Upload
adenilton-de-diva
View
406
Download
16
Embed Size (px)
Citation preview
8/11/2019 Manual FTLA2000.pdf
1/92
FTLA2000 Series Laboratory FT-IR SpectrometersUser's Guide
This manual contains: Basic information on using your FTLA2000 Series spectrometer, as well as
safety, maintenance, and troubleshooting information.
IMZ8252
Revision 1-3 December 2002 ABB
8/11/2019 Manual FTLA2000.pdf
2/92
8/11/2019 Manual FTLA2000.pdf
3/92
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide iii
WARNING! All servicing of the equipment is to be performed by
Qualified Service Personnel only.
No user/operator adjustments inside the equipment are
necessary or recommended by the manufacturer.
2002 ABB Bomem Inc. All Rights Reserved. No information contained in
this document may be reproduced in any form, in full or in part, without prior
written permission from ABB Inc.
This document contains product specifications and performance statements that
may be in conflict with other ABB Inc. published literature, such as productflyers and product catalogs. All specifications, product characteristics, and
performance statements included in this document are given as suggestive
specifications only. In case of conflict between product characteristics given in
this document and specifications given in the official ABB Inc. product catalogs,
the latter take precedence.
ABB Inc. reserves the right to make changes to the specifications of all
equipment and software, and to the contents of this document, without
obligation to notify any person or organization of such changes. Every effort has
been made to insure that the information contained in this document is current
and accurate. However, no guarantee is given or implied that the document is
error-free or that the information is accurate.
ABB Inc. provides product support services throughout the world. To receive
product support, either in or out of warranty, contact the ABB Inc. office that
serves your geographical area, or the office shown below:
ABB Inc.
Analytical and Advanced Solutions585 Charest Blvd. East, Suite 300
Qubec, QC G1K 9H4
CANADA
Phone: 418-877-2944
North America: 800-858-3847
France: 0810 020 000
Fax: 418-877-2834
Important: Please be prepared to provide the serial numbers of all units.
You can also consult ABBs web site at www.abb.com/analytical
Copyright
Disclaimer
Customer support
mailto:[email protected]:[email protected]:[email protected]8/11/2019 Manual FTLA2000.pdf
4/92
8/11/2019 Manual FTLA2000.pdf
5/92
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide v
TABLE OF CONTENTS About this Manual............................................................................................................................................ ix
1. FTLA2000 Series Laboratory FT-IR Spectrometers............................................................................. 1
1.1 Description.................................................................................................................................... 11.1.1 Arid-Zone sample station................................................................................................. 21.1.2 Detector assemblies.......................................................................................................... 31.1.3 Purging ............................................................................................................................. 31.1.4 FT-IR Ethernet Interface .................................................................................................. 3
1.2 Options.......................................................................................................................................... 41.2.1 Optical side port ............................................................................................................... 41.2.2 External source module .................................................................................................... 6
1.3 Sampling accessories .................................................................................................................... 7
1.4 Computer ...................................................................................................................................... 7
2. Safety.......................................................................................................................................................... 9
2.1 Class of equipment........................................................................................................................ 9
2.2 Certification .................................................................................................................................. 9
2.3 Power connections........................................................................................................................ 9
2.4 Fuse type....................................................................................................................................... 9
2.5 Moisture and dust........................................................................................................................ 10
2.6 Symbols ...................................................................................................................................... 10
2.7 Laser and high voltage ................................................................................................................ 102.7.1 Laser specifications........................................................................................................ 112.7.2 Laser labels..................................................................................................................... 11
2.8 Precautions.................................................................................................................................. 14
3. Operation................................................................................................................................................. 17
3.1 Control panel .............................................................................................................................. 173.1.1 Line voltage configuration ............................................................................................. 173.1.2 Ethernet cable, connector and indicators........................................................................ 183.1.3 Setting the resolution...................................................................................................... 18
3.2 Purge gas connections................................................................................................................. 18
3.3 Start up........................................................................................................................................ 193.3.1 Checking the status LEDs .............................................................................................. 19
3.3.2 Turning on the source..................................................................................................... 193.3.3 Warm up......................................................................................................................... 19
3.4 Avoiding detector saturation....................................................................................................... 19
8/11/2019 Manual FTLA2000.pdf
6/92
vi FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
3.4.1 Setting the preamplifier gain...........................................................................................20
3.5 Humidity .....................................................................................................................................21
4. Acquiring a Spectrum ............................................................................................................................. 23
4.1 Check the signal ..........................................................................................................................23
4.2 Acquire a background (reference) spectrum ...............................................................................25
4.3 Acquire a sample spectrum .........................................................................................................25
5. Validating the Spectrometer...................................................................................................................27
5.1 Required tests.............................................................................................................................. 27
5.2 Validating for quantitative analysis.............................................................................................27
6. Tips on Using your Spectrometer...........................................................................................................296.1 Power ..........................................................................................................................................29
6.2 Humidity .....................................................................................................................................296.2.1 Precautions for humidity-sensitive spectrometers ..........................................................296.2.2 Humidity-insensitive spectrometers................................................................................306.2.3 Dry air generators ...........................................................................................................30
6.3 Factors affecting the quality of spectra .......................................................................................306.3.1 Signal averaging versus time ..........................................................................................316.3.2 Variation of external conditions during acquisition........................................................316.3.3 Local environment..........................................................................................................32
6.4 Using sampling accessories.........................................................................................................32
6.5 Background (reference) spectra...................................................................................................33
7. Maintenance............................................................................................................................................. 35
7.1 Preventive maintenance schedule................................................................................................ 35
7.2 Replacing the desiccant module..................................................................................................35
7.3 Replacing the internal source ......................................................................................................36
7.4 Replacing the laser ......................................................................................................................36
7.5 Replacing the fuse.......................................................................................................................37
7.6 Cleaning or replacing windows ................................................................................................... 377.6.1 Optical side port windows ..............................................................................................37
7.6.1.1 Cleaning the windows...................................................................................377.6.1.2 Replacing the windows .................................................................................38
7.6.2 Other windows................................................................................................................387.6.2.1 Cleaning the windows...................................................................................38
8/11/2019 Manual FTLA2000.pdf
7/92
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide vii
8. Troubleshooting ...................................................................................................................................... 39
8.1 After power-up............................................................................................................................ 39
8.2 During use................................................................................................................................... 408.2.1 Software messages about hardware problems ................................................................ 42
8.2.1.1 On selecting a function in the Collectmenu................................................. 428.2.1.2 After acquiring a spectrum ........................................................................... 428.2.1.3 Other messages............................................................................................. 42
8.3 Contacting ABB.......................................................................................................................... 43
8.4 Examples of normal spectra........................................................................................................ 44
A Technical Information ............................................................................................................................ 47
A.1 Specifications.............................................................................................................................. 47A.1.1 Electrical ...................................................................................................................... 47A.1.2 Mechanical ..................................................................................................................... 47A.1.3 Environmental ................................................................................................................ 47A.1.4 Spectrometer................................................................................................................... 48A.1.5 Interferometer................................................................................................................. 48
B Installing your MB Series Spectrometer ............................................................................................... 49
Purpose .................................................................................................................................................... 49
Site preparation......................................................................................................................................... 49
Line voltage configuration........................................................................................................................ 49
Main controls and connections ................................................................................................................. 50
Purge gas connections (if used) ................................................................................................................ 51Safety .................................................................................................................................................... 52
Start up .................................................................................................................................................... 52
Ethernet configuration .............................................................................................................................. 52
Validation ................................................................................................................................................. 52
C Replacing the NIR Source in your MB Series Spectrometer........... ........... ............ ........... ........... ....... 53
Purpose .................................................................................................................................................... 53
Safety .................................................................................................................................................... 53
Removing the NIR source ........................................................................................................................ 53
Replacing the NIR source ......................................................................................................................... 56
Aligning the NIR source ........................................................................................................................... 56Validation ................................................................................................................................................. 57
8/11/2019 Manual FTLA2000.pdf
8/92
viii FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
D Replacing the MIR Source......................................................................................................................59
D.1 Purpose........................................................................................................................................ 59
D.2 Safety .......................................................................................................................................... 59
D.3 Removing the MIR source ..........................................................................................................59
D.4 Replacing the MIR source...........................................................................................................60
D.5 Validation....................................................................................................................................62
E Glossary....................................................................................................................................................63
F Index.........................................................................................................................................................77
8/11/2019 Manual FTLA2000.pdf
9/92
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide ix
ABOUT THIS MANUAL This manual is intended for personnel using an FTLA2000 Series spectrometer
for routine analysis, as well as personnel responsible for maintaining and
troubleshooting the analyzer.
