24
WARNING! STRICT ADHERENCE TO THESE INSTALLATION INSTRUCTIONS IS REQUIRED to promote the safety of those installing this product, as well as that of those ultimately using the lift for its intended purpose. Any deviation from these instructions will void the Limited Warranty that accompanies the product. Additionally, any party installing the product who deviates from the Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all loss, liability or damage, including attorney fees that might arise out of, or in connection with, such deviation. RESIDENTIAL PLATFORM LIFTS RPL400 / RPL600 Installation and Service Manual P/N: 630-00043 Rev B

Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

WARNING! STRICT ADHERENCE TO THESE INSTALLATION INSTRUCTIONS IS REQUIRED to promote the safety of those installingthis product, as well as that of those ultimately using the lift for its intended purpose. Any deviation from these instructions willvoid the Limited Warranty that accompanies the product. Additionally, any party installing the product who deviates from theInstallation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and allloss, liability or damage, including attorney fees that might arise out of, or in connection with, such deviation.

RESIDENTIAL PLATFORM LIFTS RPL400 / RPL600Installation and Service Manual

P/N: 630-00043 Rev B

Page 2: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 2

Table of Contents

This Installation and Service Manual provides step-by-step instructions on how to install a Highlander Residential Platform Lift. Read and understand the entire Manual before beginning to install the lift. Also, ensure that the end-user has a copy of the Owner’s Manual with operating instructions before it is put into service.

If you have questions, concerns or comments, please contact Harmar’s Technical Service Department at 1-866-378-6648 or email [email protected].

CONTENTS

Features ..............................................................................................3Specifications .....................................................................................3Safety & Code Requirements ..........................................................4

Site Requirements ......................................................................4Preliminary Checks ...........................................................................5

Review All Parts ..........................................................................5Tools Needed ..............................................................................5Materials Needed .......................................................................5

Preparing to Install ...........................................................................6Installing the Platform .....................................................................7

Controller Harness Connections ..............................................8Installing Outer Guard Panel ........................................................10Installing the Ramp ........................................................................11

Top Landing Gate (Optional)..................................................12Top Landing EMI Interlock (Optional) ..................................13Fascia Panel (Optional) ...........................................................14

Anchoring the Lift ...........................................................................15Setting the Limit Switches .............................................................17Verifying Lift Operation .................................................................18Manual Override .............................................................................19

Page 3: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 3

SPECIFICATIONS

Features

Payload Capacity: 600 lbs.Vertical Travel Height: RPL400: 53” RPL600: 77” Footprint: 50” x 73”

Platform: 36” x 48”Input Voltage: 115VAC - 15a grounded circuitControl Voltage: 24VACPlatform Speed: 10 ft/min.Motor: ½ HP - 90 VDC

Data / Serial Tag

Top Cover

Platform Controls

Platform Rear Guard Panel

Platform Front Guard Panel

Ramp Roller Guide

Folding Ramp

Safety Pan Assembly

Platform Assembly

Page 4: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 4

Safety

Read all instructions in this Manual before installing or operating the lift.• Always wear eye protection while installing or servicing this product.• Always disconnect this product from the electrical source before servicing it.• Do not wear loose clothing or jewelry when working on this product.• Do not disable any safety equipment or switches supplied with this lift.• Stay away from all drive train components while the lift is operating.• Do not exceed the maximum payload capacity of 600 lbs.• Do not ride on the lift until it is anchored in place.• This product is designed only for lifting people and mobility devices. Do not use for any other purpose.

Code RequirementsThis lift is designed to meet ASME A18.1 section 5 and CSA B44/ASME A17.5 with the addition of certain options. Code requirements for vertical platform lifts vary depending on location. Installers are responsible for contacting their local code enforcement office to determine all applicable regulations prior to installing vertical platform lifts.

Site Requirements• The lift requires a 115VAC 20 amp grounded circuit.• Outdoor installations require a GFI protected circuit.• Only install the lift on a 4” thick, level 3,500 psi reinforced concrete slab.• Footprint requires 49” x 69” area (this includes the folding access ramp.)

