Upload
alrns-jsm
View
43
Download
4
Tags:
Embed Size (px)
Citation preview
Model 2000 Pulse Oximeter
Operator’s / Service Manual
Ivy Biomedical Systems, Inc.
11 Business Park Drive Branford, Connecticut 06405
(203) 481-4183 • (800) 247-4614 • FAX (203) 481-8734 web:www.ivybiomedical.com e-mail:[email protected]
The following are trademarks of Masimo Corporation: Masimo® Masimo SET®, and LNOP®. Covered by one or more of the following U.S. Patents and foreign Equivalents: 5,337,744, 5,452,717, 5,482,036, 5,490,505. Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.
Copyright © 1998 by Ivy Biomedical Systems, Inc., Branford, Connecticut. All rights reserved. No part of this manual may be reproduced without the permission of Ivy Biomedical Systems, Inc.
20001199 2411-00-20 REV 05
TABLE OF CONTENTS
Model 2000 Service Manual i
TABLE OF CONTENTS WARRANTY ..............................................................................................................................................................iii Description of Warning Labels .................................................................................................................................iv INTRODUCTION .......................................................................................................................................................1 SAFETY .......................................................................................................................................................................2 Electrical.......................................................................................................................................................................2 Explosion ......................................................................................................................................................................2 Patient Connections.....................................................................................................................................................2 Measurements ..............................................................................................................................................................3 EMC .........................................................................................................................................................................3 MONITOR DESCRIPTION ......................................................................................................................................4 Specifications................................................................................................................................................................4 Classification ................................................................................................................................................................5 Controls and Indicators ..............................................................................................................................................6 Fixed Keys ......................................................................................................................................................6 On/Standby......................................................................................................................................................6 Programmable Keys ........................................................................................................................................6 Display .........................................................................................................................................................................7 Waveform Display ..........................................................................................................................................7 Alphanumeric Data .........................................................................................................................................7 Alarm Limits Display......................................................................................................................................7 Alarm Displays................................................................................................................................................7 Trend Displays ................................................................................................................................................7 Alarms .........................................................................................................................................................................8 Rear Panel ....................................................................................................................................................................8 Menu Structure............................................................................................................................................................9 MONITOR SETUP ...................................................................................................................................................10 Set Time and Date .....................................................................................................................................................10 Audio Setup................................................................................................................................................................10 Alarm Tone ................................................................................................................................................................11 Trend Display ............................................................................................................................................................11
TABLE OF CONTENTS
ii Model 2000 Service Manual
PULSE OXIMETRY MONITORING ....................................................................................................................12 Principles of Operation .............................................................................................................................................12 Warnings/Cautions....................................................................................................................................................14 Pulse Oximetry Sensors ............................................................................................................................................14 Monitoring Procedure...............................................................................................................................................14 Alarm Limits ..............................................................................................................................................................15 Response Settings.......................................................................................................................................................15 Default Values ............................................................................................................................................................15 Reset to Default Values .............................................................................................................................................15 ALARM MESSAGES ...............................................................................................................................................16 Status Messages .........................................................................................................................................................16 MONITOR TESTING ..............................................................................................................................................17 BATTERY OPERATION.........................................................................................................................................18 Battery Replacement .................................................................................................................................................18 CLEANING AND DISINFECTION........................................................................................................................19 The Monitor ...............................................................................................................................................................19 Cleaning and Reuse of Sensors.................................................................................................................................19 Patient Cables ............................................................................................................................................................20 ACCESSORIES .........................................................................................................................................................21 Reorder Numbers ......................................................................................................................................................21 Instructions for use....................................................................................................................................................22 Disposal.......................................................................................................................................................................22 TESTING & TROUBLESHOOTING .....................................................................................................................27 Required Equipment .................................................................................................................................................27 Power Supply/Charges Tests ....................................................................................................................................27 Audio Tests.................................................................................................................................................................28 Safety Tests ................................................................................................................................................................28 Rear Panel Connectors..............................................................................................................................................29 ERROR MESSAGES ................................................................................................................................................31 REPLACEMENT PARTS ........................................................................................................................................32 DISASSEMBLY & ASSEMBLY .............................................................................................................................33 SCHEMATICS ...........................................................................................................................................................A BOARD LAYOUTS ................................................................................................................................................... B
WARRANTY
Model 2000 Service Manual iii
WARRANTY All products manufactured by Ivy Biomedical Systems, Inc. are warranted to be free from defects in material and workmanship and to operate within published specifications, under normal use, for a period of one year from date of original shipment. LNOP® sensors purchased from Ivy Biomedical are warranted for 6 months. If an examination, by Ivy Biomedical Systems, Inc., discloses such products or component parts to have been defective, then our obligation is limited to repair or replacement (at our option). Fuses and batteries are not covered under this warranty.
DESCRIPTION OF LABELS
iv Model 2000 Service Manual
Description of Labels Used Attention, consult ACCOMPANYING DOCUMENTS before attempting to change
power supply selection or carry out interconnections. Equipment connected should comply with IEC-601-1 or IEC-950 with configuration to IEC-601-1-1.
⎯♥⎯ Type CF equipment, Defibrillator proof. _______ Fuse type/rating.
= Power OFF, = Power ON.
∼ Alternate Current (AC)
ON – only for a part of equipment
STBY – off, only for a part of equipment
Contrast
Protective earth (ground) adjacent to this symbol. Internal connection on mains connector/filter.
Equipotential earth connector adjacent to this symbol.
INTRODUCTION
Model 2000 Service Manual 1
INTRODUCTION This manual is to provide information on the correct use of the Model 2000 Pulse Oximeter. It is up to the user to ensure that any applicable regulations respecting the installation and operation of the monitor are observed. The Model 2000 Pulse Oximeter is to be operated by qualified personal only. Using This Manual We recommend that you read this entire manual before operating the equipment. Use the Monitor Description section for general descriptions of controls and displays. For details on pulse oximetry, refer to the section of the manual dealing with that parameter. NOTE: If the monitor beeps when you press a key, it indicates an invalid selection for the current mode or function. Boldface type is used in text to refer to the labeling on user controls. Special brackets [ ] surround menu selections used with the programmable keys. Manufacturer’s Responsibility The manufacturer of this equipment is responsible for the effects on safety, reliability, and performance of the equipment only if: • assembly operations, extensions, re-adjustments, or repairs are carried out by persons authorized by the
manufacturer • the electrical installation complies with all applicable regulations • the equipment is used in accordance with the instructions in this manual Incorrect operation or failure of the user to maintain the monitor in accordance with proper maintenance procedures relieves the manufacturer or his agent from all responsibility for consequent non-compliance, damage, or injury. For technical and service information, please refer to the Ivy Model 2000 Operator's/Service Manual or contact: Ivy Biomedical Systems, Inc. 11 Business Park Drive Branford, Connecticut 06405 (203) 481-4183 or (800) 247-4614 e-mail:[email protected] This manual explains how to set up and use the Model 2000 Pulse Oximeter. Important safety information relating to general use of the pulse oximeter appears in this manual. Other important safety information is located throughout the manual where appropriate. READ THE ENTIRE SAFETY INFORMATION SECTION BEFORE YOU OPERATE THE PULSE OXIMETER.
SAFETY
2 Model 2000 Service Manual
SAFETY
Electrical Medical Electrical Equipment. Classified by Underwriters Laboratories Inc. ® with respect to electric shock, fire, mechanical and other specified hazards only in accordance with UL2601-1 CAN/CSA C22.2 No. 601.1. This product is intended to be operated from a mains power source of nominally 100 V ~ to 230 V ~, 47-63 Hz, or from its internal battery. WARNING: For continued protection against fire hazard, replace only with same type and rating of fuse. Fuse value is T 1A 250V 5 mm x 20 mm. WARNING: To prevent electrical hazards to all personnel, this monitor must be properly grounded. The power cable supplied with this equipment provides for this protection. Do not attempt to defeat this protection by modifying the cable or by using ungrounded adapters. Do not connect to an electrical outlet controlled by a wall switch or dimmer.
WARNING: Electric shock hazard! Do not remove covers or panels. Refer service to qualified service personnel.
WARNING: To avoid electrical shock, disconnect the monitor from its power source before changing fuses.
WARNING: Do not clean monitor while it is on and/or plugged in. WARNING: When attached to other devices, insure that the total chassis leakage currents of all units do not exceed 300µA.
Explosion DANGER: Explosion hazard! Do not use this equipment in the presence of flammable anesthetics.
SAFETY
Model 2000 Service Manual 3
Patient Connections Leakage current is limited internally by this monitor to less than 10 µA. However, always consider additional leakage current that can be caused by other equipment used on the patient at the same time as this monitor. Carefully route patient cables to reduce the possibility of patient entanglement or strangulation.
