Upload
vantu
View
216
Download
0
Embed Size (px)
Citation preview
Increased Protein/Peptide Mass Transfer Rates for Improved Resolution and Faster LC
Page 2
• Proteomics studies • LC/MS studies• Combinatorial chemistry analyses• Rapid monitoring for process control• 2D-methods
Applications for Fast HPLC of Large Molecules and Proteins
Page 3
First, A Few Basics on Bio Molecule Separations
• Factors That Optimize Bio Separations• Sample Solubility• Appropriate Pore Size• Bonded Phase• Low pH Mobile Phase• High pH Mobile Phase• Temperature• Gradient
Page 4
Solvent Application Comments 0.05-5% TFA in Water General Effective solubilization of many
samples 6 M Guanidine, buffered at pH 6-8
General Very good for many proteins and peptides
5-80% Acetic Acid or Formic Acid; 0.1-0.5M Perchloric Acid
Peptides
Frequently used to extract peptides from tissues, precipitating many proteins and cellular debris
6 M Urea/ 5% Acetic aid Hydrophobic Peptides, Proteins
Useful for membrane proteins, fragments, aggregating systems
Water-Miscible Organic Solvents: Acetonitrile, Methanol, THF, Dioxane, DMSO; +/- TFA; +/- Water
Hydrophobic Peptides, Polypeptides
Limit injection volume to avoid problems; add water, as possible, to improve volume tolerance; acidify with TFA as required
Optimize SolubilitySelected Sample Solvents and Application for Protein and Peptides
Page 5
Improved Peak Shape for Large Molecules in Solution
Columns: 4.6 x 150 mm, 5 mm Mobile Phase: 60% MeOH: 40% 0.1% TFA Flow Rate: 0.75 mL/min Temperature: RT Detection: UV 282 nm Sample: Tylosin (MW 916)
SB-C3 (80Å) 300SB-C3 (300Å)
• The size of a molecule in solution determines which pore size column is best.• The narrower peak width indicates unrestricted access to the pores.
Pw1/2 = 0.442 Pw1/2 =0.125
O
O
NHOHO
CHO
HO
OHOH
O
OOO
O
O O
O
CH3
CH3CH3
OCH3OCH3
CH3
CH2
CH2
CH2
CH3
CH3
CH3
H3CH3C
H3C
0 5 10Time (min)
0 5 10Time (min)
Page 6
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
0.16
0.18
0.2
300SB-C18 (300Å)SB-C18 (80Å)
PW 1/2
LeuEnke
phalin
M.W. =
556
Angiotensin
II
M.W. =
1046
Insulin
B
M.W. =
3,49
6
CytC
M.W. =
12,32
7
RNase
M.W. =
13,68
4Lys
ozyme
M.W. =
13,90
0
Effect of Pore Size and Molecular Size on Peak Width, Gradient Separations
Wide Pore (300Å) Columns are Needed for Proteins and Polypeptides
•Proper pore size selection results in sharper peaks for large molecules
Page 7
50%
60%
70%
80%
90%
100%
110%
300SB-C8 300SB-C3 300SB-CN
ParvalbuminMyoglobinRNase AInsulinLysozymeCarbonic AnhydraseCalmodulin
Columns: 4.6 x 150 mmMobile Phase: 5 - 40% B in 20 min.
A: 0.1% TFA / WaterB: 0.1% TFA / ACN
Flow Rate: 1 mL / min.Temperature: 60°CSample: 4 µg each protein
25 µL injection
Rel
ativ
e R
ecov
ery
to 3
00S
B-C
18
•Recovery may vary depending on bonded phase.• All 300SB bonded phase generally provide good recovery.
Recovery of Polypeptides from ZORBAX 300SB Columns
Page 8
12,3
4 56
7 89
10
1 24 5
6,7
89
103
1 24 5
68
9 103
7
Ret ent ion Time (min)0 5 10 15 20 25 30 35 40
1 24 5
68
910
37
Comparison Separation of Large Polypeptides on 300SB Bonded Phases
Columns: ZORBAX StableBond 300SB4.6 x 150 mm, 5 mm
Mobile Phase: Linear Gradient, 25- 70% B in 40 minA: 0.1% TFA in WaterB: 0.09% TFA in 80% ACN/20% water
Flow Rate: 1.0 mL/minTemperature: 60°CSample: 3 mg each protein
1. RNase 6. CDR2. Insulin 7. Myoglobin3. Cytochrome C 8. Carbonic Anhydrase4. Lysozyme 9. S-100β5. Parvalbumin 10. S-100α
300SB-C18
300SB-C8
300SB-C3
300SB-CN
• Four different 300SB bonded phases allow selectivity optimization of proteins.
Page 9
Conditions:Column: ZORBAX 300SB, 4.6 x 150 mm, 5 mmMobile Phase: A – 0.1% TFA/water, B – 0.1% TFA/canGradient: 10 - 90 % B in 40 minFlow Rate: 1 mL/minTemperature: 60°C, Sample: 15 µL ( 15 µg ) of peptide in 0.1% TFA/DMSO
300SB-CN
300SB-C8
300SB-C18
Recovery 59%
Recovery 75%
Recovery 89%
Retention Time, (min.)
• Recovery on each bonded phase is not always predictable so evaluate different bonded phases.
