9
The Spine Journal 4 (2004) 379–387 Is galvanic corrosion between titanium alloy and stainless steel spinal implants a clinical concern? Hassan Serhan, PhD a, * , Michael Slivka, MSc a , Todd Albert, MD b , S. Daniel Kwak, PhD a a DePuy Spine, Inc., 325 Paramount Drive, Raynham, MA 02767, USA b Thomas Jefferson University Hospital, Philadelphia, PA 19107, USA Received 12 May 2003; accepted 2 December 2003 Abstract BACKGROUND CONTEXT: Surgeons are hesitant to mix components made of differing metal classes for fear of galvanic corrosion complications. However, in vitro studies have failed to show a significant potential for galvanic corrosion between titanium and stainless steel, the two primary metallic alloys used for spinal implants. Galvanic corrosion resulting from metal mixing has not been described in the literature for spinal implant systems. PURPOSE: To determine whether galvanic potential significantly affects in vitro corrosion of titanium and stainless steel spinal implant components during cyclical compression bending. STUDY DESIGN/SETTING: Bilateral spinal implant constructs consisting of pedicle screws, slotted connectors, 6.35-mm diameter rods and a transverse rod connector assembled in polyethylene test blocks were tested in vitro. Two constructs had stainless steel rods with mixed stainless steel (SS-SS) and titanium (SS-Ti) components, and two constructs had titanium rods with mixed stainless steel (Ti-SS) and titanium (Ti-Ti) components. METHODS: Each construct was immersed in phosphate-buffered saline (pH 7.4) at 37 C and tested in cyclic compression bending using a sinusoidal load-controlling function with a peak load of 300 N and a frequency of 5 Hz until a level of 5 million cycles was reached. The samples were then removed and analyzed visually for evidence of corrosion. In addition, scanning electron microscopy (SEM) and energy dispersive spectrometry (EDS) were used to evaluate the extent of corrosion at the interconnections. RESULTS: None of the constructs failed during testing. Gross observation of the implant components after disassembly revealed that no corrosion had occurred on the surface of the implants that had not been in contact with another component. The Ti-Ti interfaces showed some minor signs of corrosion only detectable using SEM and EDS. The greatest amount of corrosion occurred at the SS-SS interfaces and was qualitatively less at the SS-Ti and Ti-SS interfaces. CONCLUSIONS: The results from this study indicate that when loaded dynamically in saline, stainless steel implant components have a greater susceptibility to corrosion than titanium. Further- more, the galvanic potential between the dissimilar metals does not cause a discernible effect on the corrosion of either. Although the mixture of titanium alloy with stainless steel is not advocated, the results of this study suggest that galvanic corrosion is less pronounced in SS-Ti mixed interfaces than in all stainless steel constructs. 2004 Elsevier Inc. All rights reserved. Keywords: Corrosion; Dynamic testing; Pedicle screw implants FDA device/drug status: not approved for this indication (Isola/VSP Spine System). Authors SDK, HS and MS acknowledge a financial relationship (em- ployees of Depuy Spine)—TJA and HS, consultants for Depuy Spine; MS, speaker for Depuy Spine—which may indirectly relate to the subject of this research. This study was presented at the World Spine II Congress on Spine Care, Chicago, IL, 2003. * Corresponding author. 325 Paramount Drive, Raynham, MA 02767, USA. Tel.: (508) 828-3722; fax: (508) 828-3700. E-mail address: [email protected] (H. Serhan) 1529-9430/04/$ – see front matter 2004 Elsevier Inc. All rights reserved. doi:10.1016/j.spinee.2003.12.004 Introduction Titanium and its alloys are regarded as highly biocompati- ble materials and are becoming increasingly popular in spinal implant applications. The use of titanium alloys for spinal implants has increased because of their advantageous mechanical properties, biocompatibility, corrosion resistance and compatibility with magnetic resonance imaging pro- cedures. Nonetheless, with their introduction came the risk of inadvertent mixing with stainless steel components of 086455-180103

Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

The Spine Journal 4 (2004) 379–387

Is galvanic corrosion between titanium alloy and stainless steel spinalimplants a clinical concern?

