LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL team

LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

Embed Size (px)

Citation preview

Page 1: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452-v51

Pre-stabilized Laser System (PSL)

Technical StatusNSF Review of Advanced LIGO Project

Benno Willke, AEI Hannoverand the AdvLIGO PSL team

Page 2: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

Pre-stabilized Laser System (PSL) - scope

• deliver coherent radiation to allow interferometric detection of gravitational waves» 180W laser with high reliability and low fluctuations

• reduce temporal and spatial fluctuations of the light» first layer of power and frequency stabilization before light

enters the input optics (IO) subsystem» provide actuator-inputs for further frequency stabilization

layers (GW band: modecleaner, long term drift: long interferometer arms)

» stabilize laser power downstream of the suspended modecleaner

• provide control interfaces and diagnostic tools• define and implement laser safety procedures

2

Page 3: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G11004523

PSL Requirements

Power:» 165W in gaussian TEM00 mode» less than 5W in non- TEM00 modes

prestabilization

» frequency noise

» power noise

» pointing fluctuations

actuators PSL-external control:

» power actuator

» frequency actuator

Drift:

» 5% power drift over 24hr

» 100kHz frequency drift over 100s

101

102

103

104

10-2

10-1

100

Frequency (Hz)

Fre

qu

ency

No

ise

(Hz/

Hz1/

2 )

101

102

103

104

10-9

10-8

10-7

10-6

rela

tiv

e p

ow

er

no

ise

[1

/sq

rt(H

z)]

frequency [Hz]

Page 4: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

advanced LIGO laser - layout

2W 35W 180W 165W in TEM00

eLIGO laser

4

Page 5: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

purpose of PSL components I

high stability

generate high power with

low power noise

good spatial profile

preserve frequency stability of NPRO

• filter power noise at MC FSR

• filter spatial beam shape and its fluctuations

5

Page 6: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

purpose of PSL components II

6

diagnostic

frequency stabilization

powerstabilization

Page 7: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

performance of the PSL Reference System @ AEI

no data no data

7

Page 8: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

Winkelmann et. al, Appl. Phys. B. (2011) DOI 10.1007/s00340-011-4411-9

200W high power laser

High-powerFaraday isolator

beam shutter (internal)

beam shutter (external)

Long range actuator

Electrical interface

35 W MOPA

High-power oscillator

8

Page 9: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

new pre-mode-cleaner

bow tie configuration, round-trip length of 2m Finesse of 124, circulating power 9kW (200W input) Linewidth 575 kHz (HWHM) Length control with automatic lock acquisition PZT actuator / thermal actuator to off-load PZT

50 cm

Old Requirement

New Requirement

9

Page 10: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

alignment fluctuation downstream of PMC

10

Page 11: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

frequency noise in-loop performance

Reference System, AEI, (2011) 11

Page 12: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

ISS – inner loop (on PSL table)

sensing box with in-loop, out-of-loop and quadrant photodiode

low-scattering design quadrant diode for alignment and pointing

control

power stabilization sensor meets aLIGO inner-loop requirement

12

Page 13: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

power noise behind PMC

13

Page 14: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

ISS- outer loop (sensing in HAM2)

four photodiodes per detector each aligned to lowest pointing

coupling 50mA photocurrent per photodiode in dust-free environment with reduced

pointing fluctuations aLIGO parts expects to arrive soon

14

Page 15: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

lock acquisition - PSL Reference System

15

Page 16: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

first observatory laser arrived at LLO in Jan 2011

16

Page 17: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

installation in two locations:laser-diode-room and laser room

Laser Area Enclosure

Laser Diode Room

17

Page 18: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

LLO - installation laser diode room

• chiller and water lines installed and tested

• all laser diodes (#32) installed and tested

• optical fibers to laser room connected and tested

• safety interlock system tested and accepted by laser safety officer

18

Page 19: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

LLO Laser Area Enclosure

• laser room with HEPA filtered circulating air system and acoustic shielding

• high flow during installation and maintenance

• low flow to keep positive pressure in science mode

• ante room for preparation work, material transfer and storage

19

Page 20: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

current installation status LLO

20

• 35 W laser installed and characterized

• 200W laser installed and under characterization

• diagnostic breadboard installed and operable

• pre-modecleaner placed on table

• all optical mounts assembled and in position

• stabilization electronics installed and connected to realtime control and data acquisition system

