LWGY Series Turbine Flowmeter
Instruction Manual
ISO 9001-2008 Certified Company
Accurate Flowmeters & Instrumentation Pvt. Ltd.(Formerly RR Flowmeters Pvt. Ltd.)
2 | Instruction Manual of Model LWGY Flowmeter
CONTENTS
1. Summary
1.1 Attention to transportation 3
1.2 Attention to storage 3
1.3 Attention to installation place 3
1.4 Attention to limited using of wireless transceiver 4
1.5 Attention to installation of explode-proof flowmeter 4
2. Technique Performance
2.1 Technique performance of turbine flowmeter 4
2.2 Technique performance of LRT-I displayer (Supplied by lithium battery) 6
2.3 Technique performance of LRT-II displayer (outside supply) 6
3. Structure and Working Theory
3.1 The basic structure of turbine flowmeter 7
3.2 Working theory 7
4. Outline Size and Installation
4.1 Outline size 8
4.2 Installation 10
5. Wiring
5.1 Wiring of amplifier and field meter 12
5.2 Practical example 13
6. Installation and Operation of LRT-I and LRT-II Displayer
6.1 The installation of LRT-I and LRT-II displayer 14
6.2 Structure and function of displayer 15
6.3 Content of display section 16
6.4 Parameter setting 17
6.4.1 Change common mode to setting mode 17
6.4.2 Parameter code and setting of data 18
6.5 List of error code 22
6.6 Replacement of lithium battery (LRT-I) 22
7. Maintain of Transducer and Revising of K-Factor 22
8. Errors and Troubleshooting 24
Instruction Manual of Model LWGY Flowmeter | 3
1. Summary
This manual contains the standard technique specification, model, installation, operation and maintenance of LWGY turbine flowmeter. Please read the manual before using the meter. But there are neither different features of different users nor technique specification of every time or structure because some amendments will not affect the function and operation of the meter.
Model LWGY turbine flowmeter can measure low viscous liquids, both Flowrate and Total flow. It can be widely used in the fields of Petrol, Chemical industry, Metallurgy, Scientific Research Institutes, etc., for measuring and / or control of liquids flow.
Several output and display methods can be selected (detailed in the table of model code). The executed standard number of the product is: ZB N12 005-1989
1.1 Attention to transportation
In order to prevent it from being damaged, do not open the package until reach the using place.
1.2 Attention to storage
Install the instrument in time. As for LRT-I, which is supplied by battery, before using it you should put the power contact pin on position “OFF” lest the power consumption should affect the using life of the battery.
If need storage, pay attention to following items:
a) If it is possible, deposit it without opening package.b) If having opened or used the instrument, please insert the LRT-I power jumper at the position
“OFF”, and use the original package.
The conditions of storage place:
a) Rainproof and damp-proofb) Avoid collision or shockc) Temperature is between -30 to +60°C. Ideal temperature is about 25°C.d) If deposit it outside, you may affect the performance. So install it as soon as possible.
1.3 Attention to installation place
The designing of the meter has considered the case under vile environment. But to keep its long accuracy and stability, when select installation place you must pay attention to following items:
Ambient temperature: Avoid installing it at the place where the ambient temperature is changeable strongly. There should be measures of heatproof and ventilation if there is heat emission.
Ambient atmosphere: Avoid installing it at place where is full of causticity air. If must, supply ventilation measure.
Mechanical vibration and shock: Although the instrument is very solid when install it you should avoid mechanical vibration and shock. And if you install it on much bigger pipe, support the pipe.
Others: There should be enough room surrounding the instrument to install and maintain it easily.
4 | Instruction Manual of Model LWGY Flowmeter
1.4 Attention to limited using of wireless transceiver
The electrical part of flowmeter can resist high frequency electrical noise interference. However, if you use wireless transceiver too close to the instrument, the high frequency noise interference will affect the instrument. Check the installation place to see if the instrument is affected by the transceiver (move the transceiver to the instrument from some miles away). If so, put the transceiver away from the place.
1.5 Attention to installation of explode-proof meter
The designing of the flow meter can be used in dangerous area class 1 and 2 which is set by “d” and “i’ standard.
