LIGO-G1000552-V8 1
aLIGO DC Power Distribution
Contents
2
CDS Racks and Rooms
Front End Computer
Front End Computer
Timing Master
18VDC
I/O Extension
Timing Fanout
24VDC
Chassis
Chassis
Chassis
Chassis
Chassis
Chassis
DC Power
DC Power
DC Power
DC Power
DC Power
DC Power
AC
Voltage
Chassis
Chassis
Chassis
Chassis
Chassis
Chassis
MassStorageRoom (MSR)
OperatorControlRoom (OCR)
CDS ElectronicsRoom (CER
)LaserDiode
Room (LDR)
CDS Power
Room (CPR)
CDS Rack Installations
Field I/O
Cables
TimingDistribution
Cables
DCPowerRacks
Field Satellite
Racks
Subsystem
Electronics
Racks
FrontEnd
Racks
FiberOptic
Cables
DC Extension
Cables
FieldDC
ExtensionCords
Computer Users Room (CUR)
Racks and Rooms (RAR)
LIGO-G1000552-V8
CDS Electronics Room (CER) Rack Layout
3
2feet
Each Rack is 39” Deep and 24”
Wide
2 feet
L1-P
EM
-C1
L1-SUS-C1
L1-SUS-C2
L1-SUS-C3
L1-SUS-C4 L1-SUS-C5
L1-SUS-C6
L1-SUS-C7
L1-SUS-C8L1
-SEI-
C1
Fro
nt F
ron
t
Front Front
Fro
nt
Fro
nt
L1-S
EI-
C2
L1-A
OS
-C1
L1-S
EI-
C3
L1-S
EI-
C4
L1-S
EI-
C5
L1-S
EI-
C6
LIGO-G1000552-V8
L1-UNK-C3
L1-UNK-C2
L1-UNK-C1
L1-ISC-C1
L1-ISC-C2
L1-ISC-C3
L1-P
SL-
C1
CDS Electronics Room (CER)
L1-UNK-C4
L1
-UN
K-C
5
SUS ITMX, ITMY, and BS Rack 1
SUS ITM, FM, and BS
Rack 2
SUS ITM, FM, and BS
Rack 1
SUS HAM2 Rack 1
SUS HAM2 Rack 2
SUS ITMX, ITMY, and BS Rack 2
SUS HAM5 and HAM6
Rack 1
SUS HAM5 and HAM6
Rack 2
ISC LVEA Rack 1
ISC LVEA Rack 2
ISC LVEA Rack 3
PS
L R
ack 1
PEM
, D
AQ
R
ack 1
AO
S R
ack 1
Front
SEI
Rack 1
SEI
Rack 2
SEI
Rack 3
SEI
Rack 4
SEI
Rack 5
SEI
Rack 6
4
LLO LDR – Rack Layout
2feet
Each Rack is 39” Deep and 24” Wide
2 feet
Laser Diode Room (LDR)
Changing Room
480V DistributionPanel
208V DistributionPanel
DC Supplies
2feet
Fro
nt
Front
No posts
20
8V
Dis
trib
uti
on
Pan
el
20
8V
Dis
trib
uti
on
Pan
el
New Transformer and Distribution
Panel?
L1-VDC-C5
L1-VDC-C6
L1-V
DC
-C1
L1-V
DC
-C2
L1-V
DC
-C3
L1-V
DC
-C4
LDR Chiller
L1-LDR-C2
L1-LDR-C1
LIGO-G1000552-V8
<< CDS Rack Installations
CDS Power Room (CPR)
2feet
Each Rack is 39” Deep and 24” Wide
2 feet
Laser Diode Room (LDR)
Changing Room
480V DistributionPanel
208V DistributionPanel
DC Supplies
2feet
Fro
nt
Front
20
8V
Dis
trib
uti
on
Pan
el
20
8V
Dis
trib
uti
on
Pan
el
New Transformer and Distribution
Panel?
