AMSCO ® EVOLUTION STEAM STERILISERS SD862 (07/01/07) Item ________________________ Location(s)___________________ ____________________________ Typical only - some details may vary. The Selections Checked Below Apply To This Equipment MODEL NUMBER/SIZE HC600 / 660 x 660 x 991 mm (26 x 26 x 39") HC800 / 660 x 660 x 1245 mm (26 x 26 x 49") HC1000 / 660 x 660 x 1549 mm (26 x 26 x 61") HC900 / 660 x 950 x 914 mm (26 x 37-1/2 x 36") HC1200 / 660 x 950 x 1219 mm (26 x 37-1/2 x 48") HC1500 / 660 x 950 x 1524 mm (26 x 37-1/2 x 60") STEAM SOURCE House Steam Optional Integral Electric Steam Generator STERILISER ELECTRIC SERVICE 400 VAC, 50 Hz, 3-Phase, 6Amp/Phase DOOR CONFIGURATION (660 X 950 MM STERILISERS, ONLY) Single-Door Horizontal-Sliding Left-Hand Double-Door Horizontal-Sliding Door Left-Hand (Direction of door movement is right-hand to left- hand, as viewed from the steriliser's operating end.) SINGLE-DOOR MOUNTING Cabinet Enclosed/Freestanding Recessed DOUBLE-DOOR MOUNTING* Recessed through One Wall Recessed through Two Walls * 914 mm and 991 mm sterilisers not available in double door configuration ACCESSORIES 991 mm (39") Chamber Length 1245 mm (49") Chamber Length 1549 mm (61") Chamber Length 914 mm (36") Chamber Length 1219 mm (48") Chamber Length 1524 mm (60") Chamber Length Loading Car Transfer Carriage Chamber Track Assembly Single Door Double Door Loading Car, Transfer Carriage, and Track Assembly Single Door Double Door Chamber Rack and Shelf (914 mm and 991 mm Units, only) APPLICATION This steriliser is configured for prevac- uum sterilisation of heat- and moisture- stable materials used in healthcare facil- ities. These units are equipped with pre- vacuum, Bowie-Dick, and leak test cycles. DESCRIPTION Amsco ® Evolution™ Steam Sterilisers are equipped with the latest features in both state-of-the-art technology and ease of use. Primary Product Features – Interior Chamber Dimensions 660 x 660 x 991 mm (26 x 26 x 39") – 430 L capacity 660 x 660 x 1245 mm (26 x 26 x 49") – 540 L capacity 660 x 660 x 1549 mm (26 x 26 x 61") – 675 L capacity 660 x 950 x 914 mm (26 x 37-1/2 x 36") – 575 L capacity 660 x 950 x 1219 mm (26 x 37-1/2 x 48") – 766 L capacity 660 x 950 x 1524 mm (26 x 37-1/2 x 60") – 958 L capacity Horizontal-sliding door 660 x 950 mm (26 x 37-1/2") or vertical-sliding door 660 x 660 mm (26 x 26") with quiet, motor-driven cable and pulley mechanism. Door movement is con- trolled from control panel push button. » For 660 x 660 mm (26 x 26") sterilis- ers, door travels vertically down to open and up to close. » For 660 x 950 mm (26 x 37-1/2") sterilisers, door travels horizontally right to left to open. All plumbing components are mounted to a free-standing, modular rack (skid). The skid connects to the core steriliser assembly during instal- lation. IMPORTANT: This tech data sheet is intended for use outside the U.S.A. only. 660 x 660 mm (26 x 26") Steriliser 660 x 950 mm (26 x 37-1/2") Steriliser

Large Steam Sterilizers

Embed Size (px)

Citation preview

Page 1: Large Steam Sterilizers

AMSCO® EVOLUTION™

STEAM STERILISERS

SD862 (07/01/07)

Item ________________________Location(s)___________________

____________________________

Typical only - some details may vary.

The Selections Checked Below Apply To This Equipment

MODEL NUMBER/SIZE❑ HC600 / 660 x 660 x 991 mm

(26 x 26 x 39")

❑ HC800 / 660 x 660 x 1245 mm (26 x 26 x 49")

❑ HC1000 / 660 x 660 x 1549 mm(26 x 26 x 61")

❑ HC900 / 660 x 950 x 914 mm(26 x 37-1/2 x 36")

❑ HC1200 / 660 x 950 x 1219 mm(26 x 37-1/2 x 48")

❑ HC1500 / 660 x 950 x 1524 mm(26 x 37-1/2 x 60")

STEAM SOURCE❑ House Steam❑ Optional Integral Electric Steam Generator

STERILISER ELECTRIC SERVICE❑ 400 VAC, 50 Hz, 3-Phase, 6Amp/Phase

DOOR CONFIGURATION (660 X 950 MM STERILISERS, ONLY)❑ Single-Door

❑ Horizontal-Sliding❑ Left-Hand

❑ Double-Door ❑ Horizontal-Sliding Door❑ Left-Hand

(Direction of door movement is right-hand to left-hand, as viewed from the steriliser's operating end.)

