Installation Instructions SW CRE_1907
Citation preview
HEALTHCARE Imaging Services
Enclosure
Document No: DD+DIS112.11E
Dispose the enclosure document after finishing your task (observe the confidentiality note below).
CONFIDENTIALITY NOTE: Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE: The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 05-2011 printed in Germany Agfa Company Confidential
Document Node ID: 33338960 Enclosure_e_template_v07
Copyright 2011 Agfa HealthCare N.V.
This number is needed for production line only: F7.5175.9404.2 CR 30-X
Type 5175/100/105/110/120/125
Installation Instructions Software CRE_1907
Purpose of this Document
This enclosure is part of the software CRE_1907, available on the Agfa HealthCare Library, and the CD-ROM which is part of digitizer delivery.
It describes all steps for installation of software CRE_1907.
Referenced Documents Document Title
DD+DIS103.11E CR 30-X, Service Bulletin 48, Software CRE_1907 available
DD+DIS310.09E CR 30-X, Service Bulletin 34, Introduction of Speed Class free Operation
DD+DIS150.06E Chapter 3.8, Software Releases, Patches
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Manufacturer Agfa HealthCare N.V.
Publisher Agfa-Gevaert HealthCare GmbH Tegernseer Landstrae 161 D - 81539 Mnchen Germany
Copyright 2011 Agfa HealthCare N.V. All rights reserved. Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING: Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document (see Agfa HealthCare Library => General Info => Agfa HealthCare => Publications => Service Manual) prior to attempting any operation, repair or maintenance task on the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on the product.
Revision 0 CR 30-X Page 2 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 CR 30-X Page 3 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
LIST OF CONTENTS
1 INTRODUCTION ......................................................................................................................6
2 CHECKING INSTALLED DIGITIZER SOFTWARE VERSION.................................................7
3 FOR WINDOWS 7 ONLY: PREPARING THE FIREWIRE DRIVER.........................................8
4 DE-INSTALLING OLD SOFTWARE.......................................................................................11
5 INSTALLING SOFTWARE CRE_1907 AT THE CONTROL PC ............................................12
6 INSTALLING THE AGFA DEVICE DRIVER ........................................................................13
7 UPLOADING SOFTWARE TO THE DIGITIZER ....................................................................14
8 PERFORMING A BACKUP ....................................................................................................17
9 VERIFYING THE SUCCESSFUL SOFTWARE UPGRADE...................................................18
10 APPENDIX: TROUBLESHOOTING SOFTWARE INSTALLATION PROBLEMS AT THE CONTROL PC ........................................................................................................................19
10.1 After Uninstallation of the Software the Message appears: You must restart your system 19
10.2 After Uninstallation of the Software the "Found New Hardware Wizard" appears .................20
10.3 After initial Installation the "Found New Hardware Wizard" does not appear.........................21
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Subject Installation Instructions Software CRE_1907
Prerequisite Software CRE_1907 can be installed on any CR 30-X digitizer independent on the currently installed software and hardware.
IMPORTANT:
Software CRE_1907 has only been released with NX workstation software NX 2.0.6805 SU1 / NX 3.0.6805 SU1 and higher versions. If necessary upgrade to NX 2.0 6805 SU1 / NX 3.0.6805 SU1 or higher prior to installing software CRE_1907.
For details see:
DD+DIS139.07E - NX 2.0 - Service Bulletin No. 04 - Release of NX 2.0.6805 (Intranet Link / Extranet Link)
DD+DIS386.07E - NX 2.0 - Service Bulletin No. 19 - Software Update NX 2.0.6805 SU2 solves Software Defects (Intranet Link / Extranet Link)
DD+DIS125.08E - CR 30-X Technical Documentation Chapter 2 - Functional Description - Software / Hardware Compatibility Chart CR 30-X System (Intranet Link / Extranet Link)
If the feature "speed class free" operation shall be used, NX workstation software NX 2.0.8000 - NX 3.0.8000 (and higher versions) is required.