Measurements If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by alternate means and then check the Model 2000 Pulse Oximeter for proper functioning. Inaccurate measurements may be caused by:
• Incorrect sensor application or use • Significant levels of dysfunctional hemoglobins (e.g., carboxyhemoglobin or methemoglobin) – see note. • Intravascular dyes such as indocyanine green or methylene blue – see note. • Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon light source),
bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight (exposure to excessive illumination can be corrected by covering the sensor with a dark or opaque material)
• Excessive patient movement • Venous pulsations • Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line
The Model 2000 can be used during defibrillation, but the readings may be inaccurate for a short time. Loss of pulse signal can occur in any of the following situations:
• The sensor is too tight • There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp or sunlight • A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensor attached • The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia • There is arterial occlusion proximal to the sensor • The patient is in cardiac arrest or is in shock
NOTE: A pulse oximeter measures functional hemoglobin saturation. This is different to an arterial blood gas that measures fractional hemoglobin saturation. Differencies can occur when significant levels of HbCO (carboxyhemoglobin), HbMET (methomoglobin) indocynanine green or methylene blue exist. EMC This equipment has been certified to be protected to emissions and immunity according to IEC-601-1-2.
MONITOR DESCRIPTION
4 Model 2000 Service Manual
MONITOR DESCRIPTION The Model 2000 Pulse Oximeter is intended for continuous noninvasive monitoring of functional oxygen saturation or arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in a hospital and mobile environment within the hospital. Specifications Performance Range: Saturation (%SpO2) 1% - 100% Pulse Rate (bpm) 25 - 240 Accuracy: Saturation (%SpO2) – During no motion conditions Adults 70% - 100% ± 2 digits 0% - 69% unspecified Neonates 70% - 100% ± 3 digits Saturation (%SpO2) – During motion conditions Adults 70% - 100% ± 3 digits 0% - 69% unspecified Pulse (bpm) – During no motion conditions 25 to 240 ± 3 digits Pulse (bpm) – During motion conditions 25 to 240 ± 5 digits Resolution Saturation(%SpO2) 1% Pulse Rate (bpm) 1 Display Type: Backlit LCD Pixels: 240 x 64 dots Dot Pitch: 0.53 mm Data Displayed: Pulse rate, SpO2 %, Pleth wave, Alarms, Trends, Status messages
MONITOR DESCRIPTION
Model 2000 Service Manual 5
Power Requirements Voltage Input Range: 100-230 V~, 47-63 Hz Maximum AC Power Consumption: 55 VA Communications Serial Port RS232, 9600 baud Mechanical Size: 4.26in (10.65cm)H x 9.75in (24.4cm) W x 9.75in (24.4cm) D Weight: 6.5 lbs (2.95 kg) Battery Type: Sealed gel cell, DOT and IATA approved Operating Time: 2 hours Charge Time: <8 hours Classification (in accordance with IEC601-1) Protection against electric shock: Class 1 equipment Degree of protection against electric shock: Type CF equipment Degree of protection against harmful ingress of water: Drip-proof equipment Methods of sterilization or disinfection: See page 19 Degree of safety of application in the presence of a famable anaesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a fammable anaesthetic mixture Mode of operation: Continuous
MONITOR DESCRIPTION
6 Model 2000 Service Manual
Controls and Indicators Fixed Keys
Programmable keys Each of the keys below the display has a normal function, printed on the key, and several menu functions displayed on the screen in the setup mode. The normal functions are described below. Menu functions are described in the menu descriptions later in this manual.
Displays menus for all pulse oximeter settings. See the Menu Structure section of this manual for a description of the menus.
Turns the alarm audio on and off. Press once to silence the alarms for 120 seconds (two minutes). Press and hold for five seconds to silence the alarms until this button is pressed again.
On/Standby: Controls power to the monitor’s electronic circuits. The rear panel mains switch must be on for this key to be effective.
Resets the audible and visual indicators for an alarm that has been activated.
Displays Trend data. To define the trend display, see the trend display setup instructions.
Adjusts the contrast of the display. Press for small adjustments. Press and hold for larger adjustments
Initiates a printout from the optional printer.
MONITOR DESCRIPTION
Model 2000 Service Manual 7
Display Waveform Display The plethysmographic waveform is displayed from left to right. Alphanumeric Data The following data appears on the display: Pulse Rate- Lower left-hand corner of the display – values updated once per second. Oxygen Saturation (SpO2) - To the left of the waveform – values updated once per second. Alarm Limits Display SpO2 Alarm limit settings are displayed to the right of the SpO2 reading. Pulse rate alarm limit settings are displayed to the right of the pulse rate reading. Alarm Displays When an alarm setting has been exceeded, a flashing alarm tag appears on the screen, indicating which alarm has been triggered. Trend Displays Graphic display of up to 24 hours of averaged saturation. The pleth waveform is not displayed while the trend is displayed. Upper/Lower SpO2 Limits Red LED alert window
Upper/Lower Pulse Rate Limits
SpO2 %
BPM
98 70 LIMITS VOLUME DATE/TIME RESPONSE
99 90
150 50
MONITOR DESCRIPTION
8 Model 2000 Service Manual
Alarms The following is a list of alarms and ranges for which they can be set. When the alarm triggers. Audio alarm tone sounds if the ALARM PAUSE button has not been activated. Pressing ALARM RESET turns off any alarms that have triggered. If the measurement returns to within the limits, the alarm turns off. The ALARM PAUSE key disables the alarms for two minutes or until ALARM PAUSE is pressed again. High Pulse Rate: 50 to 235 bpm in 5 bpm increments Low Pulse Rate: 30 to 150 bpm in 5 bpm increments High SpO2: 70 to 99%in 1% increments Low SpO2: 50 to 95% in 1% increments Sensor Off: The sensor is not properly attached to the patient. Rear Panel The following are located on the rear panel. COM 1: A digital interface for network communication (see note). PRINTER: A connection for the optional printer (see note). MAINS (A.C.) POWER SWITCH: A switch to control mains power input to the monitor. PEQ GROUND: Potential Equalization- A ground connection that can be used to ensure that no potential differences can develop between this equipment and the other electrical equipment.
Note: When using COM1 or PRINTER rear panel connectors, the Ferrite Filter Adaptor Connector must be used. Ivy Biomedical accessory part number 150027 Note: Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards and essential performance standards, and/or the system configuration must meet the requirements of the IEC 601-1-1 medical electrical systems standard. ACCESSORIES (EQUIPMENT) – The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include: • use of the accessory in the PATIENT VICINITY; and • evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate
IEC 601-1 and/or IEC 601-1-1 harmonized national standard.
MONITOR DESCRIPTION
Model 2000 Service Manual 9
Menu Structure Settings for all options for the Model 2000 are made through a series of menus that appear at the bottom of the display. You access the menus by pressing the MENU key. When menus are displayed, press MENU again to return to normal display. NOTE: If the monitor beeps when you press a key, it indicates an invalid selection for the current mode or function. Press MENU to display the main menu, which allows you to select alarm limits, volume, date/time, and response menus. When the main menu is displayed, press MENU again to return to normal display.
Main Menu Soft Keys
SpO2%
BPM
98 70 LIMITS VOLUME DATE/TIME RESPONSE
99 90
150 50
MONITOR SETUP
10 Model 2000 Service Manual
MONITOR SETUP To setup the instrument for operation: 1. Plug the ac line cord into a power source providing the proper voltage and make sure the rear panel power
switch is in the ON position. 2. Press the ON/STBY push button at the left of the front panel. If the internal battery is completely dead wait
one minute before pressing the ON/STBY push button. 3. Connect the sensor and the pulse oximetry cable to the SpO2 connector on the front panel. Set Time and Date 1. Press the MENU key to display the main menu. Then select [DATE/TIME]. 2. The first setting is for MONTH. Use the and keys to increase or decrease the month setting. 3. Select [SET MONTH] to move to the DAY setting. Use the and keys to increase or decrease the day
setting. 4. Select [SET DAY] to move to the YEAR setting. Use the and keys to increase or decrease the year
setting. 5. Select [SET YEAR] to move to the HOUR setting. Use the and keys to increase or decrease the hour
setting. 6. Select [SET HOUR] to move to the MINUTE setting. Use the and keys to increase or decrease the
minute setting. 7. When all date and clock settings are correct, select [ENTER] to enter the settings into the monitor’s
memory. 8. Press MENU to return to normal display. Audio Setup 1. Press the MENU key to display the main menu. 2. Select [VOLUME] to display the menu for setting pulse and alarm volume. 3. Adjust the pulse volume using [PULSE and ] keys – OFF, 20%, 40%, 60%, 80% or 100%.
MONITOR SETUP
Model 2000 Service Manual 11
4. Adjust the alarm volume using [ALARM and ] keys – 20%, 40%, 60%, 80% or 100%. 5. Press MENU to return to the normal display. Alarm Tone To change the alarm tone: 1. Press the ON/STBY push button to put the monitor in Stand-by. 2. Press and hold the TREND key while simultaneously pressing ON/STBY push button. 3. Adjust the alarm tone using [ALM TONE and ] keys - LOW, MEDIUM or HIGH. 4. Press the MENU key to return to normal display. Trend Displays The Model 2000 has a trend display showing SpO2. Time selections for this trend display are 30min, 2, 8, 12 and 24 hours. Trend Setup 1. Press the TREND key to select 30min., 2, 8, 12 or 24 hour trend. 2. Press TREND to return to the normal display.
