maintenance Indexes

Used to measure different aspects of the maintenance function

Indexes (6) Broad Indexes (3) Health-Care-Specific Indexes (3)

Page 5: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index I

where

IX1 ≡ index parameter

Cm ≡ total maintenance cost

TI ≡ total investment in plant and equipment

TI

CIX m1

Relates the total maintenance cost to the total investment in plant and equipment:

Page 6: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index II

where

IX2 ≡ index parameter

Cm ≡ total maintenance cost

TSR ≡ total sale revenue

Generally, average expenditure for the maintenance activity for all industry is around 5% of the sale revenue.

TSR

CIX m2

Deals with total sale revenue:

Page 7: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index III

where

IX3 ≡ index parameter

Cm ≡ total maintenance cost

TOT ≡ total output expressed in tons, megawatts, gallons, etc.

TOT

CIX m3

Relates the total maintenance cost to the total output by the organization:

Page 8: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index IV

where

IX4 ≡ index parameter or the number of repair requests accomplished per medical device/equipment

α ≡ total number of medical devices

TNRR ≡ total number of repair requests

TNRR

IX 4

Provides number of repair requests accomplished per medical device:

Page 9: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index V

where

IX5 ≡ index parameter or the average turnaround time per repair

θ ≡ number of work orders or repairs

TATa ≡ total turnaround time

aTAT

IX 5

Measures how much time elapses from a customer request until the malfunctioning medical device/equipment is repaired and put back into full service :

Page 10: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index VI

where

IX6 ≡ index parameter

Cp≡ procurement cost or the cost at the time of purchase of equipment

Cs ≡ service cost or the total of all parts, materials, and labor costs for scheduled and unscheduled service; also includes in-house, vendor, prepaid contracts, and maintenance insurance expenses.

p

S

C

CIX 6

Defined by:

Page 11: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Index VI (Cont’d)

Advantages: Easy to compare across equipment types Takes into consideration all types of service costs Can be used with incomplete dataDisadvantages: Allows no compensation for the age of the equipment Requires a standard definition for the pricing of in-house service Possesses no mechanism for adjusting wage rates by geographic

region

p

S

C

CIX 6

Defined by:

Page 12: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Medical Equipment Computerized Maintenance Management Systems (CMMS)

• Used by hospital clinical engineering departments for• Collecting• Storing• Analyzing• Reporting

…data on the repair and maintenance performed on medical devices/equipment.

• These data are used in areas such as:• Equipment Management• Work Order Control• Quality Improvement Activities• Reliability and Maintainability Studies• Cost Control

Page 13: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Computerized Maintenance Management Systems (CMMS)

Major steps in selecting a commercial CMMS:

1) Defining the problem scope for clinical engineering department

2) Evaluating the effectiveness of the existing system

3) Evaluating the commercially available CMMS

4) Evaluating the chosen one or more CMMS with respects to identified requirements.

5) Having discussions on potential performance issues with organizations of similar size and scope

6) Evaluating the vendor support services

7) Developing preliminary budget and cost estimates

1. Define the problem scope

2. Evaluate the current system in

use3. Conduct a

preliminary study of all commercially

available systems4. Conduct a

comprehensive study of the selected CMMS

5. Discuss all potential performance issues

concerning the above chosen CMMS

6. Carefully examine all CMMS support

issues

7. Examine CMMS cost

Page 14: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Mathematical Models for Medical Equipment Maintenance Model 1

Used to determine optimum time between item/equipment replacements.

Average annual cost composed of avg. maintenance costs, avg. operating cost and avg. investment cost.

Minimizes the equipment/item average cost with respects to its life.