For details see:
DD+DIS324.08E - NX 2.0.8000 - 3.0.8000 - Service Bulletin No. 01 - Release of NX 2.0.8000 - 3.0.8000 (Intranet Link / Extranet Link)
DD+DIS298.09E - NX 2.0.8000 / 3.0.8000 Service Bulletin 09, NX 2.0.8000 / 3.0.8000. CR 30-X and CR 30-X SCF Digitizer Model Files Version 1.0 available on MedNet GSO Library (Intranet Link / Extranet Link)
DD+DIS310.09E - CR 30-X, Service Bulletin 34, Introduction of Speed Class free Operation (Intranet Link / Extranet Link)
Revision 0 CR 30-X Page 4 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
SCOPE OF DELIVERY for Software on CD-ROM (part of digitizer delivery):
CD-ROM CRE_1907 with following content: o Software CRE_1907 (Setup CR 30-X_1907.msi) o Directory "Prerequisites" with following content
(required in exceptional cases only):
- Microsoft Visual J # 2.0 Redistributable Package (vjredist.exe).
- Microsoft .NET Framework 2.0 (dotnetfx.exe)
- Java Runtime Environment 1.5.0 (jre-1_5_0-windows-i586.exe)
o readme.txt describing purpose of files in directory "Prerequisites" o This enclosure document
SCOPE OF DELIVERY for Software on the Agfa HealthCare Library:
Software CRE_1907 (Setup CR 30-X_1907.msi) Microsoft Visual J # 2.0 Redistributable Package (vjredist.exe): Required to install
software CRE_1907 on the service PC to use function Explain Error Code
This enclosure document
REQUIRED TOOLS:
If software is part of a spare part kit, and distributed on CD-ROM:
No tool required for software installation
If software is obtained via the Agfa HealthCare Library:
Service PC, to copy software on USB stick USB stick ( > 25 MB free space)
REQUIRED TIME:
Approximately 60 minutes, without possibly required software upgrade of the NX workstation
Revision 0 CR 30-X Page 5 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 0 CR 30-X Page 6 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
1 Introduction
Software CRE_1907 is the successor of software CRE_1901.
It contains the following improvements compared to software CRE_1901:
Improved self test of Slow Scan Motor Minimizing of the Error Viewer improved Following improvements are applicable for certain dealer configurations only:
Error 3192 does not occur anymore when the TWAIN client is closed Polygon and Photomultiplier high tension are switched off if the scan cycle is
cancelled (e.g. by pressing "Eject" button in the image processing application)
Additional minor changes and improvements
The required steps of software CRE_1907 installation depend on the type of installation:
Initial installation (during activation of the NX workstation software or during NX clean installation procedure) or
Upgrade installation
# Step Initial Installation
Upgrade Installation
See section
1 Check the installed software version.
2
2 For Windows 7 only: Prepare the FireWire Driver
3
3 Uninstall the old software. 4 4 Install CRE_1907 on the Control
PC. 5
5 Install the Agfa device driver 6 6 Upload the new software to the
digitizer. * 7
7 Perform a stall calibration of the fixer motor
8
8 Perform a backup. 9 9 Verify the successful software
installation. 10
* In exceptional cases, if a software mismatch error 2712 appears after connection to the digitizer, an upload of the digitizer software is required for initial installation, too.
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2 Checking installed Digitizer Software Version
Initial installation This step is required for:
Upgrade Installation
(1) Log-off the current user at the control PC.
(2) Login as user crservice. (3) Select menu: (4) Enter user name and password.
(5) Select: menu . See Figure 1.
Figure 1
(6) Evaluate as follows:
If it shows: or a lower version proceed with SW installation as described in section 3.
If it shows: stop the software upgrade: The software is up-to-date already.
(7) Close the service client.
Revision 0 CR 30-X Page 7 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3 For Windows 7 only: Preparing the FireWire Driver
Initial installation This step is required for:
Upgrade Installation
IMPORTANT:
Before starting any software installation activity at the control PC, switch off the digitizer. Otherwise the digitizer driver on the control PC might not be installed properly.
