94
It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography 1 Mark Sinnott Alexander Ucci 6 August, 2020 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

It’s Peak Season for Great Peak Shape

Tips and tricks on troubleshooting in GC chromatography

1

Mark SinnottAlexander Ucci6 August, 2020

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 2: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 2

“Everything Was Just Fine.. and Then This Happened!” “How do I troubleshoot?”

Logic=

Something changed(slowly or suddenly)

=Something is different

Track your actions/log book:• Changed column, liner, septum, syringe, etc.• Injected samples, other method, etc.• Carried out maintenance, cut column, inlet flush,

etc.

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 3: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 3

Logical Troubleshooting

Common problems

Troubleshooting tips

Examples

Agenda

Troubleshooting starts with isolating the problem.• There are five basic areas from where problems can

arise:-Injector-Flow-Column-Detector-Electronics

Or…- A combination of these

Knowing what can and cannot cause the symptom is the key, and most importantly DON'T PANIC!

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 4: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 4

• Peak tailing – flow path or activity• Bonus peaks – in sample or back flash (carry-over)• Split peaks – injector problems, mixed solvent• No peaks – wasn’t introduced, wasn’t detected• Response changes – activity, injector discrimination, detector problem• Peak fronting – overload or solubility mismatch, injector problems• Shifting retention – leaks, column aging, contamination, or damage• Loss of resolution – separation decreasing, peak broadening• Baseline disturbances – column bleed, contamination, electronics• Noisy or spiking baseline – electronics or contaminated detector• Quantitation problems – activity, injector, or detector problems

Common Peak Shape Issues

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 5: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 5

Peak Tailing

Injector or column is active• Reversible adsorption of active compounds (-OH, -

NH, -SH)

Flow problem • Dead volume, obstruction, poor installation, or severe

column contamination

Miscellaneous - overloading of PLOT columns, co-elution, polarity mismatch between phase, solute or solvent, and some compounds always tail

*Tip = Inject a light hydrocarbon, should not tail unless flow path problem.

Compound ID*. Methane1. 5-Nonanone2. Decanal3. Propionic acid4. Ethylene glycol5. Heptadecane6. Aniline

7. Methyl dodecanoate8. 2-Chlorophenol9. 1-Undecanol10. Nonadecane11. 2-Ethylhexanoic acid12. Ethyl maltol

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 6: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 6

Brochure 5991-6709EN

Agilent Inert Flow Solution

Compound ID*. Methane1. 5-Nonanone2. Decanal3. Propionic acid4. Ethylene glycol5. Heptadecane6. Aniline

7. Methyl dodecanoate8. 2-Chlorophenol9. 1-Undecanol10. Nonadecane11. 2-Ethylhexanoic acid12. Ethyl maltol

Modified Agilent J&W DB-WAX UI mix on DB-WAX UI, 122-7032UI*Every column is tested individually

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 7: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Agilent Inert Flow Solution

Agilent Ultimetal Plus– TCD, FPD, NPD/FID jets

Agilent Ultimetal Capillary Flow TechnologyDevices, Ultimate union

Agilent J&W Ultra Inert GC column

Agilent Ultra Inert gold seal

Agilent Ultra Inert inlet liner

Agilent Ultimetal Plus ferrules

Agilent Ultimetal Plus inlet weldment, shell and transfer lines

5990-8532EN brochure

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 8: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 8

Bonus or Ghost Peaks

Contamination in injector, column or flow (carrier gas)

• Carry-over from a backflash or previous sample

• Bad tank of gas or traps have expired

• Septum bleed

Tip: Run a blank run…it should be blank!

Initial Injection

Repeat Injection

Meranzin

Isomeranzin

Osthole

5991-9078EN

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 9: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What Are These Repeating Peaks?

9

Septa contamination in wash vials or inlet liners can be diagnosed by looking for siloxane polymers in your total ion chromatogram. Each peak in the chromatogram corresponds to a cyclized (ring structure) siloxane molecule. These molecules fragment with very similar patterns.

Example spectrum:

Common ions for siloxane molecules:

73

147

207

281

355

Is it column bleed? No!

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 10: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Multiple Injections from the Same Vial: Siloxanes!

10

5x10

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

2.2

2.4

2.6

2.8

3

3.2

3.4

3.6

Acquisition Time (min)

13 14 15 16 17 18 19 20 21

Run 1Run 6

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 11: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Does Your Baseline Look Like This? Do You See Extra Peaks?

