1

Instruction Manual

INDICATOR StandardIndustrial ElectronicWeight IndicatorModel I5S

Ohaus CorporationOhaus CorporationOhaus CorporationOhaus CorporationOhaus Corporation29 Hanover RoadFlorham Park NJ07932-0900

Page 3: Manual Banlanza Ohaus I5S

2

Unauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notUnauthorized changes or modifications to this equipment are notpermitted. See manual for replacement parts.permitted. See manual for replacement parts.permitted. See manual for replacement parts.permitted. See manual for replacement parts.permitted. See manual for replacement parts.

NOTENOTENOTENOTENOTE: This equipment has been tested and found to comply with the limits for a ClassA digital device, pursuant to part 15 of the FCC rules.

These limits are designed to provide reasonable protection against harmful interfer-ence when the equipment is operated in a commercial environment. This equipmentgenerates, uses, and can radiate radio frequency energy and, if not installed andused in accordance with the instruction manual, may cause harmful interference toradio communications. Operation of this equipment in a residential area is likely tocause harmful interference in which case the user will be required to correct theinterference at his own expense.

This Class A, digital apparatus meets all requirements of the Canadian Interference.Causing Equipment Regulations.

Cet appareil numerique de la classe A respecte toutes les exigences du Reglementsur le material broilleur du Canada.

This device corresponds to requirements stipulated in 90/384/EEC andfeatures radio interference suppression in compliance with valid ECRegulation 89/336/EEC. Note: The displayed value may be adverselyaffected under extreme electromagnetic influences, eg. when using a

ra- dio unit in the immediate vicinity of the device. Once the interference hasbeen rectified, the product can once again be used for its intended purpose. Thedevice may have to be switched on again.

Cet appareil correspond aux exigences selon la norme 90/384/CEE et est déparasitéconformément à la directive de la CE 89/336/CEE en vigueur. Remarque: Dans desconditions d’influences électromagnétiques extrêmes, par exemple en casd’exploitation d’un appareil radio à proximité immédiate de I’appareil la valeurd’affichage risque d’être influencée. Une fois que l’influence parasite est terminée,le produit peut être de nouveau utilisé de manière conforme aux prescriptions; le caséchéant, il est nécessaire de le remettre en marche.

Dieses Gerät entspricht den Anforderungen nach 90/384/EWG und ist funkentstörtentsprechend der geltenden EG-Richtlinie 89/336/EWG. Hinweis: Unter extremenelektromagnetischen Einflüssen z.B. bei Betreiben eines Funkgerätes in unmittelbarerNähe des Gerätes kann eine Beeinflussung des Anzeigewertes verursacht werden.Nach Ende des Störeinflusses ist das Produkt wieder bestimmungsgemäss benutzbar,ggfs. ist ein Wiedereinschalten erforderlich.

CLASSIFICATIONSCLASSIFICATIONSCLASSIFICATIONSCLASSIFICATIONSCLASSIFICATIONS The I5S has been tested and found to comply with the followingstandards:

• National Institute of Standards and Technology Handbook 44 Class III and IIIL

Page 4: Manual Banlanza Ohaus I5S

3

TABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTSTABLE OF CONTENTS

Chapter 1 Introduction and Installation 5Introduction ........................................................................................ 6Description ......................................................................................... 6Unpacking .......................................................................................... 7Assembly ............................................................................................ 7Scale Base Connections .................................................................... 7Capacity Card .................................................................................... 8Closing and Setup Protection ............................................................ 8Power Options .................................................................................... 9AC Adapter ......................................................................................... 9Battery Installation ............................................................................. 9Turning the I5S ON ............................................................................ 10Display Indicator Functions ................................................................ 10To turn the I5S OFF ........................................................................... 10

Chapter 2 Setup and Calibration....................................................... 11Setup Mode and the I5S Menu Structure ........................................... 12To Access the Main Menu and Change Settings ............................... 12To Exit a Submenu, Save Settings and Return to the Main Menu .... 12To Exit the Main Menu and Return to the Weighing Mode ................ 12I5S Menu Structure Chart .................................................................. 13Setup Submenu ................................................................................. 14Legal for Trade ................................................................................... 14Zero .................................................................................................... 15Calibration Unit .................................................................................. 16Full Scale Capacity ............................................................................ 17Graduation Size ................................................................................. 18Calibration Point ................................................................................. 19Menu Lock-out Switch Setting ........................................................... 20End/Store Settings ............................................................................. 20Scale Submenu .................................................................................. 21Reset to Factory Settings ................................................................... 21Averaging Level ................................................................................. 22Stability Level ..................................................................................... 24Unit Selection ..................................................................................... 25Auto-Zero Tracking ............................................................................ 26Auto-Off Timer .................................................................................... 27Retain Zero Data ................................................................................ 28End/Store Settings ............................................................................. 28Calibration Submenu ......................................................................... 29Procedure ........................................................................................... 29

Page 5: Manual Banlanza Ohaus I5S

4

Chapter 2 Setup and Calibration (Cont.)RS232 Submenu ................................................................................ 30Menu Protection ................................................................................. 30Legal for Trade Sealing ...................................................................... 31Audit Trail Seal ................................................................................... 32Paper seal .......................................................................................... 32Lead and Wire Seal ........................................................................... 32

Chapter 3 Operation .......................................................................... 34Selecting a Weighing Unit .................................................................. 35Weighing ............................................................................................ 35Re-Zeroing ......................................................................................... 36

Chapter 4 Troubleshooting ................................................................ 37Troubleshooting ................................................................................. 38Error Codes ........................................................................................ 39Care and Maintenance ....................................................................... 40Service Information ............................................................................ 40

Chapter 5 Specifications, Replacement Parts and Accessories ........ 41Specifications ..................................................................................... 42Replacement Parts ............................................................................. 43Accessories ........................................................................................ 43

Page 6: Manual Banlanza Ohaus I5S

5

Chapter 1Chapter 1Chapter 1Chapter 1Chapter 1IntrIntrIntrIntrIntr oduction andoduction andoduction andoduction andoduction andInstallationInstallationInstallationInstallationInstallation

If the I5S indicator has alreadybeen installed, you may skip mostof this chapter. You may howeverneed to refer back to it if you changescale bases or the power supply,or need to access the menu lock-out switch.

