26
1 Ultimate E / Ultimate RITTER DENTAL UNIT OPERATING INSTRUCTIONS V03.00.01EN Manufacturer: Ritter Concept GmbH Contact: Ritter America Jarekstr. 1 88400 Biberach Germany E-mail: [email protected]

Manual Operare - Ritter Ultimate

  • Upload
    yam39

  • View
    226

  • Download
    14

Embed Size (px)

DESCRIPTION

RITTER DENTAL UNIT ULTIMATE Operate manual for dental unit

Citation preview

  • 1

    Ultimate E / Ultimate

    RITTER DENTAL UNIT

    OPERATING INSTRUCTIONS

    V03.00.01EN

    Manufacturer: Ritter Concept GmbH

    Contact: Ritter America Jarekstr. 1 88400 Biberach Germany E-mail: [email protected]

  • 2

    CONTENTS 1. General information ...............................................................................................................4

    1.1 Symbols ..............................................................................................................................4 1.2 Purpose and use of the unit ..............................................................................................5

    1.2.1 Intended use.5 1.2.2 Classification of Equipment ..................................................................................5 1.2.3 Ambient conditions ................................................................................................6 1.2.4 Technical parameters.............................................................................................6 1.2.5 Guarantee ...............................................................................................................6

    1.3 Safety information ............................................................................................................6 1.4 Cleaning and disinfection .................................................................................................8

    1.4.1 Cleaning and disinfection instructions..8 1.4.2 Cleaning and disinfecting the instruments9

    2. Description of the unit ................................................................................................................9 2.1 Nameplate ..........................................................................................................................9 2.2Description of the main parts..........................................................................................10 2.3 Specifications ...................................................................................................................10

    3. General operations.................................................................................................................... 11 4. Operation of multifunction foot-control ................................................................................. 11 5. Operation of dental chair ......................................................................................................... 11

    5.1 Adjusting and setting chair positions ............................................................................ 11 5.1.1 Seat up/down: ....................................................................................................... 11 5.1.2 Backrest up/down................................................................................................. 11 5.1.3 Operating the programmed positions: ...............................................................12 5.1.4 Operating the rinsing position..12 5.1.5 Emergency position for the dental chair..12 5.1.6 Reset the dental chair12

    5.2 Safety devices...................................................................................................................12 5.3 Emergency stopping operation ......................................................................................13 5.4 Operating the headrest ...................................................................................................13

    5.4.1 Height adjustment................................................................................................13 5.4.2 Swiveling ...............................................................................................................13

    5.5 Mobile armrest (optional) ..............................................................................................13 6. Operating the dentists instrument board ..............................................................................13

    6.1 Dentists control console .................................................................................................14 6.1.1 Description of the buttons ...................................................................................14 6.1.2 Setting the function parameters..........................................................................15

    6.1.2.1 Mouthwash time setting...........................................................................15 6.1.2.2 Bowl rinsing time setting..15

    6.1.3 Operating the function buttons....15 6.1.3.1 Washing the rinsing bowl..15 6.1.3.2 Filling the cup and heating up the mouthwash..15 6.1.3.3 Operating light..15 6.1.3.4 X-ray image viewer..15 6.1.3.5 Optical fiber..15

  • 3

    6.2 Syringe .............................................................................................................................15 6.2.1 Components ..........................................................................................................15 6.2.2 Functioning ...........................................................................................................15 6.2.3 Replacing spout ....................................................................................................15

    6.3 Turbine .............................................................................................................................16 6.3.1 Connecting the handpiece ...................................................................................16 6.3.2 Functioning ...........................................................................................................16 6.3.3 Changing the dental burr ....................................................................................16

    6.4 Low speed air-driven handpiece 17 6.4.1 Connecting.17 6.4.2 Functioning....17 6.4.3 Changing the dental burr.....17

    7. Operating the assistants instrument board ...........................................................................18 7.1 Adjustment of height for assistants instrument board ...............................................18 7.2 Assistants control console ..............................................................................................18

    7.2.1 Layout of assistants control console ..................................................................18 7.2.2 Description of the buttons ...................................................................................18

    7.3 Operating the HV suction and saliva ejector................................................................19 7.4 Operating syringe ...........................................................................................................19

    8. Functioning of the hydro-group...............................................................................................19 8.1 Bowl and cup filler ..........................................................................................................19 8.2 water system ....................................................................................................................19