SINGLE-DOOR MOUNTING❑ Cabinet Enclosed/Freestanding❑ Recessed

DOUBLE-DOOR MOUNTING*❑ Recessed through One Wall❑ Recessed through Two Walls* 914 mm and 991 mm sterilisers not available in

double door configuration

ACCESSORIES❑ 991 mm (39") Chamber Length❑ 1245 mm (49") Chamber Length❑ 1549 mm (61") Chamber Length❑ 914 mm (36") Chamber Length❑ 1219 mm (48") Chamber Length❑ 1524 mm (60") Chamber Length❑ Loading Car ❑ Transfer Carriage ❑ Chamber Track Assembly

❑ Single Door ❏ Double Door❑ Loading Car, Transfer Carriage, and Track

Assembly ❑ Single Door ❏ Double Door

❑ Chamber Rack and Shelf (914 mm and 991 mm Units, only)

APPLICATION

This steriliser is configured for prevac-uum sterilisation of heat- and moisture-stable materials used in healthcare facil-ities. These units are equipped with pre-vacuum, Bowie-Dick, and leak test cycles.

DESCRIPTION

Amsco® Evolution™ Steam Sterilisers are equipped with the latest features in both state-of-the-art technology and ease of use.

Primary Product Features –Interior Chamber Dimensions

660 x 660 x 991 mm(26 x 26 x 39") – 430 L capacity

660 x 660 x 1245 mm(26 x 26 x 49") – 540 L capacity

660 x 660 x 1549 mm(26 x 26 x 61") – 675 L capacity

660 x 950 x 914 mm (26 x 37-1/2 x 36") – 575 L capacity

660 x 950 x 1219 mm(26 x 37-1/2 x 48") – 766 L capacity

660 x 950 x 1524 mm(26 x 37-1/2 x 60") – 958 L capacity

• Horizontal-sliding door 660 x 950 mm (26 x 37-1/2") or vertical-sliding door 660 x 660 mm (26 x 26") with quiet, motor-driven cable and pulley mechanism. Door movement is con-trolled from control panel push button.

» For 660 x 660 mm (26 x 26") sterilis-ers, door travels vertically down toopen and up to close.

» For 660 x 950 mm (26 x 37-1/2")

sterilisers, door travels horizontallyright to left to open.

• All plumbing components are mounted to a free-standing, modular rack (skid). The skid connects to the core steriliser assembly during instal-lation.

IMPORTANT: This tech data sheet is intended for use outside the U.S.A. only.

660 x 660 mm (26 x 26") Steriliser

660 x 950 mm (26 x 37-1/2") Steriliser

Page 2: Large Steam Sterilizers

2

An advanced, PC-based control system employs user-friendly interface screens, with enhanced graphics conforming to international communications standards.

• 8.4" (214 mm) color touch screen display.

• Standard communication interface with most PC-compatible peripheral devices (e.g., data collection systems, printers)

• Automatic check of control program and cycle data main-tains process integrity

• Control is designed to accommodate integrated remote monitoring and instrument tracking interfaces.

STANDARDS

Each steriliser meets applicable requirements of the following listings and standards, and carries the appropriate symbols:

• Medical Device Directive (MDD) 93/42/EEC.

• EN 285, Large Steam Sterilisers.

• Pressure Equipment Directive (PED) 97/23/EC.

• ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form U-1 is furnished. Shell and door are constructed to withstand working pressure of 3.1 bar (45 psig).

FEATURES

660 x 950 mm (26 x 37-1/2") Chamber Cross-section or 660 x 660 mm (26 x 26") Chamber Cross-section sized to allow for efficient, high-volume processing of sterilisation con-tainers, trays and packs.

Vertical-sliding Door is controlled from control panel push buttons. The door slides open, propelled by a cable and pulley driven by an electric motor. Double door configurations are supplied with controls at both ends of the steriliser to help prevent the possibility of cross-contamination.

Resistive Thermal Detectors (RTD) are installed for steriliser temperature control. The dual element chamber drain line RTD senses and controls temperature variations within the steriliser chamber. A jacket RTD provides temperature control within the jacket space. These RTD signals, converted into electrical impulses, provide accurate control inputs and readouts throughout entire cycle.

Electronic water saving control includes an RTD to control the amount of water used in condensing the exhausted cham-ber steam and condensate.

