Lab Scan

Version 1.0 | 990487 - Y9500 | September 2016

INSTRUCTION FOR USE

Page 2: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

22 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

SAFETY INSTRUCTIONS 3

INTENDED USE 4

TRANSPORT AND PACKAGING 4

IDENTIFICATION OF PARTS AND ACCESSORIES 5

USE OF EXTERNAL EQUIPMENT 7

HARDWARE SETUP 7

FIRST USE 12

OPERATION 13

ACCESSORIES 21

CARE OF THE UNIT 43

TROUBLESHOOTING 44

DISPOSAL OF THE UNIT 46

WARRANTY AND REPAIR 46

MAINTENANCE AND SERVICE 46

TECHNICAL SPECIFICATIONS 47

Prior to use, carefully read these instructions

CONTENTS

Page 3: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

33Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

CAUTION

• Follow the instructions in order to use the unit correctly and safely. Keep these instructions within easy reach of the device.• Network/UBS/Power connection may not be plugged out when the machine is ON.• Install the scanner on an appropriate, level and stable working surface, free from vibrations.• Only qualified dental personnel should use the scanner.• Prior to use, make sure that the scanner is working correctly. The unit should be recalibrated once it has been moved or repositioned• The Aadva Lab Scan is designed for use on dental models only and should not be used for anything else

other than its intended purpose. • Do not touch the device while the tilt and swivel unit is in motion. This may cause scan malfunctions, accidental squeezing of the fingers or damage

to the scanning unit itself.• Do not look directly into the light emitted by the scanning device. This may cause physical discomfort.• The unit should not be handled with wet hands. This may cause equipment malfunction.• The unit should not be exposed to direct sunlight, dust, wet environment or placed near a heater.• It is expressly prohibited to open, disassemble or remodel the unit, or use the device in a manner not

covered by these instructions, as this will lead to the termination of the warranty, guarantee and liability of the manufacturer.

• Touching/handling the sensor head may cause the unit to malfunction. • Additionally installed software can impair the operation of the system.• No additional software can be installed on the computer of the Aadva Lab Scan. The system settings on the computer of the Aadva Lab Scan may not be changed. The exceptions to this rule are the date, time, time zone and country settings.

The internet connection (minimum 128Kbs speed and ideally about 520Kbs) should be connected directly to the PC via a RJ45 crossover cable. The internet connection must be dedicated for the functionality of the scanner (including system upgrades).

WARNING • Do not use the unit if the user has a pacemaker. This device may interfere with the pacemaker.• Do not look direct into the LED light source. • Do not plug / unplug the device with wet hands. This may cause an electric shock.• If the unit should begin to smoke, emit an unusual odour, deform or discolour during scanning, calibration or storage, discontinue use immediately. • Avoid sudden or strong impact to the device. This may cause the unit to breakdown or lead to errors in the scanning function of the device. • Keep the unit away from flammable materials.• Do not store or operate the scanning device in areas where ceramic dust is produced. This may interfere with the scanning functionality and/or cause damage to the scanning device. • Do not place the scanner where liquids may accidentally get poured on it. Do not place any liquid close

to the scanner.• Do not place or store objects on top of the scanner. • Do not use the accompanying USB license dongles for data storage. This may interfere with the correct licensing and functioning of the Aadva Lab Scan.

SAFETY INSTRUCTIONS

Aadva® Lab Scan

Page 4: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

44 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

The Aadva Lab Scan is a structured light scanner for creating digital 3D dental models by scanning dental models which can then be used in the (CAD/CAM) design, manufacture and production of dental prosthesis. The device uses blue LED light and has a scan time of 980 milliseconds, making it one of the fastest dental scanners available. The included software also allows the user to directly upload edited 3D models for CAD/CAM processing and manufacture (for detailed indications see page 4, 5, 6).

The PC component of the AadvaScanner constitutes an integral part of the AadvaScanner and may not be used for any purposes other than for the functionality of the AadvaScanner itself. No components (whether hardware and/or software) and/or other applications may be changed, added or removed.

GC Europe n.v. and/or the GC Europe distributors and/or service partners approved and appointed by GCE are responsible for delivering and installing patches, fixes, service packs and software upgrades related to the installed software in order to guarantee the proper functioning of the AadvaScanner and in accordance with the maintenance contract concluded if the latter is applicable.

GC Europe n.v. is not deemed to have knowledge of or to take account of the specific use that will be made by the customer of the purchased goods, and thus GC Europe n.v. cannot be held liable for this. The customer is expected to use the purchased good in accordance with its intended use.

INTENDED USE

Prior to setting up and assembling the Aadva Lab Scan, please make sure that all parts and components are included in your delivery (refer to the accompanying delivery note). Also, check for any visible signs of damage which the device may have incurred during transport.

CONDITIONNEMENT

The Aadva Lab Scan is shipped in a standard Euro pallet. For the safe disposal and recycling of the packaging, please ensure that all packing materials are properly disposed of according to current recycling directives. In case the device has to be returned, please use the original packing material for re-packaging, as only this would ensure safe transport.

CONTENTS Quantity 1. Aadva Lab Scan 1 2. Workstation PC 1 3. Mouse 1 4. Keyboard 1 5. Power Cord 3 6. USB Cable Type A/B 1 7. LAN Cable 1

8. Hardware Dongle 1 9. Software Dongle 1 10. Model Holder Low 1 11. Model Holder High 1 12. Single Die Holder 1 13. Calibration Set 1 14. Software and documentation 1

TRANSPORT AND PACKAGING

Page 5: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

55Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

IDENTIFICATION OF PARTS AND ACCESSORIES

Correct and safe operation of the Aadva Lab Scan is only possible with original parts and accessories delivered by GC Europe and with the indicated compliant external equipment. For any use of other parts or accessories or non-compliant external equipment the user assumes the risks.

1. Aadva Lab Scan

4. ON/OFF Switch and Power Inlet on Back Panel of Aadva

Lab Scan

2. Aadva Lab Scan Power Cord

5. LAN Port on Back Panel of Aadva Lab Scan

3. Aadva Lab Scan Sensor Head

6. USB Port on Back Panel of Aadva Lab Scan

LIFTING AND CARRYING

The most practical method of moving and placing the Aadva Lab Scan will be by handling manually. Be aware of your personal limits or seek for assistance, taking account the weights of the scanner and the PC indicated in the specifications.