    8.2.1 Description............................................................................................................19 8.2.2 Filling the tank .....................................................................................................20 8.2.3 Disinfecting the water lines .................................................................................21

    9. Maintenance ..............................................................................................................................22 10. Troubleshooting.......................................................................................................................23

    10.1 Dental chair ...................................................................................................................23 10.2 Operating light ..............................................................................................................23 10.3 Instruments board ........................................................................................................24 10.4 Replacement of fuses.....................................................................................................24

    11. Technical data ..........................................................................................................................25 11.1 Water and air flow diagram .........................................................................................25 11.2 Electrical diagrams .......................................................................................................26

  • 4

    1. General information z These instructions describe the operation and maintenance procedures for the

    RITTER series dental unit. Carefully read and understand the operating instructions before using the unit in order to avoid incorrect operation and other damage.

    z These instructions apply to all versions of the RITTER Ultimate E/ Ultimate as well as all the accessories that may be provided. For this reason, the description of some parts may not correspond to your equipment.

    z Duplication and distribution of this manual require the prior consent of Ritter. z All information, specifications and illustrations contained herein are subject to change.

    Ritter reserves the right to make changes without prior notice. 1.1 Symbols Symbols used:

    General warning, caution, risk of danger

    Date of manufacture

    Manufacturer

    Manufactures serial number

    Authorized representative in the European community

    WARNING! This indicates a situation which may lead to injury to persons or damage to the unit and/or its

    parts if the information is ignored.

    NOTE!

    This indicates important information for the user and/or service personnel.

    Type B applied part

    Protective earth (ground)

    Power on

    Power off

  • 5

    Fragile; Handle with care.

    This side up

    Guard against damp

    Not to be stowed under other cargo

    Temperature limitations 1.2 Purpose and use of the unit

    1.2.1 Intended use z The dental units are intended for use in the stomatology department of a medical

    organization or other dental clinics for dental diagnoses, treatment and surgeries by trained dentists and/or dental technicians and assistants.

    z The dentists instrument board may be equipped with up to 5 instruments. z The assistants board may be provided with 2 suction tubes and 2 instruments. z The equipment bears CE mark CE0366 indicating its conformity with the provision of

    Council Directive 93/42/EEC concerning medical devices, and fulfills the essential requirement of Annex I of this directive

    1.2.2 Classification of Equipment z Type of protection against electric shock (Class I); z Degree of protection against electric shock (Type B) z Degree of protection: Ordinary equipment , not liquid preventive z Mode of operation: Continuous operation with intermittent

    Chair motor: The running with loading time is 6 minutes and the pause time is 54minutes.

    z Flammable gasses: Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide, where such gasses may accumulate in concentration (closed space)

  • 6

    1.2.3 Ambient conditions Ambient conditions for shipment and storage are required.

    z Temperature: -10- 70 z Relative humidity from 10 to 90% z Atmospheric pressure from 50kpa to 106kpa The place where the equipment is to be installed should meet the following requirements: z Temperature between 5 C and 40 C z Relative humidity from 30 to 80% z Atmospheric pressure from 86kPa to 106kPa 1.2.4 Technical parameters z Input Voltage: 220VAC z Frequency: 50/60Hz z Power Consumption: 900VA z Main fuse: 6A 250V z Heater: 24VAC, 120VA z Dental operating light: 12V AC. 50VA z DC motor: 24VDC z Water pressure: 200-400 kPa or 21-42 Bar z Air pressure: 500-550 kPa 1.2.5 Guarantee Ritter guarantees the safety of the operator and the reliability and high performance of the equipment. The guarantee will be avoided if the following precautions are not taken: z Observe the conditions specified in the guarantee certificate itself. z The unit must only be used by the following instructions contained herein. z The electric wiring in the room where the unit is installed must conform to the local

    specification (i.e. the regulations concerning the electrical wiring to be used in surgeries. The unit should be provided with a 31mm2 power cable).

    z The unit should be fixed installation. z Repairs, modifications, adjustments and any other operations which involve gaining access

    to the inside of the unit must only be carried out by Ritter technicians or authorized technicians.

    z Replacements for any spare parts and instruments on the unit should be provided by Ritter otherwise the performance and safety may be affected.

    1.3 Safety information

    WARNING! Installation of this unit should be carried out by engineers of Ritter or persons trained and authorized by Ritter.