Page 5: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 5

Review All Parts• Check the lift for shipping damage. If you find any damage contact the freight carrier to file a damage claim.• Verify the products match that described on the packing list attached to the exterior packaging.

Tools Needed• ½” Hammer Drill• 3/8” Masonry Drill Bit• Appliance Dolly• Hammer• Level• Measuring Tape• Socket Wrench Set• Screwdriver Set• Allen Wrench Set

Materials Needed• 4 Floor AnchorsHarmar recommends securing the lift using our anchor kit. If you purchase your own floor anchors they must use3/8” bolts and have a minimum tensile strength of 6,000 lbs.

Preliminary Checks

Page 6: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 6

Final Site InspectionVerify the surface of the lift will mount to is smooth and level. This surface must be made from 3,500 psi reinforced concrete with a minimum thickness of 4.”

Verify that there is enough space for the lift’s footprint. Include space for the folding access ramp, allowing 50” x 70.”

CAUTION: Verify that the running clearance around the lift complies with any codes for your area.

• The horizontal gap between the edge of the platform and the upper landing must be no less than 3/8”and no greater than ¾.”

• The horizontal gap between the guard panel and any wall or barrier must be no less than 2” and nogreater than 3.”

• The ASME A18.1 Code recommends that a gate or door be installed at the upper landing.• Verify that sufficient head room exists above the lift. The lift will require 6’8” of clearance above

the platform floor when the lift is at the upper landing.

WARNING: The area between the floor where the lift is mounted and the top landing must be covered by a smooth vertical fascia to eliminate pinch points between the platform and landing.

Connecting ElectricityThe lift requires a 115VAC 15 AMP grounded, dedicated electrical outlet within 6 feet of the lift. Plug the lift directly into the power outlet. Do not plug anything else into the same outlet as the lift.

WARNING: Do not ride on the lift until it has been anchored in place.

Preparing to Install

Page 7: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 7

Step 1 The lift will arrive on its back on a pallet. Carefully cut away and remove the shrink wrap to expose the platform, guard panels, ramp, and lift tower.

Step 2 Unpack the platform, guard panels, ramp and smaller hardware box from the lift tower. The mounting bolds, nuts an spacers that secure the platform to the lift carriage are included in the small hardware pack.

Step 3 Carefully tilt the lift tower up and remove from the pallet. Remove the top cover and the front panel from the lift tower by removing the three Phillips head screws at the top. Set the top cover aside. Grasp the loose panel by its sides and raise it up and slide it to the side, freeing it from the lift frame.

Step 4 Plug the platform control box into the Controller Harness that exits the plastic chain on the left tower.

Plug lift tower into power source and, using the rocker switch on the platform control box, carefully lower the platform carriage to the ground.

Note: If the red emergency stop button on the front of the control box is pushed in the lift will not move. Grasp the red button and pull it straight out to disengage the emergency stop.

Installing the Platform

Page 8: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 8

Preparing to Install the LiftFinal Site Inspection Verify the surface the lift will mount to is smooth and level. This surface must be made from 3,500 psi reinforced concrete with a minimum thickness of 4”. Verify that there is enough space for the lifts foot print. Be sure to include space for the folding access ramp. (49” x 69”)

Caution: Verify that the running clearance around the lift complies with any codes for your area.

• The horizontal gap between the edge of the platform and the upper landing must no less than 3/8” and no greater than 3/4”.

• The horizontal gap between the guard panel and any wall or barrier must be no less than 2” and no greater than 3”.

• It is recommended that a gate or door be installed at the upper landing. • Verify that sufficient head room exists above the lift. The lift will require 6ʼ 8” of clearance above

the platform floor when the lift is at the upper landing.

Warning: The area between the floor where the lift is mounted and the top landing must be covered by a smooth vertical fascia. This is to to eliminate any pinch points between the platform and landing.