This chapter contains the follow-ing information:

• Introduction and description.

• Unpacking the indicator.

• Scale base connections.

• Power connections.

• Turning the I5S indicator ON.

Introduction and Installation 11111

Page 7: Manual Banlanza Ohaus I5S

6

IntroductionIntroductionIntroductionIntroductionIntroduction

This manual covers installation, operation and troubleshooting for the Ohaus I5SIndicator Standard. To ensure proper operation of the indicator, please read thismanual completely.

DescriptionDescriptionDescriptionDescriptionDescription

The Ohaus I5S Indicator Standard is a rugged, reliable electronic weight indicatorthat is made in the U.S.A. and is designed to be easy to operate. It can be easilyconnected to Ohaus or other brands of Scale Bases. The I5S Indicator has beenNTEP tested and found to comply with Class III and IIIL requirements of NIST HB44for 1:5000 performance in pounds, kilograms, ounces or grams. The I5S Indicatorcan be wall, bench or tower mounted. The I5S Indicator can drive up to four 350ohm load cells and provides capacity selections up to 10,000 lbs/kgs with amaximum resolution of 1:20,000. The indicator is constructed of extrudedaluminum with removable plastic end caps and is extremely durable. The I5Soperates from six Alkaline "C" batteries and can also be powered externally usingthe AC adapter supplied. A six and a half digit LCD display which is 0.6 inches/1.5 centimeters in height provides easy visibility when working at distances fromthe indicator. Two switches mounted on the front panel enable simple set up ofconfiguration and feature parameters. A menu lock-out switch can be set to lockout various functions of the indicator to prevent settings from being changed.Accessories include: an RS232 Interface Kit, a Wall Mount Accessory kit, and aTower Mounting Accessory Kit.

I5S INDICATOR Standard

11111 Introduction and Installation

Page 8: Manual Banlanza Ohaus I5S

7

UnpackingUnpackingUnpackingUnpackingUnpacking

Unpack the I5S indicator at the installation site and verify that the following itemswere included in the carton:

• an AC Adapter • 2 security screws• a capacity card • 2 washers• a warranty card• instruction manual

It is recommended to save the carton and packing material for storing, transportingthe indicator or returning it for service.

AssemblyAssemblyAssemblyAssemblyAssembly

Assembly consists of making the necessary electrical connections, and applyingpower to the unit.

Scale Base ConnectionsScale Base ConnectionsScale Base ConnectionsScale Base ConnectionsScale Base Connections

1. Remove two screws from the rightend cap using a phillips screwdriver.

2. Slide the printed circuit board par-tially out of the housing and discon-nect the battery cable connector.

3. Pass the scale base cable throughthe liquid tight connector on the endcap.

Housing

End Cap

Liquid TightConnector

Terminal Block forScale Base WireConnections(Bottom)

PowerConnector

Battery CableConnector

Introduction and Installation 11111

Page 9: Manual Banlanza Ohaus I5S

8

Scale Base Wire Connections

Bottom ViewPC Board

+ EXCITATION

+ SIGNAL

- EXCITATION

- SIGNALGROUND

4. Connect the scale base wires to theterminal block located on the bot-tom of the printed circuit board asshown in the illustration.

Lock-outSwitch

Battery CableConnector

Slot for BatteryCable

Top ViewPC Board

5. Reconnect the battery connector tothe circuit board.

6. Push the circuit board back into thehousing.

Capacity CardCapacity CardCapacity CardCapacity CardCapacity Card

Type or write the capacity and readabil-ity data for the scale base on one of theperforated cards and slide it under thefront panel label near the display.

Closing and Setup ProtectionClosing and Setup ProtectionClosing and Setup ProtectionClosing and Setup ProtectionClosing and Setup Protection

• If the setup information (explained inChapter 2) does not need to be pro-tected from unauthorized changes,position the battery cable coming fromthe end cap into the slot on the end ofthe PC board, then replace the endcap.

• If you intend to lock out changes to thesetup selections you make, leave theend cap in position but do not tightenthe screws. You will need to accessthe menu lockout switch located onthe circuit board following the setupprocedure.

• If the unit is being powered by batter-ies during setup rather than the ACadapter, and you want to lock outsetup changes, replace the end capnow and remove it after setup to setthe menu lock-out switch.

OffOn

11111 Introduction and Installation

Page 10: Manual Banlanza Ohaus I5S

9

Power OptionsPower OptionsPower OptionsPower OptionsPower Options

The I5S may be operated using the ACAdapter supplied, or 6 Alkaline “C” bat-teries (not supplied). It is recommendedthat when the I5S is operated from bat-teries, the Auto-Off Timer feature maybe turned on to extend battery life.

AC AdapterAC AdapterAC AdapterAC AdapterAC Adapter

Plug the AC adapter connector into thereceptacle in the right end cap and plugthe adapter into a convenient AC outlet.

Battery InstallationBattery InstallationBattery InstallationBattery InstallationBattery Installation

1. Remove the two screws from theleft end cap using a phillips screw-driver.

2. Remove the end cap.

3. Insert 6 Alkaline "C" batteries in thebattery compartment orienting thepositive (+) and negative (-) ends asshown on the label at the rear of theindicator.