    WARNING! Dont place the unit on shaky floor, cart and shelf or it may tumble over and badly damaged. Dont install the unit near radiators or heaters. Dont connect unit with non- compatible wires which maybe influence performance, even cause fire or electric shock.

  • 7

    WARNING! This unit is for use by authorized and fully trained personnel only (doctors or

    assistants).

    WARNING! Before leaving the surgery, shut off the water supply, air supply and the main power.

    WARNING! The unit should always be supervised when turned on or prepared for start-up. In particular it should never be left unattended when under aged persons, persons without the required capabilities or unauthorized persons in general are nearby.

    NOTE! The manufacturer shall not be held responsible for any damage or injury caused by mishandling, neglect or improper use. z Dental chair WARNING!

    The maximum allowable load for the dental unit is 135kg. Never exceed this limit. z Tray support WARNING!

    The maximum loading of the table tray support attached to dentists instrument board is 1.5kg, which should never be exceeded. z Patients with pacemakers and/or hearing aids. WARNING!

    When treating patients with pacemakers and/or hearing aids, always consider the effects the instruments used may have on them. z Replacing the dental burrs. WARNING!

    Only replace the dental burrs when the handpieces have completely stopped otherwise the cartridge may be damaged and the dental burrs may cause personal injury.

    WARNING! Check the clamps for condition every day by making sure the dental burr is firmly secured to the instrument before starting work. z Cleaning and disinfections

    WARNING! Always switch off the dental equipment before performing any cleaning or disinfecting procedures on the external parts.

  • 8

    1.4 Cleaning and disinfection Due to the many kinds of medicaments and chemicals used in the dental surgery, upholstery or paintwork surfaces as well as plastic parts may become damaged. Since damage to surfaces caused by these products depends largely on the time taken to react, it is imperative that the places affected be immediately wiped with a damp cloth. The external parts of the equipment should be cleaned and disinfected by using a product which has proved to be effective against bacteria, fungi and viruses (including HBV and HIV). However, as the active elements in the disinfectants and sterilizing products have aggressive effects, we recommend using products which contain no more than the amounts shown below of the following products: Ethanol 96% = max. 40g per 100g disinfectant Propanol = max. 35g per 100g disinfectant 1.4.1 Cleaning and disinfection instructions

    WARNING! Always switch off the dental unit before performing any cleaning or disinfecting procedure on the external parts.

    WARNING! Make sure during cleaning that fluids do not penetrate into the equipment since they may cause malfunction.

    WARNING! In general, use only disinfectants which are approved for use on painted surfaces and plastic surfaces in dental medicine. Procedure Cleaning and disinfection are two closely related operations and should be performed together. z Cleaning: take a new piece of disposable paper or sterile gauze dampened with the

    chosen cleanser and wipe the dirty parts z Disinfection: dampen clean paper or gauze with the chosen disinfectants such as

    alcoholic disinfectants over the parts to be disinfected again. Leave to work for about 10 minutes and then remove the residue with dry disposable paper or sterile gauze.

    If the parts are extremely dirty or contaminated, conduct a pre-disinfection before the procedure of cleaning. That is, using the paper or gauze with plenty of cleanser, generously apply the disinfectant to the contaminated parts in order to cover them thoroughly. Wait for the time recommended by the manufacturer to allow the disinfectant to eliminate the bacteria and remove the organic deposits.

  • 9

    1.4.2 Cleaning and disinfecting the instruments

    WARNING! Before sterilizing, read and understand the instructions supplied with each

    Handpieces Cleaning and disinfection z Operate as the instruction of the motor provided with the motor stated. Syringe Cleaning z Use soft disposable paper dampened with detergent/disinfectant. Sterilization:

    z Syringe handles and spouts at up to 135C2Barin a water steam autoclave. Minimum length of treatment: 15 minutes.

    Suction tube and Suction tube handle Cleaning z Use soft disposable paper dampened with detergent/disinfectant. Sterilization:

    z Suction tube and Suction tube handle at up to 135C2Barin a water steam autoclave. Minimum length of treatment: 15 minutes.

