63
Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure and slide acknowledgements: Enrico Gratton

Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Embed Size (px)

Citation preview

Page 1: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Lecture 7FCS, Autocorrelation, PCH,

Cross-correlation

Joachim Mueller

Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure and slide acknowledgements:

Enrico Gratton

Page 2: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Fluorescence Parameters & Methods

1. Excitation & Emission Spectra• Local environment polarity, fluorophore concentration

2. Anisotropy & Polarization• Rotational diffusion

3. Quenching• Solvent accessibility• Character of the local environment

4. Fluorescence Lifetime• Dynamic processes (nanosecond timescale)

5. Resonance Energy Transfer• Probe-to-probe distance measurements

6. Fluorescence microscopy• localization

7. Fluorescence Correlation Spectroscopy• Translational & rotational diffusion • Concentration• Dynamics

Page 3: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Historic Experiment: 1st Application of Correlation Spectroscopy(Svedberg & Inouye, 1911) Occupancy Fluctuation

1200020013241231021111311251110233133322111224221226122142345241141311423100100421123123201111000111_211001320000010011000100023221002110000201001_333122000231221024011102_1222112231000110331110210110010103011312121010121111211_10003221012302012121321110110023312242110001203010100221734410101002112211444421211440132123314313011222123310121111222412231113322132110000410432012120011322231200_253212033233111100210022013011321131200101314322112211223234422230321421532200202142123232043112312003314223452134110412322220221

Gold particles

time

Svedberg and Inouye, Zeitschr. F. physik. Chemie 1911, 77:145

Statistical analysis of raw data required

Collected data by counting (by visual inspection) the number of particles in the observation volume as a function of time using a “ultra microscope”

Page 4: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

7

6

5

4

3

2

1

0

Part

icle

Nu

mb

er

5004003002001000

time (s)

• Autocorrelation not

available in the

original paper. It can

be easily calculated

today.

Particle Correlation

0 5 10 150.0

0.2

0.4

0.6

0 2 4 6 8100

101

102

0 200 400 600 800

0

4

C

G ()

(sec)

experiment predicted Poisson

fre

qu

en

cy

number of particles

A

nu

mb

er

of

mo

lecu

les

time (sec)*Histogram of particle counts *Autocorrelation

• Poisson

statistics

1.55N

1

1.560

NG

Page 5: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Historical Science Investigator2

Expected

μm70

6 sBk T

DR

(Stokes-Einstein)

Svedberg claimed: Gold colloids with radius R = 3 nm

2 2 Dx D

Slit

2μm

0 5 10 150.0

0.2

0.4

0.6

G()

(sec)

characteristic diffusion time

1.5sD

Experimental facts:2μm

1sExpD

200nmR

Conclusion: Bad sample preparation

The ultramicroscope was invented in 1903 (Siedentopf and Zsigmondy). They already concluded that scattering will not be suitable to observe single molecules, but fluorescence could.

Page 6: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

In FCS Fluctuations are in the Fluorescence Signal

Diffusion

Enzymatic Activity

Phase Fluctuations

Conformational Dynamics

Rotational Motion

Protein Folding

Example of processes that could generate fluctuations

Page 7: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Generating Fluctuations By Motion

Sample Space

Observation Volume

1. The Rate of Motion

2. The Concentration of Particles

3. Changes in the Particle Fluorescence while under Observation, for example conformational transitions

What is Observed?

Page 8: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Defining Our Observation Volume:One- & Two-Photon Excitation.

1 - Photon 2 - Photon

Approximately 1 um3

Defined by the wavelength and numerical aperture of the

objective

Defined by the pinhole size, wavelength, magnification and

numerical aperture of the objective

Page 9: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Brad Amos MRC, Cambridge, UK

1-photon

2-photonNeed a pinhole to define a small volume

Page 10: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Data Treatment & Analysis

0

10

20

30

40

50

0 20 40 60 80 100

Time

Co

un

ts

Time Histogram

0

0.005

0.01

0.015

0.02

0.025

0.03

0.035

0.04

0.01 0.10 1.00 10.00 100.00

Time (ms)

Au

to C

orr

ela

tio

n Fit

Data

Autocorrelation

1

10

100

1000

10000

100000

1000000

0 5 10 15

Counts per Bin

Nu

mb

er

of

Oc

cu

ran

ce

s

Photon Counting Histogram (PCH)

Autocorrelation Parameters: G(0) & kaction

PCH Parameters: <N> &

Page 11: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Autocorrelation Function

