manual Autoclave SM200,300,310,500,510-Ver.7

Embed Size (px)

Citation preview

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    1/47

     

    High Pressure Steam Sterilizer

    SM200 SM300 SM500SM310 SM510

    Version 7

     Yamato Scientific America Inc.

    Santa Clara, CA

    Read and apprehend the important warnings in

    this instruction manual prior to use.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    2/47

    Table of Contents

    1. Specifications ................................................................................................................................. 1-2

    2. Safety Information.......................................................................................................................... 3-7

    Safety Symbols.................................................................................................................................... 3Safety Precautions ........................................................................................................................... 4-6Hazardous Material ............................................................................................................................. 7

    3. Identification of Parts................................................................................................................... 8-11

    External................................................................................................................................................ 8Internal............................................................................................................................................ 9-10Control Panel..................................................................................................................................... 11

    4. Installation................................................................................................................................... 12-18

    Environmental Requirements ....................................................................................................... 12-13Power Requirements......................................................................................................................... 14Initial Set Up ................................................................................................................................. 15-18

    5. Operational Precautions............................................................................................................ 19-20

    6. Sample Run Examples.................................................................................................................... 21

    7. Program “MODE” ....................................................................................................................... 22-27

    “MODE” key Information.................................................................................................................... 22Display Symbols................................................................................................................................ 23Flowchart ........................................................................................................................................... 24Setting Sterilization Temperature and Time...................................................................................... 25Setting Dry Temperature and Time................................................................................................... 26Special Program Features................................................................................................................. 27

    8. Run “MENU” ............................................................................................................................... 28-33

    Selection of Operation....................................................................................................................... 28

    “STERILIZE” Procedure.............................................................................................................. 29-30“STERILIZE & DRY” Procedure ................................................................................................. 31-33

    9. Special Control Functions......................................................................................................... 34-35

    View Parameter during Operation..................................................................................................... 34Interrupt Program during Operation .................................................................................................. 35Pressure Relief Switch ...................................................................................................................... 35Checking chamber temperature during stand-by mode.................................................................... 35

    10. Troubleshooting Guide............................................................................................................ 36-37

    Safety................................................................................................................................................. 36Self-diagnostic Circuitry..................................................................................................................... 36Error Code List .................................................................................................................................. 36In case of power failure ..................................................................................................................... 36Problem Solving Chart....................................................................................................................... 37

    11. Wiring Diagram......................................................................................................................... 38-39

    12. Piping Chart ................................................................................................................................... 40

    13. Maintenance.............................................................................................................................. 41-42

    Daily Maintenance............................................................................................................................. 41Weekly Maintenance......................................................................................................................... 41Monthly Maintenance ........................................................................................................................ 42

    14. After Sale Service and Warranty.................................................................................................. 43

    Request for Repair ............................................................................................................................ 43

    15. Replacement Parts List ................................................................................................................ 44

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    3/47

      1

    Specifications

    SM200/300

    Model SM200 SM300 SM500

    System Automatic high pressure steam sterilizer

    Sterilization 105℃~123℃  105℃~128℃  105℃~128℃ Temperature

    Setting Range Drying 150℃~180℃ 

    Maximum Operational Pressure 0.18MPa (26.1psi) 0.2MPa (29.0psi) 0.2MPa (29.0psi)

    Power Requirement AC115V 13A

    (50/60Hz)

     AC115V 15A

    (50/60Hz)

     AC115V 15A

    (50/60Hz)

    Controller HiTec IV CR Type Microprocessor Control

    Sensor 1 (Chamber Temp.) Pt100Ωresistance thermometer sensor

    Sensor 2 (Water Temp.) Thermocouple (Type T)

    Timer 1min. ~  99hours and 59min. 100 ~  999hours

    Heater 1 (Sterilization) 1.3kW 1.7kW 1.7kW

    Heater 2 (Drying) 1.0kW 1.5kW 1.5kW

    Safety

    Main circuit breaker, Over pressure safety valve,

    Self-diagnostic circuitry; Monitors any abnormality of the temp sensor,

    Low water sensor, SSR, and heater

    External Dimensions (W×D×H)16.1”×18.5”×37.4”

    41cm×47cm×95.3cm

    17.3”×20.9”×38.0”

    44cm×53cm×96.5cm

    17.3”×20.9”×42.7”

    44cm×53cm×108.5cm

    Internal Dimensions

    (Diameter, Depth)

    9.4”, 17.5”

    24cm, 44.5cm

    11.8”, 17.5”

    30cm, 44.5cm

    11.8”, 26.2”

    30cm, 66.5cm

    Capacity (cu. ft): 0.8 1.2 1.8

    Capacity (liters): 22 36 51

    Weight143.3 pounds

    65kg

    176.4 pounds

    80kg

    187.4 pounds

    85kg

    Accessories

    Bottle 1

    Basket 2

    Bottom Plate 1

    Steam Saucer 1 

    Instruction Manual 1

    Warranty 1

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    4/47

      2

    Specifications

    SM310/510

    Model SM310 SM510

    System Automatic High pressure steam sterilizer

    Sterilization 105℃~128℃ Temperature

    Setting Range Drying 150℃~180℃ 

    Maximum Operational Pressure 0.2MPa (29.0psi)

    Power Requirement AC220V 9.5A (50/60Hz)

    Controller HiTec IV CR Type Microprocessor Control

    Sensor 1 (Chamber Temp.) Pt100Ωresistance thermometer sensor

    Sensor 2 (Water Temp.) Thermocouple (Type T)

    Timer 1min. ~  99hours and 59min. 100 ~  999hours

    Heater 1 (Sterilization) 2.0kW

    Heater 2 (Drying) 1.5kW

    Safety

    Main circuit breaker, Over pressure safety valve,

    Self-diagnostic circuitry; Monitors any abnormality of the temp sensor,

    Low water sensor, SSR, and heater

    Intend Dimensions (W×D×H):17.3”×20.9”×38.0”

    44cm×53cm×96.5cm

    17.3”×20.9”×42.7”

    44cm×53cm×108.5cm

    Internal Dimensions

    (Diameter, Depth):

    11.8”, 17.5”

    30cm, 44.5cm

    11.8”, 26.2”

    30cm, 66.5cm

    Capacity (cu. ft): 1.2 1.8

    Capacity (liters): 36 51

    Weight176.4 pounds

    80kg

    187.4 pounds

    85kg

    Accessories

    Bottle 1

    Basket 2

    Bottom Plate 1

    Steam Saucer 1 

    Instruction Manual 1

    Warranty 1

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    5/47

      3

    Safety InformationSafety Sym bols

    Safety Information

    This instruction manual contains various symbols and safety information. Ignoring safety information

    can cause such situations as listed below. Please read the following warning and caution signs in thismanual prior to use.