EAAC= EOCf + EMCf +EIC + (t-1) t 2

(i + j)

Where:EOC is the equip/item operating cost for the 1st yearEMC is the equip/item maintenance cost for the 1st yearEIC is the equip/item investment costt is the equip/item life expressed in yearsj is the amount by which maintenance cost increases annuallyi is the amount by which operating costs increases annually

Page 15: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Model 1 continued:

By differentiating the equation with respects to t ten equating it to zero we obtain the following expression:

t* = ( )

Where

t is the equipment/item optimum-replacement interval

Example assume the following: j = $500 i = $4,000 EIC = $50,000

t* = ( )

= 4.71 years

2EICi + j

1/2

2(50,000)4000 + 500

1/2

Page 16: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Model 2

Used to predict the number of spares required for a medical equipment/device in useΘ = λt + z [ λt] ½

Where

Θ is the number of spares

t is the mission time

λ is the constant failure rate of the medical item under construction

Z is associated with the cumulative normal distribution function.

Example assume that the constant failure rate of a medical device is 0.0005 failures per hour. Determine the number of spare medical devices required during a 5,000-hour time period, if the confidence level for o stock out of the devices is 0.9032.From table: z= 1.3

Θ = (0.0005)(5,000) + (1.3) [(0.0005)(5,000)] ½

= 5 medical devices are needed

Page 17: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Medical Device-Related Failure Data Sources These are some sources for obtaining medical device-

related failure data:

Sources

Hospital Equipment

Control System

(HECS™)

Medical Device

Reporting System (MDRS)

Universal Medical Device

Registration and Regulatory Management

System (UMDRMS)

MIL-HDBK-

217NUREG/CR-1278

Page 18: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Hospital Equipment Control System (HECMS™) Provides effective support and analysis of clinical

engineering-related operations through detailed work scheduling of items.

Provides objective and good quality data on the basis of satisfactory statistical foundation.

Provides cost-effective, economic, financial, productivity data within the framework of each hospital.

Provides new information support, including data on the relative reliability of competing brands and models of clinical equipment.

Page 19: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

(MDRS)

The Medical Device Reporting System It is managed by the Center for Devices and Radiological

Health, Food and Drug Administration (FDA). It contains reports filed by device manufacturers concerning

patient deaths and serious injuries allegedly involving their manufactured items, as well as failure of such items that may have caused serious injuries or deaths.

Page 20: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

(UMDRMS)

The University Medical Device Registration and Regulatory Management System It was developed by ECRI (Emergency Care Research

Institute) and is designed to facilitate items such as tracking products for recall, safety, and reliability statistics and inventory control.

Page 21: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

MIL-HDBK-217

It was developed by the U.S Department of Defense It is widely used in the industrial sector to predict the

failure rate of a given piece of equipment. It contains generic failure rates for various parts,

particularly the electronic ones, and a large number of mathematical models to predict failure rates of various types of parts in their actual use environment.

This document can also be used to predict failure rates of medical equipment/devices or their parts.

Page 22: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

NUREG/CR-1278

It was developed by the U.S. Nuclear Regulatory Commission to perform human reliability analysis with emphasis on nuclear power plant applications.

The document contains failure data on various types of tasks performed by humans.

The information contained in the document could be quite useful analyzing medical devices in regard to the occurrence of human error.

Page 23: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Organizations for Obtaining Medical Device-Related Failure Data There are many organizations that can be quite useful,

directly or indirectly, in obtaining medical device-related failure data: Center for Devices and Radiological Health, Food and Drug

Administration, 1390 Piccard Drive, Rockville, Maryland 20857 Emergency Care Research Institute (ECRI), 5200 Butler

Parkway, Plymouth Meeting, Pennsylvania 19462 Parts Reliability Information Center (PRINCE), George C.

Marshall Space Flight Center, Huntsville, Alabama 35812 Reliability Analysis Center, Griffiss AFB, U.S Department of

commerce, 5285 Port Royal, Springfield, Virginia 22161

Page 24: Medical Device Maintenance and Sources of Obtaining Medical Device- Related Failure Data

Medical Device Failure-Related Data

A wide variety of failure-related data, directly or indirectly, concerned with medical equipment/devices is available in the published literature.

There are two types, directly or indirectly, of failure-related data: Medical device specific data are concerned with the

preproduction cause of a total of 1, 143 medical device recalls for the period 1983 to 1987. The causes of these recalls were grouped under 7 distinct categories: device and process(design and software), component, label, and package (design).

The general component-related failure data.