2. Technique Performance
2.1 Technique performance of turbine flowmeter
The nominal diameter, flow range, liquid temperature, nominal pressure, ambient temperature, humidity and maximum pressure loss of turbine flowmeter is in table 2-1 while model and specification code is in table 2-2.
Table 2-1
Nominal DiameterDN(mm)
Range (m3/h)Liquid
Temperature(°C)
Nominal Pressure
PN (MPa)
Ambient Temp.
(°C)
Humidity(%)
Max. Pressure
Loss(MPa)
Error0.2%*
Error0.5%
Error1%
Min. Max. Min. Max. Min. Max.
2* 0.03 0.16
-20 to
+120
6.3
-25to
+55≤ 80
0.15
4* 0.04 0.25 0.12
6 0.1 0.6 0.1 0.66.3162532*
0.08
10 0.25 1.2 0.2 1.2 0.05
15 1.2 6 0.6 4 0.6 60.035
25 2 10 1.6 10 1 10
40 4 20 3 20 2 201.62.5
0.025
50 6 40 4 40 4 40
80 16 100 10 100 10 100
100 25 160 20 160 20 2001.62.5150 50 300 40 300 40 400
200 150 750 100 600 80 800
250* 200 1000 160 1000 120 12001.6
300* 260 1600 180 1800
Note: “*” means special order
Instruction Manual of Model LWGY Flowmeter | 5
Table 2-2: Model and Specification Code
Model Specification Code DescriptionLW..................................................................... Turbine Flowmeter
TransducerFor Liquid
G........................................................Y………………………….....
Nominal Diameter
2* ...................................4* ...................................6 .....................................10 ...................................15 ...................................25 ...................................40 ...................................50 ...................................80 ...................................100 .................................150 .................................200 .................................250 .................................300 .................................
2mm (PT G3/8”)4mm (PT G3/8”)6mm (PT G3/8”)10mm (PT G1/2”)15mm (PT G1”)25mm (PT G11/4”)40mm (Flange)50mm (Flange)80mm (Flange)100mm (Flange)150mm (Flange)200mm (Flange*)250mm (Flange*)300mm (Flange)
Type CodeA ............................B ............................C ............................
Accuracy 1%Accuracy 0.5%Accuracy 0.2%*
Output Signal
P .........................I ..........................T .........................M ........................
PulseAnalog 4 to 20mATotal display (Battery. for 2 to 3 years)Pulse out / 4 to 20mA out and LCD display (Intelligent)
Diameter Pressure
C1 .................C2 ..................C3 ..................C4 ..................C5 ..................C6 ..................C7 ...................
PN1.6MPaPN2.5MPaPN4.0MPaPN6.3MPaPN16MPa (Diameter≤25mm)PN25MPa (Diameter≤25mm)PN40MPa* (Diameter≤25mm)
Explode Request /NE/EX
Not ExplosiveEx ib I or Ex dIIBT4
Range of Temperature /NT/HT
Normal Temperature (<120°C)High Temperature (≥120 to 150°C)
Special Option / e.g.: High Temperature, Erode-resistant, Wearable etc.
Note: 1. The dimensions of connecting flange are in accordance with DIN 2. Mark “*” means special order. 3. Example: LWGY-50AIC2/EX/NT. The means Turbine flowmeter, DN50, Accuracy ± 1.0%, 4 to 20mA
analog output, nominal working pressure is less than 2.5MPa, Explosion-proof, normal temperature.