    Warning  Indicates the possibility of serious or fatal injury. (Note 1)

    Caution Indicates the possibility of injury (Note 2) or damage (Note 3) to the

    equipment.

    (Note 1) Serious injury: Bodily harm by electric shock, bone fracture or poisoning which may require

    hospitalization.

    (Note 2) Injury: Bodily harm by electric shock, bone fracture or poisoning which may require

    hospitalization.

    (Note 3) Damage: Any damage on equipment, facility, structure, etc.

    Meaning of graphic in dicat ions 

     

    Shows warnings or caution.

    Specific contents are described aside each sign.

    Shows users important information.

    Specific contents are described aside each sign.

    Shows users important information.

    Specific contents are described aside each sign.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    6/47

      4

    Safety InformationSafety Precautions

    Do not use this unit for any purpose other than its intended use, described in this manual.

    This unit is not explosion proof. Never use in flammable or explosive gas environments.

    Be sure to ground the unit. Electric leak could cause electrical shock or fire.

    Be sure to use a power supply with more than the rated capacity specified in this manual. Use

    of a power supply without the correct rated voltage and current could cause fire or electric

    shock.

    If smoke or any strange odor should disburse from this unit, turn the breaker off immediately

    and pull out the main power cord. Then contact Yamato Scientific. Neglecting this procedure

    can result in fire or electric shock. Never try repairing the unit yourself.

    Forcibly bending, pulling, wrenching or extending the power cord can cause a fire or electricshock.

    Overheat or fire can occur if the power cord is bundled or if an object is on the cord.

    The use of explosive, flammable or such compounds can cause explosion or fire. (See Pg.7)

    Disassembling this unit can cause fire, electric shock or other crisis’s.

    The exhaust is located on the right side of the unit. Possibility of personal injury if you come in

    contact with this area during operation or for a period of time after use.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    7/47

      5

    Safety InformationSafety Precautions

    Do not plug the exhaust. Any obstruction in the exhaust may cause the tank to build pressure

    and become hotter the normal.

    Do not attempt to open the sterilizer lid until the pressure is at “0psi”.

    If you attempt to open the lid when pressurized high pressure steam could

    cause serious bodily injury. Do not stand close to the lid since steam will

    exhaust once opened.

    Do not open drain when vessel is pressurized. Since sterilization water is hot just after

    sterilization, let the water cool before draining.

    The bottle, located in the inside of the front door contains hot water after operation. Let the

    water cool prior to removing the bottle. To avoid possible bodily injury, do not open the door

    during operation.

    The plate, lid and surrounding areas are hot after operation. To avoid bodily injury, do not touch

    these areas. Always wear protective gear when removing a processed load.

    Once the lid is open and all steam has exhausted remove sterilized material with heat-resistant

    gloves to prevent bodily injury.

    Hot steam will disperse out of the louvers located in the back of this unit if a malfunction

    occurs. Do not close the louvers.

    When using the pressure relief switch be sure not to touch the exhaust areas of the sterilizer

    since high pressure steam will exhaust.

    Do not operate this unit without a basket.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    8/47

      6

    Safety InformationSafety Precautions

    In the event of electrical storm turn off the main circuit breaker. Neglecting this procedure can

    result in fire, electric shock or other troubles due to thunderbolts.

    In the event of a power interruption or the main circuit breaker is switched off during operation

    the sterilizer solenoid exhaust valve will remain closed once power has been restored. The

    solenoid exhaust valve will not open until the chamber temperature reaches below boiling

    point.

    Do not attempt to open the sterilizer lid until the pressure is at 0psi.

    Use the pressure relief switch to reduce the pressure inside the chamber to “0psi”. (refer to

    page 35 for instruction how to use the pressure relief switch)

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    9/47

      7

    Safety InformationHazardous Material

    The fol lowing material is no t recomm ended for us e on the steri l izer

    Nitroglycol, Nitroglycerin, Nitrocellulose and other explosive nitric

    esters.Trinitrobenzene, Trinitrotoluene, Picric acid, and other explosive

    nitro compounds.

    Peracetic acid, Methyl ethyl ketone peroxide, Benzoyl peroxide and

    other organic peroxides.

    ExplosiveExplosive

    Substance

    Sodium azide and any other metallic azide compound.

    Combustible

    Substance

    Metallic lithium, Metallic potassium, Metallic sodium, Yellow

    phosphorus, Phosphorus sulfide, Red phosphorus, Celluloid,

    Calcium carbide(Carbide), Phosphide lime, Megnesium powder,

     Aluminum powder, other combustible metal powders and Sodium

    dithionite (Hydrosulfite).

    Potassium chlorate, Sodium chlorate, Ammonium chlorate and other

    perchlorate.

    Potassium perchlorate, Sodium perchlorate, Ammonia perchlorate,

    and other perchlorates.

    Potassium peroxide, Sodium peroxide, Barium peroxide and other

    inorganic peroxides.

    Potassium nitrate, Sodium nitrate, Ammonia nitrate and other

    nitrates.

    Sodium chlorite and other chlorites.

