Quality DeclarationCongratulations with your new hoist Invacare® RobinTM from Invacare® EC-Høng A/S.
Your new hoist is -marked in accordance with directive 93/42/EEC concerning medical devices.
Invacare® RobinTM is developed and constructed with consideration for the user and others handling or assisting with lifting. Furthermore, the hoist is developed in accordance with the European Standard EN 10535.
Invacare® RobinTM is supervised and quality controlled throughout the entire production process, and the finished hoist is inspected by our finished goods control. Identification label and QA-mark are located on the hoist confirming that the finished goods control has approved the hoist.
Please read the entire user’s manual before using the hoist.
Invacare®EC-Høng A/S is certified according to ISO 9001 and ISO 13485.
For and on behalf of Invacare®EC-Høng A/S:
R & D Manager
Brian Mundeling
DK KVALITETSDEKLARATIONTillykke med Deres nye personløfter Invacare® RobinTM fra Invacare® EC-Høng A/S.Deres nye personløfter er CE - mærket og lever op til alle krav i henhold til direktiv 93/42/EØF om medicinske anordninger.Personløfteren er udviklet og konstrueret under størst mulig hensyntagen til brugeren samt alle andre, der enten håndterer personløfteren eller hjælper til ved løfteren.Invacare® RobinTM er udviklet under hensyntagen til de sikkerhedsmæssige krav i den Europæiske Standard EN 10535. Hver personløfter har gennem hele produktionsforløbet været overvåget og kontrolleret, og den færdige personløfter er blevet inspiceret af vores færdigvarekontrol.Typeskilt og QA-mærke påsættes personløfteren som dokumentation for, at færdigvare kontrollen har godkendt personløfter. Før ibrugtagning af Deres personløfter skal De gennemlæse brugermanualen grundigt.Invacare® EC-Høng A/S er certificeret i henhold til ISO 9001 og ISO 13485.
DE QUALITETSDEKLARATIONHerzlichen Glückwunsch zu Ihrem neuen Deckenlifter Invacare® RobinTM von Invacare®EC-Høng A/S.Ihr neuer Deckenlifter ist gemäß der Richtlinie 93/42/EWG für Medizinprodukte CE-gekennzeichnet.Der Invacare® RobinTM wurde unter Berücksichtigung der Bedürfnisse von Benutzer und Pflegepersonal entwickelt und in Übereinstimmung mit der Europäischen Norm EN 10535 konzipiert und hergestellt. Während des gesamten Herstellungsprozesses unterliegt der Invacare® RobinTM einer ständigen Qualitätskontrolle und wird im Anschluss nochmals durch unsere Endkontrolle geprüft. Das Typenschild und QA-Kennzeichen sind am Deckenlifter angebracht, um die Abnahme durch unsere Qualitätskontrolle zu bestätigen.Bitte lesen die gesamte Bedienungsanleitung, bevor Sie den Deckenlifter in Gebrauch nehmen.Invacare® EC-Høng A/S ist zertifiziert nach ISO 9001 und ISO 13485.
NL KWALITEITSGARANTIEGefeliciteerd met de aanschaf van uw nieuwe plafondlift RobinTM van Invacare® EC- Høng A/S.Invacare® RobinTM is CE - gecertificeerd en goedgekeurd conform richtlijn 93/42/EEC betreffende medische hulpmiddelen.Invacare® RobinTM is ontwikkeld en geconstrueerd met inachtneming van de behoeften van de gebruiker en verzorgers.Invacare® RobinTM is ontwikkeld in overeenstemming met de Europese Standaard EN 10535.Gedurende het gehele productieproces is de RobinTM plafondlift gecontroleerd op kwaliteitsaspecten en de complete plafondlift is na productie zorgvuldig gecontroleerd. Het productlabel en de QA-markering zijn op de plafondlift bevestigd om aan te tonen dat de plafondlift is gecontroleerd en goed is bevonden door de afdeling productcontrole. Wij verzoeken u vriendelijk de gehele gebruiksaanwijzing te lezen voordat u de plafondlift in gebruik neemt.Invacare® EC-Høng A/S is ISO 9001 en ISO 13485 gecertificeerd.
IT DICHIARAZIONE DI QUALITÀComplimenti per aver scelto il sollevatore da soffito Invacare® RobinTM prodotto da Invacare® EC-Høng A/S.Il vostro nuovo il sollevatore da soffito è marcato CE - ai sensi della Direttiva 93/42/EEC relativa ai dispositivi medici.Il il sollevatore da soffito è stato progettato e costruito con un occhio di riguardo per gli utilizzatori e per i loro assistenti. Invacare® RobinTM è stato realizzato nel rispetto della Normativa Europea EN 10535.Il sollevatore da soffito è stato oggetto di accurate verifiche qualitative durante l’intero processo produttivo; una volta completato è stato controllato dal nostro servizio prodotti finiti.Un’apposita targhetta e la marcatura QA sono state apposte sul il sollevatore da soffito a conferma dell’avvenuta verifica e accettazione del prodotto da parte del servizio qualità. Prima di utilizzare il il sollevatore da soffito vi invitiamo a leggere integralmente il manuale d’uso.Invacare® EC-Høng A/S è un’azienda certificata ai sensi della Normativa ISO 9001 e ISO 13485.