)()()( tFtFtF

),()()( tCWdQtF rrr

G() F(t)F(t )

F(t)2

Q = quantum yield and detector sensitivity (how bright is our

probe). This term could contain the fluctuation of the

fluorescence intensity due to internal processes

W(r) describes our observation volume

C(r,t) is a function of the fluorophore concentration over time. This is the term that contains the “physics” of the diffusion processes

Factors influencing the fluorescence signal:

Page 12: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

2

( ) ( )( )

F t F tG

F

Calculating the Autocorrelation Function

F(t) in photon counts

26x103

24

22

20

18

16

14

12

Flu

orescence

35302520151050Time

t

t +

time

Average Fluorescence

F

Fluorescence Fluctuation

( ) ( )dF t F t F

Page 13: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

t1

t2

t3

t4

t5

0 5 10 15 20 25 30 3518.8

19.0

19.2

19.4

19.6

19.8

Time (s)

De

tect

ed

In

ten

sity

(kc

ps)

The Autocorrelation Function

G() F(t)F(t )

F(t)2

G(0) 1/NAs time (tau) approaches 0

Diffusion

Page 14: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

The Effects of Particle Concentration on theAutocorrelation Curve

<N> = 4

<N> = 2

0.5

0.4

0.3

0.2

0.1

0.0

G(t

)

10-7

10-6

10-5

10-4

10-3

Time (s)

Page 15: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Why Is G(0) Proportional to 1/Particle Number?

NN

VarianceG

1)0( 2

A Poisson distribution describes the statistics of particle occupancy fluctuations. In a Poissonian system the variance is proportional to the average number of fluctuating species:

VarianceNumberParticle _

2

2

2

2

)(

)()(

)(

)()0(

tF

tFtF

tF

tFG

2)(

)()()(

tF

tFtFG

Page 16: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

G(0), Particle Brightness and Poisson Statistics

1 0 0 0 0 0 0 0 0 2 0 1 1 1 0 0 0 0 0 0 1 0 0 0 0 0 0 0 1 0 1 0 0 0 1 0 0 1 0 0

Average = 0.275 Variance = 0.256

Variance = 4.09

4 0 0 0 0 0 0 0 0 8 0 4 4 4 0 0 0 0 0 0 4 0 0 0 0 0 0 0 4 0 4 0 0 0 4 0 0 4 0 0

Average = 1.1 0.296

296.0256.0275.0 2

2 VarianceAverageN

Lets increase the particle brightness by 4x:

Time

N

Page 17: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

What about the excitation (or observation) volume shape?

Page 18: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Effect of Shape onthe (Two-Photon) Autocorrelation Functions:

For a 2-dimensional Gaussian excitation volume:

G() N

1 8Dw

2 DG

2

1

For a 3-dimensional Gaussian excitation volume:

G() N

1 8Dw

3 DG

2

1

1 8Dz

3 DG

2

12

1-photon equation contains a 4, instead of 8

Page 19: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Additional Equations:

... where N is the average particle number, D is the diffusion time (related to D, D=w2/8D, for two photon and D=w2/4D for 1-photon excitation), and S is a shape parameter, equivalent to w/z in the previous equations.

3D Gaussian Confocor analysis:

Triplet state term:

)1

1( TeT

T

..where T is the triplet state amplitude and T is the triplet lifetime.

G() 11N

1

D

1

1 S 2

D

1

2

2

( ) ( )( )

F t F tG

F

Note: The offset of one is caused by a different definition of G() :

Page 20: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Orders of magnitude (for 1 μM solution, small molecule, water)

Volume Device Size(μm) MoleculesTimemilliliter cuvette 10000 6x1014 104

microliter plate well 1000 6x1011 102

nanoliter microfabrication 100 6x108 1picoliter typical cell 10 6x105 10-2

femtoliter confocal volume 1 6x102 10-4

attoliter nanofabrication 0.1 6x10-1 10-

6

Page 21: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

The Effects of Particle Size on theAutocorrelation Curve

300 um2/s90 um2/s71 um2/s

Diffusion Constants

Fast Diffusion

Slow Diffusion

0.25

0.20

0.15

0.10

0.05

0.00

G(t

)

10-7

10-6

10-5

10-4

10-3

Time (s)

Dk T

6r

Stokes-Einstein Equation:

3rVolumeMW

and

Monomer --> Dimer Only a change in D by a factor of 21/3, or 1.26

Page 22: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

FCS inside living cells

1E-5 1E-4 1E-3 0.01 0.10.0

0.2

0.4

0.6

0.8

1.0

in solution

inside nucleus

g()