L1-VDC-C5
L1-VDC-C6
L1-V
DC
-C1
L1-V
DC
-C2
L1-V
DC
-C3
L1-V
DC
-C4
LDR Chiller
L1-LDR-C2
L1-LDR-C1
5
LLO CDS Power Room (CPR) – Rack Layout
LIGO-G1000552-V8
<< CDS Rack Installations
CDS Power Room (CPR)
6
DC Power Rack
Kepco JQE ±24@20A
Kepco JQE ±18@20A
Kepco JQE ±24@20A
Kepco JQE ±18@20A
Kepco JQE ±24@20A
Kepco JQE ±18@20A
Kepco JQE ±24@20A
Kepco JQE ±18@20A
LIGO-G1000552-V8
Kepco JQE ±24@20A
Kepco JQE ±18@20A
Kepco JQE ±24@20A
Kepco JQE ±18@20A
Kepco JQE
Kepco JQE
Kepco JQE
Kepco JQE
DC Power Rack – VAC Wiring12 @ 125V, 20A Circuits per Rack
7LIGO-G1000552-V8
L1-VDC-C1 Power Racks – Required Supplies
L1-SUS-C2
L1-SUS-C1
L1-SUS-C2
L1-SUS-C3
L1-SUS-C4
L1-SUS-C4
L1-SUS-C5
L1-SUS-C5
L1-SUS-C6
L1-SUS-C6
L1-SUS-C7
L1-SUS-C8
L1-VDC-C1 Power Rack
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
L1-SUS-C3
L1-SUS-C1
L1-SUS-C2
L1-SUS-C3
L1-SUS-C4
L1-SUS-C4
L1-SUS-C8
L1-SUS-C5
L1-SUS-C6
L1-SUS-C6
L1-SUS-C7
L1-SUS-C8
A B
L1-SUS-C3
L1-SUS-C1
L1-SUS-C2
L1-SUS-C3
L1-SUS-C4
L1-SUS-C4
L1-SUS-C8
L1-SUS-C5
L1-SUS-C6
L1-SUS-C6
L1-SUS-C7
L1-SUS-C8
L1-SUS-C2
L1-SUS-C1
L1-SUS-C2
L1-SUS-C3
L1-SUS-C4
L1-SUS-C4
L1-SUS-C5
L1-SUS-C5
L1-SUS-C6
L1-SUS-C6
L1-SUS-C7
L1-SUS-C8
L1-VDC-C1 Power Rack
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
Kepco JQE Style
B A
Standard DC Power Connecting Scheme
88LIGO-G1000552-V8
Chassis
Chassis
Chassis
Chassis
Chassis
Chassis
Chassis
Subsystem Electronics RackDC Power Supply Rack
PP
P
P
P
P
P
P
P
P
S
S
S
S
SS
S
S
S
S
S
S
S
S
S S
PPSS
X 24
X 24
PP
PP
S S
S S
SSSS D1002209 Option
02Extension Cords
Power SuppliesPower
Supplies
Pigtail Cords
D1002189 Opt 02 24VDC Power
Strip
D1002189 Opt 01 18VDC Power
Strips
D1002209 Option 01 Extension
Cords
ACAC
D1002475 Pigtail Cords
D1002475 Pigtail Cords
D1002475 Pigtail Cords
D1002475 Pigtail Cords
Customized Extension
Cables
Chassis
Standard DC Power Connecting Scheme - Overview
Chassis
Subsystem Electronics RackDC Power Supply Rack
PP
PP
S
S
S
S
S
S
S
S S
PPSS
AC
X 24
X 24
9
D1002189 Opt 02 24VDC Power
Strip
D1002189 Opt 01 18VDC Power
Strips
Power SuppliesPower
Supplies
Pigtail Cords
D1002475 Pigtail Cords
9
S
S
S
S
S
S
S
D1002209 Option 02
Extension Cords
D1002209 Option 01 Extension
Cords
LIGO-G1000552-V8
D1002475 Pigtail Cords
D1002475 Pigtail Cords
AC
Customized Extension
Cables
Standard DC Power Strip
1010LIGO-G1000552-V8
D1002189 Opt 02 24VDC Power
Strip D1002189 Opt 01 18VDC Power
Strips
D1100034 Opt 02 24VDC Power
Strip
Custom-Length DC Power Extension Cables
LIGO-G1000552-V8 11
A1
A2
A3
A1
A2
A3
+/-18VDC 20A