ES DECLARACIÓN DE CALIDADEnhorabuena por su nueva elevador de techo Invacare® RobinTM de Invacare® EC-Høng A/S.Su nueva elevador de techo cuenta con el marcaje CE - de acuerdo a la directiva 93/42/EOF que hace referencia a los aparatos médicos.Invacare® RobinTM ha sido diseñada y fabricada teniendo en cuenta a los usuarios y las personas que los asisten.La elevador de techo Invacare® RobinTM ha sido diseñada de acuerdo a la normativa Europea EN 10535.El proceso de producción del modelo Invacare® RobinTM ha sido supervisado en su totalidad y su calidad inspeccionada, por nuestro control de producto acabado.La elevador de techo cuenta con una placa y marca CE que confirma que la misma ha sido inspeccionada por el control de producto acabado. Por favor lea el manual antes de utilizar la elevador de techo.Invacare EC-Hong A/S es una empresa certificada ISO 9001 e ISO 13485
FR DÉCLARATION DE QUALITÉFélicitations! Vous avez choisi votre nouveau lève-personnes Invacare® RobinTM de Invacare® EC-Høng A/S.Votre nouveau lève-personnes est marqué CE conformément à la directive 93/42/EEC concernant les dispositifs médicaux.Invacare® RobinTM a été développé et construit en considérant systématiquement les besoins de l’utilisateur et des tierces personnes lors de la manipulation du lève-personnes ou de son utilisation. Invacare® RobinTM a été développé conformément au Standard Européen NF EN 10535 et NF EN 60601-2-38.Invacare® RobinTM a été supervisé et contrôlé tout au long du process de fabrication et le lève-personnes achevé a été inspecté par le contrôle des produits finis. La plaque d’identification et la marque QA sont placées sur le lève-personnes attestant que le contrôle des produits finis a approuvé le lève-personnes. Le lève-personnes répond aux exigences de l’analyse de risques de la norme NF EN 14971. Nous vous remercions de lire le Manuel de l’Utilisateur dans son intégralité avant d’utiliser le lève-personnes.Invacare® EC-Høng A/S est certifiée ISO 9001 et ISO 13485.
PT DECLARAÇÃO DE QUALIDADEParabéns pela sua nova elevador de transferência Invacare® RobinTM da Invacare® EC-Høng A/S.A elevador de transferência tem a marca CE - em conformidade com a directiva 93/42/EEC referente a aparelhos médicos.A elevador de transferência foi concebida e desenhada, tendo em consideração o seu utilizador e o seu(s) assistente(s), que o ajudarão a manipular a elevador de transferência. A elevador de transferência foi concebida em conformidade com o Standard Europeu EN 10535.Durante todo o processo de fabrico e produção, a elevador de transferência Invacare® RobinTM foi supervisionada e a sua qualidade controlada, sendo o produto final inspeccionado e testado pelo nosso controlo de qualidade. As etiquetas e a marca QA, são colocadas na elevador de transferência, após a aprovação final da elevador de transferência, confirmando e garantindo a conformidade com o nosso controlo de qualidade. Por Favor, leia atentamente este manual de utilizador antes de utilizar a sua elevador de transferência.A Invacare® EC-Høng A/S está certificada em conformidade com ISO 9001 e ISO 13485.
Table of contents
1. Purpose and application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.1Manufacturer:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.2Purposeandapplication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.3Important/warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.4Sling’s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.5Unpackingandpreparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.6Installingthechargingstation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.7Hoistsysteminstallation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2. Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2.1Approvals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2.2Beforefirst-timeuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3. Description of functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 3.1Descriptionofthehoist´smainparts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 3.2Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3.3Safetyguidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.4Charging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 3.5Charginganemergencybattery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 3.6BatteryMaintanance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4. Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 4.1Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 4.2Transitcoupling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 4.3EC-GantryTM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4.3.1EC-GantryTraverseTM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 4.4Flexiscope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 4.5Mountingof4-Pointspreaderbar. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 4.6Trackswitch,manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.7Roomtoroomtransfer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5. Dismounting & transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 5.1Dismounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 5.2Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6. Maintenance & Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 6.1Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 6.2Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 6.3Maintenanceperformedbyowner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7. Service & lifetime . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 7.1Service&lifetime. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 7.2Battery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 7.3Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 7.4Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5
8. Technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 8.1Symbols............................................................. 27 8.2InstallationoptionsofthechargerIP41/IP67 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 8.2.1IP41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 8.2.2IP67. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
9. Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
10. Service schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6
1. Purpose and application
1.1Manufacturer:
Invacare®EC-HøngA/S Østergade3 DK-4270Høng Tel.+4558852722 Fax.+4558854386
1.2Purposeandapplication
PurposeRobinTMisaceilingmountedhoist,meetingtherequirementsforhospitals,housingfortheelderly,institutionsandprivatehomes.RobinTMisintendedformovingclientsweighingamaximumof200kgincludingclothing,hoistslingandotheraccessories.Thehoistisdesignedforindooruseinnormaltemperaturesandhumidity.Ifyouwishtousethehoistinanyotherenvironment,contactInvacare®foradviceandinstructions.Useofthehoistforpurposesotherthantransferofpeopleisnotadvisedandisundertakenatonesownrisk.
ConditionsItisaconditionofsalethatthepurchaseroftheceilinghoistmustreceivetrainingfromInvacare®personneloranauthoriseddealer.WhentheRobinTMceilinghoistisusedinaninstitutionorinanursinghome,itshouldalwaysbeoperatedbyspeciallytrainedstaff,andusedinrecommendedtrack.systems.ThetracksystemmustbeinstalledbyInvacare®personneloranauthoriseddealer.
Seated to seated transferDuringaseatedtransfere.g.fromawheelchair,theRobinTMshouldbepositionedclosetothepersontobelifted.Thesuspensionhooksshouldideallybepositionedlevelwiththechestandideallynofurtherthanoverthemiddleofthethigh.Placethesuspensionhooksparallelwiththeuser’sshoulders.See picture.Theslingmaynowbeattachedtothesuspensionhooks.Forplacementofthesling,see section»Placementofthesling«.
Lying to lying transferPositionthesuspensionhooksabovethepersontobelifted,alignedtotheusersshoulders(see picture).Ifpossible,raisethebackportionofthebedsothattheusermaybeliftedinaseatedposition.Theslingmaynowbeattachedtothesuspensionhooks.Forplacementofthesling,see section»Placementofthesling«.
7
Placement of slingTheslingstrapsaretobehookedinpairs(onefromthebackandonefromtheleg)ontheirownsuspensionhook.
(Attachmentoftheslingisdescribedintherespectiveusermanual,thetextandpicturebelowisonlyaprincipleexplanation).
Withthepersontobeliftedinsittingposition,theslingisplacedbehindtheuser’sback,e.g.betweenthebacksupportofthechairandtheuser.Theclientshouldnotsitonthesling.Thelegstrapsareguidedforwardalongtheoutsideofthethighandinunderthethighbetweenthekneeandthehipjoint.Thelegstrapsarecrossedbetweeneachotherinfrontoftheuser.Allfourslingstrapsarenowreadytobemountedintothesuspensionhooks.
