Upload
others
View
11
Download
0
Embed Size (px)
Citation preview
LEGENDplex™Mul�-Analyte Flow Assay Kit
Enabling Legendary Discovery™
For Accurate Quantification of Multiple Human Th(T helper Cell) Cytokines from Cell Culture Supernatant,
Serum, Plasma and Other Biological Samples
Please read the entire manual before running the assay
BioLegend.com
LEGENDplex™Mul�-Analyte Flow Assay Kit
Enabling Legendary Discovery™
Human Immunoglobulin Isotyping Mix and Match Subpanel
Please read the entire manual before running the assay.
BioLegend.com
For Research Purposes Only. Not for use in diagnostic or therapeutic procedures. Purchase does not include or carry the right to resell or transfer this product either as a stand-alone product or as a component of another product. Any use of this product other than the permitted use without the express written authorization of BioLegend is strictly prohibited.
It is highly recommended that this manual be read in its entirety before using this product. Do not use this kit beyond the expiration date.
biolegend.com1
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Table of Contents Page
Chapter 1: KIT DESCRIPTION..................................................
Introduction……………………………………………..........................
PrincipleoftheAssay……………………....……………....….…......
BeadsUsage...........................................………..……………...
StorageInformation…………………………………….......…..........
MaterialsSupplied………………….....……………….................…
MaterialstobeProvidedbytheEnd-User……...........……...
Precautions.................................……………………................
Chapter 2: ASSAY PREPARATION.............................................
SampleCollectionandHandling…………………………............
ReagentPreparation……..…………………………………...............
StandardPreparation.........................................................
SampleDilution……...........……............................................
Chapter 3: ASSAY PROCEDURE..................................................
PerformingtheAssayUsingaFilterPlate……………….........
PerformingtheAssayUsingaV-bottomPlate………………...
Chapter 4: FLOW CYTOMETER SETUP.........................................
Chapter 5: DATA ACQUISITION AND ANALYSIS.........................
DataAcquisition..................................................................
Data Analysis.....................................................................
Chapter 6: ASSAY CHARACTERIZATION..........................................
RepresentativeStandardCurve……………...………………........
AssaySensitivity...……………………………………………………..…..
Cross-Reactivity……………………………………………………..........
LinearityofDilution………………………………………………..........
Intra-AssayPrecision……………………………………...................
3
3
4
4
6
6
7
9
10
10
10
11
12
13
13
16
19
19
19
20
21
21
21
22
22
23
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
2
Inter-AssayPrecision……………………………………...................
BiologicalSamples…………………………………………….………....
TROUBLESHOOTING.........……………………………………………………....
PLATE MAP...................................................................................
24
24
26
29
biolegend.com3
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Chapter 1: KIT DESCRIPTIONIntroductionImmunoglobulins (Igs) play important roles in immune responses by recogniz-ing,bindingandneutralizingspecificantigenssuchasbacteria,virusesandtox-ins.Inaddition,immunoglobulinscanbindandmarkthepathogens,aprocesscalledopsonizationthroughwhichphagocytescanthenengulfandeliminatethepathogens.Duringimmuneresponses,plasmacellscanswitchfromproduc-ing one immunoglobulin class to another by changing the amino acid sequence intheconstantregionoftheheavychain.Therearefivedifferentclassesofimmunoglobulins: IgM, IgG, IgE, IgA and IgD. IgG and IgA can be further divided intodifferentsubclassesbasedonthedifferenceinthenumberofdisulfidebondsandthelengthandflexibilityofthehingeregion.Perturbationstotheexpression levels of the immunoglobulin classes and subclasses are associated withdifferentdiseasessuchaslympho-proliferationorimmunoglobulin-defi-ciencydisorders.Moreover,quantitationoftheseclassesandsubclassesduringandaftervaccinationcouldprovideveryusefulinformationaboutprimaryandsecondary immune responses to a vaccine.
The LEGENDplexTM Human Immunoglobulin Isotyping Panel is a bead-based multiplexassaypanel,usingfluorescence-encodedbeadssuitableforuseonvariousflowcytometers.Itallowsforsimultaneousquantificationof8humanimmunoglobulins, including IgA, IgD, IgE, IgG1, IgG2, IgG3, IgG4, and IgM. This assaypanelprovideshigherdetectionsensitivityandbroaderdynamicrangethantraditionalELISAmethods.
The LEGENDplexTMHumanImmunoglobulinIsotypingPanelsareconfiguredasshownbelow:
Catalog No.
Plex Size Targets Recommended
Sample TypeRecommended Dilution Factor
740637 740638 8-plex IgG1, IgG2, IgG3, IgE,
IgG4, IgA, IgM, IgD Tissue culture Varies
740639740640 6-plex IgG1, IgG2, IgG3, IgG4,
IgA, IgM Serum, Plasma 100,000
740714740715 4-plex IgG1, IgG2, IgG3, IgG4 Serum, Plasma 100,000
740641740642 1-plex IgE Serum, Plasma 100
740643740644 1-plex IgD Serum, Plasma 5,000
The LEGENDplexTMHumanImmunoglobulinIsotypingPanelisdesignedtoallowflexiblecustomizationwithinthepanel.Pleasevisitwww.biolegend.com/leg-endplexformoreinformationonhowtomixandmatchwithinthepanel.This assay is for research use only.
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
4
Principle of the Assay
BioLegend’s LEGENDplexTM assays are bead-based immunoassays using the samebasicprincipleassandwichimmunoassays.
Beadsaredifferentiatedbysizeandinternalfluorescenceintensities.Eachbeadsetisconjugatedwithaspecificantibodyonitssurfaceandservesasthecap-turebeadsforthatparticularanalyte.Whenaselectedpanelofcapturebeadsismixedandincubatedwithasamplecontainingtargetanalytesspecifictothecaptureantibodies,eachanalytewillbindtoitsspecificcapturebeads.Afterwashing,abiotinylateddetectionantibodycocktailisadded,andeachdetec-tionantibodyinthecocktailwillbindtoitsspecificanalyteboundonthecap-turebeads,thusformingcapturebead-analyte-detectionantibodysandwiches.Streptavidin-phycoerythrin(SA-PE)issubsequentlyadded,whichwillbindtothebiotinylateddetectionantibodies,providingfluorescentsignalintensitiesinproportiontotheamountofboundanalytes.