    Oxidant

    Calcium hypochlorite and other hypochlorites.

    Ethyl ether, Gasoline, Acetaldehyde, Propylene chloride, Carbon

    disulfide and other substance with the flash point below minus 30ºC.

    Normal hexane, Ethylene oxide, Acetone, Benzene, Methyl ethyl

    ketone and any other substance with the flash points not lower than

    minus 30ºC and below 0oC.

    Methanol, Ethanol, Xylene, Pentyl acetate (Amyl acetate) and other

    substance with the flash point not lower than 0ºC and below 30ºC.

    Ignitable

    Substance

    Kerosene oil, Light oil, Turpentine oil and Isopentyl alcohol(Isoamyl

    alcohol), Acetic acid and other substances with the flash point not

    lower than 30ºC and below 65ºC

    Flammable

    Combustible

    Gas

    Hydrogen, Acetylene, Ethylene, Methane, Ethane, Propane, Butane

    and other flammable gas at 15ºC and under 1 atmosphere.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    10/47

      8

    Identification of PartsExternal

    Handle 

    Lid 

    Exhaust (U-shaped Pipe) 

    Control Panel

     Arm

    Pressure Gauge

    Door Latch

    Bottle Water Window Viewer

    Panel Guard 

    Pressure Relief Switch 

    Circuit Breaker  

    Terminal (Optional) Power Cord 

    Louver  

    Louver  

    Louver  

    Steam Saucer  

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    11/47

      9

    Identification of PartsInternal

    Sensor 1 

    Sensor 2 

    Heater 1

    Safety Valve 

    Thermal Insulator  

    Solenoid Drain Valve 

    Solenoid Exhaust Valve

    Bottom Plate Heater 2 

    Chamber  

    Service Plug 

    Micro Switch 

    Filter  

    (For Drying)

    (For Water Temperature)

    (For Sterilization)

    (For Chamber Temperature)

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    12/47

      10

    Identification of PartsInternal

    Plug

    Silencer 

    Drain Valve 

    Micro Switch

    Exhaust Hose

    Heater 1 (For Sterilization)

    Filter

    Solenoid Exhaust Port

    Sensor 1 (For Chamber Temperature)

    Sensor 2(For Water Temperature)

    Flange

    Clamp

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    13/47

      11

    Identification of PartsContro l Panel

    ①  ENTER key Start and stop operation, fix the set point/function changed.

    ②  △▽(up down) key Change set point/function.

    ③  MENU key Choose “STERILIZE” or “STERILIZE & DRY” before operation.

    ④  MODE key Use at the beginning/end to change or confirm set point/function.

    ⑤  Main display Shows mainly chamber temperature and set temperature.

    ⑥  Sub display Shows mainly time left and set time.

    ⑦  Heater lamp Stays lit or blinks when heater is on.

    ⑧  STERILIZE lamp Stays lit or blinks in sterilization process.

    ⑨  DRY lamp Stays lit or blinks in drying process.

    ⑩  END lamp Stays lit in the end of operation.

    ⑪  STERILIZE lampBlinks when “STERILIZE” menu is selected before operation, and

    stays lit during operation.

    ⑫  STERILIZE & DRY lampBlinks when “STERILIZE & DRY” menu is selected before operation,

    and stays lit during operation.

    ⑬  TROUBLE lamp Stays lit when the unit is out of order.

    ⑭  SET lamp Stays lit when pushing MODE key to check set point etc.,

     

     

     

    ⑬ 

    ⑤ 

    ⑦ 

    ⑥ 

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    14/47

      12

    InstallationEnvironmental Requirements

    Area of instal lat ion

    Never install sterilizer near any flammable,

    explosive, or corrosive gas components.Fire, explosion, or other troubles may occur

    when turning the breaker on and off.

    Area of instal lat ion

    The sterilizer should not be installed in the

    following environments.

    ①  Under direct sun

    ② In temperatures below 5℃, or above 35℃ 

    ③  Where severe changes in temperature occur

    ④  In very humid or dusty locations

    ⑤ Under water

    ⑥  In vibrating or shock areas

    Level surface

    If the sterilizer is not on a level surface the

    heater will reach a high temperature causing

    a interruption of the operation.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    15/47

      13

    InstallationEnvironmental Requirements

    Provide ample space surroun ding th e steri lizer

    Provide ample space surrounding the unit as

    indicated below.・Space for lid to open...13 inches or more

    ・Space to open door.....10.2 inches or more

    ・Space for ventilation...5.9 inches or more

    Do not ins tal l under a low cei l ing

     An abundant amount of steam will spout out of the lid

     just after operation. Do not install the sterilizer under

    any electrical components such as an alarm.

    Lock casters

    Be sure to fasten the two front casters

    after positioning the unit. ReleaseLock

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    16/47

      14

    InstallationPower Requirements

    Ground

    To prevent shock from electric current leak, be sure to connect the ground lead.

    Contact Yamato Scientific or an electrician if a power source with a ground terminal is needed.Never connect ground to with gas pipe, water pipe, telephone line, or lighting conductor.

    SM200/300 SM300/310/500/510

    Ground outlet is recommended for use.No power plug is attached to these types of units.

    Ground properly according to power source.

    Connect to suff ic ient power source

    Type Power SourceNecessary

    Capacity

    SM200 AC115V

    single-phase13A or more

    SM300

     AC115V

    single-phase 15A or more

    SM500 AC115V

    single-phase15A or more

    SM310 AC220V

    single-phase9.5A or more

    Be sure to connect the power cord to

    sufficient power source.

    (See right table)

    SM510 AC220V

    single-phase9.5A or more

    Attaching the power cord

    Be sure that the breaker is turned off prior to connecting the power cord. The power cord plug is

    not attached to the SM310 and 510. Select an adequate capacity plug and terminal according

    to the connecting power source.

    Green (Ground Lead) 

    Black Power Side

    White Power SidePower plug

    Ground outlet 

    Ground lead

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    17/47

      15

    InstallationInit ial Set Up

    Filter Installat ion

    Be sure to install the filter prior to operation.

