Installation Instruction Replacing cover and O-Ring Promag 10, 23, 50, 51, 53, 55, Promass 40, 80, 83, 84, 8ME, CNGmass 8FF, LPGmass 8FE, Cubemass 8CM, Prosonic Flow 90, 91, 92, 93, t-mass 65, Prowirl 72, 73, Promag 400, Promag 800 EA00098D/06/A2/14.12 71191696 Instruction is valid for the following spare part sets: ! Note! The order number of the spare part set (on the packaging label) can differ from the product number (on the label directly on the spare part)! The order number of the relevant spare part set can be found by entering the product number of the spare part in the spare parts finder. We recommend that the Installation Instructions be kept with the packaging at all times. Order number Device Component Order number Device Component 50094026 Kit cover with window 1 × cover with screen complete 1 × O-ring 123.4 × 3.53 NBR 50097145 Kit cover with window 1 × cover with screen O-ring 113.90 × 3.53 NBR 50094027 Kit cover with window Exd 1 × cover complete with screen Exd 1 × O-ring 123.4 × 3.53 NBR 50097146 Kit 10 gaskets cover 10 × O-ring 113.90 × 3.53 NBR 10 × O-ring 73.00 × 3.00 NBR 50094028 Kit cover blind version 1 × cover 1 × O-ring 123.4 × 3.53 NBR 50099100 Kit cover with window stainless steel housing 1 × cover complete with screen 1 × gasket to cover D11, Silicon 50094029 Kit cover blind version Exd 1 × cover Exd 1 × O-ring 123.4 × 3.53 NBR 50103526 Kit cover with window 1 × cover with screen 1 × O-ring 113.90 × 3.53 NBR 50094038 Kit cover connection terminal 1 × cover 1 × O-ring 113.90 × 3.53 NBR 50103527 Kit cover blind version 1 × cover 1 × O-ring 113.90 × 3.53 NBR 50094039 Kit cover terminal connection Exe 1 × cover Ex 1 × O-Ring 113.90 × 3.53 NBR 50103883 Kit cover remote sensor 1 × cover Exd 1 × O-ring 73.00 × 3.00 NBR 50094040 Kit cover connection terminal Exd 1 × cover Exd 1 × O-Ring 113.90 × 3.53 NBR 50104746 Kit cover with window 1 × cover with screen 1 × O-ring 113.90 × 3.53 NBR 50094041 Kit seal cover to wall enclosure 1 × O-ring 205.00 × 3.53 NBR 1 × O-ring 180.00 × 3.53 NBR 2 × rubber buffer to D09 50104747 Kit cover blind version 1 × cover with screen 1 × O-ring 113.90 × 3.53 NBR 50094050 Kit 10 seal cover to field housing 10 × O-ring 123.4 × 3.53 NBR 10 × O-ring 113.90 × 3.53 NBR 50106386 Kit cover connection compartment Exd 1 × cover Exd 1 × O-ring 73.00 × 3.00 NBR 50097143 Kit cover blind version 1 × cover 1 × O-ring 113.90 × 3.53 NBR 71034612 Kit cover version blind Exd 1 × cover Exd 1 × O-ring 113.90 × 3.53 NBR 50097144 Kit cover terminal connection 1 × cover Exd 1 × O-ring 73.00 × 3.00 NBR

Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

  • Upload
    dolien

  • View
    221

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Installation Instruction

Replacing cover and O-RingPromag 10, 23, 50, 51, 53, 55, Promass 40, 80, 83, 84, 8ME, CNGmass 8FF, LPGmass 8FE, Cubemass 8CM, Prosonic Flow 90, 91, 92, 93, t-mass 65, Prowirl 72, 73, Promag 400, Promag 800

EA00098D/06/A2/14.12

71191696

Instruction is valid for the following spare part sets:

! Note!

• The order number of the spare part set (on the packaging label) can differ from the product number (on the label directly on the spare part)!

• The order number of the relevant spare part set can be found by entering the product number of the spare part in the spare parts finder.

• We recommend that the Installation Instructions be kept with the packaging at all times.