Aadva® Lab Scan

Page 6: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

66 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

13. Hardware Dongle 14. Model Holder Low / High 15. Keyboard

Lab Scan from GC

INSTRUCTION FOR USE

Version 1.2 | V4902 | January 2016

16. Calibration set 17. Single Die Holder 18. Documentation and Software

7. Workstation 8. Workstation Power Cord 9. Mouse

10. USB Cable Type A/B 11. LAN Cable 12. Software Dongle

Page 7: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

77Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

The Aadva Lab Scan requires the use of external equipment for proper operation. These devices must be compliant with the following standards: DIN/EN/IEC 60950, UL 60950 and CAN/CSA-C22.2NO. 60950-00. Our software is compatible with Dell and HP monitors. The customer is required to provide the following external equipment:

1. Monitor 2. Monitor Power Cord 3. VGA Monitor Cable

USE OF EXTERNAL EQUIPMENT

1. Install the Aadva Lab Scan on a flat and stable place, in a medical room or laboratory. The Aadva Lab Scan should not be located in the vicinity of patients.

The scanner will require a level surface area measuring approximately 400 x 300 mm. 2. Connect the power cord to the main power outlet.

3. Connect the LAN cable to the network plug on the Aadva Lab Scan. Then, insert Type B connector of the USB Cable into the USB port on the back panel

of the Aadva Lab Scan. 4. Install the monitor and workstation on a flat and stable place, near the Aadva Lab Scan.

HARDWARE SETUP

Aadva® Lab Scan

Page 8: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

88 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

5. Connect the power cords of the monitor and the workstation to the main power outlet (110-240V AC).

6. Connect the LAN cable to the network plug on the workstation. Then, insert the Type A connector of the USB Cable into a USB port on the workstation.

USB Cable Type A/B

Page 9: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

99Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

7. Connect the VGA cable to the VGA ports on the monitor and the workstation.

LAN Cable

Aadva® Lab Scan

Page 10: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1010 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

8. Insert the software and hardware dongle into available USB ports on the workstation (preferably at the back of the PC in order to avoid any damage).

Hardware Dongle

Software Dongle

Page 11: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1111Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

External LAN

10. Turn on the workstation and monitor. Once prompted, choose USER 1. No password is needed.

11. Do not pull out any of the connection cables while the device is still on.

9. External LAN (internet, network, …)

Aadva® Lab Scan

Page 12: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1212 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn on and the GC logo on the front panel will light up.

After a few seconds, the blue light from the LED in the sensor head will be visible. Switch on the PC.

2. Once the workstation and monitor are switched on, check to see if the LEDs on the license dongles blink a few times.

3. Start the Dental manager software via the application shortcut on the desktop.

4. Once the Aadva Lab Scan and the scanning software are running correctly, the Aadva Lab Scan must be calibrated again. This may be necessary if the Aadva Lab Scan has been transported in a horizontal position or if the scanner has undergone rough transport. The Aadva Lab Scan may have to be recalibrated even if the scan software has not indicated that calibration is necessary.

5. Make sure the calibration plate is clean before starting the different calibration steps. Use a dry and soft tissue to clean.

FIRST USE

Page 13: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1313Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

This section covers the basic operation of the Aadva Lab Scan and the scanning software. For a detailed description of all the features of the scanning software, please refer to the software documentation (http://wiki.exocad.com/). This document is permanently available on the web.

1. Start the dental manger software via the application shortcut on the desktop.

2. Once the application starts, the user has to define a scan profile, or select from a list of stored profiles.

3. To define a scan profile, the following information has to be entered:

a. Customer ID and Nameb. Patient ID and Namec. Reconstruction Type d. Technician ID and Name (optional)

The first three settings have to be entered, otherwise the Save button is deactivated and the user cannot start with the scan.

OPERATION

Aadva® Lab Scan

Page 14: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1414 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

4. Click on the Edit button in the Customer section or select an existing customer from the drop-down menu.

5. Select an existing customer or the Default customer. Alternatively, click on the New Entry button to define a new customer. Click on Save once all details have been entered. Then, click on the Close button to exit the window.

6. You can define the patient information under the Job section. Click on the Edit button to define a new patient or select an existing patient in the drop-down menu.

7. Select an existing patient or the Default patient. Alternatively, click on the New Entry button to define a new patient. Click on Save once all details have been entered. Then, click on the Close button to exit the window.

Page 15: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1515Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

8. Now, you can select the type of reconstruction for which the scan is to be made. To do this, first select the relevant position/s in the dental model and click on the corresponding number/s.

9. Once a position has been selected, define the reconstruction type from the list which appears. Click on the Save button to proceed.

For this operational guide, a scan of tooth 16 for an anatomic crown reconstruction type will be used. For a full listing of various scans and reconstruction type, please refer to the software documentation (http://wiki.exocad.com/).

Aadva® Lab Scan

Page 16: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1616 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

10. After entering the necessary information for customer, patient and reconstruction type, you can also enter the technician information. This information is not mandatory, but can be useful in identifying various scan profiles.

Click on the Edit button to start defining the technician information, or select an existing technician from the drop-down menu.

11. Select an existing technician or the Default technician. Alternatively, click on the New Entry button to define a new technician. Click on Save once all details have been entered. Then, click on the Close button to exit the window.

12. Now that the information for customer, patient and reconstruction type has been entered, the scanner software will allow you to save the profile.Click on the Save button to save this scan profile.

13. After a scan profile has been defined, you can then continue with performing a scan or single scan. Click on the appropriate buttons to start scan.

Page 17: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1717Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

14. The scan software will now connect to the Aadva Lab Scan. An image of what the sensor head finds on the scanner is displayed immediately.

If there are red areas on the dental model, you can adjust the exposure time using the adjustment setting in the lower left hand section of the sensor head image (marked by a white margin in the image below).

Select the upper box to adjust the exposure time for measurements

Select the lower box to adjust the exposure time for reference points

After selecting the exposure setting, simply hold the mouse over the image and use the scroll wheel on the mouse to make adjustments. Moving the wheel up increases the exposure time, moving it down decreases it.

To see the exact exposure time in milliseconds, click and drag the image of the wheel instead.

15. Besides adjusting the exposure setting, you can click on the dental model directly and use the scroll wheel on the mouse to zoom in and out of the object.