Sincethebeadsaredifferentiatedbysizeandinternalfluorescenceintensityonaflowcytometer,analyte-specificpopulationscanbesegregatedandPEfluorescentsignalquantified.Theconcentrationofaparticularanalyteisdeter-mined using a standard curve generated in the same assay.
Beads Usage
The LEGENDplexTM bead-based assay systemusestwosetsofbeads.Eachsethasauniquesizethatcanbeidentifiedbasedontheirforwardscatter(FSC)andsidescatter(SSC)profiles(BeadsAandBeadsB,Figure1).Eachbeadsetcanbefurtherresolvedbasedontheirinternalfluorescenceintensities.Theinternaldye can be detected using FL3, FL4, or APC channel, depending on the type of flowcytometerused.ThesmallerBeadsAconsistsof6beadpopulationsandthelargerBeadsBconsistsof7beadpopulations(Figure2-3).
Using8outofthe13beadpopulationsdistinguishedbysizeandinternalfluo-rescent dye, the Human Immunoglobulin IsotypingPanelallowssimultaneousdetectionof8immunoglobulinsinasinglesample.Eachanalyteisassociatedwithaparticularbeadsetasindicated(Figures2-3andTable1). Figure 1. Beads Differentiated by Size
Beads A = smaller beads
Beads B = larger beads
biolegend.com5
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Figure 2. Beads A Classification by FL4
Figure 3. Beads B Classification by FL4
ForBeadsusageinthepanel,pleaserefertoTable1below:
Table 1. Panel Targets and Bead ID
Target Bead IDIsotyping (8-plex)Cat No.
740637 or 740638
Isotyping (6-plex)Cat. No.
740639 or 740640
Isotyping IgGs (4-plex)
Cat. No. 740639 or
740640
Isotyping IgE (1-plex)
Cat. No. 740641 or
740642
Isotyping IgD (1-plex)
Cat. No. 740643 or
740644
Top Standard Concentrations
IgG1 A4 √ √ √ The top standard concentrationof each target may vary and
may subject to change from lot
to lot. Please refer to the lot-specificCertificateof
Analysis for this information
IgG2 A5 √ √ √
IgG3 A6 √ √ √
IgE A7 √ √
IgG4 B3 √ √ √
IgA B4 √ √
IgM B5 √ √
IgD B6 √ √
*BeadIDisusedtoassociateabeadpopulationtoaparticularanalyteintheLEGENDplexTMDataAnalysisSoftware.TheassociationofanalyteandbeadIDwillbedefinedduringthegatingstepofthedataanalysis.
When entering analyte and bead ID infomation during the gating step, always enter in the sequential order of the bead ID (e.g, A4, A5, A6...B2, B3, B4...). Please refer to the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelp for details (www.biolegend.com/legendplex).
A5 A7 A8
A4
A6
A10
B4 B5
B6 B7
B3
B9
B2
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
6
Storage Information
Recommendedstorageforalloriginalkitcomponentsisbetween2°Cand8°C.DONOTFREEZEBeads,DetectionAntibodiesorSA-PE.
• Oncethestandardshavebeenreconstituted,immediatelytransfercon-tentsintopolypropylenevials.DONOTSTORERECONSTITUTEDSTAN-DARDS IN GLASS VIALS.
• Uponreconstitution,leftoverstandardshouldbestoredat≤-70°Cforusewithinonemonth.Avoidmultiple(>2)freeze-thawcycles.Discardanyleftoverdilutedstandards.
Materials Supplied
The LEGENDplexTMkitcontainsreagentsfor100tests,listedinthetablebelow.Whenassayedinduplicate,thisisenoughforan8-pointstandardcurveand40samples.
For the Mix and Match Subpanels, individual beads are provided at 13X con-centration.TheBufferSetcontainsSetupBeads,allBuffers,PlateSealers,SA-PEandDataAnalysisSoftwareDongle.AmanualisalsoprovidedforeachMixandMatch subpanel.
Kit Components Quantity Volume Cat. #
Capture Beads* (see tables below for more information) varies varies varies
LEGENDplex™ Human Immunoglobulin IsotypingPanelDetectionAntibodies 1bottle 3.5 mL 740645
LEGENDplex™ Human Immunoglobulin Isotyping Panel Standard 1 vial lyophi-
lized 740646
LEGENDplex™BufferSetK 1 740713
Filter Plate** or V-bottomPlate*** 1 Plate 740377**or
740379***
Human Immunoglobulin Isotyping Mix and Match Subpanel Manual 1 75464
**Forassaywithfilterplate.***ForassaywithV-bottomplate.
biolegend.com7
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Capture beads for Mix and Match Subpanels*:
Bead Name Quantity Volume Cat. #LEGENDplex™ Human IgG1 Capture Bead A4, 13X 1 vial 270 µL 740647
LEGENDplex™ Human IgG2 Capture Bead A5, 13X 1 vial 270 µL 740648
LEGENDplex™ Human IgG3 Capture Bead A6, 13X 1 vial 270 µL 740649
LEGENDplex™ Human IgE Capture Bead A7, 13X 1 vial 270 µL 740650
LEGENDplex™ Human IgG4 Capture Bead B3, 13X 1 vial 270 µL 740651
LEGENDplex™ Human IgA Capture Bead B4, 13X 1 vial 270 µL 740652
LEGENDplex™ Human IgM Capture Bead B5, 13X 1 vial 270 µL 740653
LEGENDplex™ Human IgD Capture Bead B6, 13X 1 vial 270 µL 740654
* Please refer to Beads ID and Panel-Specific Target Selection table (Table 1), to see which capture beads are included in each panel.
