Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
Improving sensitivity, flexibility, and capabilities with Agilent’s ICP-MS and Triple QuadrupoleICPMS with Tandem MS/MS capabilities
For the Environmental, Petrochem and Semiconductor IndustriesJon Talbott, PhDApplication EngineerAgilent Technologies
Houston Symposium and Tradeshow, June 27, 2017
Presentation Outline
1. Background on ICP-MS and ICP-MS/MS
2. Unique product features of the Agilent 7700 ICP-MS and of the 8800 QQQ for analyzing tough samples in Your Industry
3. Environmental Regulations for Flue Gas Desulfurization Waste waters and ICPMS analyses
4. Direct Analysis for Metals in Gasoline and other organic matrices – conditions and results
5. Unique Petrochemical Applications using ICP-QQQ MS
6. Unique SemiCon Applications using ICP-QQQ MS
July 10, 20132
Page 3
General Capabilities of ICP-MS for Analyses in Aqueous Matrices
• Multi-elemental technique• High sensitivity• Low Detection Limits - ppt for most elements• Short analysis time (~ 3 min)• Wide linear dynamic range• Minimized interferences with Cell Technology• High Productivity – High Sx Throughput
Less well known that the Agilent ICP-MSs have some unique Product features that enhance their capabilities for analyzing
tough samples in the Petrochemical, Semiconductor and Environmental Industries – gasolines, kerosenes, naphthas,
FDG, SemiCon Chemicals
Agilent 7700 ICP-MS and 8800 ICP-MS/MS
7700 is Agilent’s high performance single quadrupole ICP-MS Many unique capabilities including HMI, ORS3
8800 is World’s first ICP-MS to offer MS/MS New modes of operation and performance with MS/MS modes Built upon the industry leading 7700 ICP-QMS platform with shared sample introduction, HMI, many
consumables, robustness, key hardware components and software platform
Revolutionize the MS arena!
Agilent 7700 Single Quad ICP-QMS
Agilent 8800Triple Quad ICP-MS/MS
Page 5
Agilent 7700 ICP‐MS System in Detail
High matrix introduction (HMI) dilution
gas inlet
Peltier‐cooled spray chamber
Off‐axis ion lens
Low‐flow Sample Introduction
Fast, frequency‐matching 27MHz RF generator
High‐performance vacuum system
Cell gas inlet
High‐frequency (3MHz)
hyperbolic quadrupole
Fast, simultaneous dual mode
detector (9 orders dynamic range)
High‐transmission, matrix tolerant
interface
3rd generation OctopoleReaction
System (ORS3)
Agilent 8800 ICP-MS/MS System in Detail
Low flow sample
introduction system
High matrix introduction
(HMI) technology
Fast, frequency-matching 27MHz
RF generator Efficient twin-turbo vacuum system
Dual conical Extraction and Omega lens focus ions across the mass range
9 orders dynamic range
electron multiplier (EM)
detector
Analyzer quad Q2: High frequency
hyperbolic quadrupole – selects ions that pass to detector
High-transmission, matrix tolerant interface
First quad Q1: High frequency hyperbolic quadrupole mass filter –
selects ions that enter the cell3rd generation collision/
reaction cell (ORS3) with up to 4 cell gas lines
Peltier-cooled spray
chamber
Robust, high-temperature plasma ion source
Key Product Features of the Agilent 7700x ICP-MS for Petrochemical, SemiCon and Environmental Industries
Unmatched matrix tolerance and unparalleled interference removal
ORS3 Octapole Reaction cell for removal of all polyatomic interferences
- better with both He and H2 modes for removal of polyatomics from organics
High Matrix Introduction (HMI) for unsurpassed Matrix Tolerance
- permits % level TDS samples to be run directly and routinely even the toughest samples can be handled with ease
O2 Option Gas MFC, that allows organics to be analyzed directly by ICPMS
- Auto setup of conditions
Page 7
ORS3 - a Kinetic Energy Discrimination tool for removing Polyatomic Interferences
Without it, Interferences from a complex acid matrix go from this
Page 8
… to this, with it!!
A single KED mode is very efficient at removing all polyatomic interferences!
And there’s still plenty of Sensitivity to obtain the lowest of Detection Limits
Page 9
High Matrix Introduction (HMI) is an online sample dilution technique that allows undiluted soil digestates and evenundiluted seawater to be analyzed directly.
No other MS instrument can make these claims!
Page 10
Key Benefit of High Matrix Interface (HMI)HMI shows good recoveries even in 3% Total Dissolved Solids
Analysis of Flue Gas Desulfurization Wastewater
Introduction USEPA wastewater effluent revision stricter air pollution controls applies to electric power industries requires low level determination of toxic metals
Steven M. WilburAgilent Technologies Inc.
