Upload
others
View
95
Download
1
Embed Size (px)
Citation preview
General rights Copyright and moral rights for the publications made accessible in the public portal are retained by the authors and/or other copyright owners and it is a condition of accessing publications that users recognise and abide by the legal requirements associated with these rights.
Users may download and print one copy of any publication from the public portal for the purpose of private study or research.
You may not further distribute the material or use it for any profit-making activity or commercial gain
You may freely distribute the URL identifying the publication in the public portal If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim.
Downloaded from orbit.dtu.dk on: Mar 22, 2020
Installation Qualification, Performance Qualification and Routine Control
Miller, Arne
Publication date:2011
Document VersionPublisher's PDF, also known as Version of record
Link back to DTU Orbit
Citation (APA):Miller, A. (Author). (2011). Installation Qualification, Performance Qualification and Routine Control.Sound/Visual production (digital)
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Validation and routine control ofValidation and routine control ofili i b i di iili i b i di isterilization by irradiationsterilization by irradiation
Arne MillerRi ø Hi h D R f L b tRisø High Dose Reference LaboratoryRisø DTUTechnical University of DenmarkDK-4000 RoskildeDenmark
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 11
Denmark
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Content of presentation:Content of presentation:pp
InInstallationstallation Qualification Qualification -- IQIQ
Operational Operational Qualification Qualification -- OQOQ
•• Performance QualificationPerformance Qualification -- PQPQPerformance Qualification Performance Qualification PQPQ
Process Control and MonitoringProcess Control and Monitoring
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 22
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Installation qualificationInstallation qualification
undertaken to demonstrate that the sterilization equipment has undertaken to demonstrate that the sterilization equipment has q pq pbeen supplied and installed in accordance with specification.been supplied and installed in accordance with specification.
Whether or not data are “in accordance with their specification”Whether or not data are “in accordance with their specification”Whether or not data are in accordance with their specification Whether or not data are in accordance with their specification depends on agreementdepends on agreement
Measurements often the same as for Operational QualificationMeasurements often the same as for Operational QualificationMeasurements often the same as for Operational QualificationMeasurements often the same as for Operational Qualification
Responsibility:Responsibility: Facility OperatorFacility Operator
ISO 11137ISO 11137--1, sect 9.11, sect 9.1
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 33
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Operational qualificationOperational qualification
•• -- carried out by irradiating appropriate test material to carried out by irradiating appropriate test material to demonstrate the capability of the equipment to deliver the demonstrate the capability of the equipment to deliver the sterilization process that has been defined.sterilization process that has been defined.p fp f
•• -- provides baseline data to show consistent operation of the provides baseline data to show consistent operation of the facilityfacilityfacilityfacility
Responsibility: Facility OperatorResponsibility: Facility Operator
ISO 11137ISO 11137--1, sect 9.21, sect 9.2
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 44
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Operational qualification GammaOperational qualification Gammap qp q
•• dose distribution in reference dose distribution in reference d ( )d ( )product(s)product(s)
•• dose as function of dwell timedose as function of dwell time•• process interruptionprocess interruption•• process interruptionprocess interruption•• mix of productsmix of products
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 55
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ gamma cont..OQ gamma cont..
Gamma FacilityGamma FacilityLayoutLayout
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 66
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ OQ gammagammagammagammacont..cont..
Placement ofof dosimeters
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 77
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ Gamma cont..OQ Gamma cont..
Dose distributionDose distributionin reference productin reference product
Isodose curvesIsodose curves
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 88
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ gamma cont..OQ gamma cont..
Dose mappingDose mapping
Dose maps must be made with fully loaded irradiation chamberDose maps must be made with fully loaded irradiation chamber
Range of densities of reference product must cover extremes of Range of densities of reference product must cover extremes of products to be irradiatedproducts to be irradiated
Additional measurement:Additional measurement:
ff f i iff f i i−− effect of process interruptioneffect of process interruption
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 99
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ gamma cont..OQ gamma cont..
