    MAGNETIC EYE TRACKING

    USING PLANAR TRANSMITTER

    Charusree.V.KNo:29

    S7 EC A

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    2/20

    INTRODUCTION

    Allows diagnosis of neurologic,ophthalmologic

    & vestibular disorders.

    Conventional ssc s/m employs cubical txr. Thin ,flat txr.

    Convenient in installation,operation &

    maintanance. Planar txr increases mobility of ssc s/m,

    simplifies accomodation in limited clinical

    space,enables bedside testing,no discomfortetc.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    3/20

    EYE TRACKING

    Eye movement,most frequent of all human

    movements.

    Eye tracking is the process of measuring pointof gaze or moton of eye relative to head.

    Eye tracker measures eye position & eye

    movements. Used in research on visual

    s/m,psychology,cognitive linguistic & in

    product design.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    4/20

    TECHNIQUES

    Electro oculography

    Infrared oculography

    Image based results

    Magnetic eye tracking

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    5/20

    ELECTRO OCULOGRAPHY

    Measures resting potential of retina.

    Potential difference of pairs of electrodes are

    measured.

    Eye acts as a dipole(working principle ofelectrooculography)

    Electrooculogram and tracing displayed by a pen record.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    6/20

    INFRA RED OCULOGRAPHY

    Light source is directed to eye.

    Measured the amount of light reflected back.

    Infra red not influenced by other light.

    Spatial resolution 0.1o

    Temporal resoluton 1ms.

    Drawback is 'blink' of eyes.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    7/20

    IMAGE BASED RESULTS

    Automatically extracting eye position fromimages of the eye.

    Measures vertical & horizontal eye movement.

    Temporal resolution lower than IR method.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    8/20

    MAGNETIC EYE RACKER

    Sensor & txr station.

    More accuracy than optical camera.

    It gives 3D orientaton.

    Main drawback is that,it is wired.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    9/20

    MAGNETIC EYE TRACKER

    Using Scleral Search Coils:

    Uses magnetic induction to track 'gaze' Eye position is recorded depending on voltage

    generated in coil by exposure to alternatingmagnetic field.

    Medical-grade silicon rubber suction ring

    Shape to adhere to the limbus.

    Disadvantages are invasive nature of coil,limited

    lifetime etc.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    10/20

    Scleral search coil

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    11/20

    Using Cubical Txr:

    Large cubic txr.

    Wide area of txng coil-high magnetic

    field,highS

    NR & tracking precision. High tracking speed.

    Bulky & awkward. Difficult to install,operate & maintain.

    Visual distraction.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    12/20

    Using Planar Txr:

    Coplanar array of txng coils,keeps txr thin. Minimum possible no.of txng coils.

    8 coplanar coils,provide best tracking

    precision for each txng coil.

    Each coil is excited at its own frequency.

    It excites coils continuously &

    simultaneously.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    13/20

    Construction & Working

    Txng coil:

    Coil is conneced via Ci,to half bridge excitation ckt.

    Half bridge excites txng coil.

    BS170 MOSFET translates 5v to 12v pulses.

    20 ohm resistor limits gate current. 0.1 ohm resistor to measure excitaton current.

    Ci chosen such that resonant freq. matchesexcitation freq.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    14/20

    Half bridge ckt for excitation of txng coil

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    15/20

    Rxng coil:

    Radius ofSSC is 9mm & no.of turns is 7(eyeposition)

    Custom made coil of 5mm radius 200 turns (headposition)

    Head coil mounted on mouth piece.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    16/20

    Testing applicability:

    Compared b/n Bilateral meniere affected &

    normal prson.

    Calculate Vestibular Ocular Reflex(VOR) VOR gain of healthy man would be greater

    than affected one.

    Ideally VOR would be 1.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    17/20

    a ) Ma g n e t ic t r a c k in g u s in g c u b ic t r a n s m i t t e r , b ) Us in g p l a n a r

    t r a n s m i t t e r

    a) VOR gain o f he althy o ne , b)of a no n he althy perso n

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    18/20

    APPLICATIONS

    Cognitive study

    Commercial aplictios (web,

    add.,design,marketing, automotive etc.) Communication s/m for disabled.

    Improved image & video communication.

  • 8/8/2019 Mag Eye Tracking Using Planar Txr by v.k.charUSREE S7 EC

    19/20

    CONCLUSION

    Thin flat planar txr,instead of bulky

    conventional cubic txr.

    Increased speed & precision. Diagnosis of vestibular disorders.

    Can improve tracking precision by

    increasing radius of txng coil. & increasedexcitation freq.