46
Model 2000 Pulse Oximeter Operator’s / Service Manual Ivy Biomedical Systems, Inc. 11 Business Park Drive Branford, Connecticut 06405 (203) 481-4183 (800) 247-4614 FAX (203) 481-8734 web:www.ivybiomedical.com e-mail:[email protected]

IVY 2000 Service Manual

Embed Size (px)

Citation preview

Page 1: IVY 2000 Service Manual

Model 2000 Pulse Oximeter

Operator’s / Service Manual

Ivy Biomedical Systems, Inc.

11 Business Park Drive Branford, Connecticut 06405

(203) 481-4183 • (800) 247-4614 • FAX (203) 481-8734 web:www.ivybiomedical.com e-mail:[email protected]

Page 2: IVY 2000 Service Manual

The following are trademarks of Masimo Corporation: Masimo® Masimo SET®, and LNOP®. Covered by one or more of the following U.S. Patents and foreign Equivalents: 5,337,744, 5,452,717, 5,482,036, 5,490,505. Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

Copyright © 1998 by Ivy Biomedical Systems, Inc., Branford, Connecticut. All rights reserved. No part of this manual may be reproduced without the permission of Ivy Biomedical Systems, Inc.

20001199 2411-00-20 REV 05

Page 3: IVY 2000 Service Manual

TABLE OF CONTENTS

Model 2000 Service Manual i

TABLE OF CONTENTS WARRANTY ..............................................................................................................................................................iii Description of Warning Labels .................................................................................................................................iv INTRODUCTION .......................................................................................................................................................1 SAFETY .......................................................................................................................................................................2 Electrical.......................................................................................................................................................................2 Explosion ......................................................................................................................................................................2 Patient Connections.....................................................................................................................................................2 Measurements ..............................................................................................................................................................3 EMC .........................................................................................................................................................................3 MONITOR DESCRIPTION ......................................................................................................................................4 Specifications................................................................................................................................................................4 Classification ................................................................................................................................................................5 Controls and Indicators ..............................................................................................................................................6 Fixed Keys ......................................................................................................................................................6 On/Standby......................................................................................................................................................6 Programmable Keys ........................................................................................................................................6 Display .........................................................................................................................................................................7 Waveform Display ..........................................................................................................................................7 Alphanumeric Data .........................................................................................................................................7 Alarm Limits Display......................................................................................................................................7 Alarm Displays................................................................................................................................................7 Trend Displays ................................................................................................................................................7 Alarms .........................................................................................................................................................................8 Rear Panel ....................................................................................................................................................................8 Menu Structure............................................................................................................................................................9 MONITOR SETUP ...................................................................................................................................................10 Set Time and Date .....................................................................................................................................................10 Audio Setup................................................................................................................................................................10 Alarm Tone ................................................................................................................................................................11 Trend Display ............................................................................................................................................................11

Page 4: IVY 2000 Service Manual

TABLE OF CONTENTS

ii Model 2000 Service Manual

PULSE OXIMETRY MONITORING ....................................................................................................................12 Principles of Operation .............................................................................................................................................12 Warnings/Cautions....................................................................................................................................................14 Pulse Oximetry Sensors ............................................................................................................................................14 Monitoring Procedure...............................................................................................................................................14 Alarm Limits ..............................................................................................................................................................15 Response Settings.......................................................................................................................................................15 Default Values ............................................................................................................................................................15 Reset to Default Values .............................................................................................................................................15 ALARM MESSAGES ...............................................................................................................................................16 Status Messages .........................................................................................................................................................16 MONITOR TESTING ..............................................................................................................................................17 BATTERY OPERATION.........................................................................................................................................18 Battery Replacement .................................................................................................................................................18 CLEANING AND DISINFECTION........................................................................................................................19 The Monitor ...............................................................................................................................................................19 Cleaning and Reuse of Sensors.................................................................................................................................19 Patient Cables ............................................................................................................................................................20 ACCESSORIES .........................................................................................................................................................21 Reorder Numbers ......................................................................................................................................................21 Instructions for use....................................................................................................................................................22 Disposal.......................................................................................................................................................................22 TESTING & TROUBLESHOOTING .....................................................................................................................27 Required Equipment .................................................................................................................................................27 Power Supply/Charges Tests ....................................................................................................................................27 Audio Tests.................................................................................................................................................................28 Safety Tests ................................................................................................................................................................28 Rear Panel Connectors..............................................................................................................................................29 ERROR MESSAGES ................................................................................................................................................31 REPLACEMENT PARTS ........................................................................................................................................32 DISASSEMBLY & ASSEMBLY .............................................................................................................................33 SCHEMATICS ...........................................................................................................................................................A BOARD LAYOUTS ................................................................................................................................................... B

Page 5: IVY 2000 Service Manual

WARRANTY

Model 2000 Service Manual iii

WARRANTY All products manufactured by Ivy Biomedical Systems, Inc. are warranted to be free from defects in material and workmanship and to operate within published specifications, under normal use, for a period of one year from date of original shipment. LNOP® sensors purchased from Ivy Biomedical are warranted for 6 months. If an examination, by Ivy Biomedical Systems, Inc., discloses such products or component parts to have been defective, then our obligation is limited to repair or replacement (at our option). Fuses and batteries are not covered under this warranty.

Page 6: IVY 2000 Service Manual

DESCRIPTION OF LABELS

iv Model 2000 Service Manual

Description of Labels Used Attention, consult ACCOMPANYING DOCUMENTS before attempting to change

power supply selection or carry out interconnections. Equipment connected should comply with IEC-601-1 or IEC-950 with configuration to IEC-601-1-1.

⎯♥⎯ Type CF equipment, Defibrillator proof. _______ Fuse type/rating.

= Power OFF, = Power ON.

∼ Alternate Current (AC)

ON – only for a part of equipment

STBY – off, only for a part of equipment

Contrast

Protective earth (ground) adjacent to this symbol. Internal connection on mains connector/filter.

Equipotential earth connector adjacent to this symbol.

Page 7: IVY 2000 Service Manual

INTRODUCTION

Model 2000 Service Manual 1

INTRODUCTION This manual is to provide information on the correct use of the Model 2000 Pulse Oximeter. It is up to the user to ensure that any applicable regulations respecting the installation and operation of the monitor are observed. The Model 2000 Pulse Oximeter is to be operated by qualified personal only. Using This Manual We recommend that you read this entire manual before operating the equipment. Use the Monitor Description section for general descriptions of controls and displays. For details on pulse oximetry, refer to the section of the manual dealing with that parameter. NOTE: If the monitor beeps when you press a key, it indicates an invalid selection for the current mode or function. Boldface type is used in text to refer to the labeling on user controls. Special brackets [ ] surround menu selections used with the programmable keys. Manufacturer’s Responsibility The manufacturer of this equipment is responsible for the effects on safety, reliability, and performance of the equipment only if: • assembly operations, extensions, re-adjustments, or repairs are carried out by persons authorized by the

manufacturer • the electrical installation complies with all applicable regulations • the equipment is used in accordance with the instructions in this manual Incorrect operation or failure of the user to maintain the monitor in accordance with proper maintenance procedures relieves the manufacturer or his agent from all responsibility for consequent non-compliance, damage, or injury. For technical and service information, please refer to the Ivy Model 2000 Operator's/Service Manual or contact: Ivy Biomedical Systems, Inc. 11 Business Park Drive Branford, Connecticut 06405 (203) 481-4183 or (800) 247-4614 e-mail:[email protected] This manual explains how to set up and use the Model 2000 Pulse Oximeter. Important safety information relating to general use of the pulse oximeter appears in this manual. Other important safety information is located throughout the manual where appropriate. READ THE ENTIRE SAFETY INFORMATION SECTION BEFORE YOU OPERATE THE PULSE OXIMETER.

Page 8: IVY 2000 Service Manual

SAFETY

2 Model 2000 Service Manual

SAFETY

Electrical Medical Electrical Equipment. Classified by Underwriters Laboratories Inc. ® with respect to electric shock, fire, mechanical and other specified hazards only in accordance with UL2601-1 CAN/CSA C22.2 No. 601.1. This product is intended to be operated from a mains power source of nominally 100 V ~ to 230 V ~, 47-63 Hz, or from its internal battery. WARNING: For continued protection against fire hazard, replace only with same type and rating of fuse. Fuse value is T 1A 250V 5 mm x 20 mm. WARNING: To prevent electrical hazards to all personnel, this monitor must be properly grounded. The power cable supplied with this equipment provides for this protection. Do not attempt to defeat this protection by modifying the cable or by using ungrounded adapters. Do not connect to an electrical outlet controlled by a wall switch or dimmer.

WARNING: Electric shock hazard! Do not remove covers or panels. Refer service to qualified service personnel.

WARNING: To avoid electrical shock, disconnect the monitor from its power source before changing fuses.

WARNING: Do not clean monitor while it is on and/or plugged in. WARNING: When attached to other devices, insure that the total chassis leakage currents of all units do not exceed 300µA.

Explosion DANGER: Explosion hazard! Do not use this equipment in the presence of flammable anesthetics.

Page 9: IVY 2000 Service Manual

SAFETY

Model 2000 Service Manual 3

Patient Connections Leakage current is limited internally by this monitor to less than 10 µA. However, always consider additional leakage current that can be caused by other equipment used on the patient at the same time as this monitor. Carefully route patient cables to reduce the possibility of patient entanglement or strangulation.

