19
Isolated high speed 8 channel multi function logger midi LOGGER GL900 midi LOGGER GL900 Product introduction Product introduction The data logger that meets The data logger that meets with various equipment tests with various equipment tests NEW GRAPHTEC CORPORATION

Isolated high speed 8 channel multi function logger midi LOGGER GL900 Product introduction midi LOGGER GL900 Product introduction The data logger that

Embed Size (px)

Citation preview

Isolated high speed 8 channel multi function logger

midi LOGGER GL900midi LOGGER GL900

Product introduction Product introduction

The data logger that meets The data logger that meets with various equipment testswith various equipment tests NEW

GRAPHTEC CORPORATION

Out line Out line Recently, improvement of product durability becomes more important as a development theme for the electronics equipments and industrial equipments, so the drop down test, vibration test, mechanical shock test and environment test have been performed more frequently. Especially for the durability test for mechanical movement parts and driving parts, several ten thousands times or several thousands hours durability test are often performed.Based on these backgrounds, high speed sampling and long time data capturing and simultaneous measurement for several phenomena (voltage, temperature and high voltage) are strongly required for the measurement equipments. Also for the long time durability test for multiple products, it is required that the number of test equipment needs to be increased, so compact and reasonable price measurement equipments are required for capturing of these test data.

We provide the midi LOGGER GL900 that is an 8 channels isolation input for safe measurement for these various product tests. Further more, the multifunction input which corresponds to 500V DC high voltage, humidity, pulse and logic input for measurement of various phenomena.

Target of the GL900Target of the GL900   

Company H

8835

1MS/s 100KS/s

Company K

NR600

Company K GR7000

Company Y

DL750P 

Company N RA2300

  Company H

8860-50

How many customers measure 1MS/s measurement by their measurement equipment which has the 1 MS/s sampling capability?We guess that the actual measurement 100 KS/s (10 micro second sampling) or slower, isn ’t it?Graphtec provides practical specification and price of products to meet customer’s actual frequency band.

Price(MSRP)1.1

0.3

0.6

1.0

Instantaneous phenomenon ・ Drop down test・ Mechanical shock test・ Vibration test

Long time measurement・ Durability test・ Tension/Compression test・ Environment test

Main application

GL900: Competitive price

8 channels, isolation, simultaneous sampling

Sampling Speed ( Sample / sec )

Company Y

SL1400

Company H 8841

(Million yen)

>Large size 5.7 inch color LCD for bright and easy- to- read waveform display

> Data capture for USB memory which can easily be connected to the PC

> 16 bits A/D resolution with High speed 8 channels simultaneous sampling

>8 channels isolated multi function input

Main features Main features

Isolated high speed 8 ch multi function loggerIsolated high speed 8 ch multi function logger   

The midi LOGGER GL series are popular as the handy type logger, now, the GL900 is designed as vertical operation for customer’s usage.Also the size is drastically reduced as an isolated high speed measurement equipment.

232mm

150mm

80mm

4 ch for either Logic or PulseFor pulse, Revolutions, Counts, Instant are selectable for each channels

BNC terminal for easy connection of voltage input

Screw terminal for easy connection of thermo couples

Isolated input for safe connection and variety of input ranges for corresponding various test items

Voltage : 20 mV to 500 V (Between + and - terminals : Max 500 V p-p)

Temperature : Thermocouple Humidity : Humidity sensor (B-530) Logic or Pulse : (Revolution, Integration and Instant)

4ch separated from analog 8ch input

Multi-function input with isolated 8 channels Multi-function input with isolated 8 channels

Corresponding to high voltage measurementCorresponding to high voltage measurement   

ACpower supply  

Sequence : Logic

Current : Clamp meter

Voltage  

Temperature : Thermo couple

CM-211   CM-113  

Humidity : Dedicated humidity sensor  

Wide 500 V range is provided for voltage input, so it corresponds 100 to 200 V AC power supply waveform measurement.Also, using with the logic input and clamp meters, power supply voltage and current and the sequence control can be measured as the same time.

When input each signal as voltage, it is possible to convert to original unit using the unit conversion function

Clamp meter

Max. sampling 10 sec with all channels simultaneous sampling A/D resolution is 16 bit high resolution

High speed 8ch simultaneous sampling with 16 bit high resolution

8

12

14

16

Storage Oscilloscope

100M 1M 10K 10

Data logger GL900

Memory recorder

  Sampling speed (S/s)

A/D

resolu

tion

(b

it)

1K 20M

Temperature Dynamic characteristics Control signal

Circuit signal

Speed-up

Hig

h

resolu

tion

Signal

100K

Storage destination 10 s 100 s 500 s 1ms 10ms 100ms 1s 1min

64 MB internal memory Yes Yes Yes Yes Yes Yes Yes Yes

256MB internal Flash

memory

No No No Yes Yes Yes Yes Yes

USB memory No No No Yes Yes Yes Yes Yes

Corresponding to various measurement sceneCorresponding to various measurement scene

Capture destination

10μs 100μs

500μs 1ms 10ms 100ms 1s

Internal RAM(1 million point max.)