Order number Device Component Order number Device Component

50094026

Kit cover with window

1 × cover with screen complete

1 × O-ring 123.4 × 3.53 NBR

50097145

Kit cover with window

1 × cover with screen

O-ring 113.90 × 3.53 NBR

50094027

Kit cover with window Exd

1 × cover complete with screen Exd

1 × O-ring 123.4 × 3.53 NBR

50097146

Kit 10 gaskets cover

10 × O-ring 113.90 × 3.53 NBR

10 × O-ring 73.00 × 3.00 NBR

50094028

Kit cover blind version

1 × cover1 × O-ring 123.4 × 3.53 NBR

50099100

Kit cover with window stainless steel housing

1 × cover complete with screen

1 × gasket to cover D11, Silicon

50094029

Kit cover blind version Exd

1 × cover Exd1 × O-ring 123.4 × 3.53 NBR

50103526

Kit cover with window

1 × cover with screen

1 × O-ring 113.90 × 3.53 NBR

50094038

Kit cover connection terminal

1 × cover1 × O-ring 113.90 × 3.53 NBR

50103527

Kit cover blind version

1 × cover

1 × O-ring 113.90 × 3.53 NBR

50094039

Kit cover terminal connection Exe

1 × cover Ex1 × O-Ring 113.90 × 3.53 NBR

50103883

Kit cover remote sensor

1 × cover Exd

1 × O-ring 73.00 × 3.00 NBR

50094040

Kit cover connection terminal Exd

1 × cover Exd1 × O-Ring 113.90 × 3.53 NBR

50104746

Kit cover with window

1 × cover with screen

1 × O-ring 113.90 × 3.53 NBR

50094041

Kit seal cover to wall enclosure

1 × O-ring 205.00 × 3.53 NBR

1 × O-ring 180.00 × 3.53 NBR

2 × rubber buffer to D09

50104747

Kit cover blind version

1 × cover with screen

1 × O-ring 113.90 × 3.53 NBR

50094050

Kit 10 seal cover to field housing

10 × O-ring 123.4 × 3.53 NBR

10 × O-ring 113.90 × 3.53 NBR

50106386

Kit cover connection compartment Exd

1 × cover Exd

1 × O-ring 73.00 × 3.00 NBR

50097143

Kit cover blind version

1 × cover

1 × O-ring 113.90 × 3.53 NBR

71034612

Kit cover version blind Exd

1 × cover Exd

1 × O-ring 113.90 × 3.53 NBR

50097144

Kit cover terminal connection1 × cover Exd

1 × O-ring 73.00 × 3.00 NBR

Page 2: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

2 Endress + Hauser

Confirmation whether the spare part is permitted to be used with the measuring device

The spare parts set and Installation Instructions are used to replace a faulty unit with a functioning unit of the same type. Use genuine parts from

Endress+Hauser only. Only original spare parts supplied by Endress+Hauser shall be used with the measuring device. Therefore, before use, check

whether the spare part set is compatible with the measuring device.

A spare parts overview label is located in the connection compartment cover of the measuring device. If there is no label or the spare part is not

listed the appropriate spare part can also be identified via W@M Device Viewer.

How carrying out such a check via the W@M Device Viewer is described below:

A0016264

If you have any questions, contact your Endress+Hauser service organization.

Overview of the personnel authorized to carry out repairs

Authorization to carry out a repair depends on the approval of the measuring device. The table shows the respective group of persons for each.

! Note!

The person who carries out the repair is responsible for safety during the work, the quality of work completed and safety of the device after repair.

www.endress.com/deviceviewer

Device Viewer

Ser. No.:12345i

i

Ser.No.:1234567890Ser. No.:12345

= ?

Choose the Endress+Hauser Device Viewer via web browser: www.endress.com/deviceviewer

Enter the serial number (Ser. No.) of the device into the W@M Device Viewer (on the label of the transmitter), then click on "Spare parts".

The list of the available spare parts for the device is displayed.

Check the order number on the packaging label of the spare part set.

Check whether the order number of the spare part set is specified in the list displayed by the device viewer:

= YES, the spare part can be used.

= NO, the spare part may not be used.

Approval of the measuring device Group of persons authorized to carry out repairs

without approval 1, 2, 3

with approval (for Ex. IECEx, ATEX, FM, CSA, TIIS, NEPSI) 1*, 2, 3

1 Trained customer technician

1*Trained customer technician, trained by Endress+Hauser (for repairs carried out on devices with type approval)

2 Service technician authorized by Endress+Hauser

3 Endress+Hauser (send measuring device back to manufacturer)

Page 3: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

Endress + Hauser 3

Safety instructions

• Check whether the spare part matches the identification label on the

measuring device, as explained on the first page.