Measurements If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by alternate means and then check the Model 2000 Pulse Oximeter for proper functioning. Inaccurate measurements may be caused by:

• Incorrect sensor application or use • Significant levels of dysfunctional hemoglobins (e.g., carboxyhemoglobin or methemoglobin) – see note. • Intravascular dyes such as indocyanine green or methylene blue – see note. • Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon light source),

bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight (exposure to excessive illumination can be corrected by covering the sensor with a dark or opaque material)

• Excessive patient movement • Venous pulsations • Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line

The Model 2000 can be used during defibrillation, but the readings may be inaccurate for a short time. Loss of pulse signal can occur in any of the following situations:

• The sensor is too tight • There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp or sunlight • A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensor attached • The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia • There is arterial occlusion proximal to the sensor • The patient is in cardiac arrest or is in shock

NOTE: A pulse oximeter measures functional hemoglobin saturation. This is different to an arterial blood gas that measures fractional hemoglobin saturation. Differencies can occur when significant levels of HbCO (carboxyhemoglobin), HbMET (methomoglobin) indocynanine green or methylene blue exist. EMC This equipment has been certified to be protected to emissions and immunity according to IEC-601-1-2.

Page 10: IVY 2000 Service Manual

MONITOR DESCRIPTION

4 Model 2000 Service Manual

MONITOR DESCRIPTION The Model 2000 Pulse Oximeter is intended for continuous noninvasive monitoring of functional oxygen saturation or arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in a hospital and mobile environment within the hospital. Specifications Performance Range: Saturation (%SpO2) 1% - 100% Pulse Rate (bpm) 25 - 240 Accuracy: Saturation (%SpO2) – During no motion conditions Adults 70% - 100% ± 2 digits 0% - 69% unspecified Neonates 70% - 100% ± 3 digits Saturation (%SpO2) – During motion conditions Adults 70% - 100% ± 3 digits 0% - 69% unspecified Pulse (bpm) – During no motion conditions 25 to 240 ± 3 digits Pulse (bpm) – During motion conditions 25 to 240 ± 5 digits Resolution Saturation(%SpO2) 1% Pulse Rate (bpm) 1 Display Type: Backlit LCD Pixels: 240 x 64 dots Dot Pitch: 0.53 mm Data Displayed: Pulse rate, SpO2 %, Pleth wave, Alarms, Trends, Status messages

Page 11: IVY 2000 Service Manual

MONITOR DESCRIPTION

Model 2000 Service Manual 5

Power Requirements Voltage Input Range: 100-230 V~, 47-63 Hz Maximum AC Power Consumption: 55 VA Communications Serial Port RS232, 9600 baud Mechanical Size: 4.26in (10.65cm)H x 9.75in (24.4cm) W x 9.75in (24.4cm) D Weight: 6.5 lbs (2.95 kg) Battery Type: Sealed gel cell, DOT and IATA approved Operating Time: 2 hours Charge Time: <8 hours Classification (in accordance with IEC601-1) Protection against electric shock: Class 1 equipment Degree of protection against electric shock: Type CF equipment Degree of protection against harmful ingress of water: Drip-proof equipment Methods of sterilization or disinfection: See page 19 Degree of safety of application in the presence of a famable anaesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a fammable anaesthetic mixture Mode of operation: Continuous

Page 12: IVY 2000 Service Manual

MONITOR DESCRIPTION

6 Model 2000 Service Manual

Controls and Indicators Fixed Keys

Programmable keys Each of the keys below the display has a normal function, printed on the key, and several menu functions displayed on the screen in the setup mode. The normal functions are described below. Menu functions are described in the menu descriptions later in this manual.

Displays menus for all pulse oximeter settings. See the Menu Structure section of this manual for a description of the menus.

Turns the alarm audio on and off. Press once to silence the alarms for 120 seconds (two minutes). Press and hold for five seconds to silence the alarms until this button is pressed again.

On/Standby: Controls power to the monitor’s electronic circuits. The rear panel mains switch must be on for this key to be effective.

Resets the audible and visual indicators for an alarm that has been activated.

Displays Trend data. To define the trend display, see the trend display setup instructions.

Adjusts the contrast of the display. Press for small adjustments. Press and hold for larger adjustments

Initiates a printout from the optional printer.

Page 13: IVY 2000 Service Manual

MONITOR DESCRIPTION

Model 2000 Service Manual 7

Display Waveform Display The plethysmographic waveform is displayed from left to right. Alphanumeric Data The following data appears on the display: Pulse Rate- Lower left-hand corner of the display – values updated once per second. Oxygen Saturation (SpO2) - To the left of the waveform – values updated once per second. Alarm Limits Display SpO2 Alarm limit settings are displayed to the right of the SpO2 reading. Pulse rate alarm limit settings are displayed to the right of the pulse rate reading. Alarm Displays When an alarm setting has been exceeded, a flashing alarm tag appears on the screen, indicating which alarm has been triggered. Trend Displays Graphic display of up to 24 hours of averaged saturation. The pleth waveform is not displayed while the trend is displayed. Upper/Lower SpO2 Limits Red LED alert window

Upper/Lower Pulse Rate Limits

SpO2 %

BPM

98 70 LIMITS VOLUME DATE/TIME RESPONSE

99 90

150 50

Page 14: IVY 2000 Service Manual

MONITOR DESCRIPTION

8 Model 2000 Service Manual

Alarms The following is a list of alarms and ranges for which they can be set. When the alarm triggers. Audio alarm tone sounds if the ALARM PAUSE button has not been activated. Pressing ALARM RESET turns off any alarms that have triggered. If the measurement returns to within the limits, the alarm turns off. The ALARM PAUSE key disables the alarms for two minutes or until ALARM PAUSE is pressed again. High Pulse Rate: 50 to 235 bpm in 5 bpm increments Low Pulse Rate: 30 to 150 bpm in 5 bpm increments High SpO2: 70 to 99%in 1% increments Low SpO2: 50 to 95% in 1% increments Sensor Off: The sensor is not properly attached to the patient. Rear Panel The following are located on the rear panel. COM 1: A digital interface for network communication (see note). PRINTER: A connection for the optional printer (see note). MAINS (A.C.) POWER SWITCH: A switch to control mains power input to the monitor. PEQ GROUND: Potential Equalization- A ground connection that can be used to ensure that no potential differences can develop between this equipment and the other electrical equipment.

Note: When using COM1 or PRINTER rear panel connectors, the Ferrite Filter Adaptor Connector must be used. Ivy Biomedical accessory part number 150027 Note: Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards and essential performance standards, and/or the system configuration must meet the requirements of the IEC 601-1-1 medical electrical systems standard. ACCESSORIES (EQUIPMENT) – The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include: • use of the accessory in the PATIENT VICINITY; and • evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate

IEC 601-1 and/or IEC 601-1-1 harmonized national standard.

Page 15: IVY 2000 Service Manual

MONITOR DESCRIPTION

Model 2000 Service Manual 9

Menu Structure Settings for all options for the Model 2000 are made through a series of menus that appear at the bottom of the display. You access the menus by pressing the MENU key. When menus are displayed, press MENU again to return to normal display. NOTE: If the monitor beeps when you press a key, it indicates an invalid selection for the current mode or function. Press MENU to display the main menu, which allows you to select alarm limits, volume, date/time, and response menus. When the main menu is displayed, press MENU again to return to normal display.

Main Menu Soft Keys

SpO2%

BPM

98 70 LIMITS VOLUME DATE/TIME RESPONSE

99 90

150 50

Page 16: IVY 2000 Service Manual

MONITOR SETUP

10 Model 2000 Service Manual

MONITOR SETUP To setup the instrument for operation: 1. Plug the ac line cord into a power source providing the proper voltage and make sure the rear panel power

switch is in the ON position. 2. Press the ON/STBY push button at the left of the front panel. If the internal battery is completely dead wait

one minute before pressing the ON/STBY push button. 3. Connect the sensor and the pulse oximetry cable to the SpO2 connector on the front panel. Set Time and Date 1. Press the MENU key to display the main menu. Then select [DATE/TIME]. 2. The first setting is for MONTH. Use the and keys to increase or decrease the month setting. 3. Select [SET MONTH] to move to the DAY setting. Use the and keys to increase or decrease the day

setting. 4. Select [SET DAY] to move to the YEAR setting. Use the and keys to increase or decrease the year

setting. 5. Select [SET YEAR] to move to the HOUR setting. Use the and keys to increase or decrease the hour

setting. 6. Select [SET HOUR] to move to the MINUTE setting. Use the and keys to increase or decrease the

minute setting. 7. When all date and clock settings are correct, select [ENTER] to enter the settings into the monitor’s

memory. 8. Press MENU to return to normal display. Audio Setup 1. Press the MENU key to display the main menu. 2. Select [VOLUME] to display the menu for setting pulse and alarm volume. 3. Adjust the pulse volume using [PULSE and ] keys – OFF, 20%, 40%, 60%, 80% or 100%.