PULSE OXIMETRY MONITORING
12 Model 2000 Service Manual
PULSE OXIMETRY MONITORING Overview Pulse oximetry allows you to continuously and noninvasively monitor a patient’s hemoglobin oxygen saturation.The oximetry sensor contains two light emitting diodes (LEDs) that transmit specific wavelengths of light which are received by a photodetector. Oxygen saturated blood absorbs light differently as compared to unsaturated blood. Thus the amount of light absorbed by blood can be used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood. The monitor displays this ratio as percent SpO2. Normal values typically range from 95 to 100% at sea level. Principles of Operation The Model 2000 Pulse Oximeter is based on three principles: 1. Oxyhemoglobin and deoxyhemoglobin differ in their absorption of red and infrared light (spectrophotometry). 2. The volume of arterial blood in tissue and the light absorbed by the blood changes during the pulse (plethysmography). 3. Arterio-venous shunting is highly variable and that fluctuating absorbance by venous blood is a major component of noise during the pulse. The Model 2000 Pulse Oximeter as well as traditional pulse oximetry determines SpO2 by passing red and infrared light into a capillary bed and measuring changes in light absorption during the pulsatile cycle. Red and infrared light-emitting diodes (LEDs) in oximetry sensors serve as the light sources, a photodiode serves as the photodetector. Traditional pulse oximetry assumes that all pulsations in the light absorbance signal are caused by oscillations in the arterial blood volume. This assumes that the blood flow in the region of the sensor passes entirely through the capillary bed rather than through any arterio-venous shunts. The traditional pulse oximeter calculates the ratio of pulsatile absorbance (AC) to the mean absorbance (DC) at each of two wavelengths, 660 nm and 940 nm. S(660) = AC(660)/DC(660) S(940) = AC(940)/DC(940) The oximeter then calculates the ratio of these two arterial pulse-added absorbance signals: R = S(660)/S(940) The value of R is used to find the saturation SpO2 in a look-up table built into the oximeter’s software. The values in the look-up table are based upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in induced hypoxia studies.
PULSE OXIMETRY MONITORING
Model 2000 Service Manual 13
The Model 2000 Pulse Oximeter assumes that arterio-venous shunting is highly variable and that fluctuating absorbance by venous blood is the major component of noise during the pulse. The 2000 decomposes S(660) and S(940) into an arterial signal plus a noise component and calculates the ratio of the arterial signals without the noise: S(660) = S1 + N1 S(940) = S2 + N2 R = S1/S2 Again, R is the ratio of two arterial pulse-added absorbance signals and its value is used to find the saturation SpO2 in an empirically derived equation into the oximeter’s software. The values in the empirically derived equation are based upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in induced hypoxia studies. The above equations are combined and a noise reference (N’) is determined: N' = S(660) - S(940) x R If there is no noise N' =0; then S(660) = S(940) x R which is the same relationship for the traditional pulse oximeter. The equation for the noise reference is based on the value of R, the value being seeked to determine the SpO2. The Model 2000 software sweeps through possible values of R that correspond to SpO2 values between 1% and 100% and generates an N' value for each of these R values. The S(660) and S(940) signals are processed with each possible N' noise reference through an adaptive correlation canceler (ACC) which yields an output power for each possible value of R (i.e., each possible SpO2 from 1% to 100%). The result is a Discrete Saturation Transform (DSTTM) plot of Relative output power versus possible SpO2 value as shown in the following figure, where R corresponds to SpO2 = 97%. The red light power range is 1mw to 3 mw. The infrared light power range is 0.7mw to 3mw.
The DST plot has two peaks: the peak corresponding to the higher saturation is selected as the SpO2 value. This entire sequence is repeated once every two seconds on the most recent four seconds of raw data. The Model 2000
PULSE OXIMETRY MONITORING
14 Model 2000 Service Manual
SpO2 therefore corresponds to a running average of arterial hemoglobin saturation that is updated every two seconds. WARNING: The pulse oximeter should NOT be used as an apnea monitor. WARNING: The pulse oximeter should be considered an early warning device. If a trend towards patient deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the patient’s condition. Caution: Use only sensors specified by Ivy Biomedical Systems, Inc. If you use sensors other than those specified, it might degrade perfomance and could damage the pulse oximeter. Caution: Check the sensor site frequently. Do not allow the sensor to remain on one site for a prolonged period of time, especially when monitoring neonates. Refer to the sensor’s specific instructions. Caution: Never attach a SpO2 sensor on a limb being monitored with a blood pressure cuff or a limb with restricted blood flow. Caution: A poorly applied sensor might give incorrect saturation values. Reapply the sensor. Caution: Choose a site with sufficient perfusion to ensure accurate oximetry values. Caution: Certain nail aberrations, nail polish, fungus, etc. might give inaccurate oximetry values. Remove the nail polish and/or move the sensor to an unaffected digit. Caution: Do not use the Model 2000 Pulse Oximeter or sensors during Magnetic Resonance Imaging (MRI). Pulse Oximetry Sensors To ensure conformance with all safety and performance specifications, use only the Masimo® LNOP pulse oximetry sensors. Monitoring Procedure Use the following procedure for monitoring pulse oximetry: 1. Choose a site that is well perfused and provides proper alignment of the LEDs and receiving photodetector Select a site that has unrestricted blood flow Do not restrict blood flow when securing sensor with tape Select an appropriate sensor and apply it to the patient, following the directions for use provided with the sensor Do not select a site near potential electrical interference (electrical cords, for example). 2. Connect the cable to the monitor’s SpO2 input. 3. Plug the extension cable into the sensor. 4. Use the procedures described in the following sections for alarm limit settings, response mode settings, and enabling or disabling the SpO2 alarm. Alarm Limits 1. Press the MENU key to display the main menu.
PULSE OXIMETRY MONITORING
Model 2000 Service Manual 15
2. Select [LIMITS]. 3. Press ALARM RESET to select SpO2 or PULSE PARAMETERS. 4. Use the [HIGH and ] selections to set the high limit. 5. Use the [LOW and ] selections to set the low limit. Each time you press a key, the corresponding limit changes by 1% for SpO2 and 5 BPM for Pulse. The
current settings are shown on the display. 6. Press MENU to return to the normal display. Response Mode Settings 1. Press MENU to display the main menu. Then select [RESPONSE]. 2. Use the [AVERAGING] selection to set the response time – 6, 8, 10, 12, 14 or 16 sec. 3. Use the [SENSITIVITY] selection to set the sensitivity level. (Options – Normal and High). The High
sensitivity level should be used when the clinician wants to have the absolute low profusion performance of the Masimo SET® and is willing to sacrifice some sensor off detection capability.
4. Press MENU to return to the normal display. Default Values The following defaults are set: 1. High SpO2 limit – set to OFF. 2. Low SpO2 limit – set to 85. 3. High pulse rate limit – set to 150 bpm. 4. Low pulse rate limit – set to 50 bpm. 5. Pulse volume – set to 40% 6. Alarm volume – set to 100% 7. Averaging – set to 8 sec. 8. Sensitivity – set to Normal. To reset the model 2000 to the above factory default values: 1. Press the ON/STBY push button to put monitor in stand by. 2. Press and hold ALARM RESET push button while simultaneously pressing ON/STBY push button. 3. Listen for the audio to sound 8 times in succession – defaults are now reset to factory default values.
ALARM MESSAGES
16 Model 2000 Service Manual
ALARM MESSAGES
Alarm messages indicate a problem or condition which may affect accurate monitoring values. Do not ignore these messages. Correct any fault before continuing. Whenever it is necessary to perform procedures on a patient that might cause activation of an alarm, press the ALARM PAUSE key to disable the alarms for five minutes. Alarm indications flash on the display, and the red front panel alarm indicator flashes. When the alarm triggers, the alarm indicator flashes and the audio alarm tone sounds. The displayed numeric value of the limit exceeded will be in reverse video and flashing. Pressing ALARM RESET turns off any alarms that have triggered. If the measurement returns to within the limits, the alarm turns off. These alarms will signal when: The high pulse rate limit has been exceeded for greater than three seconds. The low pulse rate limit has been exceeded for greater than three seconds. The low SpO2 limit has been exceeded for greater than three seconds. The high SpO2 limit has been exceeded for greater than three seconds.