Software calibration is performed in the Service Mode, acces-sible through the touch screen displays, and accomplished

using external or internal temperature and pressure sources. Control system provides printed record of all calibration data for verification to current readings.

Two-piece insulation sleeve is fitted around exterior of the steriliser vessel. The sleeve is sealed and held in place by hook-and-loop closures. Insulation is asbestos-free and chloride-free, silicone impregnated, oil- and water-resistant fiberglass.

Two-stage vacuum pump is supplied on all units to effectively pull chamber to specified vacuum levels, reduce cycle time by shortening conditioning and exhaust times; as well as reduce water consumption.

Pneumatic valves are fitted in piping for steam, water, and exhaust control.

Stainless-steel piping is used on all all chamber steam supply piping and is joined by sanitary fitting assemblies. Principle piping components and the primary control assembly are mounted to a separate, modular support rack (plumbing skid). The plumbing skid connects during installation to core chamber and frame assembly, allowing for increased access for service and maintenance procedures when necessary.

PROCESSING CYCLES

All processing cycles factory programmed into the steriliser control have been validated to EN 285.

NOTE: Any cycle created through changes to factory-set values requires validation by the facility or an independent authorized person before use. Refer to ISO EN 17665-1 for guidelines.

Prevacuum configuration sterilisers are factory programmed with the following cycles: 134°C (273°F) or 121°C (250°F) prevacuum cycles. Prevacuum cycles feature vacuum pulses followed by pressure pulses for porous load cycles. Prevac-uum cycles are intended for efficient, high-volume processing of heat- and moisture-stable materials, such as fabrics and wrapped hard goods. This process incorporates a series of vacuum/pressure pulses to condition the load prior to sterilisa-tion.

Sterilise temperature: 134°C (273°F)

» Sterilise time: 3 minutes and 30 seconds

» Dry time: 30 minutes or 20 minutes

or

» Sterilise temperature: 121°C (250°F)

» Sterilise time: 20 minutes

» Dry time: 30 minutes or 20 minutes

.

Internal Dimensions Cubic Cubic

Model Number Millimeters Inches Inches Feet Liters STU

HC600 660 x 660 x 991 mm 26 x 26 x 39" 26, 364 15.2 430 6

HC800 660 x 660 x 1245 mm 26 x 26 x 49" 33, 124 19.1 540 8

HC1000 660 x 660 x 1549 mm 26 x 26 x 61" 41, 236 23.8 675 10

HC900 660 x 950 x 914 mm 26 x 37-1/2 x 36" 35, 100 20.3 575 9

HC1200 660 x 950 x 1219 mm 26 x 37-1/2 x 48" 46, 800 27.0 766 12

HC1500 660 x 950 x 1524 mm 26 x 37-1/2 x 60" 58, 500 33.8 958 15

Page 3: Large Steam Sterilizers

3

TESTING CYCLES

• Vacuum Leak Test: This cycle is used for testing the vacuum integrity of the steriliser’s piping. Steriliser chamber must be empty while running this test cycle. All timing is preprogrammed and cannot be adjusted.

• Bowie-Dick Test Cycle: This cycle is used for conducting Bowie-Dick tests. Recommended load is a properly pre-pared Bowie-Dick test pack. Preprogrammed cycle param-eters cannot be adjusted by the user.

» Sterilise temperature: 134°C (273°F)

» Sterilise time: 3 minutes and 30 seconds

» Dry time: 1 minute

CONTROL SYSTEM

Design Features

The control system monitors and controls all steriliser opera-tions and functions. The control system is factory-programmed with standard sterilising cycles. Each cycle is adjustable to

meet specific processing requirements. All operator-accessi-ble control functions can be changed using the touch screen control.

IMPORTANT: If factory-programmed cycle values are changed, it is the responsibility of the healthcare facility to validate the efficacy of the changed cycle. Refer to ISO EN 17665-1 for guidelines.

Cycle values and operating features may be adjusted and verified prior to cycle operation. Cycle parameters are retained in control memory for repeated use.

Once cycle is started, cycles and cycle values cannot be changed until cycle is complete. If chamber temperature drops below set point during the exposure phase, the timer is set to stop and automatically reset once normal operating tempera-ture is reached.

Critical control system components are housed within a sealed compartment to protect the components from moisture and heat generated during the sterilisation process.

Operator interface control panel, consisting of a touch screen, is located on the operating (i.e., load or non-sterile) end of the steriliser.

• The operator interface consists of a touch-sensitive, color screen. This display allows for control communications, graphics, and excellent visibility in most environments. The display panel, in conjunction with the control, is used as the monitor for the operator. All steriliser functions, including cycle initiation and cycle configuration, are operated by pressing the touch-sensitive areas on the display. Display indicates appropriate control buttons, operator prompts, and status messages necessary to assist in steriliser oper-ation. All displayed messages are complete phrases with no codes to be cross-referenced. Display also indicates any abnormal conditions that may exist either in or out of a cycle. Control buffer memory retains up to ten previously-run cycles for later access.