Dose mappingDose mapping
Dose mapping shall be carried out on a sufficient number Dose mapping shall be carried out on a sufficient number of irradiation containers to allow determination of theof irradiation containers to allow determination of theof irradiation containers to allow determination of the of irradiation containers to allow determination of the distribution and variability of dose between containersdistribution and variability of dose between containers
Dose mapping must be repeated when significant changes Dose mapping must be repeated when significant changes of the irradiation plant are madeof the irradiation plant are made e ge g source changesource changeof the irradiation plant are made, of the irradiation plant are made, e.g. e.g. source changesource change
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1010
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Operational qualification Electron beamOperational qualification Electron beam
•• CharacteristicCharacteristic parameters toparameters to bebe measuredmeasuredCharacteristicCharacteristic parameters to parameters to bebe measuredmeasured
•• dosedose distribution reference distribution reference productproduct•• scanscan widthwidthscan scan widthwidth•• energyenergy•• dosedose as as functionfunction of speed, of speed, currentcurrent, , beambeam
widthwidth•• beambeam spotspot
i t tii t ti•• processprocess interruptioninterruption
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1111
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Placement of dosimeters Placement of dosimeters in reference productin reference product
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1212
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Dose distribution Dose distribution i fi fin reference in reference productproduct
Limits for acceptable variations can bedefineddefined.
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1313
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Scan width (=extended beam width)Scan width (=extended beam width)
Limits for acceptable variationsvariations can bedefined.
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1414
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
EnergyEnergy
Wedge and stack for energy measurementWedge and stack for energy measurementWedge and stack for energy measurementWedge and stack for energy measurement
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1515
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Typical depthTypical depth--dose curvedose curveTypical depthTypical depth--dose curvedose curve
Energy Energy –– range range relationshipsrelationshipsrelationshipsrelationships
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1616
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Dose as a function Dose as a function of of Dose = f(I/V*Sw)
-- beambeam currentcurrent (I)(I)-- convconv speed (V)speed (V)-- beambeam widthwidth ((SwSw))
y = 1129,8x - 0,1538R2 = 0 9996
7080
-- beambeam widthwidth ((SwSw)) R 0,9996
30405060
se, k
Gy
0102030
Do
00,000 0,010 0,020 0,030 0,040 0,050 0,060 0,070
I/(V*Sw)
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1717
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Beam spot patternBeam spot pattern
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1818
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
Process interruptionProcess interruption
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 1919
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
OQ EOQ E--beam cont..beam cont..
EE--beam OQ should be repeated when significant changes of thebeam OQ should be repeated when significant changes of theEE beam OQ should be repeated when significant changes of the beam OQ should be repeated when significant changes of the irradiation plant are made that may influence dose or dose irradiation plant are made that may influence dose or dose distribution.distribution.
It is recommended to repeat OQ at regular intervals, e.g. annually It is recommended to repeat OQ at regular intervals, e.g. annually i d t d t t i t t tii d t d t t i t t tiin order to demonstrate consistent operation.in order to demonstrate consistent operation.
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2020
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualificationPerformance qualificationConcerns dose mapping of real productConcerns dose mapping of real product
t id tif th l ti d it d f i i d it id tif th l ti d it d f i i d ito identify the location and magnitude of minimum and maximum to identify the location and magnitude of minimum and maximum dosesdoses
andandandand
to determine the relationship between the min and max doses and the to determine the relationship between the min and max doses and the routine doseroutine dose
Responsibility: Primary ManufacturerResponsibility: Primary Manufacturer
ISO 11137ISO 11137--1, sect 9.31, sect 9.3
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2121
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualification ..cont..Performance qualification ..cont..
Purpose of dose mappingPurpose of dose mapping
-- to demonstrate that: to demonstrate that:
•• D (min)D (min) D (Steril) and D (max)D (Steril) and D (max) D (acceptable)D (acceptable)•• D (min) D (min) D (Steril) and D (max) D (Steril) and D (max) D (acceptable)D (acceptable)
•• D (min) determined by sterilization requirementsD (min) determined by sterilization requirements
•• D (max) determined by radiationD (max) determined by radiation--induced changes in productinduced changes in product
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2222
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualification ..cont..Performance qualification ..cont..
