Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Supporting Information
A squaraine-based sensor for colorimetric detection of CO2
gas in aqueous medium through an unexpected recognition
mechanism
Jianqi Sun‡a,b, Xiangjuan Zheng‡a, Xinjie Wua, Dong Lia, Guomin Xiaa, Shuxian Yu a, Qiming Yu
and Hongming Wanga*
a College of Chemistry and Institute for Advanced Study, Nanchang University, Nanchang, Jiangxi 330031, China
b College of Chemistry and Envioronmental Engineering, Jiujiang University, Jiujiang, Jiangxi 332005, China
Detection Limit [1]. The detection limit was calculated on the basis of the UV-Vis
titration. The UV-Vis spectrum of SQM (7.5 μM) in MeCN-H2O (V : V = 90 : 10)
was measured 12 times, and the standard deviation of blank measurement was
achieved. To gain the slope, the absorbance at 627 nm versus amount of pure CO2 gas
was plotted. The detection limit was calculated using the following equation:
Detection limit = 3σ/k (1)
Where σ is the standard deviation of blank measurement, and k is the slope
between the absorbance versus the volume of pure CO2 gas (VCO2).
References
[1] S. Samanta, S. Goswami, M. N. Hoque, A. Rameshm and G. Das, Chem. Commun.,
2014, 50, 11833.
‡These authors contributed equally to this work.
Electronic Supplementary Material (ESI) for Analytical Methods.This journal is © The Royal Society of Chemistry 2017
Captions:
Fig. S1 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with 50-fold weak base EA, DEA, TEA and Py in MeCN respectively.
Fig. S2 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with DBU in MeCN. Inset: The corresponding plots of absorbance at
indicated wavelengths versus additional DBU.
Fig. S3 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with TMG in MeCN. Inset: The corresponding plots of absorbance at
indicated wavelengths versus additional TMG.
Fig. S4 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with [NBu4]F in MeCN. Inset: The corresponding plots of absorbance at
indicated wavelengths versus additional [NBu4]F.
Fig. S5 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with (a) 2-fold TBD; (b) 5-fold DBU in MeCN then bubbled with different
volumes of CO2 gas in a sealed cuvette.
Fig. S6 (a) UV-Vis spectral changes obtained during the course of titrating sensor
SQM (7.5 μM) with DBU in MeCN-H2O (V : V = 90 : 10); (b) then bubbled with
different volumes of CO2 gas in a sealed cuvette. Inset: The corresponding plots of
absorbance at indicated wavelengths versus additional DBU or volume of CO2 gas.
Fig. S7 UV-Vis spectrum of precursor SQ (7.5 μM) in MeCN-H2O (V : V = 90 : 10).
Fig. S8 The absorbance of SQM (7.5 μM) at 627 nm in MeCN-H2O with different
volume fractions of H2O after addition of 20 eq TBD (the black bar) followed by
bubbling 5 mL CO2 gas (the green bar) in a sealed cuvette.
Fig. S9 The absorbance of SQM (7.5 μM) at 627 nm with the addition of TBD (20 eq)
versus the volume of pure CO2 gas in MeCN-H2O (V : V = 90 : 10).
Fig. S10 The absorbance of SQM (7.5 μM) at 627 nm with the addition of DBU (20
eq) versus the volume of pure CO2 gas in MeCN-H2O (V : V = 90 : 10).
Fig. S11 The mass spectrometry analysis of SQM in MeCN with addition of TBD in
positive mode.
Fig. S12 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of TBD in positive mode.
Fig. S13 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of TBD followed by CO2 gas in positive mode.
Fig. S14 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of NaOH in positive mode.
Fig. S15 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of NaOH followed by CO2 gas in positive mode.
Fig. S16 The absorbance of SQM (7.5 μM) at 627 nm with the addition of NaOH
(300 eq) versus the volume of pure CO2 gas in MeCN-H2O (V : V = 90 : 10).
Fig. S17 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of TBD followed by CO2 gas in negative mode.
Fig. S18 Partial 1H-NMR spectra obtained during the course of titrating sensor SQM
(5.0 mM) with NaOH followed by CO2 in DMSO-d6.