11

Page 12: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

12

If your target ions are buried beneath matrix peaks, it might be time to trim the column or do sample clean-up

The Matrix

500 ppb ethenoproxin black tea

10 ppb linalool in shampoo

…(or improve your sample cleanup)

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 13: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

13

Beef Liver – UntreatedBeef Liver – EMR-Lipid treated

The Importance of Sample Cleanup

For sample cleanup help, please contact us! [email protected]

50 samples with clean-up

50 samples without clean-up

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 14: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Split Peaks

Injector (poor sample introduction)• Injecting the sample twice (somehow?)• Mixed sample solvent (polarity difference)• Sample in syringe needle (manual inject)Injector (activity)• Breakdown (not really a split peak, two

peaks)• Sample degradation in injector

Volatility• High boilers dropping

out on cold spots• Transfer line

temperatures• Unions or fittings not

tracking column temperature

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 15: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 15

No Peaks

Detector (not on, or not operational)

Injector (not working)

Plugged syringe/plunger not moving• Wrong injector (or detector)

• Huge leak (older systems)

• No carrier gas flow

Not the column unless…• Broken column or no column

MISSING

Last Seen Yesterday

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 16: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 16

Peak ResponseAll change in size

Detector (response problem)• Settings or flows changed

• Electronics failing

• Split ratio set incorrectly

• Wrong purge activation time

• Septum purge flow too high

• Injector temperature too low*

Injector• Leaky syringe

*Tip: Ask is it all of them or some of them, if all then injector or detector

Expected

Unexpected

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 17: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 17

Injector or column is active/contaminated• Irreversible adsorption of active compounds (-OH, -NH, -SH)

Decomposition of sample• Temperature change – Discrimination

• Evaporation from sample

Peak ResponseSome Change in Size

6.6 7 7.4 7.8 8.2 8.6 9 9.4

50000

100000

150000

200000

250000

300000

350000

400000

min

Run 1Run 7

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 18: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Change in Response: Pyraclostrobin in Spinach on Run 1

4x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

2.4

2.5Cpd 33: Pyraclostrobin: +EI MRM [email protected] (164.0000 -> 132.1000) spinach_10ppbGCPest_woolrep1.D

17.711

Counts vs. Acquisition Time (min)

17.4 17.42 17.44 17.46 17.48 17.5 17.52 17.54 17.56 17.58 17.6 17.62 17.64 17.66 17.68 17.7 17.72 17.74 17.76 17.78 17.8 17.82 17.84 17.86 17.88 17.9 17.92 17.94 17.96

18 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 19: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

4x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

2.4

2.5Cpd 33: Pyraclostrobin: +EI MRM [email protected] (164.0000 -> 132.1000) spinach_10ppbGCPest_woolrep10.D

17.710

Counts vs. Acquisition Time (min)

17.4 17.42 17.44 17.46 17.48 17.5 17.52 17.54 17.56 17.58 17.6 17.62 17.64 17.66 17.68 17.7 17.72 17.74 17.76 17.78 17.8 17.82 17.84 17.86 17.88 17.9 17.92 17.94 17.96

19

Change in Response: Pyraclostrobin in Spinach on Run 65

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 20: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

4x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

2.4

2.5Cpd 33: Pyraclostrobin: +EI MRM [email protected] (164.0000 -> 132.1000) spinach_10ppbGCPest_woolrep10.D

17.710

Counts vs. Acquisition Time (min)

17.4 17.42 17.44 17.46 17.48 17.5 17.52 17.54 17.56 17.58 17.6 17.62 17.64 17.66 17.68 17.7 17.72 17.74 17.76 17.78 17.8 17.82 17.84 17.86 17.88 17.9 17.92 17.94 17.96

20

Change in Response: Pyraclostrobin in Spinach on Run 1 vs Run 65

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 21: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

4x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

2.4

2.5Cpd 33: Pyraclostrobin: +EI MRM [email protected] (164.0000 -> 132.1000) spinach_10ppbGCPest_wool2rep1.D

17.702

Counts vs. Acquisition Time (min)

17.4 17.42 17.44 17.46 17.48 17.5 17.52 17.54 17.56 17.58 17.6 17.62 17.64 17.66 17.68 17.7 17.72 17.74 17.76 17.78 17.8 17.82 17.84 17.86 17.88 17.9 17.92 17.94 17.96

21

Change in Response: Pyraclostrobin in Spinach with New Liner

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 22: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 22

Peak FrontingShark fin-shaped or just slight

Column (contaminated)• Overload (more pronounced with large

solute and phase polarity differences)

Injector• Compound very soluble in

injection solvent (need retention gap)

• Mixed sample solvent

Other• Co-elution

• Breakdown

6x10

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

2.2

2.4

2.6

2.8

3

3.2

3.4

3.6

3.8

4

4.2

4.4

4.6

4.8

5

5.2

5.4

5.6

+ TIC Scan SVOA_30m1uL_2Aug2.D

Counts vs. Acquisition Time (min)

