Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin NCCAAPM meeting October 24, 2014

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Embed Size (px)

Citation preview

Page 1: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Larry A. DeWerd, PhD, FAAPMUW ADCL & Dept. Medical Physics

University of Wisconsin

NCCAAPM meeting October 24, 2014

Page 2: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Larry DeWerd has partial interest in Standard Imaging Inc.

Page 3: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Understanding of the TG 51 addendum The effect for kQ Reference chambers

Page 4: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

1. Reference dosimetry for linac beams based on a 60Co calibration.

2. Simple step-by-step procedure. (Much of the physics is included in the calibration and measurement parameters.)

Be careful with small fields.

Page 5: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

TG 51 gives a measure of the absorbed dose to water using an ADCL calibrated ion chamber.

kQ is the factor that converts from the calibration beam (60Co) to the user linac beam, defined by beam quality Q

Dw,Q ND,w

60Co kQ Mion

Page 6: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

6

Institutions using TG 51

Page 7: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

A. kQ factors for new chambersB. Recommendations for implementationC. Uncertainty analysis for implementation

of TG-51D. Comparison of measured and calculated

kQ factors: Measured kQ data is available for some types of chambers

Page 8: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

What are reference class ionization chambers

%dd(10) is used for kQ – No lead foil to be used anymore – caused errors. Use eq. 15 in TG 51

Use of small volume chambers in relative dosimetry

Non-water phantoms – only water for dosimetry

Development of uncertainty budget may take some time

Page 9: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

For chambers listed in both this addendum and the original TG-51 protocol, the kQ factors listed in the addendum should be used.

For chambers that are not listed in either the original TG-51 protocol or in this addendum the recommendations of Section XI of TG-51 should be followed.

Page 10: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

3 sub-types (NOTE: WGTG51 definitions) –i. 0.6 cm3 reference chambers (e.g.,

NE2571, PR-06C)ii. 0.125 cm3 scanning chambers (e.g.,

PTW31010, IBA CC13)iii. 0.02 cm3 micro chambers (e.g.,

Exradin A16, PinpointTM)

Page 11: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Examples

31010

CC01

A12

NE2577

0.125 cm3

Scanning chamber

0.01 cm3

Micro chamber

0.6 cm3

‘Farmer’ chamber

0.25 cm3

‘Short Farmer’

PTW31010

IBA CC01

Exradin A12

NE2577

Page 12: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Measurand SpecificationChamber settling Must be less than a 0.5 % change in reading from beam-on to

stabilization Pleak < 0.1 % of chamber reading Ppol < 0.4 % correction (0.996 < Ppol < 1.004)

< 0.5 % maximum variation in Ppol with energy (total range) Pion = Pinit +Pge n Pgen Pinit

Correction must be linear with dose per pulse Initial recombination must be < 0.002 at 300 V Correction follows Boag theory for chamber dimensions. Difference in initial recombination correction between opposite polarities < 0.1 %

Chamber stability Must exhibit less than a 0.3 % change in calibration coefficient over the typical recalibration period of 2 years

Page 13: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Most are Farmer typeso NE2571 and NE2611

o PTW30010, PTW30012, PTW30013, PTW31013

o Exradin A12, A12S, A19, A18, A1SL

o IBA FC65-G, FC65-P, FC23-C, CC25, CC13

o Capintec PR-06C

Page 14: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Majority are 0.6 cm3 ‘Farmer-type’ chambers

A-150 chambers explicitly excluded 5 scanning chambers, NO

microchambers (Exception A26 from some preliminary

measurements. Long term to come) No parallel plate chambers are

included

Page 15: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Recombination correction directly affects measurement of absorbed dose

Recombination correction well established butnot always straightforward

2-voltage technique as set out in TG-51 applicable only to chambers exhibiting ideal behaviour

Many examples in literature of anomalous behaviour

elecpolionTPrawwcorr PPPPMM ,

Page 16: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

B.4 Polarizing voltage

equivalent electrode separation2 (mm2)

2 4 6 8 10 12 14 16

Slo

pe(P

ion/D

pp)

0.00

0.05

0.10

0.15

0.20

0.25

0.30

"good" chambersLinear fitCC08CC13CC04

References: DeBlois et al (Med. Phys., 2000), McEwen (Med. Phys., 2010)

Page 17: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Not all chambers follow standard ‘Boag’ theory Manufacturers’ statements on voltage limits

need verifying (at least for chamber types, if not individual chambers)

Going to a higher polarizing voltage can lead to a larger uncertainty in the measurement

Recombination can be a function of the sign of the charge collected

Addendum recommends a maximum value of 300 V (lower values may be required for small-volume chambers)

Page 18: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

polarity effect is associated with a net deposition of charge in the chamber

In the situation of transient charged particle equilibrium there is no net charge deposited.

The polarity correction in photon beams should therefore be small.

Ppol values between 0.997 and 1.003 should be expected.

Provides a good check of chamber

Page 19: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

ISO Guide to the Expression of Uncertainty in Measurement

NIST has produced an explanatory document (a guide to the Guide) - NIST Technical Note 1297

Discussion of Type A and B uncertainties (distinct from ‘random’ and ‘systematic’)

Uncertainty budget broken down into: Measurement Calibration data Influence quantities

Typical values discussed but emphasis on individual users constructing site-specific uncertainty budgets for their calibration situations

Page 20: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Uncertainties subject of another lecture

An example given here.

Component of Uncertainty Type A Type B

Measurement

SSD setting 0.10%

Depth setting 0.25%

Charge measurement 0.10%

PTP correction 0.10%

Calibration data

Co-60 ND,w 0.70%

kQ factor 0.33%

Assignment of kQ factor 0.10%

Influence quantitiesPpol 0.05%

Pion 0.10%

Pre-irradiation history 0.10%Pleak 0.05%

Calibration coefficient 0.20%

Linac stability 0.10%

OVERALL 0.9%

User-dependent part 0.3%

Page 21: Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical ...chapter.aapm.org/nccaapm/z_meetings/2014-10-23_and... · Larry DeWerd has partial interest in Standard Imaging Inc. ... o IBA

Thanks are due to Students and staff

of the UW ADCL All those who send

us calibration instruments that support the research program of the UW Medical Radiation Research Center.