LEGENDplex™ Buffer Set K (Cat#: 740713)
Components Quantity Volume Part #Setup Beads 1: FITC Beads 1 vial 1 mL 77840Setup Beads 2: PE Beads 1 vial 1 mL 77842SetupBeads3:RawBeads 1 vial 2 mL 77844LEGENDplexTM SA-PE 1bottle 3.5 mL 77743LEGENDplexTMAssayBuffer 3bottle 25 mL 77562LEGENDplexTMWashBuffer,20X 1bottle 25 mL 77564DataAnalysisSoftwareDongle 1 21217Plate Sealers 4 sheets 78101
No plate is included in Buffer Set K. Plate need to be ordered separately. Please order the correct type of plate based on the preferred assay protocol (Cat# 740377 or 740378 for Filter Plate and Cat# 740379 for V-bottom Plate)
Materials to be Provided by the End-User
• Aflowcytometerequippedwithtwolasers(e.g.,a488nmbluelaseror532nmgreenlaseranda633-635nmredlaser)capableofdistinguishing575 nm and 660 nm oraflowcytometerequippedwithonelaser(e.g.,488nmbluelaser)capableofdistinguishing575nmand670nm.
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
8
Partial list of compatible flow cytometers:
Flow Cytometer
Reporter Channel
ChannelEmission
Classification Channel
Channel Emission
Compensa-tion needed?
BD FACSCaliburTM
(single laser)FL2 575 nm FL3 670 nm Yes
BD FACSCaliburTM
(dual laser)FL2 575 nm FL4 660 nm No*
BD FACSArrayTM Yellow 575 nm Red 660 nm No*
BD AccuriTM C6 FL2 585 nm FL4 675 nm No*
BD FACSCantoTM
BD FACSCantoTM IIPE 575 nm APC 660 nm No*
BDTM LSR, LSR IIBD LSRFortessaTM PE 575-585
nm APC 660 nm No*
BD FACSAriaTM PE 575 nm APC 660 nm No*
Beckman Coulter-CytoFLEX PE 585 nm APC 660 nm No*
*Compensation is not required for the specified flow cytometers when set up properly.
Forsettingupvariousflowcytometers,pleasevisit:www.biolegend.com/legendplex andclickontheInstrument Setup tab.
• Multichannelpipettescapableofdispensing5μLto200μL
• Reagentreservoirsformultichannelpipette
• Polypropylene microfuge tubes (1.5 mL)
• Laboratory vortex mixer
• Sonicator bath (e.g., Branson Ultrasonic Cleaner model #B200, or equiva-lent)
• Aluminum foil
• Absorbentpadsorpapertowels
• Plateshaker(e.g.,Lab-LineInstrumentsmodel#4625,orequivalent)
• Tabletop centrifuges (e.g., Eppendorf centrifuge 5415 C, or equivalent)
If the assay is performed in a filter plate (recommended);
• Avacuumfiltrationunit(MilliporeMultiScreen®HTSVacuumManifold,cat#MSVMHTS00orequivalent).Instructionsonhowtousethevacuummanifoldcanbefoundatthesupplier’swebsite.
• A vacuum source (mini vacuum pump or line vacuum, e.g., Millipore
biolegend.com9
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
VacuumPump,catalog#WP6111560,orequivalent)
• Ifneeded,additionalFilterplatecanbeorderedfromBioLegend(Cat#740377 or 740378)
If the assay is performed in a V-bottom plate (optional);
• Centrifugewithaswingingbucketadaptorformicrotiterplates(e.g.,Beck-man Coulter AllegraTM6RCentrifugewithMICROPLUSCARRIERadaptorforGH3.8 and JS4.3 Rotors) .
• Ifneeded,additionalV-bottomplatecanbeorderedfromBioLegend(Cat#740379)
Precautions
• All blood components and biological materials should be handled as poten-tiallyhazardous.FollowuniversalprecautionsasestablishedbytheCenterforDiseaseControlandPreventionandbytheOccupationalSafetyandHealthAdministrationwhenhandlinganddisposingofinfectiousagents.
• Sodiumazidehasbeenaddedtosomereagentsasapreservative.Al-thoughtheconcentrationsarelow,sodiumazidemayreactwithleadandcopperplumbingtoformhighlyexplosivemetalazides.Ondisposal,flushwithalargevolumeofwatertopreventazidebuild-up.
• Donotmixorsubstitutereagentsfromdifferentkitsorlots.Reagentsfromdifferentmanufacturersshouldnotbeusedwiththiskit.
• Donotusethiskitbeyonditsexpirationdate.
• SA-PEandBeadsarelight-sensitive.Minimizelightexposure.
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
10
Chapter 2: ASSAY PREPARATION
Sample Collection and HandlingPreparation of Serum Samples:
• Allowthebloodtoclotforatleast30minutesandcentrifugefor20min-utes at 1,000 x g.
• Remove serum and assay immediately or aliquot and store samples at ≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
• Whenusingfrozensamples,itisrecommendedthatsamplesarethawedcompletely,mixedandcentrifugedtoremoveparticulatespriortouse.
Preparation of Plasma Samples:
• Plasmacollectionusingananti-coagulant(e.g.EDTA,Heparin,Citrate)isrecommended. Centrifuge for 20 minutes at 1,000 x gwithin30minutesofbloodcollection.
• Remove plasma and assay immediately, or aliquot and store samples at ≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
• Whenusingfrozensamples,itisrecommendedthatsamplesarethawedcompletely,mixedwellandcentrifugedtoremoveparticulates.
Preparation of Tissue Culture Supernatant:
• Centrifuge the sample to remove debris and assay immediately. If not pos-sible,aliquotandstoresamplesat≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
Reagents PreparationPreparation of Antibody-Immobilized Beads
Theindividualbeads(13X)shouldbemixedwitheachotheranddilutedto1XfinalconcentrationwithAssayBufferpriortouse.Tomixthebeads,followthestepsbelow(a5-plexsubpanelisusedasanexample):
1. Sonicate the beads vials for 1 minute in a sonicator bath and then vortex for 30 seconds to completely resuspend the beads.
2. Calculate the amount of mixed and diluted beads needed for the assay.Prepareextratocompensateforpipettingloss.Eachreactionneeds25µLofmixedanddilutedbeads.For50reactions,prepare1.5mLofmixedbeads.For96reactions,prepare3mLofmixedbeads.
3. Tomake1.5mlof5-plex1Xdilutedbeads,transfer115µLofeachofthe 5 individual beads (13X) to a fresh tube (total bead volume = 575 µL)andadd925µLofAssayBuffertomakethefinalvolumeof1.5mL.
biolegend.com11
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Preparation of Wash Buffer
• Bringthe20XWashBuffertoroomtemperatureandmixtobringallsaltsintosolution.