• The spare parts set and Installation Instructions are used to replace a

faulty unit with a functioning unit of the same type. Use genuine parts from Endress+Hauser only.

• In the case of Ex-certified measuring devices: Only open in a de-

energized state (once a delay of 10 minutes has elapsed after

switching off the power supply) or in environments which do not

have a potentially explosive atmosphere.

• The measuring device is energized. Danger: Risk of electric shock!

Open the measuring device in a de-energized state only.

• Before removing the device: set the process in a safe condition and

purge the pipe of dangerous materials.

• Hot surfaces! Risk of injury! Before commencing work, allow the

system and measuring device to cool down to a touchable

temperature.

• Comply with national regulations governing mounting, electrical

installation, commissioning, maintenance and repair procedures.

• Requirements with regard to specialized technical staff for the

mounting, electrical installation, commissioning, maintenance and

repair of the measuring devices:

– trained in instrument safety

– familiar with the individual operation conditions of the devices

– for Ex-certified measuring devices: also trained in explosion

protection

• Follow the Operating Instructions for the device.

• Risk of damaging electronic components! Ensure you have a working

environment protected from electrostatic discharge.

• Modifications to the measuring device are not permitted.

• In the case of measuring devices in safety-related applications in

accordance with IEC 61508 or IEC 61511: After repair recommission

in accordance with Operating Instructions. Document the repair

procedure.

• Only open housing for a brief period. Avoid the penetration of foreign

bodies, moisture or contaminants.

• Replace defective seal/gaskets with genuine parts from

Endress+Hauser only.

• If threads are damaged or defective, the measuring device must be

repaired.

• Threads (e.g. of the cover for the electronics and connection

compartments) must be lubricated. Use an acid-free, non-hardening

grease if an abrasion resistant dry lubricant is non-existent.

• If spacing is reduced or the dielectric strength of the measuring device

cannot be guaranteed during repair work, perform a test on

completion of the work (e.g. high-voltage test in accordance with the manufacturer's

instructions).

• Service connector:

–do not connect in potentially explosive atmospheres.

–only connect to Endress+Hauser service devices.

• Observe the instructions for transporting and returning the device

outlined in the Operating Instructions.

• If you have any questions, contact your E+H service organization.

Tool List

3 mm PH2 T20

acid-free,

non-hardening

grease

pen/ tool

Ø < 2 mm

(0.08 inch)

Page 4: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

4 Endress + Hauser

Field housing Compact Version

A Replacing the cover, O-ring Promass 40, 80, 83, 84, 8ME, Cubemass 8CM, Promag 50, 51, 53, 55, t-mass 65, Prosonic Flow 93

B Replacing the cover, O-ring Promag 10, Prosonic Flow 91

C Replacing the cover, O-ring CNGmass 8FF, LPGmass 8FE

D Replacing the cover, O-ring Promag 23

1 2 3

A0016292 A0012704 A0012720

1 2

A0012712 A00127123

1 2

A0012026 A0012027

1 2 3

A0012707 A00127078 A0012709

3 mm

3 mm

Page 5: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

Endress + Hauser 5

Field housing Remote Version

E Replacing the cover and O-ringPromag 10, 50, 51, 53, 55, Promass 80, 83, t-mass 65, Promag 400, Promag 800

Field housing Compact Version and Remote Version

F Replacing the cover, O-ringProwirl 72, 73, Prosonic Flow 92

1 2

A0018159 A0018160

1 2 3 4

A0012506 A0012715 A0012716 A0012506

5 6

A0012718 A0012719

3 mm

3 mm

Page 6: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

6 Endress + Hauser

Stainless Steel housing

G Replacing the cover and gasket Promass 80, 83, 84, Promag 50, 53, 55

Wall mounted housing

H Replacing the O-Ring and rubber bufferPromass 80, 83, 84, Promag 50, 53, 55, t-mass 65, Prosonic Flow 90U, 93

I Re-assembly

Re-assembly is carried out in reverse order, unless otherwise instructed.The following must be noted.

1 2

A0011504 A0012714

1 2

A0012710 A0012711

6 mm

PH2

1

A0011687

Page 7: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

Endress + Hauser 7

Page 8: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Replacing cover and O-Ring

8 Endress + Hauser

Page 9: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Einbauanleitung

Austausch Deckel und O-RingPromag 10, 23, 50, 51, 53, 55, Promass 40, 80, 83, 84, 8ME, CNGmass 8FF, LPGmass 8FE, Cubemass 8CM, Prosonic Flow 90, 91, 92, 93, t-mass 65, Prowirl 72, 73, Promag 400, Promag 800

EA00098D/06/A2/14.12

71191696

Die Einbauanleitung ist für folgende Ersatzteilsets gültig:

! Hinweis!