14.7 14.72 14.74 14.76 14.78 14.8 14.82 14.84 14.86 14.88 14.9 14.92 14.94 14.96 14.98 15

1.5 μL1 μL0.5 μL

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 23: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 23

Retention Time Shift

3.25

4.75

4.00

5.50

Injector• Leak in the septum

• Change in injection solvent

• Large change in sample concentration

Flow• Change in gas velocity

Column• Contamination

• Damaged stationary phase

• Loss of stationary phase

• Change in temperature

11.40 11.45 11.50 11.55 11.60 11.65 11.70 11.75 11.80 11.85 11.90 11.95

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 24: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Thermal Stability and Retention Time Shifting on Standard WAX Column

Peak Compound1 Methanol2 Benzene3 Toluene4 Ethylbenzene5 p-Xylene6 m-Xylene7 o-Xylene

Application note 5991-9035EN

2.4 2.5 2.6 2.7 2.8 2.9 3 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5 5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 7 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 8 8.1 8.2 8.3 8.4

Initial BTEX Injection

25 hours at 280°C

50 hours at 280°C

1

2

3

4 5 6

7

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 25: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

2.8 2.9 3 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5 5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 7 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 8 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8

Initial Injection

25 hours at 280°C

50 hours at 280°C

75 hours at 280°C

100 hours at 280°C1

2

3

4 5 67

Application Note 5991-9035ENDB-HeavyWAXPeak Compound

1 Methanol2 Benzene3 Toluene4 Ethylbenzene5 p-Xylene6 m-Xylene7 o-Xylene

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 26: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 26

Loss of Resolution

Resolution is a function of separation and peak width

Separation

Peak width

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 27: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 27

Loss of Resolution - Separation Decrease (RT's change)

Column• Different column temperature

• Contamination (more phase?)

• Matrix components coeluting

Flow• Change in velocity?

Separation

Peak width

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 28: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 28

Loss of Resolution - Peak Broadening (RT's unchanged)

Flow • Make-up gas

Column• Contamination

• Phase degradation

Injector (efficiency)• Settings, liner, installation, etc.

Peak width

Separation

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 29: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Peak Broadening: Omethoate in Avocado in Run 1

4x10

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

Cpd 8: Omethoate: +EI MRM [email protected] (156.0000 -> 110.0000) avocado_10ppbGCPest_2_02rep1.D

6.809

Counts vs. Acquisition Time (min)

6.74 6.745 6.75 6.755 6.76 6.765 6.77 6.775 6.78 6.785 6.79 6.795 6.8 6.805 6.81 6.815 6.82 6.825 6.83 6.835 6.84 6.845 6.85 6.855 6.86 6.865 6.87 6.875 6.88 6.885 6.89 6.895 6.9 6.905 6.91 6.915 6.92 6.925

29 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 30: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Peak Broadening: Omethoate in Avocado in Run 65

4x10

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

Cpd 44: Deltamethrin: +EI MRM [email protected] (252.9000 -> 174.0000) avocado_10ppbGCPest_2_02rep10.D

Counts vs. Acquisition Time (min)

6.74 6.745 6.75 6.755 6.76 6.765 6.77 6.775 6.78 6.785 6.79 6.795 6.8 6.805 6.81 6.815 6.82 6.825 6.83 6.835 6.84 6.845 6.85 6.855 6.86 6.865 6.87 6.875 6.88 6.885 6.89 6.895 6.9 6.905 6.91 6.915 6.92 6.925

30 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 31: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Peak Broadening: Omethoate in Avocado in Run 1 versus Run 65

4x10

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

Cpd 8: Omethoate: +EI MRM [email protected] (156.0000 -> 110.0000) avocado_10ppbGCPest_2_02rep1.D

6.809

Counts vs. Acquisition Time (min)

6.74 6.745 6.75 6.755 6.76 6.765 6.77 6.775 6.78 6.785 6.79 6.795 6.8 6.805 6.81 6.815 6.82 6.825 6.83 6.835 6.84 6.845 6.85 6.855 6.86 6.865 6.87 6.875 6.88 6.885 6.89 6.895 6.9 6.905 6.91 6.915 6.92 6.925

31 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 32: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Peak Broadening: Recover Peak Shape with New Liner

4x10

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

Cpd 8: Omethoate: +EI MRM [email protected] (156.0000 -> 110.0000) avocado_10ppbGCPest_2_02rep1.D

6.809

Counts vs. Acquisition Time (min)

6.74 6.745 6.75 6.755 6.76 6.765 6.77 6.775 6.78 6.785 6.79 6.795 6.8 6.805 6.81 6.815 6.82 6.825 6.83 6.835 6.84 6.845 6.85 6.855 6.86 6.865 6.87 6.875 6.88 6.885 6.89 6.895 6.9 6.905 6.91 6.915 6.92 6.925

