88
The Pennsylvania State University The Graduate School Graduate Program in Acoustics MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS A Thesis in Acoustics by Ryan S. Scott 2008 Ryan S. Scott Submitted in Partial Fulfillment of the Requirements for the Degree of Master of Science May 2008

MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

The Pennsylvania State University

The Graduate School

Graduate Program in Acoustics

MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

A Thesis in

Acoustics

by

Ryan S. Scott

2008 Ryan S. Scott

Submitted in Partial Fulfillment of the Requirements

for the Degree of

Master of Science

May 2008

Page 2: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

ii

The thesis of Ryan S. Scott was reviewed and approved* by the following:

Richard J. Meyer Associate Professor of Acoustics Senior Research Associate Thesis Advisor

Stephen C. Thompson Professor of Acoustics Senior Scientist

Thomas B. Gabrielson Professor of Acoustics Senior Scientist

Anthony A. Atchley Professor of Acoustics Head of the Department of the Graduate Program in Acoustics

*Signatures are on file in the Graduate School

Page 3: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

iii

ABSTRACT

Magnetostriction can be described most generally as the deformation of a material

in response to a change in its magnetization. A magnetostrictive transducer was designed

and tested in partnership with Etrema, Inc using Galfenol as the active material. This

transducer was the first tonpilz element to be built with Galfenol. Initial attempts at

predicting the transducer’s performance using finite element analysis showed that more

accurate measurements of the magnetostrictive constant, elastic constant, permeability,

and mechanical quality factor were needed. Measurements of the transducer’s

performance also showed that these properties are nonlinear with drive level and highly

dependent on the magnetic bias field.

A simple measurement technique was needed to accurately determine and track

material properties as this unique material is developed. A method for measuring these

properties as a function of magnetic bias and AC drive amplitude was developed. A

matrix of the magnetostrictive material properties as a function of magnetic bias and

drive level was created using this method for Galfenol. An experimental sonar transducer

was built and tested to validate the measurement process. The measured performance is

compared with a finite element model using the material properties measured in this

work. Permeability is shown to be accurately measured. The predicted resonance

frequency of the transducer is 11 percent higher then measured showing that

improvements need to be made in measuring the elastic constant.

Page 4: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

iv

TABLE OF CONTENTS

LIST OF FIGURES ..................................................................................................... v

LIST OF TABLES ....................................................................................................... viii

ACKNOWLEDGEMENTS ......................................................................................... ix

Chapter 1 Introduction................................................................................................ 1

1.1 Background ..................................................................................................... 1 1.2 Magnetism ...................................................................................................... 3 1.3 Magnetostriction............................................................................................. 4 1.4 Governing Equations ...................................................................................... 6 1.5 Galfenol .......................................................................................................... 8 1.6 Experimental Galfenol Transducer ................................................................. 10 1.7 Measuring Material Properties ....................................................................... 12 1.8 Proposed Work ............................................................................................... 15

Chapter 2 Theory ........................................................................................................ 17

2.1 Magnetics........................................................................................................ 17 2.2 Magnetostriction............................................................................................. 20

Chapter 3 Experimental Method ................................................................................. 22

3.1 Requirements .................................................................................................. 22 3.2 Apparatus Design............................................................................................ 23 3.3 Procedure ....................................................................................................... 38 3.4 Data Analysis .................................................................................................. 44

Chapter 4 Results ........................................................................................................ 54

4.1 Galfenol Samples ............................................................................................ 54 4.2 Property Measurements vs DC Bias ............................................................... 57 4.3 Measurements versus AC Drive Level........................................................... 63 4.4 Uncertainty ..................................................................................................... 67

Chapter 5 ...................................................................................................................... 76

5.1 Property Discussion........................................................................................ 69 5.2 Transducer Measurements .............................................................................. 70 5.3 Future Work .................................................................................................... 77

Bibliography................................................................................................................. 78

Page 5: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

v

LIST OF FIGURES

Figure 1-1: Typical WWII magnetostrictive sonar transducer designs reproduced from [6] ................................................................................................................. 2

Figure 1-2: Orientation of magnetic domains a) paramagnetic state, b) ferromagnetic state with zero net magnetization and c) fully saturated state reproduced from [2] .............................................................................................. 5

Figure 1-3: Induced strain as a function of applied magnetic field reproduced from [3].......................................................................................................................... 6

Figure 1-4: Schematic (left) and photograph (right) of the dual rod galfenol transducer.............................................................................................................. 11

Figure 2-1: magnetic field lines representing the magnetic field generated by a (a) single coil turn b) solenoid and c) helmholtz coil................................................. 19

Figure 3-1: Galfenol strain vs magnetic field curve showing the desired operating region of magnetic excitation as measured by Etrema, Inc. ................................. 24

Figure 3-2: Schematic showing parallel sections of turns within one DC coil............ 27

Figure 3-3: AC and DC helmholtz coils ...................................................................... 32

Figure 3-4: AC and DC helmholtz coils showing mandrel geometry ......................... 33

Figure 3-5: Measured DC magnetic field vs axial center with medium DC current input ...................................................................................................................... 34

Figure 3-6: Magnetic field versus axial distance for the AC helmholtz coil pair at medium drive at 1 kHz ........................................................................................ 35

Figure 3-7: Magnetic end-piece geometry with metal tab shown................................ 37

Figure 3-8: Simulated magnetic field versus axial position generated by the DC helmholtz coil pair without the sample rod, with the rod and with the rod and magnetic end-pieces in place. Plotted magnitude is normalized to the field generated with no rod in place. ............................................................................. 37

Figure 3-9: Fixture for holding magnetostrictive sample in place with magnetic end-pieces and sense-coil in place ........................................................................ 38

Figure 3-10: Plot depicting timing of AC trigger delay as compensation for the DC bias current ramp up and overshoot ............................................................... 40

Page 6: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

vi

Figure 3-11: Block diagram showing equipment setup ............................................... 41

Figure 3-12: Input current as a function of time .......................................................... 44

Figure 3-13: Magnetic Field as a function of time ...................................................... 45

Figure 3-14: Sense-coil voltage as a function of time ................................................. 46

Figure 3-15: Velocity as a function of time ................................................................. 47

Figure 3-16: AC coil impedance measured with the DC coil leads electrically open....................................................................................................................... 48

Figure 3-17: AC coil impedance measured with the DC coil leads electrically shorted................................................................................................................... 49

Figure 3-18: Sine sweep current input from constant voltage input ............................ 50

Figure 3-19: Resultant sine swept magnetic field ........................................................ 51

Figure 3-20: Measured Velocity resulting from sine swept magnetic exciation ......... 52

Figure 4-1: Strain as a function of magnetic field measured at 1 Hz with a 2.275 MPa preload .......................................................................................................... 55

Figure 4-2: Magnetic flux as a function of magnetic field measured at 1 Hz with a 2.275 MPa preload ............................................................................................. 56

Figure 4-3: Relative permeability, µr, as a function of DC bias for rods A, B and C at low AC drive level............................................................................................ 58

Figure 4-4: Magnetostrictive constant, d, as a function of DC bias for rods A, B and C at low AC drive level ................................................................................. 59

Figure 4-5: Elastic Constant as a function of DC bias for rods A, B and C at low AC drive level....................................................................................................... 60

Figure 4-6: Mechanical Quality Factor as a function of DC bias for rods A, B and C at low AC drive level ........................................................................................ 61

Figure 4-7: Fundamental longitudinal resonance frequency as a function of DC bias for rods A, B and C at low AC drive level.................................................... 62

Figure 4-8: Relative permeability as a function of AC drive level for rods A, B and C at low optimum DC bias............................................................................. 63

Page 7: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

vii

Figure 4-9: Magnetostrictive constant as a function of AC drive level for rods A, B and C at low optimum DC bias ......................................................................... 64

Figure 4-10: Elastic constant as a function of AC drive level for rods A, B and C at low optimum DC bias ....................................................................................... 65

Figure 4-11: Mechanical quality factor as a function of AC drive level for rods A, B and C at low optimum DC bias ......................................................................... 66

Figure 4-12: Fundamental longitudinal resonance frequency as a function of AC drive level for rods A, B and C at low optimum DC bias .................................... 67

Figure 5-1: Impedance as a function of frequency for increasing drive level for the dual-rod Galfeno l transducer ................................................................................ 71

Figure 5-2: Galfenol transducer built to validate experimental results........................ 72

Figure 5-3: Galfenol transducer without coil and magnetic return path cylinder ........ 73

Figure 5-4: Final state of the Galfenol transducer as tested......................................... 73

Figure 5-5: Measured (blue) and modeled (green) impedance as a function of frequency for the single rod Galfenol transducer in air at a DC bias of 155 and 200 Oe at low AC drive level ........................................................................ 74

Page 8: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

viii

LIST OF TABLES

Table 1-1: Definition of variables used in magnetostrictive equations ....................... 7

Table 1-2: Comparison of maximum strain capability for nickel, Terfenol-D and Galfenol [3], [5], [6] ............................................................................................. 8

Table 1-3: Comparison of Galfenol and Terfenol-D material properties [3], [14] ...... 10

Table 3-1: Instruments Inc. L-10 Channel Settings ..................................................... 25

Table 3-2: DC Helmholtz coil parameters ................................................................... 27

Table 3-3: Incremental values of correction factor, K, for non- ideal solenoid inductance [13] ..................................................................................................... 29

Table 3-4: Choices of AC coil parameters and predicted magnetic field magnitude .. 30

Table 4-1: Diameter and length for each Galfenol sample .......................................... 54

Table 4-2: Maximum values for d and µ with corresponding optimum bias fields as measured at Etrema measured at 1 Hz with a prestress of 2.275 MPa [14] ..... 57

Table 4-3: Mean and Standard Deviations for 10 measurements of magnetic field, sense-coil voltage and velocity............................................................................. 68

Page 9: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

ix

ACKNOWLEDGEMENTS

I would like to take this opportunity to thank my adviser Dr. Rich Meyer for his

wonderful guidance and patience throughout the scope of this project. I would also like to

thank my family and friends for being so supportive and encouraging. I give credit to my

Grandpa Max and my father for instilling the engineering spirit in me. This work would

not be possible without the help of many people at the Applied Research Lab, including

Eric Bienert and Mark Wilson. Also, I owe a great deal to Scott Porter for helping out so

much with experiments and for being such a great peer with which I could discuss the

complex issues of magnetism. I would also like to thank all of the faculty and staff in the

Graduate Program in Acoustics at Penn State for providing such a warm atmosphere to

gain such an amazing education.