    2. Description of the unit 2.1 Nameplate The nameplate is located on the right side of hydro-group. It contains the following data: z Name of equipment z Model z Rated voltage z Rated frequency z Rated power z Operation mode z Name and address of Manufacturer z Serial number z Date of manufacture

  • 10

    2.2 Description of the main parts 7 8 9 10 11 12 13 14

    1 2 3 4 5 6

    1 Patients chair 8 Assistants control console

    2 Floor box cover 9 Saliva ejector tube

    3 Turbine 10 Rinsing bowl 4 Electric motor r 11 Hydro-group 5 Scaler 12 Instrument tray arm

    6 Dentists control console 13 Instrument tray

    7 HV suction tube 14 Syringe 2.3 Specifications z Maximum length: 2100mm z Maximum width: 1780mm z Vertical movement of the instrument tray: 480mm z The rinsing bowl can be turned 90o outwards z The rinsing bowl can be turned 90o towards the patient z Minimum, seat height: 410mm z Maximum seat height: 830mm z The backrest can turn: 75o z The seat can turn: 8.5o

  • 11

    3. General operations Main switch Press the main switch to 1, the power indicator on the switch will be on. Then, the equipment is on.

    4. Operation of multifunction foot-control (1) Pedal: When a dental instrument is taken out (or pulled out), the unit is ready for work. At

    this time, the foot-control pedal is available for controlling the instrument. Depress the pedal, the instrument begins to work.

    (2) Left button: When a dental instrument is picked up or pulled out, it is an activating/deactivating water spray switch; when no dental instrument is picked up or pulled out, it implies the rinsing position of the dental chair. That is, when the instruments are in their operating position, press the button, the water spray can be activated or deactivated; when the instruments are on the holders, press the button, the dental chair will go to the rinsing position.

    (3) Right button: When a dental instrument is picked up or pulled out, it is a air blast button;

    when no dental instrument is picked up or pulled out, it implies the reset position of the dental chair. That is, when the instruments are in their operating position, air is delivered when the button is kept pressed, and the flow of air will be interrupted when the button is released; when the instruments are on the holders, the dental chair will automatic return.

    (4) Middle regulator (Dental chair control): The following movements are performed: Seat up and down, backrest up and down. To stop the backrest and seat from moving, release the button.

    5. Operation of dental chair 5.1 Adjusting and setting chair positions

    The dental chair can perform the following movements: Seat Up/Down and Backrest Up/Down, which can be operated from dentists control console, multifunction foot-control and assistants control console.

    5.1.1 Seat up/down: 1. Push the up/down control on the multi-functional pedal.

    Upwards for the chair lift up; downwards for down movement. 2. Conduct chair up/down by control console keys

    5.1.2 Backrest up/down The backrest can perform excellent movement with the seat slope. 1. Push the up/down control on the multi-functional pedal.

    The left for backrest moves up; and the right for backrest moves down. 2. Conduct backrest up/down button by control console keys.

  • 12

    5.1.3 Operating the programmed positions: The system allows one dentist to set two sets of programmed positions, namely,

    programmed position A and programmed position B by using the function setting button

    and buttons A and B on the dentists control console.

    Operate the dental chair to the desired position and keep button A pressed, then the current chair position is automatically saved as a programmed position after an indication of audible signal. Afterwards, press button A once, the dental chair automatically goes to the saved position A and the operating light turns on corresponding with the chair movement. Apply the same method to set programmed position B.

    5.1.4 Operating the rinsing position: Operate the dental chair to the desired position. Then keep the button on the dentists control console pressed, thus the current chair position is automatically saved as a memory rinsing position after an indication of audible signal. Press this button once, then the dental chair automatically moves to the saved memory rinsing position. And at the same time, water flows to the rinsing bowl and cup; the operating light goes out automatically. After that, press the button to move chair to the last work position and the operating light turns on automatically.

    5.1.5 Emergency position for the dental chair: Operate the dental chair to the desired position. Then keep the button on the dentists control console pressed, thus the current chair position is automatically saved as an emergency position after an indication of audible signal. Then shortly press the button of emergency position to bring the chair to the emergency position and at the same time the operating light turns on automatically. After that, press the button to move the chair to the last work position and the operating light turns on automatically.

    5.1.6 Reset the dental chair: Press the automatic return button shortly, the chair moves to the reset position and the operating light goes out.

    5.2 Safety devices

    WARNING! Maximum load of the chair is 135kg.

    WARNING! When moving the dental chair down, make sure there is enough space beneath it as there

    is the risk of being injured.

    z The bottom of the dental chair is provided with a safety device which stops the chair from moving down in the presence of obstacles. The chair will move up to allow the user to remove the obstruction.

    z A safety device is installed under the rinsing bowl in the hydro-group. If the rinsing bowl is turned towards patients, the chair will be limited within a safe height.

    z When the dynamic instrument is in use, the dental chair will be under safety protection and none movement is available.

