Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
IEEE STNP Task ForceCorrosion Effects on Subsurface Transformers &
Network Protectors
Introduction to Field Corrosion Measurements
Will Elliott – General Electric, Shreveport LA
Recall:
Galvanic Potential of Metals
• Galvanic (voltage) potentials exist between dissimilar
metals in an electrolyte (“water”)
• Potentials listed in Galvanic Series shown are versus a
common reference electrode
• The corrosiveness of the electrolyte will vary with
location
• Each location or generalized area would have its own
galvanic series
• Reference Galvanic Series shown in ASTM G82-98 in
flowing seawater
Corrosion PotentialTypes of Field Measurements
Corrosion Potential• Corrosion Potential measurement is a DC voltage measurement of
the galvanic potential between a metal sample and a reference electrode
• ASTM C876-15 gives a similar procedure for testing steel rebar in concrete
• Common connected to porous electrode.
• Also uses electrolyte solution, sponge, and porous electrode(not applicable for subsurface applications electrolyte already present)
• Positive connected to exposed section of rebar imbedded in concrete
• Basic Requirements
• Voltmeter (high impedance, mV range)
• Electrical Equipment Enclosure
• Anode(s), if present
• Reference electrode
• Subsurface vault (or manhole) with standing water
Corrosion Potential• Key parameters to document & measure:
• Materials:
• Equipment enclosure material
• Anode material
• Reference electrode material
• Use a consistent electrode material
• Water (electrolyte) temperature
• Water depth in vault
• Depth of electrode in water
• Measured voltage
• Record steady-state meter readings
Corrosion Potential• Reference Electrode:
• Copper/Copper Sulfate electrode
• https://www.certifiedmtp.com/m-c-miller-ionx-
submersible-copper-copper-sulfate-electrode/
• Electrode should be designed for immersion in
liquid electrolytes
• https://www.mcmiller.com/assets/site/docs/Articles/mp
%20article%20sub.pdf
• Equipment Electrode(s):
• Connect to equipment ground
• The electrolyte between locations should be
the variable!
Electrolyte
Corrosion Potential• Development of Galvanic Series &
Corrosion Risk Map
• Measure half-cell potential from equipment
ground to reference electrode
• If present, measure half-cell potential from
anode to reference electrode
• Depending on resolution and consistency of
the data, a map can be generated based on
the scale of data
Anode Potential Measurement
Equipment Potential Measurement
Corrosion Potential• Possible Data Sets:
• Number of measurements in a vault
• Each corner or side of vault/manhole?
• Fixed spacing between measurements?
• Average data or map results
• Periodic (sample) measurements
• Perform on a fixed schedule
• Discrete data points
• Easier & cheaper but low resolution data
• Continuous measurements
• Continuous data measures dynamic changes
• Provides better data for changing
temperatures, water levels, etc.
• More complicated & expensive, but high
resolution data