24-Character ink-on-paper printer, located above touch screen, provides an easy-to-read printed record of all pertinent cycle data on 57 mm (2-1/4") wide paper. Data is automatically printed at the beginning and end of each cycle and at transition points during the cycle.Three paper tape rolls and two printer ribbons are furnished with each unit.

Non-operating end (NOE) control panel, on double-door sterilisers only, includes a touch sensitive screen similar to the operating end screen. Preprogrammed cycles can be started from the NOE control panel. Display concurrently shows the same information as the operating end screen display. Other controls located at the non-operating end include door control pushbuttons, jacket and chamber gauges and an emergency-stop button.

Cycle configuration is performed by accessing the Change Values menu through the operating end touch screen. In addition to adjustment of cycle values, the following operating parameters can also be changed through the Change Values menu:• Time display and printout units 24-hour or AM/PM.

Ready State Screen

Options Screen

Page 4: Large Steam Sterilizers

4

• Audible signals, end-of-cycle signals, and alarm signals are available through the control and display panel.

• Temperature display and printout units Celsius (°C) or Fahrenheit (°F). Temperature is set, displayed, controlled, and printed to the nearest 0.1°. Recalibration is not required when changing temperature units.

• Pressure/vacuum display and printout units bar or psig/InHg. Recalibration is not required when changing pressure units.

SAFETY FEATURES

Emergency stop button located on the front panel, below the steriliser control touch pad. When pressed, immediately shuts off all outputs on the steriliser. A key is used to reset the switch. Control lockout switch equipped on chamber door(s), senses when door seal is energized and tight against the door. Control prevents cycle from starting until the limit switch signal is received. If control loses appropriate signal during cycle, alarm activates, cycle aborts, and chamber safely vents with a controlled exhaust.Chamber float switch activates alarm, aborts cycle, and safely vents chamber with a controlled exhaust if excessive condensate is detected in the vessel chamber.Pressure relief valve limits the amount of pressure buildup so that the rated pressure in the vessel is not exceeded. The steriliser is provided with two relief valves, one for the steam jacket and one for the chamber. Power door safety feature causes door drive to slip if the sliding door encounters an obstruction during its movement.

CONSTRUCTION

Shell Assembly

Two fabricated Type 316L stainless-steel shells, welded one within the other, form the steriliser vessel. Type 316L stainless-steel end frame(s) is welded to door end. On single door units, back of chamber is fitted with welded, 316L stainless-steel dished head.

Steriliser vessel is ASME and PED rated at 3.1 bar (45 psig) and insulated. Vessel includes one 1"-NPT chamber port for customer use.

Steam-supply opening inside the chamber is shielded by a stainless-steel baffle.

Chamber Door(s)Door is constructed of Type 316L stainless steel.

During cycle operation, door is sealed by a steam-activated door seal. Door seal is constructed of a special long-life rubber compound. When sterilise cycle is complete, the seal retracts under vacuum into a machined groove in the steriliser's end frame.

A proximity switch is used by the control to determine if door is closed. An additional seal pressure switch prevents inad-vertent cycle initiation if door is not sealed.

The door assembly is equipped with a mechanical locking mechanism that ensures the door cannot be opened as long as the seal is intact and energized and more than 0.14 bar (2 psig) pressure is in the chamber.

The steriliser door is fitted with a stainless-steel panel that insulates the operator from the chamber end frame, reducing the chance of accidental contact with a hot metal surface.

Chamber Drain System

Drain system is designed to prevent pollutants from entering into the water-supply system and steriliser.

The automatic condensing system, consisting of a heat exchanger, converts chamber steam to condensate and dis-poses condensate to waste. Cooling water flow is regulated by the waste line RTD to minimize water usage. Water supply shutoff valve is located in the recessed area of the unit.

Vacuum System

Two-stage vacuum pump reduces chamber pressure during prevacuum and post-drying phases. Air is drawn from the chamber through the vacuum system. Following the dry phase, chamber vacuum is relieved to atmospheric pressure by admit-ting air through a bacteria-retentive filter.

Steam Source

Sterilisers are piped, valved, and trapped to receive building-supplied steam delivered at 3.5 to 5.6 bar (50 to 80 psig) dynamic. Standard steam piping is constructed of stainless steel and includes a shutoff valve, steam strainer, and a pressure regulator. An optional, stand-alone electric steam generator is also available.

Piping

All piping is located on a modular plumbing rack (skid). All steam to chamber connections are stainless-steel, sanitary type fittings. Plumbing skid can be located on either side of the steriliser.

