12
1 RADIOLOGY RESEARCH AAPM 2011 Annual Meeting CR/DR Image Noise Part 2 Mark P. Supanich Department of Radiology [email protected] MO-A-110-2 Learning Objectives 1. To understand the roll of measuring the Noise Power Spectrum (NPS) of a digital radiography detector in acceptance testing or routine QA 2. To understand the methodology and challenges of measuring the NPS of a digital radiography detector in a clinical setting 3. To become familiar with the NPS measurement’s sensitivity to the acquisition technique and beam conditions 2 AAPM 2011 3 AAPM 2011 Outline Motivation Setup of Beam Conditions for NPS measurement Acquisition and Export of data for NPS analysis Use of NPS Analysis software Examples of acquired NPS data Motivation Measurement of NPS provides a robust and reliable metric of digital detector performance as part of a Quality Control and Assurance Program NPS is more sensitive than visual assessment of Contrast Detail Phantoms (Rivetti et al., Med Phys V 37(2)) or the use of the Standard Deviation of the image noise (Part 1 of this session) NPS is easily repeatable and allows comparison amongst different detector systems 4 AAPM 2011

Learning Objectives RADIOLOGY - AAPM

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Learning Objectives RADIOLOGY - AAPM

1

RADIOLOGY

RESEARCH

AAPM 2011 Annual Meeting

CR/DR Image Noise –

Part 2

Mark P. Supanich

Department of [email protected]

MO-A-110-2

Learning Objectives

• 1. To understand the roll of measuring the Noise

Power Spectrum (NPS) of a digital radiography

detector in acceptance testing or routine QA

• 2. To understand the methodology and challenges

of measuring the NPS of a digital radiography

detector in a clinical setting

• 3. To become familiar with the NPS

measurement’s sensitivity to the acquisition

technique and beam conditions

2AA

PM

2011

3AA

PM

2011

Outline

• Motivation

• Setup of Beam Conditions for NPS measurement

• Acquisition and Export of data for NPS analysis

• Use of NPS Analysis software

• Examples of acquired NPS data

Motivation

• Measurement of NPS provides a robust and

reliable metric of digital detector performance as

part of a Quality Control and Assurance Program

• NPS is more sensitive than visual assessment of

Contrast Detail Phantoms (Rivetti et al., Med Phys V

37(2)) or the use of the Standard Deviation of the

image noise (Part 1 of this session)

• NPS is easily repeatable and allows comparison

amongst different detector systems

4AA

PM

2011

Page 2: Learning Objectives RADIOLOGY - AAPM

2

Motivation

• Regular Acquisition of NPS over time:

• Allows monitoring of performance of detector

from acceptance through end of life

• Normalizing NPS to the exposure at detector

allows direct comparison of data from different

time points

• Failure modes of digital detectors can be observed

through documentation and monitoring of NPS

5AA

PM

2011

Motivation

• NPS vs. Contrast Detail Analysis or Standard

Deviation of Noise for QC

• The square of the Standard Deviation of noise is

integral of NPS over frequency and cannot

differentiate between different noise textures

• Contrast Detail analysis is subjective and not

repeatable (same conditions can never be truly

replicated as observer changes over time)

• Contrast Detail analysis does not provide as clear a

visualization of differences between systems as

NPS

6AA

PM

2011

Motivation

• Plot of Contrast Detail Curve for Multiple

Detector Systems

7AA

PM

2011

Rivetti et al. Med Phys 37(2)

Motivation

• Plot of NPS for Same Detector Systems allows

better visualization of differences

8AA

PM

2011

Riv

ett

i et

al. M

ed P

hys

37(2

)

Page 3: Learning Objectives RADIOLOGY - AAPM

3

Setup of Beam Conditions for NPS measurement

• Beam Conditions for NPS of Radiographic Units

are AAPM Exposure Index Conditions defined in

TG 116 (Sheppard et al. Med Phys 36(7))

• Well established approach to reach equivalent HVL

of 6.8 mm Al for data acquisition using:

• 66-74 kVp at 0.5 mm Cu + 0-4 mm Al

• Commercial products available to assist in achieving

Standard Beam Conditions

• Radiography III (WE-A-110-1) and Part I of this

session discusses beam conditions further

• Use of same beam conditions as EI allows for easy

acquisition of NPS data at acceptance of unit

9AA

PM

2011

Setup of Beam Conditions for NPS measurement

• Setup of Standard Beam Conditions for

Radiography Systems

• Standard Beam Condition produce relatively flat

output of photons per mR vs. kVp

• Adjust each system to get same HVL at Standard

Beam Conditions so same Quanta per mR can be

used for normalization

• Adjust kVp to reach 6.8 mm Al HVL as last resort

(add Al to 0.5 mm Cu as first step)