Page 17: IVY 2000 Service Manual

MONITOR SETUP

Model 2000 Service Manual 11

4. Adjust the alarm volume using [ALARM and ] keys – 20%, 40%, 60%, 80% or 100%. 5. Press MENU to return to the normal display. Alarm Tone To change the alarm tone: 1. Press the ON/STBY push button to put the monitor in Stand-by. 2. Press and hold the TREND key while simultaneously pressing ON/STBY push button. 3. Adjust the alarm tone using [ALM TONE and ] keys - LOW, MEDIUM or HIGH. 4. Press the MENU key to return to normal display. Trend Displays The Model 2000 has a trend display showing SpO2. Time selections for this trend display are 30min, 2, 8, 12 and 24 hours. Trend Setup 1. Press the TREND key to select 30min., 2, 8, 12 or 24 hour trend. 2. Press TREND to return to the normal display.

Page 18: IVY 2000 Service Manual

PULSE OXIMETRY MONITORING

12 Model 2000 Service Manual

PULSE OXIMETRY MONITORING Overview Pulse oximetry allows you to continuously and noninvasively monitor a patient’s hemoglobin oxygen saturation.The oximetry sensor contains two light emitting diodes (LEDs) that transmit specific wavelengths of light which are received by a photodetector. Oxygen saturated blood absorbs light differently as compared to unsaturated blood. Thus the amount of light absorbed by blood can be used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood. The monitor displays this ratio as percent SpO2. Normal values typically range from 95 to 100% at sea level. Principles of Operation The Model 2000 Pulse Oximeter is based on three principles: 1. Oxyhemoglobin and deoxyhemoglobin differ in their absorption of red and infrared light (spectrophotometry). 2. The volume of arterial blood in tissue and the light absorbed by the blood changes during the pulse (plethysmography). 3. Arterio-venous shunting is highly variable and that fluctuating absorbance by venous blood is a major component of noise during the pulse. The Model 2000 Pulse Oximeter as well as traditional pulse oximetry determines SpO2 by passing red and infrared light into a capillary bed and measuring changes in light absorption during the pulsatile cycle. Red and infrared light-emitting diodes (LEDs) in oximetry sensors serve as the light sources, a photodiode serves as the photodetector. Traditional pulse oximetry assumes that all pulsations in the light absorbance signal are caused by oscillations in the arterial blood volume. This assumes that the blood flow in the region of the sensor passes entirely through the capillary bed rather than through any arterio-venous shunts. The traditional pulse oximeter calculates the ratio of pulsatile absorbance (AC) to the mean absorbance (DC) at each of two wavelengths, 660 nm and 940 nm. S(660) = AC(660)/DC(660) S(940) = AC(940)/DC(940) The oximeter then calculates the ratio of these two arterial pulse-added absorbance signals: R = S(660)/S(940) The value of R is used to find the saturation SpO2 in a look-up table built into the oximeter’s software. The values in the look-up table are based upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in induced hypoxia studies.

Page 19: IVY 2000 Service Manual

PULSE OXIMETRY MONITORING

Model 2000 Service Manual 13

The Model 2000 Pulse Oximeter assumes that arterio-venous shunting is highly variable and that fluctuating absorbance by venous blood is the major component of noise during the pulse. The 2000 decomposes S(660) and S(940) into an arterial signal plus a noise component and calculates the ratio of the arterial signals without the noise: S(660) = S1 + N1 S(940) = S2 + N2 R = S1/S2 Again, R is the ratio of two arterial pulse-added absorbance signals and its value is used to find the saturation SpO2 in an empirically derived equation into the oximeter’s software. The values in the empirically derived equation are based upon human blood studies against a laboratory co-oximeter on healthy adult volunteers in induced hypoxia studies. The above equations are combined and a noise reference (N’) is determined: N' = S(660) - S(940) x R If there is no noise N' =0; then S(660) = S(940) x R which is the same relationship for the traditional pulse oximeter. The equation for the noise reference is based on the value of R, the value being seeked to determine the SpO2. The Model 2000 software sweeps through possible values of R that correspond to SpO2 values between 1% and 100% and generates an N' value for each of these R values. The S(660) and S(940) signals are processed with each possible N' noise reference through an adaptive correlation canceler (ACC) which yields an output power for each possible value of R (i.e., each possible SpO2 from 1% to 100%). The result is a Discrete Saturation Transform (DSTTM) plot of Relative output power versus possible SpO2 value as shown in the following figure, where R corresponds to SpO2 = 97%. The red light power range is 1mw to 3 mw. The infrared light power range is 0.7mw to 3mw.

The DST plot has two peaks: the peak corresponding to the higher saturation is selected as the SpO2 value. This entire sequence is repeated once every two seconds on the most recent four seconds of raw data. The Model 2000

Page 20: IVY 2000 Service Manual

PULSE OXIMETRY MONITORING

14 Model 2000 Service Manual

SpO2 therefore corresponds to a running average of arterial hemoglobin saturation that is updated every two seconds. WARNING: The pulse oximeter should NOT be used as an apnea monitor. WARNING: The pulse oximeter should be considered an early warning device. If a trend towards patient deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the patient’s condition. Caution: Use only sensors specified by Ivy Biomedical Systems, Inc. If you use sensors other than those specified, it might degrade perfomance and could damage the pulse oximeter. Caution: Check the sensor site frequently. Do not allow the sensor to remain on one site for a prolonged period of time, especially when monitoring neonates. Refer to the sensor’s specific instructions. Caution: Never attach a SpO2 sensor on a limb being monitored with a blood pressure cuff or a limb with restricted blood flow. Caution: A poorly applied sensor might give incorrect saturation values. Reapply the sensor. Caution: Choose a site with sufficient perfusion to ensure accurate oximetry values. Caution: Certain nail aberrations, nail polish, fungus, etc. might give inaccurate oximetry values. Remove the nail polish and/or move the sensor to an unaffected digit. Caution: Do not use the Model 2000 Pulse Oximeter or sensors during Magnetic Resonance Imaging (MRI). Pulse Oximetry Sensors To ensure conformance with all safety and performance specifications, use only the Masimo® LNOP pulse oximetry sensors. Monitoring Procedure Use the following procedure for monitoring pulse oximetry: 1. Choose a site that is well perfused and provides proper alignment of the LEDs and receiving photodetector Select a site that has unrestricted blood flow Do not restrict blood flow when securing sensor with tape Select an appropriate sensor and apply it to the patient, following the directions for use provided with the sensor Do not select a site near potential electrical interference (electrical cords, for example). 2. Connect the cable to the monitor’s SpO2 input. 3. Plug the extension cable into the sensor. 4. Use the procedures described in the following sections for alarm limit settings, response mode settings, and enabling or disabling the SpO2 alarm. Alarm Limits 1. Press the MENU key to display the main menu.

Page 21: IVY 2000 Service Manual

PULSE OXIMETRY MONITORING

Model 2000 Service Manual 15

2. Select [LIMITS]. 3. Press ALARM RESET to select SpO2 or PULSE PARAMETERS. 4. Use the [HIGH and ] selections to set the high limit. 5. Use the [LOW and ] selections to set the low limit. Each time you press a key, the corresponding limit changes by 1% for SpO2 and 5 BPM for Pulse. The

current settings are shown on the display. 6. Press MENU to return to the normal display. Response Mode Settings 1. Press MENU to display the main menu. Then select [RESPONSE]. 2. Use the [AVERAGING] selection to set the response time – 6, 8, 10, 12, 14 or 16 sec. 3. Use the [SENSITIVITY] selection to set the sensitivity level. (Options – Normal and High). The High

sensitivity level should be used when the clinician wants to have the absolute low profusion performance of the Masimo SET® and is willing to sacrifice some sensor off detection capability.

4. Press MENU to return to the normal display. Default Values The following defaults are set: 1. High SpO2 limit – set to OFF. 2. Low SpO2 limit – set to 85. 3. High pulse rate limit – set to 150 bpm. 4. Low pulse rate limit – set to 50 bpm. 5. Pulse volume – set to 40% 6. Alarm volume – set to 100% 7. Averaging – set to 8 sec. 8. Sensitivity – set to Normal. To reset the model 2000 to the above factory default values: 1. Press the ON/STBY push button to put monitor in stand by. 2. Press and hold ALARM RESET push button while simultaneously pressing ON/STBY push button. 3. Listen for the audio to sound 8 times in succession – defaults are now reset to factory default values.

Page 22: IVY 2000 Service Manual

ALARM MESSAGES

16 Model 2000 Service Manual

ALARM MESSAGES

Alarm messages indicate a problem or condition which may affect accurate monitoring values. Do not ignore these messages. Correct any fault before continuing. Whenever it is necessary to perform procedures on a patient that might cause activation of an alarm, press the ALARM PAUSE key to disable the alarms for five minutes. Alarm indications flash on the display, and the red front panel alarm indicator flashes. When the alarm triggers, the alarm indicator flashes and the audio alarm tone sounds. The displayed numeric value of the limit exceeded will be in reverse video and flashing. Pressing ALARM RESET turns off any alarms that have triggered. If the measurement returns to within the limits, the alarm turns off. These alarms will signal when: The high pulse rate limit has been exceeded for greater than three seconds. The low pulse rate limit has been exceeded for greater than three seconds. The low SpO2 limit has been exceeded for greater than three seconds. The high SpO2 limit has been exceeded for greater than three seconds.