Page 10: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

Chapter 1

Introduction

1.1 Background

SONAR (SOund Navigation And Ranging) is the transmission of acoustic energy

through water. It is commonly used to transmit signals and as a means for locating

objects under water. The general operating frequency ranges from several hundred hertz

to hundreds of kilohertz. Magnetostrictive materials were commonly used in WWII sonar

transducer designs. A few examples of such designs are shown in Figure 1-1 [6].

Magnetostrictive materials undergo a body deformation in response to a change in

magnetization. This process also happens in reverse, where an imposed deformation of

the material will induce a change in magnetization. This characteristic is what enables

magnetostrictive materials to be used as the active material in transducers.

The discovery of piezoelectric materials soon replaced magnetostrictives in most

sonar applications due to the larger strain capabilities and higher energy densities. The

recent discovery of giant magnetostrictive materials such as Terfenol-D and Galfenol has

led to a renewed interest in using magnetostrictives for transducer design. Although it is

not expected that giant magnetostrictive materials will replace piezoelectrics, they do

possess characteristics which make them suitable for certain sonar applications. These

characteristics include competitive strain capabilities and in the case of Galfenol, a

physical robustness and durability which piezoelectrics do not have.

Page 11: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

2

When designing a transducer, it is useful to the engineer to predict the behavior of

the finished product using modeling techniques such as lumped parameter or finite

element computer models. Accurate measurements of the material properties must be

made in order for meaningful results to be obtained from a model or simulation. The most

important magnetostrictive material properties for sonar transducer design are the

magnetostrictive coupling coefficient (d), elastic compliance (s) and magnetic

permeability (µ). These properties are dependent on magnetic bias, AC drive amplitude,

temperature, manufacturing processes, and composition. A robust method is required to

Figure 1-1: Typical WWII magnetostrictive sonar transducer designs reproduced from [6]

Page 12: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

3

accurately measure these material properties to aid the transducer designer in modeling

how magnetostrictive materials will behave in a transducer system. The work in this

thesis will develop a method which can measure d, sH, µT and Qm simultaneously.

1.2 Magnetism

Due to the nature of magnetostrictive materials, a brief overview of magnetic

theory is presented. Oersted first discovered that a magnetic field, H, is produced in the

presence of moving charge [2]. A conductor will generate a magnetic field around it

according to the right-hand-rule. A magnetic field can also be produced by a material

which is magnetically saturated and retains a remnant flux density. To find the moving

charge in this case, one must examine the electron cloud structure of the material. To gain

a basic understanding of the field generated we first look at Ampere’s circuital law

where N is the number of current-carrying conductors, each carrying a current i amps,

and l is a line vector to the point of interest [2]. Carrying out the integration for a long

thin solenoid gives the generated magnetic field

where L is the length of the coil. This says that a magnetic field is generated within the

coil proportional to the number of turns and current in the wire and inversely proportional

to the coil’s length.

∫ ⋅= dlHNi 1-1

LNi

H = 1-2

Page 13: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

4

When a magnetic field is generated in a medium the response of the medium is a

magnetic induction B, also called the flux density. The relationship between magnetic

field and magnetic induction is called the permeability, µ, of the medium where

The permeability of a material is generally given relative to the permeability of free

space, mHo /104 7−×= πµ .

1.3 Magnetostriction

Magnetostriction can be described most generally as the deformation of a body in

response to a change in its magnetization. This linking between mechanical and magnetic

domains can be explained by analyzing the problem on an atomic level where magnetism

is an intrinsically quantum mechanical and relativistic phenomenon. The scope of this

thesis is intended to cover the behavior of magnetostrictive materials on a macroscopic

level. The reader is directed to Chapter 1 of Engdahl [3] for an in-depth analysis of the

origins of magnetostriction at the atomic level.

The phenomenon of magnetostriction was discovered in the mid nineteenth

century. In 1842 Joule carried out a number of experiments characterizing the fractional

change in length of an iron bar when magnetized. The converse effect, a change in

magnetization induced by a strain was discovered by Villari [4].

Magnetostriction can be thought of as the rotation of magnetic domains within the

material. Above the Curie temperature the material is in a paramagnetic state, meaning

HB µ= . 1-3

Page 14: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

5

that it exhibits no strain or magnetization in any direction. Below the Curie temperature

the magnetic domains randomly align and result in a net magnetization of zero. As a

magnetic field is applied to the material these domains rotate to align with the applied

magnetic field, causing a change in material volume as illustrated in Figure 1-2. This

alignment occurs until saturation is reached when all domains are oriented with the

applied magnetic field [2].

The relationship between magnetic field and strain is quadratic in nature. A

positive strain is induced for an applied magnetic field in either the positive or negative

direction. A generic plot of applied magnetic field versus induced strain is shown in

Figure 1-3. The point of saturation can be seen where the application of more magnetic

field does not result in any further strain. The section between zero applied field and

Figure 1-2: Orientation of magnetic domains a) paramagnetic state, b) ferromagnetic state with zero net magnetization and c) fully saturated state reproduced from [2]

Page 15: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

6

magnetic saturation can be modeled as linear over a fairly wide range. Sonar transducers

are designed to operate in this region due to higher efficiencies. If an oscillating magnetic

field is applied around the origin then a frequency doubling will occur in the strain

output. In order to avoid this, a magnetic bias is required around which the magnetic field

is oscillated.

1.4 Governing Equations

A first order approximation of magnetostriction can be made if we assume that

the material is being operated in the linear range and that variations in the system

parameters are small compared with the initial values of the parameters. The governing

equations can then be written as

Figure 1-3: Induced strain as a function of applied magnetic field reproduced from [3]

Page 16: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

7

where a superscript S, T, H and B denotes which quantity is constant and superscript tr

denotes the matrix transpose [10]. For example, sH is the elastic compliance with H held

constant. It should be noted that these equations only describe relative changes or

oscillations with respect to an absolute quantity. Table 1-1 defines the variables used and

their respective SI units. Only one set of equations are required to completely describe

the system. The other three sets can be derived from the first through matrix

manipulation. The first set of equations 1-4 and 1-5 will be used in this work.

HdTsS trH += , 1-4

HdTB Tµ+= , 1-5

HeScT trH −= , 1-6

HeSB Sµ+= , 1-7

BgTsS trB += , 1-8

BvgTH T+−= , 1-9

BhScT trB −= , and 1-10

BvhSH S+−= 1-11

Table 1-1: Definition of variables used in magnetostrictive equations

Quantity Symbol SI Unit Stress T N/m2 Strain S numeric Magnetic Field Strength H A/m Magnetic Flux Density B T Elastic Compliance s m2/N Permeability µ H/m Effective magnetostrictive constant d m/A

Page 17: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

8

The three properties which are needed to describe a magnetostrictive material’s

behavior using equations 1-4 and 1-5 are defined as

1.5 Galfenol

Galfenol is a magnetostrictive alloy deriving the name from its atomic makeup of

gallium and iron and the laboratory that it was developed at, the Naval Ordinance

Laboratory (now the Naval Surface Warfare Center). The material tested in this work has

a composition of 18.4 at% Ga. Terfenol-D is a magnetostrictive material made up of

terbium, dysprosium and iron which is named in similar fashion to Galfenol. A

comparison of the maximum measured strain output of Terfenol-D, Galfenol and Nickel

is shown below in Table 1-2.