• Precise positioning (and measurement) of ion

chamber location is very important for calculation

of exposure at detector

• Removal of Anti-Scatter Grid is required10A

APM

2011

Setup of Beam Conditions for NPS measurement

11AA

PM

2011

Illustration from TG116 report

Setup of Beam Conditions for NPS measurement

• Required equipment for NPS measurements on

Radiography Systems

• Ion Chamber and kVp meter and R/O box

• At least 7 mm of Al for HVL determination under

Standard Beam Conditions

• 0.5 Cu + 0-4 mm Al (or equivalent) to reach

Standard Beam Conditions

• Mammography Systems

• 40 mm of Breast Tissue Equivalent material used to

establish beam conditions of NPS measurement

12AA

PM

2011

Page 4: Learning Objectives RADIOLOGY - AAPM

4

Acquisition and Export of data for NPS analysis

• Analysis of NPS requires acquisition and exporting

of images with “For Processing” lmage Values as

defined in TG 116

• Gain and offset corrections and flat field

corrections should be applied to produce For

Processing images that can be analyzed

• Relationship between Image values and exposure

is different for different manufacturers

• Linear vs. Log relationship of pixel values to

exposure

13AA

PM

2011

Acquisition and Export of data for NPS analysis

• “Raw” data exported without corrections: Not

Useable for NPS analysis – also includes possible

lag artifacts

14AA

PM

2011

Acquisition and Export of data for NPS analysis

• Acquisition Steps

• Setup Beam Conditions

• Determine average exposure at the center of the

beam

• Determine exposure with ion chamber at the edge

of the beam (setup for image acquisition)

• Acquire 5-10 images at detector exposures ranging

from ~ 0.1 to 10 mR

• Detector exposure determined by geometric

correction of ion chamber placed at edge of field

15AA

PM

2011

Acquisition and Export of data for NPS analysis

• Approach for each system to allow for export of

for processing data (and in what form) must be

defined and supplied by vendors for

implementation of NPS (and MTF) analysis for QC

• Some systems require service level access to set

up for processing export

• Some systems can be setup once in service mode

and tech access can export data

• Radiography III (WE-A-110-1) includes

information on image export

16AA

PM

2011

Page 5: Learning Objectives RADIOLOGY - AAPM

5

Acquisition and Export of data for NPS analysis

• Challenges to acquiring data for NPS analysis

• Pixel values saturate if exposure is too high

• Gain/offset and flat field corrections not applied to

exported raw data

• Must correct for Image Values that have log

relationship to exposure (discussed in Part 1)

• Small changes in setup geometry can significantly

adjust calculated detector exposure and affect

normalization

• Solution: for repeated measurements on same

system, use average exposure value for same mAs

• Setup of system for export of For Processing

Images17A

APM

2011

Acquisition and Export of data for NPS analysis

• Exporting For Processing Images:

• AGFA required Service Access to set up export,

tech access to export once set up

• Philips required Service Access to set up and

export

• Lorad requires tech level access to export

• Exported raw

• GE Mobile unit required Physics l, uncorrected

imageevel access to export

• Exported raw, uncorrected image

18AA

PM

2011

Acquisition and Export of data for NPS analysis

AGFA DXS Example

Of Steps to Export For

Processing Images

19AA

PM

2011

Export Procedure for AGFA: 1. Go to Main Menu –select Service & Configuration Tool

20AA

PM

2011

Page 6: Learning Objectives RADIOLOGY - AAPM

6

Export Procedure for AGFA: 2. Select Device Configuration

21AA

PM

2011

Export Procedure for AGFA: 3. Select Export Destination

22AA

PM

2011

Export Procedure for AGFA: 4. Set DX for Processing –this allows export of For Processing Images to CD Writer

23AA

PM

2011

Export Procedure for AGFA: 5. Select export images from main menu and export desired images to CD as For Processing

24AA

PM

2011

Page 7: Learning Objectives RADIOLOGY - AAPM

7

Acquisition and Export of data for NPS analysis

Philips Pixium Example

Of Steps to Export For

Processing Images

25AA

PM

2011

Export Procedure for Philips: 1. Enter Service Mode and Select System Configuration

26AA

PM

2011

Export Procedure for Philips: 2. Under “Unstructured Items” tab change Unprocessed Export to “yes”

27AA

PM

2011

Export Procedure for Philips: 3. Under “DICOM” tab Enable Local Storage of DX images

28AA

PM

2011

Page 8: Learning Objectives RADIOLOGY - AAPM

8

Export Procedure for Philips: 4. After saving changes and restarting application, export images to “Storage Medium” as “Pre”

29AA

PM

2011

Acquisition and Export of data for NPS analysis

Lorad Selenia:

Select Output as -

Cdrw Raw to Export Raw

Images (uncorrected) to CD

30AA

PM

2011

Acquisition and Export of data for NPS analysis

GE AMX 700:

Select Raw processing option to

allow Export of Raw Images

(uncorrected) to PACS system

31AA

PM

2011

Use of NPS Analysis Software

• Requires DICOM input

• Define following system specific parameters

described further in Part I of this session

• Block analysis size

• Area to analyze NPS over

• Pixel size

• Correction factor for systems with log image value

to exposure relationship

• Outputs following analysis are

• 2D NPS image

• Data and plots of 1D NPS data

32AA

PM

2011

Page 9: Learning Objectives RADIOLOGY - AAPM

9

Use of NPS Analysis software

• Analysis of Data

• Plot calculated average pixel value vs. exposure

• Determine if pixel values have log or linear

relationship to exposure

• If log relationship, calculate gamma to use in NPS

calculation

33AA

PM

2011

Use of NPS Analysis Software: Q Values with log relationship to exposure

34AA

PM

2011

y = 2567.ln(x) + 19658

1.00E+04

1.20E+04

1.40E+04

1.60E+04

1.80E+04

2.00E+04

2.20E+04

2.40E+04

2.60E+04

0.010 0.100 1.000 10.000

Mean

Sig

nal

Exposure at Detector (mR)

a-Si w/ CsI phosphor #1 log Signal vs. log Exposure

Use of NPS Analysis Software: Q Values with linear relationship to exposure

35AA

PM

2011

y = 6038.x + 109.8

0.00

2000.00

4000.00

6000.00

8000.00

10000.00

12000.00

14000.00

16000.00

0.000 0.500 1.000 1.500 2.000 2.500 3.000

Mean

Sig

nal

Exposure at Detector (mR)

a-Si w/ CsI phosphor #2 Signal vs. Exposure

Use of NPS Analysis software

• Run NPS analysis using system specific parameters

• Examine 2D image of NPS to identify any NPS

trends

• Choose representative 1D plots of data if desired

• E.g. suppressed horizontal power vs. diagonal signal

• Normalize NPS data using ideal quanta per mR and

calculated exposure at detector to get nNPS

• Ideal Quanta per mR under Standard Beam

Conditions for Radiographic systems ~ 255 photons

• Ideal Quanta per mR for Mammography System w/

40 mm breast equivalent in beam at 28 kV Mo/Mo ~

45 photons

36AA

PM

2011

Page 10: Learning Objectives RADIOLOGY - AAPM

10

Storage Phosphor 2D NPS at ~ 4 mR Air Kerma

37AA

PM

2011

a-Si detector with CsI Scintillator 2D NPS at ~ 3 mR Air Kerma

38AA

PM

2011

Examples of Acquired NPS Data

39AA

PM

2011

0.00E+00

1.00E-06

2.00E-06

3.00E-06

4.00E-06

5.00E-06

6.00E-06

0 0.5 1 1.5 2 2.5 3

NP

S (

mm

^2)

Cycles/mm

Diagonal and Horizontal NPS of a-Si detector at 3 mR

Horizontal NPS

Diagonal NPS

Examples of Acquired NPS Data

40AA

PM

2011

0.00

0.50

1.00

1.50

2.00

2.50

3.00

3.50

0 0.5 1 1.5 2 2.5 3

nN

PS

Frequency (cycles/mm)

Normalized NPS of Two Radiographic Systems

A-Si w/ CsI 144

micron pixel size

Needle

Phosphor 100

micron pixel size

Page 11: Learning Objectives RADIOLOGY - AAPM

11

Examples of Acquired NPS Data

41AA

PM

2011

0.00

0.50

1.00

1.50

2.00

2.50

3.00

3.50

0 0.5 1 1.5 2 2.5 3

nN

PS

Cycles/mm

NPS Measurements of a-Si Detector w/ CsI over 60 Days

a-Se Detector 2D NPS at ~ 2 mR Air Kerma: Not Flat Field Corrected

42AA

PM

2011

Examples of Acquired NPS Data

43AA

PM

2011

0.00E+00

1.00E-06

2.00E-06

3.00E-06

4.00E-06

5.00E-06

6.00E-06

7.00E-06

8.00E-06

0 1 2 3 4 5 6

NP

S (

mm

^2)

Cycles/mm

NPS of a-Se Detector w/ Exposure of 2 mR at Detector (Not

Flat Field Corrected)

Conclusion

• NPS analysis is a robust, reliable and repeatable

assessment of detector noise performance as part

of a Quality Assurance Program

• A challenge to widespread adoption of NPS as a

regular QC metric is the difficulty in producing

and exporting gain, offset and flat field corrected

images with “for processing” image values

• Clear documentation and support from

manufacturers to allow export of images that can

be used for NPS analysis is vital

44AA

PM

2011

Page 12: Learning Objectives RADIOLOGY - AAPM

12

45AA

PM

2011

?