Powered Surgery –Electrosurgical Units

Page 2: kat-powered

High Technology Electrosurgical Unit with integrated Argon

Up to 2 Twin Foot Pedals

One for monpolar and bipolar use(REF 34-9330) and one for indepen-dentbipolar only use (REF 34-9331 –optional). Water proof (IPX7) with lowvoltage, safe medical grade design.

Neutral Electrode Alarm

When the contact surface of a dualsplit-type neutral electrode plate de creases to 50%, output powers willbe limited to max. 200W. The powerwill be totally cut off and alarm will beactivated in case contact decreasesfurther to an insufficient level or incase that the cable gets disconnectedor broken, for maximum safety.

HF Leakage Monitoring

automatically reducing the outputpowers in order to meet, even in criti-cal situations (i.e. contact between pa-tient and metallic parts of OT-table),safe values, reducing potential risksinvolved (i.e. burns).

10 Monopolar Currents

perfectly suitable for all surgical needs including TUR(monopolar resection under liquid) and vaporisation. Also available is a special A. ENDO mode for blend cut withalternate phases of pure cut and coagulation in flexibleendoscopy and A. PURE and A. BLEND modes combinedwith a constant voltage system, that automatically adjuststhe powers by keeping the electric arc constant. This provides for a very delicate way of cutting and minimal superficial necrosis.

2 Active Monopolar Outputs

for push-button handswitches with USA 3-pin connectors or monopolar electrodecables and handles. Activation even simul-taneously possible by two surgeons byusing hand switch or twin foot pedal. Aspecial universal adapter (REF 34-9590)is optionally avail able for connecting mul-tiple accessories with 2 to 8mm plugs.

Ideal for Cardio-Surgery

“Floating type” insulated at high/ lowfrequencies and protected against theuse of defibrillator and 5 preset cardioprograms (#75-79) make this unit anideal choice for cardio-surgery.

ESU-A 400HT

100 Memory Programs

Stored programs can be selected sim-ply by the push of one button. Presetprograms are available for cardio-surgery, neuro-surgery, general and laparoscopic surgery, flexible endos-copy or rigid monopolar endoscopyunder liquid (TUR) and for bipolar onlyuse. The unit always remembers thelast settings it was switched off with.

Page 3: kat-powered

Integrated Argon Gas

used especially in endoscopy suitesin laparoscopy and thoracoscopythe argon plasma coagulation withconstant gas flow provides for rapidhemostasis of large surface blee-dings without any contact, limitedpenetration depth (only max. 3mmin combination with SPRAY mode)and a low risk of perforation. Themonopolar current is sent into tissuevia an ionized argon gas stream. Asoxygen is displaced by the argongas in the working area of the activeelectrodes, smells and smoke areabsent and visibility is improved.There is no adhesion between theapplicator and the tissue. Specialaccessories are required.

7 Bipolar Currents

or all surgical needs and with special PRECISEmode for micro precise bipolar coagulation and theSEALING mode for bipolar coagulation and confi-dently ligating and sealing large vessels. AUTOmode with automatic Start/ Stop System and im-pendance sensing for manual operation activatesthe HF current automatically once the tissue isbeing touched.

Total Safety Concept

with complete self-test, hardware/software diagnosis and totalfunction and safety monitoring by two 32 bit microprocessors. Dynamic auto-adjustment of output powers for best performancewhile using the lowest possible powers and reducing damagingside effects on target points (i.e. carbonization or tissue sticking).

Powerful Performance

Output power 400W max. in monpolar pure cut and 160W max.in bipolar pure cut. Constant outputpower levels through the outputpower control system make this unitparticularly suitable for cutting underwater (e.g. during TUR). Power canbe progressively adjusted to thedesired level under visual control.

Order REF 34-9300

(230V) suggested accessories areREF 34-9320, 34-9331, 34-9408,34-9590, 34-9526, 10-7030

Neutral Electrode Connection

with safety control circuit to avoid thatburns can be caused on the patients’ skin.When dual split-type electrodes are beingused, not only the cable and its connec tionare being checked, but also the properapplication of the plate will be monitoredcontinuously.

2 Bipolar Outputs

for cables with 2x4mm USA-typebanana plug connectors (e.g.REF 34-9408) for bipolar cut andcoagulation with maximum safetyand precision.

Page 4: kat-powered

High Technology Electrosurgical Unit

Up to 2 Twin Foot Pedals

One for monpolar and bipolar use(REF 34-9330) and one for indepen-dentbipolar only use (REF 34-9331 –optional). Water proof (IPX7) with lowvoltage, safe medical grade design..

2 Active Monopolar Outputs

for push-button handswitches with USA 3-pin connectors or monopolar electrodecables and handles. Activation even simul -taneously possible by two surgeons byusing hand switch or twin foot pedal. Aspecial universal adapter (REF 34-9590)is optionally available for connecting mul-tiple accessories with 2 to 8mm plugs.

Ideal for Cardio-Surgery

“Floating type” insulated at high/ lowfrequencies and protected against theuse of defibrillator and 5 preset cardioprograms (#75-79) make this unit anideal choice for cardio-surgery.

