Upload
noelle-mcdowell
View
19
Download
1
Embed Size (px)
DESCRIPTION
ISIS Beam Protection System and MICE operation. Dean Adams 29 Nov 07 Presented by Chris Rogers. ISIS Beam Protection System. - PowerPoint PPT Presentation
Citation preview
ISIS Beam Protection System and MICE operation.
Dean Adams
29 Nov 07
Presented by Chris Rogers
ISIS Beam Protection System
• ISIS Beam Protection System (BPS) consists of hardware interlocks (magnets, RF..) and beam diagnostic measurements (Intensity and Beam loss monitors). The latter using DAQ and tolerance comparison to determine trips.
• To minimise beam off time an ‘inhibit unit’ sits above the BPS system and turns the beam off on a BPS trip for ~ 1.5 s and then automatically switches the beam back on. If another BPS trip occurs within 10s then the beam is switched off by insertion of a beam stop. This then requires a manual reset to enable beam.
Beam Intensity Monitors• 1 Current transformer
(R5IM1) sits in SP5.
• DAQ resolution 2e10 protons per pulse.
• Acceleration tolerances
2.5 – 9.8 ms
• BPS trip :
3 consecutive > 2e12 each
25 consecutive > 0.5e12 average/pulse
• BPS warning displayed in ISIS MCR @ 30 s and 20 mins if average per pulse > 0.25e12 or 0.2e12
Beam Loss Monitors
• 39 BLM’s in ring.• Each BLM signal is integrated
0-10 ms and compared to a tolerance.
• Trip issued when 20 pulses > tolerance in 15 MINUTES.
• Pulse count reset on each trip.• BLM upgrade Jan 08. Pulse
count changed to 20 CONSECUTIVE pulses.
MICE operation and the current ISIS BPS
• Assumptions: MICE at 1 Hz, ISIS at 50 Hz.
• Intensity Monitors
3 consecutive pulses : MICE cannot force this trip alone.
25 pulse average : MICE could produce one BPS trip if it intercepted more than 25* 0.5e12 = 1.25e13. This would produce an inhibit. The next MICE dip occurs in less than the 10s inhibit window so additional loss would cause another trip and turn the machine off.
• Beam Loss Monitors
20 pulses in 15 mins : MICE can inhibit but never trip
20 consecutive pulses: MICE cant inhibit or trip.
Where to go next.
• Old HEP target operated @50 Hz diping at 7 ms for ~ 200 µs. ISIS BLM tolerances were 0.2 V in super period 7. MICE target @1Hz but dips from 8-10 ms. Different machine activation source. Need to quantify..
• Current BPS protection not suitable for MICE. ISIS will need to produce some new trip mechanism on the BLM’s. Ie DAQ system measuring SP7 losses at 1 Hz connected to ISIS BPS system.
• Target tests in Jan 08 should limit loss to 0.05 V trip level. During this time machine activation can be measured. The requirement for MICE rates and ISIS activation can then be reviewed.