6 | Instruction Manual of Model LWGY Flowmeter
2.2 Technique performance of LRT-I displayer (Supplied by Lithium Battery)
Accuracy : Instant rate: 0.5%, Accumulated rate: 0.1%
Ambient temperature : -20 to +55°C
Medium temperature : -20 to +120°C (Decide medium temperature by explode temperature section)
Ambient humidity : Humidity 5 to 100% (Avoid dew)
Case material : Aluminum alloy
Power : 3.6 V Lithium battery (for 2 to 3 years)
Display : 6 LCD and project unit display Rate%, instant rate, six combined display mode
Output : No output
Memory function : Power-down, retain all the data
Error revising : 5 point (4 section) subsection revising
Shielded class : IP65, water-proof, dust-proof
Ex-proof appliance class : Ex ib I or Ex dIIBT4
Weight : 1.5 kg
2.3 Technique performance of LRT-II displayer (outside supply)
Accuracy : Instant rate: 0.5%, Accumulated rate: 0.1%
Ambient temperature : -20°C to 55°C
Medium temperature : -20°C to 120°C (Decide medium temperature by explode temperature section)
Ambient humidity : Humidity 5 to 100%(Avoid dew)
Case material of amplifier : Aluminum alloy
Power : +24 V DC
Displayer : 6 LCD and project unit display, Rate%, instant rate, six combined display mode
Output : Pulse 4 to 20 mA Analog
Memory function : Power-down, retain all the data
Error revising : Subsection revising
Shielded class : IP65, water-proof, dust-proof
Ex-proof appliance class : Ex ib I or Ex dIIBT4
Weight : 1.8 kg
Instruction Manual of Model LWGY Flowmeter | 7
3. Structure and Working Theory
3.1 The basic structure of turbine flowmeter
The brief structure of transducer is shown in Fig 3-1. It is mainly composed of housing units, rotor units, front and rear guide assembly, clamp ring and electro-magnetic amplifier.
Ampl i f i er
Body Pr e- suppor t Rot or Rear - suppor t Ri ng
Fig 3-1, The basic structure of turbine flowmeter
3.2 Working theory
As liquid running through the transducer, the rotor rotates with the help of the kinetic energy of the fluid and periodically to alter the magnetic resistance of the electro-magnetic converter. Then the magnetic flux density of the pick-up coil is altered by the passage of the magnetic blade. This change generates a pulse signal that is amplified by an amplifier and transmitted to an electric instrument to indicate or totalize the flow rate of liquid.
8 | Instruction Manual of Model LWGY Flowmeter
4. Outline Size and Installation
4.1 Outline size
a) DN2 to 25 turbine flowmeter (Nominal pressure is PN6.3MPa Fig 4-1, table 4-1)
Table 4-1Nominal DiameterDN(mm)
SizeWeight
(kg)GI
(mm)L (mm)
2*, 4*, 6 G 3/8” 11 50 0.510 G 1/2” 16 60 0.615 G 1” 18 75 1.025 G 11/4” 23 100 1.5
Note: Mark “*” means special order
Fig 4 - 1, DN2 to 25(Nominal pressure PN 6.3MPa) Outline
b) Nominal diameter DN2 to 25 high-handed turbine flowmeter (Nominal pressure PN16, 25, 32MPa as Fig 4 - 2, Table 4 - 2)
Fig 4 - 2, DN2 to 25
(Nominal pressure PN16, 25, 32MPa) outline
Table 4 - 2
Nominal diameter DN (mm)
Size (mm) Weight (kg)L0 L D0 D M
2, 4, 6 42 82 12 6 M 18 × 1.5 0.810 55 97 16 10 M 22 × 1.5 1.015 75 126 25 15 M 33 × 2 1.525 100 155 32 25 M 42 × 2 2.0
Instruction Manual of Model LWGY Flowmeter | 9
c) Nominal diameter DN40 to 300 Turbine Flowmeter (as Fig 4-3, Table 4-3)
Table 4 - 3
Nominal diameterDN (mm)
Size (mm) Weight (kg)D1 D2 d N L
40 145 11018
4140 7
50 160 125 150 880 195 160
8
200 10
100 215(230)
180(190)
18(22) 220 12
(13)
150 280(300)
240(250) 22
(26)
300 16(17)
200 340(360)
295(310)
12360 19
(20)250 405 355
26400 73
300 460 410 420 85Note:
1. The dimensions of connecting flange are in accordance with JB / T81-1994 or JB / T9119-2000.2. As to nominal diameter DN40-80, the dimensions of connecting flange of pressure capacity PN=1.6
and 2.5MPa are the same. But for DN100-200, the flange dimensions of 2.5MPa are shown in brackets.
3. Ordinarily, the products are equipped with 1.6MPa flanges.
Fig 4 - 3, DN 40 to 300, Outline
10 | Instruction Manual of Model LWGY Flowmeter
Fig 4-4, LRT-I displayer outline Fig 4-5 LRT-Ⅱ displayer outline
4.2 Installation
1. Installation area
The transducer should be used under the operating condition with the measuring medium temperature from -20 to +120℃ and ambient relative humidity no greater than 80%. Avoid installing the transducer in a corrosive atmosphere. Select an area subject to minimalize mechanical vibration or impact shock, minimalize thermal radiation, minimalize magnetic field. If the transducer is subject to the strong magnetic field, it is recommended that shield must be covered on the preamplifier.