• Dilute25mLof20XWashBufferwith475mLdeionizedwater.Storeun-usedportionsbetween2°Cand8°Cforuptoonemonth.
Standard Preparation
1. Priortouse,reconstitutethelyophilizedHumanImmunoglobulinIsotyp-ingPanelStandardCocktail,with250µLAssayBuffer.
2. Mixandallowthevialtositatroomtemperaturefor15minutes,andthen transfer the standard to an appropriately labeled polypropylene microfugetube.ThiswillbeusedasthetopstandardC7.
Note: The top standard concentrations of analytes in this panel was set at various concentrations, but may be subject to change from lot to lot (see lot-specific Certificate of Analysis provided with kit box for details).
3. Label 6 polypropylene microfuge tubes as C6, C5, C4, C3, C2 and C1, respectively.
4. Add75µLofAssayBuffertoeachofthesixtubes.Prepare1:4dilutionofthe top standard by transferring 25 µL of the top standard C7 to the C6 tubeandmixwell.ThiswillbetheC6standard.
Inthesamemanner,performserial1:4dilutionstoobtainC5,C4,C3,C2and C1 standards (see the table below using the top standard at 400,000 pg/mL as an example).AssayBufferwillbeusedasthe0pg/mLstandard(C0).
Tube/Standard ID
Serial Dilution
Assay Buffer to add (µL)
Standard to add
Final Conc. (pg/mL)
C7 -- -- -- 400,000
C6 1:4 75 25 µL of C7 100,000
C5 1:16 75 25 µL of C6 25,000
C4 1:64 75 25 µL of C5 6250
C3 1:256 75 25 µL of C4 1562.5
C2 1:1024 75 25 µL of C3 390.6
C1 1:4096 75 25 µL of C2 97.6
C0 -- 75 -- 0
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
12
Sample DilutionFor cell culture supernatant samples, the levels of analyte can vary greatly fromsampletosample.Whilethesamplescanbetestedwithoutdilutions,a preliminary experiment may be required to determine the appropriate dilutionfactor.Ifsampledilutionisdesired,dilutionshouldbedonewithcorrespondingfreshcellculturemediumorAssayBuffertoensureaccuratemeasurement.
Forserumandplasmasamples,followtargetspecificdilutionrecommen-dationsbelow.
Target Recommended Sample Type
Recommended Dilution Factor
IgG1 Serum, Plasma 100,000
IgG2 Serum, Plasma 100,000
IgG3 Serum, Plasma 100,000
IgE Serum, Plasma 100
IgG4 Serum, Plasma 100,000
IgA Serum, Plasma 100,000
IgM Serum, Plasma 100,000
IgD Serum, Plasma 5,000
• Recommended dilution scheme;
Sample 1st Dilution(1:200)
2nd Dilution(1:500)
Final dilution fold
Serum, plasma
2 µL + 398 µL AssayBuffer
2 µL1stDilution+998µLBuffer
100,000
Sample 1st Dilution(1:100)
2nd Dilution(1:50)
Final dilution fold
Serum, plasma
2 µL + 198 µL AssayBuffer
4 µL1stDilution+196µLAssayBuffer
5,000
Sample Dilution(1:100)
Final dilution fold
Serum, plasma
2 µL+198µLAssayBuffer 100
Iffurthersampledilutionisdesired,dilutionshouldbedonewithAssayBuf-fer to ensure accurate measurement.
biolegend.com13
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Chapter 3: ASSAY PROCEDURE
The LEGENDplexTM assaycanbeperformedinafilterplate,orinaV-bottomplate.
• Thein-filterplateassayprocedureisrecommendedduetoitsgoodsample to sample consistency, assay robustness and ease of handling. This procedurerequiresavacuumfiltrationunitforwashing(seeMaterials to be Provided by the End-User, page 7). If you have performed bead-based multiplexassaysbefore,yourlabmayalreadyhavethevacuumfiltrationunit set up.
• Ifthein-filterplateassayprocedureisnotpossibleorifyouprefer,theas-saycanbeperformedinaV-bottomplate.
Performing the Assay Using a Filter Plate
• Allowallreagentstowarmtoroomtemperature(20-25°C)beforeuse.
• Setthefilterplateonaninvertedplatecoveratalltimesduringassaysetupandincubationsteps,sothatthebottomoftheplatedoesnottouchanysurface.Touchingasurfacemaycauseleakage.
• Keeptheplateuprightduringtheentireassayprocedure,includingthewashingsteps,toavoidlosingbeads.
• Theplateshouldbeplacedinthedarkorwrappedwithaluminumfoilforallincubationsteps.
• Standards and samples should be run in duplicate and arranged on the plate in a vertical configuration convenient for data acquisition and analysis (as shown in attached PLATE MAP, page 29). Be sure to load standards in the first two columns. If an automation device is used for reading, the orientation and reading sequence should be carefully planned.
1. Pre-wettheplatebyadding100μLofLEGENDplexTM1XWashBuffertoeachwellandletitsitfor1minuteatroomtemperature.Toremovethe excess volume, place the plate on the vacuum manifold and apply vacuum.Donotexceed10”Hgofvacuum.Vacuumuntilwellsaredrained(5-10seconds).BlotexcessWashBufferfromthebottomoftheplatebypressingtheplateonastackofcleanpapertowels.Placetheplateontopof the inverted plate cover.
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
14
Loadtheplateasshowninthetablebelow(intheorderfromlefttoright):AssayBuffer Standard Sample*
StandardWells 25 µL 25 µL ---
Samplewells 25 µL --- 25 µL *See Sample Dilution
2. Vortexmixedbeadsbottlefor30seconds.Add25μLofmixedbeadstoeachwell.Thevolumeshouldbe75μLineachwellafterbeadsaddition.(Note:Duringadditionofthebeads,shakemixedbeadsbottleintermit-tentlytoavoidbeadsettling).