    Instal l the bo ttom plate

    The bottom plate balances the material in the

    chamber and protects heater 1 and sensor 2. Be

    sure to install the plate.

    Close the drain valveBe sure to close the drain valve.

    Bottom Plate

    (at a level)

    OPEN

    CLOSE

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    18/47

      16

    InstallationInit ial Set Up

    Install the bott le

    The sterilizer drains water forcibly during the “STERILIZE & DRY” process. Hot and high

    pressure water will spout out to the bottle. To prevent the user from operating the sterilizer

    without the drain bottle a micro switch will not allow the sterilizer to operate without the bottle.

    ①Open the front door of the sterilizer and

    remove the bottle. The display will flash

    indicating the bottle is not in place.

     

    ②Pour 1500ml of water into the bottle. This

    water is used to cool the hot steam coming

    out of the chamber.

    The unit drains water forcibly out of the

    container into the bottle during the

    “STERILIZE & DRY” process. If too much

    water is poured into the bottle, the drained

    hot water can overflow. Follow the water

    volume specified. 

    ③Put the silencer into the bottle that contains

    1500ml water.

    ④ After the silencer is placed into the bottle,

    push the bottle upward, inside the door

    while

    putting the exhaust hose into the bottle.

    ⑤Once the bottle is set, the micro switch is

    pushed, which allows the sterilizer to

    operate.

    Silencer  

    Exhaust Hose

    1500ml 

    Plug 

    Silencer  

    Exhaust Hose 

    Micro Switch 

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    19/47

      17

    InstallationInit ial Set Up

    Pour water into the chamber  

    Pour water into the chamber according to the

    gauge on the bottom plate.Failing to pour adequate water will result in a

    interruption of operation. Confirm the water

    level after each sample run.

    Requirement

    SM200 1900~2000ml

    SM300 2800~3000ml

    SM500 2800~3000ml

    SM310 2800~3000ml

    See table on right for water volume

    information. Using over the maximum level

    of water will result in hot water spouting

    from the bottle.

    SM510 2800~3000ml

    Use dist i l led w ater fo r s teri l izat ion

    Use distilled water to prevent collection of mineral deposits and corrosion of chamber

    components. Under ground water is not recommended.

    Placing materials into steri l izer

    Place the material into the basket and insert

    into the chamber. Never block the steam

    exhaust port. Always use a basket for

    material being sterilized.

    High-level

    Exhaust Port

    Low-level 

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    20/47

      18

     

    InstallationInit ial Set Up

    Close the door b efore operat ion

    Be sure to close the front door of the sterilizer before operation. An open door may cause the

    bottle to fall during operation. Do not open the front door during operation.

    Close the l id

    Be sure to close the lid completely by turning

    the handle clockwise in a 1/4-1/2 circle after

    it touches the chamber. If it is not closed

    properly, steam will escape from the

    chamber.

    Set the steam s aucer

    This unit comes equipped with a steam

    saucer to prevent any steam from dropping

    to the floor. See right illustration to place

    the saucer under the exhaust.

    CLOSE 

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    21/47

      19

    Operational Precautions

    Opening l id

    When opening the lid, be sure that the

    pressure gauge reads “0psi". Proceed to openslowly. If the lid is opened at high pressure, hot

    steam will spout out.

    Drainage

     After operation, the water in the chamber is

    very hot. Please let the water cool before

    draining. Place a drain pan less than 2 inches

    in height under the drain valve or connect a

    drain hose.

    Never drain water during operation. If you

    open the drain valve at high pressure, hot

    water will spout out. 

    Water bott le window viewer

    Check the water level window. If water level is

    close to the upper limit indicated with the caution

    label, be sure to drain some water before using.

    OPEN 

    OPEN

    Water Bottle

    Window Viewer 

    The upper limit of

    water level

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    22/47

      20

    Operational Precautions

    Certain areas become very hot

    During and following operation, the areas around

    the lid become very hot. Be careful not to touchthese areas.

    ◆Gray areas in the illustration on the right

    become especially hot.

    Lid gasket

     Any damage or dust on the lid gasket/flange of the

    chamber (gray area in illustration on right) will allow

    steam to escape. Be sure to keep this area clean and

    avoid contacting with baskets, etc. when installing or

    removing materials from the chamber.

    The gasket may deteriorate according to use. If steam escapes frequently, you may need to

    change the gasket.

    Panel guard

    The panel guard was designed to protect the

    display/keypad. Do not hold the panel guard to

    move the sterilizer.

    “ENTER” key

    The “ENTER” key will interrupt the program operation.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    23/47

      21

      Sample Run Examples

    How to op erate the “STERILIZE” pro cess

    Sterilize operation using a waste bag.① Be sure to open the mouth of the bag to let steam.

    ② Keep the bag approximately 2/3” height of the total depth of the chamber (Example: SM200 is

    17.5” deep, so the bag should be placed 11.5” from the bottom). If the bag is placed higher than

    this point, steam will not disperse easily and may result in incomplete sterilization, in addition,

    the bag can easily cover the exhaust in the upper part of the container.

    ③ Set the temperature lower than the heat resisting point of the bag.

    ④ Set time depends on quality and quantity of the material being sterilized. (See chart below)

    Example:Sample Temperature Time Notes

    Gauze 121℃  30min. 5 dry gauze’s

    Petri Dish 121℃  40min. 30 petri dishes with lid

    ◆ The above data is only for reference use. The outcome of sterilization varies according to quality

    and quantity of substance and types of containers. The result of sterilization shall be checked

    using a chemical indicator.

    NOTE: Never operate this unit without a basket.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    24/47

      22

    Program “MODE”“MODE” key Info rmation

    How to use the “MODE” key

    Turn the circuit breaker on.

    The sterilizer automatically starts in

    stand-by mode, and the display

    show current set point.