MOUNTING ARRANGEMENT

Sterilisers are arranged for either freestanding or recessed installation, as specified. Each steriliser is height-adjustable. Steriliser subframe is equipped with a synthetic rubber gasket to ensure tight fit between the cabinet panels on freestanding units or between the front cabinet panel and wall partition on recessed units.

On freestanding units, stainless-steel side panels and a lou-vered top panel enclose the steriliser body and piping.

ACCESSORIES

Material Handling Accessories include stainless-steel cham-ber tracks and stainless-steel loading cars with painted-steel or stainless-steel carriages. Stainless-steel chamber rack and shelf are available for 914 mm (36") and 991 mm (39") sterilis-ers, only. See separate product literature for details.

Optional Integral Steam Generator is constructed of 316L stainless steel. The steam generator is equipped with three main heating elements (adn one spare heatine element). Gen-erator is operated by steriliser control system. (Refer to ENGI-NEERING DATA / UTILITIES on page 6 for ultilities specifica-tions.)

Page 5: Large Steam Sterilizers

5

PREVENTIVE MAINTENANCE

A global network of skilled service specialists can provide periodic inspections and adjustments to help assure low-cost peak performance. STERIS representatives can provide infor-mation regarding annual maintenance agreements.

NOTES

1. Customer is responsible for backflow protection, if required.

2. Pipe sizes shown indicate terminal outlets only. Building service lines, provided by others, must supply the specified pressures and flow rates.

3. Disconnect switches (with OFF position lockout only; switches not supplied by STERIS) should be installed in electric supply lines near the equipment.

4. Access to the recessing area from the control end of the steriliser is recommended.

5. Clearances shown are minimal for installing and servicing the equipment.

6. Depending on the loading equipment used, additional clearance is required:

• If shelves are used, length of steriliser plus 610 mm (24") at each door (914 mm [36"] and 991 mm [39"] sterilisers, only).

• If loading car and carriage will be used, twice the length of steriliser at each door.

7. Floor drain should be provided within confines of steriliser framework.

UTILITY REQUIREMENTS

CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH APPLICABLE LOCAL AND NATIONAL CODES AND REG-ULATIONS.

Page 6: Large Steam Sterilizers

6

ENGINEERING DATA / UTILITIES

Drain: 2" ODT drain terminal (Floor drain capacity must handle peak water consumption)

Electric: Control and Vacuum Pump: 400 Volt, 3-phase, 50 Hz, 6.0 Amps/phase.

House Steam: 1" Sanitary, 3.5 to 5.6 bar (50 to 80 psig) dynamic, 97% to 100% vapor quality.

Pure Steam Pressure: 3.45 to 5.52 bar (50 to 80 psig), dynamic, condensate free, and 97% to 100% vapor quality. Steam must be made from water free of sulfites, hydrazine, anti-scaling compounds and corrosion control chemicals.

Size: 1" NPT

Consumption:

660 x 660 mm (26 x 26") 991 mm (39") 1245 mm (49") 1549 mm (61"

• Average: 47 kg/hour (104 lb/hr) • Average: 53 kg/hour (118 lb/hr) • Average: 67 kg/hour (148 lb/hr)

• Peak: 116 kg/hour (255 lb/hr) • Peak: 116 kg/hour (255 lb/hr) • Peak: 116 kg/hour (255 lb/hr)

660 x 950 mm (26 x 37-1/2") 914 mm (36") 1219 mm (48") 1524 mm (60")

• Average: 54 kg/hr (120 lb/hr) • Average: 70 kg/hr (156 lb/hr) • Average: 83 kg/hr (185 lb/hr)

• Peak: 116 kg/hr (255 lb/hr) • Peak: 116 kg/hr (255 lb/hr) • Peak: 152 kg/hr (335 lb/hr)

Feed Water Pressure: 1.38 to 3.45 bar (20 to 50 psig), dynamic. Temperature: 21°C (70°F), maximum. Water supply must be deionized, distilled or reverse osmosis water with a minimum resistivity of 1 megOhm/cm.