2. Installation direction
The transducer should be installed horizontally. The flow direction of fluids should be in accordance with the arrow point, which is on the label of the transducer for indicating the flow direction.
3. Piping
(1) For inevitable reason from installation point of view, install a fluid vibration-damping device such as strainer on the upstream side of the transducer. The upstream and downstream straight pipe requirements are generally recommended that the straight pipe 15-20 times as long as the inner pipe diameter upstream of the turbine flowmeter and 5 times as long as the inner pipe diameter downstream of the turbine flowmeter.
The length of the straightener installed on the upstream and downstream sides of the turbine flowmeter following as Fig 4-6.
Fig 4-6 Straight pipe demand
Instruction Manual of Model LWGY Flowmeter | 11
If a reducer is installed on the upstream of the pipeline: L=15DIf a single elbow is installed on the upstream of the pipeline: L=20DIf two elbows are installed on the upstream of the pipeline: L=25D (on the same plane) L=30D (on the different plane)If a right angle elbow is installed on the upstream of the pipeline: L=40DIf a shut-off valve is installed on the upstream of the pipeline: L=20D(the valve open fully) L=50D(the valve open partially)
And more, a rectifier composed by lead pipe should be installed in the straightener, which is on the upstream side of the transducer in order to clean the vortex efficiently and raise the accuracy of the transducer. After installing the rectifier, the length of straightener is (10 to 20)D
(2) A strainer with 3-9 mesh/sq. in. Screen should be installed in the pipeline before the front straightener, which is on the upstream side of the transducer in order to clean the liquid impurity, ensure the proper function and durability of the transducer. Generally, if the diameter is big the bean size is rare and the diameter is small the bean size is dense. You should select the bean size of strainer according to actual case in order to ensure the normal function of the transducer.
(3) When joint the imported flange of the transducer, you must note that there is no prominent part in the pipe and the two flanges must inosculate fully. The gasket cannot expose in pipe. The eccentric adapter will cause the flow non-uniform distribution, so do not use.
(4) In order to ensure the need of overhauls under the working condition, cut-off valve should be placed around the pipeline of transmitting instrument and by-pass pipeline also should be established. The flow control valve must be established in sensor’s downstream. When use the sensor, the cut-off valve in the upstream must be opened fully to avoid the fluid turbulization phenomenon on the upstream part.
(5) If the flow range of the transducer is oversized (surpasses the upper limit), the bearing will wear quickly because of the excessive rotational speed. Therefore, when you estimate it will be oversized you can regulate the fluid rate with the control valve, which is in the downstream part.
Fig 4-7 Match air separator and strainer Fig 4-8 Rectifier
(6) Because the air in the pipeline will bring big error to the measurement of the transducer, when install it you should pay attention to whether there is air in the measuring liquid, especially for the measurement of high-gravity oil you should install air separator. And install the pipeline
12 | Instruction Manual of Model LWGY Flowmeter
of the separator tipsily to avoid air stockpile. Besides, pay attention to the control of the back pressure of the downstream. Count the back pressure as following formula:
Pa≥ 2 △ P + 1.25Pv
In this formula, Pa – Back pressure of downstream; △P – Pressure loss of transducer at maximum flow rate; Pv – Saturation vapor pressure of medium at maximum using temperature.
(7) When install the transducer on new pipeline, you should use a blank pipe instead of the transducer to move for a period of time and confirm the impurity has been excluded to avoid them moving into transducer.
(8) When install nominal pressure PN16, 25MPa transducer, you should coat a few lubricating oil on the cutting edge of the sleeve chuck, nut’s thread and each contact spot. Then cover the nut and sleeve chuck on the pipe according to the order, insert the pipe on the awl hole base of the transducer coating, put the sleeve chuck, move the pipe with tightening the nut until it can not move and at last tighten the nut for 1-1.5 circle again.
5. Wiring
5.1 Wiring of amplifier and scene display meter
A. Wiring of pulse output amplifier ( LWGY - ooooP ) The output signal is transferred by three-core cable. You’d better use the more delicate stranded
wire. When welding or wiring, pay attention to the neighboring end will not short-circuit. As figure 5-1 shows.