32 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 33: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Peak Broadening: The Case of the Wrong Liner

33 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 34: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 34

Baseline DisturbancesSudden changes, wandering, or drifting

Column or detector• Not fully conditioned or stabilized (electronics)

• Contamination

Flow• Changes in carrier and/or detector gas flows

• Valves switching, leaks

Drifting/wandering/weird disturbances

3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.0011.0012.0013.0014.0015.00

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 35: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 35

Noisy Baseline

Flow• Contaminated gas

• Incorrect detector settings

Column• Bleed if at high temperature

• In detector flame (poor installation)

Mild

Severe

Detector• Air leak - ECD, TCD

• Electronics malfunction

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 36: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 36

Spiking Baseline

Detector• Particles entering the detector

• Random: poor connection

• Regular: nearby "cycling" equipment (electronics)Application note 5991-2975EN

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 37: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 37

Quantitation Problems

Detector• Poor stability (electronics) or baseline disturbances (contamination)

• Outside detector's linear range or wrong settings

• Integration parameters

Activity (adsorption) in injector or column

Injector• Technique, settings, conditions• Syringe worn

Other• Co-elution

• Matrix effects

• Sample evaporation – leaky vials

• Sample decomposition

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 38: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What is Not Caused by a Column?

Page 38

Not responsible

• Peaks- Any reproducible sharp chromatographed peak

• Siloxanes (even though it looks like bleed spectrally)

• Degradation product peaks: Endrin Aldehyde, endrin ketone, DDE, DDD…

• Carry-over of sample compounds

• Splitting of peaks

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 39: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 39

Bleed profile (non-injection): baseline problemsInject a nonretained peak: peak shape problemsTest mix: all problemsIsolate the components: all problems

Condensation test: baseline problemsJumper tube test: baseline problems

Troubleshooting Tools

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 40: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 40

Generating a Bleed Profile

Time (min.)0 5 10 15 20 256000

7000

8000

9000

1.0e4

1.1e4

1.2e4

1.3e4

Agilent J&W DB-1, 30 m x 0.32 mm id, 0.25 µmTemperature program // 40 °C, hold 1 min // 20 °C/min to 320 °C, hold 10 min.

Produce when the column is new (for future reference) when there is a baseline problem

(Simply remove syringe from ALS)

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 41: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 41

Inject a Nonretained Compound to Check Flow Path

Good Installation Improper Installation orInjector Leak

Potential explanations:• Injector or septum leak• Too low of a split ratio• Liner problem

- (broken, leaking, misplaced)• Column position in injector and detector

Used to checkflow path

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 42: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 42

Test Mix – Make your Own!A test mix is used to determine how “good” the column is, or if the problem is related to the chemical properties of the analytes.

It is simplest to use your own standard.

Compound Purpose

Hydrocarbons EfficiencyRetention

Alcohols Activity

FAME's, PAH's Retention

Acids Acidic CharacterActivity

Bases Basic CharacterActivity

Test ConditionsInlet: Split (250°C)Detector: FID(320°C)

Flow:37.3 cm/sec(1.8 mL/min)

Carrier gas: HydrogenHoldup compound: Methane (0.671-min)Temperature program:Isothermal (110°C)

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 43: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

ULTRA Scientific is Now Part of Agilent Technologies

Page 43

Agilent ULTRA Chemical Standards have:• Best in class online search, compare, and ordering capabilities• Rapid shipping: 99.9% of orders dispatched within 24 to 48

hours (continental US only as of now)• Custom standard solutions including our new online custom

quoting tool, enabling customers to upload recipe formulations to and to modify the recipe before submitting it

• Tool will allow customers to see the quote pricing instantly and allow them to check quote pricing based on quantity range

• Check it out at https://www.agilent.com/en/product/chemical-standards

• Rigorously tested and manufactured under ISO 9001, ISO 17025, and ISO 17034 certifications

• Sample preparation materials, columns, supplies, instrumentation, and reference materials from a single source

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 44: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 44

Agilent J&W DB-624 ColumnQC Test Mix

Column: Agilent J&W DB-624 30 m x 0.53 mm id, 3.0 µm

Carrier: Helium at 40 cm/secmeasured at 35 °C

Injector: Direct, 260 °CDetector: FID, 300 °COven: 35 °C for 1.50 min

30 °C/min to 65 °C for 10 min

1. 1,2-Dichloropropane2. Octane3. Tetrachloroethylene4. Chlorobenzene5. Nonane

5 10 15 20 25

1.0e4

2.0e4

3.0e4

4.0e4

5.0e4

6.0e4

7.0e4

8.0e4

Time (min.)

