Citation preview

Page 1: Log term test

Log term testJohan M., Vito P.16/12/2013

Page 2: Log term test

Setup

1 SRB (4chs)

MPOD

30 COVERs

Page 3: Log term test

The TestThe test consists to repeat every

minute a noise scan over 24h. HV on but no active gas

Page 4: Log term test

The TestThe test consists to repeat every

minute a noise scan over 24h. HV on but no active gas

Page 5: Log term test

Noise Scan ResultsIn order to study the behavior of our electronics in time we ar computing the weighted mean and sigma of each channel versus the time.

Page 6: Log term test

Noise Scan Results (Sigmas)

Page 7: Log term test

Noise Scan Results (Means)

Page 8: Log term test

Noise Scan ResultsIn order to study the behavior of our electronics in time we are computing the weighted mean and sigma of each channel versus the time. The mean variation is less than 1mV over 24h (no temperature or humidity monitoring).

Page 9: Log term test

Noise Scan ResultsIn order to study the behavior of our electronics in time we ar computing the weighted mean and sigma of each channel versus the time. The mean variation is less than 1mV over 24h (no temperature or humidity monitoring). Few (~8/~1800) runs are clearly deviating from the collective behavior, but inspecting them…

Page 10: Log term test

“Failing” Noise Scans

Page 11: Log term test

“Failing” Noise Scans

Page 12: Log term test

“Failing” Noise Scans

Page 13: Log term test

Future Plans The long term test we performed up

to now does not suggest problems in our apparatus.

Possible external sources of noise have been found (to be confirmed).

Possible baseline variations.The apparatus Johan prepared for the

test is ready to be extensively used. A fully comprehensive test will be

performed from January.