4. Replace the end cap and

two screws.

6 "C" Size AlkalineBatteries

End Cap

Introduction and Installation 11111

Page 11: Manual Banlanza Ohaus I5S

10

kg Lights to indicate currentweighing unit is kilograms.

g Lights to indicate currentweighing unit is grams.

oz Lights to indicate currentweighing unit is avoirdupoisounces.

Indicates displayed value isstable.

Lights when scale is at centerof zero within + 1/4 d.

lobAt Indicates low battery condi-tion.

Indicates negative value.

lb Lights to indicate currentweighing unit is pounds.

Turning The Turning The Turning The Turning The Turning The IIIII5S ONONONONON

After all connections have been made,

turn the I5S ON by pressing .

When first turned on, all display seg-ments and enabled indicators will lightbriefly, and then clear. This is followedby the software revision Sr n.n and ifLegal for Trade (LFT) is ON, a calibra-tion event counter (AC xxx) and afeature event counter (FC xxx).

Display Indicator FunctionsDisplay Indicator FunctionsDisplay Indicator FunctionsDisplay Indicator FunctionsDisplay Indicator Functions

To turn the To turn the To turn the To turn the To turn the IIIII5S OFF,5S OFF,5S OFF,5S OFF,5S OFF,

Press and hold until the display

indicates OFF, then release it.

Introduction and Installation 11111

Page 12: Manual Banlanza Ohaus I5S

11

Chapter 2Chapter 2Chapter 2Chapter 2Chapter 2Setup andSetup andSetup andSetup andSetup andCalibrationCalibrationCalibrationCalibrationCalibration

The Setup mode enables you toprogram and calibrate the I5S.

When you first install the I5S, youmust use the Setup mode to enterinformation about the scale basebeing used and to perform calibra-tion. You may also use it to pro-gram options that will customizethe I5S for your specific weighingrequirements.

This Chapter explains how to:

• Access and use the Setupmode.

• Calibrate the indicator.

• Prevent unauthorized setupchanges using the menu lock-out switch.

• Seal the indicator for Legal forTrade installations.

Setup and Calibration 22222

Page 13: Manual Banlanza Ohaus I5S

12

Setup Mode and the Setup Mode and the Setup Mode and the Setup Mode and the Setup Mode and the IIIII5S Menu Structure5S Menu Structure5S Menu Structure5S Menu Structure5S Menu Structure

Programmable features of the I5S are contained in menus which are accessed inthe Setup mode. The chart on page 2-3 shows how these menus are structured.

To access the main menu and change settings:To access the main menu and change settings:To access the main menu and change settings:To access the main menu and change settings:To access the main menu and change settings:

1. Make sure the indicator is OFF and the menu protection lockout switch on thePC board is in the OFF position.

2. Press and hold until SETUP is displayed, then release it.

For first time setup, you must go through all of the submenus to configure theindicator. After the indicator is in use, you can make changes to individual settingsand bypass others.

The following rules apply when using the setup mode:

• Press to enter the displayed submenu or select a displayed setting.

• Press to change the displayed submenu or setting.

To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:To exit a submenu, save settings and return to the main menu:

Repeatedly press until END is displayed, then press .

To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:To exit the main menu and return to the weighing mode:

You must be in the main menu with either SETUP, SCALE, CAL or RS232displayed.

Repeatedly press until END is displayed, then press to exit and return

to the weighing mode.

22222 Setup and Calibration

Page 14: Manual Banlanza Ohaus I5S

13

SCALE

Resetto FactorySettings

Averaging LevelHI or LO

Stability Level0.5, 1 or 3

Units

lb ON or OFF

kg ON or OFF

oz ON or OFF

g ON or OFF

End

Auto Zero Tracking0.5, 1, 3 or OFF*

Auto Off TimerON (2 min.) or OFF

Retain Zero DataON or OFF

End/Store Settings

SETUP

Legal For TradeON or OFF

Zero2%, 100%

Calibration Unitlb or kg

Full Scale Capacity5, 7, 10, 15, 20, 25,30, 40, 50, 60, 75,100, 125, 150, 200,250, 300, 400, 500,600, 700, 750, 1000,1500, 2000, 2500,3000, 5000, 7500,10000

Graduation Size.0005, .001, .002,.005, .01, .02, .05,.1, .2, .5, 1, 2, 5, 10

Calibration Point5, 7, 10, 15, 20, 25,30, 40, 50, 60, 75,100, 125, 150, 200,250, 300, 400, 500,600, 700, 750, 1000,1500, 2000, 2500,3000, 5000, 7500,10000

Menu Lockout Switch

Setup ON or OFFScale ON or OFFCal ON or OFFRS232 ON or OFFEnd

End/Store Settings

END

IIIII5S Menu Structure Chart5S Menu Structure Chart5S Menu Structure Chart5S Menu Structure Chart5S Menu Structure ChartMAIN MENU

NOTE:RS232 menu appears onlywhen RS232 option isinstalled.

RS232CAL

CALIBRATIONperform

• Press to enter the displayed submenu

or select a displayed setting.

• Press to change the displayed submenu

or setting.

* Not available with LFT set to ONFactory default settings are shown in underlined and boldface type.

Setup and Calibration 22222

Page 15: Manual Banlanza Ohaus I5S

14

Setup SubmenuSetup SubmenuSetup SubmenuSetup SubmenuSetup Submenu

1. With the indicator OFF, press and

hold until SETUP is dis-

played, then release it.

2. When SETUP is displayed, press

to enter the submenu. The

first parameter, Legal For Trade, willbe displayed.