Aadva® Lab Scan

Page 18: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1818 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

16. After adjusting the exposure and zoom levels, you should have an image of the dental model with all relevant positions clearly visible and without red markings, similar to the image below.

17. You are now ready to start the preparation scan. Click on the Next button to start the scan.

18. After cropping and deleting various regions from the preparation scan, you will have an image of the dental model free of any objects or parts not relevant for the current scan job, similar to the image below (keep only the relevant parts of the scan).

The scanner can now proceed with a more detailed scan.Click on the Next button to continue.

Page 19: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

1919Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

19. The Aadva Lab Scan can make several scans, depending on the angle and the level of detail which is required for the reconstruction type. For the anatomic crown reconstruction of tooth 16, the digital image has been cropped and trimmed down to the single tooth.

20. If the scanned image contains incomplete data (e.g. holes in the model surface), the scan software allows you the option to select and close any holes in the image.

Aadva® Lab Scan

Page 20: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2020 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

21. Once the dental model has been scanned sufficiently, the user is given the option to save and close the scan wizard or to start the restoration design.

22. After scanning and/or designing, the dental technician has the option to upload the scan for further CAD/CAM processing.

Page 21: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2121Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

1. SINGLE DIE HOLDER

1. Start the dental manger software via the application shortcut on the desktop.

2. How to use : - Apply an amount of putty in the central part of the holder without covering any of the reference points.- Fix the die or other scanable object firmly in the putty reassuring that the object will not move

during scanning.

2. MULTI DIE HOLDER 5 UNITS

1. Indication : scanning of up to 5 dies / abutments of different patients

2. How to use : - apply an amount of putty in the 5 pits of the holder without covering any of the reference points.- Fix the dies or scanable objects firmly in the holder reassuring that this object will not move during

scanning.

ACCESSORIES

Aadva® Lab Scan

Page 22: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2222 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

- In Aadva Manager select ‘Multiedie mode’ to visualize the holder and create the project as usual in Exocad.

Aadva LAB SCAN GC 22 Version 1.21 V4903l J u l y 2016

 

 

 1) Indication  :  scanning  of  up  to  5  dies  /  abutments  of  different  patients  

 

2) How  to  use  :  -­‐  apply  an  amount  of  putty  in  the  5  pits  of  the  holder  without                covering  any  of  the  reference  points.  

 -­‐ Fix  the  dies  or  scanable  objects  firmly  in  the  holder  reassuring  

that  this  object  will  not  move  during  scanning.  

                                                       

-­‐ In  Aadva  Manager  select  ‘Multiedie  mode’  to  visualize  the  holder  and  create  the  project  as  usual  in  Exocad.        

     

 

 

3) Multi  die  holder  10  units  :  

UnknownFormatted: Font:12 ptUnknownFormatted: Font:12 pt

Page 23: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2323Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

Aadva LAB SCAN GC 23

Aadva® Lab Scan  

 

                                                                                                                                                                         1) Indication  :  scanning  of  up  to  10  dies  /  abutments  (different  patients)  

 2) How  to  use  :-­‐  apply  an  amount  of  putty  into  the  10  pits  of  the  holder  

without  covering  any  of  the  reference  points.    

-­‐    Fix  the  dies  or  scanable  objects  firmly  into  the  holder  reassuring  that  these  objects  will  not  move  during  scanning  

                                                                 

-­‐ In  Aadva  Manager  select  ‘Multidie  mode’  to  visualize  the  holderd  and  create  the  project  as  usual  in  Exocad.    

               

4) Model  holder  low  :  

UnknownFormatted: Font:16 pt

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 ptUnknownFormatted: Font:12 pt

3. MULTI DIE HOLDER 10 UNITS

1. Indication : scanning of up to 10 dies / abutments (different patients)

2. How to use :- apply an amount of putty into the 10 pits of the holder without covering any of the reference points. - Fix the dies or scanable objects firmly into the holder reassuring that these objects will not move during

scanning

- In Aadva Manager select ‘Multidie mode’ to visualize the holderd and create the project as usual in Exocad.

Page 24: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2424

Aadva LAB SCAN GC 24 Version 1.21 V4903l J u l y 2016

 

 

                                                                                                                                                                   

1) Indication  :  scanning  of  dental  models    

2) How  to  use  :  -­‐  fix  the  model  firmly  and  as  much  to  the  back  as  possible  

                                                                                                                                                             -­‐ Screw  the  hexagonal  scanning  dot  onto  the  model  holder  in  

any  desired  position  so  that  the  reference  points  are  situated  just  above  the  occlusal  plane.  

-­‐ Execute  the  scan  as  usual  in  Exocad.    

                                                       

5) Model  holder  high  :  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 pt

Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

4. MODEL HOLDER LOW

1. Indication : scanning of dental models

2. How to use : - fix the model firmly and as much to the back as possible

- Screw the hexagonal scanning dot onto the model holder in any desired position so that the reference points are situated just above the occlusal plane.

- Execute the scan as usual in Exocad.

Page 25: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2525

Aadva LAB SCAN GC 24 Version 1.21 V4903l J u l y 2016

 

 

                                                                                                                                                                   

1) Indication  :  scanning  of  dental  models    

2) How  to  use  :  -­‐  fix  the  model  firmly  and  as  much  to  the  back  as  possible  

                                                                                                                                                             -­‐ Screw  the  hexagonal  scanning  dot  onto  the  model  holder  in  

any  desired  position  so  that  the  reference  points  are  situated  just  above  the  occlusal  plane.  

-­‐ Execute  the  scan  as  usual  in  Exocad.    

                                                       

5) Model  holder  high  :  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 pt

Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

5. MODEL HOLDER HIGH

1. Indication : scanning of dental models

2. How to use : - fix the model firmly and as much to the back as possible

- Screw the hexagonal scanning dot onto the model holder in any desired position so that the reference points are situated just above the occlusal plane.

- Execute the scan as usual in Exocad.

Page 26: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2626 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

6. MODEL PLATE WITH PUTTY

1. Indication : scanning of dental models or other dental objects that, due to their size, do not fit in the model holder low or model holder high.

2. How to use : - Fix the model with some putty in such a way that it is central to the camera of the scanner and that the

reference points of the hexagonal scanning dot are also captured by this camera. Execute the scan as usual in Exocad.