STATUS MESSAGES The following is a list of module status messages used in the Model 2000. Message Meaning NO SENSOR There is no sensor plugged into end of cable. MODULE FAIL SpO2 module failure. SENSOR FAIL SpO2 sensor failure. CHECK CABLE Interconnect cable failure. SENSOR OFF SpO2 sensor not in contact with patient. SEARCHING Unit is searching for patient's pulse. INTERFERING Outside signal or energy preventing reading. HI AMBIENT Too much light on patient. L PERFUSION Signal too small. Notes: Above messages (all messages except L PERFUSION) cause the SpO2 and PR numerics to be replaced with dashes and are accompanied by an audible alert tone. In addition, a SYS FAULT message is displayed whenever communications with MS-1 module has been lost. This message flashes in reverse video and is always accompanied by an audible alert tone.

Page 23: IVY 2000 Service Manual

MONITOR TESTING

Model 2000 Service Manual 17

MONITOR TESTING Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local or governmental regulations. In the event that internal adjustment or recalibration is necessary, refer to the Service Manual for this equipment. Alarm Testing Place the sensor on your finger. After the readings have stabilized change the SpO2 and Pulse Rate alarm limits to cause an alarm. Verify that the SpO2 and Pulse Rate alarms function when their settings are violated. The audio should sound and the red alarm LED should flash along with the limit violated.

Page 24: IVY 2000 Service Manual

BATTERY OPERATION

18 Model 2000 Service Manual

BATTERY OPERATION The model 2000 contains a 2.3 amp-hr 12V lead acid battery. Under normal use and full charge the unit will operate for greater than 2 hours continuously. The unit will recharge in less than 8 hours. In the event of mains power failure, the model 2000 will automatically switch to battery operation. When the remaining battery charge is less than 20 minutes a BAT LOW indicator in the lower right hand corner of the display will flash and the audio will sound every 15 seconds. The battery is continuously charged whenever the AC power is connected and the mains power switch on the rear panel is on. To ensure maximum battery life, it is recommended that, once a month, the unit be run on battery until it turns itself off and then connect to mains power to recharge the battery. Battery Replacement 1. Remove the 6 screws on the rear of the unit

2. Slide cover off unit

3. Locate battery tray and remove 2 screws that hold down battery

4. Lift battery and un-clip connector from edge of battery

5. Install new battery, “CAUTION, replace only with same type 2.3 amp-hr 12 Volt battery”.

6. Reverse step 1 thru 4

7. Plug unit into line and charge for at least 8 hours

8. Unplug unit and operate for 2 hours or until instrument shuts off to confirm proper operation.

NOTE: Sealed lead battery must be recycled or disposed of properly. Do not incinerate

Page 25: IVY 2000 Service Manual

CLEANING AND DISINFECTION

Model 2000 Service Manual 19

CLEANING AND DISINFECTION The Monitor When necessary, clean the exterior surfaces of the monitor with a cloth or swab dampened with a warm water and mild detergent solution. Do not allow liquids to enter the interior of the instrument. Unplug monitor from mains power. Cleaning and Reuse of Sensors Reusable sensors can be cleaned as follows: 1. Remove the sensor from the patient. 2. Disconnect the sensor from the monitor. 3. Wipe the entire sensor with a 70% isopropyl alcohol pad. 4. Allow the sensor to air dry before returning it to operation. Reattachment of Single Use Sensors Single use sensors can be reapplied to the same patient if the emitter and detector windows are clear and the adhesive still adheres to the skin. The adhesive can be partially rejuvenated by wiping with an alcohol wipe and allowing the sensor to thoroughly air dry prior to replacement on the patient. NOTE: If the sensor fails to track the pulse consistently, the sensor might be incorrectly positioned. Reposition the sensor or choose a different monitoring site. Caution: Do not soak or immerse the sensor in any liquid solution. Do not sterilize any sensor by irradiation, steam, or ethylene oxide.

Page 26: IVY 2000 Service Manual

CLEANING AND DISINFECTION

20 Model 2000 Service Manual

Patient Cables Patient cables can be cleaned using the following procedure: 1. Remove the cable form the sensor. 2. Disconnect the cable from the monitor. 3. Wipe clean with a 70% isopropyl alcohol pad. 4. Allow the cable to dry before returning it to operation. Caution: Do not soak or immerse patient cables in any liquid solution. Do not sterilize any patient cable by irradiation, steam, or ethylene oxide.

Page 27: IVY 2000 Service Manual

ACCESSORIES

Model 2000 Service Manual 21

ACCESSORIES Reusable Sensors Adult LNOP DC-1 Finger Sensor, Reusable 590227 Disposable Sensors Neonatal LNOP-Neo, Disposable Sensor, Box of 20 590223 Pediatric LNOP-Pdt, Slender Digit Disposable Sensor, Box of 20 590222 Neonatal Pre-term LNOP-NeoPt, Disposable Sensor, Box of 20 590224 Adult LNOP-Adt, Disposable Sensor, Box of 20 590221 Patient Cables Patient Cable, 8 ft PC08 with LNOP Sensor Connector 590220 Patient Cable 12 ft PC12 with LNOP Sensor Connector 590230 Starter Kits Starter Kit with 2 LNOP-Adt and 2 LNOP-Pdt sensors, and 12’ PC12 sensor cable 590225 Starter Kit with 2 LNOP-Neo and 2 LNOP-NeoPt sensors, and 8’ PC08 sensor cable 590226 Starter Kit with Adult LNOP-DC1 finger sensor, reusableand PC08 sensor cable 590228 Miscellaneous Ferrite Filter Connector Adaptor 150027 Disposal Disposal of devices or consumables must be done in accordance to natural environmental laws.

Page 28: IVY 2000 Service Manual

22

IVY LNOP® SENSORS WITH MASIMO SET® DISPOSABLE NON-STERILE SENSORS

Indications: • Continuous SpO2 monitoring • Single patient use • Use only with IVY monitors with Masimo

SET

Patient Size: Weight Sensor

Type Accuracy* Ivy Biomedical

Part Number Over 30 kg Adt +/- 2 digits 590221 10 to 50 kg Ped +/- 2 digits 590222 Under 10kg Neo +/- 3 digits 590223 Under 1 kg Neo Pt +/- 3 digits 590224 *From 70 to 100% SpO2 to +/- 1 Std.Dev.

Contraindications: • Patients allergic to adhesive tape

Adults Over 30kg – Adt Sensor Site Selection Attachment to Patient P/N: 590221 • Well perfused • Restricts movement the

least • Non dominate hand

preferred • Ring or middle finger or • Great or long toe

Step A • Remove sensor from

pouch • Hold tan side down,

remove backing • Press detector onto

fleshy tip of finger

Step B • Wrap “T” shaped

adhesive around finger

Step C • Fold emitter end

over nail

Step D • Wrap wings around

finger • Ensure emitter and

detector face each other

• Reposition if necessary

Step E • Position sensor with shiny

contacts up • Position cable with Masimo

SET logo up • Slide sensor contacts into

cable, hear click • Tug to confirm connection • Tape cable to patient if

needed

Page 29: IVY 2000 Service Manual

23

Pediatrics 10 to 50kg – Ped Sensor P/N:590222 Site Selection Attachment to Patient • Well perfused • Restricts movement the least • Non dominate hand

preferred • Ring or middle finger

Step A • Remove sensor

from pouch • Hold tan side down,

remove backing • Press detector onto

fleshy tip of finger

Step B • Wrap “T” shaped

adhesive around finger

Step C • Fold emitter end

over nail

Step D • Wrap wings around

finger • Ensure emitter and

detector face each other

• Reposition if necessary

Step E • Position sensor with shiny

contacts up • Position cable with Masimo

SET logo up • Slide sensor contacts into

cable, hear click • Tug to confirm connection • Tape cable to patient if

needed

Neonates under 10 kg or with fat feet - Neo Sensor P/N:590223 • Neonates: foot or palm • Infants: big toe • Well perfused

Step A • Remove sensor

from pouch • Hold along the “Y”,

remove backing

Step B • Locate detector • Position on fleshy

sole of foot • Align with fourth

toe

Step C • Position emitter on

top of foot • Wrap bandage

around foot • Use bandage to

secure emitter and detector

Step D • Ensure emitter and

detector face each other

• Ensure detector window is fully covered

• Reposition if necessary • Use big toe on infants

over 3 to 5 kg, or with fat feet.