STATUS MESSAGES The following is a list of module status messages used in the Model 2000. Message Meaning NO SENSOR There is no sensor plugged into end of cable. MODULE FAIL SpO2 module failure. SENSOR FAIL SpO2 sensor failure. CHECK CABLE Interconnect cable failure. SENSOR OFF SpO2 sensor not in contact with patient. SEARCHING Unit is searching for patient's pulse. INTERFERING Outside signal or energy preventing reading. HI AMBIENT Too much light on patient. L PERFUSION Signal too small. Notes: Above messages (all messages except L PERFUSION) cause the SpO2 and PR numerics to be replaced with dashes and are accompanied by an audible alert tone. In addition, a SYS FAULT message is displayed whenever communications with MS-1 module has been lost. This message flashes in reverse video and is always accompanied by an audible alert tone.
MONITOR TESTING
Model 2000 Service Manual 17
MONITOR TESTING Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local or governmental regulations. In the event that internal adjustment or recalibration is necessary, refer to the Service Manual for this equipment. Alarm Testing Place the sensor on your finger. After the readings have stabilized change the SpO2 and Pulse Rate alarm limits to cause an alarm. Verify that the SpO2 and Pulse Rate alarms function when their settings are violated. The audio should sound and the red alarm LED should flash along with the limit violated.
BATTERY OPERATION
18 Model 2000 Service Manual
BATTERY OPERATION The model 2000 contains a 2.3 amp-hr 12V lead acid battery. Under normal use and full charge the unit will operate for greater than 2 hours continuously. The unit will recharge in less than 8 hours. In the event of mains power failure, the model 2000 will automatically switch to battery operation. When the remaining battery charge is less than 20 minutes a BAT LOW indicator in the lower right hand corner of the display will flash and the audio will sound every 15 seconds. The battery is continuously charged whenever the AC power is connected and the mains power switch on the rear panel is on. To ensure maximum battery life, it is recommended that, once a month, the unit be run on battery until it turns itself off and then connect to mains power to recharge the battery. Battery Replacement 1. Remove the 6 screws on the rear of the unit
2. Slide cover off unit
3. Locate battery tray and remove 2 screws that hold down battery
4. Lift battery and un-clip connector from edge of battery
5. Install new battery, “CAUTION, replace only with same type 2.3 amp-hr 12 Volt battery”.
6. Reverse step 1 thru 4
7. Plug unit into line and charge for at least 8 hours
8. Unplug unit and operate for 2 hours or until instrument shuts off to confirm proper operation.
NOTE: Sealed lead battery must be recycled or disposed of properly. Do not incinerate
CLEANING AND DISINFECTION
Model 2000 Service Manual 19
CLEANING AND DISINFECTION The Monitor When necessary, clean the exterior surfaces of the monitor with a cloth or swab dampened with a warm water and mild detergent solution. Do not allow liquids to enter the interior of the instrument. Unplug monitor from mains power. Cleaning and Reuse of Sensors Reusable sensors can be cleaned as follows: 1. Remove the sensor from the patient. 2. Disconnect the sensor from the monitor. 3. Wipe the entire sensor with a 70% isopropyl alcohol pad. 4. Allow the sensor to air dry before returning it to operation. Reattachment of Single Use Sensors Single use sensors can be reapplied to the same patient if the emitter and detector windows are clear and the adhesive still adheres to the skin. The adhesive can be partially rejuvenated by wiping with an alcohol wipe and allowing the sensor to thoroughly air dry prior to replacement on the patient. NOTE: If the sensor fails to track the pulse consistently, the sensor might be incorrectly positioned. Reposition the sensor or choose a different monitoring site. Caution: Do not soak or immerse the sensor in any liquid solution. Do not sterilize any sensor by irradiation, steam, or ethylene oxide.
CLEANING AND DISINFECTION
20 Model 2000 Service Manual
Patient Cables Patient cables can be cleaned using the following procedure: 1. Remove the cable form the sensor. 2. Disconnect the cable from the monitor. 3. Wipe clean with a 70% isopropyl alcohol pad. 4. Allow the cable to dry before returning it to operation. Caution: Do not soak or immerse patient cables in any liquid solution. Do not sterilize any patient cable by irradiation, steam, or ethylene oxide.
ACCESSORIES
Model 2000 Service Manual 21
ACCESSORIES Reusable Sensors Adult LNOP DC-1 Finger Sensor, Reusable 590227 Disposable Sensors Neonatal LNOP-Neo, Disposable Sensor, Box of 20 590223 Pediatric LNOP-Pdt, Slender Digit Disposable Sensor, Box of 20 590222 Neonatal Pre-term LNOP-NeoPt, Disposable Sensor, Box of 20 590224 Adult LNOP-Adt, Disposable Sensor, Box of 20 590221 Patient Cables Patient Cable, 8 ft PC08 with LNOP Sensor Connector 590220 Patient Cable 12 ft PC12 with LNOP Sensor Connector 590230 Starter Kits Starter Kit with 2 LNOP-Adt and 2 LNOP-Pdt sensors, and 12’ PC12 sensor cable 590225 Starter Kit with 2 LNOP-Neo and 2 LNOP-NeoPt sensors, and 8’ PC08 sensor cable 590226 Starter Kit with Adult LNOP-DC1 finger sensor, reusableand PC08 sensor cable 590228 Miscellaneous Ferrite Filter Connector Adaptor 150027 Disposal Disposal of devices or consumables must be done in accordance to natural environmental laws.
22
IVY LNOP® SENSORS WITH MASIMO SET® DISPOSABLE NON-STERILE SENSORS
Indications: • Continuous SpO2 monitoring • Single patient use • Use only with IVY monitors with Masimo
SET
Patient Size: Weight Sensor
Type Accuracy* Ivy Biomedical
Part Number Over 30 kg Adt +/- 2 digits 590221 10 to 50 kg Ped +/- 2 digits 590222 Under 10kg Neo +/- 3 digits 590223 Under 1 kg Neo Pt +/- 3 digits 590224 *From 70 to 100% SpO2 to +/- 1 Std.Dev.
Contraindications: • Patients allergic to adhesive tape
Adults Over 30kg – Adt Sensor Site Selection Attachment to Patient P/N: 590221 • Well perfused • Restricts movement the
least • Non dominate hand
preferred • Ring or middle finger or • Great or long toe
Step A • Remove sensor from
pouch • Hold tan side down,
remove backing • Press detector onto
fleshy tip of finger
Step B • Wrap “T” shaped
adhesive around finger
Step C • Fold emitter end
over nail
Step D • Wrap wings around
finger • Ensure emitter and
detector face each other
• Reposition if necessary
Step E • Position sensor with shiny
contacts up • Position cable with Masimo
SET logo up • Slide sensor contacts into
cable, hear click • Tug to confirm connection • Tape cable to patient if
needed
23
Pediatrics 10 to 50kg – Ped Sensor P/N:590222 Site Selection Attachment to Patient • Well perfused • Restricts movement the least • Non dominate hand
preferred • Ring or middle finger
Step A • Remove sensor
from pouch • Hold tan side down,
remove backing • Press detector onto
fleshy tip of finger
Step B • Wrap “T” shaped
adhesive around finger
Step C • Fold emitter end
over nail
Step D • Wrap wings around
finger • Ensure emitter and
detector face each other
• Reposition if necessary
Step E • Position sensor with shiny
contacts up • Position cable with Masimo
SET logo up • Slide sensor contacts into
cable, hear click • Tug to confirm connection • Tape cable to patient if
needed
Neonates under 10 kg or with fat feet - Neo Sensor P/N:590223 • Neonates: foot or palm • Infants: big toe • Well perfused
Step A • Remove sensor
from pouch • Hold along the “Y”,
remove backing
Step B • Locate detector • Position on fleshy
sole of foot • Align with fourth
toe
Step C • Position emitter on
top of foot • Wrap bandage
around foot • Use bandage to
secure emitter and detector
Step D • Ensure emitter and
detector face each other
• Ensure detector window is fully covered
• Reposition if necessary • Use big toe on infants
over 3 to 5 kg, or with fat feet.