(sec)

Dsolution

Dnucleus

= 3.3

Correlation Analysis

Measure the diffusion coefficient of Green Fluorescent Protein (GFP) in aqueous solution in inside the nucleus of a cell.

objectiveCoverslip

Two-PhotonSpot

Page 23: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Autocorrelation Adenylate Kinase -EGFP Chimeric Protein in HeLa Cells

Qiao Qiao Ruan, Y. Chen, M. Glaser & W. Mantulin Dept. Biochem & Dept Physics- LFD Univ Il, USA

Examples of different Hela cells transfected with AK1-EGFP

Examples of different Hela cells transfected with AK1 -EGFP

Flu

orescence In

tensity

Page 24: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Time (s)

G(

)

EGFPsolution

EGFPcell

EGFP-AK in the cytosol

EGFP-AK in the cytosol

Normalized autocorrelation curve of EGFP in solution (•), EGFP in the cell (• ), AK1-EGFP in the cell(•), AK1-EGFP in the cytoplasm of the cell(•).

Autocorrelation of EGFP & Adenylate Kinase -EGFP

Page 25: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Autocorrelation of Adenylate Kinase –EGFPon the Membrane

A mixture of AK1b-EGFP in the cytoplasm and membrane of the cell.

Clearly more than one diffusion time

Page 26: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Diffusion constants (um2/s) of AK EGFP-AK in the cytosol -EGFP in the cell (HeLa). At the membrane, a dual diffusion rate is calculated from FCS data.

Away from the plasma membrane, single diffusion constants are found.

10 & 0.1816.69.619.68

10.137.1

11.589.549.12

Plasma MembraneCytosol

Autocorrelation Adenylate Kinase -EGFP

13/0.127.97.98.88.2

11.414.4

1212.311.2

D D

Page 27: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Multiple Species

Case 1: Species vary by a difference in diffusion constant, D.

Autocorrelation function can be used:

(2D-Gaussian Shape)G()sample fi

2 G(0) i 1 8Dw

2DG

2

1

i1

M

G(0)sample fi

2 G(0)i

G(0)sample is no longer /N !

! fi is the fractional fluorescence intensity of species i.

Page 28: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Antibody - Hapten Interactions

Mouse IgG: The two heavy chains are shown in yellow and light blue. The two light chains are shown in green and dark blue..J.Harris, S.B.Larson, K.W.Hasel, A.McPherson, "Refined structure of an intact IgG2a monoclonal

antibody", Biochemistry 36: 1581, (1997).

Digoxin: a cardiac glycoside used to treat congestive heart failure. Digoxin competes with potassium for a binding site on an enzyme, referred to as potassium-ATPase. Digoxin inhibits the Na-K ATPase pump in the myocardial cell membrane.

Binding siteBinding site

carb2

Page 29: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

120

100

80

60

40

20

0

Fra

cti

on

Lig

an

d B

ou

nd

10-10

10-9

10-8

10-7

10-6

[Antibody]free (M)

Digoxin-Fl•IgG(99% bound)

Digoxin-Fl

Digoxin-Fl•IgG (50% Bound)

Autocorrelation curves:

Anti-Digoxin Antibody (IgG)Binding to Digoxin-Fluorescein

Binding titration from the autocorrelation analyses:

triplet state

Fb mS

free

Kd S

free

c

Kd=12 nM

S. Tetin, K. Swift, & , E, Matayoshi , 2003

Page 30: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Two Binding Site Model

1.20

1.15

1.10

1.05

1.00

0.95G

(0)

0.001 0.01 0.1 1 10 100 1000Binding sites

IgG + 2 Ligand-Fl IgG•Ligand-Fl + Ligand-Fl IgG•2Ligand-Fl

[Ligand]=1, G(0)=1/N, Kd=1.0

IgG•2Ligand

IgG•Ligand

No quenching

50% quenching

1.0

0.8

0.6

0.4

0.2

0.0

Fra

cti

on

Bo

un

d

0.001 0.01 0.1 1 10 100 1000Binding sites

Kd

Page 31: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Digoxin-FL Binding to IgG: G(0) Profile

Y. Chen , Ph.D. Dissertation; Chen et. al., Biophys. J (2000) 79: 1074

Page 32: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Case 2: Species vary by a difference in brightness

The autocorrelation function is not suitable for analysis of this kind of data without additional information.