+5- 24V DC 3 2
4 1
PULSE
0V
a)Aviation Plug (thrum needs tin welding) b) Explode-proof Pulse Output
Fig 5-1 Wiring of pulse output amplifier
w Power: VDD +5 to +24VDC
wOutput frequency: (At flow rate down limit ) more than 20Hz.
Low pwl: 0 to 0.5V(Push-pull output)
High pwl: (VDD-2)V (Push-pull output)
w Ambient temperature: -250C to +550C
wHumidity is not more than 80%
Instruction Manual of Model LWGY Flowmeter | 13
B. Wiring of 4 to 20mA analog output converter ( LWGY - ooooI ) 4 to 20mA analog output connection as Fig 5-2. Actual load can be flow-integrating instrument, PLC, counter and computer etc.
Fig 5-2 4 to 20mA (two-wire system)
C. Wiring of LRT-II displayer ( LWGY - ooooM ) Two-wire system electricity output as Fig 5-3 and three-way pulse output as Fig 5-4.
Fig 5-3 Analog out wiring (two wires system)
Fig 5-4 Pulse out wiring (three wires system)
5.2 Practical example
A. Wiring of pulse output amplifier and flow controller. The connection of pulse output amplifier and flow controller as Fig 5-5.
Fig 5-5, The connection of pulse output amplifier and flow controller
14 | Instruction Manual of Model LWGY Flowmeter
B. Wiring of 4 to 20mA analog output converter and flow controllerThe connection of 4 to 20mA analog output converter and series flow controller as Fig 5-6.
Fig 5-6, The connection of 4 to 20mA analog output converter and series flow controller
6. Installation and Operation of LRT-I and LRT-II Displayer
6.1 The installation of LRT-I and LRT-II displayer
LRT-I and LRT-II displayer and turbine flowmeter connected by M14×1screw thread. You should tighten the screw thread as much as possible otherwise the frequency signal may be lost and cause measuring error.
As for LRT-I, if you want to change the direction of the monitor, loose the tightened nut of the fixed mini-connector and do not back out too much. Then move the meter head less than 360 degree until it reaches the ideal direction. At last, tighten the nut again. But for LRT-II, you can change the direction only by M14×1 screw not more than half circle. When finished, tighten the nut.
As for LRT-I meter which is supplied by battery, open the front cover pull out the power jumper from “OFF” position to position “ON". After that you can operate the meter.
Fig 6-1 Switch “on” to LRT-I
Instruction Manual of Model LWGY Flowmeter | 15
6.2 Structure and function of the displayer
This chapter contains display content and parameter setting steps of displayer ( LWGY - ooooM ). The panel is as Fig 6-2.
É Ïº £º ×¼ ª× Ô¶ »̄ Ò̄ DZ íÓ ÐÏ Þ¹ «Ë ¾
FLODSPJ oul e
%scf gal bt Nm kg/ h/ m3
SET SFI FE I NC
3
1
2
4
5
6
Fig 6-2 Panel of displayer
The contents displayed by intelligent meter liquid crystal are digital, percentage and project unit.
If you want to set a project unit, which is not in the display content, just stick the sign mark of this unit on the right part of the display window. You can set each parameter by using setting keys [SET], [SHIFT], [INC] etc.
Table 6-1 Types of display unit
DisplayUnit Direction
% Percentagel Print
t TonNm3 Standard cubic
meterm3 Cubic meterkg Kilogram/h Per hour/m Per minute
Table 6-2 Names of parts of the panelItem Direction
1 Display part : Display data, unit and number of parameter setting items2 Setting part : Setting parameter item and data of parameter (using SET, SHIFT, INC key
to set)3 Stick unit tag if set unit that outside the monitor4 Decimal point5 Determine a parameter setting number or a sign mark of parameter data6 Setting key
16 | Instruction Manual of Model LWGY Flowmeter
6.3 Content of display section
There are three modes of displayed content:
Table 6-3 Names of mode
No. Mode (state) name
Operation of keys Displayed content
1 Common mode
- Can display instant flow or integrated value
2 Setting mode SET Determine parameter content or revise data with setting part. Press SET key to enter to setting mode.
3 Alarm num-ber
Display mode
- When there is alarm in the common mode, this mode will replace. The number of alarm shows the content (about 2 seconds), then display the normal data(about 4 sec-onds).The two modes replace in turn.