3. Sealtheplatewithaplatesealer.To avoid plate leaking, do not apply posi-tive pressure to the sealer when sealing the plate.Wraptheentireplate,includingtheinvertedplatecover,withaluminumfoil.Placetheplateonaplateshaker,secureitandshakeatapproximate500rpmfor2hoursatroom temperature.
4. Do not invert the plate! Place the plate on the vacuum manifold and apply vacuumasbeforeinStep1.Add200µLof1XWashBuffertoeachwell.RemoveWashBufferbyvacuumfiltration.BlotexcessWashBufferfromthebottomoftheplatewithanabsorbentpadorpapertowels. Repeat this washingsteponcemore.
5. Add25µLofDetectionAntibodiestoeachwell.
6. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinvertedplatecover,withaluminumfoil.Placetheplateonaplateshakerandshakeatapproximately500rpmfor1houratroomtemperature.
7. Do not vacuum!Add25µLofSA-PEtoeachwelldirectly.
8. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinvertedplatecover,withaluminumfoil.Placetheplateonaplateshakerandshakeatapproximate500rpmfor30minutesatroomtemperature.
9. Repeat step 4 above.
10. Add150µLof1XWashBuffertoeachwell.Resuspendthebeadsonaplateshakerfor1minute.
11. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay (Note: Prolonged sample storage can lead to reduced signal).
Iftheflowcytometerisequippedwithanautosampler,readtheplatedirectly using the autosampler. Please be sure to program the autosampler to resuspend beads in the well immediately before taking samples. The
biolegend.com15
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
probe height may need to be adjusted when using an autosampler.
If an autosampler is not available, the samples can be transferred from the filterplateto micro FACS (or FACS) tubes and read manually.
Assay Procedure Summary for Filter PlateAdd 100 μL 1X Wash Bu�er to �lter plate wells
Vacuum to remove excess bu�er
Incubate 2 hours, RT, shaking
Capture beads
Biotinylated Detection Antibody
Analytes
Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking
Wash 2 times using vacuum �ltration unitAdd 25 μL Detection AntibodiesIncubate 1 hr, RT, shaking
Wash 2 times using vacuum �ltration unitAdd 150 μL of 1x Wash Bu�er Read on a �ow cytometer
Add to the plate:25 μL Assay Bu�er to all wells25 μL diluted standard to standard wells or 25 μL sample to sample wells25 μL pre-mixed beads to all wells
BA
C
A B C
A B C
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
16
Performing the Assay Using a V-bottom Plate
• Allowallreagentstowarmtoroomtemperature(20-25°C)beforeuse.
• Keeptheplateuprightduringtheentireassayprocedure,includingthewashingsteps,toavoidlosingbeads.
• Theplateshouldbeplacedinthedarkorwrappedwithaluminumfoilforallincubationsteps.
• Standards and samples should be run in duplicate and arranged on the plateinaverticalconfigurationconvenientfordataacquisitionandanalysis(asshowninattachedPLATEMAP,page29).Besuretoloadstandardsinthefirsttwocolumns.Ifanautomationdeviceisusedforreading,theori-entationandreadingsequenceshouldbecarefullyplanned.
1. Loadtheplateasshowninthetablebelow(intheorderfromlefttoright):AssayBuffer Standard Sample*
StandardWells 25 µL 25 µL ---
Samplewells 25 µL --- 25 µL
*See Sample Dilution
2. Vortexmixedbeadsfor30seconds.Add25μLofmixedbeadstoeachwell.Thetotalvolumeshouldbe75μLineachwellafterbeadsaddition.(Note:Duringbeadsaddition,shakemixedbeadsbottleintermittentlytoavoidbeadsettling).
3. Sealtheplatewithaplatesealer.Covertheentireplatewithaluminumfoil to protect the plate from light. Shakeat800rpmonaplateshakerfor2 hours at room temperature (Depending on the shaker, the speed may need to be adjusted. The optimal speed is one that is high enough to keep beads in suspension during incubation, but not too high so it causes spill from the wells).
4. Centrifugetheplateat1050rpm(~250g)for5minutes,usingaswingingbucketrotor(G.H3.8)withmicroplateadaptor(Please refer to Materials to be Provided by the End-User, page 7). Donotuseexcessivecentrifugationspeedasitmaymakeithardertoresuspendbeadsinlatersteps.Make sure the timer of the centrifuge works properly and standby to make sure the centrifuge reaches preset speed.
5. Immediatelyaftercentrifugation,dumpthesupernatantintoasinkbyquicklyinvertingandflickingtheplatein one continuous and forceful mo-tion.Donotworryaboutlosingbeadsevenifthepelletisnotvisible.Thebeadswillstayinthetipofthewellnicely.
biolegend.com17
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
6. Blottheplateonastackofcleanpapertowelanddraintheremainingliquidfromthewellasmuchaspossible.Becarefulnottodisturbthebeadpellet.
Alternatively,removalofthesupernatantmaybecompletedusingamultichannelpipettesetat75µL. Try to remove as much liquid as possible withoutremovinganybeads. Be sure to changepipettetipsbetweeneachroworcolumn.
7. Washtheplatebydispensing200μLof1XWashBufferintoeachwellandincubate for one minute. Repeatstep4and5above.Asecondwashisoptional,butmayhelpreducebackground.
8. Add25µLofDetectionAntibodiestoeachwell.
9. Sealtheplatewithanewplatesealer.Covertheentireplatewithalumi-num foil to protect the plate from light. Shakeat800rpmonaplateshakerfor 1 hour at room temperature.
10. Do not wash the plate!Add25µLofSA-PEtoeachwelldirectly.
11. Sealtheplatewithanewplatesealer.Wraptheentireplatewithaluminumfoilandshaketheplateonaplateshakeratapproximate800rpmfor30minutes at room temperature.
12. Repeat step 4 and 5.
13. Washtheplatebydispensing200μLof1XWashBufferintoeachwellandincubate for one minute. Repeatstep4and5above.Thiswashingstepisoptionalbutithelpstoreducethebackground.
14. Add150µLof1XWashBuffertoeachwell.Resuspendthebeadsbypipet-ting.
15. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay (Note: Prolonged sample storage can lead to reduced signal).
Iftheflowcytometerisequippedwithanautosampler,thesamplescanberead directly. Please be sure to program the autosampler to resuspend beads in the well immediately before taking samples. The probe height may need to be adjusted when using an autosampler.
