27
IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles static electric Dipole: Far away: E ~ r -3 along all directions adiating (Hertz's) Dipole: (VERY DIFFERENT!) far away: n line: zero field (n x p = 0) rthogonal: E ~ r -1 r = = distance from middle to position n = direction unit vector p = dipole vector

IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

Embed Size (px)

Citation preview

Page 1: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena1

7. Fluorescence microscopy

7.3 FRET microscopy - The different Dipoles

1. static electric Dipole:

Far away: E ~ r-3 along all directions

2. radiating (Hertz's) Dipole: (VERY DIFFERENT!)

far away:

in line: zero field (n x p = 0)orthogonal: E ~ r-1

r = = distance from middle to positionn = direction unit vectorp = dipole vector

Page 2: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena2

7. Fluorescence microscopy

7.3 FRET microscopy - The different Dipoles

3. transition Dipole moment – an analogy :

-

+

"molecular"

conductor

E = electric Field

Induced Transition Dipole

Transition Dipole Moment : a vector along who's direction the dipole will be induced

Page 3: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena3

Resolution of a light microscope is limited to several hundred nanometers

(< organelles) FRET allows detection of molecule-molecule interactions on a nanometer scale by

means of a light microscope

Decrease of donor-emission

Increase of acceptor emission

Reduction of donor fluorescence life-time

Energy transfer (FRET-efficiency) depends strongly on donor-acceptor distance

R0 = Förster-radius (distance for which energy transfer is half maximal)

7. Fluorescence microscopy

7.3 FRET microscopy

sensitizedemission

Page 4: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena4

FRET ratio imaging = acceptor emission at donor excitation (sensitized emission SAkzeptor) divided by donor

emission at donor excitation (SDonor)

Advantages: Since both donor decrease as well as acceptor increase contribute to

the signal the signal-to-noise ratio is better than for solely recording the acceptor

fluorescence

SAkzeptorSDonor

7. Fluorescence microscopy

7.3 FRET microscopy

Page 5: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena5

FRET ratio imaging – problems:

Correction for direct excitation of the acceptor when exciting donor (control measurement with YFP only) = correction factor rDE

Excitation wavelength

Correction for bleedthrough : Portion of CFP in yellow channel for blue excitation in absence of FRET (acceptor) = bleedthrough of CFP in YFP-channel (rBT,CY)

or bleedthrough of YFP in CFP-channel (rBT,YC)

FR

ET

-de

tect

ion

ch

an

ne

l

7. Fluorescence microscopy

7.3 FRET microscopy

Page 6: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena6

FRET ratio imaging – 3-filter-set:

1. Donor excitation and emission(ICFP,430)

2. Acceptor excitation and emission(IYFP,514)

3. Donor excitation and acceptor emission(IYFP,430)

7. Fluorescence microscopy

7.3 FRET microscopy

Page 7: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena7

FRET ratio imaging – 3-filter-set:

Model of FRET-detection of Src-

Csk protein interaction

(Src = protein tyrosine kinase

Csk = C-terminal Src kinase) Important signal transduction step

during blood coagulation

7. Fluorescence microscopy

7.3 FRET microscopy

No FRET

Page 8: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena8

FRET ratio imaging – 3-filter-set:Visualization of Src-Csk-interaction during aIIbß3-induced fibrinogen adhesion in a

thrombocyte model cell line (A5-CHO) by means of FRET

7. Fluorescence microscopy

7.3 FRET microscopy

superposition

Page 9: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena9

FRET ratio imaging – 3-filter-set:FRET for displaying Ca2+ in living cells via Yellow-Cameleon-2 (YC2) sensor

FRET-ratio image of HeLa-cells, expressing the YC2-sensor before and after adding ionomycin

FRET response of HEK/293 cells expressing YC2-seonsor after adding 1nM ionomycin and additional extracellular Ca21 (30 mM)

7. Fluorescence microscopy

7.3 FRET microscopy

Page 10: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

Acceptor Bleaching

Free Donor +1 Donor (Pairs)

