24
LISST-SL plots

LISST-SL plots

  • Upload
    bryga

  • View
    55

  • Download
    0

Embed Size (px)

DESCRIPTION

LISST-SL plots. R0681302. Pump takes ~2 mins to lock in Data no good. R1091106. No apparent issue with pump voltage. Apart from the 2 high conc values and open-ended PSD data looks good. R1101437. Pump not locked on for the first minute; possibly lock-on after 1 min? - PowerPoint PPT Presentation

Citation preview

Page 1: LISST-SL plots

LISST-SL plots

Page 2: LISST-SL plots

R0681302

Pump takes ~2 mins to lock inData no good

Page 3: LISST-SL plots

R1091106

No apparent issue with pump voltage.Apart from the 2 high conc values and open-ended PSD data looks good

Page 4: LISST-SL plots

R1101437

Pump not locked on for the first minute; possibly lock-on after 1 min?Upcast probably OK. Note constant tau, but variation in MC. PSD looks fine.

Page 5: LISST-SL plots

R1101500

Pump not locked on. Data not OK.

Page 6: LISST-SL plots

R1120717

Velocity very low, so pump shuts off then starts up again.

Page 7: LISST-SL plots

R1120727

Pitot tube contamination, not reliable data.

Page 8: LISST-SL plots

R1121523

Pump resetting. Data not OK.

Page 9: LISST-SL plots

R1121618

Pump resetting after ~1 minute. First minute may be OK..

Page 10: LISST-SL plots

R1160408

Pump seems to track towards the end, last ~10 data points should be OK.

Page 11: LISST-SL plots

R1301058

Pump never locked in. Data no good (velocity, depth OK).Why max velocity in the middle?

Page 12: LISST-SL plots

R1301440

Pump never locked in. Data no good (velocity, depth OK).

Page 13: LISST-SL plots

R1301447

Excellent pump lock-in and tracking of velocity. Good PSD, good concentration.

Page 14: LISST-SL plots

R1311450

Excellent pump lock-in and tracking of velocity.

Page 15: LISST-SL plots

R1311507

Good pump lock-in for the last half.

Page 16: LISST-SL plots

Kickapoo Creek nr Bloomington

Page 17: LISST-SL plots

Kickapoo Creek nr Mattoon

Page 18: LISST-SL plots

Senachwine Creek at Chillicothe

Page 19: LISST-SL plots

VC vs MC, linear

Page 20: LISST-SL plots

VC vs MC, log

Page 21: LISST-SL plots

MC/VC vs tau

Page 22: LISST-SL plots

MC/VC vs velocity

Page 23: LISST-SL plots

MC/VC vs Mass%<0.0625mm

Page 24: LISST-SL plots

MC/VC vs Volume%<0.0625mm