    Push the “MODE” key. →Main display starts blinking

    (Sterilization Temp.)

    Sett ing other parameters

    ・ Turn the circuit breaker on.

    ・ Push the “MODE” key.

    ・Use the ▽△keys to select desired parameter.

     

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    25/47

      23

    Program “MODE”Display Symbo ls

    List of Symbo ls in the display

    Symbol Word Meaning

    B BOTTLE Bottle : Bottle isn’t set.

    DRY TEMPERATURE Function of setting drying temperature.D

    DRY TIME Function of setting drying time.

    END End : To indicate the end of operation.E

    ERROR Error : To indicate the occurrence of trouble.

    OFF Off : Signal function is switched off.O

    ON On : Signal function is switched on.

    STERILIZE

    TEMPERATUREFunction of setting sterilization temperature.

    STERILIZE TIME Function of setting sterilization time.

    STOP Stop : To indicate operation stop.

    S

    SIGNAL Signal : Buzzer (at the end of operation)

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    26/47

      24

    Program “MODE”Flowchar t

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    27/47

      25

    Program “MODE”Sett ing Ster i l izat ion Temp erature and Time

    Sett ing steri l ization temp/t ime

    ◆Used to set the sterilization temp/time.

    ◆Simply push the ”MODE” key to cancel the set up at any point.◆The sterilizer automatically sets in stand-by mode if it is untouched for more than a minute during

    set up.

    Stand-

    by

    Turn the breaker on.

    The main display blinks sterilization temperature, and the sub

    display blinks sterilization time.

    Push the “MODE” key.

    The main display blinks , and the sub display flashes

    sterilization temperature.

    Push the “ENTER” key. At this time, you can change the

    sterilization temperature.

    Input the sterilization temperature desired by pushing the▽△

    keys.

    ▽key…To lower temperature.

    △key…To raise temperature.

    S.

    Temp. 

    Push the “ENTER” key.

    The changed sterilization temperature is fixed, and the panel

    shows the next function, sterilization time set-up.

    Push the “ENTER” key again.

    You can now change the sterilization time.

    Input sterilization time you desire by ▽△key.

    ▽key…To reduce time.

    △key…To increase time.

    S.

    Time 

    Push the “ENTER” key.

    The changed sterilization time is fixed, and the panel blinks the

    next function, dry temperature set-up.

    D.

    Temp. 

    Push the “MODE” if the dry mode will not be in use. If the dry

    mode will be in use, press the “ENTER” key and continue to the

    next page.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    28/47

      26

    Program “MODE”Sett ing Dry Temperature and Time

    Dry Mode Temperature and Time

    Input the dry temperature you desire by ▽△key.

    ▽key…To lower temperature to 150℃.△key…To raise temperature up to 180℃.

    D.

    Temp.Push the “ENTER” key.

    The changed dry temperature is fixed, and the panel blinks the

    next function, dry time set-up.

    Push the “ENTER” key again.

    You can now change the dry time.

    Input the dry time you desire by ▽△key.

    ▽key…To reduce time.

    △key…To increase time.

    D.

    Time

    Push the “ENTER” key. The changed dry time is fixed and the

    panel display automatically returns to stand-by. The displays

    blinks newly set sterilization temperature and time.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    29/47

      27

    Program “MODE”Special Program Features

    Buzzer

    Signal on - buzzer at the end of a cycle

    Signal off - no sound alert at the end of a cycle prior to or during operation.◆You may select the “Signal on/off”.

    Select ion of “signal on /of f”

    Push the “MODE” key.

    The main display blinks , and the sub display blinks

    sterilization temperature.

    Push either the▽△keys to search the panel display(illustrated on

    left) to the desired setting..

    Push the “ENTER” key.

    The panel display changes places and the setting can be changed.

    Push either ▽△keys to display your selection.

    Push the “ENTER” key.

    The change on signal is fixed, and the panel display returns to stand

    by.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    30/47

      28

     Run “MENU”Selection o f Operation

    Select ing operat ion

    The following procedure allows the user to select the desired operation.

    ・ Turn on main circuit breaker.・ Push “MENU” key to select sterilize. Push “MENU” key again to select sterilize &

    dry.

    ●Menu change can only be made during stand-by mode. Once operation begins

    you can not change menu setting.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    31/47

      29

    Run “MENU”“STERILIZE” Procedu re

    “STERILIZE” pro cedur e

    Check the following:

    ①  Is the drain valve closed? Pg.10②  Is the drain filter in place? Pg.15

    ③  Is the bottom plate set? Pg.15

    ④  Is the bottle set? Pg.16

    ⑤  Is water in chamber? Pg.17

    ⑥  Is the lid closed? Pg.18

    ◇Turn on the main circuit breaker(located on the left side of

    the unit).

    ◇Push the “MENU” key and select “STERILIZE”.

    (See P28) 

    ◇Set the sterilization temperature and time as you desire.

    (See P25)

    →The main display blinks the set temperature.

    →The sub display blinks the set time.

    →The sterilize menu lamp comes on and off.

    ◇Push the “ENTER” key

    Operation begins. The sterilization heater turns on.

    The solenoid exhaust valve remains open. Air purges out of

    the chamber. (Approximately 20-25 minutes)

    →The main display shows water temperature and heater

    lamp.

    →The sub display goes off.

    →Sterilize lamp comes on and off.

    →Sterilize menu lamp blinks.

     At the end of air purge, the solenoid exhaust valve closes,

    and pressurization starts. (Approximately 20-25 minutes)

    →The main display shows the chamber temperature.

    →Sterilize lamp blinks.

    →Sterilize menu lamp is lit.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    32/47

      30

    Run “MENU”“STERILIZE” Procedu re

    When the chamber temperature reaches the set temperature,

    sterilization process begins and the time counter starts. Atthis time, the sub display will show the sterilization time

    remaining.

    →The main display shows chamber temperature.

    →The sub display shows the remaining sterilization time.

    →Sterilize lamp is lit.