100 Memory Programs

Stored programs can be selected sim-ply by the push of one button. Presetprograms are available for cardio-surgery, neuro-surgery, general andlaparoscopic surgery, flexible endos-copy or rigid monopolar endoscopyunder liquid (TUR) and for bipolar onlyuse. The unit always remembers thelast settings it was switched off with.

10 Monopolar Currents

perfectly suitable for all surgical needs including TUR(monopolar resection under liquid) and vaporisation. Also available is a special A. ENDO mode for blend cut withalternate phases of pure cut and coagulation in flexibleendoscopy and A. PURE and A. BLEND modes combinedwith a constant voltage system, that automatically adjuststhe powers by keeping the electric arc constant. This provides for a very delicate way of cutting and minimal superficial necrosis.

Neutral Electrode Alarm

When the contact surface of a dualsplit-type neutral electrode plate de creases to 50%, output powers willbe limited to max. 200W. The powerwill be totally cut off and alarm will beactivated in case contact decreasesfurther to an insufficient level or incase that the cable gets disconnectedor broken, for maximum safety.

HF Leakage Monitoring

automatically reducing the outputpowers in order to meet, even in criti-cal situations (i.e. contact between pa-tient and metallic parts of OT-table),safe values, reducing potential risksinvolved (i.e. burns).

ESU-X 400HT

Page 5: kat-powered

Powerful Performance

Output power 400W max. in monpolar pure cut and 160W max.in bipolar pure cut. Constant outputpower levels through the outputpower control system make this unitparticularly suitable for cutting underwater (e.g. during TUR). Power canbe progressively adjusted to thedesired level under visual control.

Neutral Electrode Connection

with safety control circuit to avoid thatburns can be caused on the patients’ skin.When dual split-type electrodes are beingused, not only the cable and its connectionare being checked, but also the properapplication of the plate will be monitoredcontinuously.

2 Bipolar Outputs

for cables with 2x4mm USA-typebanana plug connectors (e.g.REF 34-9408) for bipolar cut andcoagulation with maximum safetyand precision.

7 Bipolar Currents

for all surgical needs and with special PRECISEmode for micro precise bipolar coagulation andthe SEALING mode for bipolar coagulation andconfidently ligating and sealing large vessels.AUTO mode with automatic Start /Stop Systemand impendance sensing for manual operationactivates the HF current automatically once thetissue is being touched.

Total Safety Concept

with complete self-test, hardware/software diagnosis and totalfunction and safety monitoring by two 32 bit microprocessors. Dynamic auto-adjustment of output powers for best performancewhile using the lowest possible powers and reducing damagingside effects on target points (i.e. carbonization or tissue sticking).

Order REF 34-9302

(230V) suggested accessories areREF 34-9320, 34-9331, 34-9408,34-9590, 34-9526, 10-7030

Page 6: kat-powered

New Technology Electrosurgical Unit

2 Active Monopolar Outputs

for push-button handswitches with US 3-pin connectors or monopolar electrodecables and handles. A special adapter isoptionally available for 2-8mm plugs (REF34-9590). Activation even simultaneouslypossible by two surgeons.

8 Monopolar Currents

perfectly suitable for all surgical needs including TUR(monopolar resection under liquid) and vaporisation. Alsoavailable is a special ENDO mode for blend cut with alter-nate phases of pure cut and coagulation in flexible endoscopy.

Neutral Electrode Connection

with safety control circuit to avoid thatburns can be caused on the patients’ skin.When split-type electrodes are being used.not only the cable and its connection arechecked, but the proper application of theplate will be monitored continuously.

Neutral Electrode Alarm

When the contact surface of a dualsplit-type neutral electrode plate decreases to 50%, output powers willbe limited to max. 200W. The powerwill be totally cut off and alarm will beactivated in case contact decreasesfurther to an insufficient level or incase that the cable gets disconnectedor broken, for maximum safety.

10 Memory Programs

Stored programs can be selected sim-ply by the push of one button. Presetprograms are available for cardio-surgery, neuro-surgery, general andlaparoscopic surgery, flexible endos-copy or rigid monopolar endoscopyunder liquid (TUR) and for bipolar onlyuse. The unit always remembers thelast settings it was switched off with.

HF Leakage Monitoring

automatically reducing the outputpowers in order to meet, even in criticalsituations (i.e. contact between patientand metallic parts of OT-table), safevalues, reducing potential risks in -volved (i.e. burns).

Order REF 34-9310

(230V) suggested accessories areREF 34-9320, 34-9331, 34-9408, 34-9590, 34-9526, 10-7030

ESU-X 300NT

Page 7: kat-powered

Powerful Performance

Output 350W max. in monpolarpure cut and 140W max. in bipolarpure cut. Constant output power levels through the output powercontrol system make this unit parti-cularly suitable for cutting under water (e.g. during TUR).

5 Bipolar Currents

for all surgical needs with bipolar cut andcoagulation. AUTO mode with automaticStart /Stop System and impendancesensing.

