Instructions for useGENTLEmini LUX 4500 BR - REF 1.007.1700

Page 2: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Distributed by:KaVo Dental Corporation11729 Fruehauf DriveCharlotte, NC 28273 USATel.: 847 550 6800Fax: 847 550 6825

Manufacturer:Kaltenbach & Voigt GmbHBismarckring 39D-88400 Biberachwww.kavo.com

Page 3: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Contents

1 User instructions ............................................................................................................................................... 52 Safety ............................................................................................................................................................... 7

2.1 Description of safety instructions .............................................................................................................. 72.2 Safety instructions .................................................................................................................................... 9

3 Product description ......................................................................................................................................... 143.1 Purpose – Proper use ............................................................................................................................. 153.2 Technical data ........................................................................................................................................ 173.3 Transportation and storage conditions ................................................................................................... 18

4 First use .......................................................................................................................................................... 204.1 Attach the MULTIflex coupling ............................................................................................................... 214.2 Check the amount of water .................................................................................................................... 224.3 Check the pressure ................................................................................................................................ 23

Contents 1

Page 4: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

4.4 Check O-rings ........................................................................................................................................ 245 Operation ........................................................................................................................................................ 25

5.1 Mount the medical device ....................................................................................................................... 265.2 Remove the medical device ................................................................................................................... 285.3 Insert the milling cutters or diamond grinders ........................................................................................ 285.4 Removing the milling tool or diamond grinder ........................................................................................ 32

6 Troubleshooting .............................................................................................................................................. 346.1 Exchanging the O-rings on the coupling on the supply hose ................................................................. 346.2 Cleaning the spray nozzle. ..................................................................................................................... 356.3 Change the water filter ........................................................................................................................... 36

7 Preparation methods according to ISO 17664 ............................................................................................... 387.1 Preparations at the site of use ................................................................................................................ 387.2 Cleaning ................................................................................................................................................. 39

7.2.1 Cleaning: Manual cleaning - external ......................................................................................... 39

Contents 2

Page 5: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.2.2 Cleaning: Automated external cleaning ...................................................................................... 407.2.3 Cleaning: Manual cleaning of the inside ..................................................................................... 417.2.4 Cleaning: Automated internal cleaning ....................................................................................... 41

7.3 Disinfection ............................................................................................................................................. 437.3.1 Disinfection: Manual disinfection - external ................................................................................ 447.3.2 Disinfection: Manual disinfection - internal ................................................................................. 457.3.3 Disinfection: Machine disinfection - external and internal .......................................................... 46

7.4 Drying ..................................................................................................................................................... 477.5 Care products and systems - Servicing .................................................................................................. 48

7.5.1 Care products and systems - Servicing: Servicing with KaVo Spray ......................................... 507.5.2 Care products and systems - Servicing: Servicing with KaVo QUATTROcare .......................... 52

7.6 Packaging ............................................................................................................................................... 547.7 Sterilisation ............................................................................................................................................. 557.8 Storage ................................................................................................................................................... 58

Contents 3

Page 6: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

8 Tools ............................................................................................................................................................... 599 Warranty terms and conditions ....................................................................................................................... 61

Contents 4

Page 7: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

1 User instructions

Dear UserCongratulations on purchasing this KaVo quality product. By following theinstructions below you will be able to work smoothly, economically andsafely.

© Copyright by KaVo Dental GmbH

Symbols

Refer to the section Safety/warning symbols

Important information for users and service technicians

User instructions 5

Page 8: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Action request

Steam sterilisable 134 °C -1 °C / +4 °C (273 F -1.6 F / +7.4 F)

Thermodisinfectable

Target group

This document is intended for dentists and their assistants. The section onstarting up is also intended for service technicians.

User instructions 6

Page 9: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

2 Safety

2.1 Description of safety instructions

Warning symbol

Structure

DANGER

The introduction describes the type and source of the hazard.This section describes the potential consequences of non-observance.

▶ The optional step includes necessary measures for hazard preven‐tion.

Safety 7

Page 10: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Description of danger levels

The safety instructions listed here, together with the three levels ofdanger will help avert property damage and injury.

CAUTION

CAUTIONindicates a hazardous situation that can cause damage to property or mildto moderate injuries.

WARNING

WARNINGindicates a hazardous situation that can cause death or serious injury.

Safety 8

Page 11: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

DANGER

DANGERindicates a hazardous situation that can directly cause death or seriousinjury.