Effect of Bonded-Phase Ligand on Recovery of a Synthetic Lipopeptide
Page 10
ZORBAX 300SB HPLC Column Developed Expressly for Bio Separations
Page 11
000336P1.PPT
Column: ZORBAX 300SB-C8, 4.6 x 150mm; Mobile Phase Gradient 0 - 60% in 120 min.A= 0.1% TFA in Water, B= 0.086% TFA in ACN Temp.: 40°C Flow Rate: 1mL/min. Det. UV 210nm Sample: 50 µg of rhGH Tryptic Digest
After 495 mL
After 13680 mL
ZORBAX 300SB Designed for Long Life in Low pH Bio Separations
rhGH Tryptic Digest
Reference 3
Page 12
High pH Can be Used for Separating Hydrophobic or Other Low-Solubility Peptides
Comparison of Aß Peptide RP-HPLC Separations at Low and High pH
Column: ZORBAX 300Extend C18 2.1 x 150 mm, 5 mm
Flow Rate: 0.25 mL/min Sample: 5 µL sample (100 pmol each)
Amyloid ß Sequences:Asp Ala Glu Phe Arg His Asp Ser Gly Tyr Glu Val His His GlnLys Leu17 Val Phe Phe Ala Glu Asp Val Gly Ser Asn Lys Gly AlaIle Ile Gly Leu Met Val Gly Gly38 Val Val40 Ile Ala42 Thr43-COOH
Abs
orba
nce
(210
nm
)TFA Conditions, 25°C
A- 0.1% TFA in waterB- 0.085% TFA in 80%AcN33-45%B in 30 min.
TFA Conditions, 80°CA- 0.1% TFA in waterB- 0.085% TFA in 80%AcN29-41%B in 30 min
.
Aβ(1-38)Aβ(1-40)
Aβ(1-42)Aβ(1-43)
Aβ(1-38) Aβ(1-40)
Aβ(1-43)
Aβ(1-42)
Aβ(1-42/3)Aβ(1-40)
Aβ(1-38)
NH4OH Conditions, 25°CA- 20 mM NH4OH in waterB- 20 mM NH4OH in 80%AcN26-38%B in 30 min.
Reference 1
• High pH and room temperature improve peak shape, recovery and change selectivity.
Page 13
pH Can Change Selectivity in Bio SeparationsComparison of TFA and NH4OH For Peptide RP-HPLC \ ESI-MS
Column: ZORBAX Extend C18, 2.1 x 150 mmFlow rate: 0.25mL/minTemp: 25°C Gradient: 5-60% B in 20 min; LC/MS: Pos. Ion ESI- Vf 70V, Vcap 4.5 kV,
N2- 35 psi, 12L/min, 300°C4 µL (50 ng each peptide);
0
.5E7
1E7
1.5E7
2E7
2.5E7
3E7
m in0 2 4 6 8 10 12 14 16 180
.5E7
1E7
1.5E7
2E7
2.5E7
3E7
TFA Conditions:A- 0.1% TFA in waterB- 0.085% TFA in 80% AcN
TIC
(150
-150
0 m
/z)
0.1% TFA
20mM NH4OH
LLG LHL
LLL-NH2LLL
NH4OH Conditions:A- 20 mM NH4OH in waterB- 20 mM NH4OH in 80% AcN
LLVF
LLVF
LLL-NH2
LLG
LHL
LLL
Page 14
Peptide RP-HPLC/ESI-MS NH4OH Mobile Phase Yields Positive and Negative Ion Spectra
Column: ZORBAX Extend C18 2.1 x 150 mm, 5 mm
Flow rate: 0.25mL/min Temp: 25°C Gradient: 5-60% B in 20 minLC/MS: Pos. Ion ESI- Vf 70V, Vcap 4.5 kV,
N2- 35 psi, 12L/min, 300°CSample: 4µL (50 ng each peptide)
0
.5E7
1E7
1.5E7
2E7
2.5E7
3E7
min0 2 4 6 8 10 12 14 16 180
.5E7
1E7
1.5E7
2E7
2.5E7
3E7
TIC
(150
-150
0 m
/z)
Positive Ions
LHL
LLG
LLL
LLVF
LLL-NH2
Negative Ions
LHL
LLG LLL
LLVFLLL-NH2
• The Extend-C18 column is ideal for LC/MS of proteins and peptides at high pH• Both positive and negative ion MS are possible with NH4OH mobile phase.
Page 15
LC/MS at Low pH on Extend-C18
Column: ZORBAX Extend-C18, 2.1 x 150 mm, 5 mm Flow Rate: 0.2 mL/min Temperature: 35°CMobile Phase Gradient: 15-50% B in 15 min. A: 0.1% TFA in water B: 0.085% TFA in 80% ACN
LC/MS: Pos. Ion ESI – Vf 70V, Vcap 4.5 kV, N2-35 psi, 12L/min., 325°C Sample: Angiotensin I, II, III 2.5 mL (50 pmol each)
• Acidic conditions can not resolve all three Angiotensins.
min0 2.5 5 7.5 10 12.50
1.0E6
2.0E6
3.0E6
4.0E6
5.0E6
12.
050
13.
261
AII+ AIII
AITIC
(200
-150
0 m/
z)
m/z500 1000
0
20
40
60
80
100Max: 10889
649
.5
433
.0
227.1
659.7
844.8
250.9
340.3
1277.
9
1179.
1
1025.
2
720.3
1326.
3
398.0
767.3
458.0
1468.
8
1373.
3
+ 3 + 2
Page 16
LC/MS at High pH on Extend-C18
• At high pH all 3 Angiotensins are resolved and the mass spectrum shows improved spectral clarity.