Ideal for Cardio-Surgery

“Floating type” insulated at high/ lowfrequencies and protected againstthe use of defibrillator and a presetcardio program make this unit anideal choice for cardio-surgery.

1 Bipolar Output

for cables with 2x4mm USA-type banana plug connectors(e.g. REF 34-9408) for bipolarcutting and coagulation withmaximum safety and precision.

Total Safety Concept

with complete self-test, hardware/software diagnosis and totalfunction and safety monitoring by two 32 bit microprocessors. Dynamic auto-adjustment of output powers for best performancewhile using the lowest possible powers and reducing damagingside effects on target points (i.e. carbonization or tissue sticking).

Up to 2 Twin Foot Pedals

One for monpolar and bipolar use(REF 34-9330) and one for inde-pendentbipolar only use (REF 34-9331 – optional). Water proof(IPX7) with low voltage, safe medi-cal grade design.

Page 8: kat-powered

New Technology Electrosurgical Unit

HF Leakage Monitoring

automatically reducing the outputpowers in order to meet, even in criticalsituations (i.e. contact between patientand metallic parts of OT- table), safevalues, reducing potential risks in-volved (i.e. burns).

8 Monopolar Currents

perfectly suitable for all surgical needs including TUR(monopolar resection under liquid) and vaporisation. Alsoavailable is a special ENDO mode for blend cut with alternate phases of pure cut and coagulation in flexibleendoscopy.

10 Memory Programs

Stored programs can be selected sim-ply by the push of one button. Presetprograms are available for cardio-surgery, neuro-surgery, general andlaparoscopic surgery, flexible endos-copy or rigid monopolar endoscopyunder liquid (TUR) and for bipolar onlyuse. The unit always remembers thelast settings it was switched off with.

Neutral Electrode Alarm

When the contact surface of a dualsplit-type neutral electrode plate decreases to 50%, output powers willbe limited to max. 200W. The powerwill be totally cut off and alarm will beactivated in case contact decreasesfurther to an insufficient level or incase that the cable gets disconnectedor broken, for maximum safety.

ESU-X 200NT

Neutral Electrode Connection

with safety control circuit to avoid thatburns can be caused on the patients’ skin.When split-type electrodes are being used.not only the cable and its connection arechecked, but the proper application of theplate will be monitored continuously.

Order REF 34-9312

(230V) suggested accessories areREF 34-9320, 34-9331, 34-9408, 34-9590, 34-9526, 10-7030

2 Active Monopolar Outputs

for push-button handswitches with US 3-pin connectors or monopolar electrodecables and handles. A special adapter isoptionally available for 2-8mm plugs (REF34-9590). Activation even simultaneouslypossible by two surgeons.

Page 9: kat-powered

Powerful Performance

Output 200W max. in monpolarpure cut and 140W max. in bipolarpure cut. Constant output power le-vels through the output power con-trol system make this unit particu-larly suitable for cutting under water(e.g. during TUR).

Up to 2 Twin Foot Pedals

One for monpolar and bipolar use(REF 34-9330) and one for inde-pendentbipolar only use (REF 34-9331 - optional). Water proof (IPX7)with low voltage, safe medical gradedesign.

Total Safety Concept

with complete self-test, hardware/software diagnosis and totalfunction and safety monitoring by two 32 bit microprocessors. Dynamic auto-adjustment of output powers for best performancewhile using the lowest possible powers and reducing damagingside effects on target points (i.e. carbonization or tissue sticking).

Ideal for Cardio-Surgery

“Floating type” insulated at high/lowfrequencies and protected againstthe use of defibrillator and a presetcardio program make this unit anideal choice for cardio-surgery.

1 Bipolar Output

for cables with 2x4mm USA-type banana plug connectors(e.g. REF 34-9408) for bipolarcutting and coagulation withmaximum safety and precision.

5 Bipolar Currents

for all surgical needs with bipolar cut andcoagulation. AUTO mode with automaticStart/ Stop System and impendancesensing.

Page 10: kat-powered

Accessories for Electrosurgical Units

34-9320 Standard Accessories Setconsisting of:34-9330 Twin Foot Pedal for monpolar and bipolar use34-9525 Electrode Holder Handle 3.5m cable for 2.3/2.4mm electrodes (2x)34-9528 Adult Stainless Steel Neutral Electrode (25x12xm) with cable34-9532 Fixing Belt 150cm 34-9550 Straight Knife Electrode (Blade) 70mm (2x)34-9554 Thick Needle Electrode 70mm (2x)34-9556 Thin Needle Electrode 70mm 34-9560 Straight Ball Electrode 2.5mm 70mm34-9562 Straight Ball Electrode 4mm 70mm (2x)

34-9321 Standard Accessories Set (Rubber Neutral Electrode)as above but instead of 34-9528 34-9530 Adult Rubber Neutral Electrode (25x12xm) with cable

34-9526 Electrode Holder with double-push buttonwith 3m cable for 2.3/2.4mm electrodes

34-9526.50 as above but with 5m long cable

34-9532

Model and color might vary from picture.