Connecting to Electricity The lift will require a 115 VAC 20 amp grounded electrical outlet within 6 feet of the lift. Plug the lift directly into the power outlet. Do not plug anything else into same outlet as the lift.

Warning: Do not ride on the lift until it has been anchored in place.

Controller Harness Connections

5

High Voltage Circuit Breaker (AC unit only)

Platform Traveling Cable

Upper Landing Interlock

Upper & Lower Limit Switches

Low Voltage Circuit Breaker

Safety Pan Jumper

Upper & Lower Call‐Sends

Controller Harness Connections

Note: Lift will be shipped with jumper in place.

Figure 1 Figure 2

Figure 3

Installing the Platform (con’t)

Page 9: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 9

Step 5 Align the four platform mounting holes with those on the lift carriage. Insert a ½” x 3” bolt into each of the lower holes. Place the spacers betweenthecarriageflangesandtheplatformarmupperholesandinsert½”x3”bolts.Installthelow-profilelocknutsontheupperboltsand the standard ones on the lower bolts.

Step 6Locate the safety pan harness exiting from the rear of the platform and plug it into the harness at the end of the plastic chain in the lift tower. Secure the platform control harness and the safety harness undertheclipatthetopofthecarriageflange.

Step 7 Remove the safety pan bypass jumper from the front of the electronics box at the top of the lift. Verify that the platform safety pan is operational by pressing up on the pan when the lift is coming down. The lift should stop but be able to move in the up direction.

Installing the Platform (con’t)

Page 10: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 10

Step 1Insert the guard panel posts in the pockets in the platform. The smooth side of the guard panel faces toward the center of the platform. Bolt each post in place with (2) 5/16-18 X 2" and (1) 1/4-20 X 2" hex head bolts. If you have a 90° exit platform, install the end guard panel using the provided bracket and bolt to inner guard panel.

Step 2Install the platform control box on the rear guard panel using screws and nuts through the panel.

Step 3 Install rear panel on platform by installing guard panel posts into the pockets at the rear of the platform and bolt in place.

Step 4Install platform control box to rear guard panel.

Installing the Outer Guard Panel

Page 11: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 11

Step 1Install the ramp roller guide on the side of the lift tower using the screws already installed in the panels.Note: The longer flange will point toward the floor.

Step 2Attach the two ramp pivot tabs to the lower landing sides of the platform. Reinsert the hex bolts, one of which will go through the guard panel post, and tighten.

Step 3Attach the ramp to the pivot tabs using shoulder bolts and locking nuts. Tighten the nuts until they seat against the bolt shoulder.

Step 4Attach the ramp roller arm by bolting it to the underside of the ramp. The ramp roller should align and make contact with the ramp roller guide. The roller should ride up the tower arm, folding the ramp when the platform is raised.

Installing the Ramp

Page 12: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 12

Top Landing Gate (optional)The optional top landing gate is provided with a combination mechanical lock and electric contact (interlock.)

The interlock:• Prevents the lift from running if the gate is open.• Prevents the gate from being opened if the platform is not at

the top landing.• Unlocks when the lift is on the upper limit switch.

WiringA length of multi-conductor wire will need to be run from the bottom of the lift tower up to the landing gate. Consult local codes for type and mounting requirements. After wiring is completed, the wiring harness must be plugged into an appropriate receptacle on the controller.

Wiring Connections

Gate HarnessBlackBlackYellowYellow

Call-Send HarnessBlue

YellowRed

Gate HarnessOrangeGreenBlackBrown

Lift HarnessBlue

WhiteRed

Top Landing Gate (Optional)

A crescent-shaped key is provided to manually unlock the gate during installation. The key is inserted from the back side to lift up on the solenoid that holds the gate locked.

Mount the gate by placing onto the upper landing making sure to align the gate opening with the platform (outer guard rail not shown for clarity.) A number of attachment holes are provided in the threshold portion of the gate for mounting using wood lag screws or concrete anchors as appropriate.