Custom length cables, metal backshells and screws on both connectors
PPSS
Orange 12/3 25A McMaster-Carr 7394K13 CableConec 3003W3PXX99A10X ConnectorConec 3003W3SXX99A10X ConnectorConec 131C11029X Male ContactsConec 132C11029X Female ContactsDigi-Key 17E-1657-15-ND Backshells
+/-24VDC 20AYellow 12/3 25A McMaster-Carr 8046K16 CableConec 303W3CPXX99A10X ConnectorConec 303W3CSXX99A10X ConnectorConec 131C11029X Male ContactsConec 132C11029X Female ContactsDigi-Key 17E-1657-15-ND Backshells
Standard DC Power Extension Cords
LIGO-G1000552-V8 12
P S
P
S
Standard 3-foot cords, backshells and screws on both connectors
+/- 24VDC 10A +/- 18VDC 10A
Standard DC Power Pigtail Cords
LIGO-G1000552-V8 13
Standard 3-foot length cables and backshells – Plug and Socket Options
PP
+/- 24VDC 20A
+/- 18VDC 20A
SS
Standard DC Power Supply Wiring Scheme
+
+
–
–
White – Positive – A1Black – Common – A2Green – Negative – A3
‘Negative’ power supply
‘Positive’ power supply
14
S S
14LIGO-G1000552-V8
1. Only connect two power supplies that are sourcing the same voltages.2. Use the correct voltage option of D1002475 DC Power Pigtail Socket Cord.
DC Power Supply Voltage and Current Monitor
LIGO-G1000552-V8 15
Pin Signal 1 Voltage + 6 Voltage – (GND) 2 Current + 7 Current – (GND) 3 N/A 8 N/A 4 N/A 9 N/A 5 N/A
Power Supply* Provide a 0-10V full-scale Voltage signal* Provide a 0-10V full-scale Current signal* Present both signals on a DB9 female bulkhead connector mounted somewhere on the supply as follows:
DC Discussions – We agreed that …
LIGO-G1000552-V8 16
•I/O Extenders will be powered separately from Electronics Chasses. This will permit the electronics to be turned off and on without interference to the I/O Extenders.
•3-conductor extension cables will be used between the DC Power Racks and the Electronics Racks. This will provide one return per set (+/-) of supplies.
•All DC Power Supply returns will be tied together (bonded) at the DC Power Rack to avoid ground loops.
•DC Power Racks will all be tied to technical ground.
•Shielding DC extension cables (from DC Power Racks to Electronics Racks) is not necessary.
•Non-standard (18/24) voltages should be a multi-conductor cable solution similar to this solution. See DC jacket colors for voltage conventions.
•Remote sensing will not be used. The length of cable is too great for effective control.
DC Wiring Color Convention
LIGO-G1000552-V8 17
Either single conductor wiring or multi-conductor cabling may be called for. These charts show the DC color code for either the single conductor wiring or the multi-conductor cabling.