Hassan Serhan, PhDa,*, Michael Slivka, MSca, Todd Albert, MDb, S. Daniel Kwak, PhDa

aDePuy Spine, Inc., 325 Paramount Drive, Raynham, MA 02767, USAbThomas Jefferson University Hospital, Philadelphia, PA 19107, USA

Received 12 May 2003; accepted 2 December 2003

Abstract BACKGROUND CONTEXT: Surgeons are hesitant to mix components made of differing metalclasses for fear of galvanic corrosion complications. However, in vitro studies have failed to showa significant potential for galvanic corrosion between titanium and stainless steel, the two primarymetallic alloys used for spinal implants. Galvanic corrosion resulting from metal mixing has notbeen described in the literature for spinal implant systems.PURPOSE: To determine whether galvanic potential significantly affects in vitro corrosion oftitanium and stainless steel spinal implant components during cyclical compression bending.STUDY DESIGN/SETTING: Bilateral spinal implant constructs consisting of pedicle screws,slotted connectors, 6.35-mm diameter rods and a transverse rod connector assembled in polyethylenetest blocks were tested in vitro. Two constructs had stainless steel rods with mixed stainless steel(SS-SS) and titanium (SS-Ti) components, and two constructs had titanium rods with mixed stainlesssteel (Ti-SS) and titanium (Ti-Ti) components.METHODS: Each construct was immersed in phosphate-buffered saline (pH 7.4) at 37 C andtested in cyclic compression bending using a sinusoidal load-controlling function with a peak load of300 N and a frequency of 5 Hz until a level of 5 million cycles was reached. The samples werethen removed and analyzed visually for evidence of corrosion. In addition, scanning electronmicroscopy (SEM) and energy dispersive spectrometry (EDS) were used to evaluate the extent ofcorrosion at the interconnections.RESULTS: None of the constructs failed during testing. Gross observation of the implant componentsafter disassembly revealed that no corrosion had occurred on the surface of the implants that hadnot been in contact with another component. The Ti-Ti interfaces showed some minor signs ofcorrosion only detectable using SEM and EDS. The greatest amount of corrosion occurred at theSS-SS interfaces and was qualitatively less at the SS-Ti and Ti-SS interfaces.CONCLUSIONS: The results from this study indicate that when loaded dynamically in saline,stainless steel implant components have a greater susceptibility to corrosion than titanium. Further-more, the galvanic potential between the dissimilar metals does not cause a discernible effect onthe corrosion of either. Although the mixture of titanium alloy with stainless steel is not advocated, theresults of this study suggest that galvanic corrosion is less pronounced in SS-Ti mixed interfacesthan in all stainless steel constructs. � 2004 Elsevier Inc. All rights reserved.

Keywords: Corrosion; Dynamic testing; Pedicle screw implants

FDA device/drug status: not approved for this indication (Isola/VSPSpine System).

Authors SDK, HS and MS acknowledge a financial relationship (em-ployees of Depuy Spine)—TJA and HS, consultants for Depuy Spine; MS,speaker for Depuy Spine—which may indirectly relate to the subject ofthis research.

This study was presented at the World Spine II Congress on Spine Care,Chicago, IL, 2003.

* Corresponding author. 325 Paramount Drive, Raynham, MA 02767,USA. Tel.: (508) 828-3722; fax: (508) 828-3700.

E-mail address: [email protected] (H. Serhan)

1529-9430/04/$ – see front matter � 2004 Elsevier Inc. All rights reserved.doi:10.1016/j.spinee.2003.12.004

086455-

Introduction

Titanium and its alloys are regarded as highly biocompati-ble materials and are becoming increasingly popular inspinal implant applications. The use of titanium alloys forspinal implants has increased because of their advantageousmechanical properties, biocompatibility, corrosion resistanceand compatibility with magnetic resonance imaging pro-cedures. Nonetheless, with their introduction came the riskof inadvertent mixing with stainless steel components of

180103

Page 2: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

H. Serhan et al. / The Spine Journal 4 (2004) 379–387380

compatible systems. Additionally, surgeons may wish tomix components of different metals in order to use the bestproperties from them (eg, stainless steel sublaminar wireswith titanium rods, which was not tested in this study).

