27
Measurements and Instrumentations Lecture 3: Digital Voltmeters and Frequency Meters Dr. Haitham El-Hussieny Electronics and Communications Engineering Faculty of Engineering (Shoubra) Benha University November 2018 Dr. Haitham El-Hussieny Measurements and Instrumentations 1 / 27

Measurements and Instrumentations Engineering/823...Measurements and Instrumentations Lecture 3: Digital Voltmeters and Frequency Meters Dr. Haitham El-Hussieny Electronics and Communications

  • Upload
    others

  • View
    8

  • Download
    2

Embed Size (px)

Citation preview

  • Measurements and InstrumentationsLecture 3: Digital Voltmeters and Frequency Meters

    Dr. Haitham El-Hussieny

    Electronics and Communications EngineeringFaculty of Engineering (Shoubra)

    Benha University

    November 2018

    Dr. Haitham El-Hussieny Measurements and Instrumentations 1 / 27

  • Lecture Outline:

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 2 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 3 / 27

  • Introduction:

    Two types will be covered: Ramp-type andDual slope Integrator DVMs.

    Digital voltmeters (DVM) are essentiallyanalog-to-digital converters with digitaldisplays to indicate the measured voltage.

    Digital Voltmeter Basic Block Diagram

    Dr. Haitham El-Hussieny Measurements and Instrumentations 4 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 5 / 27

  • Ramp Type Digital Voltmeters:

    A ramp signal is generated.

    the comparator compares theinput Vi with the ramp Vr.

    V1 =

    {1, if Vi ≥ Vr0, if Vi < Vr

    }

    If the comparator output V1 ishigh, the counting circuit willcount the pulses from clockgenerator.

    If the output V1 is low, thecounting will stop.

    Npulses ∝ Vi .The value of Vi will be displayedby the end of the ramp signal.

    Ramp type DVM block

    Dr. Haitham El-Hussieny Measurements and Instrumentations 6 / 27

  • Ramp Type Digital Voltmeters:

    Ramp type DVM block DVM waveform

    Dr. Haitham El-Hussieny Measurements and Instrumentations 7 / 27

  • Ramp Type Digital Voltmeters:

    The use of the Latch:The latch isolates the display from the counting circuit during the t1.

    It will connect the display to the counting circuit at the rising edge of the comparator output.

    Ramp type DVM block

    Dr. Haitham El-Hussieny Measurements and Instrumentations 8 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 9 / 27

  • Dual Slope Digital Voltmeters:

    Dual Slope DVM block

    Limitations of Ramp type DVM

    The ramp type DVM requiresprecise ramp voltage andprecise time periods. (Notaccurate)

    The Dual-slope-integrator DVMeliminates this requirement.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 10 / 27

  • Dual Slope Digital Voltmeters:

    Dual Slope DVM block

    An integrator (e.g. capacitor) iseither charged negatively from Vi ordischarged at a constant rateaccording to the control signal.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 11 / 27

  • Dual Slope Digital Voltmeters:

    Dual Slope DVM block

    The control signal is derived from theclock generator and a frequencydivider.

    During the charging time t1, theintegrator is charged to Vo thatdepends on Vi .

    During the discharging, the integratoris discharged in constant rate induration t2 that depends on Vo andhence on Vi .

    A voltage comparator is used aszero-crossing-detector to output highif integrator voltage is lower thanzero.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 12 / 27

  • Dual Slope Digital Voltmeters:

    Dual Slope DVM blockDVM waveform

    Dr. Haitham El-Hussieny Measurements and Instrumentations 13 / 27

  • Dual Slope Digital Voltmeters:How the Dual slope integrator DVM eliminates the need for accurate timing ?

    (1) During charging:

    Vo = −Vit1

    (2) During discharging:

    Vo = Kt2 K is constant

    So,

    Vi = −Kt2t1

    Thus the input voltage measurement is notdependent on the clock frequency, but depends on

    the ratiot1t2

    .DVM waveform

    Dr. Haitham El-Hussieny Measurements and Instrumentations 14 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 15 / 27

  • DVM Range Changing:

    The attenuation circuit is used toselect the range of input voltage:

    if Vin ≤ 1.999 V , the input isapplied directly on thecomparator.

    if 1.999 V < Vin ≤ 19.99 V , theinput is attenuated and thedecimal point is changed.

    and so on for19.99 V < Vin ≤ 199.9 V

    DVM Range ChangingDr. Haitham El-Hussieny Measurements and Instrumentations 16 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 17 / 27

  • Digital Voltmeter Accuracy:

    Accuracy in DVMs:

    Digital voltmeter accuracy is usually stated as:

    ±(0.5% rdg + 1digit)

    where 1 digit refers to the extreme right (least significant digit) that depends on the range.

