Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
1
WAT E R S SO LU T IO NS
ACQUITY UPC2 System
Empower® 3 Software
ACQUITY SQD Mass Spectrometer
K E Y W O R D S
UPC2, pharmaceutical impurities,
stability indicating methods,
degradation profiling, method
development, metoclopramide,
convergence chromatography
A P P L I C AT IO N B E N E F I T S ■■ Solve complex impurity profile challenges
■■ Approaches to aid final
method development decisions
IN T RO DU C T IO N
UltraPerformance Convergence Chromatography™ (UPC2™) exploits the benefits
of sub-2-µm particle size stationary phases, with carbon dioxide as the primary
mobile phase component. Convergence chromatography is a complementary
analytical technique to liquid chromatography as it provides orthogonal
selectivity, thereby increasing the opportunity to identify impurities present
in a sample. Mass spectral information helps analysts confirm, identify, and
characterize the quality of the pharmaceutical ingredients. Coupling UPC2 to mass
spectrometry provides an important tool for pharmaceutical analysis compared
to previously published reversed phase liquid chromatography (RPLC) impurity
analysis approaches.1-3
Anomalies were observed during the method development screening process.4
In one instance, a standard solution of impurity F was hypothesized to be unstable
after a few days. In this application, we use the ACQUITY UPC2™ System coupled
to ACQUITY® SQD to analyze the identity and relationship of the unknown peaks
observed during the method development standards and expired samples of
metoclopramide. Impurity relationship to the API are hypothesized and confirmed
with the use of the MS spectral data. Finally, the MS data from the impurity profile
was interrogated to ensure the specificity of the methodology in the presence of
these unknown peaks to aid future refinement of the final method.
Impurity Profiling Using UPC2/MSMichael D. Jones and Warren PottsWaters Corporation, Milford, MA, USA
CH3
CH3
N
O
NH
CH3ONH2
Cl
Figure 1. Chemical structure of metoclopramide.
2Impurity Profiling Using UPC2/MS
Sample description
Investigations of impurity C and impurity F standards were prepared in methanol
and explored at 0.1 mg/mL concentrations. The expired sample preparation was
extracted using methanol, and prepared at a concentration of 2 mg/mL relative to the
metoclopramide active ingredient.
R E SU LT S A N D D IS C U S S IO N
Prior to analyzing the degraded metoclopramide sample, the MS data was
examined to address questions regarding observations with the chromatography
of impurity reference standards 4-amino-5-chloro-2-methoxybenzoic acid;
“impurity C” and 4-amino-5-chloro-N-2-(diethylaminoethyl0-2-
hydroxybenzamide “impurity F.”
MS investigation of reference standard impurity C
During the screening process, two peaks were separated during the injection of
the impurity C standard using the ACQUITY UPC2 CSH™ Fluoro-Phenyl column.
This phenomenon was not observed with the other stationary phases. MS spectral
analysis of the two peaks showed similar spectra. The MS spectrum indicates
possible dimerization of the analyte. Based on this information, it was determined
that the two peaks were related to each other rather than the second peak being a
contaminant. The rapid determination of this relationship provided direction for
scoping future analysis by MS/MS and accurate mass. The data generated by those
techniques will be more useful in determining the purity of the standard.
E X P E R IM E N TA L
UPC2 conditions
System: ACQUITY UPC2 with
PDA and SQD detection
Column: ACQUITY UPC2
BEH 2-EP
3.0 x 100 mm,
1.7 µm
Mobile phase A: CO2
Mobile phase B: 1 g/L Ammonium
formate in 50:50
methanol/acetonitrile
spiked with
3% formic acid
Wash solvents: 70:30 methanol/
isopropanol
Separation mode: Gradient; 5% to
30% B over 5.0 min;
held at 30% for 1 min
Flow rate: 2.0 mL/min
CCM back pressure: 1500 psi
Column temp.: 50 °C
Sample temp.: 10 °C
Injection volume: 0.5 µL
Run time: 6.0 min
Detection: PDA 3D Channel: PDA,
200 to 410 nm; 20Hz
PDA 2D Channel:
275 nm at 4.8 nm
Resolution
(compensated
500 to 600 nm)
SQD MS:
150 to 1200 Da;
ES PosNeg
Make-up flow: None; (make-up pump
not configured in
flow splitter)
Data management: Empower 3 Software
3Impurity Profiling Using UPC2/MS
Investigation of impurity F working standard solution stability
The peak shape of impurity F was observed to degrade over time, during the method development process.