2.71

7.4310.92

12.4917.42

20.78

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 45: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 45

Example of Column Contamination Agilent J&W DB-624 QC Test MixAfter 75 injections of oily sample

0 5 10 15 20

6000

7000

8000

9000

1.0e4

1.1e4

1.2e4

1.3e4

1.4e4

1.5e4

Time (min.)

2.21

3.30

6.03

9.26

10.46 14.40 17.86

*Temperature program// 35°C hold 1.50 min // 30°/min to 65°C, hold 10 min

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 46: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 46

Example of Column Contamination

Removed 112

m from injector end

Time (min.)0 5 10 15 20 255000

6000

7000

8000

9000

1.0e4

1.1e4

1.2e4 2.80

7.3410.79

12.33

17.19

20.56

*Before column rinse and bakeTemperature program // 35 °C hold 1.50 min // 30° C/min to 65 °C, hold 10 min

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 47: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 47

Example of Column Contamination

0 10 20 30 40 50 60 705000

6000

7000

8000

9000

1.0e4

1.1e4

1.2e4

1.3e4

Time (min.)

112

m removed*QC test mix to upper temperature limit

*Before column bakeTemperature program // 35 °C, hold 1.50 min // 30 °C/min to 65 °C, hold 15 min // 20 °C/min to 260 °C, hold 50 min

We have more semivolatilecontamination!

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 48: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 48

Condensation Test

Used to isolate the cause of*:

• Erratic baselines

• Ghost peaks or carry-over

*Use when problems are worse after periods of GC non-use

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 49: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 49

Condensation Test

Procedure:

• Leave GC at 40-50 °C for > 8 hours

• Blank run

• Repeat a blank run immediately after the first blank run is complete

• Compare the two blank runs

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 50: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 50

Condensation Test

Results

First blank run is worse

• Contaminants (from injector, lines, traps or carrier gas) carried into the

column

• Blank runs the same: contaminants are not strongly focused on the front

of the column

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 51: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 51

Purpose

• Helps to locate the source of contamination or noise

• Isolates GC components

Jumper Tube Test

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 52: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 52

Isolate the detector

• Remove column from the detector

• Cap detector and turn on

• Blank run

Jumper Tube Test

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 53: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 53

Jumper Tube Test

Isolation of detector – results:

Detector is the problem

Detector OK

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 54: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 54

Jumper Tube Test

Isolate the injector

• Connect the injector and detector

- 1-2 meters deactivated fused silica tubing

• Turn on carrier gas

• Blank run

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 55: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 55

Jumper Tube Test

Isolate the injector – results:

Injector OK

Injector, lines or carriergas contaminated

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 56: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 56

Jumper Tube Test

Isolate the column

• Re-install the column

• Set up as before

• Blank run

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 57: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 57

Jumper Tube Test

Isolate the column – results:

• Problem returns? It’s the column

• Problem gone? Previous leak, solid debris, or installation problem

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 58: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 58

Troubleshooting Example

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 59: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What my TIC should look like:

What my TIC looked like:5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Everyone needs to have a reference checkout samplethat they can use to confirm whether their system is OK.

6x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest_3.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Problem: No Peaks with Semivolatiles Checkout Mixture

59 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 60: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Problem: No Peaks with Semivolatiles Checkout Mixture

What could cause this?

• The wrong vial was injected

• The sample has degraded

• The inlet is leaking

• The column is damaged

60 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 61: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Problem: No Peaks with Semivolatiles Checkout Mixture

What could cause this?

• The wrong vial was injected: Sequence and vial checked, no problem found

• The sample has degraded

• The inlet is leaking

• The column is damaged

61 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 62: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Problem: No Peaks with Semivolatiles Checkout MixtureWhat could cause this?

• The wrong vial was injected: Sequence and vial checked, no problem found

• The sample has degraded: A new vial of standard was used, no difference observed

• The inlet is leaking

• The column is damaged

62 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 63: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Problem: No Peaks with Semivolatiles Checkout Mixture.What could cause this?

• The wrong vial was injected: Sequence and vial checked, no problem found

• The sample has degraded: A new vial of standard was used, no difference observed

• The inlet is leaking: A tune was performed. O2, N2, and H2O levels were normal

• The column is damaged

63 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 64: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Problem: No Peaks with Semivolatiles Checkout MixtureWhat could cause this?

• The wrong vial was injected: Sequence and vial checked, no problem found

• The sample has degraded: A new vial of standard was used, no difference observed

• The inlet is leaking: A tune was performed. O2, N2, and H2O levels were normal

• The column is damaged: “Well, I guess I need to replace my column”

64 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 65: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What could cause this?

• The wrong vial was injected: Sequence and vial checked, no problem found

• The sample has degraded: A new vial of standard was used, no difference observed

• The inlet is leaking: A tune was performed. O2, N2, and H2O levels were normal

• The column is damaged: Well, I guess I need to replace my column

Problem: No Peaks with Semivolatiles Checkout Mixture

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

WAITTest (a few more things) before

you replace!

