4
APPLICATION NOTE >> NEUROSURGERY Mapping of the corticospinal tract and motor cortex during tumor resection Intraoperative Neuromonitoring Functional Neurosurgery Pain Treatment Neurological Diagnostics

Mapping Suction Probe by Raabe - inomed€¦ · Mapping Suction Probe by Raabe. 1 3 2 The combination of a surgical suction tube and a monopolar electrical stimu-lation probe combines

  • Upload
    others

  • View
    16

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Mapping Suction Probe by Raabe - inomed€¦ · Mapping Suction Probe by Raabe. 1 3 2 The combination of a surgical suction tube and a monopolar electrical stimu-lation probe combines

APPLICATION NOTE>> NEUROSURGERY Mapping of the cor ticospinal tract and motor cor tex during tumor resection

Intraoperative NeuromonitoringFunctional NeurosurgeryPain TreatmentNeurological Diagnostics

Mapping Suction Probe by Raabe

Page 2: Mapping Suction Probe by Raabe - inomed€¦ · Mapping Suction Probe by Raabe. 1 3 2 The combination of a surgical suction tube and a monopolar electrical stimu-lation probe combines

1 3

2

The combination of a surgical suction tube and a monopolar electrical stimu-lation probe combines suction during tumor resection with simultaneous continuous dynamic mapping of the corticospinal tract. Use of the Mapping Suction Probe makes it possible to

achieve maximal tumor removal minimising the possibility of damage to the corticospinal tract. The ability to resect a tumor down to low motor thresholds is a signi� cant re� nement of the classic subcortical mapping of the corticospinal tract.

1. Scenario preparation for the NeuroExplorer so� ware

The use of the mapping suction probe can easily be integrated into NeuroExplorer scenarios. Typically the subcortical mapping is performed in the Triggered EMG measurement window.

> Triggered EMG window settingsTo set the parameters and a train stimulation in the Triggered EMG window, the I-Check stimulation must be activated:

A. Open setup-window (right mouse click over the Triggered EMG window)B. Set HS as the stimulator typeC. Activate I-Check stimulationD. Set relevant stimulation parameters

2. Recommended stimulation parametersThe monopolar electrical stimulation is delive-red at the tip of the mapping suction probe. The sha� of the mapping suction probe is insulated to ensure that the electrical contact is restricted to the tip of the mapping suction probe only.Stimulation parameters and the routine application of the stimulation probe is identi-cal to the parameters of a standard monopolar stimulation probe.

> Stimulation parameters (*)

1 Stimulator type: HS Pulse width: 500 µs Max. stimulation current: 15 mA Stimulation frequency: 0,4 – 2 Hz (1Hz)

2 Stimulation output: 5 Max. Voltage: 80 V Pulse form: > negative for subcortical stimulation (cathodal stimulation) > positive for direct cortical stimulation (anodal stimulation)

3 Number of pulses: 5 (Train of � ve) ISI (Interstimulus Intervall): 4 ms

(*) Raabe A, Beck J, Schucht P, Seidel K: „Continuous dynamic mapping of the corticospinal tract during surgery of motor eloquent brain tumors: evaluation of a new method“ in Journal of Neurosurgery 03/2014.

Page 3: Mapping Suction Probe by Raabe - inomed€¦ · Mapping Suction Probe by Raabe. 1 3 2 The combination of a surgical suction tube and a monopolar electrical stimu-lation probe combines

> Cursor positioning

Following BEEP BOOP sound activation, a cursor is displayed on each channel. The amplitude value can be set with each cursor. The amplitude value is the trigger and determines the point at which the audio sound will be generated.The cursors should be set a� er the stimulation artefact and prior to the response signal. The amplitude of the cursors should be set just above the ambient noise level seen in the EMG signal.