For information on: Refer to Chapter/Appendix:
Installation App. B Installing your FTLA2000 SeriesLaboratory FT-IR Spectrometer
The FTLA2000 Series Spectrometers Chap. 1 FTLA2000 Series Laboratory FT-IR Spectrometers
Routine analysis Chap. 2 Safety
Chap. 3 Operation
Chap. 4 Acquiring a Spectrum
Getting the most out of yourspectrometer
Chap. 5 Validating the Spectrometer
Chap. 6 Tips on Using your Spectrometer
Maintenance and troubleshooting Chap. 7 Maintenance
Chap. 8 Troubleshooting
Software commandsand dialog box namesand are shown in boldface type.
Text to be entered into softwareis in sans serif type.
This symbol refers you to another manual or document.
Note: Supplemental information to help the reader.
Important: Information that is important, but that does not concern the safe use
of the equipment.
This symbol shows that Caution is required. Follow
the instructions carefully to avoid damage to the equipment.
WARNING! Failure to comply with warnings can result in serious injury
or loss of life.
Audience
Conventions usedin ABB manuals
8/11/2019 Manual FTLA2000.pdf
10/92
8/11/2019 Manual FTLA2000.pdf
11/92
C H A P T E R 1
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 1
FTLA2000 SERIES LABORATORY FT-IR SPECTROMETERS
FTLA2000 Laboratory FT-IR Series spectrometers are powerful and reliable
Fourier Transform Infrared (FT-IR) spectrometers designed for a wide range of
IR analyses in a variety of environments, including factories, field locations,
educational institutions, and quality control laboratories.
The FTLA2000 Series of FT-IR spectrometers includes the following models
covering both the mid infrared (MIR) and near infrared (NIR) spectral ranges:
Model Typical Spectral Range
(cm-1
)
Humidity Sensitive
(hygroscopic optics)
FTLA2000-100 6500 350 MIR Yes
FTLA2000-102 5000 200 MIR Yes
FTLA2000-104 6500 500 MIR No
FTLA2000-154 12 000 500 NIR/MIR No
FTLA2000-155 12 000 450 NIR/MIR Yes
FTLA2000-160 14 000 - 3800 NIR No
Note: These spectral ranges are approximate. Refer to published specifications
for the exact spectral range of each model. The detector and/or other
devices may reduce the spectral range.
These spectrometers provide exceptionally high stability, sensitivity andphotometric accuracy. When the necessary measures are taken, the repeatability
(the precision of repeated measurements on the same instrument) and
reproducibility (the precision of repeated measurements on the different
instruments) are outstanding. Refer to Chapter 5 Validating the Spectrometeron
page 27 for further information.
The interferometer and the input/output optics of these instruments are factory
prealigned. There are no routine adjustments required.
FTLA2000 Series spectrometers are controlled by acquisition software running
on an external PC computer running under the Microsoft Windows
2000 or
Windows
NT. ABB can supply the FTLA2000 Series spectrometers with or
without a computer. The computer requires an Ethernet network adapter.
1.1 Description
8/11/2019 Manual FTLA2000.pdf
12/92
S e c t i o n 1 . 1 D e s c r i p t i o n
2 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
All FTLA2000 Series spectrometers can be used with various optional sampling
accessories that come with their own manuals. Several additional features are
also available as options.
The following items are included with each FTLA2000 Series spectrometer:
Arid-Zonesample stationa(see Section 1.1.1)
One detector assembly (see Section 1.1.2)
Acquisition software (refer to the softwaremanuals provided)
The Arid-Zone sample station comes as a standard accessory (see Figure 1-1
below). It eliminates waiting for purge without compromising spectrometerperformance. It also eliminates the need for water vapor compensation routines
for quantitative analysis applications.
Figure 1-1. FTLA2000 Series spectrometer with Arid-Zone sample station
For MIR analysis, the telescopic purge tubes of the Arid-Zone sample
compartment adjust to the sample cell and insure continuous purge right up to
the sample. This allows purging, yet leaves a wide, open space around
the IR beam for immediate access to the sample without breaking purge, as well
as sufficient space for all standard sample accessories. The Arid-Zone sample
station is compatible with all industry-standard accessories such as ATR,DRIFT, and Liquid Cells.
aA different sample station,
such as the HOval, may be
supplied instead of, or in
addition to, the Arid-Zone.
1.1.1 Arid-Zone
sample station
Output beamcover
Humidity indicator /Desiccant module
Source adjustmentaccess plug(remove to adjustsource power)
Telescopicpurge tube (forMIR analysis)
Detector cover
Control panel
8/11/2019 Manual FTLA2000.pdf
13/92
C h a p t e r 1 F T L A 2 0 0 0 S e r i e s L a b o r a t o r y F T - I R S p e c t r o m e t e r s
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 3
For NIR analyses, where purge is not required, the telescopic purge tubes are not
used. In their place, protective windows are installed to prevent contaminant
liquids or vapors from entering the spectrometer and causing damage.
ABB can supply the FTLA2000 Series spectrometers with detector assemblies
of different types and sensing element sizes, interchangeable by the user.
The power must be turned off when detector assemblies are
being interchanged.
FTLA2000 Series spectrometers are equipped with purge inlets. Purging isrecommended for spectrometers which have hygroscopic optical components
(models FTLA2000-100, FTLA2000-102, and FTLA2000-155) or if ambient
water vapor may interfere with measurements.
Purging should be done using dry nitrogen or CO2 and H2O free air when
an application is sensitive to changes in air humidity, or when the sensitivity of
the analyzer is compromised by the absorption of CO2and H2O found in normal
concentrations in the ambient air. When water vapor alone is to be eliminated,
dry, oil-free instrumentation air can be used for purging.
The recommended purge flow rate into the spectrometer is 5 L/min. Oncethe spectrometer is purged, a continuous flow rate of 1 L/min is normallysufficient to maintain the purged condition. An automatic relief valve is installed
to prevent accidental over-pressurization of the spectrometer. This relief valve
opens if the internal pressure exceeds 1/3 psi.
The FT-IR Ethernet Interface is built into the FTLA2000 Series. It allows
communication between the FTLA Series and a computer over an Ethernet link
using standard network protocols. Communication is possible through a private
connection between the instrument and the computer, over a local area network,
or over the Internet. The computer must be equipped with a standard Ethernet
network adapter.
An Ethernet link offers many advantages over a serial or parallel link. For
example, an Ethernet link:
works over standard office and plant networks
1.1.2 Detectorassemblies
Detector assemblies can be
procured at the time of the
initial sale or later on.
1.1.3 Purging
1.1.4 FT-IR EthernetInterface
8/11/2019 Manual FTLA2000.pdf
14/92
S e c t i o n 1 . 2 O p t i o n s
4 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
allows sharing of FT-IR instruments by multiple computers overthe network
allows control of multiple FT-IR instruments by a single computer overthe network
simplifies cabling and allows greater distances between the FT-IRinstrument and the computer
does not require special hardware in the computer
can be upgraded by loading a new version of the firmware
The following are some of the features of the FT-IR Ethernet Interface:
Uses a standard RJ45 phone type connector and twisted pair wiring(10Base-T or 100Base-T)
Automatically detects network speed (10 Mbps or 100 Mbps)
Maximum network load of approximately 2 Mbps
Network traffic is not encrypted
Network traffic will not cross firewalls
Instruments are addressed by a unique name
Refer to the FT-IR Ethernet Interface and External Ethernet InterfaceUser's Guidefor complete information on the FT-IR Ethernet Interface.
The following sections describe the principle options available for the
FTLA2000 Series spectrometers.
As an option, ABB can install one optical side port on the FTLA2000 Series
spectrometer. The different types are described below:
Optical outport (Figure 1-2 on page 5). This option allows the easyinterfacing of the spectrometer to accessories mounted on the side of
the spectrometer.