Consumption

660 x 660 mm (26 x 26") 363 to 492 L/hour (96 to 130 gal/hr)

660 x 950 mm (26 x 37-1/2") 397 to 492 L/hour (105 to 130 gal/hr)

Peak Flow: 57 L/min (15 gal/min)

Size: 3/4" NPT

Compressed Air (CA) 1/4" NPT 5.52 to 8.27 bar (80-120 psig), oil free, dehumidified, 2 cfm

Steriliser Operating Weight

660 x 660 x 991 mm (26 x 26 x 39") – 1250 kg (2756 lb)660 x 660 x 1245 mm (26 x 26 x 49") – 1450 kg (3200 lb)660 x 660 x 1549 mm (26 x 26 x 61") – 1590 kg (3500 lb) 660 x 950 x 914 mm (26 x 37-1/2 x 36") – 1720 kg (3800 lb) 660 x 950 x 1219 mm (26 x 37-1/2 x 48") – 1900 kg (4200 lb)660 x 950 x 1524 mm (26 x 37-1/2 x 60") – 2125 kg (4700 lb)

Optional Integral Steam Generator Water: 3/4” NPT; hot water temperature 60°C (140°F). Optional steam generator water supply must be deionized, distilled or reverse osmosis water with a minimum resistivity of 1 megOhm/cm.

Generator Drain: 1" NPT

Electric: 400V, 50Hz, 86A, 3-phase minimum. 100A circuit breaker recommended. Conductor size AWG #1.

Boiler empty weight: 270 kg (595 lb); boiler operating weight: 363 kg (800 lb)

Page 7: Large Steam Sterilizers

7

NOTE: Each drawing includes chamber length information for the listed configuration (e.g., drawing 129390-153 includes information for 991 mm [39"], 1245 mm [49"] and 1549 mm [61"] chamber lengths, etc.).

Refer to the Following Equipment Drawings for Installation Details

Equipment Drawing Number Equipment Drawing Title

62941-091 General Notes – Applicable to Sterilizer Equipment Drawings

129390-151 660 x 660 mm (26 x 26") Evolution Steam Ster. Double Sliding Door Recessed Two Walls Electric Heat with Steam Generator

129390-152 660 x 660 mm (26 x 26") Evolution Steam Ster. Double Sliding Door Recessed Two Walls Steam Heat

129390-153 660 x 660 mm (26 x 26") Evolution Steam Ster. Single Sliding Door Recessed One Wall Electric Heat with Steam Generator

129390-154 660 x 660 mm (26 x 26") Evolution Steam Ster. Single Sliding Door Recessed One Wall Steam Heat

129390-155 660 x 660 mm (26 x 26") Evolution Steam Ster. Single Sliding Door Cabinet Steam Heat

129390-156 660 x 660 mm (26 x 26") Evolution Steam Ster. Single Sliding Door Cabinet Electric Heat with Steam Generator

129390-157 660 x 660 mm (26 x 26") Evolution Steam Ster. Double Sliding Door Recessed One Wall Steam Heat

129390-158 660 x 660 mm (26 x 26") Evolution Steam Ster. Double Sliding Door Recessed One Wall Electric Heat with Steam Generator

129390-159 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Double Sliding Door Recessed Two Walls Electric Heat with Steam Generator

129390-160 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Double Sliding Door Recessed Two Walls Steam Heat

129390-161 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Single Sliding Door Recessed One Wall Electric Heat with Steam Generator

129390-162 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Single Sliding Door Recessed One Wall Steam Heat

129390-163 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Single Sliding Door Cabinet Steam Heat

129390-164 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Single Sliding Door Cabinet Electric Heat with Steam Generator

129390-165 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Double Sliding Door Recessed One Wall Steam Heat

129390-166 660 x 950 mm (26 x 37-1/2") Evolution Steam Ster. Double Sliding Door Recessed One Wall Electric Heat with Steam Generator

Page 8: Large Steam Sterilizers

8

Recessed, Two Walls – 660 x 660 mm Double-Door Configuration

NOTE: Refer to table on page 6 for engineering data and utilities

Plan View

Service Clearance(see Table 1, page 9)

1321 mm(52")

724 mm(28-1/2")

699 mm(27-1/2")

190 mm(7-7/16")

114 mm Typical(4-1/2")

1956 mm(77") ± 6 mm (±1/4")

1130 mm(44-1/2")

2045 mm(80-1/2")

Finished Floor

Drawing not to scale. Dimensions are typical.

1270 mm (50")± 6 mm (±1/4")

Typical Wall Opening

Side View

102 - 203 mm (4 - 8") Typical Both Walls

Operating EndNon-

OperatingEnd

Page 9: Large Steam Sterilizers

9

Table 1. 660 x 660 Steriliser Service Clearance (Dimension "S")

Chamber Length Recessed, Two Walls Recessed, One Wall Cabinet Enclosed

991 mm (39") N/A 1626 mm (64") 1374 mm (54-78")

1245 mm (49") 1422 mm (56") 1880 mm (74") 1648 mm (64-78")

1549 mm (61") 1727 mm (68") 2184 mm (86") 1953 mm (76-7/8")

Recessed, One Wall – 660 x 660 mm Single-Door Configuration

NOTE: Refer to table on page 6 for engineering data and utilities

Drawing not to scale. Dimensions are typical.