HT T

BHS

d∂∂

=∂∂

= , 1-12

H

H

TS

s∂∂

= , and 1-13

T

T

HB

∂∂

=µ . 1-14

Table 1-2: Comparison of maximum strain capability for nickel, Terfenol-D and Galfenol [3], [5], [6]

Material Maximum Strain [ppm]

Nickel 40

Terfenol-D 1400

Galfenol 400

Page 18: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

9

Galfenol possesses many properties which make it unique and advantageous as

the active element in sonar transducers. Unlike Terfenol-D, it is not brittle and can thus

withstand large tensile stress. It is possible to stress anneal Galfenol such that it has full

magnetostriction far into the tensile range [5]. The need for a prestress bolt which is

required for most other sonar materials is not required for Galfenol due to its ability to be

stress annealed and its large tensile strength. This opens the door for more creativity in

the transducer design and is a feature which makes Galfenol attractive to sonar transducer

designers. The comparatively high ductility also means that it can be rolled into sheets

which allows for the previous work on nickel designs from WWII to be resurrected and

evaluated with Galfenol in mind. The permeability of Galfenol is approximately twenty

times greater then that of Terfenol-D. Because flux lines tend toward a medium of higher

permeability, Galfenol will act as a more efficient material in the magnetic return path

when compared with magnetostrictives of lower permeability. Table 1-3 compares

approximate material properties of Galfenol and Terfenol-D taken from measurements

made at Etrema, Inc and values listed in Engdahl [3]. Measurements of Galfenol’s

properties were made at Etrema by applying a 1 Hz magnetic field to full magnetic and

mechanical saturation while the induced flux and strain were recorded.

Page 19: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

10

1.6 Experimental Galfenol Transducer

A proof of concept transducer was designed and built in partnership with Etrema,

Inc. as a joint project. Galfenol was used as the active material to showcase the unique

characteristics. The transducer was designed to take full advantage of galfenol’s high

permeability and mechanical strength. A schematic and picture of the transducer are

shown in Figure 1-4. Two parallel rods of galfenol are used as the active material which

push and pull in unison. Both rods also act as sections of the magnetic circuit. A high

permeability cylinder is typically used to complete the magnetic circuit for single rod

magnetostrictive transducers. The galfenol rods are stress annealed which removes the

need for a mechanical preload typically applied by a prestress bolt. The goal was a larger

volume of active material and fewer mechanical and magnetic complications in the

design. The goal of the design was to operate in tension, have an active magnetic return

path and not require a mechanical bias bolt.

Table 1-3: Comparison of Galfenol and Terfenol-D material properties [3], [14]

Units Galfenol Terfenol-D

s33H m2/N 17E-12 42E-12

d33 m/A 18E-9 11E-9

µ33/µo --- 70 4 ρ kg/m3 7870 9200

Page 20: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

11

The work in this thesis originally began by testing and modeling this transducer

with the intent of verifying that modeling techniques can accurately predict the behavior

of a magnetostrictive transducer. Modeled impedance and experimentally measured

impedance are shown in Figure 1-5. The modeled resonance frequency is 40% higher

then measured and overall impedance magnitudes are not in agreement. It was decided

that more accurate measurements of Galfenol material properties were required to

correctly model the transducer. A further discussion of the transducer performance will

be provided in Chapter 4.

Figure 1-4: Schematic (left) and photograph (right) of the dual rod galfenol transducer

Page 21: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

12

1.7 Measuring Material Properties

Galfenol offers the sonar transducer designer many options in regards to

mechanical and magnetic design. With so many possible design variations it becomes

necessary to have accurate modeling capabilities. The use of such modeling techniques as

finite element analysis and equivalent circuits has been shown to accurately predict and

1

10

100

1000

10000

5 7 9 11 13 15 17 19

Frequency [kHz]

|Z| [

Oh

ms]

ModeledMeasured

Figure 1-5: Modeled and experimentally measured in-air impedance magnitude of the dual rod Galfenol transducer showing the need for more accurate material properties.

Page 22: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

13

model transducers [3] [4] [6]. Accurate knowledge of material properties must be used in

order for the model to produce meaningful predictions.

The methods for measuring magnetostrictive material properties for use in early

sonar transducer designs are summarized in the NRDC Summary Technical Report [6].

The method uses samples in the form of laminated stacks of rings. Permeability is first

measured using the inductance which uses one coil wound on the ring stack to generate

magnetic field and another to measure the induced magnetic flux. The ring stack is then

suspended in a toroidally wound solenoid allowing for unconstrained radial expansion

while immersed in the generated magnetic field. The motional- impedance circle and

measured permeability are used with the governing equations to determine the

magnetostrictive and stiffness coefficients at constant field. This method could be directly

applied once the manufacturing technology for production grade Galfenol becomes

capable of producing laminated rings.

Meeks and Timme developed a method which eliminates the problem of

laminating the sample under investigation [7]. A rod of magnetostrictive material is

placed in a gap of equal length in an iron core magnetic return path around which a

magnetic drive coil is wound. A sense coil is wrapped around the magnetostrictive rod.

Meeks and Timme derived an analytical expression for the impedance as seen at the coil

leads which accounts for eddy current losses in the material based on a complex

permeability. The material property values can be determined by matching the analytical

solution to the experimentally measured values. Both ends of the sample are physically

blocked by the magnetic return path so this makes direct measurement of the strain

difficult.

Page 23: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

14

Faidley and Lund devised a method which places a rod of the magnetostrictive

sample in a transducer type device [8]. The apparatus consists of a solenoid, a magnetic

return path and a mechanical bias. It should be noted that this apparatus was designed to

test Terfenol-D, which cannot be stress annealed and requires a mechanical bias. The

analysis looks at the shape of hysteresis loops in the displacement versus applied field

measurements. The susceptibility, permeability and effective magnetostrictive constant

can be determined from these measurements. These properties were measured as a

function of magnitude and frequency of the AC magnetic field and magnitude of the DC

bias. This method does not allow for the measurement of the elastic stiffness. The

measurement is also complicated by the fact that the magnetostrictive sample under

investigation is part of a transducer system, and data are not directly measured on the

material itself. The contributions of added mass from the magnetic return plate and

linearity of prestress mechanism are details which further complicate the measurement

procedure. The magnetostrictive sample is physically encased in the experimental

apparatus so it is visibly obstructed and is also close to the solenoid which raises issues of

heating while being run at high fields. This is an important aspect to note due to the

temperature dependence of the magnetostrictive constants.

Kellogg and Flatau performed experiments to measure the stiffness coefficient of

magnetostrictive materials as a function of applied DC magnetic field under quasi-static

conditions [9]. A force is applied to the sample with a load cell and the resultant strain

was measured. The sample was placed inside a magnetic drive coil which allowed for a

DC magnetic field to be applied. A cooling system was incorporated in the transducer

construction in order to eliminate temperature from the variables in the experiment.

Page 24: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

15

Because magnetostrictive materials used in sonar applications are typically run at

resonance, the stiffness coefficient should be measured under these conditions. This

method does not enable this measurement to be made.

A simple method for measuring d, µT , sH, and Qm simultaneously is not presented

in the literature. A method for measuring these properties as a pure measurement on a

material sample without the influence of a transducer device is also missing. The ultimate

purpose for performing these measurements is to acquire accurate material property data

for use in transducer design. Previous methods used cannot measure all the desired

quantities and thus require the use of multiple experimental setups. A method which

allows the quick characterization of a magnetostrictive sample from batch to batch is

desired by both manufacturers for quality control and the ability to give accurate data to

the end user.

1.8 Proposed Work

The work in this thesis develops an experimental technique that can measure the

main desired material properties as a function of DC magnetic bias and AC magnetic

drive level. The material sample will be placed in a uniform magnetic field generated by

larger drive coils which eliminates the need for a magnetic return path. By suspending the

material sample in the magnetic field generated by a physically open coil system, the

measurements will be purely dependent on the material properties themselves with no

influence from the experimental apparatus. Open coils also allow for observation of the

sample’s temperature with a thermal camera. Such independent measurements are the

Page 25: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

16

ultimate goal in characterizing magnetostrictive materials and are required to accurately

design sonar transducers with these materials.

Material properties can be measured as a function of DC bias, drive amplitude and

temperature. The test method shown in this thesis is also robust in that variation in

sample geometry is allowable.

Page 26: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

Chapter 2

Theory

2.1 Magnetics

Magnetic field, H, can be considered to be the most fundamental concept in

magnetics. A magnetic field is produced by electric charge in motion, as first discovered

by Oersted in 1819 [11]. This moving charge can be in the form of a current in a wire or

the orbital motions and spins of electrons within a permanent magnetic material. To

calculate the strength of the magnetic field Biot-Savart law can be used in the form

where i is the current flowing in an elemental length dl of conductor, r is the radial

distance, u is a unit vector along the radial direction and dH is the contribution to the

magnetic field at r due to the current in the elemental length [2]. It can be shown that

equation 2-1 is equivalent to Ampere’s circuital law. Units of magnetic field, H, are given

in A/m.

When a magnetic field, H, is generated in a medium, the response by the medium

is its magnetic induction, B, also referred to as the magnetic flux density. The units of

magnetic induction, B, are given in tesla. A tesla is defined as the strength of magnetic

induction which generates a force of 1 newton per meter on a conductor carrying a

ulir

H ×= δπ

δ 241

2-1

Page 27: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

18

current of 1 ampere perpendicular to the direction of the induction [2]. The units of one

tesla are equivalent to one Vs/m2. In free space the magnetic induction, B, is directly

proportional to H as given by

where µo is the permeability of free space defined as 7104 −×π H/m. The permeability of a

material is generally defined relative to the permeability of free space where

Using Faraday’s law and Lenz’s Law the law of electromagnetic induction can be written

as

where V is the induced voltage, N is the number of turns in the coil and A is the cross-

sectional area of the coil.