2.2 Safety instructions

WARNING

Hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, un-typical warming or when the cutter or grinder cannot be held.

▶ Do not use further and notify Service.

Safety 9

Page 12: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

WARNING

Hazard from contraindication.If the soft tissue in the oral cavity is injured, the compressed air may enableseptic substances to enter the tissue.

▶ Treatment must be discontinued with instruments operated by com‐pressed air when soft tissue is damaged in the oral cavity.

CAUTION

Premature wear and malfunctioning from improper storage during longperiods of nonuse.Reduced product life.

▶ The medical device should be cleaned, serviced and stored in a drylocation, according to instructions, before long periods of nonuse.

Safety 10

Page 13: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

CAUTION

Risk due to incorrectly stored instrument.Injury and infection caused by chucked cutters or grinders.Damage to clamping system from dropping the instrument.

▶ After treatment, place the instrument properly in the cradle, withoutthe cutter or grinder.

CAUTION

Burning hazard from hot instrument head and instruments cover.If the instrument overheats, burns may arise in the oral area.

▶ Never contact soft tissue with the instrument head.

Safety 11

Page 14: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

CAUTION

Hazard from use as a light probe.Do not use the device as a light probe since the rotating cutter grinder cancause injury.

▶ For additional illumination of the oral cavity or preparation site, use asuitable light probe such as the KaVo DIAlux 2300L.

NoteFor safety reasons, we recommend that the tool holder system be checkedannually after the warranty period expires.

To ensure proper function, the medical device must be set up according tothe reprocessing methods described in the KaVo Instructions for Use, andthe care products and care systems described therein must be used. KaVo

Safety 12

Page 15: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

recommends specifying a service interval at the dental office for a licensedshop to clean, service and check the functioning of the medical device. Thisservice interval depends on the frequency of use and should be adjustedaccordingly.Service may only be carried out by KaVo-trained repair shops using originalKaVo replacement parts.

The following individuals are authorized to repair and service KaVo prod‐ucts:

▪ Technicians at KaVo branches throughout the world▪ Technicians specially trained by KaVo

Safety 13

Page 16: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

3 Product description

GENTLEmini LUX 4500 BR (Mat. no. 1.007.1700)

Product description 14

Page 17: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

3.1 Purpose – Proper use

Purpose:

This medical device is▪ Only intended for dental treatment. Any other type of use or alteration

to the product is impermissible and can be hazardous. The medicaldevice is intended for the following uses: Removal of carious material,cavities and crown preparations, removal of fillings, processing of toothand restoration surfaces.

▪ A medical device according to relevant national statutory regulations.

Product description 15

Page 18: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Proper use:

According to these regulations, this medical device may only be used forthe described application by a knowledgeable user. The following must beobserved:

▪ the applicable health and safety regulations▪ the applicable accident prevention regulations▪ these Instructions for use

According to these regulations, the user is required to:▪ only use equipment that is operating correctly,▪ use the equipment for the proper purpose,▪ protect him or herself, the patient and third parties from danger, and▪ avoid contamination from the product..

Product description 16

Page 19: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

3.2 Technical data

Drive pressure 2.1 to 3.5 bar (30 to 51 psi)Recommended drive pressure 2.8 bar (40 psi)Air consumption 49 to 55 Nl/min.Idle speed 380,000 to 450.000 rpmRecommended pressure 2 to 3 NReturn air < 0.5 bar (7 psi)

Can be attached to all MULTIflex (LUX) / MULTIflex LED couplings.

Product description 17

Page 20: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

3.3 Transportation and storage conditions

CAUTION

It is hazardous to start up the medical device after it has been stored re‐frigerated.This can cause the medical device to malfunction.

▶ The temperature of very cold products must be raised to a range of20°C (68°F) to 25°C (77°F).

Temperature: -20°C to +70°C (-4°F to +158°F)

Relative humidity 5% RH to 95% RH

Product description 18

Page 21: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)

Protect from moisture.

Product description 19

Page 22: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

4 First use

WARNING

Hazard from non-sterile products.Infection danger to the care provider and patient.

▶ Before first use and after each use, sterilise the medical device.

CAUTION

Damage from soiled and moist cooling air.Contaminated and moist cooling air can cause malfunctions and lead topremature bearing wear.

▶ Make sure that the supply of cooling air is dry, clean and unconta‐minated according to ISO 7494-2.