1321 mm(52")

724 mm(28-1/2")

699 mm(27-1/2")

114 mm Typical(4-1/2")

1956 mm (77")± 6 mm (±1/4")

1130 mm(44-1/2")

2045 mm(80-1/2")

Finished Floor

1270 mm(50") ± 6 mm (±1/4")

Typical Wall Opening

Side View

102 - 203 mm (4 - 8") Typical Both Walls

Operating End

Service Clearance (see Table 1)

Page 10: Large Steam Sterilizers

10

Cabinet Enclosed – 660 x 660 mm Single-Door Configuration

NOTE: Refer to table on page 6 for engineering data and utilities

Drawing not to scale. Dimensions are typical.

Plan View

1524 mm(60")

1321 mm(52")

699 mm(28-1/2")

114 mm Typical(4-1/2")

1130 mm(44-1/2")

2045 mm(80-1/2")

Finished Floor

Side View

Operating End

Service Clearance(see Table 1, page 9)

Page 11: Large Steam Sterilizers

11

Table 2. 660 x 950 Steriliser Service Clearance (Dimension "S")

Chamber Length

Double DoorRecessed, One Wall

Double DoorRecessed, Two Walls

Single DoorRecessed One Wall

914 mm (36") N/A N/A 1816 mm (71-1/2")

1219 mm (48") 1657 mm (65-1/4") 1657 mm (65-1/4") 2121 mm (83-1/2")

1524 mm (60") 1962 mm (77-1/4") 1962 mm (77-1/4") 2425 mm (95-1/2")

Recessed, Two Walls – 660 x 950 mmDouble-Door Configuration

NOTE: Refer to table on page 6 for engineering data and utilities

Plan View

Service Clearance(see Table 2, page 11)

1853 mm(73")

733 mm(28-7/8")

708 mm(27-7/8")

114 mm Typical(4-1/2")

1886 mm (74-1/4") ± 6 mm (±1/4")

1067 mm(42")

1962 mm(77-1/4")

Finished Floor

Drawing not to scale. Dimensions are typical.

1803 mm (71")± 6 mm (±1/4")

Typical Wall Opening

Side View

102 - 203 mm (4 - 8") Typical Both Walls

OperatingEndNon-

OperatingEnd

Page 12: Large Steam Sterilizers

For Further Information, contact:

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 • USA440-354-2600 • 800-548-4873www.steris.com

This document is intended for the exclusive use of STERIS customers, including architects or designers. Reproduction in whole or in part by any party other than a customer is prohibited.SD862 ©2007, STERIS Corporation. All rights reserved. (07/01/07)

EC Authorized Representative

STERIS Ltd.

STERIS House

Jays Close

Viables

Basingstoke

Hampshire

RG22 4AX

Recessed, One Wall – 660 x 950 mm Double-Door Configuration

NOTE: Refer to table on page 6 for engineering data and utilities

Drawing not to scale. Dimensions are typical.

Plan View

1853 mm(73")

1803 mm (71")± 6 mm (±1/4")

708 mm(27-7/8")

114 mm Typical(4-1/2")

1067 mm(42")

1885 mm (74-1/4")± 6 mm (±1/4")

Finished Floor

Side View

OperatingEnd

Service Clearance(see Table 2, page 11)

Non-Operating

End

733 mm(28-7/8")

Wall Opening

1962 mm(77-1/4") Typical

114 mm(4-1/2")

Typical Wall Opening

102 - 203 mm (4 - 8") Typical Both Walls

Page 13: Large Steam Sterilizers

LOADING EQUIPMENT FORAMSCO® EVOLUTION™ STEAM STERILIZERS

Stationary Loading Car Stand

Loading Car and Transfer Carriage

Typical only – details may vary

Application

For loading and unloading 660 x 660 mm (26 x 26”) or 660 x 950 mm (26 x 37.5") sterilizers, and for transferring goods to and from processing areas. This equipment allows heavy loads to be transferred with low effort. Equipment design minimizes contact between operator and hot load items.

The need for protective gloves or clothing is reduced, as little or no direct contact occurs between operator and load. One or more loading transfer carriages can be used for handling loads with multiple sterilizers and load preparation areas.

IMPORTANT: This tech data sheet is intended for use outside the U.S.A. only.

Description

Loading car is a welded framework constructed of stainless steel.

Shelves – each loading car is provided with either two or three stainless-steel shelves. Shelf arrangement depends on size of sterilizer:

• 660 x 660 mm sterilizer loading cars feature two shelves: one height-adjustable, removable shelf (upper) and one fixed-height shelf (lower);

• 660 x 950 mm sterilizer loading cars feature three shelves: two height-adjustable, removable shelves (upper and middle) and one fixed-height shelf (lower).