The Biot-Savart law can be used to determine the magnetic field generated by a

single-turn circular coil of radius a. Dividing the coil into elements of arc length dl, the

magnetic field can be written as [2]

This magnetic field is valid only at the center of a coil of finite length. The magnitude of

H quickly drops off with axial distance from the coil. A solenoid is a multi- turn coil

which generates a longer area of magnetic field at its center. Helmholtz coils can be used

if a more uniform magnetic field over a larger volume is required. This configuration

HB οµ= 2-2

οµµ

µ =r 2-3

dtdB

NAV −= 2-4

∑ ==ai

lir

H2

sin4

12 θδ

π. 2-5

Page 28: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

19

consists of two flat coaxial coils with current flowing in the same direction in each coil.

The separation distance between the coils is equal to their common radius, a. Using the

Biot-Savart law, the magnetic field generated at any point along the axis of the Helmholtz

coils can be calculated as

The magnetic field lines for a single coil turn, a solenoid and a Helmholtz coil are shown

in Figure 2-1.

If a material of permeability greater than unity is placed in the generated magnetic

field, an interaction will occur. The magnetic field will tend to align the magnetic

domains in the magnetic material creating an induced magnetic pole oriented in the

opposite direction of the applied field. This induced magnetic pole tends to oppose the

applied field and reduce the overall magnitude of field seen by the magnetic material.

The severity of this trend is reduced as the ratio of length to diameter is increased. Long

( )

−++

+=

−− 5.1

2

25.1

2

2

112 a

xaax

aNi

H 2-6

(a) (b) (c)

Figure 2-1: Magnetic field lines representing the magnetic field generated by a (a) single coil turn b) solenoid and c) helmholtz coil [2]

Page 29: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

20

thin rods are not affected by demagnetization to the same degree as a short fat rod. A

torus of material has a length to diameter ratio of infinity and is not affected by

demagnetization.

2.2 Magnetostriction

The governing equations for magnetostrictive materials were outlined in Chapter 1. The

independent variables required to fully describe the magnetostrictive behavior are defined

as

As defined above, the effective magnetostrictive constant, d, is the change in

strain for a change in magnetic field with uniform stress throughout the material. It is also

defined as a change in magnetic induction for a change in stress with a uniform magnetic

field throughout the material. The elastic compliance, sH, is defined as a change in strain

for a change in stress with a constant magnetic field throughout the material. The

permeability, µT , is defined in the same manner with the additional requirement that

stress be uniform throughout the material.

It is important for these properties to be clearly defined in order for accurate

experimental measurements to be made. The proposed method in this thesis will be based

on the definitions in equations 2-7 through 2-9.

HT TB

HS

d∂∂

=∂∂

= , 2-7

H

H

TS

s∂∂

= , and 2-8

T

T

HB

∂∂

=µ . 2-9

Page 30: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS
Page 31: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

Chapter 3

Experimental Method

3.1 Requirements

A method is needed which can measure the magnetostrictive properties as defined

in Chapter 2. The goal of the work done in this thesis was to measure d, sH, Qm and µT as

a function of DC bias and AC drive level. The test method must allow for control and

variation of both of these parameters during the measurement procedure.

The geometry of the samples was chosen to be in the form of a rod to allow for

the simplified analysis that an axisymetric problem yields. The nominal sample size was

chosen to ease both the fabrication and measurement process. A longer rod will lower the

fundamental resonance frequency of the rod, thus lowering the effect of eddy current

losses during testing. A longer rod will, however, be harder to fabricate and yield fewer

samples per batch. It will also be more expensive to generate a uniform magnetic field

along the entirety of its length due to increased power requirements as the radius of the

Helmholtz coils is increased. It is also desired to allow for some flexibility in the exact

dimensions of the material sample being measured.

A length of 4.7 cm (1.85 inches) and a diameter of 0.318 cm (0.125 inches) were

chosen as a compromise between lowering the fundamental resonance frequency and

minimizing the required length of uniform magnetic field.

Page 32: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

23

3.2 Apparatus Design

The experimental method is designed to meet the specifications required for

measuring the desired parameters. Helmholtz coils were chosen to generate the magnetic

field. When compared with a solenoid of equal magnetic field generation the open area

within the coil is significantly large r. This makes exact placement of the sample in the

apparatus less critical. The geometry of the sample is also less critical allowing more

flexibility as manufacturing techniques evolve. Temperature is also more easily

controlled due to physical separation of the material sample from the ohmic heating of

the coils.

Figure 3-1 shows the typical strain versus magnetic field behavior of Galfenol as

measured under quasistatic conditions. The operating region of DC bias and maximum

AC swing which is of most interest to the sonar transducer designer is represented as the

shaded region. The generated magnetic field requirements were chosen as a DC bias of 0

to 30 kA/m and an AC magnitude up to 15 kA/m (30 kA/m peak to peak). This allows for

measurement of the material properties through the entire linear region of the strain

versus field curve and AC drive into magnetic saturation.

Page 33: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

24

The operating frequency of the AC coils was chosen to allow for a variation in

sample length. The AC coils must be able to apply an excitation field at and above the

fundamental longitudinal resonance frequency of the material sample being measured to

accurately measure the resonance frequency and mechanical quality factor. The

resonance frequency of the sample rods supplied was estimated to be about 30 kHz based

on the previously measured stiffness. A frequency range of up to 50 kHz was chosen for

the AC drive coils to allow for flexibility in the material sample sizes.

Figure 3-1: Galfenol strain as a function of applied quasistatic magnetic field as measured by Etrema, Inc (79.578 A/m = 1Oe). The optimum range of operation is highlighted.

Page 34: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

25

Litz wire is commonly used to reduce electrical losses from high frequency

signals. Litz wire consists of a number of high gage insulated wires braided together to

reduce the power losses associated with eddy currents and the “skin effect” where current

tends to be concentrated at the surface of the conductor at high frequencies. The

resistance per unit length of typical Litz wire is higher then required for the DC coils.

This choice of wire conflicted with the need for a low resistance conductor for the DC

current component due to limitations in resistance per length of the wire available from

manufacturers. The required DC current was high based on preliminary calculations so

ohmic heating and available voltage was a factor in selecting wire gage. The competing

requirements of low resistance for the DC component and high frequency for the AC

component led to the decision to generate the AC and DC magnetic field with two

separate Helmholtz coil pairs. The separate coils could then be optimized for AC and DC

drive without sacrificing one for the other.

A 40 Volt and 25 Amp Sorrenson constant voltage amplifier is used as the DC

power supply. An Instruments Inc. L-10 is used as the AC power supply. Table 3-1

shows the channel settings available on the L-10.

A nominal radius for the DC coil pair was first chosen to achieve a magnetic field

with a magnitude of equal value within 1% over a cylindrical volume of 5.08 cm (2 inch)

length and 0.635 cm (0.25 inch) diameter while maximizing the current to magnetic field

Table 3-1: Instruments Inc. L-10 Channel Settings

Load Resistance [O] 3.5 7 14 28 56 111 222 Max Amps [A rms] 30 22 15 11 7.7 5.5 3.8 Max Volts [V rms] 106 150 212 300 425 600 850

Page 35: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

26

conversion. A numerical simulation using Eq. 2-6 was used to determine the magnitude

and volume of uniform magnetic field as a function of the various coil parameters. It was

determined that a coil radius of 8.89 cm (3.5 inch) is required to achieve the DC magnetic

field.

The number of turns and wire gage were chosen next. These values were

optimized to maximize the generated magnetic field within the limits of the available DC

power supply. Increasing the number of turns will increase the value of generated

magnetic field per input current but this reaches a limit for a non- ideal wire having a

finite resistance per unit length. Choosing the wire gage required a compromise between

minimizing wire resistance or wire diameter and dimensions of the coil. Increasing the

length and thickness of the coils also means further deviation from the assumptions that a

Helmholtz coil’s radius is much larger then its length and that the ratio of outer to inner

radius is close to one. The resistance can be reduced by splitting the total turns into

several parallel sections. This is shown schematically in Figure 3-2. Each parallel section

of turns is represented as a lumped resistance, R, and an inductance, L. An interwinding

capacitance, Cwind, is also present in each of the coils. The exact value of the capacitance

is difficult to predict because it is highly dependant on how tightly the coil is wound.

Page 36: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

27

A compromise between maximizing the generated magnetic field and minimizing

the resistance and capacitance was made in selecting the number of turns and parallel

sections. The selected DC coil parameters are shown in Table 3-2.

Due to heating effects, the DC coils were implemented with a switch which turns

on just before a measurement is made. The steady state electrical impedance of the DC

coil is only dependent on the combination of parallel wire resistances. The transient

response, as the current is ramped up, will oscillate or “ring” about the steady state value.

The DC power supply contains a feedback loop which attempts to minimize this ringing

Figure 3-2: Schematic showing parallel sections of turns within one DC coil

Table 3-2: Final DC Helmholtz coil parameters

Nominal Radius 8.89 cm (3.5 in) Number of turns 800 Number of parallel sections 4 (200 turns/section) Wire Gage 18 AWG

Cwind R

L

R R R

Page 37: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

28

and reach steady state. This issue is overcome by using a trigger to delay the AC

excitation and data capture until the DC steady state is reached.

The presence of the DC coils set a limit on the size and spacing of the AC coils.