The main concern with mixing titanium and stainless steelis the fear of galvanic corrosion, a process that occurs whenmaterials of different electrochemical potential are placedin close proximity in an electrolytic environment. Galvanicpotential resulting from metal mixing may negatively influ-ence the implants by hastening their corrosion and fretting,resulting in a vicious cycle that could lead to severe compli-cations, such as aseptic loosening of the spinal implant.Corrosion can also reduce the fatigue performance of the im-plants, leading to mechanical failure, as previously report-ed [1]. Furthermore, the release of corrosion byproducts mayelicit anadverse biological reaction in the host tissue. Multipleauthors have reported increased concentrations of local andsystemic trace metals in association with metal implants [2–4]. Finally, late-appearing infection at the spinal instrumenta-tion site is of a major concern and may be exacerbated bycorrosion [5,6]. This type of infection is treated with short-term oral antibiotics, primary skin closure and even deviceremoval. Primarily, the infections affect soft tissues, not thebone, and the highest areas of metal debris concentrationare the interconnection interfaces [5–12].

Although there is no specific histological evidence ofthe slow release of metal species from all metal implants,accelerated corrosion and a tissue response (eg, discolor-ation, foreign body response) related to identifiable corrosionproducts have been demonstrated in the tissues surroundingmultiple-part devices [5–7,12–14]. Evidence of fretting andcrevice corrosion has been reported for stainless steel spinalimplants in retrieval studies [5–11]; particularly, Duboussetet al. [6] attributed the occurrence of late operative siteinfection to fretting corrosion. Even in the absence of infec-tion, corrosion products may play a role in causing localpain and swelling at the spinal implant site [7]. The presenceof particulate corrosion and wear products in the tissue sur-rounding the spinal implant may ultimately result in a cas-cade of events leading to bone loss, pseudarthrosis andpossible implant failure [5,8,14].

The typical spinal stainless steel material contains enoughchromium to give corrosion resistance by passivity. Thistype of stainless steel is vulnerable to pitting and to crevicecorrosion at the interconnections (eg, screw–rod interfaces)under certain circumstances, such as in a highly stressed andoxygen-depleted region [15]. Titanium forms an adherentporous layer (TiO2) and remains passive under physiologicalconditions. Titanium implants remain virtually unchangedin appearance and offer superior corrosion resistance [15].Titanium-based alloys have lower overall corrosion rateswhen compared with stainless steel and cobalt chromiumalloys.

Corrosion of spinal implants may occur as a result ofmechanical or electrochemical means. Mechanical forms ofcorrosion include fretting, which occurs at contact areas

086455

between materials under load subjected to repeated cyclicloading or vibration. Electrochemical means include crevice,pitting and galvanic corrosion. Crevice corrosion is a formof local corrosion occurring in a confined space as a result oflow oxygen concentration, high concentration of electrolytes,low pH or any combination of these effects caused by inhib-ited diffusion of chemicals into and out of the space, suchas in the crevices between a screw and a plate [16]. Pittingoccurs as a result of elemental impurities or corrosion-sus-ceptible alloying elements on the surface of the metal. Last,galvanic corrosion occurs when materials with different elec-trochemical potentials are in close proximity in an electrolytesolution (such as saline) [1]. One material undergoes anoxidation reaction (anode), whereas the other material under-goes a reduction reaction (cathode). Therefore, the objectiveof this study was to investigate the effect of galvanic potentialin mixed-metal spinal implant constructs, specifically tita-nium alloy and stainless steel, subjected to cyclic loadingin an in vitro environment.