We need a different type of analysis

21 DD assuming that

The quantity G(0) becomes the only parameter to distinguish species, but we know that:

G(0)sample fi

2 G(0)i

Page 33: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Photon Counting Histogram (PCH)

Aim: To resolve species from differences in their molecular brightness

Single Species: p(k) PCH(, N )

Sources of Non-Poissonian Noise• Detector Noise• Diffusing Particles in an Inhomogeneous Excitation Beam*• Particle Number Fluctuations*• Multiple Species*

where p(k) is the probability of observing k photon counts

Molecular brightness ε : The average photon count rate of a single fluorophore

PCH: probability distribution function p(k)

16000cpsm

N 0.3

Note: PCH is Non-Poissonian!

Page 34: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Photon Counts

freq

uenc

y

PCH Example: Differences in Brightness

n=1.0) n=2.2) n=3.7)

Increasing Brightness

Page 35: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Single Species PCH: Concentration

5.5 nM Fluorescein

Fit: = 16,000 cpsmN = 0.3

550 nM Fluorescein

Fit: = 16,000 cpsmN = 33

As particle concentration increases the PCH approaches a Poisson distribution

Page 36: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Photon Counting Histogram: Multispecies

Snapshots of the excitation volume

Time

Inte

nsit

y

Binary Mixture: p(k) PCH(1 , N1 ) PCH(2 , N2 )

Molecular Brightness

Concentration

Page 37: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Photon Counting Histogram: Multispecies

Sample 2: many but dim (23 nM fluorescein at pH 6.3)

Sample 1: fewer but brighter fluors (10 nM Rhodamine)

Sample 3: The mixture

The occupancy fluctuations for each specie in the mixture becomes a convolution of the individual specie histograms. The resulting histogram is then broader than

expected for a single species.

Page 38: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

-10

-8

-6

-4

-20

log(

PCH

)

0 5 10 15

k

-2

0

2re

sidu

als/s

-8

-6

-4

-2

0

log(

PCH

)

0 5 10 15

k

-2

0

2

resi

dual

s/s

Singly labeled proteins Mixture of singly ordoubly labeled proteins

Resolve a protein mixture with a brightness ratio of two

Alcohol dehydrogenase labeling experiments

+

c 1 1N 2 2N

Sample A 0.190.1826.2

0.0040.0040.540 ----- -----

Sample B 0.61.225.1

007.0002.0155.0

101056

008.0003.0006.0

Both species havesame• color

• fluorescence lifetime• diffusion coefficient

• polarization

kcpsm kcpsm

Page 39: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

0

1x104

2x104

3x104

EGFPsolution

EGFPcytoplasm

EGFPnucleus

autofluorescencenucleus

autofluorescencecytoplasm

Mo

lecu

lar

Brig

htne

ss (

cpsm

)

The molecular brightness of EGFP is a factor ten higher than that of the autofluorescence in HeLa cells

Excitation=895nm

PCH in cells: Brightness of EGFP

Chen Y, Mueller JD, Ruan Q, Gratton E (2002) Biophysical Journal, 82, 133 .

Page 40: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

100 10000

2500

5000

7500

10000

EGFP Brightness EGFP

app (

cpsm

)

Concentration [nM]

104 105 106

Brightness of EGFP2 is twice the brightness of EGFP

Brightness and Stoichiometry

Intensity (cps)

EGFP

100 10000

2500

5000

7500

100002 x EGFP Brightness

EGFP Brightness EGFP EGFP

2

app (

cpsm

)

Concentration [nM]

104 105 106

EGFP2

Chen Y, Wei LN, Mueller JD, PNAS (2003) 100, 15492-15497

Page 41: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

100 1000 10000-30000

-20000

-10000

0

10000

20000

after correction before correction

Mo

lecu

lar

bri

gh

tne

ss (

cpsm

)

EGFP concentration (nM)

Caution: PCH analysis and dead-time effects

PCH analysis assumes ideal detectors. Afterpulsing and deadtime of the photodetector change the photon count statistics and lead to biased parameters. Improved PCH models that take non-ideal detectors into account are available:Hillesheim L, Mueller JD, Biophys. J. (2003), 85, 1948-1958

Page 42: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Distinguish Homo- and Hetero-interactions in living cells

Apparent Brightness

BA

A B

A B A B A B+ A A

ε ε

ε 0

ε

ε

ε 2ε

2ε2 905nm

2 965nm

ECFP: EYFP:

• single detection channel experiment• distinguish between CFP and YFP by excitation (not by emission)!• brightness of CFP and YFP is identical at 905nm (with the appropriate filters)• you can choose conditions so that the brightness is not changed by FRET between CFP and YFP• determine the expressed protein concentrations of each cell!