Column: ZORBAX Extend-C18, 2.1 x 150 mm, 5 mm Flow Rate: 0.2 mL/min Temperature: 35°C Mobile Phase Gradient: 15-50% B in 15 min. A: 10 mM NH4OH in water B: 10 mM NH4OH in 80% ACN
LC/MS: Pos. Ion ESI – Vf 70V, Vcap 4.5 kV, N2-35 psi, 12L/min., 325°C Sample: Angiotensin I, II, III, 2.5 mL (50 pmol each)
Signal x10
m/z500 10000
20
40
60
80
100Max: 367225 6
49.0
433
.0
659
.8
129
6.6 + 1
+ 2
+ 3
min0 2.5 5 7.5 10 12.50
1.0E7
2.0E7
3.0E7
4.0E7
5.0E75.
499 8.
477AII
AITIC
(200
-150
0 m/z)
AIII9.
621
Page 17
Column: ZORBAX 300SB-C18, 4.6 x 150 mm, 5 mm Mobile Phase: A- 0.1%TFA in water, B- 0.09%TFA in acetonitrileGradient: 20-45% B in 35 min Flow Rate: 1 mL/min Sample: 10 µl injection of 5 µg peptide in 6M Urea/5% HOAc
Abs
orba
nce
(210
nm
)
25°CßAP(1-38)
ßAP(1-43)* Recovery <10%
40°C
60°C
Time (min.)0
80°C
5 10 15 20 25 30 35 40
Recovery >70%
Optimize Recovery by Changing TemperatureßAP Peptide Separation
Page 18
Protein and Peptide Retention by RP-HPLC –High Sensitivity to Changes in % Organic:
0.14 0.18 0.22 0.26 0.30 0.34 0.38 0.42
100
10
1
k«
φ
Leucine Enkephalins = 11
Lysozymes = 40
Benzenes = 2.7
log k
log k = log k0 - Sφ
φ = W ater/Organic
Large proteins like Lysozyme (14,000 MW) are more sensitive to changes in mobile phase conc of organic modifier (15X) than small molecules like benzene (78 MW) and 4X more sensitive than leucine enkephalin (600 MW).
Page 19
Longitudinal diffusion
Smaller Particles Improve Mass Transfer Effect on Column Dispersion
Plat
e H
eigh
t H
Linear Velocity u
Eddy Diffusion
Sum Curve: van-Deemter
Resistance to Mass Transfer
H = A + B/u + C uLarge Particle
SmallParticle
The smaller the plate height, the higher the plate number and the greater the chromatographic resolution
u opt
H min
Page 20
Improved Efficiency Over Wide Flow Range with Smaller Particles
0.0000
0.0005
0.0010
0.0015
0.0020
0.0025
0.0030
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
Column: ZORBAX Eclipse XDB-C18 Dimensions: 4.6 x 50/30mmEluent: 85:15 ACN:WaterFlow Rates: 0.05 – 5.0 mL/minTemp: 20°CSample: 1.0μL Octanophenone in
Eluent
Volumetric Flow Rate (mL/min)
HETP (cm
)
5.0μm3.5μm1.8μm
H = A + B/u + Cu
Smaller particle sizes yield flatter curves, minima shift to higher flow rates
Title of Presentation Agilent Restricted
0 10 20 30 40
Changing Gradient Time to AffectRetention (k*) and Resolution
Time (min)
100% B
100% B
100% B
100% B
tg= 40
tg= 20
tg= 10
tg= 5
000995P1.PPT
tg F
S Δ%B Vm
k* ∝
DF = change in volume fraction of B solventS = constantF = flow rate (mL/min.)tg = gradient time (min.)
Vm = column void volume (mL)
0% B
0% B
0% B
0% B
• S ≈ 4–5 for small molecules
• 10 < S < 1000 for peptides and proteins
Page 22
0 5 10 15 20 25 30 35 40 45 50
0 2 4 6 8 10 12 14 16 18 20 24 26 28 30
12.
16 22
0 1 2 4 5 6 8 9 1073
1 23 4 5 6 7
89 10
ZORBAX 300SB-C8 4.6 x 250 mm, 5 µm
Rapid Resolution 4.6 x 150 mm, 3.5 µm
Rapid Resolution 4.6 x 50 mm, 3.5 µm
40 min.
24 min.
9 min.
1. Met-enkephalin2. Leu-enkephalin3. Angiotensin II4. Neurotensin5. RNase6. Insulin (Bov)7. Lysozyme8. Calmodulin9. Myoglobin
10. CarbonicAnhydrase
Conditions:
Mobile Phase: A: 95:5, Water : ACN with 0.1% TFAB: 5:95, Water : ACN with 0.085% TFA
Flow: 1.0 mL/minDetection: 215nmSample: 1-10µg protein (10µL inj.) in
6M Guanidine HCL, pH7.010-60% B in 10 min.
10-60% B in 30 min.
10-60% B in 50 min.
Optimize Resolution: Column Length, Particle Size and Gradient Time
Page 23
300SB-C8, 4.6 x 150 mm, 5µm 300SB-C8, 4.6 x 50 mm, 3.5µm
Mobile Phase: A: 95% Water : 5 % ACN, 0.1% TFAB: 5% Water : 95% ACN, 0.085% TFAGradient: 10-60% B in 30 min.