Once installed, the side accessory is selected by placing the position of
a manually operated flip mirror in the appropriate position. This directsthe IR beam from the interferometer through the side port, assures that
the spectrometer selects the signal coming from the detector located on
1.2 Options
1.2.1 Optical side port
8/11/2019 Manual FTLA2000.pdf
15/92
C h a p t e r 1 F T L A 2 0 0 0 S e r i e s L a b o r a t o r y F T - I R S p e c t r o m e t e r s
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 5
the side, and also assures that the interferometer adjusts to the optimum
scanning speed for the type of detector used at that location.
When the flip mirror is in the normal position, the IR beam goes to the top
sampling station, and the detector located on the top is selected.
Figure 1-2. Optical outport
Optical inport (Figure 1-3 on page 6). This option allows you to installan external source module on the spectrometer.
Once the external source is installed, you can select the external or internal
source simply by turning one source on and turning the other off.
8/11/2019 Manual FTLA2000.pdf
16/92
S e c t i o n 1 . 2 O p t i o n s
6 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
Figure 1-3. Optical inport
Note: After switching a source on or off, allow a temperature stabilization time
from 10 minutes to 2 hours, depending on the need for stability, before
performing analyses.
ABB offers optional external source modules that mount on the optional optical
side inport (or emission port) on the left side of the FTLA2000 Series
spectrometer. The optional external source module has an external regulated
power supply and optical filter holder (a filter may be useful to prevent detector
saturation under certain conditions).
Note: The optional external source modules are not compatible with
the optional optical side outport.
For complete instructions on installing (or replacing), testing, and usinga particular external source module, refer to the documentation supplied
with that external source module.
Specifications for the optional external sources are given in the following tables.
All values are nominal. (No user settings are required.)
1.2.2 External sourcemodule
8/11/2019 Manual FTLA2000.pdf
17/92
C h a p t e r 1 F T L A 2 0 0 0 S e r i e s L a b o r a t o r y F T - I R S p e c t r o m e t e r s
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 7
External NIR source External MIR source
Rated line voltage: 90-265 VAC 90-265 VAC
Rated frequency: 50-60 Hz 50-60 Hz
Radiation source (absorption
spectroscopy)
Quartz-halogen lamp
(collimated to a 2.5 cm diameter beam)
SiC Globar
(collimated to a 2.5 cm diameter beam)
Range near IR range (15 000 cm-1to 4000 cm-1) mid IR range (7000 cm-1to 200 cm-1)
FTLA2000 Series spectrometers are compatible with industry-standard sampling
accessories from ABB and from other manufacturers. The following list shows
some accessories available from ABB.
Raw Material Identification FT-NIR Analyzer
Bag Sampling
Powder Sampling
Tablet Sampling
Vial Holder
For complete instructions on installing, testing, and using a particularsampling accessory, refer to the documentation supplied with that
sampling accessory.
ABB can supply the FTLA2000 Series spectrometer with or without a computer.If the computer is supplied by ABB, all software will be preinstalled and
configured.
The minimum configuration requirements for the computer depend on thesoftware you intend to use. Refer to your software manuals.
1.3 Samplingaccessories
1.4 Computer
8/11/2019 Manual FTLA2000.pdf
18/92
8/11/2019 Manual FTLA2000.pdf
19/92
C H A P T E R 2
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 9
SAFETYThe FTLA2000 Series spectrometer has an exposed metal chassis that is
connected directly to earth via a power supply cord and are therefore classified
as Safety Class I equipment.
WARNING! To avoid electrical shock, do not operate this equipment if it
bears any sign of damage to any portion of its exterior
surface.
Do not expose this equipment to any source of excessive
moisture.
Do not use this equipment in an explosive atmosphere.
ABB instruments receive the following safety certifications:
Canadian Standards Association (for Canada and the US)
Conformit europenne (European conformity)
In accordance with international safety standards, each powered unit uses a
three-wire power cord. When connected to an appropriate AC power receptacle,
this power cord grounds the chassis.
WARNING! To avoid the risk of injury or death:
Use only a power cord with a protective earthing
terminal. Never use an extension cord that is not
equipped with this feature.
Connect the power cord to a power outlet of the correct
voltage and that has a protective earth contact.
Disconnect all power cords before servicing any unit.
The main power fuse(s) are rated as follows:
T2A/250V (115 V line)T1A/250V (230 V line)
2.1 Class ofequipment
2.2 Certification
2.3 Powerconnections
2.4 Fuse type
8/11/2019 Manual FTLA2000.pdf
20/92
S e c t i o n 2 . 5 M o i s t u r e a n d d u s t
10 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
The FTLA2000 Series spectrometer is designed for indoor use in a laboratory-
type environment. It should not be exposed to excessive moisture and dust.
The following symbols may be used in documentation or on the instrument:
Symbol Meaning
WARNING!
or
WARNING
Failure to comply with warnings can result inserious injury or loss of life.
Ensure that all conditions necessary for safe
handling and operation are met beforeproceeding.
Danger: High voltage
Warning: Hot surface
Caution: Follow instructions carefully to
avoid damage to the equipment.
Protective earth conductor terminal
On (power)
Off (power)
Under normal conditions, the FTLA2000 Series spectrometer can be operated in
complete safety. However, since the instrument contains a laser and uses highvoltages (accessible only when the enclosure is open), observe the following
warnings:
2.5 Moisture anddust
2.6 Symbols
2.7 Laser and
high voltage
8/11/2019 Manual FTLA2000.pdf
21/92
8/11/2019 Manual FTLA2000.pdf
22/92
S e c t i o n 2 . 7 L a s e r a n d h i g h v o l t a g e
12 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
CAUTIONCLASS 3B LASER, RADIATION WHEN OPEN
AVOID EXPOSURE TO BEAMIMZ6426
Figure 2-1. Laser caution label
Figure 2-2. Locations of laser caution labels
Under the spectrometer
On top plate, underaccess cover
On access cover
Next to side port(if installed)
8/11/2019 Manual FTLA2000.pdf
23/92
C h a p t e r 2 S a f e t y
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 13
MANUFACTURE DATE:
SERIAL NO.:
MODEL:
CLASS 1LASER PRODUCT
ABB Inc.Analytical and Advanced Solutions585 Charest Blvd. East, Suite 300Qubec, QC G1K 9H4 CANADA
Phone: 418-877-2944
C US
COMPLIES WITH FDA
RADIATION PERFORMANCESTANDARDS.21 CFR CHAPTER 1SUBCHAPTER J.
Figure 2-3. Nameplate with Class 1 laser label
8/11/2019 Manual FTLA2000.pdf
24/92
S e c t i o n 2 . 8 P r e c a u t i o n s
14 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
Figure 2-4. Location of nameplate with Class 1 laser label and laser caution label
WARNING! Failing to comply with any of the instructions, precautions
or warnings contained in this manual is in direct violation of
the standards of design, manufacture, and intended use of
the equipment.
ABB Inc. assumes no liability for the users failure to
comply with any of these safety requirements.
The following precautions must be observed whenever the equipment is
operated, serviced, or repaired.
Before operating the equipment:
Inspect the equipment for any signs of damage, and read this manual
thoroughly.
Install the equipment as specified in the documentation provided.
2.8 Precautions
8/11/2019 Manual FTLA2000.pdf
25/92
C h a p t e r 2 S a f e t y
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 15
Make sure the system is configured to use the available line voltage and thatthe correct fuse(s) for that voltage is (are) installed. Use only the fuse(s)
specified as appropriate for this equipment (see Section 2.4 on page 9).
Ensure that equipment and any devices or power cords connected to it areproperly earthed.
While operating the equipment:
Do not operate the equipment in the presence of flammable gases, fumes,or dust.
Do not operate the equipment when its covers or panels have been removed.
Do not interrupt the protective earthing connection. Any such action canlead to a potential shock hazard that could result in serious personnel injury.
Do not operate the equipment if an interruption to the protective earthing issuspected. Ensure that the equipment remains inoperative.
Do not use repaired fuses and avoid any situations that could short-circuitthe fuse.
Unless absolutely necessary, do not attempt to adjust or perform anymaintenance or repair procedure when the equipment is opened and
connected to a power source at the same time. Any such procedure should
only be carried out by a qualified service professional.
Do not attempt any adjustment, maintenance, or repair procedure tothe equipment if immediate first aid is not accessible.
WARNING! High voltages may be present inside the enclosure even
when the equipment is not connected to the power source
because some of the capacitors may be charged.