Step E • Position sensor with shiny
contacts up • Position cable with Masimo
SET logo up • Slide sensor contacts into
cable, hear click • Tug to confirm connection • Tape cable to patient if
needed
24
Neonates Under 1 kg – Neo Pt Sensor P/N:590224 Site Selection Attachment to Patient • Neonates: foot or
palm • Could be to
bulky on big toe • Well perfused
Step A • Remove sensor
from pouch • Hold along the “Y”,
remove backing
Step B • Locate detector • Position on fleshy
sole of foot • Align with middle toe
Step C • Position emitter on
top of foot • Wrap sponge wrap
around foot • Use wrap to secure
emitter and detector
Step D • Ensure emitter and
detector face each other
• Ensure detector window is fully covered
• Reposition if necessary
Step E • Position sensor with shiny contacts
up • Position cable with Masimo SET
logo up • Slide sensor contacts into cable, hear
click • Tug to confirm connection • Tape cable to patient if needed
Sensor Reattachment • Sensors can be reapplied to the same patient if:
− Emitter and detector windows are clear − Adhesive still adheres
• Adhesive may be rejuvenated: − Thoroughly clean with alcohol wipe − Let sensor air dry before reattachment
Disconnecting Sensors • Squeeze gray buttons on both sides of connector • Pull out sensor tab from connector
Warnings • Check site every 8 hours for:
− Proper adhesion to patient − Alignment of emitter and detector − Reposition if required
• With poorly perfused patients: − Check site every 2 hours − Check for skin damage, pressure necrosis,
distal circulation − Reposition if required
• Do NOT: − Use damaged sensors − Immerse or soak sensors − Use with MRI scanners − Place on same limb as Art line or BP cuff
• Sources of incorrect readings
− Intravascular dyes − Elevated Carboxyhemoglobin (COHb) − Elevated Methemoglobin (MetHb) − Mis-aligned emitter and detector (must face
each other vertically) − Incompletely covered detector window
• High O2 levels may predispose premature infants to retinopathy − Set the high SpO2 alarm limit as required
25
ADULT REUSABLE NON-STERILE SENSOR P/N:590227 Indications: • Adults over 30 kg • Spot ckeck or Continuous SpO2 monitoring • Low motion, clean environments • Use only with IVY monitors with Masimo SET
Accuracy ±2 digits, from 70 to 100% SpO2 to ±1 Std.Dev.
Contraindications: • High motion • Dirty environments • Long term monitoring over 4 hours
Adults Over 30kg – DC1 Sensor Site Selection Attachment to Patient • Well perfused • Restricts movement the least • Non dominate hand preferred • Ring or middle finger or • Great or long toe
Step A • Place digit over sensor window • Fleshy tip of finger should cover window
Step B • Tip of finger should touch stop • Long nail can extend over stop
Step C • Rear hinge may open • Verify positioning • Verify detector
window is fully covered
Step D • Orient cable to exit
along top of hand
Step E • Position sensor with shiny
contacts up • Position cable with Masimo
SET logo up • Slide sensor contacts into cable,
hear click • Tug to confirm connection • Tape cable to patient if needed
26
Disconnecting Sensor • Squeeze gray buttons on both sides of connector • Pull out sensor tab from connector
Cleaning • Remove sensor from patient • Disconnect sensor form cable
• Wipe clean with 70% Isopropyl Alcohol pad • Allow to dry • Do NOT soak, immerse or sterilize sensor Warnings • Check site every 4 hours for:
− Skin damage, pressure necrosis, distal circulation − Change to disposable sensor if required
• With poorly perfused patients, check site every 2 hours
• Do NOT: − Use damaged sensors − Immerse or soak sensors − Use with MRI scanners − Place on same limb as Art line or BP cuff
• Sources of incorrect readings − Intravascular dyes − Elevated Carboxyhemoglobin (COHb) − Elevated Methemoglobin (MetHb) − Mis-aligned emitter and detector (must
face each other vertically) − Incompletely covered detector window
INSTRUMENT CABLES Directions for Use • Connect LNOP sensors to monitor • For use only with IVY monitors with Masimo SET • PC 08 is 8’ (2.4m) long –P/N:590220 • PC 12 is 12’ (3.7m) long –P/N:590230 • See above for attaching sensors
Cleaning • Wipe clean with 70% Isopropyl Alcohol pad • Allow to dry • Do NOT soak, immerse or sterilize cable
Warnings • Do NOT:
− Use cable if it is damaged − Connect to another patient cable − Connect to a monitor without Masimo SET
• Ensure cable is properly connected • Hold only by connectors when attaching
sensors
TESTING & TROUBLESHOOTING
Model 2000 Service Manual 27
TESTING & TROUBLESHOOTING Required Equipment Hipot Tester Hypotronics Model HDA5 Variable AC Supply Powerstat Model 3PN216C or equivalent Oscilloscope Tektronix Model 2236 or equivalent DMM Fluke Model 8050A or equivalent Power Supply BK Precision Model 1630 or equivalent Masimo Sensor 590227 Masimo Patient Cable 590220 Bio Tek Index 2 SpO2 Simulator Capacitor/Resistor Network 0.15µf@100V, 5% metalized polyester 1kΩ 1/4W, 1% 10kΩ 1/4W, 1% Ground Isolator Plug Power Supply/Charger Tests 1. Apply power with the MAINS power switch and check for the front panel GREEN (AC ON) LED to light. 2. Monitor the +13.7V supply at VB (TP4) with the scope. Adjust RV1 for +13.80 to +13.85V with fully
charged batteries (<.1mV across R10 DMM from F1 back to AGND). 3. Turn the front panel power ON. The voltages should decrease but must be ≥+13.60V. Verify that the front
panel GREEN (ONSTBY) LED lights. 4. Turn the MAINS power OFF. Operate the monitor on battery for approximately 1 minute. Connect the
DMM across R10 (from F1 back to AGND). Turn the MAINS power back ON. Check for the YELLOW (CHARGING LED) to light. The voltage across R10 should be 4.36mV to 5.78mV.
5. While the CHARGING LED is ON, monitor the +12V supply at P105. 6. Turn the front panel power OFF. The voltage at VB (TP4) must be +14.7V to +15.0V. 7. Check the YELLOW (CHARGING LED) to turn off when the voltage at VB (TP4) switches to +13.7
(+13.60V to +13.85V). 8. Turn the front panel power on. Measure the following supplies with the scope referenced to DGND
(TP16): +5V (+4.85V to +5.15V) VCC TP13 B++ (VB≤100mV) TP5 -12V (-11.64V to –12.36V) U13 pin 2
9. Reference the scope to ISOLATED GROUND TP11 and measure the following supplies: +6.2V (+6.16V to +6.35V) TP6 +20V (+26.5V to +32.5V) TP8 +15VI (+14.7V to +15.3V) TP10 -15VI (-14.4V to –15.63V) TP9 +5VI (+4.85V to +5.15V) TP7
TESTING & TROUBLESHOOTING
28 Model 2000 Service Manual
10. Turn the rear panel MAINS power off. Disconnect the battery. Connect the power supply to J101. Connect the positive to J101 pin 1 and negative to J101 pin 2. Adjust the external DC supply for +12.0V. Apply power with the front panel power switch.
Connect the scope to VBF (F2 front). Slowly decrease the voltage on the supply until the LOW BATTERY message appears on the display. The voltage should be +10.98V to +11.47V. Continue to decrease the voltage on the supply until the monitor shuts off. The voltage should be +9.60V to +10.26 (just before shut off). Increase the voltage on the supply (DMM) to +12V. The monitor should not turn on.
Audio Tests 1. Set the pulse volume to 100%. Connect the scope to the speaker (+) or JP4. Place the sensor on your finger
or connect it to the simulator. Check for 2.5Vp-p±.3Vp-p. Decrease the pulse volume and make sure the amplitude is gradually reduced to zero.
2. Turn the alarm audio on and set it to 100%. The amplitude at the speaker should be 3.0Vp-p±.3Vp-p.
Decrease the alarm volume to minimum and make sure the amplitude is gradually reduced to .5Vp-p±50mVp-p.
Safety Testing
BE SURE THE FRONT PANEL SWITCH IS SET TO STBY FOR THE FOLLOWING TEST 1. Connect the SpO2 hipot plug to the front panel SpO2 input. Connect the RED lead from the hipot tested to
the hipot plug. Connect the Black lead to the rear PEQ ground.
Turn the tester on and slowly increase the voltage to 4kV. There should be no arcing or leakage for one minute. Reduce the voltage to zero.
2. Replace the hipot plug with the SpO2 leakage plug. Using the DMM, select AC Volts and 200mV range.
Connect the leakage filter as shown below. CAUTION: Be careful to connect the circuit below only between the measurement points indicated in the following instructions. Do not connect this circuit across line voltage. Full line voltage would damage the measurement circuit and could possibly damage the monitor and/or the digital voltmeter Measuring instrument R1=10kΩ±5% R2= 1kΩ±1% C1=0.01µF ±5% With the monitor on, measure the leakage from the leakage plug to the HOT side of the line. The leakage must be <10µA at 120 V AC.
R2 R1 C1 V
TESTING & TROUBLESHOOTING
Model 2000 Service Manual 29
Connect the measurement circuit from the rear panel PEQ ground of the monitor to line cord ground. Measure the leakage with the following conditions: Unit Grounded Ground Broken Ground Broken, reversed polarity The leakage must be <30µA. Rear Panel Connectors COM 1: A digital interface for network communication (see note).
Pin Signal Output
1 No Connection
2 RECEIVE DATA RS-232 ±9 V (±5 V min.)
3 TRANSMIT DATA RS-232 ±9 V (±5 V min.)
4 No Connection
5 SIGNAL GROUND Reference for COM 1 signals
6 No Connection
7 REQUEST TO SEND Not Used
8 CLEAR TO SEND Not Used
9 No Connection
When in the Normal Operating Mode, the Model 2000 external RS232 COM port is configured for 9600 baud bi-directional operation with 8 data bits, no parity, 1 stop bit, and no hardware or software handshaking. The output data consists of ASCII text strings sent at one second intervals, containing the date, time, SpO2 value and pulse rate, terminated by ASCII carriage return (hex 0D) and line feed (hex 0A) characters, in the following format:
mm/dd/yy hh:mm:ss SPO2=xxx% PULSE=yyy<cr><lf> PRINTER: A connection for the optional printer (see note).