Whentheclientislyingdown,he/sheshouldbelogrolledontotheirside.Positiontheslinginsuchawaythattheupperedgeisinlinewiththecrownofthehead.Theuseristhenrepositionedontohis/herbackandlogrolledintheoppositedirection.Theslingisthenrolledtobeflatundertheuserwhenhe/sheisrolledontohis/herback.Thelegstrapsarethenguidedinunderthethighboneandcrossedinbetweeneachother.Allfour(six)slingstrapsarenowreadytobeattached.
Ifthereisdoubtastotheuseofthesling,pleasecontactInvacare®.
1.3Important/warnings
• ReadtheentiremanualbeforeusingRobinTM.• Robin’smax.loadof200kgmustnotbeexceeded.• Ifaccessoriesareratedwithalowerloadthan200kg,thisdeterminesRobin’smaximumload.• RobinTMmustonlybeusedtomoveapatient.• RobinTMmustnotbeusedwherethereisariskofcontactwithwater.• IftheRobinTMisdamaged,itmustbetakenoutofuse.CallyourlocalInvacare®dealerorInvacare’s servicedepartment.• Thecordforemergencystop/emergencyloweringistobeadjustedsoitiswithintheuser’srange.
Risks when using the hoist:
Riskofentrapmentbetweenhookandhookhousing.Riskofentrapmentbetweenhooksandsling,whenlifting.Riskofinjurytoclientandcarercausedbydanglinghooksorhandcontrol.
A caregiver should always operate the hoist.Ifthehoistinspecialcasesistobeusedbyadisabledperson,thentheinstallershallprovideacommunicationdevice,sothatintheeventofanemergencythedisabledpersonisabletocallforhelp.
8
1.4Sling’s
UseonlyCE-markedsling’switharatedloadequaltoorlowerthanthatoftheRobin’smaxload.
Theslingmustbesuitablefor2or4pointspreaderbars(pleaselookatInvacare’sbrochureconcerningslings),i.e.withsewnliftingloopsthatcanbemountedontoa2or4pointspreaderbars.
1.5Unpackingandpreparation
PerformavisualinspectionofRobinTM.Ifthepackagingisdamageduponarrival,RobinTMpartsmustbethoroughlyexaminedforvisibledamageordefects.Ifthereisanysuspicionofdamage,RobinTMmustnotbeusedbeforequalifiedservicepersonnelorservicetechnicianshaveapprovedthehoist.Itistobeexpected,thatthehoist´sbatteryisdischargedonaccountofself-discharge.Thehoistmustthereforebechargedforupto2hoursbeforefirsttimeusage,refer to section»Charging«.
The following is to be included in the package:
1. RobinTM.2. Handcontrol.3. Quickguide.(Thechargingunitmustbeorderedseparately;partno.:1446317).
1.6Installingthechargingstation
Thechargingunitshouldbeattachedtothewall,orthemostpracticalplace,withinreachofasocketoutlet,toparkthehoistwhennotinuse.Itisimportanttomakesurethatthehandcontrolcanbereadilyplacedintothechargerholderwhenbeingcharged.Thereforeplacethechargeratamaximumdistanceof85cmundertherailinwhichthehoistismounted.
Itisimportanttomakesurethatthereisaccesstothemainswherethechargerisplaced.
Tomountthecharger:,firstinsertwallfittingwith2screws,thenconnectthecablefromthechargertotheactualholder,andpresstheholderontothemountedwallsupport.
Thewallsupportforthechargermustbemountedwiththreescrewsundertheholderandatadistancethatsuitsthelengthofthechargercable.
Connectthechargertothesocketoutletandcheckthatitisoperational.
Morethanonechargingunitcanbemountedforonehoist,asthesecanworkindependently.
BeawarethatonlyanIP67chargercanbemountedinawetroom,See section8.2
9
1.7Hoistsysteminstallation
Priortofittingthehoist,ensurethatthehoisttrolleyismounted.Removetheendstopthatisattachedtotherailtrack.Thetrolleycanthenbeinsertedintotherail.(Itisunimportantwhichwaythetrolleyisfacing).Theendstopmustthenbereinstalled.See picture.
WhenmountingRobinTM,locatethehoisttrolleyattheØ32mmroundrecessintherail.See picture.
Liftthehoistandpushthesuspensionpinthroughthehoisttrolley.Turnthehoist90degreestotherail’slongitudinaldirectionandplaceitasshownbelow.(Note:Supportthehoistwhileturning).See picture.
IfRobiniscorrectlymountedintothetrolley,itisnowpossibletopullthetrolleybackandforthalongtherail.
Ifthereisexcessiveresistance,orifthetrolleyunintentionallyrollsbyitselfduetoimbalanceintherailsuspensions,adjustthefrictionbrakeusinga4mmAllenkey.See picture.
Whenthesystemisinstalled,thehandcontrolmustbeconnected,andthecordfortheemergencystop/loweringmustbeadjustedsoitiswithintheuser’sreach.
Thesystemmustbeinspectedbeforeuse; see section»Beforefirsttimeuse«.
RobinisnowreadytobeusedintheEC-TrackTMsystem.
WhenusingRobininotherrailsystems,analternativetrolleymustbeused.ContactInvacare®forinformationregardingapprovedsystems.
10
2. Inspection
2.1Approvals
TheRobinTMceilinghoisthasbeentestedandapprovedinaccordancewithLiftstandardEN/ISO10535,andisTÜVapproved.ThehoistisCE-markedinaccordancewiththemedicaldirectiveandstandardEN/ISO10535.
2.2Beforefirst-timeuse
Beforefirstuse,allfeaturesofthehoistsystemmustbeinspectedbyaauthorisedInvacare®installer.
ThehoistisnotintendedtobeusedtogetherwithanyotherMEDICALELECTRICALEQUIPMENT,meaningthatthepatientmustnotbeconnectedtoanyotherMEDICALELECTRICALEQUIPMENTduringtransferwiththeRobinTMhoist.
