SC-FAST Seawater Analysis

M. Paul Field,

Institute of Marine and Coastal SciencesRutgers The State University of New Jersey

Page 2: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-2 with FAST Flow Injection

Reduce analysis timeReduce memory effectsReduce ICP-MS cone salting and injector pluggingImprove instrument efficiency

Page 3: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-Fast Sea Water analysis on Element-XR

10% sea water (SW) diluted with 3% HNO3. (Real samples)Y internal standardAnalysis Time

SC-FAST 10 sec load time; 2 min analysis time.SIS 1 min load time; 2 min analysis time; 1 min rinse

120 unknown samples.70 standards, blanks and CRMs.Matrix match Std curve4 point curve,1 Blank, 1 NASS-5 every 12 unknownsTotal 180 injections of SW and 10 injections of acid blank (only SW injections plotted).No cleaning of cones between day

Page 4: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Pressurized Gas Delivery System of Carrier Solution

Pulse free deliveryControlled low flow ratesNo peristaltic pump tubingEnsures low background

Page 5: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

System set-ups

µFlow (160µL/min)

PFA spray Chamber

1500W

Tuned on 10% SW

10 sec uptake, 2 min analysis 10% SW

2min 10 sec/sample 120 unknowns = 7.25 hrs

µFlow (160µL/min)

PFA spray Chamber

1500W

Tuned on 10% SW

1 min uptake, 2 min analysis,1 min wash

10% SW

4 mins/sample 120 unknowns =14hrs

Sea spray (600µL/min)

Glass spray Chamber

1500W

Tuned on 10% SW

1 min uptake, 2 min analysis,1 min wash

10% SW

4 mins/sample 120 unknowns =14hrs

SC-FAST + CH4PFA MicroFlowStandard Set-up

Page 6: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

0.00

0.25

0.50

0.75

1.00

1.25

0 75 150 225 300 375 450 525 600 675 750

Number of analysis (10% SW)

Rel

ativ

e se

nsiti

v

Day 1 Day 2 Day 4Day 3

Sensitivity Drift (Y)

NebulizerSeaspray(600 µL/min)

NebulizerPFA µFlow (160µL/min)

SC-Fast + CH4(160µL/min)

Page 7: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Cone DepositionCones after Sea Spray

120 analysisCones after µFlow

540 analysisCones after CH4 SC-Fast

740 analysis

• 30% reduction in salt loading

• More effective rinsing between samples

• <20% Drift

Page 8: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Analysis Time

7.5sec25secDead Time/sample

6hr 45minTime saved per run =

7hr 15min14hrTotal time

190190No. of samples

2min 10sec4minSample Measurement time

SC-FASTStandard Introduction System

Page 9: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Tuning

1ppb Tune, PFA-ST flow rate 160µL/minPFA Spray chamber with CH4 addition

In U UO %

Page 10: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

NASS-5 (n=38) 4 Days

0

2

4

6

8

10

12

14

16

18

20

P V Mn Mo Ba U AsElement

Con

cent

ratio

n (p

p

CertifiedDet (n=38)

P = 17.3 ± 0.9 (2σ )

Ba = 5.3 ± 1.6 (2σ )

Errors are < ± 6% 2σ

Page 11: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

SC-Fast:Detection Limits

Detection limits (160µL/min, PFA spraychamber )SC-FAST with CH4 addition

Blank DL(ppt) (ppt)

P MR 67.50 5.00V MR 0.18 0.18

Mn MR 1.00 0.50Mo MR 0.64 1.11Ba MR 0.40 0.49U MR 0.00 0.04

As HR 0.21 0.30

Element Resolution

Page 12: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Summary

Maintains instrument sensitivity, accuracy and stabilityMore effective rinsing of introduction systemReduces total matrix reaching conesExtends time between cone maintenanceDramatically reduces total analysis time

Page 13: M. Paul Field, Institute of Marine and Coastal Sciences ...cfs8.blog.daum.net/attach/.../09/06/13/...Analysis.pdf · SC-Fast Sea Water analysis on Element-XR 10% sea water (SW) diluted

Future Applications SC-FAST

Automated LC-ICPMS interfaceMultiple injections per single analysis

Minimize sample loading to the ICPMS cones of aggressive samples (e.g. H3PO4)Multiple collision/reaction cell conditions