Pin Signal Output
1 No Connection
2 RECEIVE DATA Not used
3 TRANSMIT DATA RS-232 ±9 V (±5 V min.)
4 No Connection
5 SIGNAL GROUND Reference for Printer signals
6 No Connection
7 REQUEST TO SEND Not Used
8 CLEAR TO SEND RS-232 ±9 V (±5 V min.)
9 No Connection
Mating connector, use Ivy P/N 179004, 9-pin receptacle for 22-26 AWG wire. Use RFI/EMI shielded cable clamp, Ivy P/N 210030, and 5-conductor shielded cable, Ivy P/N 600002.
Mating connector, use Ivy P/N 179004, 9-pin receptacle for 22-26 AWG wire. Use RFI/EMI shielded cable clamp, Ivy P/N 210030, and 5-conductor shielded cable, Ivy P/N 600002.
TESTING & TROUBLESHOOTING
30 Model 2000 Service Manual
MAINS POWER SWITCH: A switch to control mains power input to the monitor. PEQ GROUND: Potential Equalization- A ground connection that can be used to ensure that no potential differences can develop between this equipment and the other electrical equipment.
Note: When using COM1 or PRINTER rear panel connectors, the Ferrite Filter Adaptor Connector must be used. Ivy Biomedical accessory part number 150027 Note: Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards and essential performance standards, and/or the system configuration must meet the requirements of the IEC 601-1-1 medical electrical systems standard. ACCESSORIES (EQUIPMENT) – The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include: • use of the accessory in the PATIENT VICINITY; and • evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate
IEC 601-1 and/or IEC 601-1-1 harmonized national standard.
MONITOR SETUP
Model 2000 Service Manual 31
ERROR MESSAGES Several fault conditions which cause messages and an audible alert tone to occur which are not associated with the MS-1 module, as follows: Message Meaning Caused by ERROR 1 ROM Error U31 (P/N:2356-00-04) and U27 (P/N:2357-00-04) ERROR 2 RAM Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 3 Clock Chip Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 4 Undefined Interrupt Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 5 Timing Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 6 Communications Port Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 9 Unknown Error Pulse Oximeter Board (P/N:5121-00-01)
REPLACEMENT PARTS
32 Model 2000 Service Manual
REPLACEMENT PARTS Part Ivy Part Number LCD Display 570012 Display Cable Assembly 610011 DC-AC Converter 220007 Display Window 2354-00-13 Front Panel 2343-00-10 Power Supply Module 580017 Pulse Oximeter Module, MS-1 580014 LED Board Assembly 5131-00-01 LED Alarm Lens 2401-00-01 SpO2 Input Board Assembly 5122-00-01 Fuse, 2ASB, 2AG, 250V 290003 Fuse, 5ASB, 2AG, 125V 290015 Fuse, 1ASB, Metric, (5x20mm), 250V 290014 Speaker 470000 Battery, 12 Volt, 2.3 Ah 110006
MONITOR SETUP
Model 2000 Service Manual 33
DISASSEMBLY & ASSEMBLY
Covers & Circuit Boards To remove cover: 1. Unplug unit from mains and remove line cord. 2. Locate the six (6) screws on the rear panel and remove. 3. Slide cover off. To remove motherboard: 1. Disconnect all wire sockets after noting their locations. 2. Locate and remove the six (6) mounting screws from bottom of unit. 3. Remove motherboard from bottom of unit. To remove Masimo module: 1. Remove motherboard. 2. Disconnect ribbon cable and programming voltage connector. 3. Remove interconnect PC board. 4. Remove three (3) nylon screws. To remove power supply: 1. Remove the two (2) connectors from the power supply. 2. Locate and remove the two (2) #6 nuts that retain the two ground wires and power supply mounting
bracket. 3. Remove bracket from power supply. CAUTION: When replacing power supply, be sure to replace the two (2) green ground wires to their #6 ground studs. To remove front panel assembly: 1. Remove motherboard. 2. Unplug J106. 3. Locate and remove the four (4) screws (2 on each side) that hold the front panel to side rails. 4. Remove front panel assmbly by carefully disconnecting the interconnect cable.
SCHEMATIC DIAGRAMS
Model 2000 Service Manual A
SCHEMATIC DIAGRAMS
SCHEMATIC DIAGRAMS
A Model 2000 Service Manual
This page is intentionally left blank.
BOARD LAYOUT DIAGRAMS
Model 2000 Service Manual B
BOARD LAYOUT DIAGRAMS
8
8
7
7
6
6
5
5
4
4
3
3
2
2
1
1
D D
C C
B B
A A
ISOLATED NON-ISOLATED
RECEIVE_DATATRANSMIT_DATA
RESET_INPUT-
|LINK
MASIMO
D
S
G
BAT+
BAT-
|C:\ORCAD\SHEET\SH5121\5121-002.SCH
AGND, DGND & PGNDARE ONE COMMON GNDAT J105-3,4,5
REVISIONSDATEBYDESCRIPTIONREV
01 REVISED FOR REV-01 FAB.
ALL RESISTORS ARE 1/4W, 1% U.O.S.1.
3.
NOTES:
2.L3,Q18,R92,RN5,RV1,T1,TP26,U36,Y1.
LAST USED: C81,D16,E1,F2,J110,JP4,
D
S G
02 REVISED FOR REV-02 FAB.ECN 761ECN 762
GPGP
12-22-9712-22-97
03 REVISED FOR REV-03 FAB.
|C:\ORCAD\SHEET\SH5121\5121-003.SCH
ECN 779 GP 5-26-9804 GP 5-26-98REVISED PER ECN 780.05 GP 5-26-98REVISED PER ECN 781.06 REVISED PER ECN 807. GP 7-31-9807 REVISED PER ECN 808. GP 7-31-9808 REVISED PER ECN 809. GP 8-3-9809 REVISED PER ECN 810. GP 8-3-9810 REVISED FOR REV-04 FAB. ECN 845 GP 8-19-98
GP11 REVISED PER ECN 875. 2-3-99GP12 REVISED PER ECN 887. 2-16-99
13 REVISED PER ECN 933. GP 4-9-99
NOT USED: C25,C32,C60,R52,R72,R73.
14 REVISED PER ECN 945. GP 5-20-9915 REVISED PER ECN 968. GP 8-6-9916 REVISED PER ECN 969. GP 8-6-9917 REVISED PER ECN 979. GP 9-16-99
GPGP
1819
REVISED PER ECN 1001.REVISED PER ECN 1002.
6-5-006-5-00
2120 REVISED PER ECN 1088. GP
GP5-11-015-11-01REVISED FOR REV-05 FAB. ECN 1089.