3. Continuous audio feedback

The use of the mapping suction probe in conjunction with EMG delivers continuous audio feedback during electrical stimulation. The audio feedback available during the procedure provides an indication of the patient’s electrophysiological status and a response to anato-mical changes taking place as a result of tumor resection. The BEEP BOOP synthetic sound function can be activated in the sound setting menu of the EMG window as follows: A. Open the sound menu in EMG window

by clicking on the loudspeaker symbolB. Activate BEEP BOOP C. Position cursors in triggered EMG window

The amplitude setting of the detection threshold cursor is greater than the amplitude of the signal

The amplitude setting of the detection threshold cursor is smaller than the amplitude of the signal

The amplitude setting of the detection threshold cursor is greater than the amplitude of the signal and the detection threshold cursor is positioned prior to the response signal:

The amplitude setting of the detection threshold cursor is smaller than the amplitude of the signal and the detection thres-hold cursor is positioned behind the response signal:

HIGH-PITCHED SOUNDThe signal does not exceed the amplitude of the cursor.The so� ware detects the signal not as a muscle response.

HIGH-PITCHED SOUNDThe signal does not exceed the amplitude of the cursor.In this case the high-pitched sound is delivered when current con� rm is activated and there is no motor response allowing a continuation of resection.

HIGH-PITCHED SOUNDThe signal is always detected as an artefact.

LOW-PITCHED SOUNDThe so� ware detects the signal as a muscle response.

> Failure in setting and positioning of the cursors generates always a high-pitched sound:

Page 4: Mapping Suction Probe by Raabe - inomed€¦ · Mapping Suction Probe by Raabe. 1 3 2 The combination of a surgical suction tube and a monopolar electrical stimu-lation probe combines

inomed Medizintechnik GmbHIm Hausgruen 2979312 Emmendingen (GERMANY)

Tel. +49 7641 9414-0Fax +49 7641 [email protected]

Art.-Nr. 525 650Mapping Suction Probe with connecting cable and black counter electrode

More information and further accessories: www.inomed.com

Mapping Suction Probe – order information:

5. Method>> The suction tip is at any time on the same place where the resection is performed and enables continuous mapping.The method (*) is based on resection, where no MEPs will be triggered:

>> Mapping starts with 10 mA (ca. 10 mm distance to the corticospinal tract)>> Current intensity is proportional to the distance of the corticospinal tract (Rule of thumb: 1mm ≈ 1 mA)>> If no motor response is triggeredthe resection can be continued.

>> As soon as a motor response is triggered, the resection shall continue on a more distant position to the corticospinal tract. If a motor response is triggered on further distant positions, the current should be reduced in 2 mA steps.

>> These steps can be repeated until 5 mA is reached.

>> The tumor resection should be terminated by the surgeon considering the progress of the operation and appropriate current intensity of the cortical MEPs.

Raabe A, Beck J, Schucht P, Seidel K: „Continuous dynamic mapping of the corticospinal tract during surgery of motor eloquent brain tumors: evaluation of a new method“ in Journal of Neurosurgery 03/2014.(**) CST: corticospinal tract

D020

156

/ S

ubje

ct to

cha

nge

with

out p

rior n

otic

e ©

cop

yrig

ht b

y in

omed

12.

2015

ref 1

.1

The red connection cable of the stimulator is attached directly to the mapping suction probe. The mapping suction probe is connected to stimulators with 1.5 mm Touchproof connectors

The Mapping Suction Probe is connected to the IOM Systemvia the stimulation adaptor (Art.-Nr. 540511/Art.-Nr. 540510) in channel 5.

4. Mapping suction probe connection to the IOM System

> Counter electrode

> Mapping Suction Probe > Adaptor box

CST**

CST**

15 mA

9 mA

13 mA13 mA

7 mA

11 mA11 mA

7 mA7 mA

10 mA

13 mA

5 mA

CST**

Tumor

Safe

Safe

Acoustic Feedback

Acoustic Feedback

Tumor

Tumor

> Connection cable