  • 13

    5.3 Emergency stopping operation

    WARNING! To stop the chair from moving, press any one of the dental chair control buttons.

    5.4 Operating the headrest

    5.4.1 Height adjustment

    Push headrest in or pull it out to suit the patients height.

    5.4.2 Swiveling

    Turn clamping knob counterclockwise, then bring headrest to desired position. Turn clamping knob clockwise to clamp.

    5.5 Mobile armrest (optional)

    The right armrest can be moved down so as to allow the patient to easily get on and off the dental chair.

    6. Operating the dentists instrument board Operating the instruments z The syringe is always on. z All the other instruments, once removed, are operated by using the foot control. Instruments work independently from one another. An interlocking device prevents the instruments from working at the same time. The first instrument removed is operative. Those removed afterwards are inactivated by the interlocking device, which allows the user to replace a dental burr on one instrument while another is being used on the patient. Positioning the dentists instrument board The dentist instrument board can be moved in all directions. To adjust it in height or on a horizontal plane, take hold of the handle.

  • 14

    6.1 Dentists control console Layout of the dentists control console

    6.1.1 Description of the buttons Buttons Descriptions Buttons Descriptions

    Return to Rinsing position Multi-function setting

    Bring the chair into the emergency position

    Rinsing bowl

    Retrieval of programmed position A

    Cup filling (cold water)

    Retrieval of programmed position B

    Heating mouthwash water

    Automatic return travel

    Dental operating light(ON/OFF)

    Seat UP

    Optical fibbers (ON/OFF)

    Backrest UP X-ray image viewer (ON/OFF)

    Seat DOWN Backrest DOWN

  • 15

    6.1.2 Setting the function parameters

    6.1.2.1 Mouthwash time setting: Press function setting button ; the indicator lights on,

    and then enter the setting program. Press button cup filling button to start adding water. Press again, adding stops and the setting completes. At the same time the indicator light goes out.

    6.1.2.2 Bowl rinsing time setting: Press function setting button ; the indicator lights on,

    and then enter the setting program. Press button rinsing bowl button to start rinsing. Press again, rinsing stops and the setting completes. At the same time the indicator light goes out.

    6.1.3 Operating the function buttons 6.1.3.1 Washing the rinsing bowl: press the Rinsing bowl button ; the bowl will be

    washed for the set time. Press again, the bowl will stop being washed.

    6.1.3.2 Filling the cup and heating up the mouthwash: When Cup filling button is pressed, water will be delivered to the cup for the set time. Select the warm water by using the button for heating up the mouthwash. The indicator light flickers in the preheating process. If the indicator light keeps on, it indicates heating up has completed.

    6.1.3.3 Operating light: Perform the control to the operation light by using the button operation light on/off .

    6.1.3.4 X-ray image viewer: Control the viewer by using button X-ray image viewer on/off . The viewer will work under the set brightness level and automatically closes after 3 minutes. Press the button again to turn off the viewer.

    6.1.3.5 Optical fiber: When F.O instrument is being used, turn on/off the optical fiber by using button optical fiber .

    6.2 Syringe 6.2.1 Components Spout, handle, air flow control button, water flow control button 6.2.2 Functioning z Bring the instrument into its operating position. z Press water flow button to emerge water. z Press air blast button to spray air. z Press both water flow button and air blast button to spray.

    6.2.3 Replacing spout z Press the ring on the handle to release the spout, and then pull the spout out. z To fit in a new spout, keep the ring pressed and push the spout in. Then release the

    button.

    WARNING! Use the syringe provided with the unit. The use of others incompatible would bring risk.

  • 16

    6.3 Turbine The unit provides a control system for turbines. The rotation non-load speed should not be lower than 300000r/min when the input air pressure is 250KPa.

    6.3.1 Connecting the handpiece

    z Turbine with a MIDWEST coupling: fit the turbine (a) to the end of the hose and tighten

    the ring nut (b). z Turbine with a quick coupling: fit the adapter (d) to the end of the hose and tighten the ring

    nut (b). Then fit the turbine (c) to the adapter. 6.3.2 Functioning z Method of use: operation 5 min., pause 5 min. z Bring the instrument into its operating position. z To start the instrument use the foot control. z Use the related valves to control the flow of water and spray air. WARNING!