8/11/2019 Manual FTLA2000.pdf
26/92
8/11/2019 Manual FTLA2000.pdf
27/92
8/11/2019 Manual FTLA2000.pdf
28/92
S e c t i o n 3 . 2 P u r g e g a s c o n n e c t i o n s
18 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
Two Ethernet cables are supplied (see Figure 3-2), a crossovercable for making
a private (direct) connection to the computer and astraight-pinned cable for
connecting the instrument to a network. Be sure to use the correct type of cable.
Figure 3-2. Ethernet cables for private and network connection
The Ethernet cable must be connected to the RJ45 connector identified as
10/100 BASE T on the control panel. The MAC address of the FT-IR Ethernet
Interface is indicated on the panel.
Two LEDs are associated with the Ethernet connector:
LINK is on when the instrument is properly connected.
ACT is on when there is activity on the link.
Refer to the FT-IR Ethernet Interface and External Ethernet InterfaceUser's Guide for complete information on making the Ethernet
connection and configuring the FT-IR Ethernet Interface.
The resolution can be changed at any time before making a new spectroscopic
measurement.
Make sure the same resolution setting is used for both the reference spectrum
and the sample spectra.
Refer to theInstallingmanual (reproduced on page 51) for instructions
on connecting the purge gas, if used.
3.1.2 Ethernet cable,connector andindicators
3.1.3 Settingthe resolution
3.2Purge gasconnections
For private (direct) connectionto computer
For network connection
8/11/2019 Manual FTLA2000.pdf
29/92
C h a p t e r 3 O p e r a t i o n
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 19
Turn on the spectrometer, the computer, and all other powered components.
After turning on the spectrometer, the status LED sequence should be:
LED Normal State
POWER ON
SCAN OFF (a few seconds after power on)
ZPD OFF (a few seconds after power on)
DATA Flashing (usually within one minute after power on)
This sequence confirms the correct operation of the spectrometer. Duringoperation, the POWER LED should be ON and the DATA LED should be
flashing. If you encounter problems, refer to Chapter 8 on page 39.
Set the source breaker-switch to ON. Dual source models such as the
FTLA2000-154 or the FTLA2000-155 can have an IR (MIR) and an NIR
source, each of which can be turned on or off independently.
Note: After switching on a source, allow a temperature stabilization time from
10 minutes to 2 hours, depending on the need for stability, before
performing analyses.
When turned on at room temperature, the spectrometer will need about 4 hours
to allow its infrared source and internal temperature to stabilize; for critical
measurements, more time may be required. The spectrometer can be used
immediately after being turned on but the resulting spectra may not be within
published specifications.
Important: The spectrometer should be left ON at all times.
Each time it is turned OFF then back ON, it will take several hours
for the internal temperature to stabilize.
Some sensitive near-IR detectors are provided with selectable preamplifier gain.
After installing a sampling accessory, the preamplifier gain must be set so thatsufficient signal amplitude is obtained and detector saturation is avoided
(see Section 3.4.1 on page 20).
3.3 Start up
3.3.1 Checkingthe status LEDs
3.3.2 Turning onthe source
3.3.3 Warm up
3.4 Avoiding
detectorsaturation
8/11/2019 Manual FTLA2000.pdf
30/92
8/11/2019 Manual FTLA2000.pdf
31/92
C h a p t e r 3 O p e r a t i o n
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 21
Figure 3-3. Arid-Zone sample station
Figure 3-4. Rotary switch used to set the preamplifier gain
The spectrometer may require protection from humidity.Refer to Section 6.2 on page 29.
3.5 Humidity
Detector coverscrews (2)
8/11/2019 Manual FTLA2000.pdf
32/92
8/11/2019 Manual FTLA2000.pdf
33/92
C H A P T E R 4
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 23
ACQUIRING A SPECTRUM This chapter guides you in acquiring the transmittance spectrum of a sample
using any Windows-based acquisition software program supplied by ABB Inc. It
assumes that the software is installed and configured.
Note: The commands used in this section are in the Collect menu of
the acquisition program.
For detailed information on using a particular acquisition program, referto the manual supplied with that program.
1. Make sure there are no objects in the sampling area and that the sampleholder is in position. (This condition is called open beam.)
2. Using the Setup command, select the FTLA2000 Series spectrometer and
make sure all parameters are set correctly.
3. Set the spectrometer resolution to 4 cm-1.
4. Select the alignment command to display the alignment dialog box.
Set the parameters as follows:
Spectral Range:
for MIR: 4000 to 400 cm-1 for NIR: 14 000 to 1000 cm-1
Data type: Single-Beam (or Raw spectrum)
5. Click OK Align to begin data acquisition. During data acquisition,
the display will be continuously updated until the halt command is selected.
6. Observe the displayed spectrum. The word Align, followed by a percentage,
indicates the interferogram amplitude as a percentage of the maximum
allowed. This percentage should be between 40% and 90%.
7. Check for any unusual spectral features such as a strong signal below
the detector cut-off frequency (usually caused by detector saturation).
Figure 4-1 on page 24 shows typical spectra for MIR and NIR. Figure 4-2
on page 24 presents MIR spectra showing detector saturation.
4.1 Check the signal
8/11/2019 Manual FTLA2000.pdf
34/92
8/11/2019 Manual FTLA2000.pdf
35/92
C h a p t e r 4 A c q u i r i n g a S p e c t r u m
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 25
1. Select the command for acquiring a background spectrum (Background
Scanningor Collect).
In the dialog box, set the parameters as follows:
Spectral Range (if available):
for MIR: 4000 to 400 cm-1 for NIR: 14000 to 4000 cm-1
Number of Scans: 10
File Name: (enter a file name)
If available, select the Background acquisition type.
2. Click the OK button to begin data acquisition. During data acquisitionthe spectrum of each individual scan will be displayed on the page. When
all the spectra are acquired, the coadded spectrum is displayed and saved.
1. When the background spectrum has been acquired, insert the sample in
the sample holder.
For MIR, use the polystyrene sample supplied with the spectrometer(see Figure 4-3 below).
Figure 4-3. Inserting the polystyrene sample into the sample holder
For NIR, do not use the polystyrene sample. Instead, place an ordinary
plastic bag over the sample holder.
4.2 Acquirea background(reference)spectrum
4.3 Acquirea samplespectrum
8/11/2019 Manual FTLA2000.pdf
36/92
S e c t i o n 4 . 3 A c q u i r e a s a m p l e s p e c t r u m
26 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
2. Select the command for acquiring a normal spectrum. (Scanning or
Collect).
In the dialog box, set the parameters as follows:
Spectral Range: (same as for background spectrum)
Number of Scans: 10
File Name: (enter a file name)
Data type: %Trans
If available, select the Normal acquisition type.
3. Click the OK button to begin data acquisition. During data acquisition
the spectrum of each individual scan will be displayed on the page. Whenall the spectra are acquired, the coadded spectrum is displayed and saved.
4. Observe the acquired spectrum.
Figure 4-4 shows an example of the transmittance spectrum of a polystyrene
sample in the MIR range and the transmittance spectrum of a plastic bag in
the NIR range.
Figure 4-4. Transmittance spectra
8/11/2019 Manual FTLA2000.pdf
37/92
C H A P T E R 5
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 27
VALIDATING THE SPECTROMETER FTLA2000 Series spectrometers can be used in many different applications.
However, there are three broad categories of spectrometers users, and each
category has its own requirements concerning the tests necessary to ensure
proper performance.
The following table shows the validation tests required for each type of user:
Tests required
Type of user Check for detectorsaturation*
Frequency validation
Spectral quality
All other availablevalidation tests
Users performing qualitative analysis only
Users of an ABB dedicated analyzer(e.g. PROTA)
Users performing quantitative analyses
or who may do so in the future
*Check for detector saturation before using the spectrometer and each time you install a different sampling accessory.
Adjust the preamplifier gain if necessary (see Section 3.4 on page 19).
With ABB spectrometers, models (calibrations) for quantitative analysis can be
developed using one spectrometer (e.g. in a laboratory), and then transferred to
other spectrometers (e.g. in a plant), without having to adjust the model for each
spectrometer. However, a complete performance validation of eachspectrometeris necessary to guarantee transferability of models.
Before using a spectrometer for quantitative analysis, it is recommended that
you perform allof the available validation tests.