• Die Bestellnummer des Ersatzteilsets (auf dem Produktaufkleber der Verpackung) kann sich von der Produktionsnummer (auf dem Aufkleber

direkt auf dem Ersatzteil) unterscheiden!

• Durch Eingabe der Produktionsnummer des Ersatzteiles im Ersatzteilfindetool kann die Bestellnummer des ensprechenden Ersatzteilsets ermittelt werden.

• Wir empfehlen Einbauanleitung und Verpackung immer zusammen aufzubewahren.

Bestellnummer Gerätekomponente Bestellnummer Gerätekomponente

50094026

Set Deckel mit Fenster

1 × Deckel

1 × O-Ring 123,4 × 3,53 NBR

50097145

Set Deckel mit Fenster

1 × Deckel mit Glas

O-Ring 113,90 × 3,53 NBR

50094027

Set Deckel mit Fenster Exd

1 × Deckel mit Glas komplett Exd

1 × O-Ring 123,4 × 3,53 NBR

50097146

Set 10 Deckeldichtungen

10 × O-Ring 113,90 × 3,53 NBR

10 × O-Ring 73,00 × 3,00 NBR

50094028

Set Deckel Blindversion

1 × Deckel

1 × O-Ring 123,4 × 3,53 NBR

50099100

Set Deckel mit Fenster Edelstahlgehäuse

1 × Deckel komplett mit Glas

1 × Dichtung zu Deckel D11, Silikon

50094029

Set Deckel Blindversion Exd

1 × Deckel Exd

1 × O-Ring 123,4 × 3,53 NBR

50103526

Set Deckel mit Fenster

1 × Deckel mit Glas

1 × O-Ring 113,90 × 3,53 NBR

50094038

Set Deckel Anschlussklemme

1 × Deckel1 × O-Ring 113,90 × 3,53 NBR

50103527

Set Deckel Blindversion

1 × Deckel

1 × O-Ring 113,90 × 3,53 NBR

50094039

Set Deckel Anschlußklemme Exe

1 × Deckel Ex

1 × O-Ring 113,90 × 3,53 NBR

50103883

Set Deckel Aufnehmer Getrenntausführung

1 × Deckel Exd

1 × O-Ring 73,00 × 3,00 NBR

50094040

Set Deckel Anschlußklemme Exd

1 × Deckel Exd

1 × O-Ring 113,90 × 3,53 NBR

50104746

Set Deckel mit Fenster

1 × Deckel mit Glas

1 × O-Ring 113,90 × 3,53 NBR

50094041

Set Deckeldichtungen Wandgehäuse

1 × O-Ring 205,00 × 3,53 NBR

1 × O-Ring 180,00 × 3,53 NB

2 × Gummi Puffer zu D09

50104747

Set Deckel Blindversion

1 × Deckel mit Glas

1 × O-Ring 113,90 × 3,53 NBR

50094050

Set 10 Deckeldichtungen Feldgehäuse

10 × O-Ring 123,4 × 3,53 NBR

10 × O-Ring 113,90 × 3,53 NBR

50106386

Set Deckel Anschlussraum Exd

1 × Deckel Exd

1 × O-Ring 73,00 × 3,00 NBR

50097143

Set Deckel Blindversion

1 × Deckel Proline ohne Schrift

1 × O-Ring 113,90 × 3,53 NBR

71034612

Set Deckel Blindversion Exd

1 × Deckel Exd

1 × O-Ring 113,90 × 3,53 NBR

50097144

Set Deckel Anschlussklemme

1 × Deckel Exd

1 × O-Ring 73,00 × 3,00 NBR

Page 10: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

2 Endress + Hauser

Überprüfung ob die Verwendung des Ersatzteils für das Messgerät erlaubt ist

Ersatzteilset und Einbauanleitung dienen dazu, eine defekte Einheit gegen eine funktionierende Einheit des gleichen Typs zu ersetzen. Es dürfen nur Originalteile von Endress+Hauser verwendet werden. Grundsätzlich dürfen nur Ersatzteilsets verwendet werden, die von

Endress+Hauser für das Messgerät vorgesehen sind. Vor der Verwendung ist deshalb zu überprüfen, ob das Ersatzteilsets zum Messgerät passt.