Materials and methods

Posterior spinal fixation systems generally consist of sev-eral components that connect together to form a final spinalimplant assembly. When constructed, the fixation systemprovides mechanical stability to the affected vertebrae whilefusion through arthrodesis occurs. To investigate the effectof galvanic corrosion on spinal implants, this study useda commercially available posterior lumbar spinal fixationsystem (ISOLA/VSP Spine System; DePuy Spine, Inc.,Raynham, MA) with interchangeable components availablein both stainless steel (316L) and titanium alloy (Ti-6Al-4V).

The study followed the F1717 test standard for spinal im-plants adopted by the American Society for Testing andMaterials, “Static and Fatigue for Spinal Implants in a Verte-brectomy Construct.” This test standard uses a bilateral con-struct that models a vertebrectomized spinal segment with amissing middle vertebra with the spinal implants resisting allthe applied loads. Each construct consisted of spinal implantsattached to two ultra-high molecular weight polyethylene(UHMWPE) test blocks simulating upper and lower vertebralbodies. The nominal gauge length, the distance between thepedicle screw insertion locations of the top and bottom blocks,was set at 76 mm.

For each construct, two pedicle screws were inserted inthe top test block bilaterally and another two in the bottomblock. Afterward, each pedicle screw was secured to a slottedconnector, using a nut. The two slotted connectors on theleft side, one from the top block and one from the bottom,were attached to a 6.35 mm diameter rod using a setscrew,and the other two slotted connectors on the right side wereattached to another rod, thus creating a bilateral construct.Finally, a single transverse rod connector joined the tworods. The connector consisted of a threaded rod and two setsof securing components, each containing a pair of clasps and

-180103

adsouza4
Typewritten Text
adsouza4
Typewritten Text
Page 3: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

H. Serhan et al. / The Spine Journal 4 (2004) 379–387 381

a single nut for each rod. Tightening the nut firmly securedthe clasps around both sides of the rod (Fig. 1). Followingthe manufacturer’s recommendation, the nuts connecting thepedicle screw to the slotted connector torques were tightenedto 11.3 Nm (100 inch-pounds), the set-screws connecting theslotted connector to the transverse rod to 6.8 Nm (60 inch-pounds), measured using a torque meter.

Because each implant component used in this study isavailable in both stainless steel and titanium alloy, compo-nents of different metals were intentionally mixed to producegalvanic corrosion. A total of four constructs were tested inthis study: two constructs consisted of titanium alloy rodsand titanium pedicle screws, and the other two consisted ofstainless steel rods and stainless steel pedicle screws. Tocreate mixed metal interfaces in each construct, a titaniumalloy slotted connector was used in one end of the rod, and astainless steel slotted connector was used in the oppositeend of the rod. Furthermore, because the transverse rodconnector contained two sets of securing components, atitanium set was used on one rod and a stainless steel seton the other rod. Therefore, titanium rods contacted bothtitanium components (Ti-Ti interface) and stainless steelcomponents (Ti-SS interface) in two constructs (Fig. 2), andstainless steel rods contacted titanium components (SS-Tiinterface) and stainless steel components (SS-SS interface)in another two constructs (Fig. 3).

All constructs were loaded dynamically in compressionthrough a pin in the anterior portion of each test block,creating a flexural bending of the rods. All tests were per-formed using a computer-controlled material testing machine(MTS 858 Mini-Bionix, MTS Systems Corporation, EdenPrairie, MN). Cyclic sinusoidal loads were applied at 5 Hzwith maximum and minimum compression loads of 300 N

Fig. 1. Close-up view of the transverse rod connector used in this study,showing the clamping mechanism of the device around the rod.

086455-1

and 30 N. These loads were chosen to ensure that the con-structs would reach a run-out of 5 million cycles withoutstructural failure, at which time the test was stopped. To sim-ulate the in vivo environmental conditions, all tests wereperformed in a 0.1 M phosphate-buffered saline bath (pH7.4) at a temperature of 37 C.