We strongly recommend that all validation tests be performed when
the spectrometer is first installed, and afterwards on a regular basis, even if
you do not foresee the need to transfer models to another spectrometer.
For information on performing the validation tests, refer to the SpecTestUsers Guide.
For further information on transferability of models, refer to the chapterentitled Quantitative Analysis and Calibration Transfer of the FT-IRReference Manual.
5.1 Required tests
5.2 Validating forquantitativeanalysis
8/11/2019 Manual FTLA2000.pdf
38/92
8/11/2019 Manual FTLA2000.pdf
39/92
C H A P T E R 6
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 29
T IPS ON USING YOUR SPECTROMETER Leave the spectrometer ON at all times. This not only ensures thermal stability,
but also helps protect the spectrometer from humidity. However, it is possible to
turn the source off when the spectrometer is not in use. After switching a source
on or off, allow a temperature stabilization time from 10 minutes to 2 hours,
depending on the need for stability, before performing analyses.
The average source life when left on continuously is:
MIR source: 4 years
NIR source: 1 year
Your FTLA2000 Series spectrometerb may contain some hygroscopic
components (components that can be damaged by humidity). Although the
spectrometer is sealed, it is not vacuum tight; there will always be some
exchange of air between the interferometer cavity and the outside. For this
reason, a number of precautions must be taken to ensure minimal long-term
damage due to humidity. These are described below.
For more specific information on hygroscopic materials, refer tothe Chapter Handling and Care of Hygroscopic Materials of the FT-IR
Reference Manual.
To assure maximum life for humidity-sensitive spectrometers
(models FTLA2000-100, FTLA2000-102, and FTLA2000-155), the following
precautions should be taken:
Leave the spectrometer ON at all times.
If possible, purge the spectrometer at all times using dry, oil-freeinstrumentation air or dry nitrogen. Before interrupting the purging gas,
make sure the internal desiccant module is installed and that the desiccant is
fresh.
If the spectrometer is not left ON at all times, or if the relative humidity is
extremely high:
Put external desiccant around exposed hygroscopic windows wherepossible. Use as much desiccant as possible without partially blocking
6.1 Power
6.2 Humidity
bModels FTLA2000-100,
FTLA2000-102, and
FTLA2000-155, are
sensitive to humidity,
others models are not.
6.2.1 Precautions forhumidity-sensitivespectrometers
8/11/2019 Manual FTLA2000.pdf
40/92
S e c t i o n 6 . 3 F a c t o r s a f f e c t i n g t h e q u a l i t y o f s p e c t r a
30 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
the beam. Note however that if the ambient humidity decreases rapidly,
the desiccant will release absorbed moisture that could damage
the windows.
If the spectrometer is not purged continuously:
Change the internal desiccant module regularly. Exhausted desiccantwill release substantial quantities of water when warmed by
the elevated temperature inside the spectrometer, causing rapid failure
of the beamsplitter.
Where humidity is a potential problem, we recommend the use of the
FTLA2000-104 (mid IR region) and FTLA2000-154 (mid and near IR region)spectrometers, if either of these spectrometers meets the requirements of your
application.
If a supply of dry nitrogen is not available, the spectrometer can be purged with
dry, oil-free instrumentation air. For this, ABB recommends the use of a dry air
generator.
Whatman Inc. manufactures excellent dry air generators that either work from
plant compressed air (model 75-45 is recommended) or from its own compressor
(model 74-5021 is recommended). Both units can deliver up to 14 liters per
minute of water free (-73oC dew point) and CO2 free (
8/11/2019 Manual FTLA2000.pdf
41/92
C h a p t e r 6 T i p s o n U s i n g y o u r S p e c t r o m e t e r
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 31
Understanding the effect of these factors on the resulting spectra should help to
optimize and improve the quality of acquired spectra.
The spectra acquired by all spectrometers are more or less affected by noise,
or random error introduced by variations in light and electrical signals.
The noisy signal appears as random fluctuations in the spectral response.
The amount of noise present in a spectrum is inversely proportional to the square
root of the signal averaging time, or the number of scans. For example, spectra
acquiring using 50 scans will have approximately one fifth the noise level of
spectra acquired using 2 scans.
The number of scans required will depend on various factors such as how much
energy the sample transmits and the spectral range of interest. However,
the same number of scans should usually be used to acquire the reference and
sample spectra.
Interferograms are not obtained instantaneously because of the necessity of
averaging to improve spectral quality. Unfortunately, this makes the results
dependent on variations of the instrument, sample or other conditions during
the scans.
Some examples of variations are:
A change in the fraction of water vapor or carbon dioxide in the air inthe sample path of the instrument. This can occur if you breathe or talk
near an unprotected sample during the scan.
A significant change in temperature (more than 10oC) during the scan.This can cause a change in dimensions and cause temporary optical
misalignment.
A sudden displacement or mechanical shock. This will cause a changeof scan velocity that could yield a distorted spectrum. If the shock is
severe enough, the instrument will reject the scan and thereby waste
measurement time.
Even gradual changes of conditions can cause problems whenthe background and sample spectra are acquired too far apart in time.
6.3.1 Signal averagingversus time
6.3.2 Variation ofexternalconditions duringacquisition
8/11/2019 Manual FTLA2000.pdf
42/92
S e c t i o n 6 . 4 U s i n g s a m p l i n g a c c e s s o r i e s
32 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
If conditions may vary, acquire the sample spectrum immediately after
the reference spectrum.
The FTLA2000 Series spectrometer is designed for harsh environments. There
are nevertheless limits to its adaptability.
The following conditions have a direct effect on performance:
Vibration: Excessive vibration will cause degradation of the highlyprecise scanning of the instrument, increasing the noise in
spectra. Built-in vibration isolators shield the instrument
from normal levels of vibration. However, the instrument
must not be in contact with vibrating surfaces except
through its base.
Temperature: The detector of the instrument produces more electricalnoise as the ambient temperature increases. To achieve
optimal performance, the temperature must be maintained
within 5C.c
All FTLA2000 Series spectrometers can be used with various optional sampling
accessories that come with their own manuals. Some sensitive near-IR detectors
are provided with selectable preamplifier gain. After installing or removing
an accessory, it may be necessary to adjust the preamplifier gain to obtain a
strong signal while avoiding detector saturation. Refer to Section 3.4 on page 19.
When performing quantitative analysis without normalization, a new reference
spectrum must be taken when the preamplifier gain is changed. Make sure
the same preamplifier gain setting is selected for both the reference spectrum
and the sample spectrum.
For qualitative analysis, or quantitative analysis with normalization, it is
possible to use different gain settings for reference and sample spectra.
For complete instructions on installing, testing, and using a particularsampling accessory, refer to the documentation supplied with that
sampling accessory.
6.3.3 Localenvironment
cFor example, the spectrometer
should not be located near an
outside door, a fan, or a heat
source such as an oven.
6.4 Using samplingaccessories
8/11/2019 Manual FTLA2000.pdf
43/92
C h a p t e r 6 T i p s o n U s i n g y o u r S p e c t r o m e t e r
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 33
It is good practice to acquire a new background spectrum at least every two
hours of operation. Make sure the same resolution setting is used for both
the background spectrum and the sample spectrum. For quantitative analysis
without normalization, the same gain setting must be used for both
the background and the sample spectra.
A new background spectrum must be acquired every time the IR beam geometry
is changed, such as when a different sample accessory is used, or when
the position of the sample accessory is changed or realigned.
The background spectrum should be acquired using the same number of scans
(coaddition) as will be used for the sample spectrum. It is important to acquire
the background spectrum first by running the spectrometer with the cell, orsample accessory, in place, but without the sample.
6.5 Background(reference)spectra
8/11/2019 Manual FTLA2000.pdf
44/92
8/11/2019 Manual FTLA2000.pdf
45/92
C H A P T E R 7
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 35
MAINTENANCEWARNING! Maintenance must be done by trained and qualified
personnel only.
WARNING! Laser replacement must be done by an ABB-trained Service
Engineer only.
Note: The replacement of consumables (other than the laser assembly) does not
require opening the spectrometer.