Unknown peaks would appear in the chromatogram within the course of a week. In addition, the color of the
standard solution changed from a clear solution to a solution with a brownish tint. MS interrogation revealed
the masses listed in Table 1. The masses were correlated to those found to be significant to the UV trace
(not shown) between retention time 2.0 min and 3.5 min. XIC of m/z = 330 and 296 resulted in multiple peaks.
The working standard was prepared in methanol. Many of the impurity peaks were products of methylation or
methoxylation. Based on this information, alternative diluents will be explored to inhibit the likelihood of these
transformations. Presently, the working standard solution shelf life has been decreased to three days until a suitable
diluent study can be performed. The unknown peak #8 in Table 1 with m/z = 258 has the same mass of methyl
4-(acetylamino)-5-chloro-2-methoxybenzoate, commonly referred to as “impurity B.” This unknown peak found in
the impurity F working standard was determined not to be identical to EP impurity B due to differences in retention
time; whereas, impurity B elutes at approximately 0.48 min. The origins of other impurities are still undetermined.
They are hypothesized to be more products of solution instability or present in the reference material.
2.228 Peak 1 - SQ 1: MS Scan 1: 200.00-600.00 ES+, Centroid, CV=Tune - AVG (1.4:2.1;2.4:3.1) Th: 6.000
202.1
204.1
303.3
313.2
327.2
335.3
342.1
344.2349.3
403.2
405.2
407.3
425.2427.2
482.3
Inte
nsity
0.0
7.0x105
1.4x106
2.1x106
2.8x106
2.267 Peak 2 - SQ 1: MS Scan 1: 200.00-600.00 ES+, Centroid, CV=Tune - AVG (1.4:2.1;2.4:3.1) Th: 6.000
202.2 303.3
313.2
327.2
335.2
342.2
344.1
403.2
405.2
406.3407.3
425.2427.2
482.2484.2
Inte
nsity
0.0
7.0x105
1.4x106
2.1x106
2.8x106
m/z200.00 240.00 280.00 320.00 360.00 400.00 440.00 480.00 520.00 560.00 600.0
Peak 1
Peak 2
2.2
28
2.2
70
Inte
nsity
0.0
6.0x106
1.2x107
1.8x107
2.4x107
Minutes
2.016 2.064 2.112 2.160 2.208 2.256 2.304 2.352 2.400 2.448
O H
O
C H 3ONH 2
Cl
Figure 2. MS spectral analysis of EP impurity C for the doublet peaks observed when using the UPC2 CSH Fluoro-Phenyl stationary phase.
Name Rt (min) Observed m/z Δ Mass Proposed transformation
EP Impurity F 2.924 286
Unknown 1 2.268 344 + 58 Da methoxylation + methylation
Unknowns 2 & 4 2.303 & 2.614 330 + 44 Da methoxylation
Unknowns 3 & 6 2.680 & 2.886 296 + 10 Da hydrolysis + two methylations
Unknown 5 2.864 356 + 70 Da ?
Unknown 7 3.113 252 - 34 Da Loss of Cl-
Unknown 8 3.288 258 - 28 Da Loss of two CH3 groups
Table 1. Masses found in the degraded standard solution of metoclopramide EP impurity F.
4Impurity Profiling Using UPC2/MS
Investigating an expired metoclopramide sample
The ACQUITY UPC2 System coupled to the ACQUITY SQD Mass Spectrometer controlled by Empower 3
Software provided a simple solution to profile impurities in an expired metoclopramide sample. MS spectral
extraction of peaks in the UV chromatographic trace were simply performed by using right mouse click in
the review window to rapidly confirm known impurities and identify 12 unknown impurities, as shown in
Table 2. The sensitivity of UPC2 provided s/n ≥ 10 for impurities detected with area% ≥ 0.05% in the UV
chromatographic trace. The expired metoclopramide sample was interrogated to determine if the masses
in Table 1 were present. The MS data confirmed the presence of 4 out of the 8 impurities related to EP F;
m/z =296, 344, 252, and 258. In addition to the known impurities, a total of 7 unknown impurities were
detected. The masses, retention time, and UV signal-to-noise were recorded in Table 2.