Im Anschlussraumdeckel des Messgeräts befindet sich ein Übersichtsschild Ersatzteile. Falls dieses nicht vorhanden oder das Ersatzteilset nicht

aufgelistet ist, kann eine solche Überprüfung via W@M Device Viewer durchgeführt werden.

Wie eine solche Überprüfung via W@M Device Viewer durchgeführt werden kann, ist nachfolgend beschrieben:

A0016264

Bei Fragen kontaktieren Sie bitte Ihre zuständige Endress+Hauser Serviceorganisation.

Übersicht der reparaturberechtigten Personen

Die Berechtigung zur Durchführung einer Reparatur ist von der Zulassung des Messgeräts abhängig. Die Tabelle zeigt den jeweils berechtigten

Personenkreis.

! Hinweis!

Die Person, die eine Reparatur vornimmt, übernimmt die Verantwortung für die Sicherheit während der Arbeiten, die Qualität der Ausführung

und die Sicherheit des Geräts nach der Reparatur.

www.endress.com/deviceviewer

Device Viewer

Ser. No.:12345i

i

Ser.No.:1234567890Ser. No.:12345

= ?

Über einen Webbrowser den Endress+Hauser Device Viewer aufrufen: www.endress.com/deviceviewer

Die Seriennummer (Ser. No.) des Messgeräts (vom Typenschild des Messumformers) in den Device Viewer eingeben, danach auf den Reiter

"Ersatzteile" klicken.

Auf dem Bildschirm wird eine Liste aller zur Verfügung stehenden Ersatzteilsets für das Messgerät angezeigt.

Die Bestellnummer des Ersatzteilsets ermitteln (auf dem Produktaufkleber der Verpackung).

Überprüfen, ob die Bestellnummer des Ersatzteilsets in der Liste der angezeigten Ersatzteile im Device Viewer vorhanden ist:

= JA, das Ersatzteilset darf für das Messgerät verwendet werden.

= NEIN, das Ersatzteilset darf für das Messgerät nicht verwendet werden.

Zulassung des Messgeräts Reparaturberechtigter Personenkreis

ohne Zulassung 1, 2, 3

mit Zulassung (z.B. IECEx, ATEX, FM, CSA, TIIS, NEPSI) 1*, 2, 3

1 Ausgebildete Fachkraft des Kunden

1*Ausgebildete Fachkraft des Kunden, von Endress+Hauser geschult (für Reparaturen, die an Geräten mit Bauartzulassung vorgenommen werden)

2 Von Endress+Hauser autorisierter Servietechniker

3 Endress+Hauser (Messgerät an Hersteller zurücksenden)

Page 11: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

Endress + Hauser 3

Sicherheitshinweise

• Prüfen, ob das vorliegende Ersatzteil zur Kennzeichnung auf dem

Messgerät passt, wie auf der Titelseite beschrieben.

• Ersatzteilset und Einbauanleitung dienen dazu, eine defekte Einheit

gegen eine funktionierende Einheit des gleichen Typs zu ersetzen. Nur Originalteile von Endress+Hauser verwenden.

• Bei Ex-zertifizierten Messgeräten: Nur in spannungslosem Zustand

(nach Berücksichtigung einer Wartezeit von 10 Minuten nach Ab-

schalten der Energiezufuhr) oder in Umgebungen öffnen, die keine

explosionsfähige Atmosphäre enthalten.

• Messgerät unter Spannung! Lebensgefahr durch Stromschlag.

Messgerät nur im spannungslosen Zustand öffnen.

• Vor einem Geräteausbau: Prozess in sicheren Zustand bringen und

Leitung von gefährlichen Prozessstoffen befreien.

• Verbrennungsgefahr durch heiße Oberflächen! Vor Arbeitsbeginn:

Anlage und Messgerät auf berührungssichere Temperatur abkühlen.

• Nationale Vorschriften bezüglich der Montage, elektrischen

Installation, Inbetriebnahme, Wartung und Reparatur einhalten.

• Folgende Anforderungen an das Fachpersonal für Montage,

elektrische Installation, Inbetriebnahme, Wartung und Reparatur der

Messgeräte müssen erfüllt sein:

– In Gerätesicherheit ausgebildet

–Mit den jeweiligen Einsatzbedingungen der Geräte vertraut

–Bei Ex-zertifizierten Messgeräten zusätzlich im Explosionsschutz

ausgebildet

• Die Betriebsanleitung zum Messgerät ist zu beachten.