3. Description of functions
3.1Descriptionofthehoist´smainparts
Control button UP Control button DOWN
Connection for hand control
Emergency stop/ lowering strap
Connection for emergency battery
Indicator lightHand control
Strap hooks
Lifting strapsRetainer for hand control (accessory)
11
1 2FUNCTIONSONTHEUPPERSIDE
1. Suspensionpinforconnectingthehoisttothe hoisttrolley.2. Fingerscrewfordismountingthetopsection.
3.2Operation
Operatingforcesofbuttons: Max.5N
Hand control and indicator lightRobinTMisactivatedwhenthehandcontrolbuttonsarepressed.TheblueindicatorlightilluminateswhenRobinTMisinuse,andturnsoffautomaticallyafterapprox.30seconds.Thehoistisliftedandloweredbyoperatingthehandcontrol.Thecontrolbuttonsontheundersideofthehoistmayalsobeusedforthispurpose.Thecontrolbuttonsworkindependentlyofthehandcontrol.
Note: Thebuttonsmustbedepressedinthemiddleinordertoensurecorrectfunction,whichisbestobtainedasshowninthepictures.
Thehoistwillautomaticallyrunathighspeedupwardsaswellasdownwards,whenitisnotloaded.Thestrapsmustbestraighttoensurethattheyfeedsintotheaperturescorrectly.Thestrapsareheldtaughtbytheweightofthehooks,thereforetheliftingstrapsshouldbefreelysuspendedwhentheunloadedhoistisrunningupwardsanddownwards.Whenthehoistisloadedwithaweightofmorethan10kg,thespeedwillautomaticallybeloweredtoasafelevelwhenmovinginbothdirections.RobinTMhasabuilt-insoftstart,ensuringamorecomfortableliftingandloweringmovementoftheuser.
Transport/running in the rail systemTheemptyhoistcanbemovedalongtherailsystembypullingthetwoliftingstraps.Thehandcontrolmustnotbeusedtopullthehoistalong.Duringatransfer,thecarershouldgentlypushorpulltheclientsittinginthesling.
Up
Down
12
3.3Safetyguidelines
Emergency stopIfthehoiststraps´movesunintentionallyupwardsordownwards,pullthe»emergency stop«chord.Whentheemergencystopisactivated,thehoistwillnotoperate.Iftheemergencystopbuttonispushedup,thehoistwillbereadyforuseagain.
Emergency loweringTheemergencyloweringfeaturecanbeusedinordertoloweraclienttothetransfersurfaceifthehoisthasbecomeinoperable.Emergencyloweringisoperatedbycontinuouslypullingthecord.
Whentheemergencyloweringfunctionisactivated,allendstopsandsecurityfunctionsaredeactivated.Itistherebypossibletowinduptheliftingstrapsbackup,whenactivatingtheemergencyloweringforprolongedperiods.
Strap securing/pinch protectionIftheliftingstrapsareguidedincorrectlyinandoutofthehoist,i.e.ifthestrapshasbecometwisted,thehoistwillautomaticallyturnoff.Note:Whenattachingtheslingtothehooks,andthestrapsaretwisted,thestrapswillrotateandstretchout,thusensuringtrouble-freeoperation.
Thepinchprotectionfeatureautomaticallyturnsoffthehoist,ifaclientunintentionallytrapshis/herhandintheliftingstrapduringalift.Thehoistwillstopimmediatelywhenthehandmakescontactwiththestrapinlet.
IfRobinTMisloadedwithaweightofmorethan200kg,thehoistwillautomaticallyturnofftosafeguardagainstoverloading.
Electromagneticinterferencebetweenthehoistandotherelectricalproductscanoccur.Toreduceoreliminatesuchelectromagneticinterference,increasethedistancebetweenthehoistandtheproductsorswitchofftheproducts!
Allinstallationsmustcomplywiththenationalrulesandstandards.Useonlyapprovedinstallationcomponents.Ateverypointofsuspension,theroof/ceilingstructuremustbeabletoabsorbaloadofatleast300kg.
3.4Charging
Asoundindicatorandtheflashingindicatorlightsignalswhenthehoistneedstobecharged.Thehoistmustbemovedalongtherailsystemandparkednearbythewallmountedcharger.Attachthehandcontroltothechargerunit.LEDonthechargerindicatesthatchargingistakingplace.Werecommendparkingthehoistinthechargingpositionwhennotinuse.
LED indication Function Function duration
Yellow Batterynotconnected -
Yellow Initializingandanalysisofbattery 12-14seconds
Red Fastcharge 5min-2hours
Green Tricklecharge -
Alternatingredandgreen Faultcondition(batterydefect) -
Therewillstillbesufficientpowertolowertheuserfromthetimewhenthehoistindicatesthatchargingisneeded.Ifthebatteryshouldbecometotallydischargedwhileatransferisinprogress,anexternalbatterymaybetempo-rarilyconnectedtothehoist.See the section »Battery«.
NB!Itisrecommendedtointerruptthechargingandpulltheemergencystop,whenthelifterhasnotbeenusedformorethan4-6weeks.
13
Note!Itisnotpossibletousethehoistwhilstchargingthebattery.
IfthesoundindicatoroftheRobinbeeps,andthebluefunctionlightsflashafterRobinthathasbeenfullychargedandhasbeeninuseforonlyashortperiodoftime,thebatterymaybereadytobereplacedbyanewone.See section»7.2Battery«.Anotherpossibleoptioncouldbethatthebatteryneedstobeexcercised.See section»3.6Batterymaintanance«.
3.5Charginganemergencybattery
Theemergencybatterypackageischargedbyplacingthepluginthechargersocket.Aspecialadapterisnecessary.TheLEDlightturnsgreentoindicatetheunitisfullycharged(appox.chargingtime2hours).