Flow Rate: 1.0 mL / min.Temperature: Ambient
Sample: 1. Gly-Tyr 6. Leu-Enk2. Val-Tyr-Val 7. Angiotensin II3. [Gln11] Amyloid-β- 8. Kinetensin
Protein Fragm 1-16 9. RNase4. (TYR8) Bradykinin 10. Insulin (Eq.)5. Met-Enk
0 2 4 6 8 10 12 14 16 0 2 4 6 8 10 12 14 16
1
2
3
4 56
7 8
9
101
23
45
6
7 8
9
10
000997P1.PPT
Improving Resolution Using ShortColumn Length ( Vm ) and Particle Size (N)
Page 24
Gradient Resolution and Shorter ColumnsFaster Analyses for Complex Samples: HSA Tryptic Digest
min15 20 25 30 35 40 45 50 55
min3 4 5 6 7 8 9 10
min5 10 15 20 25
2.1x150mm, 5μmP/N 883700-92270 min gradient0.2 mL/min
2.1x50mm, 1.8μmP/N 822700-90210 min gradient0.5mL/min
2.1x50mm, 1.8μmP/N 822700-90230 min gradient0.5mL/min
120 peaks60 mins
125 peaks10 mins!
156 peaks!25 mins
Conditions: M obile Phase A: W ater w/ 0.1% TFA, B: ACN w/0.1% TFA, Gradient 2%B to 50%B, Temp: 50°C UV 214 nm
Page 25
Particle Size - More Peak Capacity with 1.8 um RRHT Columns - Peptide Map of BSA
673
502
Peak CapacityColumn used
SB-C18, 2.1x150mm, 3.5µ
SB-C18, 2.1x150mm, 1.8µ
min14.5 15 15.5 16 16.5 17 17.5 18 18.5 19
mAU
-10
0
10
20
30
40
50
min15 16 17 18 19
mAU
-20
0
20
40
60
Less peak capacity, less resolution
35% More peak capacity, more resolution
Conditions: Columns: as listed, Mobile Phase: A:0.1% TFA in Water B:0.08% TFA in ACN Gradient: 5% B to 60%B in 25 min. Temperature: 80°C Sample: BSA tryptic digest
StartingPressure380 bar
StartingPressure105 barPeak Capacity
Smaller particle size = sharper peaks = greater peak capacity
Page 26
What Makes High Speed HPLC Possible?
• Small particle sizes (< 5 mm)• Short columns (10 cm or less)• Low viscosity mobile phases• Optimized HPLC – minimal extra column volume, low
volume flow cell, minimal delay volume for gradients• Detector set to fast response time• High diffusion coefficient of the sample in the mobile
phase • Rapid diffusion = less band broadening = higher
linear velocities with efficient peaks • But large molecules have small diffusion
coefficients!!
Page 27
PorousShell
SolidCore
• Increase the Diffusion Rate • Elevate operating temperature -- Need stable bonded
phase• Decrease solvent viscosity -- Helps but changes elution
• Decrease the Diffusion Distance• Develop very small particles (<2-um) • Limit diffusion distance into a particle!
Slower Diffusion of Large Molecules Limits Speed and Resolution
So, decrease the diffusion time for macromolecules!
Page 28
A solid silica core with a superficially porous surface of smaller silica sol particles derivatized with bonded phase.
Scanning Electron Micrograph of Poroshell Particles
Scale: = 1 µm
What is Poroshell 300SB?
• Poroshell 300SB superficially-porous particles, reduce diffusion distance for proteins for fast HPLC.
• A RP-HPLC phase such as C18, C8 or C3 are bound to sol particles in the porous shell.
Page 29
0.25 µm
5 µm Totally Porous Particle
2.5 µmRequired diffusion distance
for a macromoleculereduced 10 fold !
5 µm Superficially Porous Particle
Comparison of Diffusion DistanceTotally porous silica vs. superficially porous silica
Page 30
Column: Poroshell 300SB-C18, 5 μm Mobile phase: Urease, 44.7% ACN / 55.2% 0.1% TFA, Insulin1, 32.6% ACN / 67.3% 0.1% TFA Insulin 2, 28% ACN / 72% 0.1% TFA Detection:UV 214 nm Samples: in 6 M guanidine, pH 6.8 phosphate buffer Temperature: 60°C
Mobile Phase Velocity, cm/sec0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6
Pla
te H
eigh
t, cm
0.000
0.005
0.010
0.015
0.020
Data fitted to Knox equation
Urease; MW=83,000(sub-unit); column: 3 x 100 mm
Insulin1; MW=5,7003 x 100 mm column
Insulin2; MW=5,7002.1 x 75 mm column
• Slow increase in plate height with velocity means high flow rates can be used for fast separation with modest loss in resolution
• Steeper plot for urease reflects slower diffusion of much larger protein
Maintain Efficiency at High Velocity with Poroshell 300SB-C18
Van Deemter Plot for Urease (83,000D) and Insulin (5,700D)
Page 31
How Do We Use Poroshell Columns?
• Faster separations• High flow rates• High flow rates for fast analysis with high resolution in
comparison to totally porous silica• Polypeptides• Large proteins
• High recovery of large proteins• Elevated temperature
• High resolution• High resolution of protein digests• High resolution of protein impurities• High resolution for LC/MS of proteins
• Change selectivity• Different bonded phases• Comparison to totally porous 300SB materials
Page 32
High Flow Rates with 2.1 mm ID Poroshell for High Resolution and Fast Separations
12
34
5
67
8
0 0.5 1.0Time (min)
Columns: Poroshell 300SB-C182.1 x 75 mm, 5 mm
Mobile Phase:A: 0.1% TFAB: 0.07% TFA in ACN
Gradient: 5 – 100% B in 1.0 min.Flow Rate: 3.0 mL/min.Temperature: 70°CPressure: 250 barDetection: UV 215 nm
Sample:1. Angiotensin II2. Neurotensin3. Rnase4. Insulin5. Lysozyme6. Myoglobin7.Carbonic Anhydrase8.Ovalbumin
• Only Poroshell can provide high efficiency at higher flow rates for extremelyrapid separations of proteins and peptides.
