4
Measuring Hydrocarbon Oil Index according to ISO 9377-2 (DIN H53) Introduction The Hydrocarbon Oil Index (HOI) is defined as the total amount of compounds which can be extracted from the sample (potable water, surface water and waste water) with a non-polar solvent having a boiling point between 39°C and 69°C (replacing the use of halogenated solvents). In addition, the compounds must not absorb on Florisil and must elute between n-decane (C 10 H 24 ) and n-tetracontane (C 40 H 82 ) when analysed by GC using an apolar analytical column. Restrictions on the use of halogenated hydrocarbons as solvents for analytical applications combined with the increasing cost of and reduced availability of technical grade halogenated solvents as well as documented environmental Abstract Agilent Technologies developed this ana- lytical method to comply with ISO 9377-2 and DIN H53 Standards. The method employs a simple instrument configuration consisting of a Gas Chromatograph (GC) equipped with a Split/Splitless (S/SL) inlet and Flame Ionisation Detector (FID), making the method both easy to implement and robust to operate. The combination of an Agilent GC system (see configuration above), selected capillary GC column, proprietary injection port liner, appropriate chemical standards as well as method installation and support by an Agilent Applications Specialist, provide a complete solution for the measurement of the Hydrocarbon Oil Index. Gas Chromatography Bernhard Wüst Environmental Application considerations resulted in the change to a non-polar solvent for HOI determinations. This analytical method complies with the ISO Standard 9377-2 for the determination of the Hydrocarbon Oil Index by Gas Chromatography. The method is suitable for HOI determinations in concentrations above 0.1 mg/L in drinking waters, surface waters, waste waters and waters from sewage treatment plants.

Measuring Hyd Oil Index

Embed Size (px)

DESCRIPTION

Measuring

Citation preview

  • Measuring Hydrocarbon Oil Index according to ISO 9377-2 (DIN H53)

    Introduction

    The Hydrocarbon Oil Index (HOI) is defined

    as the total amount of compounds which

    can be extracted from the sample (potable

    water, surface water and waste water)

    with a non-polar solvent having a boiling

    point between 39C and 69C (replacing

    the use of halogenated solvents). In

    addition, the compounds must not absorb on

    Florisil and must elute between n-decane

    (C10H24) and n-tetracontane (C40H82)

    when analysed by GC using an apolar

    analytical column. Restrictions on the use of

    halogenated hydrocarbons as solvents for

    analytical applications combined with the

    increasing cost of and reduced availability of

    technical grade halogenated solvents as well

    as documented environmental

    Abstract

    Agilent Technologies developed this ana-

    lytical method to comply with ISO 9377-2

    and DIN H53 Standards. The method

    employs a simple instrument configuration

    consisting of a Gas Chromatograph (GC)

    equipped with a Split/Splitless (S/SL) inlet

    and Flame Ionisation Detector (FID), making

    the method both easy to implement and

    robust to operate. The combination of an

    Agilent GC system (see configuration

    above), selected capillary GC column,

    proprietary injection port liner, appropriate

    chemical standards as well as method

    installation and support by an Agilent

    Applications Specialist, provide a complete

    solution for the measurement of the

    Hydrocarbon Oil Index.

    Gas ChromatographyBernhard Wst

    Environmental Application

    considerations resulted in the change to a

    non-polar solvent for HOI determinations.

    This analytical method complies with the

    ISO Standard 9377-2 for the determination

    of the Hydrocarbon Oil Index by Gas

    Chromatography. The method is suitable for

    HOI determinations in concentrations above

    0.1 mg/L in drinking waters, surface waters,

    waste waters and waters from sewage

    treatment plants.

  • ExperimentalThe Agilent 6890 and Agilent 6850 GasChromatographs are well suited to the HOIapplication. Both instruments use the sameSplit/Splitless inlet system and identicalFlame Ionisation Detector. Discriminationeffects on higher boiling compounds are themost critical aspect of the analysis and areprevented by the use of a special injectionport liner which, in combination with theAgilent S/SL inlet system, produces robustand reliable data. Additional benefits suchas reduced maintenance time and fasteranalytical cycle times are also available byusing a S/SL inlet as compared to a Cool-on-Column inlet.The Agilent GC ChemStation standardsoftware is able to subtract backgroundsignals to produce a chromatogram withany column bleed removed. In addition, boththe Agilent 6890 and 6850 are able to savecolumn compensation signals in memoryand subsequently subtract the columnbackground from the sample data andprovide a background-subtracted signal tothe ChemStation or data integrationsystem. Standard ChemStation softwaretools are used to control the GC andperform the data analysis required of thesample data. A custom report template isincluded with the HOI method andgenerates a report conforming to the ISOStandard (see figure 3).

    Publication number 5988-0621EN Agilent Technologies Page 2

    min5 10 15 20 25 30

    pA

    10

    20

    30

    40

    50

    60

    70

    80

    90

    FID1 A, (HYC2509\007F0702.D)

    2.0

    45 2

    .158

    2.2

    50 2

    .513

    2.6

    01 2

    .783

    3.0

    11 3

    .128 3

    .377

    3.4

    94 3.

    647

    4.1

    03 4

    .207

    13.

    331

    22.