First use 20

Page 23: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

NoteAfter each treatment, KaVo recommends turning on the spray for at least20 seconds with the instrument attached to rinse the air and water chan‐nels.

4.1 Attach the MULTIflex coupling

▶ Screw the MULTIflex (LUX) / MULTIflex LED coupling onto the turbinehose and tighten with the wrench.

The amount of water in the spray can be adjusted by rotating the spray ringon the MULTIflex (LUX) coupling / MULTIflex LED coupling.

First use 21

Page 24: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

4.2 Check the amount of water

CAUTION

Overheating of the tooth due to insufficient amount of cooling water.Insufficient spray water can cause the medical device to overheat anddamage the pulp and tooth.

▶ Adjust the water amount for the spray cooling to a minimum of 50cm3/min (3.1 inch3).

▶ Check spray water channels and if necessary clean spray nozzleswith the nozzle needle (Mat. no. 0.410.0921).

▶ Check water filter and replace, if necessary.

First use 22

Page 25: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

4.3 Check the pressure

CAUTION

Compressed air connection to equipment.Dirty and moist compressed air causes premature bearing wear.

▶ Ensure dry, clean and uncontaminated compressed air according toISO 7494-2.

▶ Insert the test gauge (Mat. no. 0.411.8731) between theMULTIflex (LUX) coupling / MULTIflex LED coupling and the medicaldevice, and check the following pressures:

- Drive air: 2.1 to 3.5 bar (30 to 51 psi) (recommended: 2.8 (41 bar psi))Return air: < 0.5 bar (7 psi)- Water: 0.8 to 2.5 bar (12 to 36 psi)

First use 23

Page 26: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

- Spray air: 1.0 to 4.0 bar (15 to 57 psi)

4.4 Check O-rings

CAUTION

Missing or damaged O-rings.Malfunctions and premature failure.

▶ Make sure that all O-rings are on the coupling and undamaged.

Number of available O-rings: 5

First use 24

Page 27: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

5 Operation

NoteAt the beginning of each workday, the water-conducting systems shouldbe rinsed for at least 2 min. without the instrument being attached; if thereis a risk of contamination from reflux or back suction, the system must berinsed for 20-30 seconds.

Operation 25

Page 28: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

5.1 Mount the medical device

WARNING

Release of the medical device during treatment.A medical device that is not properly locked in place can release from theMULTIflex (LUX) / MULTIflex LED coupling during treatment.

▶ Before each use, check if the medical device is securely locked ontothe MULTIflex (LUX) / MULTIflex LED coupling by pulling on it.

Operation 26

Page 29: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

CAUTION

Damage from inaccurate coupling.Inaccurate coupling (especially during the afterglow period) can destroythe high-pressure lamp or the LED of a MULTIflex (LUX) / MULTIflex LEDcoupling or reduce its service life.

▶ Check the seat of the turbine on the coupling by pulling on it.

▶ Mount the medical device accurately on theMULTIflex (LUX) / MULTIflex LED coupling and push it backward untilthe coupling audibly locks in the medical device.

▶ Pull on it to make sure that the medical device is securely affixed to thecoupling.

Operation 27

Page 30: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

5.2 Remove the medical device

▶ Hold the coupling tight, and pull the medical device off while twistingslightly.

5.3 Insert the milling cutters or diamond grinders

NoteOnly use carbide cutters or diamond grinders that correspond to ISO1797-1 type 3, are made of steel or hard metal and meet the followingcriteria:- shaft diameter: 1.59 to 1,60 mm- Overall length: max. 19 mm- Shaft clamping length: min. 9 mm- Blade diameter: max. 2 mm

Operation 28

Page 31: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

WARNING

Use of impermissible cutters or grinders.Injury to the patient or damage to the medical device.

▶ Observe the instructions for use, and use the cutter or grinder prop‐erly.

▶ Only use cutters or grinders that do not deviate from the indicateddata.

CAUTION

Injury from using worn drill bits or burs.Drill bits or burs could fall out during treatment and injure the patient.

▶ Never use drill bits or burs with worn shafts.

Operation 29

Page 32: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

CAUTION

Hazard from defective chucking system.The cutter or grinder could fall out and cause injury.

▶ Pull on the cutter or grinder to check that the chucking system is okayand the cutter or grinder is securely held. When checking, insertingand removing, use gloves or a fingerstall to prevent an injury or in‐fection.

CAUTION

Injury hazard from cutters or grinders.Infections or cuts.

