17
Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10489600 Rev F

Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000

Ref. 10489600 Rev F

Page 2: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural
Page 3: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

i

SAFETY SUMMARY

CAUTIONS Federal Law in the U.S.A. and Canada restricts this device to sale by or on the order of a licensed practitioner. The Vacuum nipple, the vacuum adapter, the vacuum shutoff valve, and the scavenging control valve are not autoclavable (Figure 2). When autoclaving the Matrx Scavenger, Parker recommends following the specific protocol of the service provider or the sterilizer manufacturer. You can use temperatures consistent with silicone, Ultem (polyetherimide), and polysulfone, that is, temperatures in the range 120°C (248°F) to 134°C (274°F). Do not autoclave “white” single-use nasal hoods, the vacuum nipple, the vacuum adapter, the vacuum shutoff valve, or the scavenger control valve.

Page 4: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

ii

Table of Contents

Notes and Contact Information .........................................................................................iv Introduction .........................................................................................................................1 Intended Use ........................................................................................................................1 CE Notice .............................................................................................................................1

Section 1 Description ..........................................................................................................................2 Section 2 Installation ...........................................................................................................................3 2.1 Fresh Gas Tubing .................................................................................................................3 2.2 Vacuum Tubing ....................................................................................................................3 2.3 Matrx MDM® and RA® Analgesia Heads ............................................................................4 Section 3 Operation .............................................................................................................................4 Section 4 Changing the Hood ............................................................................................................5 Section 5 Cleaning Instructions .........................................................................................................6 Section 6 Assembly after Cleaning ....................................................................................................6 Section 7 Replacement Parts .............................................................................................................7

Page 5: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

iii

Table of Illustrations Figure 1. Matrx Scavenger ..................................................................................................................2 Figure 2. Shutoff Connection Options .................................................................................................3 Figure 3. Direct Connection to HVE Piping System ............................................................................3 Figure 4. Installing the Retaining Clip ..................................................................................................4 Figure 5. Removing the Y from the Hood ............................................................................................5 Figure 6. Correctly Seating the Cone in the Hood ...............................................................................5 Figure 7. Assembling the Cone ...........................................................................................................6

Page 6: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

iv

Notes and Contact Information Date Purchased Dealer Name Dealer Phone Number Notes:

Page 7: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

1

Introduction

CAUTION

Federal Law in the U.S.A. and Canada restricts this device to sale by or on the order of a licensed practitioner. Intended Use The Matrx Scavenger is intended to deliver and remove oxygen and analgesic gas to the patient’s upper airway. It is designed for use by medical professionals trained in the administration of conscious sedation.

CE Notice

The symbol on this product indicates compliance with the Medical Device Directive of the European Community. A "Declaration of Conformity" in accordance with the Directive has been made and is on file. European Communities should contact the Authorized Representative listed below regarding any Medical Device Directive (MDD) inquiries. Direct inquiries may be made to the following addresses:

Porter Regulatory Affairs:

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000

EU Authorized Representative: Parker Hannifin Ltd Instrumentation Products Division Riverside Road, Pottington Business Park Barnstaple, EX 31 1NP, England Phone: +44 (0) 1271-313131 Fax: +44 (0) 1271-373636

Page 8: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

2

Section 1 Description The Matrx Scavenger is an autoclavable nitrous oxide scavenging system consisting of four parts (see Figure 1): • a Matrx nasal hood (small, medium, or large)

• a scavenging cone system

• a vacuum tubing assembly

• a fresh gas tubing assembly The design and the small number of parts precludes missassembly during cleaning and mask changes. Figure 1. Matrx Scavenger

The vacuum ports and a one-way exhalation valve make up the scavenging cone system. The vacuum tubing assembly consists of an integral Y-piece and vacuum hose. The fresh gas tubing comprises an integral Y-piece, a fresh gas tube, a slide, and a corrugated, 22 mm fresh gas connector. All components shown in Figure 1 (except single-use nasal hoods) are autoclavable.