NOTE:
Windows 7 uses by default a driver for the FireWire interface which is not compatible with the digitizer.
The Legacy type of FireWire interface has to be activated. Otherwise image transmission will not work: The operating system of the Control PC will crash.
(1) Switch-off the digitizer.
(2) Select to log-off the current user. (3) When the log-in screen appears login as "crservice".
(4) Exit the NX software via: (5) Exit the "Error Viewer" (The icon
is shown in the task bar as long as the error viewer is active).
(6) Right-click on in the START menu.
(7) Select . See Figure 2.
Figure 2
Revision 0 CR 30-X Page 8 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(8) Select .
(9) Check the driver name under IEEE 1394 Bus host Controller:
If it is named with expression Legacy, e.g. LSI 1394 OHCI Compliant Host Controller (Legacy), the correct driver is installed already. In this case close the Computer management window and continue with section 4.
If the expression Legacy is not part of the driver name continue with the next step in this section.
Figure 3
(10) Double-click the driver (marked red in Figure 3).
(11) Select the Driver tab. See Figure 4.
(12) Select . (13) Select .
Figure 4
Revision 0 CR 30-X Page 9 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(14) Select .
Figure 5
(15) Select the Legacy driver. (16) Click : The driver gets
installed
(17) Click (2 times).
Figure 6
(18) Ensure that the Legacy driver is shown in the Computer Management window.
(19) Reboot the NX workstation.
Figure 7
Revision 0 CR 30-X Page 10 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
4 De-installing old Software
Initial installation This step is required for:
Upgrade Installation
(1) Select to log-off the current user. (2) When the log-in screen appears login as "crservice".
(3) Exit the NX software via: (4) Exit the "Error Viewer" (The icon
is shown in the task bar as long as the error viewer is active).
Figure 8
(5) Switch off the digitizer. See IMPORTANT note below.
IMPORTANT:
Before starting any software de-installation activity at the control PC, switch off the digitizer. Otherwise the driver for the digitizer on the control PC will automatically be de-installed.
(6) Deinstall the digitizer software on the control PC: Select:
(7) If the message "You must restart your system ...." appears at the end of the deinstallation process, continue with the steps described in section 11.1. Otherwise continue with next section.
Revision 0 CR 30-X Page 11 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
5 Installing Software CRE_1907 at the Control PC
Initial installation This step is required for:
Upgrade Installation
NOTE:
For the initial installation of the NX workstation steps (1) and (2) are already done by production.
(1) Copy file Setup CR 30-X_1907.msi from USB stick or CD-ROM to partition D: on the control PC.
(2) Double-click on file Setup CR 30-X_1907.msi on the D: partition. (3) Press next and follow the installation wizard. See IMPORTANT note below.
IMPORTANT:
Do not change the suggested default installation folder: C:\Program Files\Agfa\CR 30-X\
Figure 9
(4) When finished, restart the control PC.
(5) Login again as user crservice.
Revision 0 CR 30-X Page 12 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
6 Installing the AGFA Device Driver
Initial installation This step is required for:
Upgrade Installation
(1) Switch on the digitizer: The Found new hardware wizard appears
(2) Enter a Friendly Name for the new hardware, e.g. AGFA Device.
(3) Click .
Figure 10
IMPORTANT:
If the Found new hardware wizard does not appear during initial installation of the software, the driver for the digitizer possibly is not installed properly. Refer to section 11.3 for an instruction how to check and correct the problem.
(4) Click .
Figure 11
Revision 0 CR 30-X Page 13 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
7 Uploading Software to the Digitizer
Initial installation* This step is required for:
Upgrade Installation * In exceptional cases, if a software mismatch error 2712 appears after connection to the digitizer, an upload of the digitizer software is required for initial installation, too.
(1) Exit the NX software via "NX Main Menu Exit NX".
(2) Switch on the digitizer: After approximately 45 seconds, Info message 2715 synchronizing database followed by error 2712 will be displayed.
The Software Version loaded onto the digitizerand that one installed on PC do not match.Please perform the upload via Service Client.
(Error 2712)
x
Figure 12
(3) Close the error screen by clicking on the icon x .