MS TIC - SQ 1: MS Scan - 1: 200.00-600.00 ES+, Centroid, CV=Tune
Inte
nsi
ty
0.0
4.0x107
8.0x107
1.2x108
1.6x108
Minutes0.00 0.60 1.20 1.80 2.40 3.00 3.60 4.20 4.80 5.40 6.00
m/z
=28
6
m/z
=34
4m
/z=
330
m/z
=33
0
m/z
=35
6
m/z
=29
6
m/z
=25
2
m/z
=25
8
m/z
=29
6
CH3
CH3
N
O
NH
OHNH2
Cl
Impurity F: [M+H] = 286
Figure 3. MS ES+ TIC of a degraded standard solution of metoclopramide EP impurity F.
5Impurity Profiling Using UPC2/MS
: Aged Metoclop : 50:50 MeOH ACN 1g/L ammonium formate 3% formic acid
0.000
0.035
0.070
0.105
0.140
Minutes0.00 0.60 1.20 1.80 2.40 3.00 3.60 4.20 4.80 5.40 6.00
Impu
rity
B
Unk
now
n m
/z=
22
6
Impu
rity
9
Impu
rity
D
Me
tocl
op
ram
ide
Impu
rity
G
Impu
rity
H
Unk
now
n m
/z=
26
6
Impu
rity
A
Impu
rity
F
*No
MS
Unk
now
n m
/z=
34
4
Unk
now
n m
/z=
31
0
Unk
now
n m
/z=
27
2
AU
Figure 4. UV 275 nm chromatographic trace of a degraded sample of metoclopramide. The major impurities are labeled.
Rt % Area USP s/n m/z
Impurity B 0.49 0.48 148 258
Unknown #1 0.90 0.34 54 226
Impurity “9” 1.18 1.34 255 (ESI-) 180
Impurity D 1.41 1.14 229 224
Unknown #2 2.18 0.12 22 306
Unknown #3 2.33 0.02 8 358
Unknown #4 2.38 1.29 195 n/a
Impurity A (Isomer 1) 2.49 0.28 48 342
Impurity A (Isomer 2) 2.53 0.41 65 342
Rel. Impurity F (Unk. #1) 2.61 0.69 87 344
Rel. Impurity F (Unk. #6) 2.83 0.06 10 296
Impurity F 2.91 0.51 82 286
Rel. Impurity F (Unk. #7) 3.09 0.14 20 252
Unknown #5 3.14 0.55 77 310
Rel. Impurity F (Unk. #8) 3.28 0.14 20 258
Impurity G 3.37 0.87 130 316
Metoclopramide 3.43 86.07 8186 300
Unknown #6 3.60 2.83 380 266
Unknown #7 3.63 1.09 165 272
Impurity H 4.53 1.62 101 (ESI-) 194
Table 2. Peak list of the UV 275 nm chromatographic trace relative to the injection of the metoclopramide expired sample.
Waters Corporation34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Waters, Empower, and ACQUITY are registered trademarks of Waters Corporation. UltraPerformance Convergence Chromatography, UPC2, CSH, and T he Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2013 Waters Corporation. Produced in the U.S.A.February 2013 720004575EN AG-PDF
CO N C LU S IO NS
The ACQUITY UPC2 System coupled to MS provided a
comprehensive approach to impurity profiling. This configuration
enables a scientist to quickly answer questions regarding reference
standard purity, as shown with reference standard impurity C.
UPC2/MS guided the decisions to investigate diluent choices for
the impurity F working standard and adjusting the shelf life of the
working standard solution. In addition, investigating instability of
impurity F provided insight into other potential impurities that may
be present in the drug sample impurity profile. Furthermore, seven
unknown impurities were detected in the expired metoclopramide
sample. Interrogation of the UV and MS data was simply performed
using Empower 3 Software. Overall, utilizing UPC2/MS increased
the knowledge base of pharmaceutical product quality, and
improved the methodology procedures involved with achieving
the analytical goals.
References
1. Ahuja S, Alsante K. Handbook of Isolation and Characterization of Impurities in Pharmaceutical Compounds. Elsevier. 2003.
2. Jones MD et al. Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate. Waters Application Note 720003079EN. 2009 Oct.
3. Jones MD et al. A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance. Waters Application Note 720003081EN. 2009 Oct.
4. Jones MD, Potts W. UPC2 Method Development for Achiral Impurity Analysis. Waters Application Note 720004577. 2013 Jan.