Richard Burrows and Richard ClinkscalesTestAmerica Inc
Page 12
HMI Important in …. Environmental
FDGs pretty Nasty Matrices …. Synthetic FGD Interference Check Solution is a New Requirement
• Mixed Interference Check Solution (Synthetic FGD Wastewater)• Chloride, 5,000 mg/L
• Calcium, 2,000 mg/L
• Magnesium, 1,000 mg/L
• Sulfate, 2,000 mg/L
• Sodium, 1,000 mg/L
• Butanol, 2000ppmFGD-ICS-A Analyzed once per day•ISTDs must meet 60-125% requirements and analytes must be less than reporting limits
FGD-ICS-AB is “A” solution spiked with analyte elements at 40 ppb (Zn – 0.4 ppm, Al 4.0ppm). Must recover within 70–130 %
The combination of the highly robust plasma of the Agilent 7700x with HMI and ISIS discrete sampling allows routine analysis of this interference check solution and samples containing similar levels of dissolved solids UNDILUTED!
Page 14
Analysis of Flue Gas Desulfurization WastewaterLong Term Stability - Quality Control Check Recovery
Recovery of calibration check samples (CCVs) over 89 analyses
QC Summary including interference checks, matrix spike recoveries, memory
check, and continuing calibration verification
(CCV) and blank verification (CCB)
Biodiesel & KeroseneICP-QMS
Application
We examined very Challenging Aqueous Matrices (Seawater, FDGs) from the Environmental Industry ...
What about Organic Matrices?
General Problems for Organic Solvent analysis
• Plasma stability• Solvent volatility
– Cooled Spraychamber and RF generator design• Carbon deposition
– Add oxygen to plasma
• Matrix-based interferences• High carbon-based interferences plus “usual” plasma backgrounds
• “Difficult” elements often required at relatively low levels• Sulphur• Phosphorus• Magnesium• Etc…
Typical Conditions for Direct Analysis of Gasoline with Agilent 7700 ICP-MS
• Set HW Configuration for Organic Solvent Ignition Mode – Allows automated control of O2 during ignition process– Allows automatic RF Impedance matching
• Standard quartz concentric nebulizer and Scott-type double-pass quartz spray chamber used (No desolvation device necessary)
• Platinum interface cones used• O2 (as a 20% mix in Ar) added as Option Gas (to Aux gas) at torch with
optional mass flow controller (MFC) • Spray Chamber Temp set to -5 C to minimize solvent loading• Taper “organics” torch with a 1.0-mm id injector used
– Further improves plasma stability.• Carrier Gas reduced to ~ 0.7 L/min from typical 1.0 L/min
– reduces solvent loading, backs green bullet C2 off sampler
Direct Elemental Analysis of Biodiesel & KeroseneICP-QMS
Instrument Conditions
Direct Elemental Analysis of Biodiesel & KeroseneICP-QMS Detection Limits
Gasoline ApplicationICP-QMS
Application
Direct Elemental Analysis of GasolineICP-QMS
Instrument Conditions
Direct Elemental Analysis of GasolineICP-QMS Comparison of ORS Gas mode Data
Direct Elemental Analysis of GasolineICP-QMS Long Term Stability
24
Q1 Q2Cell Detector
MS/MS increases selectivity, removing isobaric interference more effectively than ever. Agilent 8800 ICP-QQQ can measure difficult
elements in challenging matrix at lower concentration.
Mass analyzer ICP
Feature of ICP-MS Agilent 8800 Triple Quad ICP-MS (ICP-QQQ)
ICP-QQQ: How Does it Work?