▶ Wear gloves or fingerstalls.

Operation 30

Page 33: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Forcefully press the push button with your thumb and simultaneouslyinsert the cutter or grinder all the way.

▶ Check that the cutter or grinder is seated by pulling on it.

Operation 31

Page 34: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

5.4 Removing the milling tool or diamond grinder

WARNING

Hazard from rotating cutter or grinder.Lacerations and damage to the chucking system.

▶ Do not touch rotating cutter or grinder.!▶ Never press the press-button while the cutter or grinder is rotating!▶ Remove the cutter or grinder from the contra-angle handpiece after

treatment to avoid injury or infection while storing it.

Operation 32

Page 35: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ After the cutter or grinder has stopped rotating, press the push buttonwith your thumb and pull out the cutter or grinder.

Operation 33

Page 36: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

6 Troubleshooting

6.1 Exchanging the O-rings on the coupling on the supply hose

CAUTION

Hazard from improper care of the O-rings.Malfunctions or complete failure of the medical device.

▶ Do not use Vaseline or other grease or oil.

NoteThe O-ring on the coupling may only be lubricated with a cotton ball wetwith KAVO spray.

▶ Press the O-ring between your fingers to form a loop.▶ Shove the O-ring to the front, and remove it.

Troubleshooting 34

Page 37: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Insert new O-rings into the grooves.

6.2 Cleaning the spray nozzle.

CAUTION

Hazard from insufficient spray water.Overheating of the medical device and damage to the tooth.

▶ Check the spray water channels and clean the spray nozzles with thenozzle needle Mat. no. 0.410.0921 if necessary.

▶ Check the water filter and exchange if necessary.

Troubleshooting 35

Page 38: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Clean the water passage in the spray nozzles by using the nozzleneedle (Mat. no. 0.410.0921).

6.3 Change the water filter

▶ Unscrew the sleeve ① anticlockwise from the insert ② and pull off.

▶ Pull back the insert ② slightly.

Troubleshooting 36

Page 39: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Use tweezers to remove the water filter ③ from the insert ②.

▶ Insert a new water filter.▶ Pull back the insert ② slightly.▶ Place the sleeve ① on the insert ②, and screw it tight clockwise.

Troubleshooting 37

Page 40: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7 Preparation methods according to ISO 17664

7.1 Preparations at the site of use

WARNING

Hazard from non-sterile products.There is a risk of infection from contaminated medical devices.

▶ Observe suitable personal protective measures.

▶ Remove all residual cement, composite or blood without delay.▶ Recondition the medical device as soon as possible after treatment.▶ Remove the cutter or grinder from the medical device.▶ The medical device must be dry when transported for reconditioning.▶ Do not place it in a solution or similar.

Preparation methods according to ISO 17664 38

Page 41: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.2 Cleaning

CAUTION

Malfunctions from cleaning in an ultrasonic unit.Defects in the product.

▶ Only clean manually or in a thermodisinfector.

7.2.1 Cleaning: Manual cleaning - external

Required accessories:▪ Tap water 30°C ± 5°C (86°F ± 10°F)▪ Brush such as a medium hard toothbrush

Preparation methods according to ISO 17664 39

Page 42: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Brush off under flowing tap water.

7.2.2 Cleaning: Automated external cleaning

KaVo recommends thermodisinfectors in accordance with EN ISO 15883that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"programme, "neodisher® mediclean" cleaning agent, "neodisher® Z" neu‐tralisation agent, and "neodisher® mielclear" rinsing agent and extends onlyto the compatibility of materials with respect to KaVo products).

Preparation methods according to ISO 17664 40

Page 43: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ For program settings as well as cleansers and disinfectants to be used,please refer to the Instructions for Use of the thermodisinfector (com‐plying with max. pH value of 10).

▶ In order to prevent negative effects on the medical device, make surethat the interior and the exterior of the medical device are dry, and thenlubricate immediately with care agents from the KaVo care system.

7.2.3 Cleaning: Manual cleaning of the inside

Not applicable.

7.2.4 Cleaning: Automated internal cleaning

KaVo recommends thermodisinfectors in accordance with EN ISO 15883that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"

Preparation methods according to ISO 17664 41

Page 44: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

programme, "neodisher® mediclean" cleaning agent, "neodisher® Z" neu‐tralisation agent, and "neodisher® mielclear" rinsing agent and extends onlyto the compatibility of materials with respect to KaVo products).