Remove the latch post cover and connect the call-send and interlock wire harnesses.

The vertical posts of the gate must be attached to a supporting structure. The gate is not designed to be freestanding.

Page 13: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 13

WiringA length of multi-conductor wire will need to be run from the bottom of the lift tower up to the interlock. Consult local codes for type and mounting requirements. After wiring is completed, the wiring harness must be plugged into an appropriate receptacle on the controller.

Wiring Connections

EMIBlack (A)Black (B)Yellow (C)Yellow (D)

Lift HarnessOrangeGreenBlackBrown

Top Landing EMI Interlock (optional)The optional EMI Interlock is provided with a combination mechanical lock and electric contact. They are to be used with existing doors.

The interlock:• Prevents the lift from running if the gate is open.• Prevents the door from being opened if the platform is not at

the top landing.• Unlocks when the lift is on the landing limit switch.

EMI Interlock1. Position interlock to door jamb and mark mounting holes2. Fasten interlock to door jamb with #8 wood screws.3. Route 4-conductor interlock cable thru hole in top of interlock

and make wire connections.4. Attach door keeper and emergency key plates to hoistway door.

Top Landing EMI Interlock (Optional)

Page 14: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 14

Fascia Panel (Optional)

Fascia Panel (optional)A fascia panel must be used. It provides a smooth surface for the platform edge to meet in order to prevent any shear or obstruction hazards.

A running clearance of no more than ¾” and no less than 3/8” must be maintain between the edge of the platform and the fascia panel.

Page 15: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 15

Step 1 • Position the lift in its final location• Verify that it is level and perpendicular to its surroundings.• Shim if necessary.

CAUTION: Verify that the running clearance around the lift complies with any codes for your area. The horizontal gap between the edge of the platform and the upper landing must be no less than 3/8” and no greater than 3/4.”

The horizontal gap between the guard panel and any wall or barrier must be no less than 2” and no greater than 3.”

It is recommended that a gate or door be installed at the upper landing.

Verify that sufficient head room exists above the lift. The lift requires 6’8” of clearance above the platform floor when the lift is at the upper landing.

Warning: The area between the floor where the lift is mounted and the top landing must be covered by a smooth vertical fascia. This is to eliminate any pinch points between the platform and landing.

Anchoring the Lift

Step 1 • Position the lift in its final location. • Verify that it is level and perpendicular to its surroundings. • Shim if necessary.

AnchorsHarmar recommends securing the lift using our Anchor Kit. If you purchase you own floor anchors they must use 1/2-13 bolts and have a minimum tensile strength of 6000 LBS.

Warning: All four floor anchors must be installed correctly in accordance to their instructions.

Step 2 Use the liftʼs base as a template. Drill 4 holes into the concrete.

Verify that the holes you have drilled are deep enough to accept the anchors.

Tip: Concrete dust may have settled into the holes you just drilled. Use a shop vacuum to clean out these holes. This will ensure the floor anchors set correctly.

Step 3 Secure the lift in place by tightening the floor anchor bolts.

9

Anchoring the Lift

Step 1 • Position the lift in its final location. • Verify that it is level and perpendicular to its surroundings. • Shim if necessary.

AnchorsHarmar recommends securing the lift using our Anchor Kit. If you purchase you own floor anchors they must use 1/2-13 bolts and have a minimum tensile strength of 6000 LBS.

Warning: All four floor anchors must be installed correctly in accordance to their instructions.

Step 2 Use the liftʼs base as a template. Drill 4 holes into the concrete.

Verify that the holes you have drilled are deep enough to accept the anchors.

Tip: Concrete dust may have settled into the holes you just drilled. Use a shop vacuum to clean out these holes. This will ensure the floor anchors set correctly.

Step 3 Secure the lift in place by tightening the floor anchor bolts.

9

Anchoring the Lift

Page 16: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 16

AnchorsHarmar recommends securing the lift using our Anchor Kit. If you purchase your own floor anchors they must use 3/8” bolts with a minimum tensile strength of 6,000 lbs.