3.6Batterymaintanance
IfthesoundindicatoroftheRobinTMbeeps,andthebluefunctionlightsflashesafterthehoisthasbeenfullychargedandhasbeeninuseforonlyashortperiodoftime,thebatterymightneedtobeexercised.Toenablethebatterytobechargedfully,itisneccessarytoexercisethebatteryfromtimetotime.Followingalongperiodofstorage,itcouldbeneccesserytoexercisedthebattery.ItisalsosuggestedthatthebatteryisexercisedinsituationswhereRobin MoverTMisusedinfrequently,and/orwherethehandcontrolisplacedinthechargerforlongperiodsoftime.Inthissituation,therewillalwaysbecapacityenoughforafewlifts,butwerecommendtoexercisethebatteryassoonaspossibleafterwards.ExerciseofthebatterycanbedonebycharginganddischargingRobinTMcompletelyseveraltimes.Thisisdonebychargingthebatteryuntilthelightindicatorturnsgreen.Thehandcontrolisthenremovedfromthecharger,andthestrapswithloadareloweredandraisedforoneminute,witha10minutesbreak,continuingthisprocedureuntilthebatteryiscompletelydischarged(continueevenwhenthesoundsignalforlowpowerbeepsandthebluelightflashes).Chargethebatteryandlowerandraisethestrapsuntilthebatteryisdechargedagain.Repeatthisprocedure2-3times.
14
4. Accessories
4.1 Accessories.1Accessories
Order no.: Article:
1493139 Battery2Ah,4/5SCNiMH.1492102-7035 ChargerkitIP41forhandcontrol,grey(with1,2/2mcable,bracketandretainers).1492047-7035 ChargerIP41forhandcontrol,grey(with1,2/2mcable).1451909-7035 ChargerkitIP67forhandcontrol,grey(with0,1/1mcable,bracketandretainers).1451910-7035 ChargerkitIP67forhandcontrol,grey(with0,1/2mcable,bracketandretainers).1451911-7035 ChargerkitIP67forhandcontrol,grey(with0,1/3mcable,bracketandretainers).1451912-7035 ChargerkitIP67forhandcontrol,grey(with1,2/1mcable,bracketandretainers).1451913-7035 ChargerkitIP67forhandcontrol,grey(with1,2/2mcable,bracketandretainers).1451914-7035 ChargerkitIP67forhandcontrol,grey(with1,2/3mcable,bracketandretainers).1450290-7035 Clampforchargerforwallmounting(withouttransformer),bracketandretainers.1449668-7035 Chargerforexternalbattery,grey.1446269-7012 Distanceclamp(Flexiscope).1488974-7012 4-pointspreaderbar,45cm.1492967-7012 2-pointspreaderbar,35cm.1492968-7012 2-pointspreaderbar,45cm.1485474-7035 Retainerforwallmountedcharger.1449228-7035 Retainerforhandcontrol,strapmounted.1450289 Gearwheelgrease(3gtube).1449383-9010 HoisttrolleyInalto/Invacare.1485183-9010 Room-toroomtrolley(white)EC-Track.1485183-9006 Room-toroomtrolley(grey)EC-Track.1451620-7012 Extendedliftingstrapswithhooksandemergencyloweringstrap(extralengthmustbe specifiedwhenordering).1493132 Ergoletspreaderbarwithweightscale.1504765-7012 Stretcherwithsling.
15
A B
C D
4.2Transitcoupling
Theroomtoroomcouplingcanbeusedintherailsystem,wherethereisarequirementtotransferapatientbetweentworooms.
Theroomtoroomcouplingcanbeusedwhenthereistraversesysteminbothroomsortraverssysteminoneoftheroomsandmonorailintheotherroom.
A) Totransferapatientbetweenrooms,placethehoistwithloadrightunderthemonosystem. See drawing A point 1. ThenmovetheRobinandtrackingsothatthetwosectionsoftracklineupwitheachother. See drawing A point 2.
B) Pressthebuttononthewallmountedcontrolbox.See drawing b point 3.Whenthetrackingislockedinplacethegreenlightilluminatesandthetransfercanbecarriedout.See drawing B point 4 & 5.
C) WhentheRobinissafelylocatedonthenewsectionoftracking,thebuttononthecontrolbox,canthenbepressedagaintounlockthetwosectionsoftracking.See drawing C point 6 & 7.
D) Thegreenlightwillthengoout,toindicatethatthetwosectionsoftrackingcanbesafelymovedawayfromeachother.See drawing D point 8 & 9.
ItisesssentialthatduringthetransferprocessyoulocatetheRobinMoverTMasindicatedintheabovediagrams,thiswillensurethesafecouplingofthetrackingsections.Ifindoubt,oryouexperiencedifficultyincouplingthetracking,contactyourInvacaredealer.
16
4.3EC-GantryTM
TheEC-GantryTMisusedwherethereisarequirementforafreestandingrailsystem.
HeightrangeforEC-GantryTMwithoutwheels:216-251cmHeightrangeforEC-GantryTMwithwheels:195-230cm
TheEC-Gantryisavailablewithorwithoutwheels.
ThewheelsoftheEC-GantryTMmustbelockedbeforeperforminganylift.
ToincreasetheheightoftheEC-Gantry,pullthelegsectionupward.TodecreasetheheightoftheEC-Gantry,pullthehandleasindicatedabove,adjusttheEC-GantryTMtothedesiredheightandthenreleasethehandletoallowthelegsectiontolockbackintoplace.
Wihregardstothesafetyofthepatient,itisrecommendedthattheEC-GantryTMisnotmovedwithapatienthanginginthesling.
ItisimportantthattheEC-GantryTMisplacedonalevelsurfacewithregardstotheriskoftipping.
PleaseobservethatthereisariskofthepatientswingingintotheEC-Gantry´slegs.ThereisariskofcarerstumblingoverfootoftheEC-GantryTMsystem.
Thereisariskoftippingifthepatientisveryunsettledorsuffersfromspasminthesling.Neverleaveapatientintheslingunattended.
17
4.3.1EC-GantryTraverseTM
TheEC-GantryTraverseTMisusedwherethereisarequirementforafreeroomcoveringstandingrailsystem.
HeightrangeforEC-GantryTraverseTM:218-253cm
ItisimportantthattheEC-GantryTraverseTMisplacedonalevelsurfacewithregardstotheriskoftipping.
PleaseobservethatthereisariskofthepatientswingingintotheGantry´slegs.
Thereisariskoftippingifthepatientisveryunsettledorsuffersfromspasminthesling.Neverleaveapatientintheslingunattended.
18
4.4Flexiscope
ScopeFlexiscopeTMisusedwiththeInvacare©RobinTMceilinghoistinconnectionwiththetransferormovingofdisabledpatientswhorequireextendeddistancebetweentheslingsshoulderstraps.Max.load:200kg.