▶ For program settings as well as cleansers and disinfectants to be used,please refer to the Instructions for Use of the thermodisinfector (com‐plying with max. pH value of 10).

▶ In order to prevent negative effects on the medical device, make surethat the interior and the exterior of the medical device are dry, and thenlubricate immediately with care agents from the KaVo care system.

Preparation methods according to ISO 17664 42

Page 45: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.3 Disinfection

CAUTION

Malfunctioning from using a disinfectant bath or disinfectant containingchlorine.Defects in the product.

▶ Only disinfect in a thermodisinfector or manually.

Preparation methods according to ISO 17664 43

Page 46: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.3.1 Disinfection: Manual disinfection - external

KaVo recommends the following products based on material compatibility.The microbiological efficacy must be ensured by the disinfectant manufac‐turer.

▪ CaviCide from Metrex

Required tools:Cloths for wiping off the medical device.

Preparation methods according to ISO 17664 44

Page 47: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Spray the disinfectant on a cloth, then thoroughly wipe down the med‐ical device and leave the disinfectant to soak in according to the in‐structions from the disinfectant manufacturer.

▶ Observe the instructions for use for the disinfectant.

7.3.2 Disinfection: Manual disinfection - internal

The efficacy of manual internal disinfection must be demonstrated by themanufacturer of the disinfection agent. With KaVo products, use only dis‐infection agents that have been released by KaVo with respect to the com‐patibility of materials (e.g. WL-cid / made by ALPRO).

▶ Immediately after internal disinfection, lubricate the KaVo medical de‐vice immediately with care agents from the KaVo care system.

Preparation methods according to ISO 17664 45

Page 48: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ Observe the instructions for use for the disinfectant.

7.3.3 Disinfection: Machine disinfection - external and internal

KaVo recommends thermodisinfectors in accordance with EN ISO 15883that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"programme, "neodisher® mediclean" cleaning agent, "neodisher® Z" neu‐tralisation agent, and "neodisher® mielclear" rinsing agent and extends onlyto the compatibility of materials with respect to KaVo products).

▶ For program settings as well as cleansers and disinfectants to be used,please refer to the Instructions for Use of the thermodisinfector (com‐plying with max. pH value of 10).

Preparation methods according to ISO 17664 46

Page 49: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

▶ In order to prevent negative effects on the medical device, make surethat the interior and the exterior of the medical device are dry, and thenlubricate immediately with care agents from the KaVo care system.

7.4 Drying

Manual drying

For manual drying, in general use dry, clean and non-contaminated coolair according to ISO 7494-2.

▶ Blow off the outside and inside with compressed air until water dropsare no longer visible.

Preparation methods according to ISO 17664 47

Page 50: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Machine drying

The drying procedure is normally part of the cleaning program of the ther‐modisinfector.

▶ Please observe the instructions for use of the thermodisinfector.

7.5 Care products and systems - Servicing

WARNING

Sharp cutters or grinders in the medical device.Risk of injury from sharp or pointed cutters or grinders.

▶ Remove cutter or grinder.

Preparation methods according to ISO 17664 48

Page 51: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

CAUTION

Premature wear and malfunctions from improper servicing and care.Reduced product life.

▶ Perform proper care regularly!

NoteKaVo only guarantees that its products will function properly when the careproducts used are those listed as accessories, as they were tested forproper use on our products.

Preparation methods according to ISO 17664 49

Page 52: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.5.1 Care products and systems - Servicing: Servicing with KaVo Spray

KaVo recommends servicing the project twice daily (at noon and in theevening after hours), after each time the machine is cleaned, and beforeeach sterilisation.

▶ Remove the cutter or grinder and close the chuck.

▶ Cover the product with the Cleanpac bag.

▶ Place the product on the cannula, and press the spray button for onesecond.

Preparation methods according to ISO 17664 50

Page 53: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Care of clamping chuck

KaVo recommends cleaning and servicing the chuck system once a week.

▶ Remove the cutter or grinder, place the spray nipple tip in the openingand spray.

▶ Carry out servicing according to instructions in the section "Care withKaVo Spray".

Preparation methods according to ISO 17664 51

Page 54: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.5.2 Care products and systems - Servicing: Servicing with KaVoQUATTROcare

Cleaning and care unit with expansion pressure for effective cleaning andcare.

KaVo recommends servicing the product after each time it is used, i.e. aftereach automatic cleaning and before each sterilisation.

▶ Remove cutter or grinder.▶ Service the product.

