View
225
Download
0
Embed Size (px)
Citation preview
Lotte Verbunt
Investigation of leaf positioning accuracy
of two types of Siemens MLCsmaking use of an EPID
Overview
• background
• goals
• calibration
• image acquisition
• method I
• method II
• measurements
• comparison
• results
• conclusions
• recommendations
Dr. Bernard Verbeeten Institute Tilburg
• independent institute for radiotherapy and nuclear medicine:
- 11 radiotherapy oncologists
- 3 nuclear physicians
- 6 medical physicists
- 1 general doctor
• 2360 new radiation treatment patients a year (2004)
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Linear accelerator and EPIDbackground
goalscalibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
accelerationbending
through patient
conversion into light
reflection by mirror
detection by camera
Multileaf collimator
two types of Siemens MLCs:
MLC 1: 29 leaf pairs; 2 mm accuracy
MLC 2: 41 leaf pairs; 1 mm accuracy
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Y jaw
leaf bank
central axis
Project goals
• development of a leaf verification method applicable to a Siemens linac using a CCD-camera based EPID
• verification of the leaf positions of two different types of Siemens MLCs and to check whether the MLCs and the current leaf calibration method are accurate enough for IMRT
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Why is this of interest?
• Intensity Modulated Radiation Therapy (IMRT):
- several beams from different directions
- each beam consist of several (abutting) segments
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Leaf calibration method: light field
• performed with EPACtool
• using light field and grid lines from graph paper
• 4-points calibration: 20, 10, 0, and -10 cm
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations left leaf bank right leaf bank
-20 -10 0 10 20
Leaf calibration method: light field
• performed with EPACtool
• using light field and grid lines from graph paper
• 4-points calibration: 20, 10, 0, and -10 cm
• uncertainties because:
– a difference between light and radiation field
– a blurred light field edge
– a possible rotation of graph paper
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Leaf verification methods: radiation field
• method I: 50 % dose value
• method II: from dose change to leaf positions
both methods use same input from the EPID
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Image acquisition
• 6 segments, 5 cm wide, 2 mm overlap of the leaves
black = blocked by the leaveswhite = unblocked irradiated part
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Image acquisition
• 6 segments, 5 cm wide, 2 mm overlap of the leaves
• barrel distortion correction
32
21 dddu RaRaRR
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Image acquisition
• 6 segments, 5 cm wide, 2 mm overlap of the leaves
• barrel distortion correction
• radiation center and pixel dimensions
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Method I
• take a mean profile for each segment and for each leaf pair
• fit a polynomial function between 20 and 80% of the dose value
• determine leaf position (50% dose value)
position [pixel]
posi
tion
[pix
el]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
500
0 100 200 300 400 5000
10
20
30
40
50
60
70
80
90
100
110
position [pixel]
rela
tive
dose
[%]
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Accuracy method I
• images:
– nine 3 cm wide segments
– flat side of the leaves
– 90 rotation
• accuracy leaf position determination: 0.08 ± 0.18 mm (1 SD)
• shift of ± 0.45 mm at abutment leaf ends for MLC 2
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Method II
• sum all segments
• take a mean profile for each leaf pair
• calculate the dose change at each abutment
• convert the dose change into gap width
position [pixel]
po
sitio
n [p
ixe
l]
100 200 300 400 500
50
100
150
200
250
300
350
400
450
5000 100 200 300 400 500
0
20
40
60
80
100
position [pixel]
rela
tive
do
se [%
]
extreme dose
average dose
ji
jiji
D
EDd
,
,, 1
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Calibration function method II
• calibration function f :
• six tests with different gap widths (-1 to 4 mm)
• plot: intended gap width versus dose change
• dose change is 15-20 % per mm gap width
gap width dose changeif
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
-2,00
-1,00
0,00
1,00
2,00
3,00
4,00
5,00
6,00
7,00
-0,15 0 0,15 0,3 0,45 0,6
dose change [-]
gap
wid
th [
mm
]
-2,00
-1,00
0,00
1,00
2,00
3,00
4,00
5,00
-0,20 0,00 0,20 0,40 0,60 0,80
dose change [-]
gap
wid
th [
mm
]
Accuracy calibration function method II
• difference between measured (line) and expected gap (dots) width.