• This is due to more rapid mass transfer of the superficially porous particle
Page 33
Flow Rates for Poroshell Columns
Column Internal Diameter
Porous Particle Flow Rate Range
Poroshell Flow Rate Range
2.1 mm 0.1 – 0.3 mL/min 0.3 – 3 mL/min
1.0 mm 30 – 60 mL/min 0.08 - 0.75 mL/min
•Very high flow rates can be used effectively with Poroshell columns
Page 34
min0 2 4 6 8 10 12 14 16 18
mAU
-1000
100200300400500
min*0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8
mAU
-1000
100200300400500
(Poroshell chromatogram - expanded)
min*0 2 4 6 8 10 12 14 16 18
mAU
-1000
100200300400500
1\DATA\PORSHL3\PrSHL010.d\ßamy-ovr.win
Poroshell 300SB-C182.1x75 mm, 5 µm
Zorbax 300SB-C18 4.6x150 mm, 5 µm
F =1 mL/min, 5-100 %B / 20 min
ß-amylase
ß-amylase
ß-amylase
Poroshell chromatogramexpanded
•Poroshell technology facilitates ultra-fast HPLC analysis of proteins Note the similar separation of minor impurities in the 2 and 20 min
separations.
12 min
2min
12min
0 1 1.4 1.80.6 0.80.40.2 1.2 1.6
Poroshell for Fast Analysis of High MW Proteins - ß-amylase (200kdal)
Agilent 1100 WPS with AutoBypass; Piston Stroke: 20µL, Column: as shown Gradient: as shown Mobile Phase: A= 95% H2O, 5% ACN with 0.1%TFA; B= 5% H2O, 95% ACN, with 0.07%TFA Flow Rate: as shown Temp.: 70°C, Detection.: UV 215 nm
F =2 mL/min, 5-100 %B / 1 min
Page 35
Poroshell Has Different Selectivity from Totally Porous 300SB
min0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
mAU
0
200
400
600
800
min0.10
0
0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
mAU
0
200
400
600
800
Agilent 1100 WPS with AutoBypass Column: 300SB-C18, 2.1 x 75 mm, 5 mm Mobile Phase: A= 95% H2O, 5% ACN with 0.1%TFA B= 5% H2O, 95% ACN, with 0.07%TFA Gradient: 5 –100%B in 0.67 min Flow Rate: 3 mL/min Piston Stroke: 20 µL Temp.: 70°C Det.: UV 215 nm
300SB-C18
Poroshell 300SB-C18
1,2 43
32
1
4
•Poroshell and 300SB-C18 can have different selectivity due to the different ratios of bonded phase on the surface.
Page 36
5 min
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
Totally Porous2.1x75mm, 5µm3 mL/min
0-100%B in 0.67 min
1. Neurotensin2. Rnase A3. Lysozyme4. Myoglobin
1 2 3 4
2 mL/min0-100%B in 1 min
1 mL/min0-100%B in 2 min
0.5 mL/min0-100%B in 4 min
Agilent 1100 WPS with AutoBypass; Piston Stroke: 20µL, Temp.: 70°C, Det.: 215 nmMobile Phase: A= 95% H2O, 5% AcN with 0.1%TFA; B= 5% H2O, 95% AcN, with 0.07%TFA
•Increased flow rate decreases run time, but, decreases resolution in protein analysis when using totally porous particles!
Effect of Increasing Flow Rate in Protein Analysis with use of Totally Porous Silica
Page 37
5 min
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
mAU
0
200
400
600
800
1000
Poroshell 300SB-C182.1x75mm, 5µm
3 mL/min0-100%B0.67 min
1. Neurotensin2. Rnase A3. Lysozyme4. Myoglobin
12 3 4
Agilent 1100 WPS with AutoBypass; Piston Stroke: 20µL, Temp.: 70°C, Det.: 215 nmMobile Phase: A= 95% H2O, 5% ACN with 0.1%TFA; B= 5% H2O, 95% ACN, with 0.07%TFA
2 mL/min0-100%B1 min
1 mL/min0-100%B2 min
0.5 mL/min0-100%B4 min
Sustained Efficiency and Resolution
• As flow rate increases, protein/peptide analysis time is dramatically reduced when using superficially porous particles!
Effect of Increasing Flow Rate in Protein Analysis with Poroshell
Page 38
min0 2 4 6 8 10 12 14 16 18
mAU
-1000
100200300400500
ß-amylase
A=0.1%TFA, H2O; B=0.07%TFA, ACN
min*0 2 4 6 8 10 12 14 16 18
mAU
-1000
100200300400500
*DAD1 A, Sig=215,16 Ref=310,10
ß-amylase(Poroshell chromatogram - expanded)
min*0 2 4 6 8 10 12 14 16 18
mAU
-1000
100200300400500
*DAD1 A, Sig=215,16 Ref=310,10
ß-amylase
1\DATA\PORSHL3\PrSHL010.d\ßamy-ovr.win
Poroshell 300SB-C182.1x75 mm, 5 µm
Temp: 70ºC
F =2 mL/min, 5-100 %B / 1 min
ZORBAX 300SB-C18 4.6x150 mm, 5 µm
Temp: 70ºC
F =1 mL/min, 5-100 %B / 20 min
ß-amylase
ß-amylase
ß-amylase
Poroshell chromatogramexpanded
• Poroshell technology facilitates ultra-fast HPLC analysis of proteins• Note the similar separation of minor impurities in the 2 and 20 min separations.