After testing, the constructs were rinsed thoroughly indistilled water several times and disassembled. Quantitativevisual observations were made of the implant surfaces, withfocus on implant interfaces. After visual observation, thecomponent surfaces were further analyzed through scanningelectron microscopy (SEM, JEOL JSM-6900LV, Tokyo,Japan) and energy dispersive spectrometry (EDS; OxfordInstruments, Concord, MA). All SEM micrographs wereobtained through a secondary electron imaging, and EDSquantitative analysis was performed using ZAF correctionand filtered least squares deconvolution, with copper as thecalibration material. The SEM and EDS analysis primarilyconcentrated on examinationof the rods, because their convexsurfaces could be most readily examined. EDS quantitativeanalysis was performed to obtain representative elementalinformation and was not statistically powered to comparedifferences between groups.

Results

For all constructs, no visual evidence of corrosion wasfound on surfaces not in contact with another surface,as confirmed using SEM and EDS (Fig. 4, A and B and Table1). Upon visual inspection, Ti-Ti interfaces appeared slightlydarker with a clearly outlined contact area. Although corro-sion was not visually evident (Fig. 2), SEM showed evidenceof minor surface cracking and peeling in some regions ofthe interconnection area (Fig. 5). Furthermore, EDS con-firmed the changes in material composition by showingan increase in oxygen content, suggesting that titanium oxidewas formed on the surface (Table 1).

For the SS-SS interface, the evidence of surface corrosionwas clearly visible as a reddish and greenish color, sug-gesting a chemical corrosion process (anodic oxidation) withiron and chromium oxides. The corrosion extended beyondthe boundary of surface contact area (Fig. 3). SEM showedextensive surface cracking and roughness throughout theSS-SS interface (Fig. 4, E), and EDS further confirmed thesurface corrosion by showing an increase in oxygen content.

When mixed-metal interfaces were examined, titaniumalloy surfaces of Ti-SS interfaces showed some visual evi-dence of corrosion with its darker appearance (Fig. 2). SEMagain confirmed the visual inspection, showing surfacecracking and peeling throughout the contact area (Fig. 4,D). Furthermore, EDS demonstrated material transfer fromthe stainless steel components to titanium rods with existenceof chromium, molybdenum and iron (Table 1), althoughEDS could not confirm whether this was the result of trans-ferring of the particles or leaching of the material from the

80103

Page 4: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

Fig. 2. Setup of test construct according to ASTM F1717 and evidence of corrosion at implant component interfaces. Two constructs in this study consistedof titanium alloy rods and pedicle screws with mixed titanium alloy and stainless steel rod–screw connectors and transverse rod connector components, asshown with arrows.

086455-180103

Page 5: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

Fig. 3. Setup of test construct according to ASTM F1717 and evidence of corrosion at implant component interfaces. Two constructs in this study consistedof stainless steel rods and pedicle screws with mixed titanium alloy and stainless steel rod–screw connectors and transverse rod connector components, asshown with arrows.

086455-180103

Page 6: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

H. Serhan et al. / The Spine Journal 4 (2004) 379–387384

Fig. 4. Scanning electron micrographs (all at ×500 magnification) of rod surfaces taken after rods were dynamically tested in vitro in spinal implantconstructs: (A) stainless steel rod surface not in contact with another surface, (B) titanium alloy rod surface not in contact with another surface, (C) stainlesssteel rod surface in contact with a titanium alloy transverse rod connector clamp, (D) titanium alloy rod surface in contact with a stainless steel transverserod connector clamp, (E) stainless steel rod surface in contact with a stainless steel transverse rod connector clamp, and (F) titanium alloy rod surface incontact with a titanium alloy transverse rod connector clamp.

opposing stainless steel surface. The EDS results alsoshowed an increase in oxygen content between the Ti rodwith Ti contact (28.4%) compared with the Ti rod with SScontact (35.4%), although it could not be confirmed whetherthis occurred as a result of transfer of SS corrosion products

086455

to the Ti rod or whether the Ti rod had accelerated corrosionresulting from the galvanic coupling.

