Is there life after LIMS? LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but thankfully there is beer.

LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Embed Size (px)

Citation preview

Page 1: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Is there life after LIMS?

LIMS for Light Stable IsotopesCreated by T.B. Coplen, USGS Reston

Benjamin Harlow, Stable Isotope Core Facility at Washington State University

No, but thankfully there is beer.

Page 2: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Is LIMS for you?

• Do you process >100 samples per year?

• Do you process samples for multiple clients?

• Do you measure multiple isotopes?

• Do you use more than one instrument?

Page 3: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Is LIMS for you…

• Do you already have a database or spreadsheets that

work for you?

• LIMS by the USGS is not the only way to go

• LIMS by the USGS is the most readily available,

supported, and FREE software for Light Stable Isotopes

• If yes, start sooner rather than later!

Page 4: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Thankfully there is beer.

Page 5: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Benefits to using LIMS

• All your information in one place

• Rapid and consistent treatment of data

• Hourly drift corrections from multiple standards

• Linearity correction

• Data revision and safeguards

• Powerful search, average and export functions

• Report and Invoice generation

Page 6: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Limitations

• Can only have as many mass spectrometers or lasers as letters in the English alphabet

• Normalization by linear regression is all that is currently possible

• Some data manipulations may still be required via spreadsheet, etc. prior to importing

Page 7: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

First things first

• MS Access supported versions– MS Office 2007, 2010, 32 bit only

• Required Security settings– Add LIMS backend database to “trusted locations”– Enable all macros

Page 8: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Set trusted locations, macros

Page 9: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Setup

• 3 folders, 4 files: – Backend DB (LM8DATA.mdb) – Frontend DB (LIMS9 for office 2010.mdb) + Preferences file – Backup

• Back up Backend DB often!• Back up Preference files periodically (especially for new

Front end versions)• Different users can have their own front end and preferences• Can install on one computer, network, cloud• WSU: dropbox; share DB with 7 computers (2 RAID), multiple

smartphones

Page 10: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Projects

View projects, print reports, and export data to Excel Import project data from Excel forms Create new projects Add new clients

Special Features

Add new instruments Add or edit laboratory standards Backup and customize the database Design analysis templates Add or edit media or procedures

Sample Information

Create sequence files that can be imported into mass spectrometers Search, view, print and edit specific client samples Search, view, and edit extractions

Analyses in Progress

Import IRMS or LAS data Determine and apply blank corrections and drift corrections Normalize results to isotope scales View or edit isotope results Print or save isotope results to Excel files Evaluate results and track laboratory QA/QC

Store final accepted results for reporting

Page 11: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but
Page 12: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Import project to LIMS

• Manual entry or Excel spreadsheet• Concept of “placeholders”• Adjusting the next “Our Lab ID”• Reference ID’s often set as a low number• Example: References <1000, samples >1000• Keeps all references in one area of the sample table

Page 13: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but
Page 14: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Enter new clients manually before creating the first project for that client

Page 15: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Adding a project manually

Date, Name, Media is the minimum information needed

More information is better; i.e. project title and account are good to reference later

Do not enter duplicate ID’s and avoid special characters for sample IDs

Page 16: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Submit to LIMS

• Simple excel templates are available for modification

• Reduces typographical errors and valuable time of lab members

• More information is good• Be unique, but avoid special

characters $/\@#$%^&*!

Page 17: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Sample excel submission form

• Customize to your lab: Letterhead, graphics, captions• Captions must match exactly to what you defined in “special features”• Use different forms for different types of samples• Protect certain cells to reduce customer “tinkering”

Page 18: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Other features

• Track lab extraction procedures– Make notes about sample prep– Created primarily for offline procedures

• Create labels for samples• Use templates to create run sequences of

samples and references

Page 19: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Selling (to) yourself

• Create a “client” entry for your lab tests and reference material

• We use “reference” and “test”• All international, commercial, and lab developed

reference material is logged here• Reference lab ID’s get defined in reference table:

Automatically fit in a normalization• Test lab ID: treat as a blind sample (Quality Control)

Page 20: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

IRMS/software setup

• Machine must use sequential analysis number, with

single letter prefix (ex: A-number)

• Isodat: Configurator > Options > Global settings

• Enter old analyses or those without prefix:

“concatenate” function in excel

• Note analysis number if you ever restore IRMS

software!