Note: The system shown by mode is in state of related setting or displaying
In common mode (state), it can display instant flow or accumulated flow. There are 6 display modes in common mode as fig 6-4.
Table 6-4 Display mode symbol
Symbol Name Content0 % display mode Instant flow by 0.0% to 110.0% display1 Display by mode of measurement
unitDisplay instant flow shown by measurement unit with 0-32000
2 Accumulated flow display mode Display accumulated flow by 0-9999993 Display flow (%) and accumulated
flow in turnDisplay instant flow (%) and accumulated flow (project unit) in turn
4 Display instant flow and accumu-lated flow in turn
Display instant flow (project unit) and accumu-lated flow (project unit)
5 Display instant flow and % in turn Display flow % and instant flow shown by project unit
* The position of point determined by setting of measuring range.
Instruction Manual of Model LWGY Flowmeter | 17
Table 6-5, Display mode changes the number
No. Operation key Display Direction
0Common mode
e.g. : Integrated value
1 SET General mode
Press SET key to enter to setting mode
Parameter number is changeable
2
INC
INC
INC
Press INC key
Press INC key
Press INC key
3
SHIFT
INC
When press SHIFT key, move the cursor to right by one point.
Press INC key, change 1 to 2
4
SET
INC
Then, press SET move the cursor to right most
You can set parameter content
5
SET
SET
Press SET key, the display can twinkle
Press SET key again, move the cursor to left-key again, move the cursor to left-most
6 SHIFT Return to common mode
6.4 Parameter setting
This chapter states how to set the parameter of flow meter by using the setting key which on the intelligent head meter’s panel. Parameter list as table 6-6�
6.4.1 Change common mode to setting mode
1 Press SET key, change “common mode” to “setting mode”. Then enter to the parameter setting state.
18 | Instruction Manual of Model LWGY Flowmeter
2. Use the sign “:” to separate the “parameter number” on the left and “parameter data content” on the right. As fig 6-3 shows.
%scf gal bt Nm kg/ h/ m3
SET SFI FE I NC
di spl ay sect i on
set t i ng sect i on
par amet er sect i on� 2 di gi t s�
dat a sect i on� 4 di gi t s�
Fig 6-4 Parameter number area and data area
6.4.2 Parameter code and setting of parameter data
Press SET key, change “common mode” to “setting mode”.
Press INC key to change parameter number(digital or letter) and press SHIFT key to move cursor.
Head meter display area
(This example means B02. If the number of parameter is B□□□B
will not be displayed.)
Press SET key, move the cursor to “parameter data area”.
Press INC key to revise parameter value and position of decimal point. Then press SHIFT to move the cursor.
After revising, press SET key, all the parameter number and data will twinkle. Confirm the content and press SET key again, the setting is over.
When finished setting, press SHIFT key, it will return to “common mode”, and then display flow or alarm.
Example of parameter setting
The flow of parameter setting is as table 6-6.
Instruction Manual of Model LWGY Flowmeter | 19
Table 6-6, Setting K-factor
Mode change and operation Display Description
SET Press [SET] to enter setting
mode↓
The item is 06 (=B06), setting K-factorSetting data and decimal point, confirm the decimal point by pressing [INC] key
SET Press [SET] again, “B”
cancelled, display “06” only↓
Press [SHIFT] to move the cursor at “6” position
↓
Data 68.6 is initial value
SET Press [SET] key to enter data
section(Change the data)
↓
Changing 68.6 to 1068.6
INC Press [INC] Moving the
cursor↓
SHIFT Press [SHIFT]
INC Repeat until1068.6 setting is performed data display circulation
SET Press [SET],
all data flashing↓
SET Press [SET] again,
saving data↓
SHIFT Press [SHIFTT] key,
Turn back to normal mode
20 | Instruction Manual of Model LWGY Flowmeter
Table 6-7 Parameter list
Item Name R/W Data range ( ) Unit Decimal point Description Initial
value
B00 SET1 R Menu B
B01 TAG NO. W 8 alphanumeric characters
Tag No.