Finally, stainless steel surfaces from SS-Ti interfacesvisually showed corrosion (Fig. 3), and the extent of corro-sion was not qualitatively different from that of SS-SS

-180103

Page 7: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

H. Serhan et al. / The Spine Journal 4 (2004) 379–387 385

Table 1Results of the energy dispersive spectrometry quantitative analysis of representative interconnection surfaces (element %)

Element Ti rod No contact Ti rod Ti contact Ti rod SS contact SS rod No contact SS rod Ti contact SS rod SS contact

C 7.10 7.60 8.75 6.02 5.02 4.75O — 28.40 35.41 — 33.00 35.76Na — 1.38 0.60 — 2.28 1.67Al 6.21 4.00 2.49 — 0.33 0.24

Si 1.20 1.07 1.18 0.98 — —P — 1.34 1.68 — 10.70 7.97K — — — — 0.42 0.33Ca — — 0.26 — — —

Ti 81.20 53.48 39.6 — — —V 4.28 2.73 1.93 — — —Cr — — 5.82 16.51 20.48 31.45Mn — — — 1.66 0.55 —

Fe — — 0.61 55.60 19.20 4.34Ni — — — 13.52 3.78 1.18Mo — — 1.68 3.18 2.37 12.31Dy — — — 2.52 1.88 —

Total 100 100 100 100 100 100

surfaces. SEM again showed extensive surface cracking androughness on the stainless steel surface (Fig. 4, C), and EDSconfirmed the extensive corrosion with a substantial increaseof oxygen. However, unlike the opposing titanium alloysurface, no evidence of titanium alloy composition was foundon the stainless steel surface.

Discussion

The results from this study indicate that when loadeddynamically in saline, stainless steel implant componentshave a greater susceptibility to corrosion than titanium.Furthermore, the galvanic potential between the dissimilarmetals does not cause a discernible effect on the corrosionof either. Previous studies have shown that the corrosion

086455-1

resistance of stainless steel is the same when measured forstainless steel alone or in contact with Ti-6Al-4V alloy [17],which agrees with the results of this study.

The greater susceptibility of stainless steel to corrosionoccurs as a result of the mechanical breakdown of thepassive oxide layer at the interface, which is much slowerto reform on stainless steel than titanium. Furthermore, lowoxygen concentration at the interface (crevice conditions)accelerates the corrosion. With the combination of highrest potential for stainless steel and crevice corrosion oc-curring, the effects of galvanic corrosion, if any, wereundetectable.

The lack of corrosion seen on Ti-Ti surfaces was notsurprising because of the high corrosion resistance oftitanium. The excellent corrosion resistance of titanium is

Fig. 5. High-magnification (×2,000) scanning electron microscopy images of titanium alloy rod surfaces: (A) titanium alloy rod surface not in contact withanother surface, and (B) titanium alloy rod surface in contact with a titanium alloy transverse rod connector clamp, showing some evidence of surface cracking.

80103

Page 8: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

H. Serhan et al. / The Spine Journal 4 (2004) 379–387386

attributable to the stable, impervious, fast-forming titaniumoxide layer on the surface. However, under reducing environ-ments, such as acidic environment or low oxygen envi-ronment, the oxide layer does not reform and thus titaniumwill become more susceptible to corrosion. These conditionsare likely to occur in crevices where the corrosion processconsumes the oxygen but oxygen cannot diffuse easily fromthe surrounding media into the crevice. Therefore, the corro-sion of stainless steel in the interconnection, which couldbe considered a crevice environment, could accelerate thecorrosion of titanium. Furthermore, because stainless steelhas twice the elastic modulus of titanium, fretting of titaniumis more likely to occur when in contact with stainless steelthan with another titanium surface. Crevice or fretting corro-sion of titanium may, in turn, reduce the fatigue life of theimplant.