12 min
2min
12min
0 1 1.4 1.80.6 0.80.40.2 1.2 1.6
A=0.1%TFA, H2O; B=0.07%TFA, ACN
Poroshell for Fast Analysis of High-Molecular Weight Proteins
ß-amylase (200kdal)
Page 39
min0 2 4 6 8 10 12 14 16 18
mAU
-50
0
50
100
150
200
250
DAD1 A, Sig=215,16 Ref=310,10 (PORSHL3\PRSHL010.D)
Poroshell 300SB-C18, 2.1x75mm, 5µm
F=2, 5-100%B/1 min, 70°C
A=0.1%TFA, H2O; B=0.07%TFA, ACN
Thyroglobulin subunits
min*0 2 4 6 8 10 12 14 16 18
mAU
-50
0
50
100
150
200
250
*DAD1 A, Sig=215,16 Ref=310,10 (PORSHL3\PRSHL010.D)
(Poroshell chromatogram - expanded)
Thyroglobulin subunits
min*0 2 4 6 8 10 12 14 16 18
mAU
-50
0
50
100
150
200
250
*DAD1 A, Sig=215,16 Ref=310,10 (PORSHL3\PRSHL042.D)
1\DATA\PORSHL3\PrSHL042.d\Thyr-ovr.win
Thyroglobulin subunits
Zorbax 300SB-C18, 4.6x150mm, 5µm
F=1, 5-100%B/20 min, 70°C
Poroshell 300SB-C182.1x75 mm, 5 µm
F =2 mL/min, 5-100 %B / 1 min
F =1 mL/min, 5-100 %B / 20 min
Thyroglobulinsubunits
Poroshell chromatogramexpanded
• Poroshell technology facilitates ultra-fast HPLC analysis of very large protein subunits.• Note the improved peak shape and recovery on Poroshell, in <2 min.
Thyroglobulinsubunits
Thyroglobulinsubunits
12 min
2min
12min
Zorbax 300SB-C18 4.6x150 mm, 5 µm
Good Peak Shape and Recovery with PoroshellAnalysis of High-Molecular Weight Proteins - Thyroglobulin 660 kDa
Page 40
Aligned
min0 1 2 3 4 5 6 7
mAU
0
50
100
150
200
min*0 1 2 3 4 5 6 7
mAU
0
50
100
150
200
Poroshell 300SB-C182.1x75 mm, 5 µm
BSA tryptic digest15hrs., 70 pmol
Agilent 1100 WPS with AutoBypass; Piston Stroke: 20µL, Temp.: 70°C, Det.: 215 nmMobile Phase: A= 95% H2O, 5% AcN with 0.1%TFA; B= 5% H2O, 95% AcN, with 0.07%TFA
1 mL/min0-100%B12 min
0.208 mL/min0-100%B120 min
Zorbax 300SB-C182.1x150 mm, 5 µm
6min
60min
• Extremely fast (6 min), high, resolution of BSA tryptic digest, using Poroshell 300SB-C18. • Under typical conditions, a comparable run takes 60min on a totally porous particle.
High Resolution and Fast AnalysisSeparation of a Protein Digest on Poroshell 300SB-C18
Page 41
UltraHigh Speed HPLC Peptide M aps of a M onoclonal Antibody on Several Zorbax Poroshell Phases
Aligned
min1 2 3 4 5
min1 2 3 4 5
min1 2 3 4 5
mAU
0
20406080
100
mAU
0255075
100125150175
mAU
0
50
100
150
200
Aligned
min1 2 3 4 5
min1 2 3 4 5
min1 2 3 4 5
mAU
0
20406080
100
mAU
0255075
100125150175
mAU
0
50
100
150
200
min1 2 3 4 5
min1 2 3 4 5
min1 2 3 4 5
min1 2 3 4 5
min1 2 3 4 5
min1 2 3 4 5
mAU
0
20406080
100
mAU
0255075
100125150175
mAU
0
50
100
150
200
mAU
0
20406080
100
mAU
0255075
100125150175
mAU
0
50
100
150
200
Conditions: M obile phase A = 0.1% TFA in water M obile phase B = 0.1% TFA in ACN; Temperature: 70°C; Detection: VW D, 210nm; Injection: 10 µl Lys-C digest of Human M onoclonal Antibody; Flow: 1.0 ml/min; Gradient: 0 min, 0% B; 5.5 min, 55% B; 5.6 min, 55% B; 7.0 min, 0% B
Zorbax Poroshell 300SB-C18, 2.1 x 75mm, 5µ
Zorbax Poroshell 300SB-C8, 2.1 x 75mm, 5µ
Zorbax Poroshell 300SB-C3, 2.1 x 75mm, 5µ
•Zorbax Poroshell technology facilitates ultra-fast HPLC analysis of peptides
Original method –120 min tG using a C18 4.6 x 250 mm –57 peaks detected
For more details see 5989-0590EN
46 to 48 peaks, 1/20 analysis time
Data courtesy of:Novartis Pharma, Biotechnology, Basel Dr. Kurt ForrerPatrikRoethlisberger
Page 42
LC/MS of Proteins with PoroshellFormic acid Provides High Sensitivity
m/z1000 1100 1200 1300 1400 1500 1600 1700 1800 1900
0
10000
20000
30000
40000
50000
60000
1254.4
1187.3
1231.2
1146.4
1303.6
1278.7
1166.5
1356.7
1208.8
1108.2
1126.9
1329.7
1090.0
1189.6
1211.9
1445.2
1128.7
1257.3
1385.5
1233.2
1414.4
1149.1
1306.8
1168.4
1331.8
1110.3
1360.0
1510.9
1545.8