    Use the turbine provided with the unit. The use of others incompatible would bring risk. 6.3.3 Changing the dental burr

    Dental burr head with button: 1) To remove the dental burr, press the button right down so as to release the burr. 2) To fit in a new dental burr, keep the button pressed and push the burr right in. Then release the button.

    WARNING! Pull the dental burr to be certain it is securely fixed in place.

  • 17

    WARNING! The shank of the dental burr should have a diameter between 1.59 and 1.60mm (ISO 1797 type 3 standard) and a maximum length of 25 mm (ISO 6360-1 standard).

    WARNING! Use the turbine provided with the unit. The use of others incompatible would bring risk. 6.4 Low speed air-driven handpiece The unit provides a control system for the low speed air-driven handpiece. The user should use air-driven handpieces compatible with the unit. The rotation non-load speed is not lower 14000r/min when the input air pressure is 250KPa.

    WARNING! Use the handpiece turbine provided with the unit. The use of others incompatible would bring risk. 6.4.1 Connecting Fit the handpiece to the end of the hose and tighten the ring nut. 6.4.2 Functioning z Method of use: operation 5 min.; pause 5 min. z Bring the instrument into its operating position. z Use the foot control to start the instrument (see section 4). z Use the related valves to control water flow and air spray. 6.4.3 Changing the dental burr

    Dental burr head with button: 1) To remove the dental burr, press the button right down so as to release the burr. 2) To fit in a new dental burr, keep the button pressed and push the burr right in. Then release the button.

    WARNING! Pull the dental burr to be certain it is securely fixed in place.

    WARNING! The shank of the dental burr should have a diameter between 1.59 and 1.60mm (ISO 1797 type 3 standard) and a maximum length of 25 mm (ISO 6360-1 standard).

  • 18

    7. Operating the assistants instrument board 7.1 Adjustment of height for assistants instrument board The assistants instrument board can be positioned vertically at three levels. Pull assistants instrument board gently upwards until the desired position is reached (audible snapping-in). Lift it fully upwards until the lock is released and then move downwards to the horizontal position. 7.2 Assistants control console 7.2.1 Layout of assistants control console

    7.2.2 Description of the buttons

    Buttons Description Buttons Descriptions

    Seat UP

    Return to Rinsing position

    Backrest UP

    Rinsing bowl

    Seat DOWN

    Cup filling (cold water)

    Backrest DOWN

    Heating mouthwash water

    Bring the chair into the emergency position

    Automatic return travel

  • 19

    7.3 Operating the HV suction and saliva ejector The suction system starts operating once the tube is pulled out from the tube holder. To vary the suction rate, use the slider on the handle on the tube holder terminal.

    NOTE! When the tube is put back in the respective holder, the suction system will stop after approx. 2 seconds. This delay stop allows the suction tubes to be dried.

    7.4 Operating syringe The operation is the same as the section 6.2 is mentioned above.

    8. Functioning of the hydro-group 8.1 Bowl and cup filler The bowl is flushed automatically for the set time when: z The Rinsing bowl is pressed; z The Cup filling is pressed; z The water is filling the cup when the Cup filling is pressed. The mouthwash time and bowl rinsing time can be set by the operator. The operating methods are described in section 6.1.3.1 and 6.1.3.2. 8.2 Water system 8.2.1 Description z All the spray nozzles located on the dentists instrument board; z The syringe on the dental assistants instrument board. When the door switch is on, the tank will automatically be pushed to the front. The door is manually opened for Ultimate. The air pressure close/open lever that is located on the left side in the front of the hydro-group (visible when the door is open) is used to let air in or out of the tank. The by-pass lever that is located on the right side in the front of the hydro-group (visible when the door is open) is used to shut off the self-contained water system when tap water is to be delivered.

  • 20

    8.2.2 Filling the tank

    1Switch the door lever to open the front door; 2Switch air pressure lever to Pressure Relief position.

    NOTE! During this operation, the air under pressure in the tank will automatically be discharged into the atmosphere.

    3Remove the tank by turning it in a clockwise direction. Pour distilled water into the tank below maximum level (approx. 1.7 liters).

    4Re-install the tank by turning it anti-clockwise. 5Switch the air pressure lever back over to the Pressure Increase position. 6Switch the door lever to close the front door and return to normal operation.

    NOTE! 1. The water tank can only be opened after the air in it completely escapes.