    535

    22.

    641

    Instrument Parameters

    Oven temperature Isothermal 35C (1.5 min), ramp 5C/min to 60C hold 0 min, profile ramp 15C/min to 350C hold 5 min.Inlet (S/SL) 375C Splitless ModeDetector (FID) 375CAnalytical Column Agilent Part Number 19095Z-221E (HP1, 15m x 0.53mm x 0,15 m) Inlet Liner Agilent Part Number 5183-4647Carrier Gas HeliumCarrier Gas 7.4 ml/min Constant Flow ModeFlow RateInjection Mode Automated (Agilent 7683) Fast InjectionInjection Volume 1.0 l

    Figure 2: 0.6 mg/ml Oil Standard

    Figure 1: Oil Standards Calibration Curve

    Calibration Curve

    Amount[mg/ml]0 0.5 1 1.5

    Area

    0

    5000

    10000

    15000

    20000

    25000

    30000

    35000

    40000

    12

    3

    4

    5

    6

    7

    Hydrocarbon Index, FID1 A

    Correlation: 0.99918

    Rel. Res%(7): -4.0289e-1

    Area = 20647.3902*Amt -375.37592Hydrocarbon Index, FID1 AArea = 20647.3902*Amt-375.37592

    Correlation: 0.99918

    Amount [mq/m]

    Area Rel. Res%(7):-4.0289e-1

  • Discrimination Test:25 sequential injections of a standardcontaining tetracontane C-40 and decane C-10 were run to test for discriminationeffects within the analytical system. Thedata obtained is presented in Figure 3 anddemonstrates the robustness of the methodand that the analytical system clearlypasses the discrimination test specified inthe ISO Standard.

    Results and DicussionThe Agilent 6890 and 6850 GasChromatographs equipped with aSplit/Splitless inlet and Flame IonisationDetector can be used for Hydrocarbon OilIndex determinations in accordance with ISO9377-2 (DIN H53) standard. In addition, theanalytical system is easy to operate andrequires minimal maintenance. Together, theAgilent GC, analytical capillary column,special inlet liner, custom report template,selected chemical standards and Agilentapplications expertise provide a completeanalytical solution for Hydrocarbon Oil Indexdeterminations.

    Publication number 5988-0621EN Agilent Technologies Page 3

    Figure 3: Discrimination Test

    Figure 4: Report Printout

    50.0

    60.0

    70.0

    80.0

    90.0

    100.0

    110.0

    120.0

    130.0

    140.0

    150.0

    1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0 21.0 22.0 23.0 24.0

    A mount / Run No. [Decane C10] A mount / Run No. [Tetracontane C40] Amount ( Mean ) A mount ( Mean )

    min5 10 15 20 2530

    pA

    10

    20

    30

    40

    50

    60

    70

    80

    FID1 A, (HYC2509\008F0802.D)

    2.0

    44 2.1

    56 2

    .248

    2.5

    12 2

    .599

    2.7

    81 3

    .009

    3.1

    27 3

    .254

    3.3

    78 3

    .492

    3.6

    45 4

    .101

    4.2

    07

    13.

    332

    22.

    533

    ============================================================================

    Hydrocarbon Index Report============================================================================

    Calib. Data Modified : Fri, 29. Sep. 2000,08:25:56 amMultiplier : 1.000000

    Hydrocarbon Index 1.052 mg/mlLow Boilers 9.061 mg/lHigh Boilers not detectable mg/l

    ============================================================================

    Detailed Report============================================================================

    Name StartTime EndTime Ret Time Total Area Res.F Amount----------------------------------------------------------------------------

    Low Boilers 2.00 4.00 - 239.48 0.03784 9.06Decane C10 - - 4.21 213.73 0.03784 8.09Hydrocarbon Index 4.31 22.37 13.33 15408.08 0.00006 1.05Tetracontane C40 - - 22.53 205.67 0.03247 7.94High Boilers - - - - - n.d.============================================================================

    Data file : C:\HPCHEM\1\DATA\HYC2509\008F0802.DSample Name : 1000 mg/l 1===================================================================

    Injection Date : 9/26/2000 3:23:50 AM Seq Line : 8Sample Name : 1.0 mg/ml Vial No. : 8Acq Operator : berwuest Inj. No. : 2

    Inj. Vol. : -Acq.Method : HYCA04.MAnalysis Method : C:\HPCHEM\1\METHODS\ISO93774.MLast Changed : Fri, 29. Sep. 2000, 10:04:10 am

    Analysis of Hydrocarbon Index according to Iso 9377-4 andDIN H53

  • Copyright 2000 Agilent TechnologiesAll Rights Reserved.

    Reproduction, adaptation or translationwithout prior written permission is prohibited, except as allowed under thecopyright laws.

    Agilent Technologies shall not be liablefor errors contained herein or forincidental or consequential damages inconnection with the furnishing,performance, or use of this material.

    Information, descriptions andspecifications in this publication aresubject to change without notice.

    Publication Number 5988-0621EN

    Bernhard Wst is an application chemistbased at Agilent Technologies GmbHHewlett Packard Str. 876337 WaldbronnGermany

    For more information on our products and services, visit our website at:www.agilent.com/chem