Warning: All floor anchors must be installed correctly in accordance to their instructions.

Step 2 Use the lift’s base as a template. Drill 4 holes into the concrete. Verify that holes you have drilled are deep enough to accept the anchors.

TIP: Concrete dust may have settled into the holes you just drilled. Use a shop vacuum to clean out these holes. This will ensure the floor anchors set correctly.

Step 3Secure the lift in place by tightening the floor anchor bolts.

Anchoring the Lift (con’t)

Page 17: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 17

This lift is equipped with upper and lower limit switches. The vertical location of these switches may be adjusted to fit your application.

Typically, the upper limit switch will need to be adjusted so the platform will stop level with the upper landing. The lower limit will typically not need adjusting.

Step 1Verify the emergency switch is in the ON position. Run the lift in the up direction until the platform floor is level with the upper landing.

Step 2Unplug the lift’s power cord from the wall before going to the next step (AC units) Unplug the power cable from the battery box before going to the next step; (DC units)

WARNING: Moving components can cut and crush. Do not operate the lift if you are in close proximity to any drive components. Be aware that loose clothing or jewelry may catch on moving parts.

Step 3Remove the front cover by tilting it forward and lifting upwards. The bottom of the front cover sets on a pin on either side of the lift frame. Set the cover aside in a safe location where it will not become damaged.

Step 4Loosen the bolts that attaches the upper limit switch assembly. Slide the assembly down the track until the lower switch in the assembly comes in contact with the stop bracket on the lift’s car. You should hear the limit switch click as contact is made.

Setting the Limit SwitchesYour lift is equipped with upper and lower limit switches. The vertical location of these switches may be adjusted to fit your application. Typically the upper limit switch will need to be adjusted so the platform will stop level with the upper landing. The lower limit will typically not need adjusting.

Step 1 Verify the emergency switch is in the ON position. Run the lift in the up direction until the platform floor is level with the upper landing.

1A Unplug the liftʼs power cord from the wall before going to the next step (AC units).

1B Uplug the power cable from the battery box before going to the next step (DC units).

Warning: Moving components can cut and crush. Do not operate the lift if you are in close proximity to any drive components. Be aware that loose clothing or jewelry may catch on moving parts.

Step 2 Remove the screws at the top of the lift that attach the front cover.

Step 3 Remove the front cover by tilting it forward and lifting upwards. The bottom of the front cover sets on a pin on either side of the lift frame. Set the cover aside in a safe location where it will not get damaged.

10

Setting the Limit SwitchesYour lift is equipped with upper and lower limit switches. The vertical location of these switches may be adjusted to fit your application. Typically the upper limit switch will need to be adjusted so the platform will stop level with the upper landing. The lower limit will typically not need adjusting.

Step 1 Verify the emergency switch is in the ON position. Run the lift in the up direction until the platform floor is level with the upper landing.

1A Unplug the liftʼs power cord from the wall before going to the next step (AC units).

1B Uplug the power cable from the battery box before going to the next step (DC units).

Warning: Moving components can cut and crush. Do not operate the lift if you are in close proximity to any drive components. Be aware that loose clothing or jewelry may catch on moving parts.

Step 2 Remove the screws at the top of the lift that attach the front cover.

Step 3 Remove the front cover by tilting it forward and lifting upwards. The bottom of the front cover sets on a pin on either side of the lift frame. Set the cover aside in a safe location where it will not get damaged.

10

Setting the Limit Switches

Page 18: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 18

Step 5Replace the front cover and secure it with the screw you removed in Step 1.

Step 6Plug the lift’s electrical cord into the wall. Verify that the emergency switch is in the ON position.

Step 7Run the lift in the down direction for several inches. Next, run the lift in the up direction. Continue to press the up button until the upper limit switch has caused the lift to stop. Be sure the lift platform stops level with the landing platform.

VERIFYING LIFT OPERATIONCAUTION: Complete the following section before training the customer to use the lift.