Legal For Trade (LFT)Legal For Trade (LFT)Legal For Trade (LFT)Legal For Trade (LFT)Legal For Trade (LFT)

This parameter is used to configure theI5S for Legal for Trade or non Legal forTrade use. Setting this parameter to ONautomatically sets certain parameters forlegal for trade requirements, disablesothers, and limits overall display resolu-tion to 1:5,000.

To access the current setting (ON or

OFF), press when LFT is dis-

played.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press . To proceed to the

next parameter (Zero), press

when LFT is displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu. To exit mainmenu and return to a weighing mode,

repeatedly press until END is dis-

played then press . The scale is

now in a weighing mode.

Legal for Trade OFFZero 100% CalibrationUnit lb Full ScaleCapacity 5000Graduation Size 1Calibration Point 5000Menu Lock-out Switch

Setup ONScale ONCalibration ONRS232 ON

SETUP MENUSETUP MENUSETUP MENUSETUP MENUSETUP MENUFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTS

22222 Setup and Calibration

Page 16: Manual Banlanza Ohaus I5S

15

ZeroZeroZeroZeroZero

Zero specifies the percentage of fullcapacity load (2% or 100%) that may be

cleared by pressing .

☞ A 2% Zero setting is used for hop-per scales or other large scaleswhere an accidental zero wouldlose the current weight. A 2% Zerois also required for Canadian Le-gal for Trade applications. Benchand counter scales normally use100% Zero setting.

To access the current setting, press

when ZERO is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Cali-

bration Unit), press when ZERO is

displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Setup and Calibration 22222

Page 17: Manual Banlanza Ohaus I5S

16

Calibration UnitCalibration UnitCalibration UnitCalibration UnitCalibration Unit

The I5S may be calibrated using poundsor kilograms. Select the desired calibra-tion unit as follows:

To access the current setting (lb or kg),

press when CAL UN is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Full

Scale Capacity), press when CAL

UN is displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

22222 Setup and Calibration

Page 18: Manual Banlanza Ohaus I5S

17

Full Scale CapacityFull Scale CapacityFull Scale CapacityFull Scale CapacityFull Scale Capacity

The maximum capacity of the load re-ceiver is entered for Full Scale Capacityusing the same weighing unit selectedfor the Calibration Unit.

☞ If Legal for Trade is selected, reso-lution will be limited to a maximumof 5,000 divisions.

To access the current setting, press

when FS CAP is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Gradu-

ation Size), press when FS CAP is

displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Available CapacitiesAvailable CapacitiesAvailable CapacitiesAvailable CapacitiesAvailable Capacities

5, 7, 10, 15, 20, 25, 30, 40,50, 60, 75, 100, 125,150, 200,250, 300, 400, 500, 600, 700,750, 1000, 1500, 2000, 2500,3000, 50005000500050005000, 7500, 10000

Setup and Calibration 22222

Page 19: Manual Banlanza Ohaus I5S

18

Graduation SizeGraduation SizeGraduation SizeGraduation SizeGraduation Size

This parameter specifies the incrementsin which the display will count.

If Legal For Trade is set to ON, resolu-tion (Capacity ÷ Graduations) is limitedto a maximum of 5000 divisions. Thegraduation selections available are lim-ited by the capacity entered under FullScale Capacity to prevent a combinationwhich exceeds 5,000 divisions.

If the graduation size selected does notmeet the resolution requirement, theSetup menu cannot be exited. An errormessage is displayed and you are auto-matically returned to the GRAD param-eter. A valid graduation value must beselected.

To access the current setting, press

when GRAD is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Cali-

bration Point), press when GRAD

is displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Complete List of GraduationComplete List of GraduationComplete List of GraduationComplete List of GraduationComplete List of GraduationSelectionsSelectionsSelectionsSelectionsSelections

.0005, .001, .002,

.005, .01, .02, .05,

.1, .2, .5, 11111, 2, 5, 10

22222 Setup and Calibration

Page 20: Manual Banlanza Ohaus I5S

19

Calibration PointCalibration PointCalibration PointCalibration PointCalibration Point

The calibration point is the weight valuethat will be used to calibrate the I5S. Theavailable values are listed in the adja-cent table. A value equal to or less thanthe full scale capacity value must beselected.

To access the current setting, press

when CAL PT is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Menu

Lockout Switch Setting), press

when CAL PT is displayed.

To exit the Setup submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Available ValuesAvailable ValuesAvailable ValuesAvailable ValuesAvailable Values

5, 7, 10, 15, 20, 25, 30, 40,50, 60, 75, 100, 125, 150,200, 250, 300, 400, 500,600, 750, 1000, 1500, 2000,2500, 3000, 50005000500050005000, 7500,10000

Setup and Calibration 22222

Page 21: Manual Banlanza Ohaus I5S

20

Menu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch SettingMenu Lockout Switch Setting

The menu lock-out switch must be in theOFF position to change settings in thismenu. This menu specifies which sub-menus will be locked out when the menulockout switch is enabled. An ON settinglocks out the selected menus. To accessthe current settings (ON or OFF), press

when LOCSW is displayed.

NOTENOTENOTENOTENOTE: When Legal for Trade is set ON,you are required to lock the Setup and Calmenus.1. The first submenu -SETUP- will be displayed.

2. Press to access the currentsetting.

3. To change the displayed setting,

press .

4. To accept the displayed setting, press

. To proceed to the next

submenu, press .

Set each submenu in the same manner.

When END is displayed, press to

return to the Setup submenu.

To exit the Setup submenu and store your

settings, press to display END, then

press to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until END

is displayed then press . The scale

is now in a weighing mode.

End/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store Settings

You must use END to exit store Setupsubmenu settings. If you do not use ENDto exit the submenu, settings you havemade will not be stored.