1477.2
1448.7
1055.5
1282.0
1621.3
1075.2
1582.6
1661.8
1704.1
1749.1
1023.3
1796.7
m/z1000 1200 1400 1600 1800 2000
0
500
1000
1500
2000
2500
3000
3500
4000
1385.2
1796.5
1513.7
1704.2
1955.1
1624.3
1749.1
1303.6
1546.2
1414.4
1665.6
1480.6
1621.8
1586.3
1388.2
1359.6
1705.9
2004.7
1510.7
1801.0
1904.5
2144.6
2171.8
2076.9
1848.8
1734.8
1254.4
min0.5 1 1.5 2 2.5 3 3.5
0
5000000
10000000
15000000
20000000
25000000
Abundance
Formic Acid
TFA
TIC
Column: Poroshell 300SB-C18, 2.1 x 75 mm, 5 mm Mobile Phase Gradient: 20-100% B in 5.5 min. A: water + 0.1% TFA or FA B: ACN + 0.1% FA or TFA Flow Rate: 500 mL/min Temperature: 60°C Injector: 1 mL Sample: 10 pmol of BSA Electrospray ionization: positive ion Vcap:6000V Drying Gas: 12L/min 350ºC Nebulizer: 45 psi Scan: 600-2500 amu Step size:0.15 amu Peak width:0.06 min
Page 43
High Flow Rates and High Sensitivity LC/MSUsing 1.0 mm ID Poroshell
min0 0.5 1 1.5 2 2.5 3 3.5
0
20000000
40000000
60000000
80000000
1E8
pmoles of protein on column0.5
2.55
0.751
•These TIC’s show good sensitivity with only 0.5 pmoles on column.
Column: Poroshell 300SB-C18, 1.0 x 75 mm, 5 mm Mobile Phase Gradient: 20-100% B in 5.5 min. A: water + 0.1% formic acid B: ACN + 0.1% formic acid Flow Rate: 600 mL/min Temperature: 80°C Injection volume: 1 mL Sample: insulin, lysozyme, cytochrome C, myoglobin, BSA, carbonic anhydraseLC/MS: Pos. Ion ESI –, Vcap 6000 V, Drying gas flow: 12 l/min Drying gas temperature: 350°C Nebulizer: 45 psi, Fragmentor volatage: 140 V Scan: 600 –2500 Stepsize: 0.15 amu Peakwidth: 0.06 min
Page 44
min0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5 2.75
mAU
10
20
30
40
50
min0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5 2.75
mAU
10
20
30
40
50
Untreated
Fast Analysis of Insulin (5.7 kDa) Impurities using Poroshell 300SB-C18
Column: Poroshell 300SB-C18, 2.1 x 75 mm, 5 mm Mobile Phase Gradient: 5-100% B in 5 min. A: water + 0.1% TFA B: ACN + 0.0.07% TFA Flow Rate: 2 mL/min Temperature: 70°C Sample: Bovine pancreas insulin
INSULIN - bovine pancreas
27 hrs. at 55°C
Page 45
8 min
min0 1 2 3 4 5 6 7
mAU
0200400600
min0 1 2 3 4 5 6 7
mAU
0200400
min0 1 2 3 4 5 6 7
mAU
0100200300
.
min0 1 2 3 4 5 6 7
mAU
0
100
min0 1 2 3 4 5 6 7
mAU
-500
50100 Impurities CA
CA
CA
CA
CA
•Poroshell technology is compatible with a wide range of gradients that result in a corresponding range of resolution and run times.
•Note the consistent separation of minor impurities even in the 1 min run at top.
tG = 2 min
tG = 4 min
tG = 1 min
tG = 8 min
tG = 16 min
Column: Poroshell 300SB-C18, 2.1 x 75 mm, 5 mm Mobile Phase Gradient: 5-100% B in tG min. A: water + 0.1% TFA B: ACN + 0.0.07% TFA Flow Rate: 2 mL/min Temperature: 70°C Sample: Carbonic Anhydrase
Increasing Gradient Times on PoroshellRapid Analysis of Fresh Carbonic Anhydrase
Page 46
4 min
0 0.5 1 1.5 2 2.5 3 3.5
mAU
0
100
200
300Carbonic Anhydrase Poroshell 300SB-C18 (2.1x75mm)
5 hrs. Incubation at 55°C
Flow=2mL/min, 5-100%B/4 min, 70°C
0 0.5 1 1.5 2 2.5 3 3.5
mAU
0
100
200
300
1 hrs. Incubation at 55°C
0 0.5 1 1.5 2 2.5 3 3.5
mAU
0
100
200
300FRESH
2
Poroshell 300SB-C182.1x75 mm, 5 µm
F=2mL/min, 5-100%B/4 min, 70°CA=0.1%TFA in H2O, B=0.07%TFA in ACN
Carbonic Anhydrase (CA)
5 hrs. incubation at 55°C
1 hr. incubation at 55°C
FRESH
CA
CA
CA
•Rapid HPLC analysis of carbonic anhydrase and its aging products is achieved in 4 minutes.
Impurities
Rapid Analysis of Carbonic Anhydrase Aging RP-HPLC on Poroshell 300SB-C18
Page 47
• Extremely fast resolution of carbonic anhydrase aging products using Poroshell under four different gradient times. High resolution is obtained in 2 or 3 min.