    →Sterilize menu lamp is lit.

    Natural cooling begins when the set sterilization time is

    complete. At this time, the heater will automatically shut off.

    →The main display shows chamber temperature.

    →The sub display blinks .

    →Sterilize lamp is lit.

    →Sterilize menu lamp is lit.

    To prevent bumping, the solenoid exhaust valve opens once

    the chamber temperature drops below boiling point.

    →The main display shows chamber temperature.

    →The sub display shows .

    →End lamp is lit.

    →Sterilize menu lamp is lit.

    ◇Push the “ENTER” key.

    The “ENTER” key resets the microprocessor. You may now

    run another cycle.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    33/47

      31

    Run “MENU”“STERILIZE & DRY” Procedure

    “STERILIZE & DRY” pro cedur e

    Check the following.

    ① Is the drain valve closed? Pg.10② Is the drain filter in place? Pg.15

    ③ Is the bottom plate set? Pg.15

    ④ Is the bottle set? Pg.16

    ⑤ Is water in chamber? Pg.17

    ⑥ Is the kid closed? Pg.18

    In “STERILIZE & DRY” operation, the water is automatically drained from the chamber to

    the bottle before the drying process. Be sure to check the water level prior to operation or

    an overflow can occur. ◇Turn on the main circuit breaker.

    ◇Push the “MENU” key to select “STERILIZE & DRY”.

    (See P28)

    ◇Set the temperature and time you desire.

    (See P25 and 26) 

    →The main display blinks the set temperature.

    →The sub display blinks the set time.

    →Sterilize & Dry menu lamp blinks.

    ◇Push the “ENTER” key.

    Operation begins. The sterilization heater turns on.

    The solenoid exhaust valve remains open to exhaust air from

    the chamber.

    →The main display shows water temperature.

    →The sub display goes off.

    →Sterilize lamp blinks.

    →Sterilize & Dry menu lamp is lit.

     After air purge, the solenoid exhaust valve is closed, and

    pressurization starts. (Approximately 20-25 minutes)

    →The main display shows chamber temperature.

    →Sterilize lamp blinks.

    →Sterilize & Dry menu lamp is lit.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    34/47

      32

    Run “MENU”“STERILIZE & DRY” Procedure

     After chamber temperature reaches the set temperature,

    sterilization process starts. The sub display shows thesterilization time.

    →The main display shows chamber temperature.

    →The sub display shows the remaining sterilization time.

    →Sterilize lamp blinks.

    →Sterilize & Dry menu lamp is lit.

    When the sterilization time is complete. The solenoid drain

    valve opens and drains all water and the dry heater turns on.

     

    →The main display shows chamber temperature.

    →The sub display blinks .

    →Dry lamp blinks.

    →Sterilize & Dry menu lamp is lit.

    When forced drainage ends, the solenoid drain valve is

    closed. The solenoid exhaust valve opens to exhaust air out

    of the chamber.

    →The main display shows chamber temperature.

    →The sub display blinks .

    →Dry lamp blinks.

    →Sterilize & Dry menu lamp is lit.

     After the air exhausted out of the chamber the dry heater

    raises the chamber temperature.

    →The main display shows chamber temperature.

    →The sub display goes off.

    →Dry lamp is lit.

    →Sterilize & Dry menu lamp is lit.

    Drying process starts when the chamber temperature

    reaches to temperature, which is less than 4 degrees C from

    the set temperature.

    →The main display shows chamber temperature.

    →The sub display shows the remaining time.

    →Dry lamp is lit.

    →Sterilize & Dry menu lamp is lit.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    35/47

      33

    Run “MENU”“STERILIZE & DRY” Procedure

    When the drying time is completed, the heater is turned off

    and natural cooling starts.

    →The main display shows chamber temperature.

    →The sub display blinks .

    →End lamp is lit.

    →Sterilize & Dry menu lamp is lit.

    ◇Push the “ENTER” key.

    The “ENTER” key resets the cycle. You may now run another

    cycle.

    Drain the water bottle after “STERILIZE & DRY” process. The remaining water can cause

    hot water to spout out during the next operation.

    Since the bottle contains hot water after operation, wait for the water to cool before

    draining it.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    36/47

      34

      Special Control FunctionsView Parameter during Operation

    To check set temperature and time during the operation, push the “MODE” key.

    Set points can only be viewed, not changed.

    ◇Push the “MODE” key.

    →The main display blinks (set Temp).

    →The sub display shows the set temperature.

    →Set lamp is lit.

    To check the other set points, push ▽△key to flash the

    indication you desire in the main display and view the set

    points in the sub display.

    To end the function, push the “MODE” key. The display

    returns to the original state.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    37/47

      35

    Special Control Functions

    In terrupt Program d ur ing Operat ion

    ◇Push the “ENTER” key.

    Operation is interrupted. Steam will exhaust when

    pressurized.

    →The main display shows current chamber temp.

    →The sub display blinks .

    ◇Push the “ENTER” key twice to resume

    operation.

    →The main display blinks the set temperature.

    →The sub display blinks the set time.NOTE:

     Always drain the excess water from the chamber after a cycle has been interrupted. Failure to drain

    the excess water from the chamber after a cycle interrupt can result in subsequent unsterile loads as a

    result of contact with water in the chamber.

    Pressure Relief Switc h

    How to use the pressure rel ief switch

    ◇Push the Pressure Relief Switch.

    By pushing this switch, open the solenoid exhaust

    valve opens. Steam inside the chamber is exhausted

    into the bottle, decreasing the chamber pressure and

    temperature.

    The pressure relief switch can be used only when the solenoid exhaust valve is closed.

    Be sure the material will not bump. Be careful not to touch the exhaust areas of the

    sterilizer since high pressure steam will exhaust. Bumping or steam spouting due to

    sudden pressure decrease could damage the container or cause bodily injury. 

    Checking chamber temp erature dur ing stand-by mode

    Push either ▽△keys.