ABB Bomem recommends that an ABB Bomem trained Service Engineer
perform all servicing of the instrument. The spectrometer is permanently alignedand should not require optical realignment during its life span. The following
items should be replaced periodically for preventive maintenance:
Item Expected lifeRecommended
replacement frequency
Laser 4 years 2 years
NIR source (if installed) 1year 6 months
MIR source (if installed) 4 years 2 years
Desiccant module
(if spectrometer is notpurged)
Depends
on use
6 to 12 months
(3 months in very humidenvironment)
Instruments use either a NIR (near infrared) or a MIR (mid infrared) source, or both.
To replace the desiccant module (standard model desiccant assembly
SPL6300G), proceed as follows:
1. Leave the spectrometer on. This will keep the internal optics warm and
gives added protection against humidity.
2. Unscrew the desiccant module (SPL6300G) on the side of the spectrometer
(see Figure 7-1 on page 36).
3. Remove the desiccant module and replace it with a new desiccant module
(SPL6300G).
7.1 Preventive
maintenanceschedule
7.2 Replacing thedesiccant module
8/11/2019 Manual FTLA2000.pdf
46/92
S e c t i o n 7 . 3 R e p l a c i n g t h e i n t e r n a l s o u r c e
36 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
4. Install the new desiccant module.
Figure 7-1. Location of the SPL6300G desiccant module
The FTLA2000 Series Spectrometer uses different types of internal sources;
make sure you replace the source with the correct replacement model (usually
the same source model as was originally installed in the spectrometer). Contact
your ABB representative for information.
Attempting to install an incompatible source in your
spectrometer may damage both the source and the
spectrometer. It will also result in poor reproducibility.
To replace a standard near-IR internal source, refer to the documentReplacing the NIR Source in your FTLA2000 Series Laboratory
Spectrometer. This document is reproduced in Appendix C of this manual
on page 53.
To replace a standard mid-IR internal source, refer to the Appendix D.
The laser assembly must be replaced by authorized ABB service personnel.Please contact your ABB representative.
7.3 Replacing theinternal source
7.4 Replacingthe laser
Desiccant module
8/11/2019 Manual FTLA2000.pdf
47/92
C h a p t e r 7 M a i n t e n a n c e
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 37
To replace the fuse:
1. Turn the power off and unplug the spectrometer.
2. Move the sliding fuse cover on the control panel to the left.
3. Remove the fuse by pulling out the black plastic handle.
4. Replace the fuse with one of the same type (please refer to Section 2.4 on
page 9 for fuse types).
5. Move the sliding fuse cover back to the right.
6. Plug in the spectrometer and turn the power on.
The surfaces of optical windows can sometimes accumulate dust, grease or other
contaminants, leading to noisy spectra and undesired absorption bands. In this
case, the windows should be cleaned or replaced.
Never touch the windows surfaces with bare hands.
Avoid exhaling near window surfaces.
Never remove the mirror or the mirror support that is under
the left (output beam) cover of the spectrometer.
WARNING! Cleaning optical elements may result in an exposure of
Class II laser radiation, do not stare into the beam or view
with optical instruments.
7.6.1.1 Cleaning the windows
If dustis on the window surfaces:
gently blow dry, oil-free instrumentation air or dry nitrogen across
the surface. Do not use a high-pressure jet.
7.5 Replacingthe fuse
7.6 Cleaningor replacingwindows
7.6.1 Optical side portwindows
8/11/2019 Manual FTLA2000.pdf
48/92
S e c t i o n 7 . 6 C l e a n i n g o r r e p l a c i n g w i n d o w s
38 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
If grease or other contaminantsare on the window surfaces:
windows replacement may be necessary. Optical side port windowsurfaces should never be wiped.
7.6.1.2 Replacing the windows
For spectrometers with hygroscopic beamsplitter (FTLA2000-
100, FTLA2000-102, and FTLA2000-155), do not replace
windows if the relative humidity is higher than 50%.
Windows are optical components and should be handled withcare. Never touch their optical surfaces with bare hands. Hold
them by the rim.
1. Leave the spectrometer on.
2. Unscrew the four screws holding the window ring.
3. Remove the window assembly from the side plate.
4. Install the new window assembly on the side plate.
5. Tighten the four screws holding the window ring.
7.6.2.1 Cleaning the windows
If dustis on the window surfaces:
Gently blow dry, oil-free instrumentation air or dry nitrogen acrossthe surface. Do not use a high-pressure jet.
If grease or other contaminants are on the windowsurfaces (except forKBr, KCl, and CsI windows):
Clean dirt off with a cotton or optical cloth, or laboratory wiping tissue,wetted with methanol or acetone.
Make sure to dry off the excess liquid.
7.6.2 Other windows
8/11/2019 Manual FTLA2000.pdf
49/92
C H A P T E R 8
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 39
TROUBLESHOOTING
WARNING! Opening the enclosure may result in exposure to laser
radiation and high voltages.
The enclosure is to be opened only by authorized ABB Inc.
Service Personnel.
Laser type: He-Ne
Laser class: 3B as per IEC-60825-1 and
Class IIIa as per 21 CFR 1040.10
Output power: 3.2 mW (maximum)
Wavelength: 632.8 nm
High voltages are present inside the enclosure. Always
proceed with care when the enclosure is open.
This chapter describes the main problems that may be encountered immediately
after applying power to the spectrometer or during its use.
The following table provides troubleshooting information for problems that may
be encountered after applying power to the FTLA2000 Series spectrometer:
Problem Possible cause Solution
POWER LED not lit Spectrometer not plugged in ornot turned on
Plug in spectrometer and turn on.
Fuse blown Replace the fuse.
(See Section 7.5 on page 37.)
Power outlet dead Test outlet.
LED sequence incorrect(on start up)
Internal problem Contact ABB Inc.
(See Section 8.3 on page 43.)
DATA LED not flashing
(after start up)
Internal problem Contact ABB Inc.
(See Section 8.3 on page 43.)
8.1 After power-up
8/11/2019 Manual FTLA2000.pdf
50/92
S e c t i o n 8 . 2 D u r i n g u s e
40 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
The following table provides troubleshooting information for problems that may
be encountered during use of the FTLA2000 Series spectrometer:
Problem Possible cause Solution
No signal Ethernet cable not properlyconnected
FT-IR Ethernet Interface orcomputer network adapter not
properly configured
Connect Ethernet cable and configure devices
as shown in theFT-IR Ethernet Interface andExternal Ethernet Interface User's Guide.
Instrument not selected Make sure the correct instrument is selected inthe software.
Network problem Check with network administrator.Detector saturation Reduce preamplifier gain.
Install a neutral density filter to attenuatethe beam.
Preamplifier / detector failure Check detector status.
Check preamplifier connections.
Contact ABB (see Section 8.3).
Noise only Source breaker switch is OFFor tripped
Set source breaker switch to ON.
Source failure Replace the source (see Section 7.3).
Weak signal IR beam blocked Remove obstacle.
Preamplifier / detectormalfunction Check detector status.Check preamplifier connections.
Contact ABB (see Section 8.3).
Low S/N Electromagnetic interference Observe raw spectra.
Separate from interfering equipment.
Mechanical vibration Isolate or remove vibration source.
Dirty windows Clean or replace windows (see Section 7.6).
Strong, undesired water lines in
spectra (in the region of 1500 cm-1
and 3700 cm-1)
Exhausted desiccant or purgegas failure
Replace desiccant (see Section 7.2).
Check supply of purge gas
Abnormal spectra(see Figure 8-1 and Figure 8-2)
Source contamination Replace the source (see Section 7.3).
If problem persists, inspect to determine sourceof contamination.
Detector saturation Adjust preamplifier gain
8.2 During use
8/11/2019 Manual FTLA2000.pdf
51/92
C h a p t e r 8 T r o u b l e s h o o t i n g
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 41
Figure 8-1. MIR spectra showing source contamination
Figure 8-2. MIR spectra showing detector saturation
Highlycontaminated
Normal
Contaminated
Normal
Saturated
Highlysaturated
8/11/2019 Manual FTLA2000.pdf
52/92
S e c t i o n 8 . 2 D u r i n g u s e
42 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
This section provides examples of typical warning and error messages that may
be displayed on the computer when a hardware problem is encountered.
Note: The actual messages may be somewhat different than those shown here,
depending on the software program and operating system you are using.
8.2.1.1 On selecting a function in the Collectmenu
Possible cause Solution
No instruments were detected by the Ethernetsoftware.
Make sure the FT-IR Ethernet Interface of the spectrometer andthe Ethernet network adapter in the computer are properlyconfigured.