B02 OUTPUT W 4~20mA DCPULSE
(0)(1)
Selection out mode(Meanwhile change the jumper)
4 to 20 mA
B03 SIZE W 15 mm (1/2”) (0) Selection of meter tube 25 mm
25 mm (1”) (1)
40 mm (1.5”) (2)
50 mm (2”) (3)
80 mm (3”) (4)
100 mm (4”) (5)
150 mm (6”) (6)
200 mm (8”) (7)
250 mm (10”) (8)
300 mm (12”) (9)
B06 K-FACTOR(KM) W 0.0001 ~ 32000 P/l 0 ~ 5 K-factor (KM at 15°C) 68.6
B35 FLOW UNIT Wm3 (0)
Selection of Flow unit m3
l (1)
B50 TIME UNIT W /s/m/h/d
(0)(1)(2)(3)
Selection of Time unit flowrate /m
B51 SPAN FACTOR W E0E+1E+2E+3E+4E+5E-5E-4E-3E-2E-1
(0)(1)(2)(3)(4)(5)(6)(7)(8)(9)
(10)
Selection of Span factorE0=1E+1=10E+2=100E-2=0.01
E0 (=1)
B52 FLOW SPAN W 0.00001 ~ 32000 m3/m 0~5 Flow span 10
B53 DAMPING W 2481632640
(0)(1)(2)(3)(4)(5)(6)
sec Selection of Damping time 4 second (1)
C00 SET2 Menu C
Instruction Manual of Model LWGY Flowmeter | 21
Item Name R/W Data range ( ) Unit Decimal point Description Initial
value
C01 TOTAL RATE W E0E+1E+2E+3E+4E+5E-5E-4E-3E-2E-1
(0)(1)(2)(3)(4)(5)(6)(7)(8)(9)
(10)
E0=1E+1=10E+2=100
E0
C02 PULSE RATE W E0E+1E+2E+3E+4E+5E-5E-4E-3E-2E-1
(0)(1)(2)(3)(4)(5)(6)(7)(8)(9)
(10)
Scaled pulse factor,E0=1E+1=10E+2=100
E0
D00 ADJUST R Menu D
D20 FLOW ADJUST W NOT ACTIVEACTIVE
(0)(1)
Selection of correcting instru-mental error
0
D21D22D23D24D25D26D27D28D29D30
FREQ1K1FREQ2K2FREQ3K3FREQ4K4FREQ5K5
WWWWWW
Hz%Hz%Hz%Hz%Hz%
0 ~ 50 ~ 50 ~ 50 ~ 50 ~ 50 ~ 50 ~ 50 ~ 50 ~ 50 ~ 5
First break-point frequency (f1)First test point K-factor (K1)Second break-point freq. (f2)Second test point K-factor (K2)Third break-point freq. (f3)Third test point K-factor (K3)Fourth break-point freq. (f4)Fourth test point K-factor (K4)Fifth break-point freq. (f5)Fifth c test point K-factor (K5)
1.01.01.01.01.01.01.01.01.01.0
E00 CONTROL R Menu E (Control signal)
E01 TOTAL RESET W NOT EXCUTEEXECUTE
(0)(1)
Resetting Totalized value NOT EXCUTE
E02 DISP SELECT W RATE (%)RATETOTALRATE (%), TOTALRATE, TOTALRATE, RATE (%)
(0)(1)(2)(3)(4)(5)
Selection of Display RATE (%)
H00H07H08H09H30
MAINTENANCEL.C. FLOWRATETRIM 4mATRIM 20mAREVISION
RWWWR
0 ~ B52-1 ~ 10-10 ~ 10
B52%%
0 ~ 50 ~ 50 ~ 5
Menu H (Maintenance)Low cut flowrateTrim 4mATrim 20mARevision number of software
0.061220.00.0
H08, H09: TRIM 4mA, TRIM 20mA Setting value in H08 or H09 equal display value. E.g. When setting H08, The display value is “4.02”, then setting “4.02” in H08, press [set] key 2 times. repeat above procedure until display “4.00”.