Galvanic effects, however, are a less likely cause of thetitanium corrosion. Titanium is considered a more noble metalthan stainless steel in a saline environment [18] and there-fore acts as the cathode while stainless steel is the anode.Therefore, titanium would be expected to accelerate the cor-rosion of stainless steel in a galvanic couple and not theopposite. However, stainless steels are very close to titaniumin the galvanic series [18] and therefore may not exhibit thedegree of galvanic corrosion as other coupled surgical alloys.In this study, the effect of galvanic potential did not appearto influence the corrosion of either metal.

Clinical situations of mixing metals in spinal surgeryinclude not only compatible interconnections of spinal im-plant systems, but also the use of interbody cages withsupplemental anterior or posterior fixation, which was nottested in this study. Interbody cages are most often titaniumor carbon fiber composite material, whereas anterior andposterior fixation systems are usually titanium or stainlesssteel. Therefore, there is a potential for galvanic corrosionwhen using, for example, a titanium interbody fusion cagewith stainless steel fixation. However, the results from thisstudy and others [17] suggest that the likelihood of adverseclinical effects is very low, particularly because interbodycages typically do not contact or come in close proximity tofixation implants.

Some limitations are important to recognize when inter-preting the results of this study. Although proteins have beensuggested to affect the corrosion of metals in saline [19],this study used a physiologically relevant electrolyte solutionof phosphate-buffered saline (pH 7.4, 37 C) without proteins;however, addition of proteins to the media would not likelychange the relative differences between metal combinations.This study also used UHMWPE blocks to simulate vertebrae.Because UHMWPE is a relatively inert material, the blocksor any particles it may have generated would have a negligi-ble effect of metal corrosion. Furthermore, the test methodused an in vitro–accelerated simulation of 5 million cyclesat 5 Hz, which takes approximately 12 days to complete.In actual use, 5 million loading cycles of average daily livingactivity occurs over approximately 2 years (ASTM F1717).

086455

This difference could affect the test results in several ways.First, because mechanical forces at the implant interconnec-tions cause physical breakdown of the passive oxide layer,loading the implants at 5 Hz could prevent reformation ofthe oxide layer, particularly for stainless steel. Furthermore,whereas galvanic corrosion is time dependent, fretting corro-sion is time independent; hence, the latter may be favoredin this dynamic test method.

Conclusions

The results from this study suggest that galvanic corro-sion does not play a significant role in mixed titanium alloy-stainless steel spinal implant constructs.

Using the dynamic, in vitro test method of this study,crevice corrosion and fretting were dominant over galvanicpotential. Based on this finding, coupling titanium compo-nents with stainless steel systems may not significantly accel-erate corrosion of stainless steel in load-bearing implants.However, coupling stainless steel components with titaniumsystems may induce fretting and crevice corrosion of tita-nium that would not otherwise occur. Further studies areneeded before it can be recommended clinically to use suchmaterials as titanium and stainless steel in direct contact forspinal implants.

References

[1] Jacobs J, Gilbert J, Urban R. Corrosion of metal orthopaedic im-plants. J Bone Joint Surg Am 1998;80(2):268–82.

[2] Pazzaglia U, Ceciliani L, Wilkinson M, Dell OC. Involvement ofmetal particles in loosening of metal-plastic total hip prostheses. ArchOrthop Trauma Surg 1985;104(3):164–74.

[3] Jacobs J, Skipor A, Patterson L, et al. Metal release in patients whohave had a primary total hip arthroplasty. A prospective, controlled,longitudinal study. J Bone Joint Surg Am 1998;80(10):1447–58.

[4] Dorr L, Bloebaum R, Emmanual J, Meldrum R. Histologic, biochemi-cal, and ion analysis of tissue and fluids retrieved during total hiparthroplasty. Clin Orthop 1990;261:82–95.

[5] Clark C, Shufflebarger H. Late-developing infection in instrumentedidiopathic scoliosis. Spine 1999;24(18):1909–12.

[6] Dubousset J, Shufflebarger H, Wenger D. Late “infection” with CDinstrumentation. Orthop Trans 1994;18:121.

[7] Vieweg U, van Roost D, Wolf H, Schyma C, Schramm J. Corrosionon an internal spinal fixator system. Spine 1999;24(10):946–51.