InstallationInstallationofFlexiscopeTMontheliftingstrapsiscarriedoutaccordingtotheillustrations.
AdjustmentAlignmentofthewidthismadebypullingthebuttonAwhilemovingthestrapbarBtothedesiredposition.Note! Ensurethatthebariscorrectlyengaged.
Maintenance and cleaning
Cleaninginconnectionwithdailyuse:Allpaintedsurfacescanbewhipedoffwithadampcloth.Onlyhouseholddetergentsareallowed.Neveruseacids,alkalinesorsolventssuchasacetoneorcellulosethinner.DesinfectionofFlexiscopeTM-itmustbewhipedoverwithofficiallyapproveddetergentsonly.TheblackcoverisremovedfromFlexiscopeTMandcanbemashinewashedat70ºC-donottumbledry,bleachordryclean.
•
••
19
4.5Mountingof4-pointand2-pointspreaderbar
LowertheRobinliftingstrapsdowntoasuitableheightandmountthespreaderbarshownontheillustrationbelow.
UseonlythespreaderbarsproducedforRobin.
Ensurethatthespreaderbarsuitstheclientandthetypeofliftrequired.
Ensurethespreaderbariscorrectlymountedbeforeuse.
•
•
•
20
4.6Trackswitch,manual
Thetrackswitchisusedintracksystemswhereachangeindirectionisrequiredduringmovementofthepatient.Thetrackswitchisoperatedbypullingthecordsoneitherside-thischangeindirectionmustbecarriedoutbeforetheclientisguidedintothedriveswitcharea.Thedirectionischanged,andthehoistwilltakethedirectiontothesidewherethecordispulled.Thetrackswitchismechanicallysecuredagainstderailing.
21
4.7Roomtoroomtransfer
1. Thetwohoistsaremovedasclosetothedoorwayaspossible. TheslingisattachedtohoistA. TheliftingstrapsonhoistAareloweredasfaraspossiblewith conciderationtothepatient´scomfort.
2. ThetwoliftingstrapsonhoistBareloweredasmuchasnecessaryto attachthemtothesling. Note! Ensurethattheliftingstraphooksareinstalledcorrectly.
3. TheliftingstrapsofhoistBareraisedwhilethestrapsonhoistAare lowered.ContinueuntilthepatientiscompletelysuspendedbyhoistB.
4. TheliftingstrapsonhoistAisnowuncoupledfromtheslingandrewound. Note!Bewareofdangelinghooksorhandcontrol.
5. TheliftingstrapsonBarerewoundtothedesiredheight.Theroomto roomtransferisnowcarriedout.
22
-30→50°C -22→122°F
10→70% 700→1.100hPa
5. Dismounting & transport
5.1Dismounting
WhendismountingRobinTM,pleasesee section »Installationofhoistsystem«.
5.2Transport
Theemergencystrapmustbepulled,whentheRobinTMmotorisdemountedfromthetracks,fortransportorstorage,otherwisetheRobinTMcanbeaccidentallyactivated.See section »6.2Storage«.Duringtransportation,thehoistmustbeplacedonasoftsurface,e.g.aclothorfoamrubbermat.MakesurethatRobinTMremainsuprightwhilstbeingmoved.
Werecommendusingtheoriginalpackagingfortransportationunderthefollowingconditions:
Temperaturebetween-30°and+50°C.Humiditybetween10og70%RH.Pressurebetween700og1.100hPa.
6. Maintenance & Storage
OnlyInvacareauthorisedpersonnelshouldperformmaintenanceandservice.Dailyinspectionshouldbecarriedoutbythecarerafterreadingthemanual.SixmonthlyinspectionaccordingtoUKLOLER1998regulationsshouldbedonealso.SparepartslistandwiringdiagramcanbeobtainedfromInvacare®orbedownloadedfromwww.invacarebeds.dk.
6.1Cleaning
Cleaning in connection with daily use:
Liftingslingsaretobewashedinaccordancewithmanufacturersinstructions.RobinTMandthehandcontrolcanbewipedwithamoistenedcloth.Onlycommoncleansingagentsmustbeused;neveruseacids,andsolvents.Thestraphooksmustbecleanedwithwaterandcommoncleansers.Ifdisinfectionisrequired,drythoroughlywithknownbrandsofliquiddetergent.
6.2Storage
RobinTMshouldbestoredinadryplaceunderconditionsmentionedinsection5.2»Transport«.PulltheemergencystrapbeforetheRobinisstored,toavoidunwantedactivationoftheRobinandfurthermoretoprotectthebatteryfromdamage,whenthehoistisstoredforalongerperiodwithoutbeingused.Iftheemergencystrapisnotpulledthecapacityofthebatterywilldepletecompletely,andwillbeirreparablydamaged.Whentheemergencystrapispulledduringstorageitwillbepossibletoregaintheentirecapacityofthebatterybyexercisingthebattery,see section»3.6Batterymaintanance«.
6.3Maintenanceperformedbyowner
Checktheliftingstrapsonaregularbasis.Iftherearesignsofdamageinanyway,donotusethehoist.ContactyourlocalInvacare®dealerforadvice.ThestrapsmustonlybereplacedbypersonsauthorisedbyInvacare®.
•••
••
••
23
1 2 3
4 5 6
7 8
7. Service & lifetime
7.1Service&lifetime
ToensuretheoptimalfunctionandsafetyofRobin,itisimportanttoperformserviceandmaintenanceinaccordancewiththemanual.
Maintenancemustbecarriedoutfollowingnationalregulations,andatleastyearly.AccordingtotheFactoriesInspectorateregulations,RobinhastoundergoanannualservicebyanInvacare®authorisedperson.DurabilityandlifetimeexpectancyisbasedonanannualserviceprogrammeperformedbyanInvacare®authoriseddealerorInvacare®.
Robinhasanexpectedlifetimeofapprox.10years.After10yearsuse,itmustbeinspectedbyqualifiedpersonnelbeforefurtheruse.Ifthehoistisusedinlocalitieswithincreasedhumidity,areductioninlifetimemustbeexpected.