Refer to theFT-IR Ethernet Interface and ExternalEthernet Interface User's Guide.
8.2.1.2 After acquiring a spectrum
Warning Possible cause Solution
Warning: background file issaturated
The detector was saturated
during acquisition of thereference spectrum.
Reduce the preamplifier gain (refer to Section 3.4 onpage 19).
Install a neutral density filter to attenuate the beam
Optical saturation detected The detector was saturated
during acquisition of thereference spectrum.
Reduce the preamplifier gain (refer to Section 3.4 onpage 19).
Install a neutral density filter to attenuate the beam
8.2.1.3 Other messages
Refer to the manuals and release notes for your software program.
8.2.1 Softwaremessages abouthardwareproblems
8/11/2019 Manual FTLA2000.pdf
53/92
C h a p t e r 8 T r o u b l e s h o o t i n g
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 43
If you are unable to solve a problem, contact ABB (see Customer support on
page iii). Please be prepared to give the serial numbers of all units.
Before contacting ABB, please check the following:
Power is applied to each device check the power outlet, the power
cable(s), and the fuse(s).
The spectrometer POWER LED is ON and the DATA LED is flashing
(see Section 3.3.1 on page 19).
The spectrometer source breaker switch (if present) is turned ON.
There is light coming from the spectrometer output(s) (unless an optical
filter is installed).
All cables are properly installed (refer to the installation guide).
All pertinent troubleshooting steps in this manual have been followed.
Note down the serial number of the spectrometer and of each
instrument and accessory.
Using Bomem-GRAMS, try to acquire a reference spectrum, and
a transmittance spectrum. Acquire both spectra open beam (i.e. with no
sample present).
Perform the BASIC test sequence in SpecTest.
Send the following to your ABB representative:
The serial numbers The two spectra The BASIC test report file The name, phone number, and e-mail address of the person to
contact
Indicate if a modem and pcAnywhere are installed on thecomputer
See note on next page.
8.3 Contacting ABB
8/11/2019 Manual FTLA2000.pdf
54/92
S e c t i o n 8 . 4 E x a m p l e s o f n o r m a l s p e c t r a
44 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
Note: SpecTest stores the instrument test reports in the subdirectories
INSTALL, EXTENDED, and BASIC. Each instrument configuration has
a separate subdirectory CONFIGn, withinwhich is a subdirectory for the
current month YYYY_MM. The instrument test report file is named
YYYY MM DD hh mm ss.QAL:
where nis the configuration number
YYYYis the year
MMis the month
DDis the day
hhis the hour
mmis the minute
ssis the second
For example, the instrument test report file for the BASIC test of
configuration 1 performed on May 8, 2000 at 11:23:45 AM is:
C:\SPECTEST\INSTALL\CONFIG1\2000_05\2000 05 08 11 23 45.QAL
Figure 8-3 and Figure 8-4 show typical examples of normal open-beam spectra
for both MIR and NIR spectrometry using different configurations.
8.4 Examples ofnormal spectra
8/11/2019 Manual FTLA2000.pdf
55/92
C h a p t e r 8 T r o u b l e s h o o t i n g
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 45
MIR NIR
Figure 8-3. Examples of normal open-beam spectra
8/11/2019 Manual FTLA2000.pdf
56/92
S e c t i o n 8 . 4 E x a m p l e s o f n o r m a l s p e c t r a
46 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
MIR NIR
Figure 8-4. Examples of normal open-beam spectra
8/11/2019 Manual FTLA2000.pdf
57/92
A P P E N D I X A
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 47
TECHNICAL INFORMATION The following information includes specifications and performance statements
that may be in conflict with other ABB published literature, such as product
flyers and product catalogs. In case of conflicts between information given
below and specifications given in the ABB official flyers and product catalogs,
the later take precedence.
Specifications common to all models are given in the following tables.
All values are nominal.
Refer to the test report provided with your analyzer for validated spectralspecifications.
Rated line voltage: 100-120 VAC or 220-240 VAC(manually selectable)
Rated frequency: 50-60 Hz
Rated input current: 1 A max.
Rated power consumption: 150 VA
Fuse type: T2A/250V (115 V line)T1A/250V (230 V line)
Installation category II
Enclosure: Sealed and desiccated aluminum cast
Overall dimensions (WDH): 20 in. 22.25 in. 12 in.(50.8 cm 56.5 cm 30.4 cm)
Weight: 96 lb. (44 kg)
Operating temperature range: 0C to 30C, stable within 5C
Storage temperature range: -15C to 50C
Relative humidity:
FTLA2000-104, FTLA2000-154,and FTLA2000-160
0% to 95% noncondensing
FTLA2000-100, FTLA2000-102,and FTLA2000-155
0% to 50% noncondensing (operating)
0% to 35% noncondensing (nonoperating)
Pollution degree 2
A.1 Specifications
A.1.1 Electrical
A.1.2 Mechanical
A.1.3 Environmental
8/11/2019 Manual FTLA2000.pdf
58/92
A p p e n d i x A T e c h n i c a l I n f o r m a t i o n
48 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
Resolution: 1 to 128 cm-1in steps of 2
(2 to 128 cm-1in steps of 2for NIR analysis)
Scanning speed: 1cm/sec OPD (typical)
(Speed may vary depending on detector used andfactory spectrometer configuration.)
Purge: Dry, oil-free instrumentation air or nitrogen
5 L/min
ADC: 16-bit with built-in sample and hold,100 kHz maximum sampling rate
Stabilization time: approximately 4 hours after power-up
(8 hours, e.g. overnight, if possible)
Principle: Patented Michelson-type with 2 corner-cuberetroreflectors mounted on wish-bone scan arm
Retroreflectors precision: 2-arc sec corner-cubes (FTLA2000-154,FTLA2000-155, FTLA2000-160)
5-arc sec corner-cubes (FTLA2000-100,FTLA2000-102, FTLA2000-104)
Internal reference: Embedded He-Ne Laser (Class 3B)
A.1.4 Spectrometer
A.1.5 Interferometer
8/11/2019 Manual FTLA2000.pdf
59/92
I ns ta l l i ng Your FTLA2000 Ser iesLabora to ry FT- IR Spec t romete rs Append ix B
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 49
This guide shows how to install your FTLA2000 Series Laboratory FT-IR
spectrometer and connect it to a computer using the built-in FT-IR Ethernet
Interface. It assumes that the computer is supplied by ABB Inc., in which case
the software and the Ethernet network adapter are preinstalled in the computer.
If you are supplying your own computer, refer to the software manuals toinstall the software. The computer must have an Ethernet network
adapter. If necessary, refer to its manual for installation instructions.
Make sure the site meets the requirements shown below.
Condition RequirementsAmbienttemperature
Maintained within 5C
Humidity Models FTLA2000-100, FTLA2000-102, and FTLA2000-155:
0% to 50% noncondensing (operating)
0% to 35% noncondensing (nonoperating)
Other models:
0 to 95% noncondensing
Vibration Some small-amplitude vibration in the supporting table can beaccepted.
Electrical power 100-120 VAC (60 Hz) or 220-240 VAC (50 Hz),
150 VAClean and constant power source, without glitches or failures(no large electrical motors on the same circuit).
Purge gas(if used)
Dry, oil-free instrumentation air or nitrogen
Flow rate: 5 L/min
Tubing: -in (supplied by the customer)
Before plugging in any equipment, make sure that each
individual component is properly configured for the available
line voltage.
On the spectrometer, the currently selected line voltage is visible on the voltageconfiguration plate (behind the sliding fuse cover inFigure B-1on page 50).
Purpose
This symbol refers you to
another manual.
Site preparation
Line voltageconfiguration
8/11/2019 Manual FTLA2000.pdf
60/92
I ns ta l l i ng Your FTLA2000 Ser iesLabora to ry FT- IR Spec t romete rs Append ix B
50 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
All controls and connections, except for the preamplifier gain adjustment, are
located on the control panel (see Figure B-1).
Figure B-1. Control panel
For a network connection, connect the Ethernet connector of thespectrometer to a network using the suppliedstraight-pinnedEthernet cable.
For a private (direct) connection between the spectrometer and computer,
use the supplied crossovercable (see Figure B-2).
Plug the spectrometer power cord into a suitable electrical outlet.