(when option is installed)

22222 Setup and Calibration

Page 22: Manual Banlanza Ohaus I5S

21

Scale SubmenuScale SubmenuScale SubmenuScale SubmenuScale Submenu

If you are already in the main menu, you

can repeatedly press until SCALE

is displayed. If you are not already in the

main menu, proceed as follows:

1. With the indicator OFF, press and

hold until SETUP is dis-

played, then release it.

2. Press to display SCALE.

3. When SCALE is displayed, press

to enter the Scale submenu.

The first parameter, Reset to Fac-tory Settings, will be displayed.

Reset to Factory SettingsReset to Factory SettingsReset to Factory SettingsReset to Factory SettingsReset to Factory Settings

This parameter is used to reset all scalesubmenu parameters to the factory de-fault settings outlined in the adjacenttable.

To reset all scale submenu settings,

press when RESET is displayed.

When is released, DONE will be

displayed briefly, indicating RESET iscompleted.

To proceed to the next parameter (Av-

eraging Level), press when RE-

SET is displayed.

To exit the Scale submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

SCALE MENUSCALE MENUSCALE MENUSCALE MENUSCALE MENUFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTSFACTORY DEFAULTS

Averaging Level LOStability Level 0.5dUnits:

lb ONkg ON g ONoz ON

Auto Zero Tracking 0.5dAuto Off Timer OFFRetain Zero Data OFF

Setup and Calibration 22222

Page 23: Manual Banlanza Ohaus I5S

22

Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)Reset to Factory Settings (Cont.)

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Averaging LevelAveraging LevelAveraging LevelAveraging LevelAveraging Level

Averaging level compensates for vibra-tion or excessive air currents on thescale base. During operation, the indi-cator continually takes weight readingsfrom the load cell. Successive readingsare then digitally processed to achieve astabilized display. Using this feature,you specify how much processing youneed.

HI and LO settings are available.

HI setting:More processing, greater stability andslower stabilization time.

LO setting:Less processing, less stability and fasterstabilization time.

To access the current setting, press

when AL is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

22222 Setup and Calibration

Page 24: Manual Banlanza Ohaus I5S

23

To proceed to the next parameter (Sta-

bility Level), press when AL is

displayed.

To exit the Scale submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Setup and Calibration 22222

Page 25: Manual Banlanza Ohaus I5S

24

Stability LevelStability LevelStability LevelStability LevelStability Level

The stability level specifies how much adisplayed weight may change (in gradu-ations) while the stability indicator re-mains ON. When the displayed weightchanges beyond the allowable range,the stability indicator turns OFF. Stabil-ity values of 0.5, 1 or 3 graduations maybe selected.

To access the current setting, press

when STAB is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Unit

Selection), press when STAB is

displayed.

To exit the Scale submenu and store

your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

22222 Setup and Calibration

Page 26: Manual Banlanza Ohaus I5S

25

Unit SelectionUnit SelectionUnit SelectionUnit SelectionUnit Selection

The I5S permits you to display weight inpounds, kilograms, ounces and grams.To enable or disable various weighingunits, use the following procedure.

To access the current settings, press

when UNITS is displayed.

The first unit, lb will be displayed.

1. Press to access the current

setting.

2. To change the displayed setting,

press .

3. To accept the displayed setting,

press .

To proceed to the next unit, press .

Set each unit in the same manner.

When END is displayed, press to

return to the Scale submenu.

To proceed to the next parameter (Auto

Zero Tracking), press when UNITS

is displayed. To exit the Scale submenuand store your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

NOTE: ON is the default settingfor lb, kg, oz, and g.

Setup and Calibration 22222

Page 27: Manual Banlanza Ohaus I5S

26

Auto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero TrackingAuto Zero Tracking

Auto Zero Tracking compensates forchanges in zero reading due to temper-ature changes or build up of smallamounts of debris on the scale baseplatform. Select (0.5, 1, or 3 divisions orOFF), the value of weight change thatthe indicator will disregard in order tomaintain a zero reading.

Auto Zero Tracking occurs only whenthe display indicates zero regardless ofthe load on the scale base.

NOTENOTENOTENOTENOTE: If LFT is set ON, the Auto-ZeroTracking OFF setting is not available.

To access the current setting, press

when Auto-0 is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Auto

Off Timer), press when Auto-0 is

displayed.

To exit the Auto-Zero Tracking submenuand store your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

22222 Setup and Calibration

Page 28: Manual Banlanza Ohaus I5S

27

Auto Off TimerAuto Off TimerAuto Off TimerAuto Off TimerAuto Off Timer

This parameter can be set to automati-cally turn the I5S off when it is idle formore than 2 minutes. This is especiallyuseful when the unit is being batteryoperated and will extend the battery lifesignificantly. This feature can also beswitched off

To access the current setting, press

when A OFF is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To proceed to the next parameter (Re-

tain Zero Data), press when

A OFF is displayed.

To exit the Auto-Off Timer submenu andstore your settings, repeatedly press

until END is displayed, then press

to return to the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

Setup and Calibration 22222

Page 29: Manual Banlanza Ohaus I5S

28

Retain Zero DataRetain Zero DataRetain Zero DataRetain Zero DataRetain Zero Data

Set this parameter to ON to save zerodata in permanent storage in case theindicator is turned off or power is lost.This feature is used primarily in hopperand tank scale applications where lossof weight data will have serious conse-quences.

To access the current setting, press

when RETAIN is displayed.

1. To change the displayed setting,

press .

2. To accept the displayed setting,

press .

To exit the Retain Zero submenu and

store your settings, press to dis-

play END, then press to return to

the main menu.

To exit main menu and return to a weigh-

ing mode, repeatedly press until

END is displayed then press . The

scale is now in a weighing mode.