Fig. S19 1H NMR spectrum of SQM
Fig. S20 13C NMR spectrum of SQM
Fig. S21 The mass spectrometry analysis of SQM in MeCN in positive mode.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0Ab
sorb
ance
Wavelength/nm
SQM SQM+50eq EA SQM+50eq DEA SQM+50eq TEA SQM+50eq Py
Fig. S1 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with 50-fold weak base EA, DEA, TEA and Py in MeCN respectively.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0Ab
sorb
ance
Wavelength/nm
SQM Current SQM+5eq DBU
0 1 2 3 4 50.0
0.5
1.0
1.5
2.0
2.5
3.0
At 627nm
Abso
rban
ce
Equiv.DBU
Fig. S2 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with DBU in MeCN. Inset: The corresponding plots of absorbance at
indicated wavelengths versus additional DBU.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0Ab
sorb
ance
Wavelength/nm
SQM Current SQM+20eq TMG
0 5 10 15 200.0
0.5
1.0
1.5
2.0
2.5
3.0
At 627Ab
sorb
ance
Equiv.TMG
Fig. S3 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with TMG in MeCN. Inset: The corresponding plots of absorbance at
indicated wavelengths versus additional TMG.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0Ab
sorb
ance
Wavelength/nm
SQM Current SQM+50eq F-
0 10 20 30 40 500.0
0.5
1.0
1.5
2.0
2.5
3.0
At 627nmAb
sorb
ance
Equiv.[NBu4]F
Fig. S4 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with [NBu4]F in MeCN. Inset: The corresponding plots of absorbance at
indicated wavelengths versus additional [NBu4]F.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0Ab
sorb
ance
Wavelength/nm
SQM SQM+2eq TBD Current SQM+2eq TBD+5ml CO2
(a)
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0
Abso
rban
ce
Wavelength/nm
SQM SQM+5eq DBU Current SQM+5eq DBU+5ml CO2
(b)
Fig. S5 UV-Vis spectral changes obtained during the course of titrating sensor SQM
(7.5 μM) with (a) 2-fold TBD; (b) 5-fold DBU in MeCN then bubbled with different
volumes of CO2 gas in a sealed cuvette.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0Ab
sorb
ance
Wavelength/nm
SQM Current SQM+50eq DBU
0 10 20 30 40 500.0
0.5
1.0
1.5
2.0
2.5
3.0
At 627nmAb
sorb
ance
Equiv.DBU
(a)
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
3.0
Abso
rban
ce
Wavelength/nm
SQM SQM+50eq DBU Current SQM+50eq DBU+10ml CO2
0 1 2 3 4 5 6 7 8 9 100.0
0.5
1.0
1.5
2.0
2.5
3.0
At 627nm
Abso
rban
ce
VCO2(ml)
(b)
Fig. S6 (a) UV-Vis spectral changes obtained during the course of titrating sensor
SQM (7.5 μM) with DBU in MeCN-H2O (V : V = 90 : 10); (b) then bubbled with
different volumes of CO2 gas in a sealed cuvette. Inset: The corresponding plots of
absorbance at indicated wavelengths versus additional DBU or volume of CO2 gas.
300 400 500 600 7000.0
0.5
1.0
1.5
2.0
2.5
Abso
rban
ce
Wavelength/nm
Fig. S7 UV-Vis spectrum of precursor SQ (7.5 μM) in MeCN-H2O (V : V = 90 : 10).
0 10 20 30 40 500.0
0.5
1.0
1.5
2.0
2.5
Abso
rban
ce
Water Volume Fraction (%) Fig. S8 The absorbance of SQM (7.5 μM) at 627 nm in MeCN-H2O with different volume fractions of H2O after addition of 20 eq TBD (the black bar) followed by bubbling 5 mL CO2 gas (the green bar) in a sealed cuvette.
0.00 0.05 0.10 0.15 0.200.4
0.5
0.6
0.7
0.8
0.9Ab
sorb
ance
(a.u
.)
VCO2 (ml)
At 627 nm Y=0.44747+2.12145X,
R=0.99738
Fig. S9 The absorbance of SQM (7.5 μM) at 627 nm with the addition of DBT (20 eq)
versus the volume of pure CO2 gas in MeCN-H2O (V : V = 90 : 10).
The detection limit in MeCN-H2O (V : V = 90 : 10) (3.0 ml) was calculated to be
about 1.59*10-6 M (ca. 39.0 ppm, 1atm, 25°C)
0.0 0.1 0.2 0.3
0.2
0.4
0.6
0.8
VCO2 (ml)
Abso
rban
ce (a
.u.)
At 627 nm Y=0.26298 + 1.68529X,
R=0.99447
Fig. S10 The absorbance of SQM (7.5 μM) at 627 nm with the addition of DBU (20
eq) versus the volume of pure CO2 gas in MeCN-H2O (V : V = 90 : 10).
The detection limit in MeCN-H2O (V : V = 90 : 10) (3.0 ml) was calculated to be
about 2.0*10-6 M (ca. 49.1 ppm, 1atm, 25°C)
SQ-OCH3+TBD #23-76 RT: 0.09-0.24 AV: 54 NL: 4.69E3T: ITMS + c ESI Full ms [100.00-1000.00]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e A
bund
ance
749.09610.24
594.31140.06750.11
592.27
611.24
338.34
439.33 578.28
717.14 751.12612.26210.20 815.51564.44 683.55 859.62339.36 810.32 903.57624.29440.38 854.27 947.48498.45 546.41 772.66 860.65 976.00652.43141.10 539.39315.07242.35 369.19 457.39167.02 534.50424.38281.14139.13
Fig. S11 The mass spectrometry analysis of SQM with addition of TBD in MeCN in
positive mode.