• accuracy calibration functions: – 0.09 mm (1 SD) for linac 1 – 0.12 mm (1 SD) for linac 2
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
1,50
2,00
2,50
3,00
3,50
4,00
4,50
5,00
5,50
0,38 0,42 0,46 0,50 0,54 0,58
dose change [-]
ga
p w
idth
[m
m]
Measurements
gap width and gap position:
• long-term reproducibility: segments from left to right (20 times in 10 weeks)
• short-term reproducibility: segments from left to right (5 times in succession)
• hysteresis:
– segments from right to left (5 times in 10 weeks)
– segments in random order (5 times in 10 weeks)
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Comparison
• t-test for comparison of both methods:
– difference is 0.004 ± 0.14 mm (1SD)
• statistical: different because of many measurements
• clinical: difference not relevant
• only results obtained with method I are shown
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Results
• results of MLC 2 are split because of calibrations
* Vieira et al. Radiother. Oncol. (2005)
• mean gap width error: 1.26 mm due to:
– a difference between light and radiation field
– a blurred light field edge
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
gap width error [mm] gap position error [mm]
MLC 1 1.12 ± 0.43 -0.10 ± 0.32
MLC 2 (1-7) 1.26 ± 0.49 0.43 ± 0.30
MLC 2 (8-13) 1.30 ± 0.35 -0.07 ± 0.60
MLC 2 (14-20) 1.61 ± 0.51 -0.32 ± 0.58
Vieira et al. * 1.21 ± 0.22
Results
• criterion dose error: 3%
criterion position error: 3 mm
• mean gap width error: 1.26 mm
under dosage of 19-25% (15-20% per mm)
4-5% per plan (if plan consists of 5 beams)
• calibration method is not accurate enough for IMRT
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Results: dependency
• gap width and gap position not dependent on:
– abutment position
– leaf pair number
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Results: gap width error in time
MLC 1
MLC 2 drift (0.35 mm)
-0,50
0,00
0,50
1,00
1,50
2,00
2,50
3,00
3,50
0 2 4 6 8 10 12 14 16 18 20 22
days of measurement [-]
gap
wid
th e
rro
r (m
eau
red
-exp
ecte
d)
[mm
]
-0,50
0,00
0,50
1,00
1,50
2,00
2,50
3,00
3,50
0 2 4 6 8 10 12 14 16 18 20 22
days of measurement [-]
gap
wid
th e
rro
r (m
eau
red
-exp
ecte
d)
[mm
]
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Results: gap position error in time
MLC 1
MLC 2
-1,50
-1,00
-0,50
0,00
0,50
1,00
1,50
0 2 4 6 8 10 12 14 16 18 20 22
days of measurement [-]
gap
po
siti
on
err
or
(mea
ure
d-e
xpec
ted
) [m
m]
-1,50
-1,00
-0,50
0,00
0,50
1,00
1,50
0 2 4 6 8 10 12 14 16 18 20 22
days of measurement [-]
gap
po
siti
on
err
or
(mea
ure
d-e
xpec
ted
) [m
m]
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
rotation: 0.6 *
* Bayouth et al. Med. Phys. 30 (2003)
Results: reproducibility gap width
• short-term reproducibility (1 SD):
0.17 mm and 0.12 mm for MLC 1 and MLC 2
• long-term reproducibility (1 SD):
0.22 mm and 0.18 mm for MLC 1 and MLC 2
3% dose error criterion 15% dose error in beam (using 5 beams) maximum allowed gap width error: 0.75 mm (3 SD) (using 20% per mm)
maximum gap width deviation for each leaf pair from its average: 0.75 mm and 0.70 mm for MLC 1 and MLC 2
MLCs are accurate enough for IMRT
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Results: hysteresis
• gap width:
– no hysteresis
• gap position:
– small hysteresis for MLC 1 ( 0.40 mm)
– no hysteresis for MLC 2
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Conclusions
• an accurate leaf verification method has been developed
• the current leaf calibration method using the light field is not accurate enough for IMRT
• the day-to-day variation in gap width is accurate enough for typical IMRT plans
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Recommendations
• not possible to calibrate the leaves using this leaf verifation method
• method can be used for:
– checking the accuracy of the leaf calibration
– verifying the leaf positions in order to change the encoder value using MLCCHK
reduce calibration frequency
• investigate hysteresis for different gantry angles
backgroundgoals
calibrationacquisitionmethod Imethod II
measurementscomparison
resultsconclusions
recommendations
Questions?