22 REVISED PER ECN 1115. GP 8-6-01
5121-00-01 (-03 SCHEMATIC) 22
PULSE OXIMETER, MODEL 2000
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVEBRANFORD, CT. 06405(203) 481-4183
D
1 2Tuesday, March 08, 2005
Title
Size Document Number Rev
Date: Sheet of
CHARGLED
/BATLO
B+
ON/STBY
TXMASI
RXMASI
AGND
DGND
VCC
UNIT_ON
PGND
B++
BLINV
-12V
MASIRS
EPRM1
I
I
I
D
I
I
I
I
I
A
A
A
A
A
A
AA
A
AA
A
A
A A
A
A
A
A
A
A A
A
A
A
A
D
A
D
A
A
A
A
A
I
D
AAA
I
I
I
A
A
A
+5VI
VCC
VCC
B+
-12A
B+
VB
B++
VCC
B+
B++
VR
VBFVBB+
VBF
VBF
VBF
VBF
VBF
VCC
+5VI
+5VI
VCC
B+
-12A
-12A
+5VI
+6.2V+6.2V
VR
VR
B+
B+
VCC
B+
VBF
+15VI
-15VI
+5VI
+15VI
-15VI
+5VI
VCC
B++
B++
B++
-12V
VCC
Q1
2N3903
13
2
U6PT5101N
1
2
3VI
GN
D
VO
C24
.47UF MP
C38.015UF1KV CD
D12
1N4148
L3
25UH
+
C9100UF25V
TP7+5VI
C111UFMP
+C31
1000UF16V
C6
1UF MP
TP11IGND +C39
100UF50V
+ C37100UF25V
R13100 OHM
Q32N5460
3
21
+C43100UF
50V
TP10+15VI
C181UF MP
Q2
PN43933
12
R37200K
+
C106.8UF
TANT
TP8+20V
RN4
1K5 6
TP9-15VI
Q7RFP30P05S D
G1
3 2
R15499K
R11
200K
D16
1N4937
R8499K
U21CNY17-3
21
45
U4ICL7662
1
2
4
3
6
7
5
8
TEST
CAP+
CAP-
GN
D
LV
OSC
VOUT
V+
R5110K
+C3047UF
35V
R14
10K
Q8
2N3903
13
2
R49
100K
R20
1K
Q13
2N3906
13
2
R90
60.4K
+C406.8UF
TANT
R50
100K
R38
200K
C801UF MP
+
-U11A
LM393
3
21
84
+C42
6.8UFTANT
C44
.1UF CER
+
C141500UF
25V
+
-U11B
LM393
5
67
J110EPRMV
123
J1041639-02-10
SaO2 DIGITALINTERFACE
123456789
1011121314
TP26-12V
C34
.1UFCER
R18
10K
+ C416.8UFTANT
Q92N3906
13
2
U15LM78L12ACZ
VI
GN
D
VO
L1
220UH52
R4510K
R61200K
5ASB2AG
F1
TP6+6.2V
R41200K
D14
1N4148
R44
10K
U5LT1070
3 1 2
45
GN
D
VC FB
VSWVIN
R249.9K
R40
604K
R4
200K
R19
2W 5%150 OHM
R60
698K
R7100K
C4
.1UFCER
U10AD680
3
1
2VIN
GND
VOUT
C5
.1UFCER
R92220 OHM
2W 5%
Q18IRFD120
D
SG23
1
R39200K
R594.99K
R9
1K
R16
1K
U12ACD4013B
5 1
3
2
14 647
D Q
CLK
Q
VD S
RVS
U12BCD4013B
9 13
11
12
810
D Q
CLK
Q
SR
R51M
+C36
100UF25V
730MA
FC
U13PT78NR112
1
3
2VI
CO
M
VO
FC
+ C48100UF 25V
+C35100UF 25V
730MA
D4
MBR350
VCC
GND
U19HCPL-2601-020
87
6
2
35
+C8100UF
25V
+C12100UF
25V
C7
.1UF MP
R32
16.2K
D2
1N4148
R35
10K
R10.01 OHM
3W 10%
C27
.1UF MP
RV15K TOP
13.7V ADJ
U8LT1431CN8
1234
8765
COLLECTORCOMP
V+R_TOP
REFR_MIDGND-FGND-S
+ C29100UF 25V
R171.2K .1%
J101
173001BATTERY #1
123
R33
3.74K
R46
10K
R474.99K
R34
2.49K
R6
100K
R1
10K
U9LM2940T-5.0
H.S.
1
2
3VI
AD
J
VO
+ C168UFTANT
D1
1N4148
+
C13100UF25V 730MA
D15
1N4937
R91
30.1K
U22CNY17-3
21
45
D11
1N4937
+ C794.7UFTANT
D51N4937
R1212.7K
U14LM340LAZ15
3
2
1 VI
GN
D
VO
R31K
U16LM320LZ-15
2
1
3 VI
GN
D
VO
T11640-00-10
8
6
9
10
7
4
2
D3
MBR350 J105POWER IN
12345
R2710K
R25
1K
+C28100UF
25V
U7CNY17-3
21
45
+C16100UF25V730MA
Q4
2N3903
13
2
+ C17100UF25V
R236.04K
RN4
1K7 8
C201000PF
NPO
L2
25UH2.5A
R2610K
2ASB2AG
F2
FC
RN4
1K3 4
R21330 OHM
1/2W 10%
FC
D71N6263
RN4
1K12
R2910K
RN4
1K910
+ C266.8UFTANT
C23
470PF NPO
R48475 OHM
RN2
1K1 2
C2
.1UFCER
R284.99K
RN2
1K34
+C1547UF
35V
C3
.1UFCER
RN2
1K5 6
Q5
2N3903
13
2
TP13VCC
RN2
1K7 8
Q6IRF522 1
32
C78.1UF CER
TP2BI
RN2
1K9 10
U3AICM7556
6
3
5
1
2
414
7
TR
CV
Q
DIS
THR
RV+
V-
Q102N6400
2
3
1
VCC
GND
U20HCPL-2601-020
87
6
2
35
U2
OP07CP3
261
8
74
+
-U1B
TL0627
5
6
H.S.
U3BICM7556
8
11
9
13
12
10
TR
CV
Q
DIS
THR
R
D10
1N4148
+C4947UF 35V
TP4VB
+
-U1A
TL062
1
84
3
2
R24100 OHM
D9
1N751A
+ C216.8UF TANT
R3011.5K
C221000PF NPO
TP316us
R3110K
D8
1N4148
+C196.8UF
TANT
TP5B++
R22
100K
D6
1N5819
C81.1UF
CER
8
8
7
7
6
6
5
5
4
4
3
3
2
2
1
1
D
C
B
A
HI VOLT KEEP ISOLATED
JP1JP2
POWER
POWERLED
AUDIO_A
AUDIO_B
REVISIONSDATEBYDESCRIPTIONREV
01 REVISED FOR REV-01 FAB.
3.6864MHZ
JP3
02 REVISED FOR REV-02 FAB.ECN 761ECN 762
GPGP
12-22-9712-22-97
NOTES:
03 REVISED FOR REV-03 FAB. ECN 779 GP 5-26-9804 GP 5-26-98REVISED PER ECN 780.
GP 5-26-9805 REVISED PER ECN 781.06 REVISED PER ECN 807. GP 7-31-9807 REVISED PER ECN 808. GP 7-31-9808 REVISED PER ECN 809. GP 8-3-9809 REVISED PER ECN 810. GP 8-3-9810 REVISED FOR REV-04 FAB.
+
-
ECN 845 GP 8-19-98GP11 REVISED PER ECN 875. 2-3-99GP12 REVISED PER ECN 887. 2-16-99
13 REVISED PER ECN 933. GP 4-9-9914 REVISED PER ECN 945. GP 5-20-9915 REVISED PER ECN 968. GP 8-6-9916 REVISED PER ECN 969. GP 8-6-9917 REVISED PER ECN 979. GP 9-16-99
GPGP
1819
REVISED PER ECN 1001.REVISED PER ECN 1002.
6-5-006-5-00
ALL RESISTORS ARE 1/4W, 1% U.O.S.1.
3.
2.L3,Q18,R92,RN5,RV1,T1,TP26,U36,Y1.
LAST USED: C81,D16,E1,F2,J110,JP4,
NOT USED: C25,C32,C60,R52,R72,R73.
2120 REVISED PER ECN 1088. GP
GP5-11-015-11-01REVISED FOR REV-05 FAB. ECN 1089.
22 REVISED PER ECN 1115. GP 8-6-01
5121-00-01 (-03 SCHEMATIC) 22
PULSE OXIMETER, MODEL 2000
IVY BIOMEDICAL SYSTEMS INC.