ICP (plasma) and Interface: Forms and extracts ions from the sample (just like the 7700)
EM (detector): Measures the ions that are scanned by Q2 (just like the 7700)
25
Q1 – controls ions that enter the cell
Mass filter• Consistent reactions w/iORS even if sample composition changes
ORS3 – collision/ reaction gas added• Ion size and/or reactions filter and are neutralized or moved
• Product ions are formed
Q2 – selects the target analyte mass • Interference-free analyte ions passed to EM
ASTS 2012
Page 26
MS/MS modes – filter ions entering ORS cell 1. On‐Mass Mode (Q1 and Q2 both set to target mass)
2. Mass‐Shift Mode (Q1 and Q2 set to different masses)
• Advanced Scanning Modes ‐ for research and method development
A. Precursor Ion ScanB. Product Ion Scan C. Neutral Gain Scan
ICP-QQQ: Modes of Operation
ASTS 2012
1. On-Mass Measurement: Unreactive analyte does not react with chosen cell gas, remains at original m/z and so can be separated from reactive interferences. No new cell-formed interferences can occur at the analyte mass, since all non-target masses are rejected by Q1
With ICP-MS/MS, Q1 rejects all non-target masses, ensuring no new analyte/matrix product ions can form new overlaps on original analyte mass
How Reaction Mode Works in ICP-MS/MS
Reaction product ion
On-mass interference
AnalyteOff-mass
interference
Reaction gas
Interference M+MR+
Q1 set to analyte mass – rejects all non-target masses
Q2 set to original analyte mass – rejects any off-mass product ion(s)
Analyte
Interference reacts to form product ion
All non-target masses
27 July 10, 2013
2. Mass-Shift Measurement: Reactive analyte promoted to a new product ion mass and separated from unreactive interferences.
With ICP-MS/MS, Q1 rejects all non-target masses, ensuring no existing ions (analyte, matrix, or polyatomic) can overlap new analyte product ion
How Reaction Mode Works in ICP-MS/MS
Original interfering ion
On-mass interference
AnalyteOff-mass
interference
Reaction gas
Analyte M+MR+
Q2 set to analyte product ion mass – rejects original interfering ions
Analyte product ion
Analyte reacts to form product ion
Q1 set to analyte mass – rejects all non-target masses
All non-target masses
28 July 10, 2013
Oxygen Mass Shift for 34SWith MS/MS
50Cr+/50V+/50Ti+38Ar12C+
13C37Cl+
34S+
17O2+
16O18O+
O2 reaction gas
34S16O+
Q1MS/MS34amu
Q2
50amu
The mass difference between Q1 and Q2 is fixed (16) therefore a single transition is observed – the other oxygen isotopes are eliminated so the
original isotopic pattern is preserved!
Sulfur – Measured as SO+ in O2 ModeIsotopic Abundance of Mass Shifted Ions Maintained
Q1 – Q2 mass difference is 16, so only the + 16O transition is measuredEnsures S isotope abundance is maintained – no overlap from 32S18O+
on 34S16O+, for example
Supports confirmatory isotopes and IR/IDMS
32S16O+
33S16O+
34S16O+
Excellent isotopic template match for 32S, 33S, 34S (~ 30ppb S)
High background at 52 is due to 36Ar16O – S isotope at m/z 36 is too low to be analytically useful (except as a spike)
31 July 10, 2013
Unique ICP-MS/MS Analysesfor
Sulfur, Silicon, Phosphorus, Titanium & Arsenic
in Organic Matrices
Sulfur in Pure Ethanol
Instrument was set to monitor S and SO in a number of modes for comparison
• Single Quad (NoGas)• Single Quad (O2 Gas) – Q1 set as a bandpass filter• MS/MS (O2 Gas) – Q1 set at unit mass resolution
Tune Scan Type Q1 Q2 Name R DL BECNoGas Single Quad 32 S 0.910 11.806 1383.11 ug l-1NoGas Single Quad 33 S 0.752 157.615 4568.56 ug l-1NoGas Single Quad 34 S 0.971 10.550 615.15 ug l-1O2 Q Single Quad 32 S O/R O/R O/R ug l-1O2 Q Single Quad 33 S -0.675 -177.597 -1557.31 ug l-1O2 Q Single Quad 34 S 0.677 191.277 12336.87 ug l-1O2 Q Single Quad 48 32SO 0.999 3.555 53.80 ug l-1O2 Q Single Quad 49 33SO 0.975 99.790 1442.79 ug l-1O2 Q Single Quad 50 34SO 0.947 19.460 1849.35 ug l-1O2 MS/MS MS/MS 32 48 S 1.000 0.358 19.03 ug l-1O2 MS/MS MS/MS 33 49 S 0.999 2.860 19.84 ug l-1O2 MS/MS MS/MS 34 50 S 1.000 0.852 18.99 ug l-1
Calibration “spikes” in Ethanol32S, 33S and 34S Mass Shift Data
100 µgl-1
50 µgl-1
10 µgl-1
5 µgl-1
Blank
Silicon & Phosphorus in EthanolMS/MS Mode
• Optimal gases for silicon and phosphorus differ
– although P, S, As, Se can all be determined under O2
• Si provides better BEC’s under H2 MS/MS on-mass measurement
Mass Shift Mode with O2 On Mass Mode with H2
Ti and As in XyleneMass Shift Mode with NH3 and O2
Ammonia-Ti cluster measured at 132
TiNH2(NH3)4
Arsenic measured as AsO
As measured at 91
ICP-MS/MS Analysis of Sulfur, Phosphorus, Silicon
and Chlorine in a
N-methyl-2-pyrrolidone (NMP) Matrix
S, P, Si and Cl measurement in NMP
S, P, Si and Cl measurement in NMPMS/MS Mass Shift with O2 and H2
(Naoki Sugiyama Poster FP-1)
Applications of ICP-QQQ for SemiConductor Industry
Ultra trace Calcium measurement P and Ti measurement in Si matrix Ti and Zn in high purity Sulfuric Acid
40
Three challenging applications to conventional ICP-MS are introduced today
SemiCon Application 1Ultra trace Calcium measurement
41
Alkali metals and alkaline earth metals are elements that are strictly controlled in semiconductor manufacture since
contamination of those elements significantly deteriorates reliability of the insulating layer, affecting yield and performance
of final product. Argide overlap isotopes of the elements such as K+ and Ca+ causing big spectral interference problems. Agilent
8800 allows an ultra low BEC for Ca to be reached by effectively removing the interference due to the unique
reaction cell using MS/MS.