Aadva LAB SCAN GC 26 Version 1.21 V4903l J u l y 2016

 

 

                                                                 

1) Indication  :  scanning  of  dental  models  or  other  dental  objects  that,  due  to  their  size,  do  not  fit  in  the  model  holder  low  or  model  holder  high.    

2) How  to  use  :  -­‐  Fix  the  model  with  some  putty  in  such  a  way  that  it  is  central  to  the  camera  of  the  scanner  and  that  the  reference  points  of  the  hexagonal  scanning  dot  are  also  captured  by  this  camera.    Execute  the  scan  as  usual  in  Exocad.  

                                                         

   

7) Impression  model  holder  :  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

Page 27: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2727Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

7. IMPRESSION MODEL HOLDER

1. Indication : Scanning of physical dental impressions

2. How to use : - Assemble the impression model holder onto a regular model holder low by removing the anterior retention

screws and using the posterior retention screws to fix the impression holder.

- Insert the impression by means of the impression tray handle into the holder and fix with the black fixation screw on top of the impression holder. Apply scan spray if necessary to scan silicon impression.

Remove antorior retention pins

Fix impression holder with posterior retention pins

Fix retention pins into the retention holes of the impression holder

Page 28: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2828 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

- After having created the Exocad project, open Aadva Scanner with ‘Impression scan’, as indicated in image below.

- Make sure to select ‘Impression’ in the Scan Wizard.

Aadva LAB SCAN GC 28 Version 1.21 V4903l J u l y 2016

 

 

-­‐ After  having  created  the  Exocad  project,  open  Aadva  Scanner  with  ‘Impression  scan’,  as  indicated  in  image  below.    

                                                                                     -­‐ Make  sure  to  select  ‘Impression’  in  the  Scan  Wizard.  

               

                                         

 

 

 

 

 

8)                      Wax  up  frame  :    

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:Bold, Underline

Aadva LAB SCAN GC 28 Version 1.21 V4903l J u l y 2016

 

 

-­‐ After  having  created  the  Exocad  project,  open  Aadva  Scanner  with  ‘Impression  scan’,  as  indicated  in  image  below.    

                                                                                     -­‐ Make  sure  to  select  ‘Impression’  in  the  Scan  Wizard.  

               

                                         

 

 

 

 

 

8)                      Wax  up  frame  :    

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:Bold, Underline

Page 29: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

2929Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

8. WAX UP FRAME

1. Indication : Scanning (both sides) of dental objects, like diagnostic wax-ups, dentures, provisional crowns, etc.

2. How to use : - Before using the wax up frame it is necessary to do the wax up frame calibration process in ‘Kiosk mode’ :

1) Open ‘Kiosk mode’ in Aadva Manager

Aadva LAB SCAN GC 29

Aadva® Lab Scan  

 

 

1) Indications  :  Scanning  (both  sides)  of  dental  objects,  like  diagnostic  wax-­‐ups,  dentures,  provisional  crowns,  etc.  

2) How  to  use  :  -­‐  Before  using  the  wax  up  frame  it  is  necessary  to  do  the  wax  up  frame  calibration  process  in  ‘Kiosk  mode’  :  

 1) Open  ‘Kiosk  mode’  in  Aadva  Manager  

 2) Select  ‘Waxup  frame  calibration’  and  fill  in  the  serial  number  of  

the  waxup  frame  holder  (situated  on  the  bottom)  

                   

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

Minale, Mario 23/6/16 16:40Formatted: Dutch

Aadva LAB SCAN GC 29

Aadva® Lab Scan  

 

 

1) Indications  :  Scanning  (both  sides)  of  dental  objects,  like  diagnostic  wax-­‐ups,  dentures,  provisional  crowns,  etc.  

2) How  to  use  :  -­‐  Before  using  the  wax  up  frame  it  is  necessary  to  do  the  wax  up  frame  calibration  process  in  ‘Kiosk  mode’  :  

 1) Open  ‘Kiosk  mode’  in  Aadva  Manager  

 2) Select  ‘Waxup  frame  calibration’  and  fill  in  the  serial  number  of  

the  waxup  frame  holder  (situated  on  the  bottom)  

                   

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

UnknownFormatted: Font:12 pt, Font color:Custom Color(RGB(0;176;80))

Minale, Mario 23/6/16 16:40Formatted: Dutch

Page 30: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3030 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

2) Select ‘Waxup frame calibration’ and fill in the serial number of the waxup frame holder (situated on the bottom)

3) Scan the waxup frame one sided

4) Check the reference points of the scan and click ‘OK’

Aadva LAB SCAN GC 30 Version 1.21 V4903l J u l y 2016

 

 

 

3) Scan  the  waxup  frame  one  sided

 4) Check  the  reference  points  of  the  scan  and  click  ‘OK’  

 

5) Rotate  the  upper  part  of  the  wax  up  frame,  so  that  the  arrows  are  again  facing  each  other.    

Aadva LAB SCAN GC 30 Version 1.21 V4903l J u l y 2016

 

 

 

3) Scan  the  waxup  frame  one  sided

 4) Check  the  reference  points  of  the  scan  and  click  ‘OK’  

 

5) Rotate  the  upper  part  of  the  wax  up  frame,  so  that  the  arrows  are  again  facing  each  other.    

Aadva LAB SCAN GC 30 Version 1.21 V4903l J u l y 2016

 

 

 

3) Scan  the  waxup  frame  one  sided

 4) Check  the  reference  points  of  the  scan  and  click  ‘OK’  

 

5) Rotate  the  upper  part  of  the  wax  up  frame,  so  that  the  arrows  are  again  facing  each  other.    

Page 31: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3131Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

5) Rotate the upper part of the wax up frame, so that the arrows are again facing each other.

6) Check the alignment of the reference points and dlick ‘OK’

7) Approve component if alignment is ok. The waxup frame is now calibrated and will be recognized when scanning.

- When creating the project, it is only possible to choose a job definition in ‘Digital copy milling’ as shown below.

Aadva LAB SCAN GC 31

Aadva® Lab Scan  

 

 6) Check  the  alignment  of  the  reference  points  and  dlick  ‘OK’  

 

7) Approve  component  if  alignment  is  ok.    The  waxup  frame  is  now  calibrated  and  will  be  recognized  when  scanning.    

-­‐ When  creating  the  project,  it  is  only  possible  to  choose  a  job  definition  in  ‘Digital  copy  milling’  as  shown  below.  