Page 9: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

3

Section 2 Installation 2.1 Fresh Gas Tubing Connect the Matrx Scavenger fresh gas tubing to the fresh gas outlet of the analgesia flowmeter. The 22 mm fresh gas connector of the Matrx Scavenger connects directly to a Matrx analgesia flowmeter or to one from a different manufacturer. 2.2 Vacuum Tubing Vacuum tubing lacks corrugations on the hose. A white, hard, threaded fitting, called the vacuum nipple, attaches to the tubing’s termination. The vacuum nipple connects to the vacuum shutoff valve or to the optional scavenging control valve. The shutoff valve or the scavenging control valve attaches to the vacuum adapter. See Figure 2.

Figure 2. Shutoff Connection Options

CAUTION The vacuum nipple, the vacuum adapter, the vacuum shutoff valve, and the scavenging control valve are not autoclavable (Figure 2).

The four methods to connect the vacuum adapter to a vacuum source are: 1. Direct connection to a high volume evacuation

(HVE) line. The leading diameter of the vacuum adapter is sized to fit the HVE line inlet. Remove the HVE tip and insert the vacuum adapter.

2. Vacuum source in a junction box. Manufacturers

provide junction boxes that have at least two vacuum terminals. You must use dealer-supplied hose and/or adapters to make this connection.

3. Tee into an HVE line. The vacuum adapter

diameter is stepped to accept either 1/2” or 7/16” inside diameter hose. The properly sized tee (dealer supplied) can be inserted into the HVE line ahead of the tip. Using the appropriately sized tubing supplied by the dealer, connect the scavenger to the free leg of the tee.

4. Direct connection to an HVE piping system (see

Figure 3)

a. PVC Plumbing. The scavenger vacuum shutoff valve has 1/4 NPT female pipe threads. From a plumbing supply house, purchase the proper saddle (to fit the PVC pipe), bushing, and nipple to reduce the 1/4 NPT. Following the plumbing instructions, install the saddle and the bushings. Remove the vacuum adapter from the scavenger’s vacuum shutoff valve and thread the purchased nipple into the vacuum shutoff valve. Connect the nipple to the bushing.

b. Hard copper plumbing. From a plumbing supply house, obtain a copper flush bushing with 1/4 NPT female pipe threads. Measure the inside diameter to determine the proper drill size. At the position where you plan to place the bushing, center punch the main copper pipe. Drill an 1/8” diameter hole through one wall only. Use successively larger drills until the hole’s diameter matches the outside diameter of the flush bushing. Insert the bushing, with the nipple in place, and attach the scavenging shutoff valve to the nipple.

Figure 3. Direct connection to HVE Piping System

Page 10: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

4

2.3 Matrx MDM® and RA® Analgesia Heads If you purchase the scavenger with the shutoff valve, the supplied clip is used to support the valve at the back of the MDM® or the RA® analgesia head. See Figure 4. 1. From the rear of the head, remove the lower left

hand cover retaining screw. 2. Insert the screw through the mounting hole in the

retaining clip. 3. Install the screw, securing the retaining clip to the

machine head. 4. Snap the scavenger vacuum shutoff valve into the

retaining clip. 5. Using dealer-supplied vacuum hoses, make the

vacuum connection as previously described. Section 3 Operation 1. Check the system before operation. Ensure that:

• The fresh gas tubing is connected to the analgesia flowmeter.

• The vacuum system is connected to the vacuum

shutoff valve or the scavenger control valve.

• The vacuum shutoff valve or the scavenger control valve is connected to the vacuum system.

• The vacuum system is operating.

2. Open the vacuum shutoff valve completely.

Figure 4. Installing the Retaining Clip

Or, open the scavenger control valve, adjusting the vacuum within the green band. Note that the line between red and green is in the American Dental Association’s (ADA) recommended 45 LPM scavenging flow. Any adjustment above this level improves the scavenging efficiency.