34-9330 Twin Foot Pedalfor monopolar and bipolar use

34-9331 Twin Foot Pedal (Bipolar)optional second pedal for independent bipolar use only

34-9528 Neutral Electrode stainless steel (25x12cm) complete with cable (REF 34-9540)

Disposable Split-type Neutral Electrode (pack of 25)SW-1211 Adult >15kg, 104x215mm, SW-1222 Child 5-15kg, 90x144mmSW-1223 Infant <5kg, 74x165mm

34-9540 Cable for connecting Neutral Electrodes

Rubber Neutral Electrodes 34-9530 Adult >15kg (25x12xm), with 2.5m cable34-9531 Child 5-15kg (19x8cm), with 2.5m cable

34-9590 Adapterfor monopolar outlet, for plug sizes 2-8mm

Page 11: kat-powered

Active Electrodesreusable, with 2.3-2.4mm stems34-9550 Straight Knife (Blade) Electrode 70mm 34-9551 Angled Knife (Blade) Electrode 70mm 34-9550K Straight IMA Knife (Blade) Electrode 70mm (5mm distal tip)34-9550L Straight Lancet Electrode 70mm34-9551L Angled Lancet Electrode 70mm34-9554 Straight Thick Needle Electrode 70mm34-9555 Angled Thick Needle Electrode 70mm34-9556 Straight Thin Needle Electrode 70mm 34-9557 Angled Thin Needle Electrode 70mm34-9555K Straight IMA Thin Needle Electrode 70mm (5mm distal tip)34-9560 Straight Ball Electrode 2.5mm 70mm34-9561 Angled Ball Electrode 2.5mm 70mm34-9562 Straight Ball Electrode 4mm 70mm 34-9563 Angled Ball Electrode 4mm 70mm34-9564 Straight Ball Electrode 6mm 70mm 34-9565 Angled Ball Electrode 6mm 70mm34-9568 Diamond Loop Electrode 5x10mm 70mm34-9569 Diamond Loop Electrode 10x10mm 70mm34-9571 Round Wire Loop Electrode 10mm 70mm34-9572 Round Wire Loop Electrode 15mm 70mm34-9570N Round Ribbon Loop Electrode 5mm 70mm34-9571N Round Ribbon Loop Electrode 10mm 70mm34-9572N Round Ribbon Loop Electrode 15mm 70mm34-9573 Plate Electrode 70mm34-9651 Electrode Extension 15cm (2.3-2.4mm electrodes)34-9583 Straight Thin Needle Electrode 120mm 34-9582 Straight Thick Needle Electrode 120mm34-9580 Straight Knife (Blade) Electrode 120mm34-9584 Round Wire Loop Electrode 5mm 120mm34-9585 Round Wire Loop Electrode 10mm 120mm34-9585X Round Wire Loop Electrode 15mm 120mm34-9586 Straight Ball Electrode 2.5mm 120mm34-9588 Straight Ball Electrode 4mm 120mm34-9589 Straight Ball Electrode 6mm 120mm34-9580K Straight IMA Knife (Blade) Electrode 120mm (5mm distal tip)34-9583K Straight IMA Thin Needle Electrode 120mm (5mm distal tip)

34-9321 Round Box 10cm diameter, 12cm high, with insert for 10 active electrodes 2.3-2.4mm stem

34-9550

34-9551

34-9550K

34-9550L

34-9551L

34-9554

34-9555

34-9556

34-9557

34-9555K

34-9560

34-9561

34-9562

34-9563

34-9564

34-9565

34-9568

34-9569

34-9570

34-9571

34-9572

34-9571N

34-9572N

34-9573

34-9651

34-9583

34-9582

34-9580

34-9584

34-9585

34-9585X

34-9586

34-9588

34-9589

34-9580K

34-9583K

34-9525 Electrode Holder with 3.5m cable for 2.3/2.4mm electrodes

34-9421 Monpolar Cablefor connecting Laparoscopic Instrumentsto monopolar outlet of ESU

Page 12: kat-powered

Bipolar Scissorsfor cutting and dissection with simultaneous bipolar coagulation andfor precise pinpoint or zone coagulation of blood vessels and tissue. 26-7918 Bipolar Scissors 18cm regular tip26-7918.FN Bipolar Scissors 18cm fine tip, 26-7920 Bipolar Scissors 20cm regular tip26-7920.FN Bipolar Scissors 20cm fine tip26-7923 Bipolar Scissors 23cm regular tip26-7923.FN Bipolar Scissors 23cm fine tip26-7928 Bipolar Scissors 28cm regular tip26-7928.FN Bipolar Scissors 28cm fine tip

Endoscopic Bipolar Forceps with Power Grip Handleprecise opening and closing of the jaws as well as applying very highpressure when grasping and cutting while precisely coagulating.34-2980 Bipolar Power Grip Handle with Euro-plug (flat)34-2983 Shaft for 34-2980 34-2985 Bipolar Metzenbaum Scissors Insert for 34-298034-2986 Bipolar Grasping Forceps Insert for 34-298034-2987 Bipolar Maryland Dissector Insert for 34-298034-9408 Cable for Bipolar Forceps with Euro-plug (flat)