SparepartlistforRobinandEC-TrackcanbeobtainedfromInvacare®.
7.2Battery
To change battery.Ifbatteryreplacementisnecessary,RobinTMmustberemovedfromthetracksystem.Pleasesee section»dismounting«.Releasetheupperpartofthehoistbyremovingtwoscrews.See picture.
Removethebatterypackinsidethehoistbyunpluggingthewires,andreplaceitwithanewone.Reattachtheupperpartofthehousing.
Important!UseonlyInvacare®recommendedbatteries,see section »Accessories«.Thebatterymustbelocatedasshowninpicture7.Ifthebatteryhousingisdamaged,donotusethebattery.
Itisnotnecessaryforanapprovedservicepersontocarryoutthebatterychange.
24
Use of emergency batteriesShouldtheneedforimmediateuseofRobinariseatatimewhenthehoisthasadischargedbattery,aInvacare®emergencybatterycanbeused.
Thisbatteryisconnectedtotheundersideofthemotor,hangingfreelyduringclienttransportation.
Usetheemergencybatteryonlywhenabsolutelynecessary.Alwaysrechargethisbatteryafteruse,pleasesee section »Charging«.
TheemergencybatteryisanaccessorythatcanbeorderedfromyourInvacare®dealer,see section »Accessories«.
Important!UseonlyInvacare®recommendedemergencybatteries.
7.3 Maintenance.3Maintenance
Whenperformingannualorregularmaintenance,allpartsdesignedtocarryloadmustbetestedatmaximumcapacity(bothhoistandtracksystem).Allsafetyfeaturesmustbechecked.Afunctiontestofthesystemwith1.5xmaximumloadmustalsobeperformed.
Duringtheservice,allpartsthatisexchangedandallworkcarriedout,mustbedocumented.ThewornanddefectivepartsmustbereplacedwithnewsparepartsfromInvacare.InspectionofthehoistandtracksystemmustbecarriedoutbyaqualifiedInvacare®authoriseddealer.Torecordthedetails,pleaseusetheservicescheduleonpage30.
EC-Track system inspection
Allbolts/fixturesinwallsandceilingmustbecheckedandretightenedifnecessary.Theendstoponthetracksischecked,adjustedifnecessaryandretightened.Systemswithtrackconnectionsshouldbecheckedandrealignediftheconnectionhasbecomemis-alignedduringuse.Testthetrolleyandthetraversetrolley.Checkfornoiseandsmoothmovement.Foratraversesystem,checkallfunctionsandconnections.Checkthefrictionbrakeonboththehoist-andtraversetrolley,andadjustaccordingly.
Hoist inspection
Mechanicalparts:Checkthesuspensionpinandthecrossaxle.Checkthesteel-andplasticgearwheelsforwearanddamage.Ifthegearwheelsaredry,usethegearwheelgrease(see»accessories«).Checktheloosestrapsensorandadjust,ifnecessary.Checktheguidecylinderandtheendstop.Checkthestrapsforwearanddamage.Checkthehooksfordamage.Checkthescrewsandpins.
•••
•••
••••••••
25
Electricalparts:• Checkthewiringfordamageandsecureconnections.• Checkthemicroswitchesforclampingandtheirmechanicalfunction.• Checkthehandcontrolfordamageandfunction.Makesurethatthehandcontrolissecurelyconnected.• Checktheemergencystop/lowering.
7.4Troubleshooting
ServiceandmaintenanceofRobinTMmustonlybeperformedbypersonnelwhohavereceivedthenecessaryinstructionandtraining.
Symptom Possible cause Action
Thehoistisnotrespondingcorrectlywhenpressingthehandcontrolbuttons,andtheup/downbuttonsonthehoistarenotfunctioning.
1)Theemergencystopis activated.2)Thebatteryisdischarged.
3)Thefuseisblown.
1)Depresstheemergencystop button.2)Putthehoistonchargeoruse theemergencybattery.3)Replacethefuse.
Thehoistisnotrespondingcor-rectlywhenpressingthehandcontrolbuttons,buttheup/downbuttonsonthehoistarefunctioning.
1)Thehandcontrolisnot connectedcorrectly.2)Thehandcontrolisdefect.
1)Securetheplug.
2)Replacethehandcontrol.
Thestrapswillonlymovedownwardsandnotupwards.
1)Thebatteryisnearlydischarged.
2)Theloadistohigh.
1)Putthehoistonchargeoruse theemergencybattery.2)Reducetheload(thehoist isequippedwithaover- currentsafetysystem).Check thehandcontrol,it´s connectionandthere-entering endstop.
Thestrapswillonlymoveup-wardsandnotdownwards.
1)Thehandcontrolisdefect.2)Theloosestrapfunctionis activated.
1)Replacethehandcontrol.2)Checktheloosestrapswitch.
Thehoistspeedisthesamewithandwithoutload.
1)Thenon-loadquickmovement functionisactivatedpermanently.
1)Checkthequickmovement switch.
Theliftingstrapsaretwisted,anddonotstraightenout.
1)Theliftinghooksdonotturn freelyenough.2)Thestrapsarefrayed.
1)Cleanorreplacethehooks.
2)Replacethestraps.
Thehoistemitsexcessivenoisewhenactivated.
1)Defectbearings,gearwheelsor motor.
1)Replacethehoist.
Iftheabovementioneddoesnotsolvetheproblem,pleasecontactInvacare®ServiceDepartment.
26
= 200 kg
8. Technical specifications. Technical specifications
Dimensions(HxBxD) 19x47x25cm.
Max.liftingcapacity 200kg.
Numberofliftspercharge 65liftsof0.5mwith80kg.
Liftingvelocityat200kgload 2cm/s.
Maxliftinginterval 250cm.
Heightfromcellingtoslinghook 41cm.
Totalweightwithoutsling 7kg.
Electricalconnectionofcharger 230VAC50Hz
Voltage/currentbetweencharger/hoist 24VDC900mA/50mA.
Chargingtime(with2Ahbattery) Max2hours
Battery 20pcs.1.2V/2.0Ah4/5SCNiMH
Intermittence 8%
Soundpressure 50-55dB(A)
Protectionclass IPX4
8.1Symbols
Warning.