0 0.5 1 1.5 2 2.5 3 3.5
mAU
50100150200250
0 0.5 1 1.5 2 2.5 3 3.5
mAU
5075
100125150175200
0 0.5 1 1.5 2 2.5 3 3.5
mAU
020406080
100
min0 0.5 1 1.5 2 2.5 3 3.5
mAU
020406080
100
Carbonic Anhydrase (CA)Several M onths at 4°C
1 min
2min
3 min
Fast Separation of Aged Carbonic AnhydraseRP-HPLC on Poroshell 300SB-C18
Column: Poroshell 300SB-C18, 2.1 x 75 mm, 5 μm M obile Phase Gradient: 5-100% B in tG min. A: water + 0.1% TFA B: ACN + 0.0.07% TFA Flow Rate: 2 mL/min Temperature: 70°C Sample: Carbonic Anhydrase1 mg/mL Injection Volume: 2 μL
tG = 2 min
tG = 4 min
tG = 8 min
tG = 1 min
Page 48
More Poroshell Bonded Phases Provide Selectivity Options to Enhance Resolution
min0 0.5 1 1.5 2 2.5
mAU
0
100
200
300
400
Poroshell SB-C18, 2.1 x 75 mm
1
2 3 4
5, 6
7
8
9 0.9
76
1.0
77
1.2
89
1.3
95
1.4
63
1.5
23
1.7
41
1.8
60
min0 0.5 1 1.5 2 2.5
mAU
0
100
200
300
400
512
34
6
7
8
9
Poroshell SB-C3, 2.1 x 75 mm
0.9
01
1.0
36 1.2
64
1.3
54 1.4
30 1
.449
1.4
97
1.7
39
1.8
37
Column: Agilent Poroshell (2.1 x 75 mm); Temp.: 70 0C; Flow: 0.5 mL/min; Det: UV 215 nmMobile Phase: A= 0.1% TFA/H2O, B= 0.07% TFA/ACN; Gradient: 5-100% B in 3.0 min
1.5
5
4 6
7
1.5
5
4 6
7
1.5
4
5, 6
7
1.5
4
5, 6
7
Samples:1. Angiotensin II2. Neurotensin3. RNase A4. Insulin B Chain5. Insulin6. Cytochrome C7. Lysozyme8. Myoglobin9. Carbonic
Anhydrase
Poroshell 300SB-C182.1 x 75 mm
Poroshell 300SB-C32.1 x 75 mm
•Changing from SB-C18 to SB-C3, within the Poroshell family results in resolution of peaks 5 and 6, still in 3 min!
Page 49
Summary
• Fundamentals of Bio Separations apply whether using conventional columns or columns that optimize mass transfer
• These fundamentals can be used to optimize bio separations
• 1.8 µm particles can be used for much faster gradients peptide maps
• Poroshell columns have greatly reduced surface area and markedly reduce resistance to large molecule mass transfer
• This allows for fast analysis of proteins – including very large proteins with high recovery.
• Optimized bonded phases for low and high pH offer more options for RPLC and Mass Spec assays
Page 50
Column Description
Monolith stationary phase.
Page 51
Column Description
• Currently there are 4 Bio-Monolith Products:
• Bio-Monolith QA (strong anion-exchanger)
• Bio-Monolith DEAE (weak anion-exchanger)
• Bio-Monolith SO3 (strong cation-exchanger)
• Bio-Monolith Protein A (immunoaffinity)
Further products to expand the product portfolio (CM: weak cation-exchanger; Protein G (?), Specialty columns (pDNA analytics, virus analytics).
Page 52
Catalog number 5069-36395
Column chemistry QA: Strong anion exchanger
Matrix Poly(glycidyl methacrylate – ethylene dimethacrylate)
Ligand density 2.0 ± 0.2 mmol/g
Matrix dimensions Diameter: 5.2 mm; Length: 4.95 mm;
Bed volume: 0.1 mLDynamic binding capacity > 25 mg BSA/ml
Maximum loading capacity 1.2 to 1.5 mg
Working flow rates 0.2 to 2 ml/min (max. 3 mL/min)
Maximum allowed pressure over the column 150 bar
Temperature stability 4 to 40 °C
Recommended pH 2 to 13 working
1 to 14 CIP
Bio-Monolith Columns: Data SheetsBasic characteristics
Page 53
Column Column LLengthength
Bio-Monolith Columns are characterized by a high resolving power at extremely short column lengths – only 5 milimeter.
10 ml/min = 4500 cm/h
= 360 CV/min (res. time: 0,1 s)
Page 54
Different Loadings on Bio-Monolith QA
10uL 50uL 100uL Gradient
Page 55
Consecutive injections Bio Monolith QAAnalytical Column
Page 56
Dynamic Binding Capacity for DNA
Dynamic Capacities
1% 5% 10%
1%5%
10%
1% 5% 10%
0,0
2,0
4,0
6,0
8,0
10,0
12,0
14,0
16,0
DBC
mg
DNA
QFF – porous particles Q nano - membrane QA - monolith
Note 50 times higher dynamic binding capacity than particle based resinwhile operating at 4-fold higher flow rate!
Courtesy of Pete Gagnon, for details visit validated.com
Page 57
Bio Monoliths
• Monolithic supports are highly porous rigid polymers with:
• High porosity (over 60 %)
• Flow-through channels (“pores”) having large diameter (1.5 µm)
• Uniform channel connectivity in 3D (homogeneous structure).
10 100 1000 10000
Pore diameter [nm]