65 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 66: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What does a working GC/MS

look like?

Half-split the problem

Make repairs, as necessary

Put the system back together

Develop steps to prevent re-

occurrence

Follow a Logical Troubleshooting Procedure!

Start Finish

Step 1 Step 1 Step 1 Step 1 Step 1

66 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 67: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Troubleshooting Step 1: What is the “Working System”?

What are the components of the GC/MS system (follow the sample flow-path)• Agilent 7693A autosampler + 10 μL syringe• Agilent 7890B GC • Agilent MultiMode Inlet (with CO2 cryo)• Agilent J&W HP-5ms UI, 30 m x 0.25 mm x 0.25 μm• Agilent 5977A Series Extractor GC/MSD

GCelectronics

Autosampler

Vacuum system Detector

Power

Gases

Inlet

Interface

Syringe

SourceQuadrupoleDetector

MSelectronics

Sample

67 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 68: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Compare your current data to known good data, when possible.Use over-lay to zero-on on differences• How does your background compare to normal?

• Does the problem occur for every run, every analyte, every method?Only affects certain samples/analytes/Instruments?

• Are the peaks smaller or larger than normal?

• Is the peak shape gaussian, or are the peaks splitting, tailing, or saturated?

Troubleshooting Step 1: What is the “Working System”?What does a working

GC/MS look like? Half-split the problem Make repairs, as necessary

Put the system back together

Develop steps to prevent re-occurrence

68 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 69: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Compare your current data to known good data, when possible. • How does your background compare to normal?

Background looked a LOT bigger than peaks in the good TIC

• Does the problem occur for every run, every analyte, every method?Only affects certain samples/analytes?

Occurring on all checkout sample runs attempted

• Are the peaks smaller or larger than normal?Definitely smaller

• Is the peak shape gaussian, or are the peaks splitting, tailing, or saturated?Let’s find out

Troubleshooting Step 1: What is the “Working System”?

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

69 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 70: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

5x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

6x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest_3.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Good, previous run from April

Bad, recent run

Compare your current data to known good data. Now, the data is much clearer, and the background is not significantly higher. Signals in separate scales:

8.0 x 105

8.0 x 106

Troubleshooting Step 1: What is the “Working System”?

Page 70 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 71: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

6x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

Compare your current data to known good data. Now, the data is much clearer, and the background is not significantly higher. Signals with linked Y axis:

Similar background profiles

Good, previous run from April

Bad, recent run

6x10

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

6.5

7

7.5

8

+ TIC Scan SVOA_10ppm_CF_sampletest_3.d

Counts vs. Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35

8.0 x 106

8.0 x 106

Troubleshooting Step 1: What is the “Working System”?

Page 71 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 72: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Troubleshooting Step 2: Break Apart (Half-Split) the Problem

GCelectronics

Autosampler

Vacuum system Detector

Power

Gases

Inlet

Interface

Syringe

SourceQuadrupoleDetector

MSelectronics

Sample

Think of a set of tests that will break the system into smaller pieces.

1. Try a new sample.

2. Tune the MS to half-split the detector from the GC.

3. Perform a manual injection with a new syringe to split autosampler and inlet/column.

72 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

NOT THEPROBLEM

MSD ?

ALS ?

Page 73: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Troubleshooting Step 2: Break Apart (Half-Split) the Problem

GCelectronics

Autosampler

Vacuum system Detector

Power

Gases

Inlet

Interface

Syringe

SourceQuadrupoleDetector

MSelectronics

Sample

NOT THEPROBLEM

NOT THEPROBLEM

Think of a set of tests that will break the system into smaller pieces.

1. Try a new sample.

2. Tune the MS to half-split the detector from the GC.

3. Perform a manual injection with a new syringe to split autosamplerand inlet/column.

73 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 74: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Troubleshooting Step 2: Break Apart (Half-Split) the Problem

GCelectronics

Autosampler

Vacuum system Detector

Power

Gases

Inlet

Interface

Syringe

SourceQuadrupoleDetector

MSelectronics

Sample

NOT THEPROBLEM

NOT THEPROBLEM

Think of a set of tests that will break the system into smaller pieces

1. Try a new sample.

2. Tune the MS to half-split the detector from the GC.

3. Perform a manual injection with a new syringe to split autosamplerand inlet/column.

74 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 75: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Let’s focus on the autosampler and syringe:

While sample was new, what is the solvent?

What kind of syringe?

Does the autosampler work?

Does the syringe pull up liquid?

Autosampler

SyringeDichloromethane

Agilent 10 µL syringe, 23-26s/42/cone (G4513-80204)

Autosampler turns and moves plunger up and down

No, it doesn’t

We may have found the problem!