 -­‐ Before  scanning  check  the  box  ‘Scan  waxup  both  way’  in  order  to  

generate  the  additional  scan  of  the  object  in  the  waxup  frame.    

Aadva LAB SCAN GC 31

Aadva® Lab Scan  

 

 6) Check  the  alignment  of  the  reference  points  and  dlick  ‘OK’  

 

7) Approve  component  if  alignment  is  ok.    The  waxup  frame  is  now  calibrated  and  will  be  recognized  when  scanning.    

-­‐ When  creating  the  project,  it  is  only  possible  to  choose  a  job  definition  in  ‘Digital  copy  milling’  as  shown  below.  

 -­‐ Before  scanning  check  the  box  ‘Scan  waxup  both  way’  in  order  to  

generate  the  additional  scan  of  the  object  in  the  waxup  frame.    

Aadva LAB SCAN GC 31

Aadva® Lab Scan  

 

 6) Check  the  alignment  of  the  reference  points  and  dlick  ‘OK’  

 

7) Approve  component  if  alignment  is  ok.    The  waxup  frame  is  now  calibrated  and  will  be  recognized  when  scanning.    

-­‐ When  creating  the  project,  it  is  only  possible  to  choose  a  job  definition  in  ‘Digital  copy  milling’  as  shown  below.  

 -­‐ Before  scanning  check  the  box  ‘Scan  waxup  both  way’  in  order  to  

generate  the  additional  scan  of  the  object  in  the  waxup  frame.    

Page 32: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3232 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

- Before scanning check the box ‘Scan waxup both way’ in order to generate the additional scan of the object in the waxup frame.

- When executing the ‘Waxup scan in holder’ the dental object needs to be well fixed with the metal pins, so it will not move during scanning.

- After having scanned the upper part, the software will ask to rotate the holder in order to scan the lower part.

Aadva LAB SCAN GC 32 Version 1.21 V4903l J u l y 2016

 

 

 

 

-­‐ When  executing  the  ‘Waxup  scan  in  holder’  the  dental  object  needs  to  be  well  fixed  with  the  metal  pins,  so  it  will  not  move  during  scanning.    

     

-­‐ After  having  scanned  the  upper  part,  the  software  will  ask  to  rotate  the  holder  in  order  to  scan  the  lower  part.  

 Aadva LAB SCAN GC 32

Version 1.21 V4903l J u l y 2016

 

 

 

 

-­‐ When  executing  the  ‘Waxup  scan  in  holder’  the  dental  object  needs  to  be  well  fixed  with  the  metal  pins,  so  it  will  not  move  during  scanning.    

     

-­‐ After  having  scanned  the  upper  part,  the  software  will  ask  to  rotate  the  holder  in  order  to  scan  the  lower  part.  

 Aadva LAB SCAN GC 32 Version 1.21 V4903l J u l y 2016

 

 

 

 

-­‐ When  executing  the  ‘Waxup  scan  in  holder’  the  dental  object  needs  to  be  well  fixed  with  the  metal  pins,  so  it  will  not  move  during  scanning.    

     

-­‐ After  having  scanned  the  upper  part,  the  software  will  ask  to  rotate  the  holder  in  order  to  scan  the  lower  part.  

 

Page 33: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3333Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

9. SCANFLAGS

1. Indication : Automated detection of implant brand, type, diameter and position. Please note that scan flags are implant specific. Make sure you select the proper scan flag and diameter. For the complete list of implants supported please refer to the price list.

2. How to use : - Choose the dedicated scanflag for the specific implant interface and fix it with an appropriate screw into

the model analogue. Do not use the same scanflag in two different positions.

- Open the scan server with scan and check in the scan flow if the first scan to be done is ‘Implant position scan ….’

- If not all scanflags are detected during the scan, the software will indicate this with a warning : ‘Scan adapter number mismatch : found yyy Expected xxx’,. In this case an add scan needs to be done in order to detect all scanflags.

Aadva LAB SCAN GC 34 Version 1.21 V4903l J u l y 2016

 

 

1)  2) How  to  use  :  -­‐  Choose  the  dedicated  scanflag  for  the  specific  

implant  interface  and  fix  it  with  an  appropriate  screw  into  the  model  analogue.  Do  not  use  the  same  scanflag  in  two  different  positions.  

       -­‐ Open  the  scan  server  with  scan  and  check  in  the  scan  flow  if  the  

first  scan  to  be  done  is  ‘Implant  position  scan  ….’  -­‐ If  not  all  scanflags  are  detected  during  the  scan,  the  software  will  

indicate  this  with  a  warning  :  ‘Scan  adapter  number  mismatch  :  found  yyy  Expected  xxx’,.  In  this  case  an  add  scan  needs  to  be  done  in  order  to  detect  all  scanflags.    

       

10)                  Scanbodies  

                                                                     

                                                                               

UnknownFormatted: Font:Bold, Underline

Page 34: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3434

Aadva LAB SCAN GC 35

Aadva® Lab Scan  

 

1) Indication  :  -­‐  Detection  of  an  implant  position  Please    Note  that  scan  bodies  are  implant  specific.  Make  sure  you  select  the  proper  scan  body  and  diameter.  For  the  complete  list  of  implants  supported  please  refer  to  the  price  list        

2) How  to  use  :  -­‐  Before  scanning  a  scanbody,  make  sure  that  the  specific  shape  is  in  the  implant  library.  -­‐  Fix  the  scanbody  with  an  appropriate  screw  into  the  model              analogue.  -­‐    Open  the  scan  server  with  ‘Scan  Optical  Abutment’  and  check  if  the  first  scan  in  the  scan  flow  is  ‘Scan  Abutment  scan’    

   -­‐ For  further  progress  of  the  design  proces,  follow  the  Exocad  

training  syllabus  

Articulator  frame  calibration    You  will  receive  the  following  items  (the  baseplate  in  the  frame  will  depend  on  your  articulator  type)    

         

UnknownFormatted: Font:Bold, Underline

UnknownFormatted: Font:14 pt, Kern at 18 ptUnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

10. SCANBODIES

1. Indication : Detection of an implant position Please Note that scan bodies are implant specific. Make sure you select the proper scan body and diameter. For the complete list of implants supported please refer to the price list.