The AC drive coils were designed to fit within the DC coils to allow for more leeway in

the choice of AC coil radius. The inner radius of the DC coils restricted the maximum

outer radius of the AC coils. This meant that maintaining the magnetic field magnitude

within 1% of the center value was not possible. This requirement was relaxed to 3% and

was met by choosing a radius of 6.35 cm (2.5 inch) for the AC coils. This also allowed

for a reasonable thickness in the coil turns of both sets of coils. The frequency range up to

50 kHz demanded that Litz wire be used as the AC coil windings. Litz wire is sold as

bundles of a specific wire gage. The first step in Litz wire selection is to choose the gage

based on the upper frequency limit. In this case 36 AWG was chosen to allow for the

desired frequency range. The resistance per unit length and current carrying capacity is

set by the number of parallel strands in the bundle. The wire was chosen to be 20/36

AWG litz wire, which is made of 20 strands of 36 AWG wire in the braided

configuration which reduces surface losses.

As the drive frequency is increased the electrical impedance will become

dominated by the inductance of the coils. The inductance of a coil, L, can be calculated as

where A is the area of the coil and K is a correctional factor for non- ideal solenoids.

Values of K for various increments of radius/length are shown in Table 3-3.

lAN

L ro2µµ

κ= 3-1

Page 38: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

29

The inductance is directly dependent on the core permeability, coil area and the

number of turns squared and is inversely proportional to the coil length. The cross

sectional area of Galfenol covers less then 1% of the coil’s cross sectional area. The core

permeability is assumed to be that of air in order to estimate the impedance of the AC

coils during the design procedure. This was verified experimentally after constructing the

Helmholtz coils by measuring impedance with and without a Galfenol rod present. It is

important to note that the same trick of parallel sections of windings per coil does not

work when attempting to reduce the overall inductance of a coil. The parallel sections

would share the same magnetic core and act as one inductor with a distributed wire cross

section. It can be imagined that the coil wire’s cross sectional area is split into smaller

wires, but the number of turns and total current still remains the same. This fact greatly

limited the AC magnetic field which could be generated.

The number of layers in each coil, along with the number of turns per layer was

simultaneously selected. Table 3-4 shows the combinations of parameters which were

Table 3-3: Incremental values of correction factor, K, for non- ideal solenoid inductance [13]

Radius/Length K 0 1

0.2 0.85 0.4 0.74 0.6 0.65 0.8 0.58 1.0 0.53 1.5 0.43 2.0 0.37 4.0 0.24 10.0 0.12

Page 39: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

30

selected. The electrical impedance at 50 kHz was calculated from the coil inductance.

Resistive impedance was neglected due to the relatively short length of wire being used.

The coil length, l, was experimentally measured by wrapping the 20/36 AWG litz wire

and averaging several iterations. The maximum current was selected based on the

available channel settings of the AC amplifier. The calculation was done with the coil

pair driven in series. The AC magnetic field magnitude at 50 kHz was used as a

comparison point to choose the best coil parameter combination.

Based on the predicted values, it can be seen that the optimum AC coil design

could only reach slightly more than 1850 A/m in peak magnitude. This does not reach the

goal of 15 kA/m peak due to limitations in AC current generation. The final coil design

was chosen to be 60 turns per coil, with two layers of turns per coil. This was selected

because the required current was lower then for the 40 and 50 turn designs and still gave

similar results in magnetic field generation.

The coil mandrels were designed next. The dimensions were chosen such that the

AC coils fit inside the DC coils. A means for adjusting the coil spacing was incorporated

to account for deviation from the ideal spacing due to non ideal coil aspect ratios. The

Table 3-4: Possible AC coil parameters and corresponding predicted AC magnetic field magnitude with available power supply (79.578 A/m = 1Oe).

Turns Length Radius/Length K Inductance Zcoil@50kHz imax H [cm] [µH] [Ohm] [A] [A/m]

20.00 1.60 3.97 0.24 96.32 30.26 5.45 1228.2 30.00 2.54 2.50 0.34 190.33 59.79 3.85 1301.4 40.00 3.30 1.92 0.37 289.19 90.85 3.85 1734.8 50.00 3.81 1.67 0.39 409.00 128.49 3.3 1858.9 60.00 2.54 2.50 0.34 761.32 239.18 2.5 1690.2 80.00 3.30 1.92 0.37 1156.75 363.40 1.65 1487.3

100.00 3.81 1.67 0.39 1636.02 513.97 1.17 1318.4

Page 40: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

31

DC coil spacing can be adjusted by moving and remounting them to the table. The AC

coils were attached to the DC coils using guide pins and a set of locking screws to allow

for spacing adjustments independent of the DC coils. The mandrels were machined out of

aluminum to withstand possible coil stresses from interactions in the generated magnetic

fields. The geometry was selected such that no radial dimension was larger then the

critical dimension for eddy currents, defined as

where eρ is the electrical resistivity, fc is the maximum frequency operation in kilohertz,

and t is the material thickness in millimeters. This was done by minimizing the cross

sectional area of the mandrel, and by cutting a slit in the mandrel to remove the

circumferential path for eddy currents.

The coil mandrels were covered in a layer of kapton tape to prevent arcing and

reduce the interwinding capacitance. Each layer of wire turns was also covered with a

layer of kapton tape. Figure 3-3 and Figure 3-4 show the finished drive coils. The AC

coils sit further inside the DC coils when the proper spacing between coil pairs is set.

cr

e

ft

µπρ

2

10

210

= 3-2

Page 41: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

32

Figure 3-3: AC and DC helmholtz coils

AC Coils

DC Coils

Page 42: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

33

The separation between coil pairs was determined experimentally. The magnetic

field at medium drive level was measured throughout the volume inside the Helmholtz

coils as the coil spacing was adjusted to yield the highest uniformity possible. It was

assumed that the variation in magnetic field throughout the coil volume is not a function

of drive level. Magnetic field was measured along the coil axis for varying DC and AC

coil spacing. Once the optimum spacing was found the coils were mounted to the test

bench. The magnetic field versus axial distance for the DC coils is shown in Figure 3-5.

The initial spacing of 8.89 cm (3.5 inch) from the center of each DC coil mass was too

small. As the coils were brought together the magnetic field tends to flatten and slightly

Figure 3-4: AC and DC helmholtz coils showing mandrel geometry

Page 43: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

34

decrease in magnitude. A final spacing of 10.48 cm (4.125 in) was chosen to give

magnetic field profile close to that predicted for ideal Helmholtz coils.

The AC coils were initially spaced 6.35 cm (2.5 inch) apart, but this changed as

the DC coils were moved. This meant that the AC coils started out with spacing larger

than required. Magnetic field versus axial center was measured with a 1 kHz sine wave

input. The trend in magnetic field as a function of spacing for the AC coils can be seen in

Figure 3-6. As the spacing of the helmholtz coils approaches infinity, the measured

Figure 3-5: Measured DC axial magnetic field as a function of axial distance with 10 A DC input current showing the curve flattening as the coil spacing is increased (79.578 A/m = 1Oe).

Page 44: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

35

magnetic field will be that of two separate coils. As the spacing approaches zero the

measured magnetic field will be that of a solenoid.

Initial test results for d, sH and µT and Qm were measured to be significantly lower

then previous methods showed. Demagnetization was discovered to be the root cause of

this issue. This caused the magnetic field seen by the sample in the experiment to be

significantly lower then was measured for the same drive current without the rod in place.

The initial calculation of the magnetostrictive parameters, ratios of strain to field and flux

Figure 3-6: Measured AC axial magnetic field as a function of axial distance with 1 kHz input current showing the curve flattening as the coil spacing is decreased (79.578 A/m = 1Oe).

Page 45: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

36

to field, were done using the predicted magnetic field applied. The calculated value of d

and µ were artificially reduced because the actual field was lower then initially predicted.

A computer model of the magnetic field generated by the helmholtz coil apparatus

was created using finite element analysis. The FEMM software package is freely

available and is made specifically for analysis of electro-magnetic problems. The model

includes the coil geometry, number of turns and wire density of both the DC and AC coil

pairs. The magnetic field magnitude is simulated with and without the rod sample in

place. Figure 3-8 shows that the overall magnitude is significantly reduced along with the

field uniformity. The effects of demagnetization are reduced by placing rods of high

permeability on both ends of the rod sample. This increases the effective magnetic length

of the sample. The predicted magnetic field with the magnetic end-pieces is also plotted

below to show the improvement of nearly 6 kA/m in DC field.

A small metal stainless steel tab must be glued to one end of the rod allowing for

the laser vibrometer to measure the velocity of the rod end while maintaining alignment

of the rod and magnetic end-pieces. Figure 3-7 shows the configuration of this setup with

the laser beam depicted as a dashed line. Velocity measurements were made with the tab

glued to the sample rod to ensure that the tab could be assumed to be rigid. The velocity

at the rod center and tab tip were measured to be in phase and of equal magnitude.

