RIGOL TECHNOLOGIES,INC.
Introduction to DL3000 Programmable DC
Electronic Load
Edward Pan
Rev.1.0
2017-Dec
Fundamental Concept of E-Load
What is Electronic Load?
It absorbs the energy from Power Supply and transfers to another type of energy
and to be a storage or dissipated equipment.
PowerSupply
Electronic Load
Current
Application:
1. ON/OFF Power Supply
2. linear Power Supply
3. UPS
4. Transformer
5. Rectifier
6. Battery and Charger etc
Fundamental Concept of E-Load
What kind of functions in E-Load?
Static Mode
1. Constant Current(CC),
2. Constant Voltage(CV),
3. Constant Resistor(CR) and
4. Constant Power modes(CP).
Transient Mode
1. Continuous,
2. Pulse,
3. Toggled
4. List Function
Static Transient
DL3000 Introduction
DL3021/DL3021A: Single Output, 150V/40A, 200W
DL3031/DL3031A: Single Output, 150V/60A, 350W
Up to 30kHz Dynamic Frequency
Adjustable Current Rising Speed 0.001A/us~5A/us
The minimum reading resolution : 0.1mV, 0.1mA
4.3 inch TFT LCD,Parameters and Status can be
displayed simultaneously.
OV/OC/OP/OT/Reverse Connection Protection.
4 Static Mode: CV/CC/CR/CP
3 Transient Mode: Continuous, Pulse, Toggling
List functions can support up to 512 steps
Current Waveform Display
Battery Test Function, OC Protection Test Function, OP
Protection Function, Factory Test Function and Short
Circuit Test Function.
Shut down maintain and memory functions
Build-in RS232、USB、LAN Ports
USB-GPIB Interface Module ( Option )
DL3000 Introduction
Input Terminal
SenseTerminal
Waveform Display Button
Application Function Buttons:Battery, OCP, OPP
Main Function Buttons:CC, CV, CR, CPCon, Pul, Tog, List
System Button:Local, Utility, Option, Store, Help
DL3000 Introduction
• Innovative Waveform Display Quick Button, it is
adjustable under any modes.
• The Sampling Period can be up to 4µs, more
detailed waveforms can be addressable.
• The Max Timing Windows is 800s,longer
waveform trend can be observed.
• Time base is adjustable.
Waveform Display Quick Button
RIGOL TECHNOLOGIES, INC.
Feature/Advantage/Benefit
Feature:1. Large 4.3 inch LCD Display
2. CC,CV,CR,CP Mode for Static state
3. CON, Pulse, Toggle, List function embedded
Advantage:1. Dynamic Frequency up to 30KHz.
2. Slew Rate up to 5A/us
3. Monitor waveform intuitively.
Benefit:1. Easy to monitor Voltage and Current status.
2. Convenient to observe Current Waveform directly on display
RIGOL TECHNOLOGIES, INC.
Required Equipment List for Test
1. DL3031/A/3021/A Electronic Load
2. DP832A Power Supply
3. Testing cables
4. Ultra Sigma Software
5. Ultra Load Software
6. PC Laptop x 1
RIGOL TECHNOLOGIES, INC.
DL3000 Test Procedure
1. Connect the test cable from Power Supply to Eload input terminals, turn
on the CH1 of DP832A and set the Voltage value to 5V.
2. The S+, S- terminals should be also connected because it can make
the measurement more accurate and even compensate the voltage drop
when the DUT offers large current through the load.
RIGOL TECHNOLOGIES, INC.
DL3000 Test_CC Mode
3. Press the “CC” button on DL3000 and setup the current limitation value
on the left bottom slide of the display.
4. Press “ON/OFF” button on DL3000 to activate CC Mode operation.
Set the current limitation value to 2A
Constant current value will be shown on Display
RIGOL TECHNOLOGIES, INC.
DL3000 Test_CV Mode
5. Press the “CV” button on DL3000 and setup the Voltage limitation value
on the left bottom slide of the display.
6. Press “ON/OFF” button on DL3000 to activate CV Mode operation.
Set the constant Voltage value to 5V
Constant voltage value will be shown on Display
RIGOL TECHNOLOGIES, INC.
DL3000 Test_CR Mode
7. Press the “CR” button on DL3000 and setup the constant resistor value
on the left bottom slide of the display.
8. Press “ON/OFF” button on DL3000 to activate CR Mode operation.
Set the constant resistor value to 100ohm
Constant resistor value will be shown on Display
RIGOL TECHNOLOGIES, INC.
DL3000 Test_CP Mode
9. Press the “CP” button on DL3000 and setup the constant power value
on the left bottom slide of the display.
10. Press “ON/OFF” button on DL3000 to activate CP Mode operation.
Set the constant power value to 0.5W
Constant power value will be shown on Display
RIGOL TECHNOLOGIES, INC.
DL3000 Test_Continuous Mode
11. Press the “CON” button on DL3000 and enter the setup environment.
12. Set the parameters as below indicated table and then press “Apply” on
the bottom side of the display, then press “TRAN”button on DL3000,
Users can observe the loading average voltage and current value on
display.
Set the parameters in this table
Average Voltage & Current will be shown on Display
RIGOL TECHNOLOGIES, INC.
DL3000 Test_Continuous Mode
13. Press the “Eye” button on DL3000 and users can monitor the current
waveform in advance.
RIGOL TECHNOLOGIES, INC.
DL3000 Test_Pulse Mode
14. Press the “PUL” button on DL3000 and enter the setup environment.
15. Set the parameters as below indicated table and then press “Apply” on
the bottom side of the display, then press “TRAN”button on DL3000.
16. Continuously press the “TRAN”, users can see the Pulse current
waveform on display
RIGOL TECHNOLOGIES, INC.
DL3000 Test_LIST Mode
17. Press the “LIST” button on DL3000 and enter the setup environment.
18. Set the mode as “CR”, Steps for 10 and Cycles to 1, and then press
“TRAN” button on DL3000, it can be done.
RIGOL TECHNOLOGIES, INC.
Ultra Load Software Application
19. Connect the A to B USB cable between DL3000 and Laptop.
20. Open the Ultra Sigma Software to check whether DL3000 is connected
or not.
21. Start the Ultra Load Software as below indicated
RIGOL TECHNOLOGIES, INC.
Ultra Load Software application
22. Select “Interface” item on top of software environment and connect the
instrument address in advance.
RIGOL TECHNOLOGIES, INC.
Ultra Load Software Application
23. Then the remote mapping diagram will be shown on Ultra Load Software
24. Users can setup all the application mode via this interface.