AcceptorRemoved

Free Donor +Donor (Pairs)

Donor excitation,Donor emission detection

1 can be calibrated

De-Quenching

Page 11: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

Principle ofAcceptor-bleaching-FRET microscopy

Donor fluorescenceshould increase afteracceptor bleaching

PrincipleDonor fluorescence should

increase (dequenching) after“removal” of the acceptor

Acceptor depletion FRET

Page 12: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

CFPExc 457Em 470-500

‘Sens-YFP’Exc 457Em 535-570

YFPExc 514Em 535-570

Combined

100 µm

Data of Dorus Gadella

Donor Emission channel

Acceptor Emission channel

Page 13: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

CFPExc 457Em 470-500

‘Sens-YFP’Exc 457Em 535-570

YFPExc 514Em 535-570

Combined

100 µm

Data of Dorus Gadella

BUT is it:High concentration, low efficiency?

Low concentration, high efficiency?

Donor Emission channel

Acceptor Emission channel

Page 14: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena14

FRET fluorescence life-time microscopy:

In case of FRET the donor fluorescence life-time is reduced. Determination of this

donor life time reduction yields a quantitative FRET measurement which is

independent of dye concentration or spectral contamination (crosstalk, bleedthrough).

Dimerization of C/EBP® – proteins in GHFT1-5 cell nuclei(donor/acceptor CFP/YFP-C/EBP®)

7. Fluorescence microscopy

7.3 FRET microscopy

Page 15: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena15

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Laser pulse

Longer fluorescence life-time

Shorter fluorescence life-time

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements

Sample is excited by a short laser pulse

Sample molecules relax individually according to the transition probability of the different relaxation pathways to the ground state

Fluorescence intensity exhibits mono-, multi or non-exponential decay depending on nature and number of fluorescence contributions.

Page 16: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena16

Time-resolved measurements

Intensity integrating measurementsThe determination of the fluorescence decay time ¿ or times ¿i and relative amplitudes ®i in

case of multiple contributions is possible by recording the fluorescence signal for several measurement points after the excitation pulse. For a mono-exponential decay behavior or to determine the average decay time ¿ two sampling points are sufficient

For two times t1 and t2 after the excitation pulse the detector signal is integrated for a sampling window ¢ T. The ratio of the measurement signals D1 und D2 can be used to calculate the decay-time ¿ or the average decay-time ¿ :

Methods of time-resolved fluorescence diagnostics

7. Fluorescence microscopy

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

Page 17: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena17

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements Gated fluorescence detection

Gated optical image intensifiers (GOI) are capable of taking pictures with high (sub-nanosecond) time resolution i.e. camera with ultrafast shutter (gate < 100 ps) which can be opened and closed for different delay-times after the sample has been excited with an ultrashort laser-pulse. By collecting a series of time-scanned fluorescence intensity images for different delay-times after excitation the fluorescence decay profile for every pixel in the field of view can be accessed and displayed as false color plot = fluorescence life-time image

7. Fluorescence microscopy

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

Page 18: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena18

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements Gated fluorescence detection

Tissue section of a rat ear: (a) Brightfield microscopy image stained with orcein(b) Fluorescence intensity- and (c) FLIM images of an unstained parallel sample (tissue autofluorescence) (excitation 410 nm; FLIM false color plot from 200 ps (blue) to 1800 ps (red)

(a) (b) (c)

(Top) FLIM image of an unstained human pancreas section (tissue autofluorescence) with an endocrine tumor (below) Brightfield image of the same section after conventional histopathological staining

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Page 19: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena19

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements

TCSPC = time-correlated single photon countingIn case of intensive excitation light many electrons of the dye are getting excited for every laser pulse i.e. the average life-time can be deduced from the fluorescence decay-time after every pulse (multiple photon emission).