Strategies for dose mappingStrategies for dose mapping-- based on:based on:
OQ measurementsOQ measurements Inhomogeneous product distribution, orientation, voids, Inhomogeneous product distribution, orientation, voids,
i t fi t finterfaces.interfaces. ExperienceExperience
Processing categoriesProcessing categories
Limitations of dose mappingLimitations of dose mapping
Ability to measure “the true” value of D(min) or D(max)Ability to measure “the true” value of D(min) or D(max) Location of dosimeterLocation of dosimeter
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2323
Location of dosimeterLocation of dosimeter Resolution of dosimeter systemResolution of dosimeter system
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualification ..cont..Performance qualification ..cont..Performance qualification ..cont..Performance qualification ..cont..
Examples from real lifeExamples from real lifeExamples from real life …..Examples from real life …..
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2424
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Dose map examplesDose map examplesDose map examplesDose map examples
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2525
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Measurement UncertaintyMeasurement Uncertainty
Performance qualification ..cont..Performance qualification ..cont..Measurement UncertaintyMeasurement Uncertainty
Dose mapping shall be carried out using representative irradiation Dose mapping shall be carried out using representative irradiation containers sufficient in number to determine the variability of dose containers sufficient in number to determine the variability of dose ff y fff y fbetween containersbetween containers
Practical approach:Practical approach:
Either dose map in detail 3Either dose map in detail 3--10 containers10 containers
orororor−− dose map in detail one dose map in detail one cntainercntainer−− repeat measurement of D (min) and D (max) in several (e.g. 10) repeat measurement of D (min) and D (max) in several (e.g. 10) p ( ) ( ) ( g )p ( ) ( ) ( g )
containers containers −− evaluate measurement uncertainty taking into account all evaluate measurement uncertainty taking into account all
t i t tt i t t
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2626
uncertainty components.uncertainty components.
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualification ..cont..Performance qualification ..cont..
Setting Process LimitsSetting Process Limits
Select process parametersSelect process parametersp pp p
gamma:gamma: e.g. timer settinge.g. timer setting ee--beam:beam: e.g. conveyor speed e.g. conveyor speed
so that D(min) will exceed D(sterile) and so that D(min) will exceed D(sterile) and so that D(max) will not exceed D(acceptable)so that D(max) will not exceed D(acceptable)
Select process limits based on the measured uncertainty.Select process limits based on the measured uncertainty.
Practical approachPractical approach select process parameters so that D(min) is two standard select process parameters so that D(min) is two standard
d i ti hi h th D(d i ti hi h th D( t ilt il))
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2727
deviations higher than D(deviations higher than D(sterilsteril))
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualification ..cont..Performance qualification ..cont..
Selection of uncertainty Selection of uncertainty limits for process controllimits for process control
Normal distribution
1
limits for process controllimits for process control0,8
0,4
0,6
Prob
abili
ty
0,2Limit based onLimit based on2 standard deviations2 standard deviations
0-4 -3 -2 -1 0 1 2 3 4
Standard deviations
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2828
Medical Devices Sterilization for ManufacturersMedical Devices Sterilization for Manufacturers
Cameron Quality ConsultantsCameron Quality Consultants
Performance qualification ..cont..Performance qualification ..cont..
Repetition of Performance Qualification Dose MapRepetition of Performance Qualification Dose Map
Only needed if product is changed Only needed if product is changed or if facility is changedor if facility is changedy gy g
review of documentation at defined intervals is recommended, review of documentation at defined intervals is recommended, e g every 3 yearse g every 3 yearse.g. every 3 yearse.g. every 3 years
Copenhagen Medical Devices consulting apsCopenhagen Medical Devices consulting aps 22 September 200322 September 2003 2929