Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
INVASIVE MONITORING
AIMS OF THE SESSION
• Enable you to care for both CVP and arterial lines
safely
• Be able to assist with insertion
• Be able to safely remove a CVP
CVP’S
• A central line is an intravenous line where the tip sits
just above the right atrium in either the Superior
Vena Cava or Inferior Vena Cava.
• Purpose: Reflects the volume of fluid returning to
the right side of the heart (Right atrial pressure) and
the ability of the heart to pump the blood in the
arterial system
CVP
• Reasons to have:
• Measurement of central venous pressure (CVP)
• Administration of drugs (inc those that cannot be
given peripherally- inotropes, amiodarone)
• Parenteral Feeding (TPN)
• Obtain blood tests
• Venous access problems/ long term IV access
INSERTION
• CVP pack
• Chlorhex 2% solution or snap stick
• US machine
• Sterile gloves, gown, mask
• Local anaesthetic
• Transducer set run through from 500ml 0.9% N saline
bag
• Pressure bag set to 300mmHg
• Transducer cable/ plate/ appropriate monitor
USES
• Estimate circulating volume
• Guide fluid administration
• Assist in assessment of cardiac function and
vascular tone
• Aid assessment of treatment response
• Note:
• Trend more useful than single reading
• Should be considered alongside BP, UO, patient
assessment
TAKING MEASUREMENTS/ WAVEFORMS
• Explain to woman what doing
• Flush line to ensure patency
• Position to supine if possible, align transducer to fifth intercostal space (mid axilla)
• Check the CVP trace
• Document reading and refer on as appropriate
• Increase in readings:
• Hypervolemia
• Forced exhalation
• Tension
Pneumothorax
• Heart Failure
• Pleural effusion
• Decreased cardiac
output
• Cardiac tamponade
• Decrease in readings:
• Hypovolemic
• Deep inhalation
CARE OF THE PERSON WITH A CVP
• Close monitoring for signs
of complications
• Documents any
interventions, changes in
site/ length at site
• Renew dressing as per
local guidance
• If not in use flush
regularly/ remove
• Transducer set to be
changed every 48 hours
• Ensure regaularly all ports
secure
• Removal: • As per local guidance • Check clotting prior to removal • Explain procedure • Detach any lines
• Aseptic technique • Position supine with slight head
down tilt • Remove Stitches • Slowly remove catheter
• Apply pressure (min 5 mins or until bleeding stops)
• Dress with Gauze and clear dressing
• Check complete (inc tip)- send for MC&S of infection screen needed
• Document removal
ARTERIAL LINES
• Advantages:
• Beat to beat measurement of BP
• Frequent arterial blood gases
• Frequent blood sampling
• More accurate if arrhythmias or hypotension
• Useful where NIBP is difficult e.g. obesity
• Information about cardiovascular status
HOW TO SET UP ARTERIAL LINES
• Intra-arterial cannulae
• Fluid filled tubing: • NaCl 0.9%
• No bubbles in system.
• No 3 way taps.
• Pressurised to 300mmHg
• Transducer: right atrium.
• Monitor: zeroing.
• Common sites
• Radial artery
• Brachial artery
• Dorsal pedis artery
• Femoral artery
• Potential complications
• Bleeding if tap left
open
• Injection of drugs into
it
• Incorrect siting
• Tissuing
• Infection
• Arterial damage
• Haematoma
• Air embolism
CARE OF WOMEN WITH ARTERIAL LINES
• Observe limb for perfusion
• Observe cannula site for infection
• Take blood samples using sterile technique
• Ensure all ports secure with an intervention
• Change fluid bag every 48hours
• Change giving set every 5 days
• Change dressing as per local guidance
• Ensure no blood/air bubbles in circuit
• 1:2 ratio to look after arterial line.
ZEROING
• Every four hours both the CVP and Arterial line should be zeroed to calibrate them with zero pressure
• Position patient supine
• Flush the system using the pull flush on the transducer
• Level the transducer (align with mid axilla fifth intercostal space)
• Turn the tap on the port closest to the transducer so it is OFF to the patient
• Remove the cap so it is ‘open to air’
• Press zero on the monitor
• Ensure zero appears, then replace the cap and re-open to patient
• Record time of zeroing
RECORDINGS
• Explain what going to do
• Position as we did with CVP reading
• Check the flush system is pressurised to 300mmHg ( inflate bag
over the 500ml bag of saline, ensuring pressure in green zone) and flush the line
• Zero the transducer (on the machine)
• Check the arterial trace
• Document the reading, refer as necessary
ARTERIAL LINE TRACE
Normal Overdamped
Underdamped
Dicrotic notch
TROUBLESHOOTING
Difficulty Zeroing • Check all equipment and connections
• Ensure all roller clamps open • Check system for blood clots/ air bubbles • Check flush bag volume and pressure • Replace transducer, cable, monitor, replace
Unable to aspirate • Check lines for kinks • Apply traction to cannula • Gently try to flush (anaesthetist) • Replace line
Falsely high readings • Incorrect placement of transducer (below level of heart) • Calibration issue
• Under damped trace • Swapped invasive pressure cables
Falsely low readings • Incorrect placement (above level of heart) • Kinked cannula • Over damped trace
• Swapped cables
Dampened Trace • Check position of transducer • Re-Zero • Remove kink • Remove all bubbles/ clots
REMOVAL
• Check clotting prior to removal
• Ensure peripheral access patent
• Aseptic procedure
• Remove dressing
• Slowly remove and apply pressure (for a minimum
of five minutes or until bleeding stops)
• Dress with gauze and clear dressing
• Check catheter tip complete and skin for signs or
pressure/ infection
• Document removal
QUESTIONS?
TAKE HOME POINTS
• Never use fluid other than Normal saline 0.9%
• NEVER inject into arterial line
• It is only used for BP measurement and blood sampling.
• If problems ask for anaesthetist/ ODP to review