Page 46: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

37

Figure 3-7: Magnetic end-piece geometry with metal tab shown

-25

-20

-15

-10

-5

0

5

0 0.5 1 1.5 2 2.5 3 3.5 4 4.5

Axial Position [cm]

No

rmal

ized

Mag

net

ic F

ield

[kA

/m]

no rodrodrod & endpieces

Figure 3-8: Simulated magnetic field versus axial position generated by the DC helmholtz coil pair without the sample rod, with the rod and with the rod and magnetic end-pieces in place. Plotted magnitude is normalized to the field generated with no rod in place (79.578 A/m = 1Oe).

magnetic end-pieces

sample rod metal tab w/ reflective tape

36% 7%

Page 47: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

38

From this analysis it can be seen that the desired uniformity in magnetic field was

not achieved. The average magnetic field in the rod is calculated to be 7% lower then the

simulated field with no rod in place. This percent reduction in magnetic field seen by the

sample was used as a first order approximation for the demagnetization factor. Figure 3-9

shows the final test fixture for holding the magnetostrictive sample with the magnetic

end-pieces and sense-coil in place. Onion skin paper was placed in the gap between the

rod and the magnetic end-pieces to simulate a zero force boundary condition.

3.3 Procedure

Applied magnetic field, magnetic induction, fundamental rod resonance, and

strain in the rod must be measured in order to calculate d, s, µ and Qm. Measurements

Figure 3-9: Fixture for holding magnetostrictive sample in place with magnetic end-pieces and sense-coil in place

Page 48: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

39

were made as a function of DC bias and AC drive level. The relationships and equations

used to calculate the desired parameters will be described in this section.

The experimental procedure consists of three steps to capture the required data.

First, the DC bias is applied to the sample. There is a small amount of “ringing” which

occurs as the current is quickly increased in the DC coils. These decaying oscillations are

due to the coil inductance and capacitance interacting with the feedback loop of the DC

power supply. The DC bias is monitored using a current probe and set by the operator

manually. An AC field is then applied to the sample. The triggering of the AC excitation

is delayed by a few milliseconds to allow for the ringing in the DC coils to decay.

Figure 3-10 graphically represent the timing of the DC bias application and delay of the

AC drive. The AC excitation starts as a sine wave with a frequency far below resonance.

This ensures that the strain is uniform throughout the sample during measurement of

permeability and the magnetostrictive constant. The AC excitaiton then uses a sine sweep

which passes through the longitudinal resonance frequency of the rod sample being

measured to determine the exact resonance frequency and mechanical quality factor, Qm.

Page 49: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

40

A block diagram is shown in Figure 3-11 which outlines the equipment used and the

signal path for the experiment.

Magnetic field is measured directly using a Lakeshore 475DSP axial hall probe.

The probe is mounted on an adjustable platform which allowed fo r axial and radial

positioning of the probe tip throughout the control volume. The hall probe measures flux

density perpendicular to the tip surface and outputs a proportional voltage. The magnetic

field can be determined using the permeability of free space.

Figure 3-10: Plot depicting timing of AC trigger delay as compensation for the DC bias current ramp up and overshoot

Page 50: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

41

The magnetic induction in the rod sample is measured using a sense coil wrapped

around the center of the rod. Ten turns of 37 AWG wire were used. The resultant voltage

in the sense coil is recorded as a magnetic field is applied. Starting with Eq. 2-4 and

noting that

substitution leads to

An applied periodic magnetic field excitation will induce a voltage of the same

frequency, f, in the sense coil. Substituting and taking derivatives leads to a direct

Figure 3-11: Block diagram showing the equipment setup and signal flow for the measurement procedure.

T

T

HB

3

333 ∂

∂=µ , 3-3

T

T

tH

NAV∂

∂= 3

33µ . 3-4

Page 51: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

42

calculation of permeability from the measured parameters. This relationship is expressed

as

where V is the induced voltage in the sense coil measured in volts, and H is the applied

magnetic field measured in A/m.

The axial velocity of the rod is measured using a laser-vibrometer. For a periodic

excitation,

where v is the measured velocity and x is the displacement of the rod end. The resultant

strain can be calculated using the relationship where

if the length of the sample, l, is known. An expression for d from the measured

parameters can be expressed by substituting into Eq. 2-7 as

where v3 and H3 are the magnitudes of the periodic velocity and magnetic field in the

axial direction. The coupling coefficient can be defined as [3]

HV

fNAT

πµ

21

33 = 3-5

fjv

xπ2

= 3-6

lx

strain2

= 3-7

THfl

vd

3

333

= 3-8

HT sd

k3333

2332

33 µ= . 3-9

Page 52: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

43

Woolett defines the longitudinal wave speed, c3, in a rod of magnetostrictive material to

be

where ρ is the density [12]. Using the relationship between sH and sB [3]

and the equation for the fundamental longitudinal resonance, fo, of a rod defined as

an equation for the magnetostrictive elastic constant based on measured parameters can

be written as

The fundamental resonance of the rod can be determined by observing the frequency at

which maximum displacement of the rod end occurs during the swept sine excitation. The

mechanical quality factor, Qm, can be calculated from the relationship

where f+ and f- are the frequencies above and below the resonance frequency where the

magnitude of the velocity spectrum is 3 dB less than the magnitude at resonance.

ρBsc

333

1= 3-10

)1( 2333333 kss HB −= 3-11

321

cl

f o = , 3-12

ρµ 2233

233

33 41

oT

H

fld

s += . 3-13

−+ −=

fff

Q om 3-14

Page 53: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

44

3.4 Data Analysis

Input current, generated magnetic field, sense-coil voltage and velocity were

measured simultaneously. The measured signals are shown below for rod A at optimum

bias and low AC drive level.

-1

-0.5

0

0.5

1

0 10 20 30 40

Time [ms]

Cu

rren

t [A

]

Figure 3-12: 1 kHz sine wave input current as measured through the AC helmholtz coils.

Page 54: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

45

-400

-300

-200

-100

0

100

200

300

400

0 10 20 30 40

Time [ms]

Mag

netic

Fie

ld [

A/m

]

Figure 3-13: Measured magnetic field generated from 1 kHz sine wave input current as shown in Figure 3-12 (1 Oe = 79.578 A/m).

Page 55: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

46

-0.015

-0.01

-0.005

0

0.005

0.01

0.015

0 10 20 30 40

Time [ms]

Vo

ltag

e [V

]

Figure 3-14: Measured sense-coil voltage induced from magnetic field excitation shown in Figure 3-13.

Page 56: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

47

During the initial testing procedure it was noted that a magnetic coupling exists

between the AC and DC coils. This coupling acts to reduce the amount of AC magnetic

field which can be generated by the apparatus. The DC coils are attached to a constant

voltage power supply, which acts as an electrical short circuit. The magnetic field

generated by the AC coils couples with the DC coils and induces a proportional current in

the DC coils which can flow unimpeded due to the short. This induced current flows in a

direction to oppose the change in magnetic flux, thus reducing the generated AC

magnetic field. Using a constant current DC amplifier would reduce this problem because

-1

-0.5

0

0.5

1

0 10 20 30 40

Time [ms]

Vel

oci

ty [

mm

/s]

Figure 3-15: Measured velocity induced from magnetic field excitation shown in Figure 3-13.

Page 57: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

48

the load impedance acts like an open circuit of infinite impedance and would act to block

this induced current. Figure 3-16 and Figure 3-17 shows the impedance of the AC coils

with the DC coil leads left electrically open and closed respectively.

0

20

40

60

80

100

120

140

160

180

0 20000 40000 60000 80000 100000

Frequency [Hz]

|Z| [

Ohm

s]

0

10

20

30

40

50

60

70

80

Pha

se [

degr

ees]

Impedance MagnitudePhase

Figure 3-16: AC coil impedance magnitude measured with the DC coil leads left open-circuit

Page 58: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

49

A second product of the magnetic coupling between the AC and DC coil pairs

was an effect on the generated magnetic field amplitude which is dependent on

frequency. A constant input current from 5 to 50 kHz results in a magnetic field of

increasing amplitude. This rise in magnetic field was approximately inversely

proportional to the rise in impedance of the AC coils. Therefore, a constant voltage sine

sweep was applied resulting in the current waveform shown in Figure 3-18. The

decreasing magnitude in the current drive offsets the increasing magnitude of the

generated magnetic field, leading to a generated magnetic field of nearly equal magnitude

through the entirety of the sine sweep.

0

200

400

600

800

1000

1200

0 20000 40000 60000 80000 100000

Frequency [Hz]

|Z| [

Ohm

s]

-60

-40

-20

0

20

40

60

80

100

Pha

se [

degr

ees]

Impedance MagnitudePhase

Figure 3-17: AC coil impedance magnitude measured with the DC coil leads left electrically short-circuit

Page 59: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

50

Figure 3-19 shows the waveform of the sine swept magne tic field. The downward slope

is a product of the hall probe’s internal processing and filters. In AC detection mode the

internal hardware integrates the measured signal and this causes a constant field to

gradually reduce in time. The graphical scalloping just after 30 milliseconds is a product

of aliasing in the hall probes internal processing.

-1

-0.5

0

0.5

1

0 10 20 30 40

Time [ms]

Cur

rent

[A]

Figure 3-18: Measured sine sweep current as input through the AC helmholtz coils using a constant voltage input.

Page 60: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

51

The measured velocity resulting from the applied frequency swept magnetic field

is shown in Figure 3-20. It can be seen that the velocity of the rod is mechanically

amplified by the fundamental longitudinal resonance of the rod.

-400

-300

-200

-100

0

100

200

300

0 10 20 30 40

Time [ms]

Mag

netic

Fie

ld [A

/m]

Figure 3-19: Resultant sine swept magnetic field as generated by the input current shown in Figure 3-18.