End/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store SettingsEnd/Store Settings

You must use END to exit store Scalesubmenu settings. If you do not use ENDto exit the submenu, settings you havemade will not be stored.

22222 Setup and Calibration

Page 30: Manual Banlanza Ohaus I5S

29

Calibration SubmenuCalibration SubmenuCalibration SubmenuCalibration SubmenuCalibration Submenu

The calibration menu is used to performcalibration.

If you are already in the main menu, you

can repeatedly press until CAL is

displayed and proceed with step 3 of thefollowing procedure. If you are not al-ready in the main menu, start with step 1.

ProcedureProcedureProcedureProcedureProcedure

Before beginning, make sure there is noload on the scale base and calibrationweight equal to the Calibration Point isavailable.

1. With the indicator OFF, press and

hold until SETUP is dis-

played, then release it.

2. Press twice to display CAL.

3. When CAL is displayed, press

to begin calibration.

4. C 0 will be displayed. With no load

on the platform, press .

5. When is released, -C- will be

displayed followed by C n indicat-ing the calibration weight value.

6. Place the weight on the load re-

ceiver and press to accept

the calibration data.

7. The display will show -C-, return tothe weighing mode and display thecalibration weight value with thecurrent unit indicator.

Setup and Calibration 22222

Page 31: Manual Banlanza Ohaus I5S

30

RS232 SubmenuRS232 SubmenuRS232 SubmenuRS232 SubmenuRS232 Submenu

The RS232 submenu only appears inthe main menu if the option is installed inthe indicator. Refer to the instructionsprovided with the accessory for setupand operating information.

Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)Menu Protection (Lockout Switch)

The lockout switch located on the circuitboard can be switched to the ON posi-tion to disable unauthorized changes tothe indicator setup. Only those submenusthat are turned ON in the LOCSW pa-rameter of the SETUP submenu are dis-abled. This is indicated by -SAFE- whenentering a locked out submenu.

If -SAFE- is displayed, submenu settingsmay be reviewed but cannot be changed.

Set the switch as shown in the illustrationand replace the end cap.

Menu Lock-out switch shown inON position to lock out menus.

ON

OFF

22222 Setup and Calibration

Page 32: Manual Banlanza Ohaus I5S

31

Legal For Trade SealingLegal For Trade SealingLegal For Trade SealingLegal For Trade SealingLegal For Trade Sealing

The I5S has been tested and found tocomply with the Class III and IIIL require-ments of NIST Handbook 44.

After the indicator has been tested andfound to comply with applicable regula-tions by a local Weights and Measuresofficial, it may be sealed as follows:

1. Remove the right end cap and makesure that the menu lockout switchon the printed circuit board is set tothe OFF position.

2. Enter the Setup submenu and go tothe LOCSW parameter. Set theSETUP and CAL LOCSW param-eters to ON. Refer to the MenuLock-out Switch Setting on page2-10 for detailed procedures.

3. Set the menu lockout switch on theprinted circuit board to the ON posi-tion.

4. If an audit trail or paper seal will beused, install the End Cap and tightenthe two 6-32 phillips pan headscrews.

5. If a lead and wire seal will be used,replace the 6-32 phillips pan headscrews with the 6-32 hex socketsecurity screws and washers en-closed in the plastic bag containingthe accessories.

Setup and Calibration 22222

Page 33: Manual Banlanza Ohaus I5S

32

22222 Setup and Calibration

Audit Trail SealAudit Trail SealAudit Trail SealAudit Trail SealAudit Trail Seal

Calibration Event CounterCalibration Event CounterCalibration Event CounterCalibration Event CounterCalibration Event Counter

A calibration event counter incrementsthe calibration event number by oneeach time the indicator is calibrated.

This number can be verified each timethe indicator is turned on. It will bedisplayed in the following format wherexxx is the calibration event number.

AC xxx

Feature Event CounterFeature Event CounterFeature Event CounterFeature Event CounterFeature Event Counter

The feature event counter incrementsthe feature event number by one eachtime any of the following features (pa-rameters) are changed. Changing oneor any combination of these features willconstitute one event. The list of featuresmonitored is as follows:

LFTLFTLFTLFTLFT ZEROZEROZEROZEROZERO CAL UNITSCAL UNITSCAL UNITSCAL UNITSCAL UNITSRETAINRETAINRETAINRETAINRETAIN STB LEVELSTB LEVELSTB LEVELSTB LEVELSTB LEVEL AVG LEVELAVG LEVELAVG LEVELAVG LEVELAVG LEVELAZT LEVELAZT LEVELAZT LEVELAZT LEVELAZT LEVEL AUTO PRINTAUTO PRINTAUTO PRINTAUTO PRINTAUTO PRINT STABLE PRINTSTABLE PRINTSTABLE PRINTSTABLE PRINTSTABLE PRINTFULL SCALEFULL SCALEFULL SCALEFULL SCALEFULL SCALE CAL POINTCAL POINTCAL POINTCAL POINTCAL POINT GRADGRADGRADGRADGRAD

This number can be verified each timethe indicator is turned on. It will bedisplayed in the following format wherexxx is the calibration event number.

FC xxx

These numbers should be recorded onaccompanying documentation or on apaper seal that is affixed to the balancefor future verification.

Page 34: Manual Banlanza Ohaus I5S

33

Setup and Calibration 22222

Paper SealPaper SealPaper SealPaper SealPaper Seal

Apply the paper seal over one of the 6-32phillips pan head screws on the right endcap.

Lead and Wire SealLead and Wire SealLead and Wire SealLead and Wire SealLead and Wire Seal

After the indicator has been inspected bya Weights and Measures official, it maybe sealed using a lead and wire seal asfollows:

1. Replace the 6-32 phillips pan headscrews in the right end cap with thewashers and security screws provid-ed in the I 5S packaging.