Custom DC Power Extension Cables – Components
LIGO-G1000552-V8 18
Connectors
Custom length cables, metal backshells and screws on both connectors
Plug Connector Conec 3003W3PXX99A10XConec 131C11029X Male Contacts
Socket Connector Conec 3003W3SXX99A10XConec 132C11029X Female Contacts
5VDC @13A White 16/3 McMaster-Carr 7589K11
12VDC @13A Green 16/4 McMaster-Carr 7673K63
18VDC @25A Orange 12/3 McMaster-Carr 7394K1318VDC @13A Orange 16/3 McMaster-Carr 7394K33
24VDC @25A Yellow 12/3 McMaster-Carr 8046K1624VDC @13A Yellow 16/3 McMaster-Carr 8046K12
36VDC @13A Black 16/3 McMaster-Carr 8157K64
>100VDC @13AGray(Marked) 16/4 McMaster-Carr 9936K26
Cables
BackshellsAll Plug and Socket Backshells Digi-Key 17E-1657-15ND
Custom DC Extension Cable - BOM
LIGO-G1000552-V8 19
+/- 18VDC 20A (Used between DC Power Racks and Electronics Racks)
18V 20A Custom Extension Cables
Item Part Type Description Designator Manufacturer SupplierSupplier Part
NumberQuantit
y Unit
1 Cable Orange SJTW AWG 12/3 25A Stranded McMaster-Carr 7394K13 100 ft
2 Connector Shell 3W3C Male Combo Shell - No Power Conec Conec 3003W3PXX99A10X 1 @
3 Connector Contact 20 AMP Male Power Contact Conec Conec 131C11029X 3 @
4 Connector Shell 3W3C Female Combo Shell - No Power Conec Conec 3003W3SXX99A10X 1 @
5 Connector Contact 20 AMP Female Power Contact Conec Conec 132C11029X 3 @
6 Backshell Connector Backshell DB15 DIE CAST ZINC Amphenol Digi-Key 17E-1657-15-ND 2 @
7 Standoff4-40 D-Sub Hardware Hex Standoffs, .232
inch Hex L-com SDH 2 @
8 Heat Shrink Heat Shrink Insulation Lab Supplies 1 @
P
S
Orange 12/3 25A McMaster-Carr 7394K13 CableConec 3003W3PXX99A10X ConnectorConec 3003W3SXX99A10X ConnectorConec 131C11029X Male ContactsConec 132C11029X Female ContactsDigi-Key 17E-1657-15-ND Backshells L-Com SDH Standoff
Custom length Cables, backshells and screws on both connectors, standoffs on socket
Custom DC Extension Cable - BOM
LIGO-G1000552-V8 20
+/- 24VDC 20A (Used between DC Power Racks and Electronics Racks)
24V 20A Custom Extension Cables
Item Part Type Description Designator Manufacturer SupplierSupplier Part
NumberQuantit
y Unit
1 Cable Yellow SJEOOW AWG 12/3 25A Stranded McMaster-Carr 8046K16 100 ft
2 Connector Shell 3W3C Male Combo Shell - No Power Conec Conec 303W3CPXX99A10X 1 @
3 Connector Contact 20 AMP Male Power Contact Conec Conec 131C11029X 3 @
4 Connector Shell 3W3C Female Combo Shell - No Power Conec Conec 303W3CSXX99A10X 1 @
5 Connector Contact 20 AMP Female Power Contact Conec Conec 132C11029X 3 @
6 Backshell Connector Backshell DB15 DIE CAST ZINC Amphenol Digi-Key 17E-1657-15-ND 2 @
7 Standoff4-40 D-Sub Hardware Hex Standoffs, .232
inch Hex L-com SDH 2 @
8 Heat Shrink Heat Shrink Insulation Lab Supplies 1 @
PS
Yellow 12/3 25A McMaster-Carr 8046K16 CableConec 303WC3PXX99A10X ConnectorConec 303WC3SXX99A10X ConnectorConec 131C11029X Male ContactsConec 132C11029X Female ContactsDigi-Key 17E-1657-15-ND Backshells L-Com SDH Standoff
Custom length Cables, backshells and screws on both connectors, standoffs on socket
LIGO-G1000552-V8 21
Custom 24VDC Adapter
P P
+/- 24VDC 10A
+/- 24VDC 10A (Used between Old eLIGO 24VDC Connector and New Standard)
New 24VDC Standard
Old eLIGO 24VDC Standard
This adapter is basically a 4” M-M cable that is intended to be attached to a chassis that does not conform to the new connector standard for 24VDC.
LIGO-G1000552-V8 22
Custom Socket Pigtail CablesUsed between DC Power Supplies and Custom Extension Cables
+/- 12VDC 10A
This is a sample custom socket pigtail for use on a +/-12VDC set of power supplies. It really is a dark green cable jacket.
LIGO-G1000552-V8 23
24VDC Adapter – IO Chassis to Power Strip
New 24VDC Power Supply
24VDC Spades to Plug Adapter
Adapter construction
This adapter is basically a 3.5” +/-24VDC 10A Plug Pigtail that is intended to be attached to the 24VDC input screws of IO Chasses.It will connect to a +/-24VDC Extension Cord.