Bipolar Scissors/Clamps for Vessel SealingThese instruments enable powerful grasping and compression of the tissue, combined withbipolar coagulation or sealing of blood vessels. Then the tissue can be cut without changing the instrument.26-7950 Bipolar Scissors/Clamps 14cm straight26-7951 Bipolar Scissors/Clamps 14cm curved26-7952 Bipolar Scissors/Clamps 19cm straight26-7953 Bipolar Scissors/Clamps 19cm curved

with plastic shanks and ratchet

Bipolar Clamps for Vessel Sealingwithout scissors25-9730 Bipolar Clamps 16cm, 30mm jaw, 4mm tip, curved25-9733 Bipolar Clamps 23cm, 40mm jaw, 3mm tip, curved25-9736 Bipolar Clamps 26cm, 45mm jaw, 4mm tip, curved26-9743 Bipolar Clamps 23cm, 40mm jaw, 4mm tip, curved

26-7923.FN

26-7918.FN

34-2980

34-2983

34-2986

26-7952

25-9733

25-9743

26-7920

Accessories for Electrosurgical Units

34-9408.BC Bipolar Cable for Bipolar Scissors/Clamps and Bipolar Clamps

34-9423 Bipolar Cable for Bipolar Scissors

Page 13: kat-powered

10-7124

10-7030

10-7201.US

10-7012

Bipolar Tissue Forcepswith flat Euro-plug, straight tip10-7021 Bipolar Tissue Forceps 15cm, tip 2.0mm10-7020 Bipolar Tissue Forceps 15cm, tip 1.0mm10-7022 Bipolar Tissue Forceps 15cm, tip 0.3mm10-7031 Bipolar Tissue Forceps 18cm, tip 2.0mm10-7030 Bipolar Tissue Forceps 18cm, tip 1.0mm10-7032 Bipolar Tissue Forceps 18cm, tip 0.3mm10-7041 Bipolar Tissue Forceps 20cm, tip 2.0mm10-7040 Bipolar Tissue Forceps 20cm, tip 1.0mm10-7042 Bipolar Tissue Forceps 20cm, tip 0.3mm10-7051 Bipolar Tissue Forceps 22cm, tip 2.0mm10-7050 Bipolar Tissue Forceps 22cm, tip 1.0mm10-7052 Bipolar Tissue Forceps 22cm, tip 0.3mm10-7061 Bipolar Tissue Forceps 24cm, tip 2.0mm10-7060 Bipolar Tissue Forceps 24cm, tip 1.0mm10-7062 Bipolar Tissue Forceps 24cm, tip 0.3mm

Bipolar Jeweler`s Forcepswith flat Euro-plug, straight tip10-7001 Bipolar Jeweler`s Forceps 7cm straight, tip 0.25mm10-7000 Bipolar Jeweler`s Forceps 7cm straight, tip 0.45mm10-7003 Bipolar Jeweler`s Forceps 7cm angled, tip 0.25mm10-7002 Bipolar Jeweler`s Forceps 7cm angled, tip 0.45mm10-7012 Bipolar Jeweler`s Forceps 11cm straight, tip .25mm10-7010 Bipolar Jeweler`s Forceps 11cm straight, tip 0.60mm10-7016 Bipolar Jeweler`s Forceps 11cm angled, tip 0.25mm10-7014 Bipolar Jeweler`s Forceps 11cm angled, tip 0.60mm

Bipolar ENT Tissue Forcepswith flat Euro-plug, knee-bent, 17cm, straight tip10-7201 Bipolar ENT Tissue Forceps, tip 2.0mm10-7200 Bipolar ENT Tissue Forceps, tip 1.0mm10-7202 Bipolar ENT Tissue Forceps, tip 0.3mm

Bipolar Bayonet Tissue Forcepswith flat Euro-plug, straight tip 10-7114 Bipolar Tissue Forceps 18cm bayonet, tip 2.0mm10-7110 Bipolar Tissue Forceps 18cm bayonet, tip 1.0mm10-7111 Bipolar Tissue Forceps 18cm bayonett, tip 0.3mm10-7124 Bipolar Tissue Forceps 20cm bayonet, tip 2.0mm10-7120 Bipolar Tissue Forceps 20cm bayonet, tip 1.0mm10-7121 Bipolar Tissue Forceps 20cm bayonett, tip 0.3mm10-7134 Bipolar Tissue Forceps 22cm bayonet, tip 2.0mm10-7130 Bipolar Tissue Forceps 22cm bayonet, tip 1.0mm10-7131 Bipolar Tissue Forceps 22cm bayonett, tip 0.3mm10-7144 Bipolar Tissue Forceps 24cm bayonet, tip 2.0mm10-7140 Bipolar Tissue Forceps 24cm bayonet, tip 1.0mm10-7141 Bipolar Tissue Forceps 24cm bayonett, tip 0.3mm