Step 1Run the lift up and down for 5 complete cycles. Hold the direction button down and allow the limit switches to stop the lift. At the top, verify that the platform stops level with the upper landing. At the bottom, verify the access ramp unfolds and rests on the ground.

Step 2Verify the Emergency Stop Switch operation. When Emergency Stop is pushed, the lift should not run in either direction. When the switch is pulled back out, the lift should operate normally.

Step 3Verify the operation of the sensor pan underneath the platform. Start with the lift at the top landing. Press up on the sensor pan. While holding the pan in this location, press the down switch on the platform. The lift should not run. Check all sides and middle of the safety pan.

Verifying Lift Operation

Page 19: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

RPL Installation and Service Manual | | www.Harmar.com | 866-351-2776 19

In case of power failure, this lift is equipped with a Manual Override Crank that permits it to be raised or lowered by hand. Before using, check to ensure that the manual system is required.

Verify the Need for Manual OverrideBefore using the Manual Override Crank, try trouble-shooting by verifying the following:• Check that the Emergency Stop Switch is pulled out.• Check that the electrical cord is plugged into the wall.• Check that the building’s circuit breaker has not tripped.• Try to run the lift by pushing both the up and down buttons.

Manual Override ProcedureIf, after Troubleshooting as above, the lift still will not run, complete the following steps:

Step 1 Unplug the lift from the wall. Push in the Emergency Stop Button.

WARNING: Do not service or operate the Manual Override Crank while the lift is connected to electricity.

Step 2Loosen the screws and remove the top cap at the top of the tower.

Step 3Insert the manual hand crank onto the upper nut on the top of the drive screw. Rotate the crank to raise or lower the platform.

WARNING: Never operate the lift while the Manual Override Crank is inserted into the lift.

For your safety and that of the end-user, if you have any questions about installation or service procedures, please contact Harmar’s Technical Service Department at 1-866-378-6648 or email [email protected].

© 2012 Harmar MKT-IM-HIGHLANDER-1214

Manual Override

Page 20: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

1 1

2 2

3 3

4 4

AA

BB

CC

DD

90

V

DC

Moto

r

Up

Top

L

mt

UP

A1

A2

Do

wn

DN

A1

A2

VA

CV

AC

24

Top Interlock

Sole

noid

Fin

al L

mt

CN

C

11

0

Ke

y

Sw

itch

SA

F

A1

A2

Em

erg

ency

Sto

p

SA

F

21

Sa

fe

ty P

an S

witche

s

CN

CC

NC

CN

CC

NC

110 VAC

2

3

8

4

10

11

1

5

DN

52

51

UP

12

DN

22

21

UP

22

21

15

A

mp

C.B

.

3 A

mp

C.B

.

+

-

~ ~

50

W-5

DN

21

UP

51

52

Gate

Closed

67

CN

C

CN

C

Btm

Lm

t

DN

34

UP

43

RE

DB

LK

Sa

fe

ty

Nu

t

CN

O

CN

C

NO

L1

L2

G

4T

erm

ina

l S

trip

C

on

nection

Sw

itch - N

.C

.

Sw

itch - N

.O

. held

closed

Push B

utton

Con

tact - N

.O

.

Con

tact - N

.C

.

Circuit B

reaker

Leg

en

d:

Op

tiona

l E

qu

ipm

ent

Call S

end

Ke

y

Sw

itch

joe.brock
Typewritten Text
SCHEMATIC, RPL AC 44111 REV B
Page 21: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

1 1

2 2

3 3

4 4

AA

BB

CC

DD

C

NO

VAC

VAC

24

110

15 Amp

C.B.

3 Amp

C.B.