Step E • Position sensor with shiny

contacts up • Position cable with Masimo

SET logo up • Slide sensor contacts into

cable, hear click • Tug to confirm connection • Tape cable to patient if

needed

Page 30: IVY 2000 Service Manual

24

Neonates Under 1 kg – Neo Pt Sensor P/N:590224 Site Selection Attachment to Patient • Neonates: foot or

palm • Could be to

bulky on big toe • Well perfused

Step A • Remove sensor

from pouch • Hold along the “Y”,

remove backing

Step B • Locate detector • Position on fleshy

sole of foot • Align with middle toe

Step C • Position emitter on

top of foot • Wrap sponge wrap

around foot • Use wrap to secure

emitter and detector

Step D • Ensure emitter and

detector face each other

• Ensure detector window is fully covered

• Reposition if necessary

Step E • Position sensor with shiny contacts

up • Position cable with Masimo SET

logo up • Slide sensor contacts into cable, hear

click • Tug to confirm connection • Tape cable to patient if needed

Sensor Reattachment • Sensors can be reapplied to the same patient if:

− Emitter and detector windows are clear − Adhesive still adheres

• Adhesive may be rejuvenated: − Thoroughly clean with alcohol wipe − Let sensor air dry before reattachment

Disconnecting Sensors • Squeeze gray buttons on both sides of connector • Pull out sensor tab from connector

Warnings • Check site every 8 hours for:

− Proper adhesion to patient − Alignment of emitter and detector − Reposition if required

• With poorly perfused patients: − Check site every 2 hours − Check for skin damage, pressure necrosis,

distal circulation − Reposition if required

• Do NOT: − Use damaged sensors − Immerse or soak sensors − Use with MRI scanners − Place on same limb as Art line or BP cuff

• Sources of incorrect readings

− Intravascular dyes − Elevated Carboxyhemoglobin (COHb) − Elevated Methemoglobin (MetHb) − Mis-aligned emitter and detector (must face

each other vertically) − Incompletely covered detector window

• High O2 levels may predispose premature infants to retinopathy − Set the high SpO2 alarm limit as required

Page 31: IVY 2000 Service Manual

25

ADULT REUSABLE NON-STERILE SENSOR P/N:590227 Indications: • Adults over 30 kg • Spot ckeck or Continuous SpO2 monitoring • Low motion, clean environments • Use only with IVY monitors with Masimo SET

Accuracy ±2 digits, from 70 to 100% SpO2 to ±1 Std.Dev.

Contraindications: • High motion • Dirty environments • Long term monitoring over 4 hours

Adults Over 30kg – DC1 Sensor Site Selection Attachment to Patient • Well perfused • Restricts movement the least • Non dominate hand preferred • Ring or middle finger or • Great or long toe

Step A • Place digit over sensor window • Fleshy tip of finger should cover window

Step B • Tip of finger should touch stop • Long nail can extend over stop

Step C • Rear hinge may open • Verify positioning • Verify detector

window is fully covered

Step D • Orient cable to exit

along top of hand

Step E • Position sensor with shiny

contacts up • Position cable with Masimo

SET logo up • Slide sensor contacts into cable,

hear click • Tug to confirm connection • Tape cable to patient if needed

Page 32: IVY 2000 Service Manual

26

Disconnecting Sensor • Squeeze gray buttons on both sides of connector • Pull out sensor tab from connector

Cleaning • Remove sensor from patient • Disconnect sensor form cable

• Wipe clean with 70% Isopropyl Alcohol pad • Allow to dry • Do NOT soak, immerse or sterilize sensor Warnings • Check site every 4 hours for:

− Skin damage, pressure necrosis, distal circulation − Change to disposable sensor if required

• With poorly perfused patients, check site every 2 hours

• Do NOT: − Use damaged sensors − Immerse or soak sensors − Use with MRI scanners − Place on same limb as Art line or BP cuff

• Sources of incorrect readings − Intravascular dyes − Elevated Carboxyhemoglobin (COHb) − Elevated Methemoglobin (MetHb) − Mis-aligned emitter and detector (must

face each other vertically) − Incompletely covered detector window

INSTRUMENT CABLES Directions for Use • Connect LNOP sensors to monitor • For use only with IVY monitors with Masimo SET • PC 08 is 8’ (2.4m) long –P/N:590220 • PC 12 is 12’ (3.7m) long –P/N:590230 • See above for attaching sensors

Cleaning • Wipe clean with 70% Isopropyl Alcohol pad • Allow to dry • Do NOT soak, immerse or sterilize cable

Warnings • Do NOT:

− Use cable if it is damaged − Connect to another patient cable − Connect to a monitor without Masimo SET

• Ensure cable is properly connected • Hold only by connectors when attaching

sensors

Page 33: IVY 2000 Service Manual

TESTING & TROUBLESHOOTING

Model 2000 Service Manual 27

TESTING & TROUBLESHOOTING Required Equipment Hipot Tester Hypotronics Model HDA5 Variable AC Supply Powerstat Model 3PN216C or equivalent Oscilloscope Tektronix Model 2236 or equivalent DMM Fluke Model 8050A or equivalent Power Supply BK Precision Model 1630 or equivalent Masimo Sensor 590227 Masimo Patient Cable 590220 Bio Tek Index 2 SpO2 Simulator Capacitor/Resistor Network 0.15µf@100V, 5% metalized polyester 1kΩ 1/4W, 1% 10kΩ 1/4W, 1% Ground Isolator Plug Power Supply/Charger Tests 1. Apply power with the MAINS power switch and check for the front panel GREEN (AC ON) LED to light. 2. Monitor the +13.7V supply at VB (TP4) with the scope. Adjust RV1 for +13.80 to +13.85V with fully

charged batteries (<.1mV across R10 DMM from F1 back to AGND). 3. Turn the front panel power ON. The voltages should decrease but must be ≥+13.60V. Verify that the front

panel GREEN (ONSTBY) LED lights. 4. Turn the MAINS power OFF. Operate the monitor on battery for approximately 1 minute. Connect the

DMM across R10 (from F1 back to AGND). Turn the MAINS power back ON. Check for the YELLOW (CHARGING LED) to light. The voltage across R10 should be 4.36mV to 5.78mV.

5. While the CHARGING LED is ON, monitor the +12V supply at P105. 6. Turn the front panel power OFF. The voltage at VB (TP4) must be +14.7V to +15.0V. 7. Check the YELLOW (CHARGING LED) to turn off when the voltage at VB (TP4) switches to +13.7

(+13.60V to +13.85V). 8. Turn the front panel power on. Measure the following supplies with the scope referenced to DGND

(TP16): +5V (+4.85V to +5.15V) VCC TP13 B++ (VB≤100mV) TP5 -12V (-11.64V to –12.36V) U13 pin 2

9. Reference the scope to ISOLATED GROUND TP11 and measure the following supplies: +6.2V (+6.16V to +6.35V) TP6 +20V (+26.5V to +32.5V) TP8 +15VI (+14.7V to +15.3V) TP10 -15VI (-14.4V to –15.63V) TP9 +5VI (+4.85V to +5.15V) TP7

Page 34: IVY 2000 Service Manual

TESTING & TROUBLESHOOTING

28 Model 2000 Service Manual

10. Turn the rear panel MAINS power off. Disconnect the battery. Connect the power supply to J101. Connect the positive to J101 pin 1 and negative to J101 pin 2. Adjust the external DC supply for +12.0V. Apply power with the front panel power switch.

Connect the scope to VBF (F2 front). Slowly decrease the voltage on the supply until the LOW BATTERY message appears on the display. The voltage should be +10.98V to +11.47V. Continue to decrease the voltage on the supply until the monitor shuts off. The voltage should be +9.60V to +10.26 (just before shut off). Increase the voltage on the supply (DMM) to +12V. The monitor should not turn on.

Audio Tests 1. Set the pulse volume to 100%. Connect the scope to the speaker (+) or JP4. Place the sensor on your finger

or connect it to the simulator. Check for 2.5Vp-p±.3Vp-p. Decrease the pulse volume and make sure the amplitude is gradually reduced to zero.

2. Turn the alarm audio on and set it to 100%. The amplitude at the speaker should be 3.0Vp-p±.3Vp-p.

Decrease the alarm volume to minimum and make sure the amplitude is gradually reduced to .5Vp-p±50mVp-p.

Safety Testing

BE SURE THE FRONT PANEL SWITCH IS SET TO STBY FOR THE FOLLOWING TEST 1. Connect the SpO2 hipot plug to the front panel SpO2 input. Connect the RED lead from the hipot tested to

the hipot plug. Connect the Black lead to the rear PEQ ground.

Turn the tester on and slowly increase the voltage to 4kV. There should be no arcing or leakage for one minute. Reduce the voltage to zero.

2. Replace the hipot plug with the SpO2 leakage plug. Using the DMM, select AC Volts and 200mV range.

Connect the leakage filter as shown below. CAUTION: Be careful to connect the circuit below only between the measurement points indicated in the following instructions. Do not connect this circuit across line voltage. Full line voltage would damage the measurement circuit and could possibly damage the monitor and/or the digital voltmeter Measuring instrument R1=10kΩ±5% R2= 1kΩ±1% C1=0.01µF ±5% With the monitor on, measure the leakage from the leakage plug to the HOT side of the line. The leakage must be <10µA at 120 V AC.

R2 R1 C1 V

Page 35: IVY 2000 Service Manual

TESTING & TROUBLESHOOTING

Model 2000 Service Manual 29

Connect the measurement circuit from the rear panel PEQ ground of the monitor to line cord ground. Measure the leakage with the following conditions: Unit Grounded Ground Broken Ground Broken, reversed polarity The leakage must be <30µA. Rear Panel Connectors COM 1: A digital interface for network communication (see note).