    The main display shows temperature inside the chamber. If you release

    your finger from the key, the display returns stand-by.

    Pressure Relief switch

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    38/47

      36

    Troubleshooting GuideSafety

    Self-d iagnost ic Circuit ry

    The Yamato sterilizer was designed with self-diagnostic circuitry capable of monitoring operation of

    the sterilizer. If a malfunction occurs, the display panel blinks the error code and an alarm is sounded.The microprocessor disables power to the heater element. If this occurs, check the error code and

    immediately shut off the main breaker.

    The safety valve is activated if pressure inside the chamber is abnormal. No special alarm is given

    when the safety valve is activated. However, a significant amount of steam will spout out of the louvers

    or spaces on the exterior. In this situation, there can be some problems such as an air purge or

    problems with the temperature controller. Push the “ENTER” key immediately to stop the operation.

    Turn off the main breaker after the steam spouting out settles down.

    Error Code Lis t

    Indication Cause Alarm in emergency Quick measure Final measure

    Sensor-related

    troubleDisplay and buzzer

    Disconnected heater

    circuitExchange sensor etc.,

     Triac-related

    troubleDisplay and buzzer

    Disconnected heater

    circuit

    Exchange triac

    etc.,

    Heater-related

    troubleDisplay and buzzer

    Disconnected heater

    circuit

    Exchange heater

    etc.,

    Main relay-related

    troubleDisplay and buzzer

    Disconnected heater

    circuit

    Exchange relay

    etc.,

    Run-down of

    back-up batteryDisplay and buzzer

    The unit stays stand-by

    condition.

    Turn on the breaker

    and charge the battery

      Heating low water Display and buzzerDisconnected heater

    circuitSupply water

    In case of power failure

    If the main circuit breaker is turned off during operation or main power is temporally interrupted

    during operation.

    1. The sterilzier will remain in the state it was disabled at. A power failure will result in the

    chamber remaining pressurized if it was under pressure at the time of the failure.

    2. When power is restored the operation program is lost and the solenoid exhaust valve will

    remain closed until the chamber temperature is below boiling point.

    Do not open the lid in this situation or hot and high pressure steam will spout out.

    If you want to open the lid, use the pressure relief switch to reduce pressure inside the chamber

    to “0psi”.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    39/47

      37

    Troubleshooting GuideProblem Solv ing Chart

    Fault Indication Check Points

    Display does not come up when

    the breaker is turned on

    ・Is the power supply cord connected?

    ・Is there any power failure?

    ・Is the voltage of the power source adequate?

     Air is not exhausted.

    The safety valve is activated.

    ・Is the hose to the bottle bent or clogged?

    ・Is the exhaust clogged by material etc.,?

    Water is not drained.

    Water is not drained in drying

    process.

    ・Is the filter clogged?

    Sterilization temperature does not

    go up.The pressure does not increase.

    ・Is the lid securely closed?

    ・Is the packing or flange damaged?

    The pressure increases when the

    solenoid valve is not closed.・Is the exhaust vent clogged?

    The temperature changes during

    operation.・Is there considerable changes of the outside temperature?

    Steam spouts rapidly.・Is there water in the bottle?

    ・Is the exhaust hose out of place or damaged?

    Water leaks.・Is the drain valve securely closed?

    ・Is there too much water in the bottle?The operation does not start from

    stand-by.・Is bottle installed?

    Noise during air purge is loud. ・Is the silencer out of place?

    ◆If you require further technical assistance, please call Yamato Scientific at (800) 292-6286 Ext. 235

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    40/47

      38

    Wiring Diagram

    SM200

    SM300/500

    ◆Inside of dashed line is option

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    41/47

      39

    Wiring Diagram

    SM310/510

    X1

     

    ◆Inside of dashed line is option 

    Symbol Name of Parts Symbol Name of Parts

    Tr Transformer PIO2 Pio2 Board

    X Relay POWER2 Power2 Board

    CT Current Transformer SW1 Pressure Relief Switch

    SSR Solid State Relay SW2 Micro Switch

    MCB Circuit Breaker TC Sensor 2 (Thermocouple)

    T Terminal PT Sensor 1 (Pt100 resistance thermometer)

    MV1 Solenoid Exhaust Valve H1 Heater 1 (Sterilization)

    MV2 Solenoid Drain Valve H2~4 Heater 2 (Drying)

    PLANAR Planar Board

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    42/47

      40

    Piping Chart

    Chamber

    Heater 2

    Sensor 1 

    Sensor 2

    Heater 1

    Drain Valve

    Silencer

    Bottle 

    SafetyValve 

    Heater 2

    Solenoid Drain Valve (MV2)

    Pressure Gauge

    Solenoid Exhaust Valve (MV1)

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    43/47

      41

    Maintenance

    For m aintenance

    For maintenance, turn the breaker and power supply off for safety. If the chamber is hot, wait for

    it to cool. Wipe off any dirt on the exterior with a damp cloth. Do not use benzene, thinner orcleanser to wipe, nor scrub with a brush.

    Daily Maintenance

    Cleaning after l iqu id loads

    Biological media tends to boil at a higher rate compared to other liquids. This can cause media

    to be splattered inside the chamber. Therefore, the chamber must be cleaned daily with a damp

    cloth after temperature has cooled.

    Weekly Maintenance

    Cleaning the inside of the chamber

    Clean the inside of the chamber with a clean damp cloth. Do not pull the filter out, located in the

    bottom center when cleaning the inside of the chamber. If cleaned without the filter inside, the

    pipe can get clogged with debris.

    Filter cleaning

    If the filter in the bottom center of the chamber is

    clogged, drainage is prevented. Clean the filter in

    appropriate intervals, according to frequency of

    use.

    Pull the filter out, from the bottom drain port. The

    filter may be cleaned under the water faucet.

    Cleaning the Door Gasket

    The door gasket, located under the lid of the sterilizer can be wiped clean with a clean damp

    cloth. The gasket should be examined for cracks or damage that can result in a poor pressure

    seal. If replacement is needed, contact Yamato Scientific.