2. How to use : - Before scanning a scanbody, make sure that the specific shape is in the implant library.- Fix the scanbody with an appropriate screw into the model analogue.- Open the scan server with ‘Scan Optical Abutment’ and check if the first scan in the scan flow is ‘Scan

Abutment scan’

- For further progress of the design proces, follow the Exocad training syllabus

Aadva LAB SCAN GC 35

Aadva® Lab Scan  

 

1) Indication  :  -­‐  Detection  of  an  implant  position  Please    Note  that  scan  bodies  are  implant  specific.  Make  sure  you  select  the  proper  scan  body  and  diameter.  For  the  complete  list  of  implants  supported  please  refer  to  the  price  list        

2) How  to  use  :  -­‐  Before  scanning  a  scanbody,  make  sure  that  the  specific  shape  is  in  the  implant  library.  -­‐  Fix  the  scanbody  with  an  appropriate  screw  into  the  model              analogue.  -­‐    Open  the  scan  server  with  ‘Scan  Optical  Abutment’  and  check  if  the  first  scan  in  the  scan  flow  is  ‘Scan  Abutment  scan’    

   -­‐ For  further  progress  of  the  design  proces,  follow  the  Exocad  

training  syllabus  

Articulator  frame  calibration    You  will  receive  the  following  items  (the  baseplate  in  the  frame  will  depend  on  your  articulator  type)    

         

UnknownFormatted: Font:Bold, Underline

UnknownFormatted: Font:14 pt, Kern at 18 ptUnknownFormatted: Font:14 pt, Kern at 18 pt

ARTICULATOR FRAME CALIBRATION

You will receive the following items (the baseplate in the frame will depend on your articulator type)

LOWER articulator frame (SAM) UPPER articulator frame (SAM)

Page 35: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3535Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

UPPER and LOWER calibration piece

Baseplate for Adesso split Baseplate for Artex

1. The first step is to fix the articulator pieces into the articulator (with the matching base plate)

In order to use the new articulator functions you will have to do an articulator calibration first.

Page 36: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3636 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

This calibration has to be performed only once!

2. Open the ‘Kiosk’ mode by clicking on the new GC_KIOSK icon that you will find in the Aadva Manager.

3. Once the Kiosk software is open select ‘Articulator frame calibration’ and enter in ‘Case ID:’ a name or number under which you would like the file to be saved. Then continue with ‘Start’.

Aadva LAB SCAN GC 38 Version 1.21 V4903l J u l y 2016

 

 

 

 

3. Once  the  Kiosk  software  is  open  select  ‘Articulator  frame  calibration’  and  enter  in  ‘Case  ID:’  a  name  or  number  under  which  you  would  like  the  file  to  be  saved.  Then  continue  with  ‘Start’.                                      

4. Follow  the  instructions  on  the  screen.    

a) First  scan  will  be  the  empty  LOWER     b)    Scan  the  LOWER  frame  with  the  articulator  frame.  Enter  the  frame                calibration  object  inserted.  in  the  scanner  and  click  on  ‘Ok’.    

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN GC 38 Version 1.21 V4903l J u l y 2016

 

 

 

 

3. Once  the  Kiosk  software  is  open  select  ‘Articulator  frame  calibration’  and  enter  in  ‘Case  ID:’  a  name  or  number  under  which  you  would  like  the  file  to  be  saved.  Then  continue  with  ‘Start’.                                      

4. Follow  the  instructions  on  the  screen.    

a) First  scan  will  be  the  empty  LOWER     b)    Scan  the  LOWER  frame  with  the  articulator  frame.  Enter  the  frame                calibration  object  inserted.  in  the  scanner  and  click  on  ‘Ok’.    

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Page 37: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3737Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

Aadva LAB SCAN GC 39

Aadva® Lab Scan  

 

                       

   

c) Scan  UPPER  frame  with  calibration     d)    Scan  the  empty  UPPER  frame  object        

   

 

 

 

 

           

e) Scan  the  bottom  of  the  articulator            f)    If  satisfied  with  the  result                        base  plate.  Fix  the  calibration                                      continue  by  clicking  on  the  green                      object  (with  articulator  plate)  on                            check  mark.                      a  model  holder.  

                   

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN GC 39

Aadva® Lab Scan  

 

                       

   

c) Scan  UPPER  frame  with  calibration     d)    Scan  the  empty  UPPER  frame  object        

   

 

 

 

 

           

e) Scan  the  bottom  of  the  articulator            f)    If  satisfied  with  the  result                        base  plate.  Fix  the  calibration                                      continue  by  clicking  on  the  green                      object  (with  articulator  plate)  on                            check  mark.                      a  model  holder.  

                   

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN GC 39

Aadva® Lab Scan  

 

                       

   

c) Scan  UPPER  frame  with  calibration     d)    Scan  the  empty  UPPER  frame  object        

   

 

 

 

 

           

e) Scan  the  bottom  of  the  articulator            f)    If  satisfied  with  the  result                        base  plate.  Fix  the  calibration                                      continue  by  clicking  on  the  green                      object  (with  articulator  plate)  on                            check  mark.                      a  model  holder.  

                   

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN GC 39

Aadva® Lab Scan  

 

                       

   

c) Scan  UPPER  frame  with  calibration     d)    Scan  the  empty  UPPER  frame  object        

   

 

 

 

 

           

e) Scan  the  bottom  of  the  articulator            f)    If  satisfied  with  the  result                        base  plate.  Fix  the  calibration                                      continue  by  clicking  on  the  green                      object  (with  articulator  plate)  on                            check  mark.                      a  model  holder.  

                   

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN GC 39

Aadva® Lab Scan  

 

                       

   

c) Scan  UPPER  frame  with  calibration     d)    Scan  the  empty  UPPER  frame  object        

   

 

 

 

 

           

e) Scan  the  bottom  of  the  articulator            f)    If  satisfied  with  the  result                        base  plate.  Fix  the  calibration                                      continue  by  clicking  on  the  green                      object  (with  articulator  plate)  on                            check  mark.                      a  model  holder.  

                   

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Aadva LAB SCAN GC 39

Aadva® Lab Scan  

 

                       

   

c) Scan  UPPER  frame  with  calibration     d)    Scan  the  empty  UPPER  frame  object        

   

 

 

 

 

           

e) Scan  the  bottom  of  the  articulator            f)    If  satisfied  with  the  result                        base  plate.  Fix  the  calibration                                      continue  by  clicking  on  the  green                      object  (with  articulator  plate)  on                            check  mark.                      a  model  holder.  