10 sec

100 sec

8 min20 sec

16 min40 sec

2 hours46 min

1 day3 hours

11 days13

hours

256 MB internal flash memory

× × × 11 hours 4 days 49 days493 days

512 MB USB memory

× × × 22 hours 8 days 98 days986 days

Application・ Drop down test・ Shock test・ Vibration test

Application・ Durability test・ Tension/Compression test・ Environment test

High speed phenomena such as the drop down test and the mechanical shock test can be measured with 1ms or faster high speed sampling then stored to internal RAM for max. 1 million points. Also all channels are simultaneously sampled without time difference between channels, it is possible to check the correlation of each parts.Further more, 1ms to 1min slow sampling is for the durability test, and large size 256MB internal flash memory or external USB memory can store long time measurement.

・ Bright and easy-to-read, Large size 5.7 inch color LCD ・ Simple setting operation by cursor keys ・ High speed waveform can be enlarged or reduced, Oscilloscope like operation

Large size 5.7 inch color LCD for bright and easy-to- read waveform display

Simple setting operation by cursor keys

Easy-to-read large size 5.7 inch color TFT LCD

Magnifying of time axis

Compression of time axis

TIME/DIV key

10ms/ DIV to

1 min / DIV

Previously, measurement equipment needs to record the input signal once to confirm the waveform before start measurement.But the GL series installs the free running display function, so the input signal is displayed on the screen soon after power on, you can check the input signal soon, and easily set the measurement condition within short period of time.

Easy setting operation with Easy setting operation with the free running display functionthe free running display function   

Previous models  

Press START key to start measurement,check the input signal and press STOP key  

Set input range  

Press START key again to start measurement,check the input signal and press STOP key   

Press START key to start actual measurement

If necessary, repeat above steps

Check input signal by free running display

Set the input range

Press START key for actual measurement

GL900  

Waveform display area  

Waveform is displayed during setting for easy operation

TIME/DIV, Position and Span setting can be done in the free running display  

Display scrollMax..: 10ms/DIV

Following two file formats are selectableGBD (Graphtec binary data) format : data size is smaller than CSV formatCSV format : You can open it by Excel or other software without using any dedicated software

The USB memory can be used as the external memory device.If you specify the USB memory as the capture destination, the data is directly stored to the USB memory.Also, for the long time measurement such as the durability test, you can measure it just by use the USB memory which has necessary data capacity.The transfer to PC is also possible just by insert the USB memory to the PC.* The max. file size for measurement data is 2GB. Please use standard type USB memory which does not support the functions such as the finger print recognition.

Data can be stored to USB memory Data can be stored to USB memory which easily can be connected to the PCwhich easily can be connected to the PC   

External USB memory terminal  

Data can be transferred to the PC easily.

Using the combination of trigger function and timer function, only necessary data can be stored without capturing wasted data.

Wealth of trigger and timer function  

Jan. 1 Jan. 5 Jan. 10 Jan. 15 Jan. 20 Jan. 25 Jan. 30

Timer Mode : Day/Time Start Settings : Date 2008 – 01-10 Time 00 : 00Start Settings : Date 2008 – 01-14 Time 23 : 59

Trigger Settings Start Source : OffStop Source : Off

Timer Settings

Timer Mode Off , Day/Time , Daily , Hourly

Trigger Settings

Start Source Off , Level , Ext.

Stop Source Off , Level , Ext. , Time

Pre Trigger 0 to 100 %

Repeat measurement On/Off and repeat interval

: Data capture

: Timer is activated

Example 1 : When measure 4 days from Jan. 10

Only when the capturing

destination is internal RAM

Start key On

Wealth of trigger and timer function  

00 :00 04 :00 08 : 00 12 : 00 16 : 00 20 : 00 24 : 00

Timer Mode : OffTrigger Settings Start Source : Off

Stop Source : TimeTime Settings : 1 h 00 min 00 sec

Repeat : OnRep. int. : 0004 h 00 min

: Data capture

Example 3 : When measure 1 hour in 4 hours interval every day

Repeat interval is counted from the time

when trigger is activated

Repeat interval : 4 hours

00 : 00 12 : 00 00 : 00 12 : 00 00 : 00

Start Trigger : 3V

Stop Trigger : 2V

Timer Mode : Daily Start Settings : 09 : 00Stop Settings : 17 : 00

Trigger Settings Start Source : LevelLevel Settings : CH1 , Mode : H , Level : 3 V