Pin Signal Output

1 No Connection

2 RECEIVE DATA RS-232 ±9 V (±5 V min.)

3 TRANSMIT DATA RS-232 ±9 V (±5 V min.)

4 No Connection

5 SIGNAL GROUND Reference for COM 1 signals

6 No Connection

7 REQUEST TO SEND Not Used

8 CLEAR TO SEND Not Used

9 No Connection

When in the Normal Operating Mode, the Model 2000 external RS232 COM port is configured for 9600 baud bi-directional operation with 8 data bits, no parity, 1 stop bit, and no hardware or software handshaking. The output data consists of ASCII text strings sent at one second intervals, containing the date, time, SpO2 value and pulse rate, terminated by ASCII carriage return (hex 0D) and line feed (hex 0A) characters, in the following format:

mm/dd/yy hh:mm:ss SPO2=xxx% PULSE=yyy<cr><lf> PRINTER: A connection for the optional printer (see note).

Pin Signal Output

1 No Connection

2 RECEIVE DATA Not used

3 TRANSMIT DATA RS-232 ±9 V (±5 V min.)

4 No Connection

5 SIGNAL GROUND Reference for Printer signals

6 No Connection

7 REQUEST TO SEND Not Used

8 CLEAR TO SEND RS-232 ±9 V (±5 V min.)

9 No Connection

Mating connector, use Ivy P/N 179004, 9-pin receptacle for 22-26 AWG wire. Use RFI/EMI shielded cable clamp, Ivy P/N 210030, and 5-conductor shielded cable, Ivy P/N 600002.

Mating connector, use Ivy P/N 179004, 9-pin receptacle for 22-26 AWG wire. Use RFI/EMI shielded cable clamp, Ivy P/N 210030, and 5-conductor shielded cable, Ivy P/N 600002.

Page 36: IVY 2000 Service Manual

TESTING & TROUBLESHOOTING

30 Model 2000 Service Manual

MAINS POWER SWITCH: A switch to control mains power input to the monitor. PEQ GROUND: Potential Equalization- A ground connection that can be used to ensure that no potential differences can develop between this equipment and the other electrical equipment.

Note: When using COM1 or PRINTER rear panel connectors, the Ferrite Filter Adaptor Connector must be used. Ivy Biomedical accessory part number 150027 Note: Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards and essential performance standards, and/or the system configuration must meet the requirements of the IEC 601-1-1 medical electrical systems standard. ACCESSORIES (EQUIPMENT) – The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include: • use of the accessory in the PATIENT VICINITY; and • evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate

IEC 601-1 and/or IEC 601-1-1 harmonized national standard.

Page 37: IVY 2000 Service Manual

MONITOR SETUP

Model 2000 Service Manual 31

ERROR MESSAGES Several fault conditions which cause messages and an audible alert tone to occur which are not associated with the MS-1 module, as follows: Message Meaning Caused by ERROR 1 ROM Error U31 (P/N:2356-00-04) and U27 (P/N:2357-00-04) ERROR 2 RAM Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 3 Clock Chip Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 4 Undefined Interrupt Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 5 Timing Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 6 Communications Port Error Pulse Oximeter Board (P/N:5121-00-01) ERROR 9 Unknown Error Pulse Oximeter Board (P/N:5121-00-01)

Page 38: IVY 2000 Service Manual

REPLACEMENT PARTS

32 Model 2000 Service Manual

REPLACEMENT PARTS Part Ivy Part Number LCD Display 570012 Display Cable Assembly 610011 DC-AC Converter 220007 Display Window 2354-00-13 Front Panel 2343-00-10 Power Supply Module 580017 Pulse Oximeter Module, MS-1 580014 LED Board Assembly 5131-00-01 LED Alarm Lens 2401-00-01 SpO2 Input Board Assembly 5122-00-01 Fuse, 2ASB, 2AG, 250V 290003 Fuse, 5ASB, 2AG, 125V 290015 Fuse, 1ASB, Metric, (5x20mm), 250V 290014 Speaker 470000 Battery, 12 Volt, 2.3 Ah 110006

Page 39: IVY 2000 Service Manual

MONITOR SETUP

Model 2000 Service Manual 33

DISASSEMBLY & ASSEMBLY

Covers & Circuit Boards To remove cover: 1. Unplug unit from mains and remove line cord. 2. Locate the six (6) screws on the rear panel and remove. 3. Slide cover off. To remove motherboard: 1. Disconnect all wire sockets after noting their locations. 2. Locate and remove the six (6) mounting screws from bottom of unit. 3. Remove motherboard from bottom of unit. To remove Masimo module: 1. Remove motherboard. 2. Disconnect ribbon cable and programming voltage connector. 3. Remove interconnect PC board. 4. Remove three (3) nylon screws. To remove power supply: 1. Remove the two (2) connectors from the power supply. 2. Locate and remove the two (2) #6 nuts that retain the two ground wires and power supply mounting

bracket. 3. Remove bracket from power supply. CAUTION: When replacing power supply, be sure to replace the two (2) green ground wires to their #6 ground studs. To remove front panel assembly: 1. Remove motherboard. 2. Unplug J106. 3. Locate and remove the four (4) screws (2 on each side) that hold the front panel to side rails. 4. Remove front panel assmbly by carefully disconnecting the interconnect cable.

Page 40: IVY 2000 Service Manual

SCHEMATIC DIAGRAMS

Model 2000 Service Manual A

SCHEMATIC DIAGRAMS

Page 41: IVY 2000 Service Manual

SCHEMATIC DIAGRAMS

A Model 2000 Service Manual

This page is intentionally left blank.

Page 42: IVY 2000 Service Manual

BOARD LAYOUT DIAGRAMS

Model 2000 Service Manual B

BOARD LAYOUT DIAGRAMS

Page 43: IVY 2000 Service Manual
Page 44: IVY 2000 Service Manual

8

8

7

7

6

6

5

5

4

4

3

3

2

2

1

1

D D

C C

B B

A A

ISOLATED NON-ISOLATED

RECEIVE_DATATRANSMIT_DATA

RESET_INPUT-

|LINK

MASIMO

D

S

G

BAT+

BAT-

|C:\ORCAD\SHEET\SH5121\5121-002.SCH

AGND, DGND & PGNDARE ONE COMMON GNDAT J105-3,4,5

REVISIONSDATEBYDESCRIPTIONREV

01 REVISED FOR REV-01 FAB.

ALL RESISTORS ARE 1/4W, 1% U.O.S.1.

3.

NOTES:

2.L3,Q18,R92,RN5,RV1,T1,TP26,U36,Y1.

LAST USED: C81,D16,E1,F2,J110,JP4,

D

S G

02 REVISED FOR REV-02 FAB.ECN 761ECN 762

GPGP

12-22-9712-22-97

03 REVISED FOR REV-03 FAB.

|C:\ORCAD\SHEET\SH5121\5121-003.SCH

ECN 779 GP 5-26-9804 GP 5-26-98REVISED PER ECN 780.05 GP 5-26-98REVISED PER ECN 781.06 REVISED PER ECN 807. GP 7-31-9807 REVISED PER ECN 808. GP 7-31-9808 REVISED PER ECN 809. GP 8-3-9809 REVISED PER ECN 810. GP 8-3-9810 REVISED FOR REV-04 FAB. ECN 845 GP 8-19-98

GP11 REVISED PER ECN 875. 2-3-99GP12 REVISED PER ECN 887. 2-16-99

13 REVISED PER ECN 933. GP 4-9-99

NOT USED: C25,C32,C60,R52,R72,R73.

14 REVISED PER ECN 945. GP 5-20-9915 REVISED PER ECN 968. GP 8-6-9916 REVISED PER ECN 969. GP 8-6-9917 REVISED PER ECN 979. GP 9-16-99

GPGP

1819

REVISED PER ECN 1001.REVISED PER ECN 1002.

6-5-006-5-00

2120 REVISED PER ECN 1088. GP

GP5-11-015-11-01REVISED FOR REV-05 FAB. ECN 1089.

22 REVISED PER ECN 1115. GP 8-6-01

5121-00-01 (-03 SCHEMATIC) 22

PULSE OXIMETER, MODEL 2000

IVY BIOMEDICAL SYSTEMS INC.