• Beschädigungsgefahr elektronischer Bauteile! Eine ESD-geschützte

Arbeitsumgebung herstellen.

• Nach Entfernen der Elektronikabdeckung: Stromschlaggefahr durch

aufgehobenen Berührungsschutz! Messgerät ausschalten, bevor interne Abdeckungen entfernt werden.

• Änderungen am Messgerät sind nicht zulässig.

• Bei Messgeräten in sicherheitstechnischen Applikationen gemäß IEC 61508 bzw. IEC 61511: Nach Reparatur Neuinbetriebnahme

gemäß Betriebsanleitung durchführen. Reparatur dokumentieren.

• Gehäuse nur kurzzeitig öffnen. Eindringen von Fremdkörpern,

Feuchtigkeit oder Verunreinigung vermeiden.

• Defekte Dichtungen nur durch Original-Dichtungen von

Endress+Hauser ersetzen.

• Defekte Gewinde erfordern eine Instandsetzung des Messgeräts.

• Gewinde (z.B. von Elektronikraum-und Anschlussraumdeckel)

müssen geschmiert sein. Säurefreies, nicht härtendes Fett

verwenden, sofern keine abriebfeste Trockenschmierung vorhanden

ist.

• Wenn bei den Reparaturarbeiten Abstände reduziert oder die

Spannungsfestigkeit des Messgeräts nicht sichergestellt werden kann: Prüfung nach Abschluss der Arbeiten durchführen (z.B. Hochspannungstest gemäß Herstellerangaben).

• Servicestecker:

–nicht in explosionsfähiger Atmosphäre anschließen.

–nur an Servicegeräte von Endress+Hauser anschließen.

• Die in der Betriebsanleitung aufgeführten Hinweise zum Transport

und zur Rücksendung beachten.

• Bei Fragen kontaktieren Sie bitte Ihre zuständige Endress+Hauser

Serviceorganisation.

Werkzeugliste

3 mm PH2 T20

Säurefreies,

nicht härtebndes

Fett

Stift/ Werkzeug

Ø < 2 mm

(0,08 inch)

Page 12: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

4 Endress + Hauser

Feldgehäuse Kompaktausführung

A Austausch Deckel, O-Ring Promass 40, 80, 83, 84, 8ME, Cubemass 8CM, Promag 50, 51, 53, 55, t-mass 65, Prosonic Flow 93

B Austausch Deckel, O-Ring Promag 10, Prosonic Flow 91

C Austausch Deckel, O-Ring CNGmass 8FF, LPGmass 8FE

D Austausch Deckel, O-Ring Promag 23

1 2 3

A0016292 A0012704 A0012720

1 2

A0012712 A00127123

1 2

A0012026 A0012027

1 2 3

A0012707 A00127078 A0012709

3 mm

3 mm

Page 13: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

Endress + Hauser 5

Feldgehäuse Getrenntausführung

E Austausch Deckel, O-Ring Promag 10, 50, 51, 53, 55, Promass 80, 83, t-mass 65, Promag 400, Promag 800

F Feldgehäuse Kompaktausführung und Getrenntausführung

G Austausch Deckel, O-Ring Prowirl 72, 73, Prosonic Flow 92

1 2

A0018159 A0018160

1 2 3 4

A0012506 A0012715 A0012716 A0012506

5 6

A0012718 A0012719

3 mm

3 mm

Page 14: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

6 Endress + Hauser

Edelstahlgehäuse

H Austausch Deckel, DichtungPromass 80, 83, 84, Promag 50, 53, 55

Wandgehäuse

I Austausch O-Ring und Gummi-PufferPromass 80, 83, 84, Promag 50, 53, 55, t-mass 65, Prosonic Flow 90U, 93

J Zusammenbau

Der Zusammenbau erfolgt in umgekehrter Reihenfolge, wenn nicht anders angegeben.Folgendes ist zu beachten:

1 2

A0011504 A0012714

1 2

A0012710 A0012711

6 mm

PH2

1

A0011687

Page 15: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

Endress + Hauser 7

Page 16: Installation Instruction Replacing cover and O-Ring · Replacing cover and O-Ring 2 Endress + Hauser Confirmation whether the spare part is permitted to be used with the measuring

Austausch Deckel und O-Ring

8 Endress + Hauser