11 BUSINESS PARK DRIVEBRANFORD, CT. 06405(203) 481-4183
D
2 2Tuesday, March 08, 2005
Title
Size Document Number Rev
Date: Sheet of
LCDVEEWRTOT RDOTCSDISP A1
RESETD8 D9D10 D11D12 D13D14 D15
B++
VCC
DGNDCSCONT
WRTOT
D15D14D13D12D11D10D9D8
-12V
ACON
RXCOM RXDATPTXCOM TXDATP
RESET RTSCOM RTSPHALT RDOT CTSCOM
A0 BERRDSACK0 BRDSACK1 BGACK
DSACK1 CSUARTDSACK0 A1 D0 D15
A1 D0 A2 D1 D14A2 D1 A3 D2 D13A3 D2 A4 D3 D12 REDOTA4 D3 A5 D4 D11 A16 AS TXDOTA5 D4 A6 D5 A3 UBWE WRTOT CTSOTA6 D5 A7 D6 D10 A2 RDOTA7 D6 A8 D7 D9 A1 LBWEA8 D7 A9 CSVOL
CS0 A9 A10 CS1 D8 CSUART SIZ0CS1 A10 CS0 A11 RDOT WRTOT R/W
R/W CS2 A11 RDOT A12 LBWE CSDISP CS2ACON A12 A13 CSCONT A17
SIZ0 CS3 A13 A14 A0A14 A15
AS A15BGACK A16
DONE2 CSUARTBR DREQ2BERR DONE2 DONE1HALT DREQ1 RTSOTRESET DREQ2 RDOT RXPRNT
DONE1 CSDISP TXPRNTWRTOT RTSPRNT
DREQ1 CTSPRNT
CTSOTRTSOTTXDOTREDOTEPROM1ALRTL
A0 D8 A1 D8A1 D9 A1 D8 A2 D9A2 D10 A2 D9 A3 D10A3 D11 A3 D10 A4 D11A4 D12 A4 D11 A5 D12A5 D13 A5 D12 A6 D13A6 D14 A6 D13 A7 D14 D8A7 D15 A7 D14 A8 D15 D9A8 A8 D15 A9 D10A9 A9 A10 CS1 D11A10 A10 CS0 A11 RDOT D12
A11 RDOT A12 UBWE D13A12 A13 D14
RDOT A13 A14 D15WRTOT A14 A15 A1
A15A16 WRTOT
CSVOL
KESFT1KESFT2KESFT3KESFT4KEMENUKEALRO
CSCONT
ALRT-LIT
DATA [0..15]
ADDR [1..23]
KE
SF
T1
KE
SF
T2
KE
SF
T3
KE
SF
T4
KE
ME
NU
A17
KE
ALR
O
D15
A16
D14
A15
D13
A14
D12
A13
A12
A11
D11
D10
D9
A10
D8
A9
D7
A8
D6
A7
D5
A6
D4
A5
A4
D3
D2
A3
D1
A2
D0
A1
RX
PR
NT
TX
PR
NT
RT
SP
RN
TC
TS
PR
NT
CS
3
KE
SF
T1
KE
SF
T2
KE
SF
T3
KE
SF
T4
KE
ME
NU
KE
ALR
O
ALR
TL
CSCONT
CSVOL
A15
EP
RO
M1
CTSP
VCC
DGND
AGND
B+
-12V
/BATLO
TXMASI
RXMASI
CHARGLED
ON/STBY
UNIT_ON
PGND
BLINVB++
EPRM1
MASIRS
D
D
D
A
D
D
A
AA
A
A
D
DD
D D
D
D
D
DD
D
D
D
D
D
DD
D
A
A
A
D
DD
D
D
D
D
D
D
D
A
D
D D
A
D
D
A
A
A
D
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
-12V
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
B+
B+
VCC
B++
B++
B++
VCC
-12V
-12V
-12V
VCC
-12V
VCC
B++
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
VCC
C541UF MP
C71.1UF CER
TP20RD
R66
10.7K
C75.1UF CER
TP21CSV
R63
4.99K
TP22CSU
C64.1UF CER
+ C3333UF TANT
C72.1UF CER
U28SCC2691AC1A28
151617181920212223242526
1234567891011121314
2728
GNDINTRNCEND7D6D5D4D3NCD2D1NC
VCCRDNRXDTXDMP0MP1
NCNCA2A1A0
X1/CLKX2
RESETD0WRN
C69
.1UF CER
C76.1UF
CER
C70
.1UFCER
R68
9.31K50PPM
R83
9.31K50PPM
U34MAX690CPA
4
6
2
1
5
7
3
8
PFI
WDI
VC
C VO
PFO
RSGN
D
VB
Q162N3903
13
2
Q152N3906
13
2
J109
172001ALRT-LITE
12
R864.75K
R71
4.75K
U24MAX238
123456789
101112 13
1415161718192021222324T2OUT
T1OUTR2INR2OUTT1INR1OUTR1INGNDVCCC1+V+C1- C2+
C2-V-
R4INR4OUT
T2INT3IN
T4OUTT4IN
R3OUTR3IN
T3OUT
C551UF MP
C561UF MP
C68470PFNPO
C511UF
MP
U272357-00-04
(330099, AM28F512)
12111098765
272623254
282932
1314151718192021
2224
13130
16 32
A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14A15NC
O0O1O2O3O4O5O6O7
CEOE
VPPPGM
NC
GND VCC
R87
100 OHM
R7730.1K
U26MC68340FE16
123456789
101112131415161718192021222324252627282930313233343536
37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72
737475767778798081828384858687888990919293949596979899100101102103104105106107108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
RMCR/WSIZ1SIZ0DSASBGACKBGBRBERRHALTRESETGNDCLKOUTVCCXFCVCCEXTALVCCSYNXTALGNDMODCKVCCIPIPEIFETCHBKPTFREEZETIN1TOUT1TGATE1TCKTMSTDITDOVCCGND
FC
3F
C2
FC
1F
C0
VC
CG
ND
A23
A22
A21
A20
A19
A18
A17
VC
CG
ND
A16
A15
A14
GN
DA
13A
12A
11V
CC
GN
DA
10A
9A
8A
7A
6A
5A
4V
CC
GN
DA
3A
2A
1
TGATE2TOUT2
TIN2RXDATXDAVCCGND
RTSACTSA
RXRDYATXRDYA
RXDBTXDBRTSBCTSBSCLK
X2VCCGND
X1DREQ1DACK1DONE1DREQ2DACK2DONE2
IRQ7IRQ6IRQ5VCCGND
CS3/IRQ4IRQ3
CS2/IRQ2CS1/IRQ1
CS0/AVEC
D0
D1
D2
D3
VC
CG
NDD4
D5
D6
D7
D8
D9
D10
D11
VC
CG
ND
D12
GN
DD
13D
14D
15A
24A
25/IA
CK
1A
26/IA
CK
2A
27/IA
CK
3V
CC
GN
DA
28/IA
CK
4A
29/IA
CK
5A
30/IA
CK
6A
31/IA
CK
7A
0D
SA
CK
0D
SA
CK
1V
CC
GN
D
U312356-00-04
(330099, AM28F512)
12111098765
272623254
282932
1314151718192021
2224
13130
16 32
A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14A15NC
O0O1O2O3O4O5O6O7
CEOE
VPPPGM
NC
GND VCC
R88
100 OHM
R7847.5 OHM
R76
301K
JP4
172001
SPEAKER
R79
499 OHM
JP2
SPKR+ORN
+
-U29
LF3513
2
1 5
6
74
JP1
U30MK48T02-12
87654321
232219
182021
910111314151617
2412
A0A1A2A3A4A5A6A7A8A9A10
CEOEWE
D0D1D2D3D4D5D6D7
VC
CG
ND
SPKR-GRN
C61.1UF
CER
R5310K
JP3DO NOT INSTALL JP3DO NOT INSTALL JP3DO NOT INSTALL JP3
C62.1UF
CER
Q142N3903
13
2
R54
10K
R8047.5 OHM
R69
14.3K
R6728.7K
+ C6722UF TANT
R75
80.6K
R84
11KC59
.0047UFMP
R74
150K
R8228.7K
C66.001UF MP
R62
8.25K
+C73
22UF TANT
R5810K
-
+U32C
TL064
10
98
+
-
U32D
TL064
12
1314
C74.001UF
MP
TP12MRXDAT
TP15MTXDAT
RN11K
21
345678910
TP14TSTOT
SOCKET32 PIN
TP23CSD
SOCKET32 PIN
TP24CSC
R8510K
TP19WR
TP18ALARM
TP1AGND
Q172N3906
13
2
SOCKET28 PIN
R7010K
TP25/RESET
R811K
TP16DGND
U33AD7528JN
14131211109876
1615
5
17
234
201918
1
DB0DB1DB2DB3DB4DB5DB6DB7DA/DB
WRCS
DGND
VDD
OUT ARFB A
VREF A
OUT BRFB B
VREF B
AGND
Y17.3728MHZ
5
48
OUT
GN
D+
5VD
C
+
-U17B
TL062
75
6U17A
TL062
1
84
3
2
-
+U32B
TL064
5
67
-
+U32ATL064 3
21
411
J107174010
LCDRIBBON
1234
R89100 OHM
R361K
MPIE1
TP17PULSE
U25MT5C2568-15
109876543
252421232
261
1112131516171819
202227
14 28
A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14
D0D1D2D3D4D5D6D7
CEOEWE
DGND VCC
U23MT5C2568-15
109876543
252421232
261
1112131516171819
202227
14 28
A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14
D0D1D2D3D4D5D6D7
CEOEWE
DGND VCC
J102
179015RS232
123456789
J103
179015PRINTER
123456789
C47.1UF
CER
J106LCD RIBBON
135791113151719
2468
101214161820
C77.01UF MP
C45
.001UF MP
R57
10K
U352364-00-04
123456789
101112 13
1415161718192021222324GND
A16UBWERDOTLBWECSVOLCSUARTCST02CSDISCSCONTA15GND GND
A0A17CS2R/WSIZ0
7.4MHZ3.6MHZ1.8MHZWRTOT
ASVCC
R5510K
R56
10K
Q112N3906
13
2R65
1K
C46.001UF MP
Q122N3903
13
2
U36CXA-L10A
1
2
3
4
5
VIN
GND
OUT1
OUT2
OUTGND
C57.1UF CER
C58.1UF CER
U18AD7524 14
16
1
2
15
3
12
13
4
65
789
1011
VD
D RFB
OUT1
OUT2
VREF
GN
D
CS
WR
DB7
DB5DB6
DB4DB3DB2DB1DB0
R64
150 OHM
C65.1UF
CER
D131N6263
C53.1UF CER
R43200K
C501UFMP
C52.1UF CER
RN310K
2 13456789
10
R42100K
RN510K
2 13456789
10
J108
FRONT PANEL
1 2 3 4 5 6 7 8 9 10 11 12 13 14
C63.1UF CER
BOARD LAYOUT DIAGRAMS
B Model 2000 Service Manual
This page is intentionally left blank.