Figure B-2. Private connection between spectrometer and computer
Main controlsand connections
Many different private and
network connection scenarios
are possible. Refer to the FT-
IR Ethernet Interface and
External Ethernet Interface
User's Guidefor complete
information.
Ethernet connector
Resolution switch
Status LEDs
Fuse and voltageconfiguration plate(behind slidingfuse cover)
Power switch
1 or 2 Sourcebreaker-switches
Power connector
8/11/2019 Manual FTLA2000.pdf
61/92
I ns ta l l i ng Your FTLA2000 Ser iesLabora to ry FT- IR Spec t romete rs Appendix B
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 51
The FTLA2000 Series spectrometers are equipped with purge inlets. Purging is
recommended for spectrometers that have hygroscopic optical components
(models FTLA2000-100, FTLA2000-102, and FTLA2000-155) or if ambient
water vapour may interfere with measurements.
Install the female Swagelok connector (included with the spectrometer) anda T-connector on your source of purge gas.
Connect the female Swagelok connector to the main purge inlet connectoron the spectrometer (see Figure B-3).
Connect the purge inlets for the Arid-Zone sample station as shown.
When using a purged sampling accessory, connect the middle purge inletas well.
The recommended purge flow rate into the spectrometer is 5 L/min. Oncethe spectrometer is purged, a continuous flow rate of 1 L/min is normallysufficient to maintain the purged condition.
When a purged sampling accessory is not being used, disconnect the tubing from
this inlet, and block the tubing to prevent purge gas from escaping.
Figure B-3. Purge inlet connectors
Purge gasconnections (if used)
8/11/2019 Manual FTLA2000.pdf
62/92
I ns ta l l i ng Your FTLA2000 Ser iesLabora to ry FT- IR Spec t romete rs Append ix B
52 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
Before using the spectrometer, read the safety information
presented in the FTLA2000 Series Laboratory FT-IR
Spectrometers Users Guide.
Turn the power ON for the spectrometer, the computer and monitor, and
peripheral components, where applicable.
Note: Leave the spectrometer ON at all times. This not only ensures thermal
stability, but also helps protect the spectrometer from humidity.
Immediately after turning on the FTLA2000 Series spectrometer, check that
the sequence of status LEDs on the control panel is as follows:
POWER ON
SCAN ON then OFF (it may take a few minutes to turn OFF)
ZPD ON then OFF (it may take a few minutes to turn OFF)
DATA Flashing ON and OFF (usually after a few minutes).
Note: During operation, the POWER LED should be ON and the DATA LED
should be flashing.
If you encounter problems after applying power to the spectrometer, referto Troubleshooting in the FTLA2000 Series Laboratory FT-IR
Spectrometers Users Guide.
If the computer is supplied by ABB, the equipment is preconfigured for a private
connection between the spectrometer and computer (see Figure B-2).
For any other type of connection, or if you are supplying your owncomputer, refer to the FT-IR Ethernet Interface and External Ethernet
Interface User's Guidefor configuration information.
Before using the spectrometer for quantitative analysis, or as part of an ABB
dedicated analyzer(e.g. PROTA), you must validate it.
Refer to the section entitled Validation in the FTLA2000 SeriesLaboratory FT-IR Spectrometers Users Guide for information on thevalidation tests required.
Safety
Start up
Ethernetconfiguration
Validation
8/11/2019 Manual FTLA2000.pdf
63/92
8/11/2019 Manual FTLA2000.pdf
64/92
Replac ing the NIR Sourcein Your FTLA2000 Ser ies
Labora to ry FT- IR Spec t romete r Append ix C
54 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
4. Remove the source assembly holding screws located under the rear heat
sink portion of the spectrometer (see Figure C-3 below). Lift the source
assembly slightly out of the spectrometer.
Figure C-3. Bottom of source assembly
Remove only the four screws located toward the outside of
the support plate.d The screws located near the center of
the plate are used for the alignment of the source.
5. Loosen the two screws that fasten the electrical leads of the source assembly
(see Figure C-4 on page 55) and disconnect the two leads.
6. Unscrew the screws that hold down the shield on the source assembly
(see Figure C-4 on page 55). Remove the shield from the source assembly.
dThe source assembly comes
prealigned on its circular
support plate. However, fine-
tuning of the source is
required after installation.
To minimize the adjustment
required, do not move
the alignment screws until you
are ready to align the source.
Source assemblyholding screws (4)
Sourcealignmentscrews (4)
Support plate
8/11/2019 Manual FTLA2000.pdf
65/92
Replac ing the NIR Sourcein Your FTLA2000 Ser ies
Labora to ry FT- IR Spec t rometer Append ix C
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 55
Figure C-4. Removing the source assembly
7. Remove the NIR bulb from its socket. The bulb is held by two clips
(see Figure C-5 below).
Figure C-5. NIR bulb on the source assembly
Electrical leadfasteningscrews (2)
Shield holding screws (2)
Shield
Electrical leads (2)
NIR source bulb
Source holding clips (2)
8/11/2019 Manual FTLA2000.pdf
66/92
8/11/2019 Manual FTLA2000.pdf
67/92
Replac ing the NIR Sourcein Your FTLA2000 Ser ies
Labora to ry FT- IR Spec t rometer Append ix C
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 57
Figure C-6. Aligning the NIR source (with enlargement of the target)
3. Once the NIR source is aligned, make sure the alignment screws are tight.
Check that the image has not moved and remove the target from the sample
holder of the spectrometer (see Figure C-6 above).
Validate the spectrometer and/or the application.
Refer to the Validationsection of the FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide, and to your Operation Qualification
procedure.
Validation
Target
8/11/2019 Manual FTLA2000.pdf
68/92
8/11/2019 Manual FTLA2000.pdf
69/92
Rep lac ing the MIR sourceI N YO U R FTLA2000 S E R I E S
LA B O R AT O R Y FT- IR S P E C T R O M E T E R AP P E N D I X D
FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide 59
Rep lac ing the MIR Source
This appendix shows how to replace the internal MIR (mid infrared) source of
your FTLA2000 Series Spectrometer.
Before proceeding, read the safety information presented in
the FTLA2000 Series Laboratory FT-IR Spectrometers Users
Guide.
Do not replace the source while the source power is on;
damage to electronic components may occur.
Note: Replacement of the source does not require opening the spectrometer.
To remove an internal MIR source assembly and install a new one, proceed as
follows:
1. Turn the source and the spectrometer power off (see Figure D-1 below
Figure D-1 below). Unplug the spectrometer.
2. Make sure no accessories are attached to the spectrometer.
3. Carefully lift up the front end of the spectrometer (see Figure D-1 below)
and place it on its heat sink (see Figure D-2 below).
Figure D-1. FTLA2000 spectrometer Figure D-2. Spectrometer resting on its heat sink
D.1 Purpose
D.2 Safety
D.3 Removingthe MIR source
Power andsourceswitches Source
assemblysupport plate
Heat sink
8/11/2019 Manual FTLA2000.pdf
70/92
Rep lac ing the MIR sourcein Your FTLA2000 Ser ies
Labora to ry FT- IR Spec t romete r Append ix D
60 FTLA2000 Series Laboratory FT-IR Spectrometers Users Guide
4. Remove the source assembly holding screws located under the rear heat
sink portion of the spectrometer (see Figure D-3 below). Lift the source
assembly slightly out of the spectrometer.
Figure D-3. Bottom of source assembly
Remove only the four screws located toward the outside of
the support plate.f The screws located near the center of
the plate are used for the alignment of the source.
5. Loosen the two screws that fasten the electrical leads of the source assembly
(see Figure D-4 below) and disconnect the two leads.
Figure D-4. Removing the source assembly
To install a new source and adjust the source voltage, proceed as follows:
1. Connect the new source assembly and a voltmeter as shown in Figure D-5
on page 61.
fThe source assembly comes
prealigned on its circular
support plate.
D.4 Replacingthe MIR source
Electrical leadfasteningscrews (2)
Source assemblyholding screws (4)
Sourcealignmentscrews (4)
Support plate
8/11/2019 Manual FTLA2000.pdf
71/92
8/11/2019 Manual FTLA2000.pdf
72/92
Rep lac ing the MIR sourcein Your FTLA2000 Ser ies
Labora to ry FT- IR Spec t romete r Append ix D
62 FTLA2000 Series Laboratory FT-IR Spectrometer