Page 61: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

52

A computer program was written which imports the recorded 1 kHz sine pulse

waveforms and outputs the signal amplitude with appropriate conversions. The measured

velocity resulting from the 5 to 50 kHz sine sweep is also imported allowing for the

resonance frequency along with the 3 dB down points needed to calculate Qm. It should

be noted that the experimental procedure currently uses two separate pulses, a 1 kHz sine

wave and a sine sweep. These were initially done separately to allow for independent

control of each waveform. Combining these two pulses into a single input waveform

should be trivial and would lower test time in a manufacturing application.

-80

-60

-40

-20

0

20

40

60

80

0 10 20 30 40

Time [ms]

Vel

oci

ty [

mm

/s]

Figure 3-20: Measured velocity resulting from sine swept magnetic excitation shown in Figure 3-19.

Page 62: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS
Page 63: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

Chapter 4

Results

4.1 Galfenol Samples

Measurements were made on three samples of Galfenol (18.4 at% Ga) in the

shape of a rod. The samples were produced using the Free Stand Zone Melt technique at

Etrema Products. The samples were stress annealed to 45 MPa. Each sample was then

sliced axially and glued back together with non-conductive epoxy to prevent eddy

currents. A lamination thickness of 0.04064 cm (0.016 inch) was chosen to prevent eddy

currents up to 50 kHz. The dimensions for each rod are shown below in Table 4-1.

Quasistatic measurements of d and µ were made at Etrema on the three provided

sample rods with a prestress of 2.275 MPa. Plots of strain and magnetic flux as a function

of magnetic field for all three rods are shown in Figure 4-1 and Figure 4-2 respectively.

Table 4-1: Diameter and length for each Galfenol sample

Diameter [cm] Length [cm]

Rod A 0.3175 4.722

Rod B 0.3175 4.684

Rod C 0.3175 4.702

Page 64: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

55

Figure 4-1: Strain as a function of magnetic field measured at 1 Hz with a 2.275 MPa preload as measured by Etrema, Inc (79.578 A/m = 1 Oe).

Page 65: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

56

The measurement technique used at Etrema, Inc. required the use of a preload.

The measurements were made with a minimum possible preload of 2.275 MPa and

averaged over several sweeps through magnetic and mechanical saturation. Maximum

reported material property values along with the corresponding optimum bias fields are

shown in Table 4-2.

Figure 4-2: Magnetic flux as a function of magnetic field measured at 1 Hz with a 2.275 MPa preload as measured by Etrema, Inc (79.578 A/m = 1 Oe).

Page 66: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

57

4.2 Property Measurements vs DC Bias

Data was taken for a DC bias of 9.5 to 22 kA/m. This covered the entire linear

region of the strain versus field curve as shown in the quasistatic measurements made at

Etrema. Measurements were made with low AC drive level as a function of the DC bias.

The DC values were initially chosen based on quasistatic data and modified after initial

testing was done with this experiment. Eq. 3-5, 3-8, 3-13 and 3-14 are used with the

appropriate measurements to determine the desired parameters. Measured properties as a

function of DC bias are shown in the plots below. The magnetic field used to derive these

parameters was measured with no sample rod in place. The measured field was then

reduced by 7% to compensate for the demagnetization effects as previously discussed.

Figure 4-3 shows the measured relative permeability of the rods A, B and C as a

function of DC bias. The measurement shows that the optimum bias field is not identical

for each rod even though they are taken from the same batch. Figure 4-4 shows the

Table 4-2: Maximum values for d and µ with corresponding optimum bias fields measured at 1 Hz with a prestress of 2.275 MPa as measured at Etrema [14] (79.578 A/m = 1 Oe).

Rod max µr33 max µr33 Field max d33 max d33 Field

[kA/m] [m/A] [kA/m]

A 72 15.2 18.8E-9 15.2

B 67 16.3 17E-9 17.3

C 68 14.5 17.5E-9 14.5

Page 67: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

58

magnetostrictive constant measured as a function of DC bias for the three galfenol

samples tested.

Figure 4-3: Relative permeability, µr, as a function of DC bias for rods A, B and C at low AC drive level of 430 A/m peak to peak (79.578 A/m = 1 Oe).

Page 68: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

59

Figure 4-5 shows the elastic constant, s, measured as a function of DC bias.

Figure 4-4: Magnetostrictive cons tant, d, as a function of DC bias for rods A, B and C at low AC drive level of 430 A/m peak to peak (79.578 A/m = 1 Oe).

Page 69: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

60

Figure 4-6 shows the mechanical quality factor, Qm, measured as a function of DC bias

for each rod. Figure 4-7 shows the resonance frequency of each rod measured as a

function of DC bias.

Figure 4-5: Elastic Constant as a function of DC bias for rods A, B and C at low AC drive level of 430 A/m peak to peak (79.578 A/m = 1 Oe).

Page 70: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

61

Figure 4-6: Mechanical Quality Factor as a function of DC bias for rods A, B and C at low AC drive level of 430 A/m peak to peak (79.578 A/m = 1 Oe).

Page 71: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

62

Figure 4-7: Fundamental longitudinal resonance frequency as a function of DC bias for rods A, B and C at low AC drive level of 430 A/m peak to peak (79.578 A/m = 1 Oe).

Page 72: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

63

4.3 Measurements versus AC Drive Level

Optimum DC bias fields were selected based on the measurements in the previous

section. This bias was then applied while measurements were made as the AC drive level

was increased. Figure 4-8 shows the relative permeability at optimum DC bias as the AC

drive level is incrementally increased from 430 A/m to 2.15 kA/m peak to peak. This

measurement is also made for the magnetostrictive constant, elastic constant, mechanical

quality factor and resonance frequency and shown in Figure 4-9 through 4-12.

Figure 4-8: Measured relative permeability, µ33/µo, as a function of AC drive level for rods A, B and C at optimum magnetic bias (79.578 A/m = 1 Oe).

Page 73: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

64

Figure 4-9: Measured magnetostrictive constant, d33, as a function of AC drive level for rods A, B and C at optimum magnetic bias (79.578 A/m = 1 Oe).

Page 74: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

65

Figure 4-10: Measured elastic constant, s33, as a function of AC drive level for rods A, B and C at optimum magnetic bias (79.578 A/m = 1 Oe).

Page 75: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

66

Figure 4-11: Measured mechanical quality factor, Qm, as a function of AC drive level for rods A, B and C at optimum magnetic bias (79.578 A/m = 1 Oe).

Page 76: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

67

4.4 Uncertainty

The uncertainty was evaluated by repeating a single measurement and calculating

the mean value and standard deviation. This experiment was performed on the sense-coil

voltage, magnetic field, sine swept velocity and resonance frequency. Each measurement

was repeated 10 times on rod A. Table 4-3 shows a summary of this study.

Figure 4-12: Measured fundamental longitudinal resonance frequency, fr, as a function of AC drive level for rods A, B and C at optimum magnetic bias (79.578 A/m = 1 Oe).

Page 77: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

68

The velocity measurement proved to have the highest amount of uncertainty.

Several factors contribute to this. The material sample tends to align itself with the

applied magnetic field as the DC bias is turned on. The force of this repositioning is

sensitive to the placement of the material sample in the center of the rods. This movement

of the rod oscillates until friction of the fixture causes the movement to decay and the rod

to center itself. Care was taken to build a fixture which supports the sample without

influencing its strain while maintaining a steady geometrical centering. A slight offset

from the center of the magnetic field will cause unwanted motion of the sample and

influence the measurement of velocity.

It should be noted that the measured property values at optimum bias in the study

as a function of DC bias did not match with the same measurement in the measurement as

a function of AC drive level. This discrepancy should be further investigated. It may be

due to a slight change in magnetic conversion of the Helmholtz coils from a change in

spacing or alignment.

Table 4-3: Mean and standard Deviations for 10 measurements of magnetic field, sense-coil voltage and velocity made on rod A

Measured Property Mean Standard Deviation %

Magnetic Field [kA/m] 1.397 0.0293 0.17

Sense-Coil Voltage [V] 0.1441 0.0011 0.76

Velocity [mm/s] 3.529 0.0308 0.87

Resonance Frequency [Hz] 29414 138.65 0.47

Page 78: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

69

4.5 Property Discussion

The measurements of Galfenol’s material properties showed the basic trends of

magnetostrictive materials. It was shown that µ, d and s have a maximum value at a

specific DC magnetic bias field. The mechanical quality factor reaches a minimum value

at the optimum bias value. The mechanism behind this phenomenon should be further

investigated.

It is interesting to note that the optimum bias for permeability was not the same

for the three measured material samples. It can also be seen that for a single rod the

optimum bias for permeability is not always the same as the optimum bias for the

magnetostrictive constant. Looking at rod A, for example, we see that the permeability

reaches a maximum at 165 Oe whereas the magnetostrictive constant has a maximum at

about 175 Oe. These differences may be a result of the lamination process which occurs

after the samples are stress annealed.

The results also show that the fundamental longitudinal resonance frequency of

the Galfenol samples reached a minimum at optimum magnetic bias. This is important to

note when operating a magnetostrictive transducer which is designed to operate at a

specific frequency. This fact can be used in further experiments and transducer design to

quickly determine the optimum bias field at the desired drive level by tracking the

resonance frequency.