    Cleaning th e Pipe Heater

    The pipe heater should be cleaned twice a month.

     Accumulation of scale build up results in poor heat transfer and high sheath temperatures.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    44/47

      42

    MaintenanceMonth ly Maintenance

    Cleaning the silencer

    The unit is set up with a silencer at the end of the

    exhaust hose to reduce noise occurring from air purge.Remove the silencer and clean with water once a month.

    ◆A ball is set in the silencer to prevent back flow of

    water. Do not misplace the ball when removing the

    silencer. 

    Silencer  

    Ball 

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    45/47

      43

    After Sale Service and WarrantyRequest for Repair

    When you request repair

    If any troubles should occur, stop the operation immediately, turn the breaker off, pull the power cord

    out and contact Yamato Sciectific’s Technical Service Department.

    Necessary information

    ●Model Number

    ●Serial Number

    ●Date of Purchase

    ●Distributor Name

    ●Information on difficulties

    To reach Yamato’s Technical Service Department, please call (800) 292-6286 Ext. 235

    Warranty

    ●Keep your warranty card for future references. Check the name of the distributor, date of purchase

    and any other contents of the warranty.

    ●The terms of the warranty is two year limited commencing the date of purchase. Repair is made

    without charge according to the contents of warranty.

    ●Decontamination Statement:

    We can not accept any products or parts returned to us for repair or credit that is contaminated with

    or has been exposed to potentially infectious agents or radioactive materials.

    If you need repair, please call Yamato Scientific at (800) 292-6286 Ext. 221 for a Return

     Authorization Number. No product will be accepted without this number.

    See the warranty or nameplate on the unit 

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    46/47

      44

      Replacement Parts List

    Common for al l SM seriesPart Name Part No. Specifications

    Planar Board 1-24-000-0059 ⅣCR 

    PIO Board 1-24-000-0028 ⅣFR 

    Power Board 1-24-000-0026 ⅣCR 

    SSR1 (SM300) 2-16-000-0020 YLT-SSR-01A

    SSR2 (SM300) 2-16-000-0021 YLT-SSR-01B

    SSR (SM200/310/510) 2-16-000-0010 YLT-SSR-01

    CPU 1-18-001-0006

    Transformer(Tr1; SM200/300) 2-18-000-0035 AD21-100A 115V

    Transformer(Tr2; SM200/300) 2-18-000-0022 AC100V

    Transformer(Tr1; SM310/510) 2-18-000-0035 AD21-100A 220V

    Transformer(Tr2; SM310/510) 2-18-000-0023 AC200V

    Relay (SM200/310/510) 2-05-000-0031 LY2F-DC12V

    Relay (SM300) 2-05-000-0019 HEla-DC12VRelay 2-05-012-0001 JRlaF-TM-DC12V

    Stick Keypad SM51A-30490 W467

    Current Transformer 2-17-001-0002 URP CTL-6-5-400

    Pressure Relief Switch 2-01-001-0014 A2A-4W

    Circuit Breaker (SM200/310/510) 2-06-001-0003 BS2021

    Circuit Breaker (SM300) 2-06-001-0004 BS2022

    Heater 1 (SM200) 2-24-000-0057 AC115V 1.3kw

    Heater 1 (SM300) 2-24-000-0058 AC115V 1.7kw

    Heater 1 (SM310/510) 2-24-000-0059 AC220V 2.0kw

    Heater 2 (SM200) 2-26-000-0003 AC115V 500W

    Heater 2 (SM300/310/510) 2-26-000-0004 AC115V 750W

    Sensor 1 1-16-003-0035 Pt100 Resistance Thermometer SensorSensor 2 1-16-003-0034 T type Thermometer

    Solenoid Exhaust Valve (SM200/300) 3-02-001-0016 AB41-02-7-C4A PT1/4 AC100V CKD

    Solenoid Exhaust Valve (SM310/510) 3-02-001-0017 AB41-02-7-C4A PT1/4 AC200V CKD

    Solenoid Drain Valve (SM200/300) 3-02-001-0016 AB41-02-7-C4A PT1/4 AC100V CKD

    Solenoid Drain Valve (SM310/510) 3-02-001-0017 AB41-02-7-C4A PT1/4 AC200V CKD

    Micro Switch 2-02-001-0005 ABV163661

    Safety Valve (SM200) 3-18-001-6003 M3D-B1.6±0.2Kgf/cm2 

    Safety Valve (SM300/310/510) 3-18-001-6002 M3D-B2.0±0.2Kgf/cm2 

    Plug SM500-30280

    Silencer SM500-30340

    Lid Gasket (SM200) 241022-180 SH75UN

    Lid Gasket (SM300/310/510) 241024-180 SH75UN

    Bottom Plate (SM200) SM200-30750

    Bottom Plate (SM 300/310/510) SM500-30750

    Filter SM500-30700

    Drain Valve 3-15-0003-6002 Type BSB PT3/8

    Mesh (WG)253003-172-2

    Bottle 7-26-000-0006 5000cc

    Pressure Gauge 5-05-000-0002 GS58-201

    ●The solenoid valve is usable for exhaust band drain.

  • 8/20/2019 manual Autoclave SM200,300,310,500,510-Ver.7

    47/47

    Responsibility

    Please follow the instructions in this document when using this unit. Yamato Scientific has no

    responsibility for the accidents or breakdown of device if it is used with a failure to comply.

    Never conduct what this document forbids. Unexpected accidents or breakdown may result in.

    Note

    ◆  The contents of this document may be changed in future without notice.

    ◆  Any books with missing pages or disorderly binding may be replaced.

    Instruction Manual for

    High Pressure Steam Sterilizer

    Model SM200/300/500/310/510

    Third Edition March. 26, 2009

     Yamato Scientific America, Inc.

    925 Walsh Avenue, Santa Clara,

    CA 95050 U S A