30
Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

  • View
    225

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Lotte Verbunt

Investigation of leaf positioning accuracy

of two types of Siemens MLCsmaking use of an EPID

Page 2: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Overview

• background

• goals

• calibration

• image acquisition

• method I

• method II

• measurements

• comparison

• results

• conclusions

• recommendations

Page 3: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Dr. Bernard Verbeeten Institute Tilburg

• independent institute for radiotherapy and nuclear medicine:

- 11 radiotherapy oncologists

- 3 nuclear physicians

- 6 medical physicists

- 1 general doctor

• 2360 new radiation treatment patients a year (2004)

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 4: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Linear accelerator and EPIDbackground

goalscalibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

accelerationbending

through patient

conversion into light

reflection by mirror

detection by camera

Page 5: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Multileaf collimator

two types of Siemens MLCs:

MLC 1: 29 leaf pairs; 2 mm accuracy

MLC 2: 41 leaf pairs; 1 mm accuracy

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Y jaw

leaf bank

central axis

Page 6: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Project goals

• development of a leaf verification method applicable to a Siemens linac using a CCD-camera based EPID

• verification of the leaf positions of two different types of Siemens MLCs and to check whether the MLCs and the current leaf calibration method are accurate enough for IMRT

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 7: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Why is this of interest?

• Intensity Modulated Radiation Therapy (IMRT):

- several beams from different directions

- each beam consist of several (abutting) segments

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 8: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Leaf calibration method: light field

• performed with EPACtool

• using light field and grid lines from graph paper

• 4-points calibration: 20, 10, 0, and -10 cm

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations left leaf bank right leaf bank

-20 -10 0 10 20

Page 9: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Leaf calibration method: light field

• performed with EPACtool

• using light field and grid lines from graph paper

• 4-points calibration: 20, 10, 0, and -10 cm

• uncertainties because:

– a difference between light and radiation field

– a blurred light field edge

– a possible rotation of graph paper

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 10: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Leaf verification methods: radiation field

• method I: 50 % dose value

• method II: from dose change to leaf positions

both methods use same input from the EPID

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 11: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Image acquisition

• 6 segments, 5 cm wide, 2 mm overlap of the leaves

black = blocked by the leaveswhite = unblocked irradiated part

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 12: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Image acquisition

• 6 segments, 5 cm wide, 2 mm overlap of the leaves

• barrel distortion correction

32

21 dddu RaRaRR

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 13: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Image acquisition

• 6 segments, 5 cm wide, 2 mm overlap of the leaves

• barrel distortion correction

• radiation center and pixel dimensions

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 14: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Method I

• take a mean profile for each segment and for each leaf pair

• fit a polynomial function between 20 and 80% of the dose value

• determine leaf position (50% dose value)

position [pixel]

posi

tion

[pix

el]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

500

0 100 200 300 400 5000

10

20

30

40

50

60

70

80

90

100

110

position [pixel]

rela

tive

dose

[%]

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 15: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Accuracy method I

• images:

– nine 3 cm wide segments

– flat side of the leaves

– 90 rotation

• accuracy leaf position determination: 0.08 ± 0.18 mm (1 SD)

• shift of ± 0.45 mm at abutment leaf ends for MLC 2

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 16: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Method II

• sum all segments

• take a mean profile for each leaf pair

• calculate the dose change at each abutment

• convert the dose change into gap width

position [pixel]

po

sitio

n [p

ixe

l]

100 200 300 400 500

50

100

150

200

250

300

350

400

450

5000 100 200 300 400 500

0

20

40

60

80

100

position [pixel]

rela

tive

do

se [%

]

extreme dose

average dose

ji

jiji

D

EDd

,

,, 1

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 17: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Calibration function method II

• calibration function f :

• six tests with different gap widths (-1 to 4 mm)

• plot: intended gap width versus dose change

• dose change is 15-20 % per mm gap width

gap width dose changeif

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

-2,00

-1,00

0,00

1,00

2,00

3,00

4,00

5,00

6,00

7,00

-0,15 0 0,15 0,3 0,45 0,6

dose change [-]

gap

wid

th [

mm

]

-2,00

-1,00

0,00

1,00

2,00

3,00

4,00

5,00

-0,20 0,00 0,20 0,40 0,60 0,80

dose change [-]

gap

wid

th [

mm

]

Page 18: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Accuracy calibration function method II

• difference between measured (line) and expected gap (dots) width.