The trends in material properties were also tracked as the AC drive level is

increased. The permeability and elastic constant increase in magnitude as the AC drive

level is increased. It is assumed that the magnetostrictive constant also follows this trend,

Page 79: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

70

but uncertainty in the velocity measurement hindered accurate results. It was also shown

that the resonance frequency is decreased as the magnetostrictive material is driven at

higher levels. A decrease by approximately 1 kHz was measured. Quantifying these

trends is important for transducer design and operation. Previous measurements were

made at low drive levels which could not predict the effects shown in this experiment.

The knowledge gained from this measurement would be greatly improved by increasing

the magnitutude of the AC drive level beyond what is currently capable with available

equipment. This would show a more general trend in the properties as a function of drive

level.

4.6 Transducer Measurements

The performance of the dual-rod transducer described in Section 1.6 was

measured. The transducer was tested in water with increasing drive level. A plot of

impedance as a function of frequency is shown in Figure 4-13. The transducer was driven

with constant voltage and the measured input current at resonance for each drive level

was converted to magnetic field. This measurement also shows the trend of the elastic

constant softening as the AC drive level is increased. The impedance of the transducer is

dominated by its inductance below resonance. The inductance is proportional to the slope

of the impedance at this point. The increasing slope with increasing drive level points to

the fact that permeability is also increasing as the drive level is increased.

Page 80: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

71

A transducer was built using rod A to assess the validity of the results decribed in

Ch. 4. The transducer design consisted of a single rod of Galfenol and a high

permeability magnetic return path. The geometry is shown in Figure 4-14. The transducer

was designed to have a fundamental resonance well below the first plate mode resonance

of the head mass to avoid “flapping”. The head mass material is aluminum, the tail mass

material is made of stainless steel, and the magnetic return path is an iron composite

which has a high permeability and high resistivity to prevent eddy currents. The goal of

this transducer was only to compare the model using measured material properties and

the experimental performance. The transducer was not optimized for acoustic

50

150

250

350

450

550

650

750

5 10 15 20 25Frequency [kHz]

|Z|

[Ohm

s]

0.1 kA/m0.4 kA/m1 kA/m3.6 kA/m

Figure 4-13: Impedance magnitude of the dual- rod Galfenol transducer measured as a function of frequency for increasing drive level in water at optimum DC bias field (79.578 Oe = 1 A/m).

Page 81: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

72

performance. The actual transducer is shown first in Figure 4-15 before the coil is wound

and the magnetic return path cylinder is attached. It is then pictured in Figure 4-16 in its

final state.

Figure 4-14: Galfenol transducer built to validate experimental results

Galfenol Rod

Tail Mass

Head Mass

Magnetic Return Path

Page 82: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

73

Figure 4-15: Galfenol transducer without coil and magnetic return path cylinder

Figure 4-16: Final state of the Galfenol transducer as tested

Page 83: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

74

A finite element computer model was created using the measured properties with

the goal of comparing measured and modeled performance. The model does not include

the glue joints between the Galfenol rod and the head and tail masses. The impedance of

the transducer was measured experimentally with a magnetic bias field of 14 kA/m and

an AC drive level of 1.4 kA/m. Figure 4-17 shows a comparison of the measured and

modeled impedance magnitude as a function of frequency.

The resonance frequency of the model is higher then the experimentally measured

value by 11 percent. This is an indication of the error in the measured value of the elastic

compliance used in the model. This error may be reduced by fixing the demagnetization

0

50

100

150

200

250

0 2000 4000 6000 8000 10000

Frequency [Hz]

Imp

edan

ce M

agn

itu

de

[Oh

ms]

ModeledMeasured

Figure 4-17: Measured (pink) and modeled (black) impedance magnitude as a function of frequency for the single rod Galfenol transducer in air at a DC bias of 14 kA/m and an AC drive level of 1.4 kA/m.

Page 84: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

75

problems in the Helmholtz coil apparatus. Demagnetization causes the applied magnetic

field at the ends of the material sample to be lower in magnitude then the at the samples

center. This acts to reduce the length of material which is at the intended magnetic bias

and thus reduces the length of active material. An artificially reduced length would cause

the measured elastic compliance to appear greater in magnitude.

The model and measured slope of the impedance above resonance show a good

match. This shows that the permeability measured using the experiment in Chapter 4 is

valid. The coupling coefficient, defined in Eq. 3-9 can also be defined as

where fr is the resonance frequency and fa is the antiresonance frequency. Calculating the

experimentally measured coupling coefficient gives a value of 0.397. Calculating the

coupling coefficient of the model yields a value of 0.418. If we assume that the measured

values for permeability and elastic coefficient used in the model are accurate we can use

the coupling coefficient as a validation of the magnetostrictive constant measurement.

Comparing the coupling coefficients of the modeled and measured impedance shows that

the measured magnetostrictive coupling coefficient is relatively accurate.

The initial goal of this experiment was to accurately characterize Galfenol for

transducer design. The comparison of the measured and modeled impedance of the

transducer shows that the experimental method described in this thesis can measure the

permeability, magnetostrictive constant and elastic constant of Galfenol as a function of

DC magnetic bias and AC drive level.

2

2 1

−=

a

r

ff

k 4-1

Page 85: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

Chapter 5

5.1 Conclusions

A magnetostrictive transducer was designed and tested in partnership with

Etrema, Inc. Initial attempts at predicting the transducer’s performance using finite

element analysis showed that more accurate measurements of the magnetostrictive

constant, d, elastic constant, s, permeability, µ, and mechanical quality factor, Qm, are

needed. Measurements of the transducer’s performance also showed that these properties

are nonlinear with drive level.

A method for measuring d, s, µ and Qm as a function of DC magnetic bias and AC

drive level was developed without a transducer- like apparatus used in previous methods.

The method is robust in allowable geometry and can be adapted for quality control in an

industrial setting. A matrix of Galfenol material properties was derived using this method

that will help to better understand the trends as a function of bias and drive level.

A transducer was built to validate the results of this work using one of the

Galfenol samples. A finite element model using the measured material properties showed

good correlation with the experimentally measured transducer performance. The

permeability measurement was shown to be accurate. An 11% difference between the

modeled and measured resonance frequency of the transducer shows that improvements

Page 86: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

77

need to be made in measuring the elastic constant. This discrepancy is attributed to

demagnetization effects in the measurement procedure.

5.2 Future Work

In order to further improve the accuracy of the experiment a few improvements to

the technique can be made. Modifications to the magnetic end-pieces should be

investigated to improve the magnetic field magnitude throughout the sample. Deviations

from constant magnetic field throughout the volume of the material sample are a source

of error in all measurements made thus far.

Improvements in the AC power supply need to be addressed. The goal stated in

the design of the apparatus called for an AC drive level of 16 kA/m. A maximum of 21.5

kA/m AC drive level was reached with the available amplifiers. Increasing the drive level

capabilities will help further characterize the nonlinearities of the material properties as a

function of drive level.

A few steps can be adopted for use in a manufacturing environment. The 1 kHz

sine pulse and the sine sweep pulse should be combined into one waveform. Automating

the process of magnetic excitation, data collection and data processing can be done with

one of several commercial softare and hardware packages.

Page 87: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

Bibliography

1 M. Velazquez, “Analysis of the Effective Coupling Coefficient of a Terfenol-D Tonpilz Transducer”, The Pennsylvania State University, 2005.

2 David Jiles. Introduction to Magnetism and Magnetic Materials. Chapman and Hall, 1991.

3. Goran Engdahl. Handbook of Giant Magnetostrictive Materials. Academic Press, 2000.

4. Frederick V. Hunt. Electroacoustics: The Analysis of Transduction, and its Historical Background. Acoustical Society of America, 1982.

5 R. Meyer, J. Slaughter, “Galfenol Transducer Technology”, Emes, Iowa, Etrema Products, October 5, 2005.

6 NDRC Summary Technical Report, Division 6, Volume 13, (1946), The Design and Construction of Magnetostrictive Transducers.

7. S. Meeks, R. Timme, Acoustic Resonance of a Rare Earth Iron Magnetostrictive Rod in the Presence of Large Eddy Currents, IEEE Transactions on Sonics and Ultrasonics, Vol. SU-27, No. 2, March 1980.

8. L. Faidley et al, Terfenol-D elasto-magnetic properties under varied operating conditions using hysteresis loop analysis, SPIE Symposium on Smart Structures, March 1998 Paper No. 3329-92.

9. R. Kellogg, A. Flatau, Blocked force investigation of a Terfenol-D transducer, SPIE Symposium on Smart Structures, March 1999, Paper No. 3668-19.

10. IEEE Standard on magnetostrictive materials: Piezomagnetic nomenclature, IEEE Std 319-1990, February 1991.

11. H. C. Oersted, Experiment on the effects of a current on the magnetic needle, Annals of Philosophy, 16, 1820.

12. R. Woollett, Sonar Transducer Fundamentals, Naval Underwater Systems Center, 1987.

13. F. Grover, Inductance Calculations: Working Formulas and Tables, Dover Publications, 1973.

Page 88: MEASUREMENTS ON GALFENOL MATERIAL AND TRANSDUCERS

79

14. Correspondence with Julie Slaughter and Eric Summers of Etrema, Inc. via telephone and email from October 2005 to December 2008