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Troubleshooting Step 2: Narrow Focus of the Problem

75 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 76: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Syringe

Troubleshooting Step 3: Make the Repair

Replace the syringe with a 10 μL PTFE tipped plunger syringe (G4513-80203) – a much easier repair than venting and changing the column.

PTFE tipped syringes are more chemically resistantand offer a reduced chance of carry over and longer syringe lifetime.

Proper syringe maintenance must still be performed. Cleanand refill syringe wash vials frequently.

Beware highly concentrated samples and samples withparticulates (organic material, salts, etc.)

PTFE plunger tip

76 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 77: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

SUCCESS!

What happened with a new syringe?

What does a working GC/MS look like? Half-split the problem Make repairs, as

necessaryPut the system back

togetherDevelop steps to

prevent re-occurrence

Troubleshooting Step 4: Put the System Back Together

77 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 78: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

78

Have a Good Troubleshooting Story? Let Us Know!

Please call or email us today to share a troubleshooting success story or if you need help troubleshooting!

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 79: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 79

Troubleshooting Tips

1. Isolate the problem – half-split the system into its component parts

(blank run, inject unretained compound, jumper tube test)

2. Change only one variable at a time

3. Compare before/after chromatograms

(Peak shape, response, retention, baseline rise, background, look for trends, etc.)

4. Utilize technical support

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 80: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 80

Remember

• Multiple cause and effect

• Do not change too many variables at once

Complete system = carrier gas + injector + column + detector + data system

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 81: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Contact Agilent Chemistries and Supplies Technical Support

1-800-227-9770 Option 3, Option 3:Option 1 for GC and GC/MS columns and suppliesOption 2 for LC and LC/MS columns and supplies

Option 3 for sample preparation, filtration, and QuEChERSOption 4 for spectroscopy supplies

Option 5 for chemical standardsAvailable in the USA and Canada 8–5, all time zones

[email protected]@agilent.com

[email protected]@[email protected]

81

August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593

Page 82: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Page 82

“Everything was Just Fine and then this Happened!” “How do I go about Troubleshooting?”

Logic = Something changed (slowly or sudden) = Something is different

Agilent Restricted

Track events- log book• Changed column, liner, septum, syringe, etc.• Injected samples, other method, etc.• Did maintenance, cut column, inlet flush, etc.

Page 83: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Hexane blanks (testing vial storage over time)

Immediate1 day1 week

Hexane Solvent Impurities

Page 84: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

If GC/MS was off for 1+ week (no carrier gas flow)…

August 6, 2020 Title Confidentiality label Regulatory statement (if applicable)84

5x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

System “open” for 1+ week

TIC looks okay (I think). How does it compare to a previous run of the same sample?

Page 85: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

If GC/MS was off for 1+ week (no carrier gas flow)… zoom out

August 6, 2020 Title Confidentiality label Regulatory statement (if applicable)85

5x10

0

0.5

1

1.5

2

2.5

3

3.5

Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

System “open” for 1+ week (Dec)System under vacuum (Nov)

“Open system” TIC is ~10x lower than good run in the previous month. What happens if we replace the column and line

Page 86: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

If GC/MS was off for 1+ week (no carrier gas flow)…

August 6, 2020 Title Confidentiality label Regulatory statement (if applicable)86

“Open system” TIC is ~10x lower than good run in the previous month. What happens if we replace the column and liner?

6x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

System “open” for 1+ week (Dec)System under vacuum (Oct)

Page 87: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Recover peak response with new column and liner

August 6, 2020 Title Confidentiality label Regulatory statement (if applicable)87

6x10

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

Acquisition Time (min)

3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

System “open” for 1+ week (Dec)System under vacuum (Oct)New column, liner, gold seal (Dec)

Try a new liner and re-conditioning column first.If response doesn’t recover, a new column may be required.

Page 88: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Brand X-5ht Peak Symmetry DegradationPeak Name

0 Methane1 Decane2 1-Octanol3 2,6-Dimethylphenol4 2,6-Dimethylaniline5 Naphthalene6 1-Decanol7 Tridecane8 Methyl Decanoate

min0 2 4 6 8 10 12 14

pA

0

20

40

60

80

10

30

50

70

901

2 3

45

6 78

min0 2 4 6 8 10 12 14

pA

0

20

40

60

80

10

30

50

70

90 1

23

4 5

6 78

40 Hours at 400°C

05

101520253035404550

0 10 20 30 40 50

Taili

ng F

acto

r

Time at 400°C (hours)

Brand X-5ht 1-Octanol Tridecane Methyl Decanoate

0

5

10

15

20

25

30

35

40

45

50

0 5 10 15 20 25 30 35 40

Taili

ng F

acto

r

Time at 400°C (hours)

