IEC SC62A MT 23Liaison Report to C63® SC84 May 2011
Jeffrey L. SilberbergFDA/CDRH
C63 SC8 meeting 2011-05-04 Cedar Park, TX 2
IEC SC62A MT23 Liaison Report MT23 met February 2011 in Carlsbad, CA
Prepared second CD of IEC 60601-1-2 Ed. 4 for circulation
IEC 60601-1-2 Ed. 4 circulated 2011-03-04 as 62A/746/CD Comment period closes 2011-06-10 US SC62A TAG meeting 2011-05-13 to
prepare US comments
C63 SC8 meeting 2011-05-04 Cedar Park, TX 3
Changes made from previous draft Performance (EMC) requirements were
moved to an informative annex Several of the proposed new test methods
were deleted Low-frequency magnetic field emissions Low-frequency magnetic field immunity* Close-distance RF immunity*
* Under consideration
C63 SC8 meeting 2011-05-04 Cedar Park, TX 4
Changes made from previous draft (cont’d) Several requirements were deleted
Harmonics and flicker emissions for equipment rated at > 16 A
Harmonics and interharmonics immunity Patient cable conducted emissions test is
now informative Deleted requirements and published
standards listed as things that should be considered in the Risk Management process
C63 SC8 meeting 2011-05-04 Cedar Park, TX 5
Changes made from previous draft (cont’d) Some general immunity test levels were
lowered Radiated RF Home and Transport ESD
Higher immunity test levels in frequency bands where sources are ISM Amateur radio Portable communications equipment
C63 SC8 meeting 2011-05-04 Cedar Park, TX 6
Changes made from previous draft (cont’d) The close distance RF test was replaced with
an interim test using IEC 61000-4-3 at 30 V/m to approximate portable transmitters Recommended separation distances, but
smaller distances Warning symbol required if you don’t do the
test
C63 SC8 meeting 2011-05-04 Cedar Park, TX 7
Changes made from previous draft (cont’d) An MT23 member is writing proposals for
SC77A for circulation as an NWIP Proximity RF immunity Proximity magnetic field immunity
Normative references are now dated unless there is a reason otherwise
PORT is now a defined term, figure “Clinic” environment renamed to “Small clinic”
Categories: Hospital, Small clinic, Home, Transport, Special
C63 SC8 meeting 2011-05-04 Cedar Park, TX 8
Example test levels for Enclosure port
Phenomenon
Immunity test levels
Hospital Small clinic Home Transport
Electrostatic discharge
± 8 kV contact± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
EM Field a) 3 V/m f) g)
80 MHz – 1000 MHz, 1,4 GHz - 2,7 GHz and 5 GHz – 6 GHz b)
80% AM @ 1 kHz c)
3 V/m f) g)
80 MHz – 1000 MHz, 1,4 GHz - 2,7 GHz and 5 GHz – 6 GHz b)
80% AM @ 1 kHz c)
10 V/m f)
80 MHz – 1000 MHz b)
3 V/m 1,4 GHz -
2,7 GHz and 5 GHz – 6 GHz
80% AM @ 1 kHz c)
20 V/m f) 80 MHz – 1000
MHz b)
10 V/m f)
1,4 GHz - 2,7 GHz
3 V/m f)
5 GHz – 6 GHz)80% AM @ 1 kHz
c)
C63 SC8 meeting 2011-05-04 Cedar Park, TX 9
MT23 Next steps
Finish (revising) responses to 460 comments from first CD
Respond to 217 comments on questionnaire (62A/700/Q)
MT23 meeting scheduled for August 2011 To respond to comments on 62A/746/CD and
prepare CDV Probably in White Oak / Silver Spring, MD Goal for CDV: November 2011