SWL(SafeWorkingLoad)=Max.load(patient+sling+otheraccessories).
ClassIIequipment.
TypeBequipment-patientnotseparatedfromground.
CE-marking.
Theproductshouldbereusedwherepossible.
27
8.2InstallationoptionsofthechargerIP41/IP67
8.2.1IP41
Maynotbeusedinawetroomenvironment.
8.2.2IP67
ThechargerisdesignedaccordingtoIP67,andcanthereforebeusedinwetlocationsaccordingtotheseillustrations:
Installationofthechargerisrecommendedinzone2and3.
Bath tub
Shower without basin Shower with fixed partition
Showerhead Showerhead
Bath tubBath tub
Shower basin
28
9. Disposal
ThisproducthasbeensuppliedfromanenvironmentallyawaremanufacturerthatcomplieswiththeWasteElectricalandElectronicEquipment(WEEE)Directive2002/96/CEThisproductmaycontainsubstancesthatcouldbeharmfultotheenvironmentifdisposedofinplaces(landfills)thatarenotappropriateaccordingtolegislation.The»crossedoutwheeliebin«symbolisplacedonthisproducttoencourageyoutorecyclewhereverpossible.
Pleasebeenvironmentallyresponsibleandrecyclethisproductthroughyourrecyclingfacilityatitsendoflife.
Allelectricalpartsmustberemovedanddisposedofaselectricalcomponents.Plasticpartsmustberecycledorincinerated.Steel-andaluminiumpartsmustbedisposedofasmetalscrap.
Batteries: Localandnationalregulationsforenvironmentalrecyclingmustbefollowed.Batteriesmustalwaysbedeliveredtoanapprovedcollectionplace.
Disposalmustalwaysfollowtherelevantlocallawsandregulationsforwasteandscrap.
29
10. Service schedule
Service and maintenance of the hoist must be carried out only by trained/experienced personnel
Personnel hoist serial number: __________________________________
Inspection point:Date:
Inspection of the track system
Allbolts/fixturestowallandceilingmustbecheckedandre-tightenedifnecessary.
Endstopsonthetracksmustbechecked,movedifnecessaryandre-tightened.
Systemswithtrackconnectionsshouldbecheckedandre-alignediftheconnectionhasbecomemis-alignedduringuse.
Checktrolleyandtraversetrolley.
Checktrolleywheelsforsmoothmovementandabnormalnoise.
Checkfrictionbreakadjustmentontrolleys.Adjustifnecessary.
Inspection of hoist
Mechanical parts:
Checksuspension-andcrosspin.
Checksteel-andplasticgearwheelsforwearanddamage.
Ifgearwheelsaredry,applygearwheelgrease(see»accessories«)
Checkloosestrapsensorandadjustifnecessary.
Checkguidecylinderandendstops.
Checkstrapsforwearanddamage.
Checktheliftinghooks,screwsandpinsforwearanddamage.
Electrical parts:
Checkthewiringfordamageandcorrectconnections.
Checkmicroswitchesforclampingandmechanicalfunction.
Checkhandcontrolfordamageandfunction.Makesurethatthehandcontrolisconnectedcorrectly.
Checkemergencystopandemergencylowering.
30
Customer Sales and Service
Portugal
Denmark
Sweden & Finland
Norway & Iceland
Spain
Belgium & Luxemburg
Netherlands
Germany
Portugal
France
Italy
United Kingdom & Ireland
Australia
New Zealand
Switzerland Austria
Manufacturer:
INVACARE EC-Høng A/SØstergade 3DK-4270 Høngwww.invacarebeds.dk
Invacare® EC-Høng A/SIdent. no.: 1485245Version: 06Date: 05.2008
INVACARE A/SSdr. Ringvej 39DK-2605 BrøndbyPhone: +45 36 90 00 00Fax: +45 36 90 00 [email protected]
INVACARE ASGrensesvingen 9Postbox 6230 EtterstadN-0603 OsloPhone: +47 22 57 95 00Fax: +47 22 57 95 [email protected]
INVACARE N.V.Autobaan 22B-8210 Loppem, BrüggePhone: +32 50 83 10 10Fax: +32 50 83 10 [email protected]
INVACARE ABFagerstagatan 9 / Box 66S-163 91 SpångaPhone: +46 8 761 70 90Fax: +46 8 761 81 [email protected]
INVACARE B.V.Celsiusstraat 46NL-6716 BZ EdePhone: +31 318 695 757Fax: +31 318 695 [email protected]
INVACARE LTDSouth RoadBridgend Industrial EstateUK-Bridgend, CF31 3PYPhone: +44 1 656 664 321Fax: +44 1 656 667 [email protected]
INVACARE LdaRua Senhora de Campanhã 105P-4369-001 PortoPhone: +351 225 1059 46/47Fax: +351 225 1057 [email protected]
INVACARE Poirier S.A.SRoute de St. RochF-37230 FondettesPhone: +33 2 47 62 64 66Fax: +33 2 47 42 12 [email protected]
INVACARE S.A.C/Areny S/NPoligon Industrial de Celrà E-17460 Celrà (Girona)Phone: +34 972 49 32 00Fax: +34 972 49 32 [email protected]
AQUATEC GmbHAlemannenstrasse 10D-88316 IsnyPhone: +49 75 62 700 0Fax: +49 75 62 700 [email protected]
INVACARE MECC SAN S.R.L.Via dei Pini 62I-36016 Thiene (VI)Phone: +39 0445 38 00 59Fax: +39 0445 38 00 [email protected]
INVACARE Australia Pty Ltd1 Lenton Place, North RocksNSW 2151Phone: +61 2 8839 5333Fax: +61 2 8839 [email protected]
INVACARE NZ4 Westfield Place,Mt. WellingtonAucklandPhone: +64 9 917 3939Fax: +64 9 917 [email protected]
MOBITEC REHAB AGBenkenstrasse 260CH-4108 WitterswilPhone: +41 61 487 70 80Fax: +41 61 487 70 [email protected]
MOBITEC MOBILITÄTSHILFEN GmbHHerzog Odilostrasse 101A-5310 MondseePhone: +43 6232 5535 0Fax: +43 6232 5535 [email protected]