Page 21: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Know your media

• KISS IT• More classifications can be useful• Can add more at any time• Tailor specific to your lab

Page 22: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

• Media originally designed for 2 isotopes only!

• It is possible to store, edit and average 4.

• How? Stay tuned

Media

Page 23: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Procedure Codes

Page 24: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Low and High procedure Codes

Page 25: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Running samples

• LIMS ID# “Our lab ID” is most important identifier to use in sample sequences

• Use LIMS template or design your own in excel

• Template: LIMS automatically designs sample sequence

Page 26: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Other features: Creating a sample list/template

Page 27: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Another way to make a run sequence

• Import excel file and generate LIMS ID’s

• Copy/Paste sample information into a blank tab of the submission worksheet

•Insert rows for running standards and QC’s

• Can customize to instrument service state, sample types, desired timing, etc.

Page 28: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Then you run the samples

What Paul said…

Page 29: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Exporting results and importing to LIMS

• Must export specific column headings required by LIMS

• Line #, Analysis #, time code, gasconfig, area, peak#, identifier 1, delta columns, is ref, comment, Rt

• All peaks for reference gas must be marked “Is Ref = 1” or removed

• Perform data preprocessing before importing (i.e. your own linearity, drift or mass fraction corrections)

Page 30: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Export tips

• Isodat: If a project got created with some special character, don’t export that column

• Check chromatographs prior to exporting. Is reprocessing necessary?

• Garbage in garbage out

Page 31: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Importing data

Option to apply linearity correction at this point

Page 32: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but
Page 33: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

3 point normalization example

Page 34: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Add elemental and blank corrections

Page 35: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Normalize std gasses and data

• Seldom used

• Must import and properly code ref gasses

• Provides option to salvage data

Page 36: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Applying an hourly drift correction

Use drift from any single standard

Override: combination of all accepted standards, or custom

Apply two drifts: must break up the sequence into two corrections

Page 37: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Evaluating/editing samples in progress

Mean, SD and outlier information

Note: When samples are “in progress” mean, SD are available

Some may choose to report at this point

Page 38: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Rapid tracking of QC/QA

Page 39: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Export full results for samples

Page 40: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Store final results

Stores a single value from all accepted reps

Marks samples as completed

Provides an extra layer of protection to data

Necessary to use LIMS report function

Page 41: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Add stored results to in process

Unlocks samples for editing

Can be done at any time

Store function may be unnecesary for some users

Page 42: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Storing an additional isotope or other numeric data

• The “comment” field now averages numeric values and

allows editing

• The high and low comment provide two additional storage

points

• Include data calculated by a different detector, i.e. CNHOS%

• Use with procedures that generate data for more than two

isotopes, i.e. thermal conversion of N2O.

Page 43: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Invoice generation and payment tracking

Page 44: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Can I use LIMS for GCC-IRMS?

• Yes you can!

• Treat each vial as a project and target compounds are

samples in the project

• Use Rt or component ID feature in LIMS (extraction ID,

Comment, Rt)

• Internal standards should be in each vial

Page 45: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

• Alternative method: Modify file to import

• Sort by component (Rt) and time code

• Assign the same analysis number

• Create unique peak number for every compound

in a sample

Can I use LIMS for GCC-IRMS?

Page 46: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Can I use LIMS for Lasers?

• Yes you can!• See Len Wassenaar’s session on Wednesday

Page 47: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

More Advanced

• Open the tables and look in there• Cut, copy, paste; but watch for linked tables• Make your own queries

Page 48: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Lab specific queries

• Annual reports, audits and cost evaluations• # days runtime/machine/peripheral/Fiscal year• # samples/machine/peripheral/year• Count of customers by country, state, institution, and

break down by sample type• # of samples billed by type• Average turnaround by sample type• Long term QC/QA

Page 49: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Lab specific queries

• Annual reports, audits and cost evaluations• # days runtime/machine/peripheral/Fiscal year• # samples/machine/peripheral/year• Count of customers by country, state, institution, and

break down by sample type• # of samples billed by type• Average turnaround by sample type• Long term QC/QA

Page 50: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

Questions?

Page 51: LIMS for Light Stable Isotopes Created by T.B. Coplen, USGS Reston Benjamin Harlow, Stable Isotope Core Facility at Washington State University No, but

What if…

• I need to delete a project• I entered the wrong media code• I can’t find my samples in LIMS