3. Operate the analgesia machine, according to the

manufacturer’s instructions. 4. At the completion of the procedure, shut off the

vacuum shutoff valve or the scavenger control valve.

Page 11: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

5

Section 4 Changing the Hood 1. Hold the nasal hood in one hand. 2. With your other hand, gently pull the scavenging

cone from the hood. 3. Disconnect the long sections of the gas Y and the

vacuum Y from the hard plastic cone. 4. Pinching the gas Y just behind the hose connector,

work the hood free from the gas Y. Repeat for the vacuum Y. These procedures leave the hose connectors connected to the Ys. See Figure 5.

Figure 5. Removing the Y from the Hood 5. Remove the hood from the scavenging cone. 6. Turn the new hood upside down. Place your thumbs

inside the hood.

Figure 6. Correctly Seating the Cone in the Hood 7. Gently stretch the hole in the hood over the retaining

tabs on the scavenging cone. When the cone is seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6.

8. Note the natural bend of the long tubing sections of

the gas Y and the vacuum Y. Use this bend as a guide to determine a natural fit to the left and the right sides of the hood and cone assembly.

9. Pinching the hose connector in the gas Y, insert it

into the hood fully. Repeat for the vacuum Y. 10. Connect the long sections of the Y pieces to the two

ports on the scavenging cone.

Page 12: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

6

Section 5 Cleaning Instructions You can autoclave all Matrx Scavenger components, except the components of the vacuum shutoff valve and the scavenging control valve. To ensure proper cleaning, follow the directions below.

CAUTION

When autoclaving the Matrx Scavenger components, Parker recommends following the specific protocol of the service provider of the sterilizer manufacturer. You can use temperatures consistent with silicone, Ultem (polyetherimide), and polysulfone, that is, in the temperature range from 120°C (248°F) to 134°C (274°F).

1. Remove the vacuum Y from the hood and cone. Remove the tubing from the vacuum nipple.

NOTE

Do not unthread the vacuum nipple from the shutoff valve or the scavenging control valve.

Section 6 Assembly after Cleaning

Figure 7. Assembling the Cone 1. Attach the exhalation valve to the exhalation valve

cage (see Figure 7). 2. With the exhalation valve facing the cone, insert the

assembled exhalation valve and exhalation valve cage into the cone.

3. Turn the cage counterclockwise. 4. Continue turning until the tabs engage and the pinch

bar snaps closed. When completely engaged, the cone and the cage cannot disengage.

2. Remove the fresh gas Y from the hood. Remove the

tubing from the fresh gas outlet of the gas mixing apparatus.

3. Wash each tube with a mild alkali detergent. 4. Wash the scavenging cone system with a mild alkali

detergent. Run a soft bristle brush around the inside of the cone to remove any debris. If necessary, you can detach the cone from the cage by bending the pinch bar back and turning clockwise (see Figure 7).

CAUTION

Do not autoclave “white” single-use nasal hoods, the vacuum nipple, the vacuum adapter, the vacuum shutoff valve, or the scavenger control valve.

5. Autoclave all parts not included in the CAUTION

above. Use autoclavable protocol. 5. Turn the hood upside down. Place your thumbs

inside the hood. 6. Gently stretch the hole in the hood over the retaining

tabs on the scavenging cone. When the cone is seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. (see Figure 6).

7. Note the natural bend of the long tubing sections of

the gas Y and the vacuum Y. Use this bend as a guide to determine a natural fit to the left and right sides of the hood and cone assembly.

8. Pinching the hose connector in the gas Y (Figure 5),

insert it into the hood fully. Repeat for the vacuum Y. 9. Connect the ling section of the Y pieces to the two

ports on the scavenging cone. 10. Attach the free end of the fresh gas tubing to the

analgesia head. 11. Attach the free end of the vacuum tubing to the

vacuum nipple on the vacuum shutoff valve or on the scavenging control valve.