Tridecane Tailing Factor

Brand X-5ht

DB-5ht

Page 89: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Column Efficiency Over120 hours at 400°CpA

0

20

40

60

80

10

30

50

70

90

min0 2 4 6 8 10 12 14 16 18

min0 2 4 6 8 10 12 14 16 18

pA

0

20

40

60

80

10

30

50

70

90

0

500

1000

1500

2000

2500

3000

3500

4000

0 20 40 60 80 100 120

plat

es/m

eter

Time at 400°C (hours)

Tridecane Peak Name0 Methane1 Decane2 1-Octanol3 2,6-Dimethylphenol4 2,6-Dimethylaniline5 Naphthalene6 1-Decanol7 Tridecane8 Methyl Decanoate

1

2 3

4 5

6 78

12

3

4 5

6 78

Agilent J&W DB-5ht30m x 0.25 mm x 0.10 µm

Brand X-5ht30m x 0.25 mm x 0.10 µm

Agilent Publication 5994-1013EN

Page 90: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Column Efficiency at 430°CPeak Name

0 Methane1 Decane2 1-Octanol3 2,6-Dimethylphenol4 2,6-Dimethylaniline5 Naphthalene6 1-Decanol7 Tridecane8 Methyl Decanoate

1pA

0

20

40

60

80

10

30

50

70

90

min0 2 4 6 8 10 12 14

Agilent J&W DB-5ht

23

45

6 7 8

pA

0

20

40

60

80

10

30

50

70

90

min0 2 4 6 8 10 12 14

1

23

45

6 78 Brand X-5ht

0

0

0

500

1000

1500

2000

2500

3000

3500

0 2 4 6 8 10

plat

es/m

eter

Time at 430 °C (hours)

Tridecane

Agilent Publication 5994-1013EN

Page 91: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Phase Degradation Increases Retention

5.0

5.5

6.0

6.5

7.0

7.5

8.0

8.5

9.0

9.5

0 20 40 60 80 100 120

k

Time (hours)

Tridecane "k"

min0 2 4 6 8 10 12 14 16 18

Initial

120 Hours at 400°C1

2 3

45

6 78

12

3

4 5

6 7 8

Peak Name0 Methane1 Decane2 1-Octanol3 2,6-Dimethylphenol4 2,6-Dimethylaniline5 Naphthalene6 1-Decanol7 Tridecane8 Methyl Decanoate

Brand X-5ht

Agilent J&W DB-5ht

Brand X-5ht

𝑘𝑘 =𝑡𝑡𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐 − 𝑡𝑡 𝑐𝑐𝑚𝑚𝑡𝑡𝑚𝑚𝑚𝑐𝑐𝑚𝑚

𝑡𝑡 𝑐𝑐𝑚𝑚𝑡𝑡𝑚𝑚𝑚𝑐𝑐𝑚𝑚

Page 92: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

“Potholes” Created as the Phase DegradesRaw Fused Silica exposed…..

The more heat you add the more “potholes” you create

Page 93: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Recover peak shape with new liner (black)

August 6, 2020 Title Confidentiality label Regulatory statement (if applicable)95

4x10

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

Cpd 44: Deltamethrin: +EI MRM [email protected] (252.9000 -> 174.0000) avocado_10ppbGCPest_3_02rep1.D

Counts vs. Acquisition Time (min)

6.74 6.745 6.75 6.755 6.76 6.765 6.77 6.775 6.78 6.785 6.79 6.795 6.8 6.805 6.81 6.815 6.82 6.825 6.83 6.835 6.84 6.845 6.85 6.855 6.86 6.865 6.87 6.875 6.88 6.885 6.89 6.895 6.9 6.905 6.91 6.915 6.92 6.925

Page 94: It’s Peak Season for Great Peak Shape - Agilent · It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography. 1. Mark Sinnott Alexander Ucci

Recover peak shape with new liner (black)

4x10

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

1.1

1.15

1.2

1.25

1.3

1.35

1.4

1.45

1.5

1.55

1.6

Cpd 8: Omethoate: +EI MRM [email protected] (156.0000 -> 110.0000) avocado_10ppbGCPest_3_02rep1.D

6.805

Counts vs. Acquisition Time (min)

6.74 6.745 6.75 6.755 6.76 6.765 6.77 6.775 6.78 6.785 6.79 6.795 6.8 6.805 6.81 6.815 6.82 6.825 6.83 6.835 6.84 6.845 6.85 6.855 6.86 6.865 6.87 6.875 6.88 6.885 6.89 6.895 6.9 6.905 6.91 6.915 6.92 6.925

August 6, 2020 Title Confidentiality label Regulatory statement (if applicable)96