If an autosampler is not available, the samples can be transferred from the plate to micro FACS (or FACS) tubes and read manually.
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
18
Assay Procedure Summary for V-bottom Plate
Incubate 2 hours, RT, shaking
Capture beads
Biotinylated Detection Antibody
Analytes
Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking
Spin down beads, remove supernatant Wash 1 time Add 25 μL Detection AntibodiesIncubate 1 hr, RT, shaking
Spin down beads, remove supernatant Wash 1 time (optional)Add 150 µL of 1x Wash Bu�er Read on a �ow cytometer
Add to the plate:25 μL Assay Bu�er to sample wells25 μL diluted standard to standard wells or 25 μL sample to sample wells25 μL mixed beads to all wells
BA
C
A B C
A B C
biolegend.com19
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Chapter 4: FLOW CYTOMETER SETUP
Inordertogeneratereliabledata,theflowcytometermustbesetupproperlybeforedataacquisition.
Thesetupinstructionshavebeenremovedfromthismanualanduploadedontoourwebsitetosavepaper.
Toaccessthesetupinstructions,pleasevisit:www.biolegend.com/legendplex andclickontheInstrument Setup tab.
Chapter 5: DATA ACQUISITION AND ANALYSIS
Data Acquisition
1. Beforereadingsamples,makesurethattheflowcytometerissetupprop-erly.
2. Createanewtemplateoropenanexistingtemplate(fordetailsonhowtocreateacytometer-specifictemplate,pleaserefertotheFlowCytometerSetup Guide).
3. Vortex each sample for 5 seconds before analysis.
4. Settheflowratetolow.Setthenumberofbeadstobeacquiredtoabout300 per analyte (e.g., acquire 2,400 beads for a 8-plex assay or 4000 beads for a 13-plex assay). Do not set to acquire total events as samples may contain large amounts of debris. Instead, create a large gate to include both Beads A and Beads B (gate A+B) and set to acquire the number of events in gateA+B.Thiswillexludemajorityofthedebris.
Note:Donotacquiretoofewortoomanybeads.ToofewbeadsacquiredmayresultinhighCVsandtoomanybeadsacquiredmayresultinslowdata analysis later.
5. Read samples.
Whenreadingsamples,settheflowcytometertosetupmodefirstandwaituntilbeadpopulationisstabilizedbeforerecordingorswitchingtoacquisi-tionmode.
To simplify data analysis using the LEGENDplexTMDataAnalysisSoftware,readsamplesinthesameorderasshownonthePLATEMAPattachedatthe end of the manual. For an in-plate assay, read column by column (A1, B1, C1...A2, B2, C2...).
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
20
Whennamingdatafiles,trytousesimplenameswithaconsecutivenum-bering for easy data analysis (e.g. for standards, C0.001, C0.002, C1.003, C1.004, C2.005, C2.006, C3.007, C3.008, ... C7.015, C7.016; for samples, S1.017,S1.018,S2.019,S2.020,S3.021,S3.022…)
StoreallFCSfilesinthesamefolderforeachassay.Ifrunningmultipleas-says, create a separate folder for each assay.
6. Proceed to data analysis using LEGENDplexTMDataAnalysisSoftwarewhendataacquisitioniscompleted.
Data Analysis
• TheFCSfilegeneratedonaflowcytometershouldbeanalyzedusingBioLegend’s LEGENDplexTMDataAnalysisSoftware.TheLEGENDplexTM Data AnalysisSoftwarecanbedownloadedforfreehere:www.biolegend.com/legendplex.
• ForPCusers,installthesoftwareonaPCrunningWindows7orWindows8anduseitinconjunctionwiththeDataAnalysisSoftwareDongleincludedinthiskit.Thedonglehasalicensekeystoredinitandisneededtorunthesoftware.Tousethedongle,simplyplugitintheUSBportofthecomputeronwhichthedataanalysissoftwareisinstalled,priortolaunchingthesoftware.
• ForMacusers,installonaMacrunningMacOSXversion10.7(Lion)orlaterandyouwillbepromotedtorequestasoftwarelicensekeyafterthesoftwareinstallation.
• FollowtheLEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelptousethesoftware(www.bioLegend.com/legendplex; or press F1 for online help at any step of the data analysis).
biolegend.com21
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Chapter 6: ASSAY CHARACTERIZATION
Representative Standard Curve ThisstandardcurvewasgeneratedusingtheLEGENDplexTM Human Immu-
noglobulinIsotypingPanelfordemonstrationpurposesonly.Astandardcurvemustberunwitheachassay.
Assay SensitivityTheminimumdetectableconcentration(MDC)isthetheoreticallimitofdetectioncalculatedusingtheLEGENDplexTMDataAnalysisSoftwarebyapplyinga5-parametercurvefittingalgorithm.Assaysensitivitypresentedhere is MDC + 2 STDEV.
AnalyteSensitivity
(pg/mL)(N=6)
Human IgG1 21.36 + 39.70
Human IgG2 468.22 + 33.65
Human IgG3 2.28 + 2.97
Human IgE 19.03 + 51.52
Human IgG4 6.32 + 1.47
Human IgA 33.10 + 28.82
Human IgM 177.33 + 36.53
Human IgD 53.93 + 4.89
1
10
100
1000
10000
1 10 100 1000 10000 100000 1000000 10000000
MFI
Concentration (pg/mL)
A4.IgG1
A5.IgG2
A6.IgG3
A7.IgE
B3.IgG4
B4.IgA
B5.IgM
B6.IgD
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
22
Cross-ReactivityCrossreactivitywastestedinthepanelusingtheLEGENDplexTM Human ImmunoglobulinIsotypingPanel.Thefollowinghumanimmunoglobulinsweretestedattherespectiveconcentrationasshowninthetablebelow.IgG3crossesupto3%withIgG1assay.Noornegligiblecross-reactivitywasfoundforalltheothertestedanalytes.