A common FLIM method is the measurement of the life-time for single fluorescence photons. In doing so the dye is excited by light pulses of extremely low intensity in a way that at most one electron per pulse gets excited. The individual life-time of every photon is measured and the average life-time is determined staistically.

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Page 20: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena20

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements

TCSPC = time-correlated single photon counting Detection of single photons of a periodic light

signal

Light intensity is so weak, that the probability to detect a photon within one period is very small.

Periods with more than one photon are extremely rare

For every detected photon its delay time with respect to the excitation pulse is determined

A delay distribution builds up over many pulse

Time resolution up to 25 ps

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Page 21: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena21

Probe Stop watch = TAC: Time-to-Amplitude Converter

converts time between a start and a stop pulse by charging

a capacitor with constant current

Start can be reference (from laser) and photon is stop

-> Problem is loss of much time (due to reset time)

-> Reverse counting (start = photon, stop=next laser pulse)

Histogram of arrival times after excitation

-> fluorescence life-time.

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements

TCSPC = time-correlated single photon counting

7. Fluorescence microscopy

Page 22: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena22

Fluorescence intensity image of a vacuole which is labeled byfluorescent phospholipids

FLIM image and corresponding distribution of life-times. Long life-times (red) are found in the cell membrane while the cytoplasma exhibits shorter life-times pointing towards a less ordered environment.

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

Methods of time-resolved fluorescence diagnostics

Time-resolved measurements

TCSPC = time-correlated single photon counting

7. Fluorescence microscopy

Page 23: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena23

Methods of time-resolved fluorescence diagnostics

Steady state measurements Phase modulation

Intensity of a continuous wave (CW) source is modulated at high frequency by

a standing wave acousto-optic modulator ( 50 MHz) which will

modulate the excitation intensity at double frequency.

Detected fluorescence is modulated at the same frequency.

The observed phase shift with respect to the excitation and

the modulation depth M (ratio of Ac signal to DC signal)

depends on the fluorescence life-time of the excited fluorophores.

Fluorescence lifetimes phase and phase can be calculated and should be

identical for single exponential decays.

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Page 24: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena24

Methods of time-resolved fluorescence diagnostics

Steady state measurements Phase modulation

Measurement values:Demodulation (modulation depth) M Phase shift

Modulation of excitation light with , which is characterized by modulation depth ME = a/d and E :

leads to an accordingly modulated fluorescence signal F(t) with demodulation MF = A/D und phase F

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

Time

Inte

nsity Excitation

light

Fluorescence

7. Fluorescence microscopy

Page 25: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena25

Methods of time-resolved fluorescence diagnostics

Steady state measurements Phase modulation

Rate equation of change of number of excited molecules

F(t) ~ N(t) Relationship between fluorescence life-time and

fluorescence emission behavior upon intensity modulated excitation light Relationship between measurement parameters:

M = MF / ME as well as = E - E and life-time :

Absorption rate:

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Page 26: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena26

Methods of time-resolved fluorescence diagnostics

Steady state measurements Phase modulation

Continuous intensity modulated excitation of fluorescence transforms the

determination of fluorescence decay-times to measurements of phase shifts and

demodulation of the fluorescence signal

Demodulation M and phase shift of the fluorescence depend on the fluorescence life-time as well as on the modulation frequency = 2 of the excitation light. The simulation shows the dependency of M and for a decay time of = 4 ns and a frequency of = 40 MHz

Choosing frequency at 1

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy

Page 27: IPC Friedrich-Schiller-Universität Jena 1 7. Fluorescence microscopy 7.3 FRET microscopy - The different Dipoles 1. static electric Dipole: Far away: E

IPC Friedrich-Schiller-Universität Jena27

Methods of time-resolved fluorescence diagnostics

Steady state measurements Phase modulation

Frozen section of portio biopsies in the spectral region of Em>500 nm (exc = 457 nm; = 40 MHz)Top right: Fluorescence intensity Bottom right: corresponding HE stain image.

7.4 Fluorescence Life-Time Imaging Microscopy (FLIM)

7. Fluorescence microscopy