Preparation methods according to ISO 17664 52

Page 55: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

KaVo QUATTROcare plus spray can

KaVo recommends cleaning and servicing the chuck system once a week.

See the KaVo QUATTROcare instructions for use

▶ Remove the cutter or grinder, place the spray nipple tip in the openingand spray.

▶ Subsequently treat with the care products and care systems specified.

See also: Care with KaVo QUATTROcare

Preparation methods according to ISO 17664 53

Page 56: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.6 Packaging

NoteThe sterilisation bag must be large enough for the instrument so that thebag is not stretched.The quality and use of the sterilised product packaging must satisfy ap‐plicable standards and be suitable for the sterilisation procedure.

▶ Weld each medical device individually in a sterilised item package!

Preparation methods according to ISO 17664 54

Page 57: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.7 Sterilisation

Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060/ISO 17665-1 (e.g. KaVo STERIclave B 2200 / 2200 P)

CAUTION

Premature wear and malfunctions from improper servicing and care.Reduced product life.

▶ Before each sterilisation cycle, service the medical device with KaVocare products.

Preparation methods according to ISO 17664 55

Page 58: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

CAUTION

Contact corrosion due to moisture.Damage to product.

▶ Immediately remove the product from the steam steriliser after thesterilisation cycle!

The KaVo medical device has a maximum temperature resistance up to138 ℃ (280.4 °F).

Preparation methods according to ISO 17664 56

Page 59: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

(Depending on the available autoclave,) select a suitable procedure fromthe following sterilisation processes:

▪ Autoclave with three times initial vacuum:– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4

°F)Drying time: 20 min.

▪ Autoclave using the gravitation method:– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4

°F) or alternativelyDrying time: 30 min.

– at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4°F)Drying time: 30 min.

▶ Use according to the manufacturer's Instructions for Use.

Preparation methods according to ISO 17664 57

Page 60: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

7.8 Storage

▶ Reconditioned products should be stored protected from dust withminimum exposure to germs in a dry, dark and cool space.

▶ Comply with the expiry date of the sterilised item.

Preparation methods according to ISO 17664 58

Page 61: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

8 Tools

Available from dental suppliers.

Material summary Mat. no.Replacement turbine with key 1.006.5535Replacement turbine without key 1.006.4564Water filter with nipple 1.000.4823Instrument stand 2151 0.411.9501Insert for turbines 0.411.9902Cleanpac 10 units 0.411.9691MULTIflex spray head (nozzle) 0.411.9921QUATTROcare coupling 0.411.7991

Tools 59

Page 62: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

Only for the USA

Material summary Mat. no.KaVo Spray America 2113 A 0.411.9660QUATTROcare plus Spray America2141 P

1.005.4524

Only for Canada

Material summary Mat. no.KAVOspray Canada 2114 A 0.411.9680QUATTROcare plus Spray Canada2149 P

1.005.4523

Tools 60

Page 63: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

9 Warranty terms and conditions

The following warranty conditions apply to this KaVo medical device:

KaVo provides the end customer with a warranty of proper function andguarantees zero defects in respect of material and processing for a periodof 24 months from data of invoice, subject to the following conditions:In case of justified complaints, KaVo will honour its warranty with a freereplacement or repair. Other claims of any nature whatsoever, in particularwith respect to compensation, are excluded. In the event of default, grossnegligence or intent, this shall only apply in the absence of mandatory legalregulations to the contrary.KaVo cannot be held liable for defects and their consequences that havearisen or may arise from to natural wear, improper handling, cleaning ormaintenance, non-compliance with operating, maintenance or connectioninstructions, calcination or corrosion, contaminated air or water supplies orchemical or electrical factors deemed abnormal or impermissible in ac‐

Warranty terms and conditions 61

Page 64: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

cordance with KaVo's instructions for use or other manufacturer's instruc‐tions.No liability is assumed when defects or their consequences are derivedfrom manipulations or changes to the product by the customer or a thirdparty.Service warranty claims will only be accepted if the product is submittedalong with proof of purchase in the form of a copy of the invoice/deliverynote.

Warranty terms and conditions 62

Page 65: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked
Page 66: Instructions for use - Dentalcompare.com LU… · suitable light probe such as the KaVo DIAlux 2300L. Note For safety reasons, we recommend that the tool holder system be checked

1.00

7.36

78 ·

kb ·

2010

0924

- 04

- U

S · e

n