2. Pass the wire through the holes in thesecurity screws on the right end cap.

3. Pass the end of the wire through thelead seal and fasten the seal.

LEAD AND WIRE SEAL

PAPER SEAL

Page 35: Manual Banlanza Ohaus I5S

34

Chapter 3Chapter 3Chapter 3Chapter 3Chapter 3OperationOperationOperationOperationOperation

Once the indicator has been setup and scale base connectionshave been made, you can performweighing operations.

If the RS232 interface accessoryis being used, refer to the operat-ing instructions supplied with thataccessory.

33333 Operation

Page 36: Manual Banlanza Ohaus I5S

35

Operation 33333

Selecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing UnitSelecting a Weighing Unit

Repeatedly pressing will cause

the displayed unit indicator to change.Only the units which are set to ON in thesetup mode will appear. If the desiredunit indicator does not appear, refer toChapter 2 for setup instructions.

WeighingWeighingWeighingWeighingWeighing

1. Make sure there is no load on thescale base.

2. Press to obtain a zero read-

ing if necessary.

3. Add the load to the scale base andread the weight on the display.

When the stability indicator ( )ap-pears in the display, the reading isstable.

*

Page 37: Manual Banlanza Ohaus I5S

36

33333 Operation

Re-ZeroingRe-ZeroingRe-ZeroingRe-ZeroingRe-Zeroing

Use the RE-ZERO switch to reset thedisplay to zero. When weighing materialor objects that must be held in a contain-er, re-zeroing enables the indicator tosave the container’s weight in memoryand subtract it from the gross load on thescale base.

1. Press to obtain a zero read-

ing.

2. Place an empty container on thescale base and read its weight onthe display.

3. Press . The display will show

zero and the container weight willbe stored in memory.

4. Add the material to the container.As material is added its net weightwill be displayed.

5. When the loaded container is re-moved, the weight of the containerwill be displayed as a negative (-)value.

Page 38: Manual Banlanza Ohaus I5S

37

Chapter 4Chapter 4Chapter 4Chapter 4Chapter 4TTTTTrrrrroubouboubouboub leshootingleshootingleshootingleshootingleshooting

The information in this chapter isintended to help identify and cor-rect errors that may be made ininstalling or operating the I5S. Thischapter includes:

• A troubleshooting chart.

• A description of messages thatmay appear on the display.

• Information on proper care andmaintenance.

Troubleshooting 44444

Page 39: Manual Banlanza Ohaus I5S

38

TroubleshootingTroubleshootingTroubleshootingTroubleshootingTroubleshooting

PROBABLE CAUSE(S)

Adapter not plugged inor properly connected.

Batteries dead or notproperly connected.

Load on scale base ex-ceeds allowable zero %entered in ZERO pa-rameter of Setup menu(see Chapter 2).

Retain Zero Data is en-abled in Scale menu.

Scale base motion or dis-turbances exceed cen-ter of zero criteria.

Desired unit not set toON in Scale menu.

RS232 communicationparameters set up incor-rectly.

Cable connections.

Scale base discon-nected.

LOCSW set to ON andmenu protection switchset to ON position.

REMEDY

Check power cord con-nections.

Make sure adapter con-nector is plugged all theway into the indicator.

Check battery connector.

Check orientation of thebatteries.

Make sure end caps aresecure on indicator.

Replace batteries.

Remove load on scalebase to less than enteredzero %.

Change allowable zero% in “ZERO” parameterof Setup menu (seeChapter 2).

Normal operation whenthis feature is enabled.

Remove disturbances orreduce motion.

Increase averaging levelin AL menu.

Enable desired unit (seeChapter 2).

Verify communicationparameters.

Check cable connec-tions.

Check connections.

Set menu protectionswitch to OFF position.

SYMPTOM

Unit will not turn on.

Cannot rezero indicator,or will not re-zero whenturned on.

Center of Zero displayindicator erratic or doesnot appear with no loadon scale base.

Cannot display weight indesired weighing unit.

RS232 not working.

Unable to calibrate unit.

44444 Troubleshooting

Page 40: Manual Banlanza Ohaus I5S

39

Troubleshooting 44444

Error CodesError CodesError CodesError CodesError Codes

DISPLAYEDMESSAGE DESCRIPTION

Flashes at 10 second intervals when batteries are weak. Approxi-mately 2 minutes of operating time remain.

Underload condition.

Overload condition.

Err 0: Invalid memory checksum, on either cal data, parameters, or both.Checksum is verified every time I5S is turned on. Check setupparameters in menus and recalibrate. If error persists, a hardwareproblem may exist and unit must be serviced.

Err 1: Overload error. (Interface only). This error shows as all uppersegments on the display.

Err 2: Underload error. (Interface only). This error shows as all lowersegments on the display.

Err 4: Invalid calibration error. This error usually occurs during thecalibration procedure when there is insufficient signal for the ScaleBase as compared to the indicator set up parameters.

In LFT operation, this error will occur if an attempt to calibrate theindicator when the active range of the load cell is less than1.2 mv/v for the active full-scale load. Recheck the load cell rating,set up parameters, and calibration weights.

Err 14: Attempt to zero balance when zero criteria are not met. Zerocriteria can change depending on user settings and whether or notthe indicator is doing a power-on zero or a normal zero. See power-on zero operation for more details.

Err 21: Calibration data does not match current Full Scale, Grad and CalPoint settings. Settings must be restored or the indicator must berecalibrated using the current settings.