Page 21: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

HPL production continues at LZH

• almost all parts delivered• all HPL components pre-assembled• second HPL in final test phase

(since two weeks)• third HPL assembled

21

Page 22: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

second and third observatory laser at LZH

22

Page 23: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

OBS1 – LLO1 OBS2 – LHO1 (preliminary data)

7.4% 6.9%

power noise and higher order mode content

23

Page 24: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

status summary

• we fully commissioned one PSL at AEI (Reference System)» laser and all control loops operational» fully automated lock acquisition (including FSS)» could demonstrate to compensate diode aging» continue long term operation test

• installation of aLIGO PSL at LLO underway» Laser Area Enclosure (LAE) finished» laser diodes, chiller and Beckhoff control installed» 35W and 200W laser in operation» commissioning of is stabilization progressing well » realtime control and data acquisition fully functional

• fabrication of H2 laser finished, will be shipped mid June 2011• all components for photo diode array for outer loop ISS delivered

24

Page 25: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

PSL - project organization

• AEI scope

» laser will come as a ready-to-install package to LIGO sites

» refurbishment of the 35W lasers used for eLIGO

• LIGO scope / Advanced LIGO project

» laser area enclosure (cleanliness, laser safety) and laser diode room

» power supplies and data acquisition and controls computer hardware

» outer loop power stabilization.

» Safety procedures and interfaces

• AEI and LIGO

» once rooms and infrastructure is ready, laser will be installed and stabilization will be commissioned

» reuse laser tables and frequency stabilization (with new interface cards)

25

Page 26: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

PSL - labor

Hannover» AEI (5 FTE)

– Kwee, Pickenpack, Pöld, Krämer, Willke– support mechanical and electronic workshop (1.5 dedicated FTEs

)» LZH (3 FTE)

– Weßels, Winkelmann, Puncken, Kluzik, Damjanic» neoLASE (2 FTE)

– Frede, Schulz, Jansen

LIGO – King, Savage, Fyffe, Rodruck– technical support– CDS support

26

Page 27: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

PSL – risks, QA and near term activities

• quality management» diagnostic breadboard will be used for fast and frequent performance tests

– after fabrication in Hannover– after installation at sites– during operation

» same personnel for laser fabrication as for development (Winkelmann, Puncken, Wessels

» electronic fabrication by neoLASE (LZH spin off company, personnel that developed laser (Frede, Schulz) and wrote control software (Veltkamp))

• risk registry» PSL has no entry in major threat list» all „high risk entries“ retired to med / low

• near term plan» install LLO laser» finish laser and PSL fabrication until Oct 2011» install H2 laser at LHO (Oct-Dec 2011)» continue long term performance test of Ref. Sys. at AEI

27

Page 28: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

Safety

• As a Project, we follow the standards suggested by ANSI Z136.1 (2007).» Laser safety is important!» Established Project-wide laser safety policies and

procedures.

• Standard Operation Procedure (SOP) in place for PSL operation at LLO» access control to Laser Hazard Areas» tested protective eye wear is provided» safety interlock system fully functional and accepted by LLO

laser safety officer

2828

Page 29: LIGO-G1100452-v5 1 Pre-stabilized Laser System (PSL) Technical Status NSF Review of Advanced LIGO Project Benno Willke, AEI Hannover and the AdvLIGO PSL

LIGO-G1100452

PSL summary

• finished detailed characterization of the aLIGO PSL reference system, could demonstrate almost all requirements

• infrastructure modifiacitons are finished at LLO and have started at LHO

• first aLIGO PSL was shipped to LLO and is currently commissioned

• fabrication of the second aLIGO PSL is finished, installation will start at LLO in October 2011

29