Human
Target IgG1 IgG2 IgG3 IgE IgG4 IgA IgM IgD
Concentration (ng/mL)
400 3200 100 800 50 400 1600 400
Mouse
Target IgG1 IgG2a IgG2b IgG3 IgA IgM IgE
Concentration (ng/mL)
500 500 500 500 500 500 500
Linearity of Dilution
Fortestinglinearityofdilution,serumandplasmasampleswerefirstdilutedtotheirrespectivedilutionfactorsasmentionedinSampleDilu-tion,thenseriallydiluted1:2,1:4,1:8withAssayBufferandassayed.Themeasuredconcentrationsofseriallydilutedsampleswerethencomparedwiththatoftheinitiallydilutedsamples.
Analyte Linearity of Dilution (Serum)
Linearity of Dilution (Plasma)
Human IgG1 101% 99%Human IgG2 66% 69%Human IgG3 89% 91%
Human IgE 95% 104%Human IgG4 102% 96%Human IgA 95% 96%Human IgM 98% 98%Human IgD 135% 108%
biolegend.com23
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Intra-Assay Precision
Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedinoneassaywith16replicatesforeachsample.Theintra-assayprecisionwascalculatedasbelow.
Analyte Sample Mean (pg/mL) STDEV %CV
Human IgG1Sample 1 2020.94 71.32 4%Sample 2 7792.56 162.44 2%
Human IgG2Sample 1 14394.23 363.14 3%Sample 2 55494.89 764.85 1%
Human IgG3Sample 1 113.57 3.62 3%Sample 2 447.21 8.29 2%
Human IgESample 1 4190.66 100.12 2%Sample 2 15597.76 285.13 2%
Human IgG4Sample 1 227.49 6.42 3%Sample 2 886.90 16.59 2%
Human IgASample 1 1797.31 62.38 3%Sample 2 6983.05 172.75 2%
Human IgMSample 1 7089.48 220.19 3%Sample 2 26832.52 600.57 2%
Human IgDSample 1 1964.74 44.88 2%Sample 2 7456.15 210.68 3%
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
24
Inter-Assay Precision
Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedinfourindependentassayswith4replicatesforeachsample.Theinter-assayprecisionwascalculatedasbelow.
Analyte Sample Mean (pg/mL) STDEV %CV
Human IgG1Sample 1 1946.37 146.36 8%Sample 2 7423.18 1006.81 14%
Human IgG2Sample 1 15042.84 2441.91 16%Sample 2 55360.31 5595.00 10%
Human IgG3Sample 1 127.44 14.41 11%Sample 2 450.99 51.87 12%
Human IgESample 1 4673.86 578.75 12%Sample 2 15875.11 1572.28 10%
Human IgG4Sample 1 267.82 20.92 8%Sample 2 927.87 89.81 10%
Human IgASample 1 2152.27 151.25 7%Sample 2 6888.56 747.07 11%
Human IgMSample 1 8269.99 694.93 8%Sample 2 27319.01 2690.52 10%
Human IgDSample 1 2149.19 225.79 11%Sample 2 7260.50 763.26 11%
Biological Samples
Serum and Plasma (Samples are paired)
Normalhumanserumsamples(n=20)weretestedforendogenouslevelsoftheimmunoglobulinisotypes.Theconcentrationsmeasuredareshownbelow.
biolegend.com25
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
AnalyteRange
(µg/mL)% of
DetectableMedian (µg/mL)
Human IgG1 3239.07 - 20869.79 100% 8006.32
Human IgG2 272.80 - 2086.20 100% 873.30
Human IgG3 140.60 - 1218.38 100% 595.91
Human IgE 0.27 - 7.67 100% 0.73
Human IgG4 45.17 - 3774.34 100% 366.14
Human IgA 1836.76 - 12384.43 100% 4361.83
Human IgM 449.76 - 4175.35 100% 1293.44
Human IgD 0.64 - 37.08 100% 13.59
Normalhumanplasmasamples(n=20)weretestedforendogenouslevelsoftheimmunoglobulinisotypes.Theconcentrationsmeasuredareshownbelow.
AnalyteRange
(µg/mL)% of
DetectableMedian (µg/mL)
Human IgG1 2097.85 - 18171.98 100% 4862.87
Human IgG2 122.26 - 1905.86 100% 579.21
Human IgG3 61.62 - 1281.92 100% 504.39
Human IgE 0.09 - 5.71 100% 0.73
Human IgG4 3.83 - 2805.58 100% 233.07
Human IgA 914.15 - 8485.23 100% 3268.81
Human IgM 313.30 - 2250.03 100% 1088.25
Human IgD 0.57 - 55.41 100% 12.37
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
26
TROUBLESHOOTING
Problem Possible Cause Solution
Bead popula-tionshiftingupwardordownwarddur-ingacquisition
The strong PE signal from high concentra-tionsamplesorstan-dards may spill over to classificationChannel(e.g.,FL3/FL4/APC)and mess up the bead separation.
OptimizeinstrumentsettingsusingKitSetupBeads,andmakeappropriatecom-pensationbetweenchannels.
Filterplatewillnot vacuum orsomewellsclogged
Vacuum pressure is insufficientorvacuummanifold does not seal properly.
Increase vacuum pressure such that 0.2 mLbuffercanbesuctionedin3-5seconds.Cleanthevacuummanifoldandmakesurenodebrisonthemanifold.Pressdowntheplateonthemanifoldtomakeagoodseal.
Samples have insoluble particlesorsampleistoo viscous (e.g., serum and plasma samples)
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Ifsomewellsarestillcloggedduringwash-ing,trythefollowing:
1).Addbuffertoallthewells,pipetteupanddownthecloggedwellsandvacuumagain.
2).Useapieceofcleanwipe,wipetheun-dersideofthecloggedwellsandvacuumagain.
3).Takeathinneedle(e.g.,insulinneedle),whileholdingtheplateupward,pokethelittleholeundereachofthecloggedwellsandvacuumagain.Donotpoketoohardortoodeepasitmaydamagethefilterandcauseleaking.
Filterplatewasusedwithoutpre-wet.
Pre-wetplatewithwashbufferbeforerun-ning the assay.
biolegend.com27
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
Insufficientbead count or slowreading
Beads inappropriately prepared
Sonicate bead vials and vortex just prior toaddition.Agitatemixedbeadsintermit-tentlyinreservoirwhilepipettingthisintothe plate.