(4) On the control PC select menu:
(5) Enter user name and password.
(6) Select the menu . (7) Ensure that Version on PC CRE_1907 for all 3 nodes is displayed,
see Figure 13.
(8) Select the button . See IMPORTANT note below.
IMPORTANT:
Do not upload the nodes individually, but use the functionality, to markall three nodes for upload automatically. Reason: The software for the PMI board has to be uploaded first before the software for the two other boards is uploaded.
(9) Select button . See Figure 13.
Get installed version Upload selected nodes
Select All
Clear Selection
Node Installed Version Version on PC Status
517501df.cdr
unknown/unknownunknownunknown
CRE_1907CRE_1907
differentPMI Board/FPGAHDL BoardSSC Board
differentdifferent
Figure 13
Revision 0 CR 30-X Page 14 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(10) Wait for approximately 7 minutes until the message "Versions are current" appears in the message box. Refer to Figure 14. If message "SW upload required" appears instead, refer to IMPORTANT note next page.
Get installed version Upload selected nodes
Select All
Clear Selection
CRE_1907/2104 CRE_1907/2104CRE_1907 CRE_1907CRE_1907 CRE_1907
okokok
NodePMI Board/FPGAHDL BoardSSC Board
Installed Version Version on PC Status
Activity
Final SW upload success message
Upload to selected nodes completedWaiting for PMI to restart...Finalizing Upload.Waiting for restart...Retrieving actual version information.Starting version check of nodesPMI: installed version 'CRE_1907/1096'; file version 'CRE_1907/2104Controller HDL Board(0x780): installed version 'CRE_1907'; file version 'CRE_1907'Controller SSC Board(0x784): installed version 'CRE_1907'; file version 'CRE_1907'
Versions are current
Figure 14
(11) Switch off the digitizer.
Revision 0 CR 30-X Page 15 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
IMPORTANT:
After upload of the selected boards, the PMI board is restarted. In some cases the version check of the nodes fails like in this example for the PMI2 board: Starting version check of nodes PMI: installed version 'unknown/unknown'; file version 'CRE_1907/2104' No controller with CAN-ID 0x780 found in machine Controller HDL Board(0x780): installed version 'unknown'; file version 'CRE_1907' No controller with CAN-ID 0x784 found in machine Controller SSC Board(0x784): installed version 'unknown'; file version 'CRE_1907' SW upload required.
Proceeding to fix the problem: (1) Press button "Get installed versions to trigger the versopn check once more.
Message "Versions are current" appears (see Figure 14).
(2) If message "Versions are current" does not appear: Repeat the software upload: Go back to step (8).
8 Performing a Stall Calibration for the Fixer Motor
Initial installation* This step is required for:
Upgrade Installation * during initial installation (= activation of the NX workstation software or during NX clean installation procedure) this step only has to be performed, if the software on the digitizer was below CRE_1907.
NOTE:
Together with software CRE_1907 a new type of fixer motor was introduced with different motor characteristics.
Software CRE_1907 in general supports all types of fixer motors. To ensure that the previous type of fixer motor runs at optimal speed a stall calibration of the fixer motor is required however.
If the stall calibration is not performed during software upgrade, throughput would be slightly decreased, as some movements of the fixer motor would be slower.
(1) Switch on the digitizer.
(2) In the CR 30-X Service Client select "Preventive Maintenance - Stall Calibration".
(3) Select "Calibrate Fixer".
(4) Wait until the stall calibration is finished: This takes approximately 1 minute.
Revision 0 CR 30-X Page 16 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
9 Performing a Backup
Initial installation This step is required for:
Upgrade Installation
(1) In the CR 30-X service client select the menu .
(2) Keep the default settings as shown in Figure 15.
Figure 15
(3) Select the button to perform a backup.
(4) Select the partition of the control PC as storage place.
(5) Select : The screen "Performing backup ..." with a progress bar appears.
(6) Wait till the backup is finished.
(7) Insert a backup USB stick at the USB port of the control PC.