11 BUSINESS PARK DRIVEBRANFORD, CT. 06405(203) 481-4183

D

1 2Tuesday, March 08, 2005

Title

Size Document Number Rev

Date: Sheet of

CHARGLED

/BATLO

B+

ON/STBY

TXMASI

RXMASI

AGND

DGND

VCC

UNIT_ON

PGND

B++

BLINV

-12V

MASIRS

EPRM1

I

I

I

D

I

I

I

I

I

A

A

A

A

A

A

AA

A

AA

A

A

A A

A

A

A

A

A

A A

A

A

A

A

D

A

D

A

A

A

A

A

I

D

AAA

I

I

I

A

A

A

+5VI

VCC

VCC

B+

-12A

B+

VB

B++

VCC

B+

B++

VR

VBFVBB+

VBF

VBF

VBF

VBF

VBF

VCC

+5VI

+5VI

VCC

B+

-12A

-12A

+5VI

+6.2V+6.2V

VR

VR

B+

B+

VCC

B+

VBF

+15VI

-15VI

+5VI

+15VI

-15VI

+5VI

VCC

B++

B++

B++

-12V

VCC

Q1

2N3903

13

2

U6PT5101N

1

2

3VI

GN

D

VO

C24

.47UF MP

C38.015UF1KV CD

D12

1N4148

L3

25UH

+

C9100UF25V

TP7+5VI

C111UFMP

+C31

1000UF16V

C6

1UF MP

TP11IGND +C39

100UF50V

+ C37100UF25V

R13100 OHM

Q32N5460

3

21

+C43100UF

50V

TP10+15VI

C181UF MP

Q2

PN43933

12

R37200K

+

C106.8UF

TANT

TP8+20V

RN4

1K5 6

TP9-15VI

Q7RFP30P05S D

G1

3 2

R15499K

R11

200K

D16

1N4937

R8499K

U21CNY17-3

21

45

U4ICL7662

1

2

4

3

6

7

5

8

TEST

CAP+

CAP-

GN

D

LV

OSC

VOUT

V+

R5110K

+C3047UF

35V

R14

10K

Q8

2N3903

13

2

R49

100K

R20

1K

Q13

2N3906

13

2

R90

60.4K

+C406.8UF

TANT

R50

100K

R38

200K

C801UF MP

+

-U11A

LM393

3

21

84

+C42

6.8UFTANT

C44

.1UF CER

+

C141500UF

25V

+

-U11B

LM393

5

67

J110EPRMV

123

J1041639-02-10

SaO2 DIGITALINTERFACE

123456789

1011121314

TP26-12V

C34

.1UFCER

R18

10K

+ C416.8UFTANT

Q92N3906

13

2

U15LM78L12ACZ

VI

GN

D

VO

L1

220UH52

R4510K

R61200K

5ASB2AG

F1

TP6+6.2V

R41200K

D14

1N4148

R44

10K

U5LT1070

3 1 2

45

GN

D

VC FB

VSWVIN

R249.9K

R40

604K

R4

200K

R19

2W 5%150 OHM

R60

698K

R7100K

C4

.1UFCER

U10AD680

3

1

2VIN

GND

VOUT

C5

.1UFCER

R92220 OHM

2W 5%

Q18IRFD120

D

SG23

1

R39200K

R594.99K

R9

1K

R16

1K

U12ACD4013B

5 1

3

2

14 647

D Q

CLK

Q

VD S

RVS

U12BCD4013B

9 13

11

12

810

D Q

CLK

Q

SR

R51M

+C36

100UF25V

730MA

FC

U13PT78NR112

1

3

2VI

CO

M

VO

FC

+ C48100UF 25V

+C35100UF 25V

730MA

D4

MBR350

VCC

GND

U19HCPL-2601-020

87

6

2

35

+C8100UF

25V

+C12100UF

25V

C7

.1UF MP

R32

16.2K

D2

1N4148

R35

10K

R10.01 OHM

3W 10%

C27

.1UF MP

RV15K TOP

13.7V ADJ

U8LT1431CN8

1234

8765

COLLECTORCOMP

V+R_TOP

REFR_MIDGND-FGND-S

+ C29100UF 25V

R171.2K .1%

J101

173001BATTERY #1

123

R33

3.74K

R46

10K

R474.99K

R34

2.49K

R6

100K

R1

10K

U9LM2940T-5.0

H.S.

1

2

3VI

AD

J

VO

+ C168UFTANT

D1

1N4148

+

C13100UF25V 730MA

D15

1N4937

R91

30.1K

U22CNY17-3

21

45

D11

1N4937

+ C794.7UFTANT

D51N4937

R1212.7K

U14LM340LAZ15

3

2

1 VI

GN

D

VO

R31K

U16LM320LZ-15

2

1

3 VI

GN

D

VO

T11640-00-10

8

6

9

10

7

4

2

D3

MBR350 J105POWER IN

12345

R2710K

R25

1K

+C28100UF

25V

U7CNY17-3

21

45

+C16100UF25V730MA

Q4

2N3903

13

2

+ C17100UF25V

R236.04K

RN4

1K7 8

C201000PF

NPO

L2

25UH2.5A

R2610K

2ASB2AG

F2

FC

RN4

1K3 4

R21330 OHM

1/2W 10%

FC

D71N6263

RN4

1K12

R2910K

RN4

1K910

+ C266.8UFTANT

C23

470PF NPO

R48475 OHM

RN2

1K1 2

C2

.1UFCER

R284.99K

RN2

1K34

+C1547UF

35V

C3

.1UFCER

RN2

1K5 6

Q5

2N3903

13

2

TP13VCC

RN2

1K7 8

Q6IRF522 1

32

C78.1UF CER

TP2BI

RN2

1K9 10

U3AICM7556

6

3

5

1

2

414

7

TR

CV

Q

DIS

THR

RV+

V-

Q102N6400

2

3

1

VCC

GND

U20HCPL-2601-020

87

6

2

35

U2

OP07CP3

261

8

74

+

-U1B

TL0627

5

6

H.S.

U3BICM7556

8

11

9

13

12

10

TR

CV

Q

DIS

THR

R

D10

1N4148

+C4947UF 35V

TP4VB

+

-U1A

TL062

1

84

3

2

R24100 OHM

D9

1N751A

+ C216.8UF TANT

R3011.5K

C221000PF NPO

TP316us

R3110K

D8

1N4148

+C196.8UF

TANT

TP5B++

R22

100K

D6

1N5819

C81.1UF

CER

Page 45: IVY 2000 Service Manual

8

8

7

7

6

6

5

5

4

4

3

3

2

2

1

1

D

C

B

A

HI VOLT KEEP ISOLATED

JP1JP2

POWER

POWERLED

AUDIO_A

AUDIO_B

REVISIONSDATEBYDESCRIPTIONREV

01 REVISED FOR REV-01 FAB.

3.6864MHZ

JP3

02 REVISED FOR REV-02 FAB.ECN 761ECN 762

GPGP

12-22-9712-22-97

NOTES:

03 REVISED FOR REV-03 FAB. ECN 779 GP 5-26-9804 GP 5-26-98REVISED PER ECN 780.

GP 5-26-9805 REVISED PER ECN 781.06 REVISED PER ECN 807. GP 7-31-9807 REVISED PER ECN 808. GP 7-31-9808 REVISED PER ECN 809. GP 8-3-9809 REVISED PER ECN 810. GP 8-3-9810 REVISED FOR REV-04 FAB.

+

-

ECN 845 GP 8-19-98GP11 REVISED PER ECN 875. 2-3-99GP12 REVISED PER ECN 887. 2-16-99

13 REVISED PER ECN 933. GP 4-9-9914 REVISED PER ECN 945. GP 5-20-9915 REVISED PER ECN 968. GP 8-6-9916 REVISED PER ECN 969. GP 8-6-9917 REVISED PER ECN 979. GP 9-16-99

GPGP

1819

REVISED PER ECN 1001.REVISED PER ECN 1002.

6-5-006-5-00

ALL RESISTORS ARE 1/4W, 1% U.O.S.1.

3.

2.L3,Q18,R92,RN5,RV1,T1,TP26,U36,Y1.

LAST USED: C81,D16,E1,F2,J110,JP4,

NOT USED: C25,C32,C60,R52,R72,R73.

2120 REVISED PER ECN 1088. GP

GP5-11-015-11-01REVISED FOR REV-05 FAB. ECN 1089.

22 REVISED PER ECN 1115. GP 8-6-01

5121-00-01 (-03 SCHEMATIC) 22

PULSE OXIMETER, MODEL 2000

IVY BIOMEDICAL SYSTEMS INC.

11 BUSINESS PARK DRIVEBRANFORD, CT. 06405(203) 481-4183

D

2 2Tuesday, March 08, 2005

Title

Size Document Number Rev

Date: Sheet of

LCDVEEWRTOT RDOTCSDISP A1

RESETD8 D9D10 D11D12 D13D14 D15

B++

VCC

DGNDCSCONT

WRTOT

D15D14D13D12D11D10D9D8

-12V

ACON

RXCOM RXDATPTXCOM TXDATP

RESET RTSCOM RTSPHALT RDOT CTSCOM

A0 BERRDSACK0 BRDSACK1 BGACK

DSACK1 CSUARTDSACK0 A1 D0 D15

A1 D0 A2 D1 D14A2 D1 A3 D2 D13A3 D2 A4 D3 D12 REDOTA4 D3 A5 D4 D11 A16 AS TXDOTA5 D4 A6 D5 A3 UBWE WRTOT CTSOTA6 D5 A7 D6 D10 A2 RDOTA7 D6 A8 D7 D9 A1 LBWEA8 D7 A9 CSVOL

CS0 A9 A10 CS1 D8 CSUART SIZ0CS1 A10 CS0 A11 RDOT WRTOT R/W

R/W CS2 A11 RDOT A12 LBWE CSDISP CS2ACON A12 A13 CSCONT A17

SIZ0 CS3 A13 A14 A0A14 A15

AS A15BGACK A16

DONE2 CSUARTBR DREQ2BERR DONE2 DONE1HALT DREQ1 RTSOTRESET DREQ2 RDOT RXPRNT

DONE1 CSDISP TXPRNTWRTOT RTSPRNT

DREQ1 CTSPRNT

CTSOTRTSOTTXDOTREDOTEPROM1ALRTL

A0 D8 A1 D8A1 D9 A1 D8 A2 D9A2 D10 A2 D9 A3 D10A3 D11 A3 D10 A4 D11A4 D12 A4 D11 A5 D12A5 D13 A5 D12 A6 D13A6 D14 A6 D13 A7 D14 D8A7 D15 A7 D14 A8 D15 D9A8 A8 D15 A9 D10A9 A9 A10 CS1 D11A10 A10 CS0 A11 RDOT D12