• accuracy calibration functions: – 0.09 mm (1 SD) for linac 1 – 0.12 mm (1 SD) for linac 2

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

1,50

2,00

2,50

3,00

3,50

4,00

4,50

5,00

5,50

0,38 0,42 0,46 0,50 0,54 0,58

dose change [-]

ga

p w

idth

[m

m]

Page 19: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Measurements

gap width and gap position:

• long-term reproducibility: segments from left to right (20 times in 10 weeks)

• short-term reproducibility: segments from left to right (5 times in succession)

• hysteresis:

– segments from right to left (5 times in 10 weeks)

– segments in random order (5 times in 10 weeks)

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 20: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Comparison

• t-test for comparison of both methods:

– difference is 0.004 ± 0.14 mm (1SD)

• statistical: different because of many measurements

• clinical: difference not relevant

• only results obtained with method I are shown

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 21: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results

• results of MLC 2 are split because of calibrations

* Vieira et al. Radiother. Oncol. (2005)

• mean gap width error: 1.26 mm due to:

– a difference between light and radiation field

– a blurred light field edge

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

gap width error [mm] gap position error [mm]

MLC 1 1.12 ± 0.43 -0.10 ± 0.32

MLC 2 (1-7) 1.26 ± 0.49 0.43 ± 0.30

MLC 2 (8-13) 1.30 ± 0.35 -0.07 ± 0.60

MLC 2 (14-20) 1.61 ± 0.51 -0.32 ± 0.58

Vieira et al. * 1.21 ± 0.22

Page 22: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results

• criterion dose error: 3%

criterion position error: 3 mm

• mean gap width error: 1.26 mm

under dosage of 19-25% (15-20% per mm)

4-5% per plan (if plan consists of 5 beams)

• calibration method is not accurate enough for IMRT

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 23: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results: dependency

• gap width and gap position not dependent on:

– abutment position

– leaf pair number

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 24: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results: gap width error in time

MLC 1

MLC 2 drift (0.35 mm)

-0,50

0,00

0,50

1,00

1,50

2,00

2,50

3,00

3,50

0 2 4 6 8 10 12 14 16 18 20 22

days of measurement [-]

gap

wid

th e

rro

r (m

eau

red

-exp

ecte

d)

[mm

]

-0,50

0,00

0,50

1,00

1,50

2,00

2,50

3,00

3,50

0 2 4 6 8 10 12 14 16 18 20 22

days of measurement [-]

gap

wid

th e

rro

r (m

eau

red

-exp

ecte

d)

[mm

]

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 25: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results: gap position error in time

MLC 1

MLC 2

-1,50

-1,00

-0,50

0,00

0,50

1,00

1,50

0 2 4 6 8 10 12 14 16 18 20 22

days of measurement [-]

gap

po

siti

on

err

or

(mea

ure

d-e

xpec

ted

) [m

m]

-1,50

-1,00

-0,50

0,00

0,50

1,00

1,50

0 2 4 6 8 10 12 14 16 18 20 22

days of measurement [-]

gap

po

siti

on

err

or

(mea

ure

d-e

xpec

ted

) [m

m]

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

rotation: 0.6 *

* Bayouth et al. Med. Phys. 30 (2003)

Page 26: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results: reproducibility gap width

• short-term reproducibility (1 SD):

0.17 mm and 0.12 mm for MLC 1 and MLC 2

• long-term reproducibility (1 SD):

0.22 mm and 0.18 mm for MLC 1 and MLC 2

3% dose error criterion 15% dose error in beam (using 5 beams) maximum allowed gap width error: 0.75 mm (3 SD) (using 20% per mm)

maximum gap width deviation for each leaf pair from its average: 0.75 mm and 0.70 mm for MLC 1 and MLC 2

MLCs are accurate enough for IMRT

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 27: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Results: hysteresis

• gap width:

– no hysteresis

• gap position:

– small hysteresis for MLC 1 ( 0.40 mm)

– no hysteresis for MLC 2

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 28: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Conclusions

• an accurate leaf verification method has been developed

• the current leaf calibration method using the light field is not accurate enough for IMRT

• the day-to-day variation in gap width is accurate enough for typical IMRT plans

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 29: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Recommendations

• not possible to calibrate the leaves using this leaf verifation method

• method can be used for:

– checking the accuracy of the leaf calibration

– verifying the leaf positions in order to change the encoder value using MLCCHK

reduce calibration frequency

• investigate hysteresis for different gantry angles

backgroundgoals

calibrationacquisitionmethod Imethod II

measurementscomparison

resultsconclusions

recommendations

Page 30: Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID

Questions?