Page 43: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

0 4000 8000 120000.0

0.5

1.0

1.5

2.0

2.5

3.0

- RXR agonist + RXR agonist

app/ m

onom

er

total protein concentration [nM]0 1000 2000 3000 4000

0.0

0.5

1.0

1.5

2.0

2.5

3.0

- RXR agonist + RXR agonist

app/ m

onom

er

RXRLBD-YFP [nM]

PCH analysis of a heterodimer in living cells

We expect: RAR

RXR

2 965nm 2 905nm

The nuclear receptors RAR and RXR form a tight heterodimer in vitro. We investigate their stoichiometry in the nucleus of COS cells.

Chen Y, Li-Na Wei, Mueller JD, Biophys. J., (2005) 88, 4366-4377

Page 44: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Two Channel Detection:Cross-correlation

Each detector observes the same particles

Sample Excitation Volume

Detector 1 Detector 2

Beam Splitter1. Increases signal to noise

by isolating correlated signals.

2. Corrects for PMT noise

Page 45: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Removal of Detector Noise by Cross-correlation

11.5 nM Fluorescein

Detector 1

Detector 2

Cross-correlation

Detector after-pulsing

Page 46: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

)()(

)()()(

tFtF

tdFtdFG

ji

ji

ij

Detector 1: Fi

Detector 2: Fj

26x103

24

22

20

18

16

14

12

Flu

orescence

35302520151050Time

26x103

24

22

20

18

16

14

12

Flu

orescence

35302520151050Time

t

Calculating the Cross-correlation Function

t +

time

time

Page 47: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Thus, for a 3-dimensional Gaussian excitation volume one uses:

21

212

1

212

1212

81

81)(

z

D

w

D

NG

Cross-correlation Calculations

One uses the same fitting functions you would use for the standard autocorrelation curves.

G12 is commonly used to denote the cross-correlation and G1 and G2 for the autocorrelation of the individual detectors. Sometimes you will see Gx(0) or C(0) used for the cross-correlation.

Page 48: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Two-Color Cross-correlation

Each detector observesparticles with a particular color

The cross-correlation ONLY if particles are observed in both channels

The cross-correlation signal:

Sample

Red filter Green filter

Only the green-red molecules are observed!!

Page 49: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Two-color Cross-correlation

G ij() dF

i(t)dF

j(t )

Fi(t) F

j(t)

)(12 G

Equations are similar to those for the cross

correlation using a simple beam splitter:

Information Content Signal

Correlated signal from particles having both colors.

Autocorrelation from channel 1 on the green particles.

Autocorrelation from channel 2 on the red particles.

)(1 G

)(2 G

Page 50: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Experimental Concerns: Excitation Focusing & Emission Collection

Excitation side: (1) Laser alignment (2) Chromatic aberration (3) Spherical aberration

Emission side:(1) Chromatic aberrations(2) Spherical aberrations(3) Improper alignment of detectors or pinhole

(cropping of the beam and focal point position)

We assume exact match of the observation volumes in our calculations which is difficult to obtain experimentally.

Page 51: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Uncorrelated

Correlated

Two-Color Fluctuation Correlation Spectroscopy

1)()(

)()()(

tFtF

tFtFG

ji

jiij

Interconverting

2211111 )( NfNftF 2221122 )( NfNftF

Ch.2 Ch.1

450 500 550 600 650 7000

20

40

60

80

100

Wavelength (nm)

%T

For two uncorrelated species, the amplitude of the cross-correlation is proportional to:

2

222212112212211

2

11211

222211121112

)()0(

NffNNffffNff

NffNffG

Page 52: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Does SSTR1 exist as a monomer after ligand binding whileSSTR5 exists as a dimer/oligomer?

Somatostatin

Fluorescein Isothiocyanate (FITC)

Somatostatin

Texas Red (TR)

Cell Membrane

R1

R5 R5

R1

Three Different CHO-K1 cell lines: wt R1, HA-R5, and wt R1/HA-R5Hypothesis: R1- monomer ; R5 - dimer/oligomer; R1R5 dimer/oligomer

Collaboration with Ramesh Patel*† and Ujendra Kumar**Fraser Laboratories, Departments of Medicine, Pharmacology, and Therapeutics and Neurology and Neurosurgery, McGill University, and Royal VictoriaHospital, Montreal, QC, Canada H3A 1A1; †Department of Chemistry and Physics, Clarkson University, Potsdam, NY 13699

Page 53: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Green Ch. Red Ch.

