Investigation of the properties of diamond radiation detectors

K. Afanaciev, I. Emeliantchik, A. Ignatenko,E. Kouznetsova, W. Lohmann

FCAL Workshop, TelAviv, 2005

Page 2: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

Possible ApplicationsPossible Applications

ILC Detector forward calorimeter

Sampling diamond-tungsten calorimeter

Expected dose from beamstrahlungis about 10 MGy per year

Page 3: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

Diamond as detectorDiamond as detector

Diamond have very high radiation hardness(doses up to 10 MGy)

This makes diamond very promising material for the next generation colliders

It also have unique properties, such as low conductivity, high thermoconductivity, wide bandgap, chemical inertness

Page 4: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

Detection mechanismDetection mechanismDiamond works as a solid state ionization chamber

Charge carriers, generated by ionizing particle move in an external electric field and induce charge on the surface electrodes

This charge is detected by a charge-sensitive preamplifier

Page 5: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

The sapmlesThe sapmlesChemical vapor deposited (CVD) polycrystalline samples:

A large selection of samples from Fraunhofer IAF, Freiburg

12x12 mm plates with thickness 200 - 700 m

Metallization: 10 nm Ti + 400nm Au

1 pad or 4 pad segmentation

2 samples from Element Six (former De Beers Industrial Diamonds)

HPHT synthetic monocrystalline samples:From ADAMAS BSU

Typical size 4x4 mm, thickness 300 - 600 m Metallization: Fe-Ni with B ion implantation

1 sample of natural monocrystalline diamond

Page 6: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

Charge collection distanceCharge collection distance

Performance of diamond as particle detector is usually characterized by charge collection efficency or distance (CCD)

= Qmeas/Qind

CCD = *d

Efficency is given by

where Qmeas is measured charge and Qind - total charge generated

Charge collection distance equals

where d is the thickness of the sample

CCD could be considered as separation between charge carrier pair before trapping

Page 7: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

CCD MeasurementCCD Measurement

To measure CCD we need to know the exact amount of charge carriers generated

But we know that a MIPgenerates about 36 e--h pairsper 1 m pathlength in diamond

&Gate

PA

discr

discr

delay

ADC

Sr90

PM1

PM2

diamond

Scint.

In this case CCD = Qmeas /36

Scintillator allows triggeringonly from particles which couldbe considered MIPS

Page 8: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

CCD MeasurementCCD Measurement

Pedestal (random trigger) Signal (MIP trigger)

Signal charge value could be calculated from this pair of spectra by subtracting pedestal and multiplying by charge/ADC channelvalue from calibration.

pedestal

Page 9: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

CCD vs HV and timeCCD vs HV and time

Charge collection distance depends on bias voltageapplied to the sample

Saturation is observed atfield strength of about 1V/m

There is also dependenceof CCD on time with HV on

This is probably due to some kind of space-charge effectfrom trapped charge carriers

Page 10: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

IV MeasurementIV Measurement

Most of the samples show ohmicbehavior and very high resistancein the order of 1013-1014 ohm

Settling time of the current is in theorder of 100 s. Hysteresis is probablydue to polarisation.

Page 11: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

CCD vs DoseCCD vs Dose

Diamond is stable under irradiation with doses up to 10 MGy

But in the process of irradiation detector properties could change

Most of the samples show similar behaviorCDD is rising with the dose by approx. 10-20%and then stabilises

But some sapmles show complete degradation of signal and sharp riseof the current under irradiation

Page 12: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

LinearityLinearity• A linear behavior over high dynamic range is required for the

calorimetry application.• We have checked this at a beamtest at CERN

Setup consists of a boxwith diamond and preamplifier andscintillator trigger box

Incident hadronic beam with approx. 4 GeV energyand intensityfrom 105 - 106 / sup to 105 - 107 / 10ns

Page 13: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

LinearityLinearity

Signal at high intensity clearly visible w/o preamplifier

Good linearity for two samples over the dynamic range of 102

Page 14: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

TrappingTrapping• Trapping of the charge carriers is the main cause of low

CCD.• Traps could be impurities, defects and other

irregularities of crystalline structure.• Filling of traps produces space charge in the crystal and

could give rise to polarisation effects• If we have several trapping levels with different

activation energy, density and cross section, the dynamics of trapping will be really weird.

• On the other hand filling of deep traps leads to “pumping effect”

Page 15: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

TSC measurementsTSC measurements• TSC means thermally stimulated currents• The idea is that if we have charge carrier trapping in the

diamond at room temperature, then heating the sample up could free trapped charge carriers and produce detectable current. The current peak position vs temperature gives the idea about energy levels of traps.

Setup consists of heating plateelectronic thermometer with thermopair, HV supply and picoammeter

Page 16: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

TSC measurementsTSC measurements• TSC data for different diamonds

TSC “resets” sample, but the shape is similar for different samples

Peak position correspondsto energy of about 0,8-1 eV

Page 17: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

PL SpectraPL Spectra• Photoluminescence spectra of different samples

There seems to be a correlation between detector properties and PL

300 400 500 600 700 800 900-2000

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

20000

22000

24000 ASC-D-00013510K, HeCd-laser

Inte

nsity

[a.u

.]Wavelength [nm]

Page 18: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

Raman SpectraRaman Spectra

0 1000 2000 3000 4000-0.002

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

ASC-D-000135-2.0

Raman-Güte : 1010Halbwertsbreite : 3.5435

Wavenumber [cm-1]

0 1000 2000 3000 4000

0.000

0.002

0.004

0.006

0.008

0.010

DESY#1_n

Raman-Güte : 230.1Halbwertsbreite : 5.4658

Wavenumber [cm-1]

Page 19: Investigation of the properties of diamond radiation detectors K. Afanaciev, I. Emeliantchik, A. Ignatenko, E. Kouznetsova, W. Lohmann FCAL Workshop, TelAviv,

ConclusionConclusion

•Need to establish close contact with manufacturer, find a method to characterize samples (TL, raman?) and get interpretetion from the material scientists•Further study of the influence of metallization and surface treatment is needed•A study of radiation hardness with large doses is planned

Future work

Diamond sensors work and we have results both from CVD and synthetic monocrystalline diamonds.The sensors have good radiation hardness and linear response.But the CCD vs time behavior and dependence of CCD on dose complicates use of diamond sensors in calorimetry.In the present state diamond sensors already could be used for beam monitoring.