Bipolar Tissue Forcepswith flat Euro-plug, angled tip10-7025 Bipolar Tissue Forceps 15cm angled, tip 2.0mm10-7024 Bipolar Tissue Forceps 15cm angled, tip 1.0mm10-7026 Bipolar Tissue Forceps 15cm angled, tip 0.3mm10-7035 Bipolar Tissue Forceps 18cm angled, tip 2.0mm10-7034 Bipolar Tissue Forceps 18cm angled, tip 1.0mm10-7036 Bipolar Tissue Forceps 18cm angled, tip 0.3mm10-7045 Bipolar Tissue Forceps 20cm angled, tip 2.0mm10-7044 Bipolar Tissue Forceps 20cm angled, tip 1.0mm10-7046 Bipolar Tissue Forceps 20cm angled, tip 0.3mm10-7055 Bipolar Tissue Forceps 22cm angled, tip 2.0mm10-7054 Bipolar Tissue Forceps 22cm angled, tip 1.0mm10-7056 Bipolar Tissue Forceps 22cm angled, tip 0.3mm

34-9408 Bipolar Cablefor Bipolar Forceps with Euro-plug (flat)

Most Bipolar Forceps are also available with USA-plug (2 pins) on request. For ordering simply add „.US“ behind article number (e.g. 10-7110.US).

34-9408.US Bipolar Cablefor Bipolar Forceps with USA-plug

(2 pins)

10-7034

Page 14: kat-powered

Accessories for Electrosurgical Units

GEISTER® XSMOKE Smoke Evacuatordesigned to evacuate and filter the smoke produced during all electro-surgical procedures, thus eliminating bad smells or the bacteriologicalrisks, and improving significantly the visual field during laparoscopic pro-cedures. The GEISTER® XSMOKE is equipped with an electric suctionwith high flow and a low noise level, a special circuit for the connec tionto a central vacuum supply system, and another circuit for the controlledoutput of the CO2 which is used in the laparoscopic procedures forpneumoperitoneum. The unit is ideal for the usage in the operatingtheatre, thanks to its low level of noise and the wide range of applications.It is usable for evacuation of the smoke during open surgery, with anaspiration system on the electrodes-handle or with an independent cannulae, adjustable evacuation of the smoke during laparoscopic pro-cedures with a pneumoperitoneum, by connection to the trocar providingfor the controlled output of the gas by itself, and adjustable evacuationof the smoke during gas-less laparoscopic procedures, by connectionwith the trocar, through suction All the models can be used as a “shelf”for the electro surgical units; there are also specific trolleys available,both for the out-patient treatment and the operating theatre. Deliveredwith filter group (antibacterial Maxipleat/2, antibacterial Deas1911, anti-smell with active carbons Scavenger), non sterile tubing T19-1/5 (for afixed suction system, or extension of T11-1/8), disposable sterile tubingT11-1/8 for the connection to the speculum, to the suction set for theelectrodes-handle, to the sterile handle with suction system and to thetrocar in laparoscopy, tubing TC/CV for the connection to the externalcentral vacuum system. *

* Microprocessor control system with self-test when switching-on. Three stages filter: two antibacterial Hepa-type, with atotal efficiency 99,99999% for 0.3 micron particles, and 1 active carbon anti-smell filter, controlled by micro-processor forfilter efficacy (2 hour time limit of usage). Automatic Start /Stop System: through non-contact sensing which eliminates theneed of connections to the ESU and pedal switch (optional). Programmable Stop Delay: Adjusting the delay for switchingoff the suction system from 0 to 30 sec, to completely eliminate all smoke. Optional Foot Pedal switch: Pneumatic, water -proof, anti-explosion, for operating theatre IPX7/AP/APG. Cooling: by convection without fan. Classification: IEC 60601-1- Class I, type BF (leakage to the housing: less than 25microA; leakage to the patient: absent), EC 93/42, CE-marked-Class IIA, safety meets IEC 601-1. Dimensions: 38cmx35cmx16cm. Weight: 17kg. Suction System: Rotative, lubricated- flow: 50l/min; max. vacuum: 720/730mmHg. Mains/Absorption/Fuses/Switch: 230V-50/60Hz/250VA/ 2x1,6A/Bipolar,green. Control of the CO2 output: Circuit with pressure tight electro-valve. Technical alterations and all rights reserved.

Internal Filters34-9355 SCA – SCAVENGER, Anti-smell filter with active carbons 34-9356 MAX2 – MAXIPLEAT/2, Antibacterial/Antiviral filter

External Sterile Filters for suction tubing34-9357 DE19 – DEAS1911, Sterile Antibacterial/Antiviral filter for

suction tubing

Suction tubing34-9360 Disposable Sterile Tubing, 1.8m – Ø 11mm34-9361 Disposable Extension Tubing for 34-9361, 1.5m – Ø 19mm34-9362 Tubing, 2.5m – Ø 7x12mm with connection to the unit34-9363 Disposable Sterile Tubing 5m with antibacterial filter and

luerlock connection for laparoscopy or general applications 34-9364 Smoke Evacuation Sleeve-Suction set

for disposable selectrode-handles with suction system34-9365 Connection tubing, 5m, with quick connector (not equipped

with any connector for the centralized system)34-9370 Mobile Arm for rigid suction system 34-937134-9371 Rigid Suction System