A11 RDOT A12 UBWE D13A12 A13 D14

RDOT A13 A14 D15WRTOT A14 A15 A1

A15A16 WRTOT

CSVOL

KESFT1KESFT2KESFT3KESFT4KEMENUKEALRO

CSCONT

ALRT-LIT

DATA [0..15]

ADDR [1..23]

KE

SF

T1

KE

SF

T2

KE

SF

T3

KE

SF

T4

KE

ME

NU

A17

KE

ALR

O

D15

A16

D14

A15

D13

A14

D12

A13

A12

A11

D11

D10

D9

A10

D8

A9

D7

A8

D6

A7

D5

A6

D4

A5

A4

D3

D2

A3

D1

A2

D0

A1

RX

PR

NT

TX

PR

NT

RT

SP

RN

TC

TS

PR

NT

CS

3

KE

SF

T1

KE

SF

T2

KE

SF

T3

KE

SF

T4

KE

ME

NU

KE

ALR

O

ALR

TL

CSCONT

CSVOL

A15

EP

RO

M1

CTSP

VCC

DGND

AGND

B+

-12V

/BATLO

TXMASI

RXMASI

CHARGLED

ON/STBY

UNIT_ON

PGND

BLINVB++

EPRM1

MASIRS

D

D

D

A

D

D

A

AA

A

A

D

DD

D D

D

D

D

DD

D

D

D

D

D

DD

D

A

A

A

D

DD

D

D

D

D

D

D

D

A

D

D D

A

D

D

A

A

A

D

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

-12V

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

B+

B+

VCC

B++

B++

B++

VCC

-12V

-12V

-12V

VCC

-12V

VCC

B++

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

VCC

C541UF MP

C71.1UF CER

TP20RD

R66

10.7K

C75.1UF CER

TP21CSV

R63

4.99K

TP22CSU

C64.1UF CER

+ C3333UF TANT

C72.1UF CER

U28SCC2691AC1A28

151617181920212223242526

1234567891011121314

2728

GNDINTRNCEND7D6D5D4D3NCD2D1NC

VCCRDNRXDTXDMP0MP1

NCNCA2A1A0

X1/CLKX2

RESETD0WRN

C69

.1UF CER

C76.1UF

CER

C70

.1UFCER

R68

9.31K50PPM

R83

9.31K50PPM

U34MAX690CPA

4

6

2

1

5

7

3

8

PFI

WDI

VC

C VO

PFO

RSGN

D

VB

Q162N3903

13

2

Q152N3906

13

2

J109

172001ALRT-LITE

12

R864.75K

R71

4.75K

U24MAX238

123456789

101112 13

1415161718192021222324T2OUT

T1OUTR2INR2OUTT1INR1OUTR1INGNDVCCC1+V+C1- C2+

C2-V-

R4INR4OUT

T2INT3IN

T4OUTT4IN

R3OUTR3IN

T3OUT

C551UF MP

C561UF MP

C68470PFNPO

C511UF

MP

U272357-00-04

(330099, AM28F512)

12111098765

272623254

282932

1314151718192021

2224

13130

16 32

A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14A15NC

O0O1O2O3O4O5O6O7

CEOE

VPPPGM

NC

GND VCC

R87

100 OHM

R7730.1K

U26MC68340FE16

123456789

101112131415161718192021222324252627282930313233343536

37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72

737475767778798081828384858687888990919293949596979899100101102103104105106107108

109

110

111

112

113

114

115

116

117

118

119

120

121

122

123

124

125

126

127

128

129

130

131

132

133

134

135

136

137

138

139

140

141

142

143

144

RMCR/WSIZ1SIZ0DSASBGACKBGBRBERRHALTRESETGNDCLKOUTVCCXFCVCCEXTALVCCSYNXTALGNDMODCKVCCIPIPEIFETCHBKPTFREEZETIN1TOUT1TGATE1TCKTMSTDITDOVCCGND

FC

3F

C2

FC

1F

C0

VC

CG

ND

A23

A22

A21

A20

A19

A18

A17

VC

CG

ND

A16

A15

A14

GN

DA

13A

12A

11V

CC

GN

DA

10A

9A

8A

7A

6A

5A

4V

CC

GN

DA

3A

2A

1

TGATE2TOUT2

TIN2RXDATXDAVCCGND

RTSACTSA

RXRDYATXRDYA

RXDBTXDBRTSBCTSBSCLK

X2VCCGND

X1DREQ1DACK1DONE1DREQ2DACK2DONE2

IRQ7IRQ6IRQ5VCCGND

CS3/IRQ4IRQ3

CS2/IRQ2CS1/IRQ1

CS0/AVEC

D0

D1

D2

D3

VC

CG

NDD4

D5

D6

D7

D8

D9

D10

D11

VC

CG

ND

D12

GN

DD

13D

14D

15A

24A

25/IA

CK

1A

26/IA

CK

2A

27/IA

CK

3V

CC

GN

DA

28/IA

CK

4A

29/IA

CK

5A

30/IA

CK

6A

31/IA

CK

7A

0D

SA

CK

0D

SA

CK

1V

CC

GN

D

U312356-00-04

(330099, AM28F512)

12111098765

272623254

282932

1314151718192021

2224

13130

16 32

A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14A15NC

O0O1O2O3O4O5O6O7

CEOE

VPPPGM

NC

GND VCC

R88

100 OHM

R7847.5 OHM

R76

301K

JP4

172001

SPEAKER

R79

499 OHM

JP2

SPKR+ORN

+

-U29

LF3513

2

1 5

6

74

JP1

U30MK48T02-12

87654321

232219

182021

910111314151617

2412

A0A1A2A3A4A5A6A7A8A9A10

CEOEWE

D0D1D2D3D4D5D6D7

VC

CG

ND

SPKR-GRN

C61.1UF

CER

R5310K

JP3DO NOT INSTALL JP3DO NOT INSTALL JP3DO NOT INSTALL JP3

C62.1UF

CER

Q142N3903

13

2

R54

10K

R8047.5 OHM

R69

14.3K

R6728.7K

+ C6722UF TANT

R75

80.6K

R84

11KC59

.0047UFMP

R74

150K

R8228.7K

C66.001UF MP

R62

8.25K

+C73

22UF TANT

R5810K

-

+U32C

TL064

10

98

+

-

U32D

TL064

12

1314

C74.001UF

MP

TP12MRXDAT

TP15MTXDAT

RN11K

21

345678910

TP14TSTOT

SOCKET32 PIN

TP23CSD

SOCKET32 PIN

TP24CSC

R8510K

TP19WR

TP18ALARM

TP1AGND

Q172N3906

13

2

SOCKET28 PIN

R7010K

TP25/RESET

R811K

TP16DGND

U33AD7528JN

14131211109876

1615

5

17

234

201918

1

DB0DB1DB2DB3DB4DB5DB6DB7DA/DB

WRCS

DGND

VDD

OUT ARFB A

VREF A

OUT BRFB B

VREF B

AGND

Y17.3728MHZ

5

48

OUT

GN

D+

5VD

C

+

-U17B

TL062

75

6U17A

TL062

1

84

3

2

-

+U32B

TL064

5

67

-

+U32ATL064 3

21

411

J107174010

LCDRIBBON

1234

R89100 OHM

R361K

MPIE1

TP17PULSE

U25MT5C2568-15

109876543

252421232

261

1112131516171819

202227

14 28

A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14

D0D1D2D3D4D5D6D7

CEOEWE

DGND VCC

U23MT5C2568-15

109876543

252421232

261

1112131516171819

202227

14 28

A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14

D0D1D2D3D4D5D6D7

CEOEWE

DGND VCC

J102

179015RS232

123456789

J103

179015PRINTER

123456789

C47.1UF

CER

J106LCD RIBBON

135791113151719

2468

101214161820

C77.01UF MP

C45

.001UF MP

R57

10K

U352364-00-04

123456789

101112 13

1415161718192021222324GND

A16UBWERDOTLBWECSVOLCSUARTCST02CSDISCSCONTA15GND GND

A0A17CS2R/WSIZ0

7.4MHZ3.6MHZ1.8MHZWRTOT

ASVCC

R5510K

R56

10K

Q112N3906

13

2R65

1K

C46.001UF MP

Q122N3903

13

2

U36CXA-L10A

1

2

3

4

5

VIN

GND

OUT1

OUT2

OUTGND

C57.1UF CER

C58.1UF CER

U18AD7524 14

16

1

2

15

3

12

13

4

65

789

1011

VD

D RFB

OUT1

OUT2

VREF

GN

D

CS

WR

DB7

DB5DB6

DB4DB3DB2DB1DB0

R64

150 OHM

C65.1UF

CER

D131N6263

C53.1UF CER

R43200K

C501UFMP

C52.1UF CER

RN310K

2 13456789

10

R42100K

RN510K

2 13456789

10

J108

FRONT PANEL

1 2 3 4 5 6 7 8 9 10 11 12 13 14

C63.1UF CER

Page 46: IVY 2000 Service Manual

BOARD LAYOUT DIAGRAMS

B Model 2000 Service Manual

This page is intentionally left blank.