50W

-5

90 VDC

Motor

Call Send

SOL

Call Send

NC

NC

NO

NC

Emergency

Stop

Platform

Control

C

NO

CNC

CNC

CNC

CNC

Safety

Nut

Safety Pan

Gate

Closed

Lower

Limit

Upper

Limit

Final

Limit

Tra

vel C

able #4

40

29

Platfo

rm

C

on

trol

#4

92

18

Safety N

ut H

RN

S #4

40

09

Safety Nut HRNS #44011

Ca

ll S

en

d #4

93

00

Gate E

xten

sion

H

RN

S #44

01

4

Gate C

all

Sen

d H

RN

S

#4

401

2

Gate Interlo

ck

HR

NS

#

44

01

3

Lim

it S

witch

H

RN

S #

44

007

joe.brock
Typewritten Text
WIRING DIAGRAM, RPL AC 44111 REV B
Page 22: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

1 1

2 2

3 3

4 4

AA

BB

Up

Top

L

mt

UP

A1

A2

Do

wn

DN

A1

A2

Top Interlock

Solenoid

Fina

l L

mt

CN

C

Ke

y

Sw

itch

SA

F

A1

A2

Em

erg

ency

Sto

p

4T

erm

in

al S

trip

C

on

ne

ctio

n

Sa

fe

ty P

an S

witche

s

CN

CC

NC

CN

CC

NC

2

3

8

4

10

11

1

13

Sw

itch

- N

.C

.

Sw

itch

- N

.O

. h

eld

clo

sed

Pu

sh B

utto

n

Co

nta

ct - N

.O

.

Co

nta

ct - N

.C

.

Circuit B

re

ake

r

Le

ge

nd

:

30

A

mp

3 A

mp

Ga

te

Closed

6

7

CN

C

CN

C

Btm

L

mt

Op

tion

al E

quip

me

nt

CN

C

NO

U2

A1

A2

D2

A1

A2

SA

F

U2

D2

+ +

12

V

DC

BA

TT

ER

Y

CH

AR

GE

R

24

V

DC

--

12

V

DC

Sa

fe

ty N

ut

CN

O

Ca

ll S

en

d

Ke

y

Sw

itch

24

VD

C

Moto

r

DN

UP

DN

UP

RED

DN

UP

BLK

-M

+1

2V

+M

G

PW

M

MO

TO

R

SP

EE

D

CO

NT

RO

L

joe.brock
Typewritten Text
SCHEMATIC, RPL DC 44121 REV C
Page 23: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

1 1

2 2

3 3

4 4

AA

BB

123456789101112

G-

+

+

+

12 V

DC

12 V

DC

25 A

BA

TT

ER

Y

CH

AR

GE

R

24 V

DC

--

CR1 - "UP"

+

-

C

NC

NO

C

NC

NO

CR2 - "DOWN"

+

-

COM

NO NC

24 V

DC

Motor

A1A2

SAFETY

C

NO

+

-

A1A2

50W

-3Ω

3 A

+

-

C

NC

NO

C

NC

NO

A1A2

+

-

COM

NO NC

C

NO

Call Send

SOL

Call Send

NC

NC

NO

NC

Emergency

Stop

Platform

Control

C

NO

CNC

CNC

CNC

CNC

Safety

Nut

Safety Pan

Gate

Closed

Lower

Limit

Upper

Limit

Final

Limit

Travel C

able #44029

Platform

C

ontrol

#49218

Safety N

ut H

RN

S #44009

Safety Nut HRNS #44011

Call S

end #49300

Gate E

xtension H

RN

S #44014

Gate C

all

Send H

RN

S

#44012

Gate Interlock

HR

NS

#44013

Lim

it S

witch H

RN

S #44007

-M +12V+MG

PWM

MOTOR

SPEED

CONTROL

joe.brock
Typewritten Text
WIRING DIAGRAM, RPL DC 44121 REV C
joe.brock
Typewritten Text
joe.brock
Typewritten Text
Page 24: Installation and Service Manual - Harmar · 7/24/2017  · Installation Instructions shall be taken to agree to INDEMNIFY, SAVE AND HOLD HARMLESS the manufacturer from any and all

Highlander Residential Platform LiftInstallation Manual

P/N: 630-00043 Rev B