                   

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

4. Follow the instructions on the screen.

a) First scan will be the empty LOWER articulator frame. Enter the frame in the scanner and click on ‘Ok’.

b) Scan the LOWER frame with the calibration object inserted.

c) Scan UPPER frame with calibration object

d) Scan the empty UPPER frame

e) Scan the bottom of the articulator object (with articulator plate) on a model holder.

f) If satisfied with the result check mark.

Page 38: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3838 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva LAB SCAN GC 40 Version 1.21 V4903l J u l y 2016

 

 

       

 

g) If  the  base  plate  has  been                                                h)  The  file  will  be  saved  on  C:  Drive  calculated  by  the  system  choose                                    in  the  Kiosk-­‐data.  the  articulator  type  and  click  on  ‘Export’.    

               

                         

 

 

 

5. Now  you  can  use  these  articulator  frames  in  the  Aadva  software  to  do  the  articulator  scan.  

 

SEND  FILE  VIA  DSP    

1) Once  the  design  is  finished  and  saved  open  DentalDB  and  click  on  GC_DSP    

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Tomelleri, Marco 22/6/16 16:02Formatted: English

Aadva LAB SCAN GC 40 Version 1.21 V4903l J u l y 2016

 

 

       

 

g) If  the  base  plate  has  been                                                h)  The  file  will  be  saved  on  C:  Drive  calculated  by  the  system  choose                                    in  the  Kiosk-­‐data.  the  articulator  type  and  click  on  ‘Export’.    

               

                         

 

 

 

5. Now  you  can  use  these  articulator  frames  in  the  Aadva  software  to  do  the  articulator  scan.  

 

SEND  FILE  VIA  DSP    

1) Once  the  design  is  finished  and  saved  open  DentalDB  and  click  on  GC_DSP    

UnknownFormatted: Font:14 pt, Kern at 18 pt

UnknownFormatted: Font:14 pt, Kern at 18 pt

Tomelleri, Marco 22/6/16 16:02Formatted: English

5. Now you can use these articulator frames in the Aadva software to do the articulator scan.

g) If the base plate has been system choose the articulator type and click on ‘Export’.

h) The file will be saved on C: Drive calculated by the in the Kiosk-data.

Page 39: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

3939Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

SEND FILE VIA DSP

1) Once the design is finished and saved open DentalDB and click on GC_DSP

Aadva LAB SCAN GC 41

Aadva® Lab Scan  

 

 

 2) Enter  username  and  password  and  check  ‘I  accept  the  license  agreement’.  

Then  click  on  Upload.  

 

 

 

3) The  file  will  now  be  uploaded  to  the  DSP  account.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

Aadva LAB SCAN GC 41

Aadva® Lab Scan  

 

 

 2) Enter  username  and  password  and  check  ‘I  accept  the  license  agreement’.  

Then  click  on  Upload.  

 

 

 

3) The  file  will  now  be  uploaded  to  the  DSP  account.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

2) Enter username and password and check ‘I accept the license agreement’. Then click on Upload.

Page 40: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4040 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

3) The file will now be uploaded to the DSP account.

4) After the upload is finished the DSP account will open automatically

Aadva LAB SCAN GC 42 Version 1.21 V4903l J u l y 2016

 

 

 

 

4) After  the  upload  is  finished  the  DSP  account  will  open  automatically  

 

 

5) Open  the  Tab  ‘Home’.  You  can  find  the  uploaded  case  under  ‘My  Cases’.  

 

6) Click  on  the  case  to  open  further  information/options.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

Aadva LAB SCAN GC 42 Version 1.21 V4903l J u l y 2016

 

 

 

 

4) After  the  upload  is  finished  the  DSP  account  will  open  automatically  

 

 

5) Open  the  Tab  ‘Home’.  You  can  find  the  uploaded  case  under  ‘My  Cases’.  

 

6) Click  on  the  case  to  open  further  information/options.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

5) Open the Tab ‘Home’. You can find the uploaded case under ‘My Cases’.

Aadva LAB SCAN GC 42 Version 1.21 V4903l J u l y 2016

 

 

 

 

4) After  the  upload  is  finished  the  DSP  account  will  open  automatically  

 

 

5) Open  the  Tab  ‘Home’.  You  can  find  the  uploaded  case  under  ‘My  Cases’.  

 

6) Click  on  the  case  to  open  further  information/options.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

Page 41: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4141Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

6) Click on the case to open further information/options.

Aadva LAB SCAN GC 43

Aadva® Lab Scan  

 

 

 

7) Click  on  the  icon  GC  Factory  and  GC  Milling  to  send  the  file  to  the  GC  milling  center.  

 

 

8) Confirm  your  choice  by  clicking  on  ‘Yes’  

 

9) You  can  find  all  your  sent  cases  under  the  Tab  ‘Transactions’.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

Tomelleri, Marco 22/6/16 16:05Formatted: English

7) Click on the icon GC Factory and GC Milling to send the file to the GC milling center.

Aadva LAB SCAN GC 43

Aadva® Lab Scan  

 

 

 

7) Click  on  the  icon  GC  Factory  and  GC  Milling  to  send  the  file  to  the  GC  milling  center.  

 

 

8) Confirm  your  choice  by  clicking  on  ‘Yes’  

 

9) You  can  find  all  your  sent  cases  under  the  Tab  ‘Transactions’.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

Tomelleri, Marco 22/6/16 16:05Formatted: English

8) Confirm your choice by clicking on ‘Yes’

Aadva LAB SCAN GC 43

Aadva® Lab Scan  

 

 

 

7) Click  on  the  icon  GC  Factory  and  GC  Milling  to  send  the  file  to  the  GC  milling  center.  

 

 

8) Confirm  your  choice  by  clicking  on  ‘Yes’  

 

9) You  can  find  all  your  sent  cases  under  the  Tab  ‘Transactions’.  

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

UnknownFormatted: Font:12 pt

Tomelleri, Marco 22/6/16 16:05Formatted: English

9) You can find all your sent cases under the Tab ‘Transactions’.

Page 42: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4242 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

1. Keep the device dust-free and covered when not in use.

2. The LED in the scan head is not susceptible to damage caused by prolonged use. The Aadva Lab Scan can be kept running if necessary, but the device can be turned off as an energy-

saving measure.