To remove 40Ar+ interference on 40Ca+, cool plasma ( RF=600-700W ) is an effective technique. MS/MS scan of ICP-QQQ further increases the
performance removing unwanted ions before cell.
42
Application #1: Ultra trace measurement of Calcium Effect of MS/MS scan in cool plasma
parameter unit
RF power W 600
Sampling Depth mm 18
Carrier gas flow rate L/min. 0.7
Make up gas flow rate L/min. 1.0
Cool plasma condition BEC was 6.8ppt with single Quad scan.
BEC was decreased to 1.4ppt with MS/MS scan.
Single Quad scan : Q1 operates as ion guide. It emulates conventional ICP-MS.MS/MS scan : Q1 operates as 1-amu band-path. Unique to 8800 ICP-QQQ.
Addition of small amount of H2 into cell further improves the BEC. H2 totally scavenges remaining 40Ar+ ( slightly produced in cool plasma ) and provides an ultra low BEC.
43
Application #1: Ultra trace measurement of Calcium Ultra low BEC for Ca
Cool plasma
BEC of 41ppq was achieved.
Calibration curve of Ca
SemiCon Application 2P and Ti measurement in Si matrix
44
Silicon is the major substrate material used in semiconductor industry. Many ICP-MS systems are used in research lab but also in
QC labs, since impurity and contamination of metals causes significant impact on yield and performance of the final product.
In Si matrix sample such as Photovoltaic Silicon, VPD sample and TCS , Si polyatomic ions can cause spectral interference problems.
He collision cell and H2 reaction cell can measure trace level of Fe, Ni, Cu and Zn in 2000ppm Si removing interference .
But trace level analysis still challenging for P and Ti.
45
Application #2: P and Ti measurement in Si matrixSpectra interference by Si matrix isotope Abundance % Interference
31P 100 30SiH+, 29H2+, 28SiH3+
46Ti 8.25 30SiO+
47Ti 7.44 28SiF+, 30SiOH+
48Ti 73.72 29SiF+, 28SiFH+
49Ti 5.41 30SiF+, 29SiFH+
56Fe 91.75 28Si2+
58Ni 68.08 30Si28Si+
60Ni 26.22 28SiO2+
63Cu 69.15 29Si16O18O+, 28SiOF+
65Cu 30.85 30SIOF+
64Zn 48.27 29SiOF+, 28SiOFH+
66Zn 27.98 28SiF2+, 30SiOFH+
68Zn 19.02 30SiF2+
Remaining problems
Concentration of Si in VPD sample varies with thickness of the oxide layer; 20-30ppm for naturally oxidized wafer and up to 2000ppm for thermally oxidized wafer. We dissolved bulk Si with HF/HNO3 to make 2000ppm synthetic VPD sample. Normal volume of VPD sample is not more than 0.5ml, so we used low flow PFA nebulizer ( C-flow 50; sample up take rate is 50μL/min ) with Agilent 8800. Three cell gas modes were used as shown . H2 and O2 cell gas are used for P and Ti, respectively.