34-9350 GEISTER® XSMOKEElectrosurgical Smoke Evacuator 230V

Monopolar electrodes-handles with suction system 34-9450 Disposable sterile handle with regular shaft 40-60mm,

double-push-buttons34-9452 Disposable sterile handle with long extendeable shaft

90-110mm, double-push-buttons

Page 15: kat-powered

34-9730 Hand-switch for rigid electrodes with cable34-9740 Rigid Electrode fo coagulation 25mm 34-9741 Rigid Electrode fo coagulation 100mm 34-9742 Rigid Electrode fo coagulation 320mm 34-9745 Rigid Needle Electrode 40mm 34-9746 Rigid Needle Electrode 115mm 34-9750 Rigid Knife Electrode 40mm 34-9751 Rigid Knife Electrode 115mm

34-9760 Connecting Cable for flexible electrodes34-9761 Flexible Electrode 1.5m 1.5mm34-9762 Flexible Electrode 1.0m 2.3mm34-9763 Flexible Electrode 2.2m 2.3mm34-9764 Flexible Electrode 3.2m 2.3mm

34-9772 Argon Gas cylinder 5l with gas (transport restrictions for certain countries might apply – medical argon gas tanks can be also purchased or refilled by your local medical gas suppliers of choice)

34-9773 Pressure Reducer34-9774 Tube for gas input34-9785 Spare Filter (not illustrated)34-9786 Metal connector for filter 34-9785 (not illustrated)

34-9742

34-9740

34-9594

34-9746

34-9741

34-9745

34-9751

Electrodes for Argon Gas enhanced surgeryrapid hemostasis of large surface bleedings without any contact, limitedpenetration depth and a low risk of perforation. The monopolar currentis sent into tissue via an ionized argon gas stream. As oxygen is displaced by the argon gas in the working area of the active electrodes smells and smoke are absent and visibility is improved. There is no adhesion between the applicator and the tissue.

Trolleys34-9325 Trolley with antistatic wheels with brakes, 3 shelves for ESU, smoke evacuator, accessories, pedals, Dimensions: 50cmx60cmx100cm.34-9326 as 34-9325 above, but with compartment/holder for 2 Argon gas bottles34-9595 Trolley with antistatic wheels with brakes, 50x50x75cm34-9594 Economy Trolley with 2 shelves 40x40x75cm

34-9325 34-9595

34-9763

34-9774

34-9773

34-9760

34-9730

34-9750

34-9772

Page 16: kat-powered

Technical Specifications ESU HT/NT Series

Mains Power: 230/115 Volt ± 10%/ Hz: 50/60 Phases: alternate single-phase. Electrical input: at max. power: 828 VA – at switching on: 840 VA –in stand by: 100 VA. Mains switch: bipolar with green light, positioned on rear. Supply cable: detachable with standard SCHUKO plug (differentplugs inquiry), length 3m. Fuses: 2 fuses in the appliance inlet. Protective earth resistance: Lower than 0,1 ohm. Equipotential connection: by DIN42801 intl. plug. Dimensions: WxDxH 38cm x 35cm x 16cm, Weight :15 kgs (ESU-A400HT: 16kgs). Installation: units can be used on a trolley, shelfor hanging arm. Special conditions/facilities for the installation: None (no water, gas, compressed air, etc. required). Electric supply must meet theapplicable safety standards. Loss of heat in the environment: None. Limits of environmental temperature/humidity during the functioning or thestore: standard IEC limits – temp from 10° to 40 °C / humid. from 35 to 75% / press. from 700 to 1060 hPa. Cooling: by convection without fan.Twin foot-switch pedals: electrical type, water-proof (IPX7) for operating rooms, with low, medical, voltage 12VDC. Argon gas section (only for model ESU-A 400HT): Gas supply: by 1/2 cylinders with 5lt (cylinder full = pressure 200 ATM) with pressure reducer or by external central pipe-line – Connection for gas supply: Quick-action coupling connector – Max working pressure and maximum gas flow: 1atm, 15l/min – Adjustment ofgas flow: from 1 to 15 l/min, setting by push-buttons and displayed on progressive LED-bar. Warranty 2 years. Errors, technical alterations and allrights reserved.

Model Order REF Order REFType 230V 115V

ESU-A 400HT 34-9300 34-9300.USESU-X 400HT 34-9302 34-9302.USESU-X 300NT 34-9310 34-9310.USESU-X 200NT 34-9312 34-9312.US

© G

eist

er 2

009

–P

rinte

d in

Ger

man

y –

Ord

er N

o. 9

9-91

75B

–A

ll rig

hts

rese

rved

.

