Measurement based MU verification tool for lung SBRT

Lihong Qin, PhDDepartment of Radiation Oncology

Minneapolis VA Health Care System (MVAHCS)

Page 2: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Disclosure

• I’m a contract employee working at MVAHCS• Have no conflict interest

Page 3: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Outline

• Introduction and Background• Lung Insert (phantom)• Measurement and Result• Discussion

Page 4: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Introduction• SBRT program at Minneapolis VA:

– Only linac based lung SBRT .– Follow guidelines:

o VA National Radiation Oncology Program (NROP)o ACRo AAPM

– SBRT training from the University of Texas Southwestern Medical Center (Robert Timmerman, MD, Timothy Solberg, PhD, Paul Medin, PhD)

• We are ACR accredited facility • VA NROP contracted IROC Houston (RPC) to do a

“On-site Audits” every two years

Page 5: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Lung SBRT at MVAHCS

Immobilization /localizationBody Fix Philips big bore CT/ 4D

Planning Treat/imagingMonitor unit verification

Page 6: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

TPS independent assessment• RPC lung phantom study for IMRT plan

Page 7: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

TPS independent assessment

• RPC head/neck phantom: excellent agreements

• IROC on-site audit measurements (in water)all data IROC measured /our TPS calculation are in agreement

• Trust our TPS more than other calculations available in our clinic

Page 8: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Lung SBRT at MVAHCS

• 3D conformal noncoplanar• Field size: collimator size no

smaller than ~3.5 cm• Planning: average CT, 6x

photon, cc convolution, 2mm dose grid, density correction: heterogeneous

• Monitor Unit verification: program

Page 9: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

MU verification for lung SBRT TG 114 guideline

• Simple effective depth puts our SBRT MU constantly above action level

• We try everything to bring them within 5% /

– TG 114, TG 71, TG101,TG 65– Calling friends – Staring at the 20 something % differences

Page 10: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Lung Insert Phantom

• We have an ArcCHECK with PMMA (Acrylic) insert and ion chamber holder

PMMA insert Diameter is 15 cmion chamber holder diameter 3cm

CC 13 chamber

Find a better way for MU verification of SBRT

1.18g/cc0.3~0.5 g/cc

Page 11: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Make Lung Insert Phantom

• Tissue characterization phantom specification from Gammex:– LN-300 lung 0.31 g/cm3

– LN-450 lung 0.47 g/cm3

• Western red cedar: nominal density 0.38 g/cm3

• Get the cedar

Page 12: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Make Lung Insert Phantom

Page 13: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Make Lung Insert Phantom

• Machine Shop

This is not my garage

Page 14: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Making Lung Insert Phantom

Steve Morin making the phantom in “his”shop at the VA He can make lots of things

Page 15: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

LungInsert phantom

LungInsert

Page 16: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

ArcCHECK with LungInsert

Page 17: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

ArcCHECK with LungInsert

• Lung insert: 15 cm• 3 cm diameter PMMA

chamber holder

PMMA: 1.18 g/ccLung Insert: 0.33 g/cc

Good geometry/density with diodes and ion chamber

Page 18: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Measurement• Commissioned for clinic use : (save CT image as

QA phantom, copy patient plan to QA phantom)• 5 cases measured using ArcCHECK/LungInsert for

commissioning:– 4 with planned gantry/collimator angles but with couch

angle 0.– 1 case with all planned angles, including couch

• Obtained both point dose at Iso-center (Rx point) and dose distribution with Gamma analysis

• Point dose is calculated per beam.• 2 cases measured for clinic MU verification

Page 19: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Results: exampleMUCheck with effective depth

Page 20: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Results: exampleArcCHECK with LungInsert Measurement

Page 21: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Results

• 7 SBRT plans tested • ~ over 70 dose points measured• All below 5%, most within 2%

between TPS/measurement • 2 points above 4% in 1 case

(happens to be the case measured with all planned angles, including couch angles)

Page 22: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Dose distribution: the case measured with all planned angles

Page 23: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

ResultsGamma Analysis 3%/3mm

% passed

91

92

93

94

95

96

97

98

99

100

1 2 3 4 5 6 7

Series1

7 case results % passed from 94.4 to 99.2

Page 24: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Discussion

• The LungInsert for ArcCHECK can be used as measurement based MU verification tool for lung SBRT

• It’s good to have the Insert made as two halves. Easier In/Out from the ArcCHECK cavity.

Page 25: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Discussion

• Pros: – The results obtained from our 7 cases all below

AAPM guideline action levels. – Also get a dose distribution gamma analysis – Make it in your garage

• Cons:– Machine time– Measurement? I do the measurement now

Page 26: Measurement based MU verification tool for lung …chapter.aapm.org/nccaapm/z_meetings/2015-04-17/04_Agenda...TG 114 guideline • Simple effective depth puts our SBRT MU constantly

Acknowledgement

• Special thanks go to Steve Morin who made the LungInsert

• Thanks rad onc and biomed at VA• Questions?

Thank you