    Example

    If the accuracy is ±(0.5% rdg + 1 digit)What is the maximum error of reading1.800 V on:(1) the 2 V scale.(2) the 20 V scale.

    Solution:(1) error =±[0.5%× 1.8V + 0.001] = ±0.01V(2) error =±[0.5%× 1.8V + 0.01] = ±0.019V

    Dr. Haitham El-Hussieny Measurements and Instrumentations 18 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 19 / 27

  • Types of Digital Multi-meters:

    Hand-held Multimeter Bench-type Multimeter

    Dr. Haitham El-Hussieny Measurements and Instrumentations 20 / 27

  • Types of Digital Multi-meters:Clamp Meters:

    Advantage:

    It is an electrical device having two jawswhich open to allow clamping around anelectrical conductor. This allows tomeasure electric current throughconductor, without having to makephysical contact with it, or to disconnectit for insertion through the probe.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 21 / 27

  • Types of Digital Multi-meters:How Do Clamp Meters Operate ?:

    A current transformer(CT) used to pick up magnetic flux generated as a result of currentpassing through a conductor.A secondary winding generates a current by electromagnetic induction that is proportionalto the primary current.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 22 / 27

  • Table of Contents

    1 Introduction.

    2 Ramp Type Digital Voltmeters.

    3 Dual Slope Digital Voltmeters.

    4 DVM Range Changing.

    5 Digital Voltmeter Accuracy.

    6 Types of Digital Multi-meters.

    7 Basic Digital Frequency Meters (DFM).

    Dr. Haitham El-Hussieny Measurements and Instrumentations 23 / 27

  • Introduction:

    1 Frequency of a periodic signal is defined as: the number of occurrences of a repeatingevent per unit time = Number of signal’s cycles per one second.

    Principle of Frequency Meters:

    To measure the frequency of a certain periodicsignal, the waveform of that signal is used totoggle a counter for a certain fixed time.The number of counted cycles per unit timeindicates the signal frequency.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 24 / 27

  • Basic Digital Frequency Meters (DFM):

    Register

    BCD-to-seven-

    segment drivers

    AND

    gate

    Amplifier/

    attenuator

    Display

    Time

    base

    Wave

    shaping

    circuit

    Register reset

    Latch

    Flip flop

    Q

    Latch trigger

    Amplified/

    attenuated input

    Shaped waveform

    _Q

    1 s 1 sTime

    base

    output

    _Q output

    Q output

    Register reset

    Counting

    No

    counting

    Latch triggered

    To correct display

    Input

    Basic Digital Freq. Meter

    The basic DFM consists of:

    Accurate timing source(time base) withfrequency of 1 Hz.

    Digital counting circuitto count the inputwaveform cycles.

    Amplifier/Attenuationcircuit to amplify orattenuate the inputsignal.

    Waveform shapingcircuit to convert theinput signal to squarewave.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 25 / 27

  • Basic Digital Frequency Meters (DFM):

    The operation of basic DFM:

    1 The input signal is amplified or attenuated as necessary.

    2 The input signal is converted to a square wave and is fed to one terminal of the AND gate.

    3 The time base signal with 1 Hz. freq. is fed to a flip-flop.

    4 The flip-flop changes its state at each falling-edge of the time base. It divides thefrequency by 2 giving a high on the Q terminal for 1 s and a low for another 1 s. Theterminal Q̄ is an inverted version of Q.

    5 One terminal of the AND gate is fed from the flip-flop Q output and the other terminal isfed from the shaped input signal. So, the counter circuit will count the input pulses forthe duration of 1 s. (Frequency).

    6 The counter will reset to zero at each negative (falling) edge of the Q̄.

    7 The latch will isolate the counting from the display during the first 1 s and will updatethe display on the rising edge of Q̄ output.

    Dr. Haitham El-Hussieny Measurements and Instrumentations 26 / 27

  • End of Lecture

    Best Wishes

    [email protected]

    Dr. Haitham El-Hussieny Measurements and Instrumentations 27 / 27

    Introduction.Ramp Type Digital Voltmeters.Dual Slope Digital Voltmeters.DVM Range Changing.Digital Voltmeter Accuracy.Types of Digital Multi-meters.Basic Digital Frequency Meters (DFM).