22 | Instruction Manual of Model LWGY Flowmeter
6.5 List of error code
When there is error in instrument, the displayed error is as table 6-8:
Table 6-8 Error code list
No. Errorname Display Possible
reasonCurrent output
Pulse output Output%
Project unit
output
Accumulation output Remove way
1Output
over range
No display
Output signalMore than
110%
Fixed at110% Normal Fixed
110% Normal NormalChange
parameter or flow range
2 Data overflow HHHH
Instant flow or accumulation flow overflow
Normal Normal Normal Normal Not Normal Change B51 or C01
6.6 Replacement of lithium battery (LRT-I)
The steps of changing Li-ion battery are as following:
» Unscrew the back cover.
» Put the power jumper at “OFF” position.
» Loose the two bolts of the fixed battery with screwdriver (as Fig 6-1), draw the battery pin and unload the Battery.
» Rip away the new battery lead wire for about 10 millimeter with wire stripper; insert the battery in the connection terminal.
» (Pay attention to the polarity) and tighten the screw.
» Put the power jumper at “ON” position (at “OFF” when you don’t use it for saving power).
» Cover the back cover.
7. Maintenance of Transducer and Revising of K-Factor
a) You should install the transducer according to the flow range, nominal pressure and flow direction mark on the checkup certificate.
b) The transducer should work under following conditions:
Liquid temperature: -20 to +120°C
Ambient temperature: -20 to +55°C
Ambient humidity is not more than 80% (Except high temperature).
c) When the transducer leaves the factory, it is calibrated with normal temperature water. If the measuring liquid is not the same with normal temperature water you should revise the parameter or recalibrate it with actual measuring liquid. But for liquid whose viscosity is lower than 5×10-6m2/s (5MPa.s), you should not recalibrate. The example of instrument parameter revising is as following:
Instruction Manual of Model LWGY Flowmeter | 23
Example: There is a LWGY-25A, K-factor is 327.5834P/L (calibrated with water), use it to measure 30cP hydraulic pressure oil, find there have 100 liter (QS) oil in the container, while flow displayer displays 93.5 liter (Qa), at this time you should adjust the K-factor as follow:
LPQQQKKs
saN /2905.306
100
1005.9315834.3271 =÷
�
���
� �+�÷÷
�
����
�+� £½
£-£½
Reset the new K-factor into the flow displayer, repeat above procedure, observe the displayed value, if there still have errors, you may revise for several times until you are satisfied.
d) The usage period of transducer is usually from half a year to one year according to the working
condition. And you should wash it regularly. If find the axle or axletree is wearing seriously, change it and recalibrate it.
Note: Take care of the product calibrate certificate properly to avoid losing the constant of the instrument or other data.
8. Error and Troubleshooting
The troubles of transducer can be concluded as three points:a) There is no output signal in the transducer or monitor;b) There is still output signal when the flow is zero;c) Indicative flow is not in agreement with actual flow.
The reasons of these troubles are as table 8-1:Table 8-1
TroublePhenomena
TroubleReason
RemoveMethod
No output
Wiring is not correct Check wiring
Sludge adhering to the rotor Clean the pipeline or rotor
Turnoff or short Change the amplifier
Preamplifier failed Change the amplifier
No power or voltage error Check the power supply
Error in itself Check the instrument
Has output when no flow
Outside magnetic field Check the shield or exclude obstruction
Pipeline vibration Eliminated the vibration
Large flow errors
Outside magnetic field and pipelin vibration Check the shield and eliminated the vibration
Preamplifier failed Check the amplifier
Vapor in the pipeline Install air eliminator
Back pressure low Increase back pressure
Bearing weared out Change the bearing
Sludge adhering to the rotor Clean the pipeline or rotor
Pipeline error Select proper pipe
Error in itself Check the instrument
Accurate Flowmeters & Instrumentation Pvt. Ltd.(Formerly RR Flowmeters Pvt. Ltd.)
Head Office : G1, Sai Thirumala Heights, Plot No. 54, 55 & 91, Bhandari Layout, Nizampet, Hyderabad - 500090, INDIA. Tel : +91 - 40 - 2980 3333, Telefax : +91 - 40 - 2980 2323, Mobile : +91- 9490 262626
Factory : Plot No. 71 & 82, Sy. No. 139, Aminpur Village, Patancheru Mandal, District Medak - 502032
Web : www.accurateflowmeters.com E-mail : [email protected]
ISO 9001-2008 Certified Company