Samples cause beads aggregationduetoparticulatematterorviscosity.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Beadswerelostduringwashingforin-tubeassay
Makesurebeadsarespundownbyvisu-allycheckthepellet(beadsareinlightblueorbluecolor).Beverycarefulwhenremovingsupernatantduringwashing.
Probe might be par-tiallyclogged.
Sample probe may need to be cleaned, or if needed, probe should be removed and sonicated.
Plateleaked
Vacuum pressure set too high
Adjust vacuum pressure such that 0.2 mL buffercanbesuctionedin3-5seconds.Donot exceed 10” Hg of vacuum.
Plate set directly on tableorabsorbenttow-elsduringincubationsorreagentadditions
Set plate on plate holder or raised edge sobottomoffilterisnottouchinganysurface.
Liquid present on the under side of the plate aftervacuum
Afterwashing,pressdownplatefirmlyonastackofcleanpapertowelstodrytheunderside of the plate.
Pipettetouchinganddamagedplatefilterduringadditions.
Pipettetothesideofwells.
HighBack-ground
Backgroundwellswerecontaminated
Avoidcross-wellcontaminationbychang-ingtipsbetweenpipettingwhenperform-ingtheassayusingamultichannelpipette.
InsufficientwashesThebackgroundmaybeduetonon-specificbindingofSA-PE.Increasenumberofwashes.
Debris(FSC/SSC) during sample acquisi-tion
Debris or platelet may exist in sample solu-tion.
Centrifuge samples before analyzing samples. Remove platelet as much as possible.
Tel: 858-768-5800
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
28
Variationbe-tweenduplicate samples
Beadsaggregation Sonicate and vortex the Beads prior to use.
Multichannelpipettemay not be calibrated or inconsistent pipet-ting
CalibratePipette.Ensuregoodpipettingpractice.Primepipettebeforeusemayhelp.
Platewashingwasnotuniform
Makesureallreagentsarevacuumedoutcompletelyinallwashsteps.
Samples may contain particulatematters.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Loworpoorstandard curve signal
Thestandardwasin-correctlyreconstituted,stored or diluted
Followtheprotocoltoreconstitute,storeanddilutestandard.Doublecheckyourcalculation.
Wrongorshortincuba-tiontime
Ensurethetimeofallincubationswasappropriate.
Signals too high, standard curves satu-rated
PMTvalueforFL2/PEset too high
MakesurethePMTsettingforthere-porter channel is appropriate
Plateincubationtimewastoolong Useshorterincubationtime.
Sample read-ings are out of range
Samples contain no or belowdetectablelevelsof analyte
Makesuretheexperimenttogeneratethesamplesworked.Useproperpositivecontrols.
Samplesconcentrationshigher than highest standard point.
Dilute samples and analyze again.
Standardcurvewassaturated at higher end of curve.
MakesurethePMTsettingforthere-porter channel is appropriate. Use shorter incubationtimeifincubationtimewastoolong
Missed beads populationsduring reading, ordistributionis unequal
Sample may cause some beads to ag-gregate.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Beadspopulationsarenot mixed properly
Makesureallbeadpopulationsaremixed.and in similar numbers.
biolegend.com29
LEGENDplexTM Human Immunoglobulin Isotyping Mix and Match Subpanel
PLA
TE M
AP
(for
in-p
late
ass
ay)
1
2 3
4 5
6 7
8 9
10
11
12
A
C0
C4
Sa
mpl
e1
Sa
mpl
e5
Sa
mpl
e9
Sa
mpl
e 13
Sa
mpl
e17
Sa
mpl
e21
Sa
mpl
e25
Sa
mpl
e29
Sa
mpl
e33
Sa
mpl
e 37
B
C0
C4
Sa
mpl
e1
Sa
mpl
e5
Sa
mpl
e9
Sa
mpl
e 13
Sa
mpl
e17
Sa
mpl
e21
Sa
mpl
e25
Sa
mpl
e29
Sa
mpl
e33
Sa
mpl
e37
C
C1
C5
Sa
mpl
e2
Sa
mpl
e6
Sa
mpl
e10
Sa
mpl
e 14
Sa
mpl
e18
Sa
mpl
e22
Sa
mpl
e26
Sa
mpl
e30
Sa
mpl
e34
Sa
mpl
e38
D
C1
C5
Sa
mpl
e2
Sa
mpl
e6
Sa
mpl
e10
Sa
mpl
e 14
Sa
mpl
e18
Sa
mpl
e22
Sa
mpl
e26
Sa
mpl
e30
Sa
mpl
e34
Sa
mpl
e38
E
C2
C6
Sa
mpl
e3
Sa
mpl
e7
Sa
mpl
e11
Sa
mpl
e 15
Sa
mpl
e19
Sa
mpl
e23
Sa
mpl
e27
Sa
mpl
e31
Sa
mpl
e35
Sa
mpl
e39
F
C2
C6
Sa
mpl
e3
Sa
mpl
e7
Sa
mpl
e11
Sa
mpl
e 15
Sa
mpl
e19
Sa
mpl
e23
Sa
mpl
e27
Sa
mpl
e31
Sa
mpl
e35
Sa
mpl
e39
G
C3
C7
Sa
mpl
e4
Sa
mpl
e8
Sa
mpl
e12
Sa
mpl
e 16
Sa
mpl
e20
Sa
mpl
e24
Sa
mpl
e28
Sa
mpl
e32
Sa
mpl
e36
Sa
mpl
e40
H
C3
C7
Sa
mpl
e4
Sa
mpl
e8
Sa
mpl
e12
Sa
mpl
e 16
Sa
mpl
e20
Sa
mpl
e24
Sa
mpl
e28
Sa
mpl
e32
Sa
mpl
e36
Sa
mpl
e40
LEGENDplex™ Kits are manufactured by BioLegend 9727 Pacific Heights Blvd.San Diego, CA 92121Tel: 1.858.768.5800Tel: US & Canada Toll-Free: 1.877.Bio-Legend (1.877.246.5343)Fax: 1.877.455.9587Email: [email protected]
For a complete list of world-wide BioLegend offices and distributors, please visit our website at: biolegend.com
Enabling Legendary Discovery™
75464_V01