TM

Geister Medizintechnik GmbHFoehrenstrasse 2D-78532 Tuttlingen/ GermanyPhone +49-(0)7461-966 240Telefax +49-(0)7461-966 2422E-mail: [email protected]: www.geister.com

Distributor

ESU-A 400HT ESU-X 400HT

Currents Output P/P Outout P/P Nominal Frequency(F)- Crest Factor(CF) - Acoustic/ VisualPower Voltage Power Voltage Load Modulation(M)- Duty Cycle (DT) Signal

MONOPOLAR:Pure 400 W 3465 V 400 W 3465 V 360 Ω F: 435 kHz- CF: 1.6 - M: N/A – DT: N/A Low tone, yellow lightBlend 1 300 W 3595 V 300 W 3595 V 360 Ω F: 435 kHz- CF: 2.3 - M: 29 kHz – DT: 65 % Low tone, yellow lighBlend 2 140 W 7600 V 140 W 7600 V 620 Ω F: 435 kHz- CF: 8.1 - M: 19 kHz – DT: 9 % Low tone, yellow lightA. Pure 400 W 1470 V 400 W 1470 V 360 Ω F: 435 kHz- CF: 1.6 - M: N/A – DT: N/A Low tone, yellow lightA. Blend 300 W 1930 V 300 W 1930 V 360 Ω F: 435 kHz- CF: 2.3 - M: 29 kHz – DT: 65 % Low tone, yellow lightA. Endo 250 W 1890 V 250 W 1890 V 360 Ω Alternating 50% Pure and Blend1 - CF: 2,2 Low tone, yellow lightFulg 150 W 4700 V 150 W 4700 V 345 Ω F: 435 kHz- CF: 4.5 - M: 78 kHz – DT: 3,5 % High tone, blue lightPin Point 250 W 3460 V 250 W 3460 V 250 Ω F: 435 kHz- CF: 2.6 - M: 29 kHz – DT: 50 % High tone, blue lightSoft 280 W 3440 V 280 W 3440 V 250 Ω F: 435 kHz- CF: 2.5 - M: 29 kHz – DT: 56 % High tone, blue lightSpray 140 W 7660 V 140 W 7660 V 630 Ω F: 435 kHz- CF: 8.1 - M: 19 kHz – DT: 9 % High tone, blue light

BIPOLAR:Pure 160 W 850 V 160 W 850 V 310 Ω F: 435 kHz- CF: 1.5 – M: N/A – DT: N/A Low tone, yellow lightBlend 130 W 1000 V 130 W 1000 V 310 Ω F: 435 kHz- CF: 1.8 – M: 29 kHz – DT: 75 % Low tone, yellow lightMicro 130 W 450 V 130 W 450 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A Low tone, yellow lightPrecise 130 W 760 V 130 W 760 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A Low tone, yellow lightAuto 130 W 450 V 130 W 450 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A Low tone, yellow lightMacro 130 W 760 V 130 W 760 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A High tone, blue lightSealing 130 W 710 V 130 W 710 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A High tone, blue light

ESU-X 300NT ESU-X 200NT

Currents Output P/P Output P/P Nominal Frequency(F)- Crest Factor(CF) - Acoustic/ Visual Power Voltage Power Voltage Load Modulation(M)- Duty Cycle (DT) Signal

MONOPOLAR:Pure 350 W 3455 V 200 W 3455 V 360 Ω F: 435 kHz- CF: 1.6 - M: N/A – DT: N/A Low tone, yellow lightBlend 1 300 W 3595 V 200 W 3530 V 360 Ω F: 435 kHz- CF: 2.3 - M: 29 kHz – DT: 65 % Low tone, yellow lighBlend 2 140 W 7600 V 140 W 7600 V 620 Ω F: 435 kHz- CF: 8.1 - M: 19 kHz – DT: 9 % Low tone, yellow lightEndo 220 W 1720 V 200 W 1700 V 360 Ω Alternating 50% Pure and Blend1 - CF: 2,2 Low tone, yellow lightFulg 150 W 4700 V 150 W 4700 V 345 Ω F: 435 kHz- CF: 4.5 - M: 78 kHz – DT: 3,5 % High tone, blue lightPin Point 250 W 3460 V 200 W 3450 V 250 Ω F: 435 kHz- CF: 2.6 - M: 29 kHz – DT: 50 % High tone, blue lightSoft 280 W 3440 V 280 W 3435 V 250 Ω F: 435 kHz- CF: 2.5 - M: 29 kHz – DT: 56 % High tone, blue lightSpray 140 W 7600 V 140 W 7600 V 630 Ω F: 435 kHz- CF: 8.1 - M: 19 kHz – DT: 9 % High tone, blue light

BIPOLAR:Pure 140 W 840 V 140 W 840 V 310 Ω F: 435 kHz- CF: 1.5 – M: N/A – DT: N/A Low tone, yellow lightBlend 120 W 990 V 120 W 990 V 310 Ω F: 435 kHz- CF: 1.8 – M: 29 kHz – DT: 75 % Low tone, yellow lightMicro 120 W 440 V 120 W 440 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A Low tone, yellow lightAuto 120 W 440 V 120 W 440 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A Low tone, yellow lightMacro 120 W 750 V 120 W 750 V 100 Ω F: 435 kHz- CF: 1.7 – M: N/A – DT: N/A High tone, blue light