46
Application #2: P and Ti measurement in Si matrixSi analysis
mode Gas change time (s) element
NoGas 5 Li,be,B,Mg,Rb,Ru,Ag,Cs,Re,Hf,Pt, Au,Tl,Pb,Bi,Th,U
H2 15 Na,Al,P,K,Ca,Mn,Fe,Ga,Sr,Rh
O2 15 S,Ti,V,Cr,Co,Ni,Cu,Zn,Ge,As,Se,Zr,Nb,Mo,Pd,Cd,Sn,Sb,Te,Ba,Hf,Ta,W
Method for high conc. Si analysis
Agilent webinar 201347
Application #2: P and Ti measurement in Si matrixP method; measure P as PH4
+
1%HNO3 100ppm Si 100ppm Si +10ppb P
SiH2
SiH3
SiH
PH4PH3
SiH2+PH
SiH3+PH2
SiH4, SiH5
Q1=31 Ionic Species in H2 mode
Q2 1%HNO3 Si solution P solution31 NO, NOH 30SiH 31P32 NOH 30SiH2 31PH33 30SiH3 31PH234 31PH335 31PH4
SiH+P
In H2 reaction cell, SiH4+ and SiH5
+
are NOT formed. PH3+ and PH4
+
are free from interference of Si+.
48
Application #2: P and Ti measurement in Si matrixP and Ti measurement in 2000ppm Si
Phosphorus TitaniumCell gas H2 O2Q1/ Q2 31/35 48/64
Analyte ion PH4+ TiO+
BEC in 2000ppm Si (ppt) 348 2.8DL in 2000ppm Si (ppt) 213 1.9
P Ti
P calibration plot in 2000ppm Si Ti calibration plot in 2000ppm Si
SemiCon Application 3Ti and Cr measurement in H2SO4
49
Many chemicals are used in Semiconductor industry device for the cleaning process. The highest purity is required to those
chemicals and the required purity is getting stricter year by year with narrowing geometries of advanced semiconductor devices.
H2SO4 is a popular acid as well as HCl for the cleaning, but Sulfur polyatomic ions cause spectra interference problem on many
elements. It is especially challenging to measure Ti and Cr at trace level in high purity H2SO4.
He collision cell and H2 reaction cell can easily measure trace levels of Cu and Zn in a Sulfur matrix by removing the interference.
Trace level measurement of Ti and Cr is more challenging.
50
Application #3: Ti and Cr measurement in H2SO4
Spectra interference by S matrix
isotope Abundance % Interference
46Ti 8.25 32SN+
47Ti 7.44 32SNH+
48Ti 73.72 32SO+, 34SN+
49Ti 5.41 32SOH+
52Cr 83.79 34S18O+
53Cr 9.501 34S18OH+
63Cu 69.15 32SNOH
65Cu 30.85 32S2H+
64Zn 48.27 32S2+, 32SO2+
66Zn 27.98 34SO2+
68Zn 19.02 34S2+
Remaining problems
51
Application #3: Ti and Cr measurement in H2SO4
Ti reaction method Ti+ forms cluster ions with NH3. Below shows “product ion scan” of 48Ti + with NH3
cell gas. Preliminary test showed that a product ion TiNH(NH3)3+ ( Q2 = 114)
provided us the best BEC of Ti in a S matrix.
Sig
nal c
ts
Q2
Product ion of 48Ti+ ; 50ppb Ti std was nebulized.
52
Application #3: Ti and Cr measurement in H2SO4
Cr reaction method Reaction of Cr+ with O2 is endothermic as shown below. Due to excess collision
energy, the oxidation reaction proceeds, allowing to detect Cr+ as CrO+. Cr+ + O2 → CrO+ + O ∆Hr = 1.36eV
Interfering SO+ is removed via chain reaction given below.
SO+ + O2 → SO2+ +O ∆Hr = 1.48eV
SO2+ + O2 → SO2 + O2
+ ∆Hr = - 0.25eV
Reaction cell method for Cr in S matrix
53
Titanium ChromiumCell gas NH3 O2Q1/ Q2 48/114 52/68
Analyte ion TiNH(NH3)3+ CrO+
BEC in 10% H2SO4 (ppt) 2.0 4.0DL in 10% H2SO4 (ppt) 2.3 2.2
Ti Cr
Ti calibration plot in 10% H2SO4 Cr calibration plot in 10% H2SO4
Application #3: Ti and Cr measurement in H2SO4
Ti and Cr BEC in 10% H2SO4
Conclusions With it’s unique HMI and ORS3 capabilities, the Agilent
7700 ICP-MS is well suited for analysing the most challenging aqueous matrices in the Environmental Industry
Both the 7700 ICP-MS and the 8800 QQQ can also be used to determine metals directly in Petrochemicals.
MS/MS uniquely preserves isotopic information when using mass transitions
MS/MS offers truly unique capability and unprecedented control over cell-based reaction chemistry
July 10, 201354
Page 55