    2. As the tank cap connects with air and water lines, use your left hand to fix the cap and your right hand to rotate the tank. Dont rotate the cap.

    3. We suggest using distilled water for treatment in order to avoid damage of the units pipeline and the cycle life of the instruments which may be caused by the tap water impurities.

  • 21

    8.2.3 Disinfecting the water lines With this disinfecting system, disinfectant is let into the water lines leading to the spray nozzles of all the instruments on the dentists board and the syringe on the assistants board. The disinfectant should be left to work for a certain amount of time inside the lines in order to disinfect them. To perform the disinfecting operation, proceed as follows:

    1. Keep the water bottle empty and switch the air pressure lever to Pressure Increase to draw away the filth and dry the tubes.

    2. Switch the air pressure lever to Pressure Relief. Remove the white distilled water bottle by turning it in a clockwise direction and install the pink bottle containing disinfectant.

    3. Make sure the water regulators for the spray nozzles, which are located at the bottom of the board, are open (otherwise no water- or not enough water- will come out).

    4. Take out the syringe and drain the water line. When air comes out of the line, it means it has been fully drained.

    5. Repeat the same procedure for the syringe on the assistants board and for the rest of the dynamic instruments.

    6. Switch the air pressure lever to Pressure Increase. 7. Operate the dental instruments until disinfectant has been sprayed out for seconds one

    by one.

    8. Keep the unit off for one or two hours.

    9. Switch the air pressure lever to Pressure Relief, remove the pink disinfectant bottle and reinstall the white distilled water bottle.

    10. Take out the syringe and drain the water line. When air starts to come out of the line it means that it has been fully drained.

    11. Repeat these operations to drain the disinfectants out of the lines leading the instruments. 12. Switch the air pressure lever to Pressure Increase. 13. Operate the instruments and wait until the distilled water starts to come out.

    NOTE! Allow the distilled water to come out for at least 6-7 seconds.

  • 22

    9. Maintenance

    NOTE! Proper maintenance can extend the cycle life of the unit! In order to keep the dental unit in proper working order it is essential to perform the routine maintenance operations as indicated in the following. z Maintenance of the instruments Maintenance instructions are supplied for each instrument.

    WARNING! We advise performing the maintenance operations for the instruments with the equipment turned off. z Check the voltage for the power supply at intervals. The air pressure and water pressure must

    conform to the requirements of this unit. And the pipes must be smooth. z Check the sensitivity for the movement of the dental unit parts and the control buttons. z At the end of each work day, use liquid germicide to clean ceramic bowl and spout. And clean

    bowl filter under running water and with detergents available on the market.

    WARNING! Never use acids or abrasive products.

    WARNING! Always wear gloves to perform this operation. z At the end of each work day, clean the suction filter. Pull out the suction filter and drop the dirty.

    Reinstall it after the cleaning and make sure the installment will not cause air leaks.

    WARNING! Always wear gloves to perform this operation. z Cleaning the air return line. Every month, check the filter of the oil collector. If necessary,

    replace the filter element. z Air filter: In order to guarantee the proper operation of the unit, an air filter is designed to locate

    at the air inlet so as to prevent the impurity in air. The filter collector must be cleaned or replaced timely to avoid abnormal function of the unit. Generally, the filter should be replaced every one month.

    z Water filter: In order to guarantee the proper operation of the unit, a water filter is designed to locate at the water inlet so as to prevent the impurity in water. After a long time operation, the dirty may jam the filter collector and the water flow may be obstructed. Therefore, the filter collector must be cleaned or replaced timely. Generally speaking, it should be done every one month. And if any one of the following case happened, the filter also should be cleaned or replace.

    1. The filter collector has been used for more than 1 month; 2. The filter collector is polluted; 3. The water coming out becomes turbid.

  • 23

    The way to replace the filter collector: open the hydro-group and disassemble the water filter. Then remove the shell of the filter by turning it anti-clockwise. Take out the seal ring and collector in turn. After cleaning or replacing, put them in the original place. Please take care of the seal to avoid leaks.

    z Working instruments For the maintenance operations, refer to the manual supplied with each instrument.