Stop Source: LevelLevel Settings : CH1 , Mode: L , Level : 2 V

Repeat : On

5V

0 V

: Data capture

: Timer is activated

Example 2 : When measure abnormal signal of equipment during working time 

: Trigger point  Measurement finishes by the priority for timer

setting

Start key On

Start key On

Wealth of trigger and timer function

00:00 01:00 02:00 03:00 04:00 05:00 06:00

5V

0V

: Data capture

: Timer is activated

Example 4 : When measure for 20 minutes in every hour

00:00 01:00 02:00 03:00 04:00 05:00 06:00

Start Trigger 3V

Stop Trigger 2V

5V

0V

Example 5 : When measure only abnormal signal within 20 minutes of every hour  

Measurement stops by the priority for

timer setting

Start key On

Start key On

Timer Mode : Hourly Start Settings : 00 min 00 secStart Settings : 20 min 00 sec

Trigger Settings Start Source : OffStop Source : Off

: Data capture

: Timer is activated

: Trigger point  

Timer Mode : Hourly Start Settings : 00 min 00 secStop Settings : 00 min 20 sec

Trigger Settings Start Source : LevelLevel Settings : CH1 , Mode : H , Level : 3 V

Stop Source: LevelLevel Settings : CH1 , Mode: L , Level : 2 V

Repeat : On

Possible to use as X-Y recorderPossible to use as X-Y recorder   Duplicating the X-Y recorder’s action.Pen up/down action is duplicated, so it is possible to use like an analog X-Y recorder.Corresponding to Max. 4 pen X-Y recorder.Also, the data is digital format, so you can easily check the data and create the report after finish the measurement.

Point moves as if pen moves If capturing the screen copy then paste on the Excel, you can use it as report.

Pen moving can be checked with pen-up condition  

Waveform is displayed by pen-down. Data capture starts

by measurement start.  

After finish measurement, data can be check by digital

value using cursor.

NTP server Ethernet

The GL900’s time can be synchronized to the NTP server’s time at periodic intervals. ( NTP : Network Time Protocol )

Setting, control and data can be transferred without using the standard attached software  

Ethernet

FTP client function

The data captured to the GL900’s internal memory or the USB memory stick can be automatically backed up to the FTP server at fixed intervals

USB drive mode

USB drive mode is a fast and easy way to transfer data from internal memory to an external computer

NTP client server function

WEB server function

Waveform display and GL900 settings can be made via a web browser such as Internet Explorer

Ethernet USB

USB DRIVEMODE

USB and Ethernet is provided as standard (dedicated PC software is attached)Using USB dive mode, WEB server and FTP server function, you can transfer the data to PC, and control and setting of GL900 from PC is possible.

Remote control and data transfer is possible

Ethernet / USB   USB I/F  

Ethernet I/F  

Item Description

Number of input channels 8ch

Type of input terminal BNC and M3 Screw type terminal

Method All channels isolated, Simultaneous sampling

Measurement range

Voltage 20mV to 500V / F.S. (Between + and – terminals, 500 V p-p max.)

Temperature Thermocouple: K, J, E, T, R, S, B, N, W(WRe5-26)

Humidity 0 to 100% ( When using optional B-530 humidity sensor is used )

External input /

output

Input Trigger: 1ch, Selectable between pulse and Logic: 4ch

Output Alarm: 4ch

A / D converter 16bit(out of which 14 are internally acknowledged)

Maximum sampling interval 10 micro-second

Internal memory size RAM : 1 million data, 256MB Flash memory

External memory USB memory (1 file 2 GB)

Calculation function Statistics calculation, Searching function

Display 5.7 inch TFT color LCD

Interface to PC Ethernet ( 10BASE-T / 100BASE-TX ) , USB (Compatible with full speed)

Operating environmentTemperature : 0 to 40 (15 to 40 when operating by battery ) , Humidity : 5℃ ℃ ~ 85 % R.H.

Power supplyAC adapter ( 100 to 240V AC ,50 to 60Hz ) , DC input ( 8.5 to 24V DC)Battery pack : 2 batteries installable

External dimensions (approx.) 232W x 150H x 80D mm

Weight (approx.) 1.1kg ( Excluding the AC adapter and battery )

Vibration resistance Compatible with JIS Vibration testing methods for automobile Type 1 Class A-equivalent

GL900 Specifications

Accessories Accessories

Item Model Remark

Battery pack B-5172 batteries need to be installed when operate

Logic alarm cable B-513 2m

DC power cable B-514 2m

Humidity sensor B-530 3m

Safe probe RIC-141 1:1 , 42pF

BNC – BNC cable RIC-112 1.5m

BNC – Banana cable RIC-113 1.5m

BNC –Alligator Clip cable RIC-114 1.5m