(8) Copy the backup file stored on the partition on the control PC to the USB stick.
(9) Unmount and remove the USB stick from the control PC and store it on a save place.
(10) Close the service client.
Revision 0 CR 30-X Page 17 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(11) Answer the dialogue popping up did you perform any repair or maintenance activities with "Yes".
Figure 16
10 Verifying the successful Software Upgrade
Initial installation* This step is required for:
Upgrade Installation * These verification steps are also part of the digitizer installation instructions.
(1) Insert cassette with unexposed IP in the digitizer.
(2) At the NX workstation select examination type System diagnosis Flatfield.
(3) Select ID to identify the image: Scanning starts.
(4) Confirm that the image arrives at the NX workstation.
(5) Reject the image.
(6) Log-off user crservice.
Result
The digitizer software CRE_1907 is installed.
Revision 0 CR 30-X Page 18 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
11 Appendix: Troubleshooting Software Installation Problems at the Control PC
The following problems may occur during software installation at the control PC:
# Symptom See Section:
1 After uninstallation of the Software the Message appears: You must restart your system
11.1
2 After uninstallation of the Software the "Found New Hardware Wizard" appears
11.2
3 After initial installation the "Found New Hardware Wizard" does not appear: The connection between digitizer and control PC cannot be established
11.3
11.1 After Uninstallation of the Software the Message appears: You must restart your system
Symptom After uninstallation of the old digitizer software on the control PC following message pops-up:
Figure 17
Cause The NX software or the error viewer have not been closed before software de-installation.
Solution Restart the control PC. Continue with software installation.
See section 5.
Revision 0 CR 30-X Page 19 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
11.2 After Uninstallation of the Software the "Found New Hardware Wizard" appears
Symptom After uninstallation of the Software followed by a reboot of the control PC the screen "Found New Hardware Wizard" appears.
Figure 18
Cause Digitizer was not switched off during software uninstallation. In this case the digitizer driver on the control PC will also be uninstalled.
Solution (1) Cancel the "Found new Hardware Wizard" and continue with software installation. See section 5.
(2) Install the driver via "found new hardware wizard" which pops up automatically after software upload to the digitizer (and restart of the control PC).
A
C D
B
Figure 19
Revision 0 CR 30-X Page 20 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
DD+DIS112.11E Enclosure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
11.3 After initial Installation the "Found New Hardware Wizard" does not appear
Symptom After initial installation the "Found New Hardware Wizard" does not appear. The connection between digitizer and control PC cannot be established: The status LED at the digitizer stays red blinking.
Cause Device driver for the digitizer not installed properly.
Solution (1) At control PC select: Start - Control Panel - Tab "Hardware" - Device Manager.
(2) Check whether Imaging device PMI or Agfa or CR30 is available.
Result: A)
Imaging devicesPMI
Imaging device PMI or Agfa or CR30 available: This is the normal situation.
B)
??Other devices?? PMI!
PMI or Agfa or CR30 listed under other devices. In this case the driver is not properly installed. Proceed with next step.
(3) This step has only to be performed for result B), PMI or Agfa or CR30 listed under other devices: - Right-click on the driver named PMI or Agfa or CR30. - Select uninstall. - Be sure to have digitizer switched on and the FireWire cable connected. - Reboot the control PC. - Re-install the digitizer driver as described in Figure 19.
Revision 0 CR 30-X Page 21 of 21 05-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
1 Introduction2 Checking installed Digitizer Software Version3 For Windows 7 only: Preparing the FireWire Driver4 De-installing old Software5 Installing Software CRE_1907 at the Control PC6 Installing the AGFA Device Driver7 Uploading Software to the Digitizer8 Performing a Stall Calibration for the Fixer Motor9 Performing a Backup10 Verifying the successful Software Upgrade11 Appendix: Troubleshooting Software Installation Problems at the Control PC11.1 After Uninstallation of the Software the Message appears: You must restart your system 11.2 After Uninstallation of the Software the "Found New Hardware Wizard" appears11.3 After initial Installation the "Found New Hardware Wizard" does not appear