Page 13: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

7

Section 7 Replacement Parts You can obtain the following replacement parts from Parker.

Matrx Part No. Description Size 91515201 Hose connectors (package of 2)

91515202 Vacuum tubing and Y assembly

91515203 Fresh gas tubing and Y assembly (including slide)

91515177 Replacement cone assembly

10489600 Instruction manual

91515199 Extension kit

91515158 Vacuum nipple (with small orifice)

91515159 Shutoff valve

91515096 Reusable hood Large

91515095 Reusable hood Medium

91515094 Reusable hood Small

See pages 8 & 9 for single-use hoods

Scavenging Control Valve (above)

91515083 For MDM® or RA®

91525109 For Digital MDM®

91515084 For Centurion Mixer

91515085 Adapter with Ohio Quick Connect

91515086 For cabinet mounting

Page 14: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

8

DynoMite Single-use Nasal Hoods (package of 12)

Matrx Part No. Description Size

91316496 Bubble gum Small

91316503 Bubble gum Medium

91316510 Bubble gum Large

91316497 Strawberry Small

91316504 Strawberry Medium

91316511 Strawberry Large

91316498 Orange Small

91316505 Orange Medium

91316512 Orange Large

91316499 Vanilla Small

91316506 Vanilla Medium

91316513 Vanilla Large

91316500 Plain Small

91316507 Plain Medium

91316514 Plain Large

91316501 Assorted scents Small

91316508 Assorted scents Medium

91316515 Assorted scents Large

91316516 Assorted scents Various

DynoMite Single-use Nasal Hoods (package of 24)

Matrx Part No. Description Size

91316461 Bubble gum Small

91316468 Bubble gum Medium

91316475 Bubble gum Large

91316462 Strawberry Small

91316469 Strawberry Medium

91316476 Strawberry Large

91316463 Orange Small

91316470 Orange Medium

91316477 Orange Large

91316464 Vanilla Small

91316471 Vanilla Medium

91316478 Vanilla Large

91316465 Plain Small

91316472 Plain Medium

91316479 Plain Large

91316466 Assorted scents Small

91316473 Assorted scents Medium

91316480 Assorted scents Large

91316481 Assorted scents Various

Page 15: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

9

DynoMite Single-use Nasal Hoods (24 in a decorative canister, right)

Matrx Part No. Description Size

91316482 Bubble gum Small

91316519 Bubble gum Medium

91316489 Bubble gum Large

91316483 Strawberry Small

91316520 Strawberry Medium

91316490 Strawberry Large

91316484 Orange Small

91316521 Orange Medium

91316491 Orange Large

91316485 Vanilla Small

91316522 Vanilla Medium

91316492 Vanilla Large

91316486 Plain Small

91316523 Plain Medium

91316493 Plain Large

91316487 Assorted scents Small

91316524 Assorted scents Medium

91316494 Assorted scents Large

91316495 Assorted scents Various

Page 16: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural
Page 17: Matrx Breathing Circuit And Scavenger Brochure · seated in the nasal hood properly, three segments of the cone base are visible inside the nasal hood. See Figure 6. 8. Note the natural

DATE: 4/30/09 ECN: 09-0090 DR: MAM CH: AP: SHT 17 OF 17 Scavenger Instruction NUMBER: 10489600-F REV.

Manual 10489600 F

REVISION HISTORY

Rev. Date Ecn # Description of Changes Dr Ch Appr D 4/30/09 09-0090 Updated to have Parker information. Created in

word document from PDF. MAM AEM SDL

E 3/11/10 10-0073 Changed Authorized Rep. For Ce Marked

Products From Heatherside-April Fay To Parker Hannifin Ltd Instrumentation Products Division

ANB RAR ALE

F 8/19/10 10-0235 Updated CE Mark ANB