SSTR1 CHO-K1 cells with SST-fitc + SST-tr

• Very little labeled SST inside cell nucleus• Non-homogeneous distribution of SST• Impossible to distinguish co-localization from molecular interaction

Page 54: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Minimum

MonomerA

Maximum

DimerB

= 0.22G12(0)

G1(0)

= 0.71G12(0)

G1(0)

Page 55: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Experimentally derived auto- and cross-correlation curves from live R1 and R5/R1 expressing CHO-K1 cells using dual-color two-photon FCS.

The R5/R1 expressing cells have a greater cross-correlation relative to the simulated boundaries than the R1 expressing cells, indicating a higher level of dimer/oligomer formation.

R1 R1/R5

Patel, R.C., et al., Ligand binding to somatostatin receptors induces receptor-specific oligomer formation in live cells. PNAS, 2002. 99(5): p. 3294-3299

Page 56: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

- 4 Ca2+ + 4 Ca2+

CFP YFP

calmodulin M13

CFP

“High” FRET

“Low” FRET

trypsin

CFP YFP+

NO FRET

(a)

(b)

(c)

Molecular Dynamics

What if the distance/orientation is not constant?

• Fluorescence fluctuation can result from FRET or Quenching

• FCS can determine the rate at which this occurs

• This will yield hard to get information (in the s to ms range) on the internal motion of biomolecules

Page 57: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

450 500 550 600 6500

10000

20000

30000

40000

50000

60000

Fluo

resc

ence

Int

ensi

ty (

cps)

Wavelength (nm)

trypsin-cleaved cameleon

CFP cameleon

Ca2+-depleted cameleon

Ca2+-saturated YFP

Page 58: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

A)

10-3 10-2 10-1 100 101

0

40

80

120

160

200

(s)

G( )

B)

. A) Cameleon fusion protein consisting of ECFP, calmodulin, and EYFP.

[Truong, 2001 #1293] Calmodulin undergoes a conformational change that allows the

ECFP/EYFP FRET pair to get cl ose enough for efficient energy transfer. Fluctuations

between the folded and unfolded states will yield a measurable kinetic component for the

cross-correlation. B) Simulation of how such a fluctuation would show up in the

autocorrelation and cross-correlation. Red dashed line indicates pure diffusion.

Page 59: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Crystallization And Preliminary X-Ray Analysis Of Two New Crystal Forms Of Calmodulin, B.Rupp, D.Marshak and S.Parkin, Acta Crystallogr. D 52, 411 (1996)

Ca2+ Saturated

Are the fast kinetics (~20 s) due to conformational changes or to fluorophore blinking?

In vitro Cameleon Data

Page 60: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Cross-Correlation

Dual-Color PCHF

luor

esce

nce

F(t

)

<FA>

FA

Time t

<FB>FB

Time t

Flu

ores

cenc

e F

(t)

Dual-color PCH analysis (1)

Page 61: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Tsample

Sig

nal A

Tsample

Sig

nal B

Brightness in each channel: A, B

Average number of molecules: N

Page 62: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Tsample

Sig

nal A

Tsample

Sig

nal B

Brightness in each channel: A, B

Average number of molecules: N

Dual-color PCH analysis (2)

Single Species: ( , ) ( , , )A B A Bp k k PCH N

Page 63: Lecture 7 FCS, Autocorrelation, PCH, Cross-correlation Joachim Mueller Principles of Fluorescence Techniques Laboratory for Fluorescence Dynamics Figure

Dual-Color PCHfluctuations

2 channels

2 species model 2 = 1.01

2000 3000 4000 5000 6000 7000 8000 9000100002000

3000

4000

5000

6000

7000

8000

9000

10000

EYFP alone ECFP alone species 1 from fit species 2 from fit

A (

cpsm

)

B (cpsm)

1 species model

2 = 17.93

ECFP & EYFP mixture resolved with single histogram.

Resolve Mixture of ECFP and EYFP in vitro

Note: Cross-correlation analysis cannot resolve a mixture of ECFP & EYFP with a single measurement!

Chen Y, Tekmen M, Hillesheim L, Skinner J, Wu B, Mueller JD, Biophys. J. (2005), 88 2177-2192