Err 22: The current combination of Grad and Full Scale settings exceedsrated resolution (1:5,000 LFT, 1:20,000 non-LFT). This can occurif the user modifies the Full Scale setting, then attempts to exit theSETUP menu without entering the Grad submenu. Once the user

Page 41: Manual Banlanza Ohaus I5S

40

44444 Troubleshooting

entered the Grad submenu, graduation selection will be limited to choices that donot exceed rated resolution.

Err 23: Retain is enabled in the Scale menu, but valid Retain data does notexist. Normal power-on zero operation will occur. The new power-on zero, since Retain is enabled, will be written to the EEPROM andused as valid retain data the next time the indicator is turned on.

Err 24: An attempt was made to process an interface command that iscurrently locked by the lock switch.

Err 25: Attempt to leave the SEL submenu with all units turned off.

Care and MaintenanceCare and MaintenanceCare and MaintenanceCare and MaintenanceCare and Maintenance

To keep your I5S operating properly, observe the following:

• Remove the power pack before cleaning.

• Use only a damp cloth and mild detergent for cleaning. Abrasives may damagethe display window.

• Be careful not to scratch the display window.

• Do not allow any liquid to flow inside the I5S.

• Do not store long periods of time with the batteries installed.

Service InformationService InformationService InformationService InformationService Information

If the troubleshooting section does not resolve or describe your problem, you willneed to contact an authorized Ohaus Service Agent. For Service assistance in theUnited States, call Ohaus Corporation toll-free at (800) 526-0659. An OhausProduct Service Specialist will be available to help you.

Page 42: Manual Banlanza Ohaus I5S

41

Chapter 5Specifications,ReplacementPartsand Accessories

Specifications, Replacement Parts and Accessories 55555

Page 43: Manual Banlanza Ohaus I5S

42

55555 Specifications, Replacement Parts and Accessories

SpecificationsSpecificationsSpecificationsSpecificationsSpecifications

Capacity (Ib or kg) 5 to 10,000*

Graduations (readability) (lb or kg) 0.0005 to 10*

Displayed resolution 1:5,000 LFT or 1:20,000 maximum* (Non LFT)

Weighing modes lb, kg, oz, g*

Over range capacity Capacity plus 9d Stab i l i za -

tion time 3 seconds Auto-zero tracking

capture range

OFF, 0.5, 1 or 3 divisions* Zeroing range

2% or 100% of capacity* Internal resolution 1:200,000

Operating temperature range 32° to

104°F/0° to 40°C (Non LFT) Operating humidity range

10% to 95% RH Span calibration

push-button (selectable up to 100% of scale base

capacity) Weighing system Strain gauge

Load cell excitation voltage 5 V dc

Load cell input sensitivity Up to 3 mV/V (Non LFT)

Load cell drive 60mA at 5V dc (drives up to four 350 ohm

load cells)

Display (in/cm) LCD (0.6/1.5) high

Power requirements AC adapter: 100, 120, 220, 240 V ac,

50/60 Hz or 6 alkaline "C" batteries

Typical battery life 125 hours with one 350 ohm load cell

(without RS232) 55 hours with four 350 ohm load cells

Dimensions

(WxHxD) (in/cm) 7.5 x 6 x 3/19.1 x 15.2 x 7.6

Net weight (Ib/kg) 2.7/1.2

Shipping weight (lb/kg) 4/2

NIST/NTEP NIST HB44 Class III and IIIL nmax=5000d

Certificate of conformance No. 93-112A1

Operating temperature

50° to 104°F/10° to 40°C

Load cell input sensitivity 1.2 to 3 mV/VElectrical ratings (for power 100/120 V UL1910, CSA C22.2 no. 66 andadapters only) 107, 230 V (Europe)/ 240 V (UK) EN60950,

240 V (Australian) IEC 950

CE electromagnetic compatiblity Directive No. 89/336/EEC

EN50081-1 and EN50082-1

* User selectable

Page 44: Manual Banlanza Ohaus I5S

43

Specifications, Replacement Parts and Accessories 55555

Replacement PartsReplacement PartsReplacement PartsReplacement PartsReplacement Parts

OHAUSDESCRIPTION PART NO.

AC Adapters:100/120 V ac North American Plug 90524-77230 V ac, 50 Hz, Euro Plug 90524-75230/240 V ac, 50 Hz, UK Plug 90524-79240 V ac Australian Plug 90524-76

AccessoriesAccessoriesAccessoriesAccessoriesAccessories

OHAUSDESCRIPTION PART NO.

Wall Mount Accessory Kit 77688-01Tower Accessory Kit 77689-01RS232 Interface Kit 77694-01

Page 45: Manual Banlanza Ohaus I5S

44

Limited WarrantyLimited WarrantyLimited WarrantyLimited WarrantyLimited Warranty

Ohaus products are warranted against defects in materials andworkmanship from the date of delivery through the duration of thewarranty period. During the warranty period Ohaus will repair, or,at its option, replace any component(s) that proves to be defectiveat no charge, provided that the product is returned, freight prepaid,to Ohaus.

This warranty does not apply if the product has been damaged byaccident or misuse, exposed to radioactive or corrosive materials,has foreign material penetrating to the inside of the product, or asa result of service or modification by other than Ohaus. In lieu of aproperly returned warranty registration card, the warranty periodshall begin on the date of shipment to the authorized dealer. Noother express or implied warranty is given by Ohaus Corporation.Ohaus Corporation shall not be liable for any consequentialdamages.

As warranty legislation differs from state to state and country tocountry, please contact Ohaus or your local Ohaus dealer forfurther details.

Page 46: Manual Banlanza Ohaus I5S

45

Ohaus Corporation29 Hanover Road,Florham Park, NJ 07932, USATel: (973) 377-9000,Fax: (973) 593-0359

With offices worldwide.

P/N 77680-02 © Ohaus Corporation 1995, 1999 all rights reserved.