SQ-OCH3+TBD+CH3CN-H2O #187-266 RT: 0.57-0.80 AV: 80 NL: 2.06E4T: ITMS + c ESI Full ms [100.00-1000.00]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e A
bund
ance
140.03
564.40
565.39428.88
141.08439.30
515.89 749.01566.41376.98210.21 771.62440.34 859.55739.20341.31 605.10 815.56 903.55652.34 683.53516.96 947.55 991.79378.06315.04142.07 211.23 498.57112.04 278.86
Fig. S12 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of TBD in positive mode.
SQ-OCH3+TBD+CO2+CH3CN+H2O #75-180 RT: 0.25-0.55 AV: 106 NL: 2.91E4T: ITMS + c ESI Full ms [100.00-1000.00]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100R
elat
ive
Abu
ndan
ce140.02
564.38
428.87565.37
141.07749.03210.20 515.86376.98 750.06566.39429.92341.28 605.11 739.18315.02 687.83516.92 887.83 976.03865.25211.21 915.96378.03 768.07675.02439.25278.81142.09112.04 498.41
Fig. S13 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of TBD followed by CO2 gas in positive mode.
SO-OCH3+NaOH+CH3CN-H2O #33-231 RT: 0.12-0.66 AV: 199 NL: 6.49E3T: ITMS + c ESI Full ms [100.00-1000.00]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e A
bund
ance
871.03
872.06
140.07
727.60683.58 771.58 815.59
639.56
859.60 903.48
595.55369.14
986.69942.99793.34855.06
728.65 985.56551.52 705.41904.67 947.68640.65
447.28926.64 948.71661.42507.53 596.61
969.21428.90 552.57 617.38370.20 479.17141.09 508.58 586.21313.19 362.69 419.44202.20 294.74226.98180.15138.18
Fig. S14 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of NaOH in positive mode.
SQ-OCH3+NaOH+CO2+CH3CN+H2O #50-225 RT: 0.17-0.65 AV: 176 NL: 1.56E4T: ITMS + c ESI Full ms [100.00-1000.00]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100R
elat
ive
Abu
ndan
ce871.02
872.03
873.02140.06 815.60771.59727.59683.57 859.58447.31
639.58 903.55369.15 816.63595.55 772.61728.65425.27 947.60684.62551.55452.37 640.62 735.69 991.65596.61 930.54817.64424.39 507.58293.36 487.47253.30 657.72 685.65141.10 332.76180.11128.94
Fig. S15 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of NaOH followed by bubbling CO2 gas in positive mode.
0.00 0.05 0.10 0.15 0.20 0.25
0.2
0.4
0.6
0.8
1.0
At 627 nm Y=0.25472 + 2.81449X
R=0.99871
VCO2(ml)
Abso
rban
ce (a
.u.)
Fig. S16 The absorbance of SQM (7.5 μM) at 627 nm with the addition of NaOH
(300 eq) versus the volume of pure CO2 gas in MeCN-H2O (V : V = 90 : 10).
The detection limit in MeCN-H2O (V : V = 90 : 10) (3.0 ml) was calculated to be
about 1.20*10-6 M (ca. 29.4 ppm, 1atm, 25°C)
SQ-OCH3+TBD+CH3CN+H2O- #16 RT: 0.09 AV: 1 NL: 1.23E1T: ITMS - c ESI Full ms [100.00-1000.00]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rel
ativ
e A
bund
ance
809.17
863.04690.82598.37864.55
766.11 893.10
943.79
778.45499.15 946.33
695.97
902.22844.23637.16
610.36 663.53310.83 913.75591.46 955.55
710.72
524.07 558.26 997.10
417.40
985.41
Fig. S17 The mass spectrometry analysis of SQM in MeCN-H2O (V : V = 90 : 10)
with addition of TBD followed by CO2 gas in negative mode.
N O
O Na
b
cd
e
N O
ON
a′b′
c′d′
e′OH-
e′′
ae
bd
d′′
b′′a′
OH
Fig. S18 Partial 1H-NMR spectra obtained during the course of titrating sensor SQM
(5.0 mM) with NaOH followed by CO2 in DMSO-d6.
Fig. S19 1H NMR spectrum of SQM
N O
O N
SQM
Fig. 20 13C NMR spectrum of SQM
N O
O N
SQM
CF3SO3
SQ-OCH3 #85-161 RT: 0.25-0.44 AV: 77 NL: 1.27E5T: ITMS + c ESI Full ms [200.00-1000.00]
200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100R
elat
ive
Abu
ndan
ce
439.31
440.26
441.30727.55 771.56683.54 815.57639.53595.52 859.56551.51 903.56507.50 947.56463.50424.40 741.39 785.40 991.60697.44641.58252.21 369.26339.38222.22 277.30
Fig. 21 The mass spectrometry analysis of SQM in MeCN in positive mode.