3. Unplug the device and remove any objects in the scanning tray prior to routine cleaning and maintenance of the unit.

4. Do not attempt to clean the sensor head inside the casing of the scanner, as this may damage the device.

5. Use a soft cloth moistened with a small amount of water or other detergents used for cleaning computers. Wipe the cloth on exposed surfaces of the Aadva Lab Scan to remove dust or other particles. Do not use

cleaning sprays on the surface of the scanner.

6. The dental clamps and trays can also be cleaned using a cloth moistened with a small amount of water.

7. While cleaning the device with a moistened cloth, make sure that the capacitors or other electrical parts of the Aadva Lab Scan do not come into contact with any moisture.

8. Do not place any objects on top of or above the scanner, as this may interfere with the scanning functionality.

Make sure that no liquids could accidentally spill on the device.

CARE OF THE UNIT

Page 43: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4343Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

If any difficulty with the Aadva Lab Scan unit is experienced, please check below for possible causes of the problem and suggested corrective actions before seeking further advice or repair from an authorized GC Dealer or Service Center.

Problem Check Corrective Action

When the Aadva Lab Scan is switched ON, no light appears.

Is the power cord correctly attached to the Aadva Lab Scan and to the power outlet?

Switch the device OFF. Unplug and plug the power cord correctly into the Aadva Lab Scan and the power outlet. Then switch the device ON again and wait for a few seconds.

The Aadva Lab Scan application does not start. The following error message appears:

Is the software dongle (page 7) available and correctly inserted at the USP port of the workstation?

Make sure that the software dongle (page 7) is attached. Also, make sure that your license is currently valid. (how?)

The Aadva Lab Scan software is running, but the scan job does not start (or the scanner software indicates that no Aadva Lab Scan was found). The following error message appears:

Is the hardware dongle (page 7) available and correctly inserted at the USP port of the workstation?

Make sure that the hardware dongle (page 7) is attached. Also, make sure that your license is currently valid.

TROUBLESHOOTING

Page 44: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4444 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

An error occurs during the preparation scan:

The scan wizard shows a red button, implying that there is no connection to the device.

Is the USB cable connected to the Aadva Lab Scan and the workstation?

Make sure that the USB cable has been attached correctly. The Type B end of the USB cable connects to the Aadva Lab Scan, while the Type A connector is inserted into the workstation.

Is the Aadva Scan switched on? Check to see if the Aadva Lab Scan is emitting a blue light. Otherwise, turn the unit on.

During the scan, an error message appears during scan data triangulation:

Has the device completed scanning the dental model?

Adjust the view of the model. If necessary, trim or crop any parts of the initial scan which are not relevant for the reconstruction type.

The device has completed the scan but the scanned object appears incomplete.

Has the device completed scanning the dental model?

Adjust the view of the model. If necessary, trim or crop any parts of the initial scan which are not relevant for the reconstruction type.

The power supply is interrupted during the scanning procedure.

Restart the scanner. If necessary reload the scan profile or setup a new one.

Page 45: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4545Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva® Lab Scan

Your product is marked with this symbol for an environmentally responsible recycling/disposal of your product. Do not discard it with normal waste. The black line under the crossed-out waste bin symbol signifies that this product has been put into circulation after 13.08.2005 (see Directive 2012/19/EU and EN 50419:2005).

Please be aware that this product has been submitted to the European Directive 2012/19/EU (WEEE) and the laws applicable in your country for environmentally responsible recycling/disposal. Please ask GC Europe or your GC dealer when and where your product will have to be disposed of.

DISPOSAL OF THE UNIT

Your Aadva Lab Scan has a warranty period of two years from the date of purchase.

In case of a breakdown during this period, repair will be carried out free of charge provided that the unit has been used under normal conditions and according to the instructions for use. In case of damage under these conditions stop using the instrument immediately and contact GC Europe (contact details see last page of this instructions for use).

WARRANTY AND REPAIR

Only GC Europe may perform the maintenance, service and repair. In case of damage please contact GC Europe (contact details see front page of this instructions for use).

Aadva Lab Scanner Rating Plate

MAINTENANCE AND SERVICE

Page 46: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

4646 Aadva LAB SCAN from GC Version 1.0 | 990487 - Y9500 | September 2016

Aadva Lab Scanner

Name/Model Aadva Lab Scan

Light source LED (Blue)

Resolution 2 megapixels

Number of cameras 2

Positioning unit 2 axis (tilt & swivel)

Accuracy 6 microns

Scan time < 60 seconds

Scan accuracy < 10 µm

Supply voltage, frequency range and current 110/220 V CA, 50-60 Hz, 3,5/2 A

Dimensions (w x h x l) 400 x 600 x 300 mm

WeightScanner 27 kg

PC 20 kg

Environmental conditions in use

For indoor use only

Temperature 5°- 40 °C (41-104°F)

Altitude up to 2000 m or

Barometric pressure 700 hPa – 1060 hPa

Relative Humidity 30-75%

Main supply voltage fluctuation +/- 10%

Pollution degree 2

Temperature -10°- +40 °C (14-104°F)

Environmental conditions at transport and storage

Altitude up to 2000 m or

Barometric pressure 700 hPa – 1060 hPa

Relative Humidity 30-75%

Protection class XP00

StandardsEN 61010-1

EN 61326-1

Regulatory Conformity

RoHS2

WEEE

Note: The specifications of the unit may be subject to change without notice.

TECHNICAL SPECIFICATIONS

Page 47: Lab Scan - GC EUROPE Aadva LAB SCAN from GC Version . 9048 Y950 September 016 1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel. The ventilation fans should turn

9904

87 -

Y95

00

Manufactured & distributed by: GC Europe N.V.Researchpark Haasrode-Leuven 1240Interleuvenlaan 33B-3001 LeuvenTel: +32.16.74.10.00Fax: [email protected]

Distributed by:GC UNITED KINGDOM Ltd.Coopers CourtNewport Pagnell UK-Bucks. MK16 8JSTel. +44.1908.218.999 Fax. +44.1908.218.900 [email protected]://uk.gceurope.com

The Aadva® Lab Scan instructions for use can be downloaded at the following web address http://www.gceurope.com or can be requested at no charge at following contacts: [email protected] 0032 16742600