    10. Troubleshooting

    NOTE! This part of instructions is intended to be an aid to self-help. If faults cannot be localized with the aid of this troubleshooting scheme, an engineer specially trained or authorized by Ritter must be contacted for assistance. 10.1 Dental chair

    Fault Cause Rectification

    Nothing functions Main switch off Main fuse blown

    Switch on main switch Check main fuse and if necessary

    replace a fuse of the same kind

    Dental chair does not move up or down

    Safety device activated Motor damaged, loose

    connection or other causes

    Check safety devices and eliminate cause of shutoff Check the power lines and call technician

    The backrest does not move up or down

    Motor damaged, loose connection or other causes

    Check the power lines and call technician

    Movements are in disorder

    Chips damaged Absorb moisture and call technician

    10.2 Operating light

    Fault Cause Rectification Operating light doesnt function

    Fuse blown or lamp faulty

    Replace fuse or bulb

    Light doesnt concentrate

    Lamp was not installedto the focus

    Adjust the lamp

  • 24

    10.3 Instruments board

    Fault Cause Rectification Poor air flow at the handpiece

    Air lines distorted or air filter jammed

    Adjust the air line or clean the filter

    Poor spray at the syringe

    Improper adjustment of the throttle valve

    Adjust the throttle valve

    Suction tube doesnt work

    Solenoid valve damaged

    Call technician to replace it

    Poor suction of saliva elector

    O-ring damaged or water pressure was not high enough

    Replace the O-ring or adjust the water pressure

    Poor suction of HV suction

    O-ring damaged or water pressure was not high enough

    Replace the O-ring or adjust the water pressure

    No warm water Heater damaged Call technician to replace it 10.4 Replacement of fuses

    Turn off the power supply, and open the hydro-group cover. Use medium size screwdriver to remove the fuse cup. Then install a new fuse. Fuse Specification Main fuse 6A 250V

  • 25

    11. Technical data 11.1 Water and air flow diagram

  • 26

    11.2 Electrical diagrams

    12

    34

    56

    78

    A B C D

    87

    65

    43

    21

    DCBA

    Title

    Num

    berRevision

    SizeA2

    Date:

    30-Oct-2007

    Sheet of File:

    F:\work\hardw

    are\siger\siger.DD

    BD

    rawn By

    :

    J43 AC23VAC24V

    AC8VAC0VAC10VAC13V/AC17V

    operation light

    AC_0VAC_10VAC13V/AC17V

    J47

    AC

    24VA

    C24V

    _0V

    J45

    vacuumseparator

    AC

    24VA

    C24V

    _0V

    signal

    J49J51J48J54A

    C24V

    AC

    24V_0V

    J3J8J5J6J4

    HV suction

    work valve

    saliva ejector

    rinse solenoid

    cup-water solenoid

    A -+

    A

    -

    +

    J40J41

    chair seat engine connection

    chair backrest engine connection

    water heater

    AC23V

    J52J53J31J32

    floor box protection switch

    cuspidor safety sw

    itch

    Hydro-group control PCB

    J81A

    J19B

    J34-6J34-7

    J34-8

    J82-2A

    J82-2B

    AC

    24V

    curing light

    J58AC24V Assistant side

    Food board

    3*4KEY

    BOA

    RD

    J22-2BJ22-2D

    J20-2A

    J20-1A

    J82-3AJ82-1A

    LOW1

    LOW2

    LOW3

    LOW4

    ROW1

    ROW2

    ROW3

    LED

    NC

    GND

    X-ray viewer

    Hydro-group

    Instrument table

    Junction boxFuse

    Transformer

    J37J20-1B

    power port

    S model

    J29AJ28A

    PWM

    DC5V

    KG

    1K

    G2

    KG

    3K

    G4

    KG

    5

    X model

    X-ray

    J69

    4*6 key board

    ROW1

    ROW2

    ROW3

    ROW4

    ROW5

    ROW6

    LOW1

    LOW2

    LOW3

    LOW4

    JR_LED

    VDD33V

    J36

    DC12V

    J74

    J13-3J13-2

    J13-1

    LINE3 LINE2 LINE1

    J14-2J14-1

    water air

    scaler water and power solenoid IO . D

    C24V

    J13-3

    J73J80

    scaler

    J201J76J77

    turbine

    turbine

    J82-3BJ82-1B

    J57

    Instrument table PCB

    R?

    S30_OLD

    _BYQ

    J44

    Intra-oral camera

    comm

    J34-1J34-2J34-3J34-4

    AC24

    AC24

    J13-1,J13-2,J13-3 control main

    electric valve of 3 instruments w

    ater ,

    main air valve

    main water valve

    fiber optical output

    fiber optical output