No tools are required for adjusting shelf heights.

Transfer carriage is constructed of welded stainless steel with four swiveling casters. Casters are equipped with long-wear-ing synthetic-material tires. Rear casters are provided with a locking mechanism. Transfer carriage is height-adjustable to align with sterilizer chamber. Transfer carriage is equipped with a hand wheel for purposes of lowering or raising guide rails under loading car in sterilizer chamber.

During loading and unloading, transfer carriage is held to the sterilizer by a floor-level docking interface that releases trans-fer carriage after car is fully loaded (lowered) into chamber (unloading process is similar). Loading car is aligned upon the transfer carriage by two guide rails.

Transfer carriage is available constructed from acrylic-coated mild steel for use at the sterilizer’s clean side, or is also available constructed from brushed stainless steel for use at the sterilizer’s sterile side.

The Selections Checked Bel

❑ Sterilizer Chamber Size

❑ 660 x 660 x 991 mm(26 x 26 x 39")

❑ 660 x 660 x 1245 mm(26 x 26 x 49")

❑ 660 x 660 x 1549 mm(26 x 26 x 61")

❑ 660 x 950 x 914 mm(26 x 37-1/2 x 36")

❑ 660 x 950 x 1219 mm(26 x 37-1/2 x 48")

❑ 660 x 950 x 1524 mm(26 x 37-1/2 x 60")

❑ Transfer Carriage

❑ Stainless Steel❑ Acrylic-Coated M

❑ Docking Interface

❑ Loading Car Shelves

❑ One, for 660 x 66(26 x 26") Sterili

❑ Two, for 660 x 9(26 x 37-1/2") St

❑ Inter-Shelf Condens

The separately-available docking interface is installed for use at the bottom of the sterilizer paneling.

Stationary loading car stand allows staging of load near preparation and packing areas or loading/unloading area. Stand is available constructed of either stainless steel or mild steel.

Inter-Shelf Condensate Shield – is an optional item that can be installed directly below the upper (and/or middle) shelf. The shield is designed to deflect falling condensate droplets from reaching items located on lower shelves.

SD869 (07/01/07)

Item ____________

Location(s)_______

______________

ow Apply To This Equipment

ild Steel

0 mm zers50 mmerilizers

ate Shield

❑ Stationary Loading Car Stand

❑ Stainless Steel ❑ Mild Steel

Page 14: Large Steam Sterilizers

Table 1. Loading Equipment Capacity

Sterilizer Model SterilizerChamber Size

Approximate Shelf Length/Number of Shelves

Sterilization Modules per Shelf

Total Modules when All Shelves Used

HC-600 660 x 660 x 991 mm(26 x 26 x 39") 991 mm (39") / 2 3 6

HC-800 660 x 660 x 1245 mm(26 x 26 x 49") 1245 mm (49") / 2 4 8

HC-1000 660 x 660 x 1549 mm(26 x 26 x 61") 1549 mm (61") / 2 5 10

HC-900 660 x 950 x 914 mm(26 x 37-1/2 x 39") 914 mm (36") / 3 3 9

HC-1200 660 x 950 x 1219 mm(26 x 37-1/2 x 48") 1219 mm (48") / 3 4 12

HC-1500 660 x 950 x 1524 mm(26 x 37-1/2 x 60") 1524 mm (60") / 3 5 15

Table 2. Loading Equipment Specifications(See Figure on Facing Page)

Loading Carriage and Car

for use in Sterilizer Model

A

(Transfer Carriage Height)

B

(Loading Car Shelf Dimensions)

C

(Loading Car Height)

HC-600

800 mm (32")

602 x 956 mm(23-11/16 x 37-5/8")

505 mm (19-7/8")HC-800 602 x 1211 mm(23-11/16 x 47-11/16")

HC-1000 602 x 1540 mm (23-11/16 x 60-5/8")

HC-900

590 mm (23")

602 x 1037 mm(23-11/16 x 40-13/16")

778 mm (30-5/8")HC-1200 602 x 1210 mm(23-11/16 x 47-5/8")

HC-1500 602 x 1664 mm(23-11/16 x 65-1/2")

2

Page 15: Large Steam Sterilizers

Hand Wheel

Loading Car Guide Rail

B

A

Loading Car

B

Caster

Docking Interface

C

Loading Equipment for Amsco Evolution Steam Sterilizers

Transfer Carriage

3

Page 16: Large Steam Sterilizers

For Further Information, contact:

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 ? USA440-354-2600 ? 800-548-4873www.steris.com

This document is intended for the exclusive use of STERIS customers,including architects or designers. Reproduction in whole or in part

by any party other than a customer is prohibited.SD869 © 2007, STERIS Corporation. All rights reserved. (07/01/07)