[8] Prikryl M, Srivastava S, Viviani G, Ives M, Purdy G. Role of corrosionin Harrington and Luque rods failure. Biomaterials 1989;10(2):109–17.

[9] Bidez M, Lucas L, Lemons J, Ward J, Nasca R. Biodegradation pheno-mena observed in vivo and in vitro spinal instrumentation systems.Spine 1987;12(6):605–8.

[10] Aulisa L, di Benedetto A, Vinciguerra A, Lorini G, Tranquilli LP. Cor-rosion of the Harrington’s instrumentation and biological behaviour ofthe rod-human spine system. Biomaterials 1982;3(4):246–8.

[11] Sim E, Stergar P. The fixateur interne for stabilising fractures of thethoracolumbar and lumbar spine. Int Orthop 1992;16(4):322–9.

[12] Sim E. [Spinal internal fixator—analysis of problems in 28 cases.]Wien Klin Wochenschr 1992;104(12):349–55.

[13] Brayda-Bruno M, Fini M, Pierini G, Giavaresi G, Rocca M, GiardinoR. Evaluation of systemic metal diffusion after spinal pedicular

-180103

Page 9: Is galvanic corrosion between titanium alloy and stainless ... CA… · pain and swelling at the spinal implant site [7]. The presence of particulate corrosion and wear products in

H. Serhan et al. / The Spine Journal 4 (2004) 379–387 387

fixation with titanium alloy and stainless steel system: a 36-monthexperimental study in sheep. Int J Artif Organs 2001;24(1):41–9.

[14] Urban R, Jacobs J, Gilbert J, Galante J. Migration of corrosion productsfrom modular hip prostheses. Particle microanalysis and histopatho-logical findings. J Bone Joint Surg Am 1994;76(9):1345–59.

[15] Atkinson J, Jobbins B. Properties of engineering materials for usein the body. In: Dowson D, Wright V, editors. Introduction to thebiomechanics of joints and joint replacement. London: MechanicalEngineering Publications Ltd, 1981.

SeventyYears Agoin Spine

The August 2, 1934 issue of the NewEngland Journal of Medicine contained the article thatushered into spine care the “era of the disc.” WilliamJason Mixter, then Chief of Neurosurgery at Massachu-setts General Hospital and Professor of Neurosurgery atHarvard Medical School, and Joseph Seaton Barr, thenOrthopaedic Surgeon to Outpatients at the same institu-tion (later Chief of the Orthopaedic Service and Professorof Orthopaedics at Harvard), had read the paper the

086455-18

[16] Bowden F, Williamson J, Laing P. The significance of metal transferin orthopaedic surgery. J Bone Joint Surg Br 1955;37B:676–90.

[17] Griffin C, Buchanan R, Lemons J. In vitro electrochemical corrosionstudy of coupled surgical implant materials. J Biomed Mater Res1983;17(3):489–500.

[18] Donachie M. Corrosion resistance. Titanium: a technical guide. MetalsPark, OH: ASM International, 1989:207–16.

[19] Brown S, Merritt K. Electrochemical corrosion in saline and serum.J Biomed Mater Res 1980;14(2):173–5.

previous fall to the Annual Meeting of the New EnglandSurgical Society. The paper cited the work of Schmorlin describing Knorpel-knochen, prolapsed discs, in patho-logic specimens and the several previous clinical observa-tions of disc herniations and often-made erroneouslabelling of them as enchondromas or chondromas. Theirown work described 19 cases of ruptured discs, includingtwo detailed clinical descriptions, one cervical and onelumbar. Among their conclusions was, “ … the treatmentof this disease is surgical … ”. Dr. John Homans in theDiscussion commented, “ … I should like to ask … whyoperation should be of any value whatever. I can’t getthat through my head.”

Reference

[1] Mixter WJ, Barr JS. Rupture of the intervertebral disc with involve-ment of the spinal canal. N Engl J Med 1934;211:210–5.

0103