    Trainees Intellectual Forum

    Medical Linear Accelerators(Linacs)

    Medical Physics, Aga Khan University-Karachi 1

    Ghufran Ali

    Trainee Medical Physicist

    ACADEMIC SESSION

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    2/43

    Contents

    1. Particle Accelerators

    2. Medical Linacs

    Schematic diagram

    Linac configurations Linac components

    Beam interlocks

    Medical Physics, Aga Khan University-Karachi

    2

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    3/43

    PARTICLE ACCELERATORS

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    4/43

    PARTICLE ACCELERATORS

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    5/43

    PARTICLE ACCELERATORS

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    6/43

    Betatron

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    7/43

    Cyclotron

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    8/43

    Microtron

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    9/43

    Medical Linacs

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    10/43

    Medical Linacs

    Medical Physics, Aga Khan University-Karachi

    10

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    11/43

    Medical Physics, Aga Khan University-Karachi

    11

    Schematic Diagram (simple)

    1-80keV electrons

    pulse- injected into

    accelerator tube

    Evacuated Cu tube;

    Transfer microwave

    energy (from magnetron

    or klystron) to electrons

    X-ray target, flattening

    filter, electron scattering

    foil, collimators, etc

    Emits high voltage pulses (PRF: 30-300 Hz) to magnetron & e - gun; to

    synchronize electron injection with

    microwave power.

    Source of microwave power

    (~3000MHz);

    Magnetrons up to ~20 MeV;

    Klystrons at higher energies

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    12/43

    Medical Physics, Aga Khan University-Karachi

    12Schematic diagram (detailed)

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    13/43

    Linac Configurations:

    vertical mounting (LA 1 at AKUH)

    Medical Physics, Aga Khan University-Karachi

    13

    For 4-6 MV linacs

    e-gun & x-ray target

    part of waveguide

    aligned with isocentreno bending magnet

    & transport system

    X-ray target embedded into waveguide; therefore, this design

    cannot produce electron beam

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    14/43

    Linac Configurations:

    horizontal mounting(LA 2 at AKUH)

    Medical Physics, Aga Khan University-Karachi

    14

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    15/43

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    16/43

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    17/43

    RF Power Generation

    Medical Physics, Aga Khan University-Karachi

    17

    RF power amplifier

    Source of microwave power (~3000MHz);

    Magnetrons up to ~20 MeV; Klystrons at

    higher energies

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    18/43

    RF Power Generation

    Medical Physics, Aga Khan University-Karachi

    18

    to synchronize electron injection with microwave power.

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    19/43

    Waveguides

    Medical Physics, Aga Khan University-Karachi

    19

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    20/43

    RF Power transport

    Medical Physics, Aga Khan University-Karachi

    20

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    21/43

    Travelling Waveguide

    Medical Physics, Aga Khan University-Karachi

    21

    Above 8 MeV, accelerator structures becomes too long (> 1 m)

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    22/43

    Standing Waveguide

    Medical Physics, Aga Khan University-Karachi 22

    More efficient; half as long;

    but more expensive

    Most modern linacs

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    23/43

    Auxiliary System

    Medical Physics, Aga Khan University-Karachi

    23

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    24/43

    Electron Beam Transport

    Medical Physics, Aga Khan University-Karachi

    24

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    25/43

    Coils & Bending Magnet

    Medical Physics, Aga Khan University-Karachi

    26

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    26/43

    Electron Beam Transport

    Medical Physics, Aga Khan University-Karachi 27

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    27/43

    Linac Treatment Head

    Medical Physics, Aga Khan University-Karachi 28

    d

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    28/43

    Medical Physics, Aga Khan University-Karachi 29

    Linac Treatment HeadElectron

    Beam

    IonChamber

    Secondary

    Collimators

    Electron

    applicator

    Patient

    Primary

    Collimator

    Flattening Filter

    Patient

    Target

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    29/43

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    30/43

    Target and Flattening Filter

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    31/43

    Carousel

    Medical Physics, Aga Khan University-Karachi

    32

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    32/43

    Photon Beam Collimation

    Medical Physics, Aga Khan University-Karachi 33

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    33/43

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    34/43

    Medical Physics, Aga Khan University-Karachi

    35

    Production of Electron Beams

    6 to 30 MeV energy beams are available

    Electron exits through a thin beryllium window

    Two techniques for producing uniform beam

    Pencil beam scattering (Pb or Cu foils)

    Pencil beam scanning

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    35/43

    Electron Beam Collimation

    Medical Physics, Aga Khan University-Karachi 36

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    36/43

    Electron Mode

    INOR

    Electron

    Beam

    Scattering Foil

    IonChamber

    SecondaryCollimator

    Electron

    applicator

    Patient

    Primary

    Collimator

    Scattering Foil

    Patient

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    37/43

    Dose Monitoring

    Medical Physics, Aga Khan University-Karachi 38

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    38/43

    Dose Monitoring

    Medical Physics, Aga Khan University-Karachi 39

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    39/43

    Dose Monitoring

    Medical Physics, Aga Khan University-Karachi 40

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    40/43

    Dose Monitoring

    Medical Physics, Aga Khan University-Karachi 41

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    41/43

    Dose Monitoring

    Medical Physics, Aga Khan University-Karachi 42

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    42/43

  • 8/12/2019 Linacs on 11-11-2013 - Copy

    43/43

    References

    Chapter 4:The Physics of radiation Therapy, Faiz M.KHAN

    Sections 4.3 - 4.8

    Chapter 5:

    Radiation Oncology Physics: A Handbook

    for Teachers and Students, IAEA, Vienna,2007

    Section 5.5